biosan BS-010526 Assist Pipettes with Volume Lock Instruction Manual
- September 27, 2024
- bioSan
Table of Contents
- biosan BS-010526 Assist Pipettes with Volume Lock
- Product Usage Instructions
- . Product Description
- Packing
- Pipette Design
- Safety Recommendations
- Specifications
- Pipette Operation
- Operating Instructions
- Aspiration and Dispensing Instructions
- Checking Pipetting Accuracy Parameters and Pipette Recalibration
- Pipette Maintenance
- Troubleshooting
- Spare Parts
- Limited Warranty
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
biosan BS-010526 Assist Pipettes with Volume Lock
Specifications
- Product Name: Biosan Assist Series Pipettes with Volume Lock
- Available Volumes: Single-channel pipettes from 0.1 µL to 10,000 µL; 8-channel and 12-channel pipettes for microplate filling
Product Description
The Biosan Assist VL single-channel pipette is a volumetric instrument designed to measure and transfer liquids precisely and safely. It is available in volumes from 0.1 µL to 10,000 µL. The Biosan Assist VL 8-channel and 12-channel pipettes are designed for microplate filling. The pipettes enable precise and simultaneous delivery of preset volumes of liquid in different ranges.
Single-channel Pipettes
Cat. No. | Volume Range (µL) |
---|---|
BS-010526 | 0.1 – 2 |
8-channel and 12-channel Pipettes
Model | Volume Range (µL) | Color Code |
---|---|---|
8-10 | 0.5 – 10 | Red |
Packing
The pipettes come with a quick start guide, quality control certificate,
calibration key, pipette hanger, identification labels, lubricant (for single-
channel pipette), ejector cap, filters, and color identification rings.
Pipette Design
-
* Pipettes operate using an air-cushion system.
- The volume range is displayed on the pushbutton and tip ejector bar.
Product Usage Instructions
-
Preparation
Ensure the pipette is clean and calibrated before use. Attach the disposable tip securely. -
Volume Adjustment
Use the volume adjustment knob to set the desired volume. Lock the volume setting in place using the locking ring. -
Liquid Transfer
Aspirate the liquid by pressing the pushbutton gently. Make sure the liquid enters the disposable tip, not the shaft or plunger. -
Ejection of Tips (Multi-channel Pipettes)
To eject tips in multi-channel pipettes, press the ejector pushbutton sequentially to release tips one by one.
FAQ
- Q: When should I replace the filter in the 5,000 L and 10,000 L pipettes?
A: The filter should be replaced if it becomes wet or if there is a risk of contamination during liquid aspiration and dispensing.
Medical-Biological
Research & Technologies
Assist Series Pipettes with Volume Lock
Instruction Manual
. Product Description
The Biosan Assist VL single-channel pipette is a volumetric instrument designed to measure and transfer liquids precisely and safely. It is available in volumes from 0.1 μL to 10,000 μL.
The Biosan Assist VL 8-channel and 12-channel pipettes are designed for microplate filling. The pipettes enable precise and simultaneous delivery of 8 or 12 preset volumes of liquid. These pipettes are available in four-volume ranges: 0.5-10 μL, 5-50 μL, 20-200 μL, and 50-300 μL.
Nominal Volume (µL)|
Cat. No.
| Range of Pipette Volumes (µL)|
Color Code
---|---|---|---
Single-channel Pipettes
2| BS-010526| 0.1 – 2|
| | | Red
10| BS-010527| 0.5 – 10|
20| BS-010528| 2 – 20|
100| BS-010530| 10 – 100| Yellow
200| BS-010531| 20 – 200|
1,000| BS-010532| 100 – 1,000| Blue
5,000| BS-010533| 500 – 5,000|
| | | White
10,000| BS-010534| 1,000 – 10,000|
8-channel and 12-channel Pipettes
8-10
12-10
| BS-010535 BS-010539|
0.5 – 10
|
–
8-50
12-50
| BS-010536 BS-010540|
5 – 50
|
| | | –
8-200
12-200
| BS-010537 BS-010541| 20 – 200|
8-300
12-300
| BS-010538 BS-010542| 50 – 300|
–
Biosan pipettes operate using an air-cushion (i.e., the aspirated liquid does not come into contact with the shaft or plunger of the pipette). The liquid is drawn into the disposable tip attached to the pipette.
Packing
The pipettes are delivered with the following:
Description | Qty/Pk |
---|---|
Quick start guide | 1 |
Quality Control certificate | 1 |
Calibration key | 1 |
Pipette hanger | 1 |
Identification labels | 1 |
Lubricant (for single-channel pipette) | 1 |
Ejector cap (for pipette models APL2, APL10, APL5000, APL10000) | 1 |
Filters (for pipette models APL5000, APL10000) | 3 |
Color identification rings (for pipette models APL2 – APL1000) | 6 |
Filters
The 5,000 μL and 10,000 μL pipettes include a replaceable filter, fitted at
the bottom of the shaft which prevents aspirated liquid from entering the
shaft and contaminating the inner surface and plunger. Use of the filter is
particularly important when aspirating and dispensing large volumes
of liquid. The filter should be replaced if it becomes wet.
Pipette Design
Model Identification
The volume range of the pipette is shown on the pushbutton and on the tip
ejector bar in multi-channel pipettes.
Single- and Multi-channel Pipettes
- A. Pipetting pushbutton: Consists of 2 parts: the pushbutton (A1) and the knob (A2).
- B1. Volume adjustment knob: Used to set the volume.
- B2. Locking Ring: Allows the volume setting to be locked by pushing the ring upwards.
- C. Shaft: Manufactured using high quality plastic, ensuring high chemical and mechanical strength.
- D. Tip ejector: For multi-channel pipettes, the tips are ejected sequentially, reducing the required force.
- F. Shaft nut
- N. Ejector pushbutton
- P. Color identification ring
- R. Multi-channel manifold: Contains a set of plungers and a set of flexible-suspension shafts, which function to reduce the force required to attach the tip.
Safety Recommendations
Long-term use of the pipette relies on correct method of use. Please read and
follow the instructions for use carefully.
Symbols used:
Danger, risk of injury.
Risk of damage to the pipette or errors in pipetting.
NOTE:
- The pipette is designed for the transfer of liquids only using the tip. Do not aspirate liquids without the tip attached. The aspirated liquid should not enter the pipette, as it may cause damage.
- Single-use tips reduce the risk of contamination of samples.
- Keep the pipette clean, avoiding the use of abrasive or corrosive cleaning agents (e.g., acetone).
- Keep the pipette upright when there is liquid in the tip.
- Only using the pipette in accordance with the manufacturer’s instructions ensures the correct pipette parameters are maintained.
- After replacing the plunger or the shaft, the pipette should be calibrated.
- In the case of incorrect operation, the device should be cleaned in accordance with the Instructions for Use or transferred to a service point.
- Ambient operating temperature is +5°C to 45°C.
- Ambient storage conditions (in the original packaging during transport and short storage) is -25°C to 55°C.
When working with the pipette:
- Follow general work safety regulations regarding hazards related to work in the laboratory.
- Take special care when pipetting aggressive substances.
- Use appropriate protective attire (e.g., clothing, goggles, and gloves).
- Avoid pointing the pipette at yourself or others during use.
- Only use parts and accessories recommended by the manufacturer.
Specifications
The pipette is a high quality instrument which offers excellent accuracy and precision. The accuracy and precision (repeatability) of the liquid volume depends on the quality of pipette tips used. The values for systematic error and random error shown in the table below were obtained using manufacturer non-filter pipette tips. Those tips are recommended for use to ensure compatibility, accuracy and precision when pipetting.
Nominal Volume (μL)|
Volume (μL)
| Systematic Error (%)| Random Error (%)|
Non-filter Tips (μL)
---|---|---|---|---
Single-channel Pipettes
| Min.| 0.1| ±40.0| ≤12.0|
2| | 0.2| ±12.0| ≤6.0| 10
| 1| ±2.7| ≤1.3
| Max.| 2| ±1.5| ≤0.7|
| Min.| 0.5| ±4.0| ≤2.8|
10| | 1.0| ±2.5| ≤1.8| 10
| 5.0| ±1.0| ≤0.6
| Max.| 10.0| ±0.5| ≤0.4|
| Min.| 2| ±3.0| ≤1.5|
20| | 10| ±1.0| ≤0.5| 200
| Max.| 20| ±0.8| ≤0.3|
| Min.| 10| ±1.6| ≤0.80|
100| | 50| ±0.8| ≤0.24| 200
| Max.| 100| ±0.8| ≤0.20|
| Min.| 20| ±1.2| ≤0.60|
200| | 100| ±0.8| ≤0.25| 200
| Max.| 200| ±0.6| ≤0.20|
| Min.| 100| ±1.6| ≤0.40|
1,000| | 500| ±0.7| ≤0.20| 1,000
| Max.| 1,000| ±0.6| ≤0.15|
| Min.| 500| ±1.2| ≤0.50|
5,000| | 1,000| ±0.6| ≤0.25| 5,000
| 2,500| ±0.6| ≤0.20
| Max.| 5,000| ±0.5| ≤0.15|
| Min.| 1,000| ±2.5| ≤0.6|
10,000| | 5,000| ±0.8| ≤0.3| 10,000
| Max.| 10,000| ±0.5| ≤0.2|
Nominal Volume (μL)|
Volume (μL)
| Systematic Error (%)| Random Error (%)|
Non-filter Tips (μL)
---|---|---|---|---
8-channel and 12-channel Pipettes
| Min.| 0.5| ±10.0| ≤8.0|
8 – 10
12 – 10
| | 1| ±8.0| ≤6.0| 10
| 5| ±4.0| ≤2.0
| Max.| 10| ±2.0| ≤1.2|
| Min.| 5| ±4.0| ≤2.5|
8 – 50
12 – 50
| | 25| ±3.0| ≤1.2| 200
| Max.| 50| ±1.6| ≤0.6|
| Min.| 20| ±3.0| ≤1.5|
8 – 200
12 – 200
| | 100| ±1.5| ≤0.8| 200
| Max.| 200| ±1.0| ≤0.6|
| Min.| 50| ±1.6| ≤1.5|
8 – 300
12 – 300
| | 150| ±1.2| ≤1.0| 300
| Max.| 300| ±1.0| ≤0.6|
The systematic error and random error were obtained gravimetrically using manufacturer tips performing at least 10 measurements of distilled water at a temperature of 20°C ± 1°C according to EN ISO 8655 standards.
The use of tips from other manufacturers or filter tips may result in
incorrect liquid aspiration and require pipette recalibration.
The design of the pipette enables the user to recalibrate it according to the
information presented in Section 9.
Pipette Operation
Attaching the Tips
- Attach the correct tip corresponding to the model number displayed on the pipette pushbutton (Section 5).
- Position the pipette vertically when attaching tips.
- Single-channel pipettes: push the pipette tip on firmly using a slight twisting motion to ensure an airtight seal.
- Multi-channel pipettes: press the pipette against the tips positioned in the rack box until the shaft retracts approximately 1.5mm into the manifold. The suspension system ensured even and airtight sealing of the pipette tips. The rocking motion does not have to be performed to seal the pipette tips tightly.
NOTE:
- Do not attach the tips with a rocking movement, as this may damage the shaft or plunger. Observe this rule particularly with single-channel pipettes of low volume range.
- Never draw liquids directly into the pipette without the tip attached.
Adjusting the tip ejector
The tip ejector can be adjusted by the user and can accommodate most types of
tip available on the market. When using narrow tubes, it may be necessary to
remove the tip ejector.
2-1,000 μL pipettes
- Tip ejector disassembly
- Press the tip ejector pushbutton.
- Turn the metal tip ejector 1/3 turn counter-clockwise.
- Slide the ejector from the plastic arbor.
- Tip ejector assembly
- Press the tip ejector pushbutton.
- Line the metal tip ejector up with the plastic arbor.
- Turn the metal tip ejector clockwise until it has fully latched.
- Tip ejector adjustment
- Remove the tip ejector pushbutton.
- To increase the ejector length turn the plastic bush counter-clockwise.
- To decrease the ejector length turn the plastic bush clockwise.
5,000 and 10,000 μL pipettes
-
Tip ejector disassembly
- Remove the tip ejector pushbutton.
- Using a screwdriver, turn the metal arbor counter-clockwise to unscrew the ejector from the arbor.
- Release the ejector.
-
Tip ejector assembly
- Remove the tip ejector pushbutton.
- Using a screwdriver, turn the metal arbor clockwise to secure the ejector and adjust its length.
NOTE: Leave a space of at least 1 mm between the shaft and tip ejector collar.
-
Tip ejector adjustment
- Remove the ejector pushbutton.
- Use the screwdriver to increase or decrease the ejector length.
- Following ejector assembly or adjustment, ensure the ejector pushbutton is relocated.
If the tip ejector cannot be adjusted sufficiently, or if the diameter of the
ejector is insufficient to eject the tip, it may be necessary to attach the
ejector cap “M” to the ejector.
In 2 and 10 PL pipettes place the cap, supplied with the pipette, on the
bottom of the pipette shaft and slide the cap upwards to the bottom of the tip
ejector.
Volume Setting
- To enable volume selection, set the locking ring to the lower position.
- Aspiration volume can be set by either turning the pipetting pushbutton knob, or the adjustment knob. The volume setting is displayed as a three digit counter, which should be read from top to bottom. The smallest volume increment is printed on the bottom counter drum.
- After the volume has been set, set the locking ring to the upper position to avoid inadvertent changing of volume.
Examples of counter indications
The volume display shown by the counter has three digits. These should be read
from top to bottom. The lowest counter dial shows a scale which allows volume
setting in the minimum graduation range.
Volume Ranges (µL)| Counter Readings|
Set Volume
| Increment (µL)
---|---|---|---
Single-channel Pipettes
0.1 – 2
| 1
2
5
|
1.25 µL
|
0.002
| | |
| | |
0.5 – 10
| 0
7
5
|
7.5 µL
|
0.02
| | |
| | |
2 – 20
| 1
2
5
|
12.5 µL
|
0.02
| | |
| | |
10 – 100
| 0
7
5
|
75 µL
|
0.2
| | |
| | |
20 – 200
| 1
2
5
|
125 µL
|
0.2
| | |
| | |
100 – 1,000
| 0
7
5
|
0.75 mL
|
2
| | |
| | |
500 – 5,000
| 1
2
5|
1.25 mL
|
10
| | |
| | |
1,000 – 10,000
| 0
7
5|
7.5 mL
|
20
| | |
8-channel and 12-channel Pipettes
0.5 – 10
| 0
3
5|
3.5 µL
|
0.02
| | |
| | |
5 – 50
| 0
6
5|
6.5 µL
|
0.10
| | |
| | |
20 – 200
| 0
8
5| 85.0 µL|
0.20
| | |
| | |
50 – 300| 2
5
0| 250.0 µL|
1.0
For maximum accuracy, the set volume must be approached from a higher value by decreasing counter readings. Before reaching the desired value, reduce the speed of turning the adjustment knob to avoid inadvertently exceeding the desired value.
Operating Instructions
Observing the following recommendations will ensure maximum possible accuracy and precision of liquid sampling.
- During operation, the volume setting should be locked, with black adjustment knob in the lower position.
- Ensure smooth and slow operation of the pipette.
- Immersion of the tip into the sample liquid should be kept to a minimum depth, which should remain constant during aspiration. The recommended immersion depths are given in the table below:
Model Volume Range (µL) | Immersion Depth (mm) |
---|---|
0.1-1 | ≤1 |
1-100 | 2-3 |
101-1,000 | 2-4 |
5,000 | 3-6 |
10,000 | 5-7 |
- The pipette should be held in a vertical position.
- The pipette tip should be changed whenever the volume setting is altered, and when a different liquid is to be aspirated.
- The pipette tip should be changed if a droplet remains on the end of the tip from the previous pipetting operation.
- Each new pipette tip should be pre-rinsed with the liquid to be pipetted.
- Liquid should never enter the pipette shaft. To prevent this:
- Press and release the pushbutton slowly and smoothly.
- Never turn the pipette upside down.
- Never lay the pipette on its side when there is liquid in the tip.
- Never force the volume setting beyond the recommended limits.
- When pipetting liquids with a different temperature from the ambient temperature, it is recommended to pre-rinse the tip several times prior to use.
- Do not pipette liquids with temperatures above 70°C.
NOTE: When pipetting acids or corrosive solutions which emit vapors, it is recommended to disassemble the shaft and rinse the plunger and O-ring with distilled water after finishing the pipetting operation.
Aspiration and Dispensing Instructions
**Aspirating Liquid**
-
Press the pushbutton to the first stop point. Holding the pipette vertically, immerse the pipette tip into the sample liquid to the recommended depth (for recommended values see Section 7). If the pipette tip is not immersed to the recommended depth or if the pipetting pushbutton is rapidly released, air may enter the pipette tip.
-
Release the pipetting pushbutton slowly and smoothly to aspirate the sample. Wait one second and then withdraw the pipette tip from the liquid.
CAUTION: Do not touch the used tip.
Dispensing Liquid
-
Place the end of the pipette tip against the inside wall of the vessel at an angle of 10° to 40°. Press the pushbutton smoothly to the first stop. Wait one second.
-
Press the pushbutton to the second stop to expel any remaining liquid. While keeping the pushbutton depressed, remove the pipette from the vessel by drawing the pipette tip against the inside surface of the vessel.
-
Release the pushbutton to its starting position.
-
Eject the pipette tip by pressing the tip ejector pushbutton.
Remember to change the pipette tip whenever a different type of liquid is to
be sampled.
Aspirating High-density Liquids
When pipetting liquids of higher viscosity or lower surface tension than water
(e.g., sera or organic solvents), a film of liquid may be formed on the inside
of the pipette tip which may produce erroneous results. As the film remains
relatively constant in successive pipetting operations with the same tip, this
error can be eliminated by pre-rinsing the tip and allowing a film to form
before transferring the first sample.
This is achieved by aspirating a sample and dispensing it back into the same
vessel. Allowing a film to form prior to sampling ensures optimal accuracy and
repeatability. This pre-rinsing operation should be repeated when the volume
to be aspirated is changed or when a new pipette tip is used.
NOTE : Normally the degree of error resulting from viscous liquids is
negligible if pipetting is performed slowly and carefully, however can be
minimized further by holding the pipette tip in position for at least 2
seconds after aspiration to allow the liquid time to react to the change in
pressure before it is dispensed.
- If the above method does not result in accurate values, recalibrate the pipette in accordance with Section 9.
- It is recommended to record recalibration and correction values, in order to facilitate reverse calibration to a standard liquid.
Filters
The 5,000 μL and 10,000 μL pipettes include a replaceable filter (L), fitted
at the bottom of the shaft which prevents aspirated liquid from entering the
shaft and contaminating the inner surface and plunger. Use of the filter is
particularly important when aspirating and dispensing large volumes of liquid.
The filter should be replaced if it becomes wet.
Checking Pipetting Accuracy Parameters and Pipette Recalibration
The pipettes have been factory-calibrated using gravimetric methods with manufacturer pipette tips and distilled water, in accordance with ISO 8655 guidelines for the maximum (nominal) liquid volume drawn by the pipette and for 10% of the maximum or minimum liquid volume according to the values given (Section 5).
The pipettes are designed to enable recalibration and adaptation to different
pipetting techniques and liquid properties (e.g., temperature, density, and
viscosity).
Periodic checks of the operation of the pipette are recommended at least once
per year. Frequency of checks should be increased depending on workload,
sterilization or autoclave processes, and frequency of replacement of
component parts.
If during pipette operation the systematic error (the difference between the real aspirated volume and the preset volume) exceeds the permissible value given in the table in Section 5, pipette recalibration should be carried out.
Recalibration of the pipette involves volume adjustment based on the value obtained by weighing. Recalibration of the pipette is performed only for one liquid volume drawn by the pipette.
The recommended volume for recalibration is 10% of the maximum (nominal) value or minimum volume depending on which of these values is greater. See table on page 16 for more details.
Parameters for Checking the Pipetting Accuracy
The pipetting accuracy is influenced by factors such as: tips used,
characteristics of the pipetted liquid (density, viscosity), and operating
conditions (ambient temperature, pressure).
To determine the systematic error of the pipette, the following conditions should be met:
- Ambient temperature and temperature of the pipette, pipette tips, and liquid should be within the range of 20°C to 25°C and stabilized during weighing within ±0.5°C.
- Measurements should be conducted using distilled water.
- Balance sensitivity should be suitable for the volume “V” to be measured.
Volume Checked (V, μL) | Balance Sensitivity (mg) |
---|---|
0.1≤V≤10 | 0.001 |
10≤V≤100 | 0.01 |
100≤V≤1,000 | 0.1 |
V>1,000 | 0.1 |
- When calculating the liquid volume aspirated by the pipette, the conversion factor (Z) [μL/mg] for distilled water or a liquid with comparable density should be taken into account. Sample values of conversion factors are given in the following table.
Temperature (°C)| | Pressure (kPa)|
---|---|---|---
95.0| 101.3| 105.0
20| 1.0028| 1.0029| 1.0029
21| 1.0030| 1.0031| 1.0031
22| 1.0032| 1.0033| 1.0033
23| 1.0034| 1.0035| 1.0036
24| 1.0037| 1.0038| 1.0038
25| 1.0039| 1.0040| 1.0040
See ISO 8655 for the full table of conversion factors (Z).
NOTE: Pipetting should be performed in accordance with guidance described
in Sections 7 and 8.
Checking the Pipetting Accuracy Parameters
- Set the preset volume depending on the pipette volume according to the following table.
Nominal volume (µL)
|
Preset volume (μL)
|
Permissible values (μL)
| Volume Change ΔV (μL) for a Turn of a Calibration
Key by
---|---|---|---
1 Turn| 1 Increment
Single-channel Pipettes
2| 0.2| 0.176 – 0.224| 0.06| 0.0025
10| 1| 0.975 – 1.025| 0.33| 0.0137
20| 2| 1.94 – 2.06| 0.63| 0.0262
100| 10| 9.84 – 10.16| 2.50| 0.104
200| 20| 19.76 – 20.24| 6.30| 0.262
1,000| 100| 98.4 – 101.6| 25.00| 1.04
5,000| 500| 494 – 506| 125.00| 5.2
10,000| 1,000| 975 – 1,025| 250.00| 10.4
8-channel and 12-channel Pipettes
10| 1| 0.92 – 1.08| 0.33| 0.0137
50| 5| 4.8 – 5.2| 1.67| 0.070
200| 20| 19.4 – 20.6| 6.30| 0.262
300| 50| 49.2 – 50.8| 10.00| 0.42
- Perform 10 aspirations, and calculate the average value in [mg].
- Calculate the volume in [μL] by multiplying the value in [mg] by the conversion factor Z [μL/mg].
If the average aspirated volume exceeds the permissible values of the range, the pipette should be recalibrated.
Pipette Recalibration
-
Remove the pipetting pushbutton.
CAUTION: The pipetting pushbutton consists of 2 parts: the knob and the pushbutton. After removal of the pushbutton, both parts are separated. -
Holding the volume setting knob to prevent rotation, insert the calibration key into the calibration screw.
-
Turn the key clockwise to reduce the aspirated volume, or counter-clockwise to increase the aspirated volume. Use the values given in the above table to precisely adjust the volume.
-
Remove the key and replace the pipetting pushbutton by first replacing the knob, then the pushbutton.
Determine the average aspirated volume. The average volume should be within the permissible range given in the table. If the volume exceeds the values stated, the recalibration procedure should be repeated.
When pipetting liquids with physical properties considerably different from
those of water, follow the guidance given in Section 5.
More information on the calibration procedure can be found at
www.biosan.lv.
Pipette Maintenance
Depending on the applications and intensity of use, the pipette requires periodic maintenance. The components exposed to corrosive vapors, such as shaft elements, should be regularly checked and cleaned.
- Do not use sharp tools for pipette maintenance. It may cause damage to the device and affect the user’s safety.
Cleaning
External surfaces of the pipette such as the pushbutton, ejector pushbutton,
handgrip, shaft nut, and adjustment knob may be cleaned using a cloth dampened
in isopropyl alcohol. The remaining parts removed from the pipette during
pipette disassembly may be washed with distilled water or isopropyl alcohol.
NOTE: Before using cleaning agents other than those recommended by the
manufacturer, check the compatibility charts and consider chemical resistance
of the following plastics which form components of the pipette: PP, PC, POM,
PA, PPS, PVDF.
Sterilization
Sterilization using an autoclave
The pipette can be sterilized in an autoclave at 121°C for 20 minutes.
Sterilization under other conditions may cause damage to the pipette. It is
recommended to:
-
Unscrew the shaft nut slightly in the APL2-APL1000 pipettes, and unscrew the shaft slightly in the
APL5000 and APL10000. After autoclaving these parts should be screwed tight again. -
Set the locking ring to the lower (unlocked) position prior to sterilization.
-
Sterilize the pipettes using an autoclave with an initial vacuum and drying cycle.
-
After sterilization, the pipette should be dried and cooled to room temperature.
-
The shaft of the APL5000 and APL10000 models should be autoclaved without the filter.
Precision and accuracy should not alter if the pipetting processes including autoclaving are carried out as described in this manual. If a change in accuracy occurs, it is recommended to:
- Check the calibration of the pipette after the first, third, and fifth autoclaving cycles and then after every 10 autoclaving cycles.
Ultra Violet (UV) Sterilization
The pipettes are UV resistant. The distance from the radiation source to the
exposed element of the pipette should be at least 50 cm. Prolonged or intense
UV exposure can cause discoloration of pipette parts but does not affect its
performance.
Troubleshooting
If a problem is encountered during pipette operation, use the following table to identify and eliminate the fault following the instructions provided. Replacement of parts should be required only occasionally and should not be required with normal pipette use.
Problem | Cause | Solution |
---|---|---|
Droplets of liquid remain in the pipette tip. | The pipette tip is emptied too | |
fast. | Decrease the speed |
of pressing the pipette pushbutton.
The pipette tip wettability has increased due to extensive use.| Replace the
tip with a new one.
| The pipette tip immersion depth is too shallow.| Immerse the tip to the
recommended depth according to the instructions.
Droplets of air appear in the liquid aspirated into the pipette tip.|
The pipette tip is incorrectly pressed onto the pipette shaft.|
Press the pipette tip firmly onto the pipette.
| The tip is damaged or worn out due to extensive use.| Replace the tip with a
new one.
The pipette incorrectly aspirates the liquid or the liquid drops out from the
tip.| The pipette tip is incorrectly pressed onto the pipette shaft.| Press
the pipette tip firmly onto the pipette.
Problem| Cause| Solution
---|---|---
The pipette incorrectly aspirates the liquid or the liquid drops out from the
tip.| The shaft surface is damaged or contaminated at the sealing site.| Clean
the shaft or replace it with a new one.
The pipette incorrectly aspirates the liquid or the liquid drops out from the
tip.| The plunger or the O-ring is damaged due to prolonged aspiration of
corrosive liquids.| Disassemble the shaft set; wash the shaft, the plunger and
the seal (Section 10: Cleaning). Replace the elements with new ones if
necessary. Apply a small amount of lubricant onto the plunger and reassemble
the set in the correct order.
The inside of the pipette is contaminated.
| The sealing elements are not sufficiently lubricated.
Uneven work of the pipetting set, the pipetting pushbutton gets blocked.| The
inside of the pipette is contaminated due to aspiration of corrosive
substances.| Unscrew the shaft set, wash the parts. Replace the elements with
new ones if necessary. Apply a small amount of lubricant on the plunger and
reassemble the set in the correct order.
The inside of the pipette is contaminated due to the liquid entering the
pipette.
The sealing elements are not sufficiently lubricated, e.g. after repeated
autoclaving procedures.
| Liquid with properties other than water (density, viscosity).| Calibrate the
pipette using the liquid which is to be pipetted.
Incorrect
aspiration.
| | Calibrate the pipette using the tips which are to be used for pipetting.
| Tips with a filter with increased flow resistance.
If the problem persists after carrying out the above steps, contact our
regional representative.
Before returning the pipette, please ensure the pipette is completely free of
any chemical, radioactive, or microbiological contamination which could pose a
threat during transport and repair.
Spare Parts
Contact our Customer Service to inquire about availability of spare parts.
Pipette model and name of the part required should be specified. The most
common parts are depicted below.
NOTE: The replacement of the plunger requires conducting the calibration
procedure according to Section 9.
Seal (J) used only in models APL2, APL5000 and APL10000 µL
Item | Description | Model | Cat. No. | Qty/Pk |
---|---|---|---|---|
B | Volume adjustment knob set | All | SP29327 | 1 |
C, I,
J*
| Shaft with O-ring and seal| APL2| SP19111| 1
APL10| SP19112| 1
| | APL20| SP19113| 1
| | APL100| SP19114| 1
| | APL200| SP19115| 1
| | APL1000| SP19116| 1
| | APL5000| SP19118| 1
| | APL10000| SP19119| 1
Item| Description| Model| Cat. No.| Qty/Pk
---|---|---|---|---
D| Tip ejector| APL2, APL10| SP19051| 1
| | APL20, APL100| SP19052| 1
| | APL200| SP19054| 1
| | APL1000| SP19055| 1
| | APL5000| SP19452| 1
| | APL10000| SP19453| 1
F| Shaft nut| APL2-APL1000| SP9759| 1
G| Plunger assembly| APL2| SP19381| 1
| | APL10| SP19382| 1
| | APL20| SP19383| 1
| | APL100| SP19384| 1
| | APL200| SP19385| 1
| | APL1000| SP19386| 1
| | APL5000| SP19388| 1
| | APL10000| SP19389| 1
I, J*| O-ring (and seal)| APL2| SP19101| 10
| | APL10| SP19102| 10
| | APL20| SP19103| 10
| | APL100| SP19104| 10
| | APL200| SP19105| 10
| | APL1000| SP19106| 10
| | APL5000| SP19108| 5
| | APL10000| SP19109| 5
K| Calibration key| All| SP29334| 1
L| Shaft filter| APL5000, APL10000| SP19476| 10
M| Tip ejector cap| APL2, APL10| SP19378| 3
| | APL5000| SP19466| 3
| | APL10000| SP19467| 3
P| Color indentification ring| APL2-APL1000| SP19630| 6
S| Shelf clip| All| SP19483| 1
Seal (J) used only in models APL2, APL5000 and APL10000 µL
Limited Warranty
Biosan SIA (Biosan) warrants that this product will be free from defects in material and workmanship for a period of three (3) years from date of purchase. BIOSAN DISCLAIMS ALL OTHER WARRANTIES WHETHER EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE. Biosan’s sole obligation shall be to repair or replace, at its option, any product or part thereof that proves defective in material or workmanship within the warranty period, provided the purchaser notifies Biosan of any such defect. Biosan is not liable for any incidental or consequential damages, commercial loss or any other damages from the use of this product.
This warranty is valid only if the product is used for its intended purpose and within the guidelines specified in the supplied instructions manual. This warranty does not cover damage caused by accident, neglect, misuse, improper service, natural forces or other causes not arising from defects in the original material workmanship. This warranty does not cover the O-ring, or shaft. Claims for transit damage should be filed with the transportation carrier.
In the event this product fails within the specified period of time because of
a defect in material or workmanship, contact Biosan Customer Service at
www.biosan.lv.
Biosan Customer Service will help arrange local service where available or
coordinate a return authorization number and shipping instructions. Products
received without proper authorization will be returned.
All items returned for service should be sent using prepaid postage in the original packaging or other suitable carton, padded to avoid damage. Biosan will not be responsible for damage incurred by improper packaging. Biosan may elect for onsite service for larger equipment.
Some states do not allow limitation on the length of implied warranties or exclusion or limitation of incidental or consequential damages. This warranty gives you specific legal rights. You may have other rights which vary from state to state. No individual may accept for, or on behalf of Biosan, any other obligation of liability, or extend the period of this warranty.
For your reference, make a note of the model number, serial number, date of purchase, and supplier here.
- Model No. ___ Serial No. ____
- Date Purchased _ Supplier ___
Warranty/Disclaimer: Unless otherwise specified, all products are for
research use only. Not intended for use in diagnostic or therapeutic
procedures. Biosan makes no claims regarding the performance of these products
for clinical or diagnostic applications
For additional product or technical information, visit
www.biosan.lv or contact your local sales office.
Biosan SIA
Ratsupites 7 k-2, Riga,
LV-1067, Latvia
Phone: +371 67426137 Fax: +371 67428101
Made in Poland
© 2024 Biosan SIA. All rights reserved. 5/24 BSL-5-2024-I
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>