biosan FTA Series Aspirator with Trap Flask Instruction Manual

June 9, 2024
bioSan

biosan FTA Series Aspirator with Trap Flask

biosan-FTA-Series-Aspirator-with-Trap-Flask-PRODUCT

The FTA-1 and FTA-2i aspirators with trapping flask are designed for liquid removal from micro-test tubes, ELISA plates, cell culture flasks, and other vessels during various laboratory techniques. The FTA-1 is a low-cost solution that has a 1 L glass trapping flask and is suitable for aspiration with a single standard 200 L tip and the 8-tipped MA-8 adapter. The FTA-2i comes with a 2 L polypropylene trapping flask and a liquid overflow sensor that stops collected material from damaging the filter. Aspiration is possible with single 200 and 1000 L tip using the MA-U adapter or with the handheld vacuum controller HAS-1 with different attachments such as an 8-channel adapter with stainless steel tips, 8channel adapter for 200 L tips, single channel adapter for 200 L tips and others.

Safety Precautions

Before using the equipment, make sure you have fully read and understood the user manual. Pay special attention to sections marked by the caution symbol. The unit and packaging feature CE marking, which affirms conformity with European health, safety, and environmental protection standards, and WEEE directive marking.

Product General Information

In order to prolong performance and efficiency of the device, it is not allowed to use aggressive organic compounds (e.g., phenol) and strong conjugated acids (e.g., guanidine) for aspiration. Please contact the manufacturer if you have any questions regarding the operation of the device.

Getting Started

Unpacking

Remove packing materials carefully and retain them for future shipment or storage of the unit. Examine the unit carefully for any damage incurred during transit. The warranty does not cover in-transit damage. Warranty covers only the units transported in the original package.

Complete Set

The FTA-1 package contents include:

  • Power button
  • Filter-to-bottle tube adapter
  • Tip holder microtubes
  • Filter-to-bottle tube adapter
  • Hydrophobic filter set (note the IN marking)
  • Lid with centre opening
  • Lid insert, with tip-to-bottle tube adapter
  • Glass trapping flask

Refer to Figure 1 for reference.

No package contents are listed for the FTA-2i. To use the device, refer to the specific section in the user manual that corresponds to your model and version.

If you have any feedback on our products or services, we would like to hear from you. Please send all feedback to:

Manufacturer

About this edition of user instructions

The current edition of the user instructions applies to the following models:

Model and name Version
FTA-1 , aspirator with trap flask V.5AW
FTA-2i , aspirator with trap flask V.1A01

Edition: 1.-5.01 – March of 2023

Safety precautions

Symbols used in these user instructions.

Caution

  • Make sure you have fully read and understood the present Manual before using the equipment. Please pay special attention to the sections marked by this symbol.

Icons used on the unit and packaging

  • CE marking, manufacturer affirms conformity with European health, safety, and environmental protection standards, see 12.1
  • WEEE directive marking, see 12.1
  • Polarity of the power connector
  • Equipment uses direct current
  • Microtube placements on FTA-1, see fig. 1/3
  • Pump exhaust on FTA-2i, connect to the means of air filtration, see 8.1

General safety

  • The protection provided can be ineffective if the operation of the appliance does not comply with the manufacturer’s requirements.
  • Save the unit from shocks and falling.
  • Store and transport the unit as described in section Storage and transportation.
  • Use only original parts and accessories, provided by manufacturer for this product.
  • Before using any cleaning or decontamination methods except those recommended by the manufacturer, check with the manufacturer that the proposed method will not damage the equipment.
  • Do not make modifications in design of the unit.

Electrical safety

  • Connect only to the mains with voltage listed on the serial number label.
  • Use only the external power supply provided with this product.
  • Ensure that the power plug is easily accessible during use.
  • Disconnect the unit from the mains before moving.
  • If liquid penetrates into the unit, disconnect it from the mains and have it checked by a repair and maintenance technician.
  • Do not operate the unit in premises where condensation can form. Operating conditions of the unit are defined in section Specifications.

During operation

  • Do not operate the unit in environments with aggressive or explosive chemical mixtures. Please contact manufacturer for possible operation of the unit in specific atmospheres.
  • Do not operate the unit if it is faulty or has been installed incorrectly.
  • Do not use outside laboratory rooms.
  • Do not use the unit for aggressive organic compounds, chlorinated solvents and strong conjugated acids, see section Chemical compatibility.

Biological safety

  • The user is responsible to carry out appropriate decontamination if hazardous material spills on or penetrates into the equipment.
  • Empty the trapping flask when the liquid level reaches maximum. Take necessary precautions utilizing waste liquid in accordance with general laboratory standards.

General information

FTA-1 & FTA-2i aspirators with trapping flask have been designed for liquid removal from micro-test tubes (e.g., for supernatant removal during DNA/RNA extraction), ELISA plates (e.g., performing several repeated well washing cycles), cell culture flasks (cell media exchange and washing during cell growth) and other vessels during various laboratory techniques. Model FTA-1 has a 1 L glass trapping flask and is designed to be a low-cost solution. Aspiration is possible with a single standard 200 μL tip and the 8-tipped MA-8 adapter. Model FTA-2i comes with a 2 L polypropylene trapping flask and a liquid overflow sensor that stops collected material from damaging the filter. Aspiration is possible with single 200 and 1000 μL tip using the MA-U adapter or with the handheld vacuum controller HAS-1 with different attachments such as an 8-channel adapter with stainless steel tips, 8-channel adapter for 200 μL tips, single channel adapter for 200 μL tips and others.

Microbiologic suction filter eliminates risks of contamination caused by bacteria, viruses, and infected particles from samples. Such contamination can be distributed by a suction pump or central vacuum units. Microbiological suction filter is hydrophobic and has very high contamination removal bacterial efficiency – up to 99.99% of particles bigger than 0.027 micrometers (e.g., Hepatitis A, B and C, HIV, etc.). In order to prolong performance and efficiency of device it is not allowed to use aggressive organic compounds (e.g., phenol) and strong conjugated acids (e.g., guanidine) for aspiration. Please, contact manufacturer in case of any questions regarding operation of device.

Getting started

Unpacking

Remove packing materials carefully and retain them for future shipment or storage of the unit. Examine the unit carefully for any damage incurred during transit. The warranty does not cover in-transit damage. Warranty covers only the units transported in the original package.

Complete set

Package contents

  • FTA-1:
    • FTA-1, aspirator with trap flask ……………………………………………………………… 1 pce.
    • Hydrophobic filter set …………………………………………………………………………….. 1 set
    • Aspiration tips, 200 μL ………………………………………………………………………… 50 pcs.
    • Holder tubes (1.5 ml) …………………………………………………………………………. 10 pcs.
    • Spare hydrophobic filter ……………………………………………………………………….. 1 pce.
    • External power supply …………………………………………………………………………. 1 pce.
    • Operating instructions, certificate …………………………………………………………. 1 copy
    • MA-8, 8-channel aspiration tip with holder ……………………………….. 1 set, on request
  • FTA-2i:
    • FTA-2i, aspirator with flask trap ……………………………………………………………… 1 pce.
    • MA-U, universal adapter ………………………………………………………………………. 1 pce.
    • Hydrophobic filter set …………………………………………………………………………….. 1 set
    • Aspiration tips, 200 μL ………………………………………………………………………… 50 pcs.
    • Aspiration tips, 1000 μL ………………………………………………………………………. 10 pcs.
    • Holder tubes (2 ml) ……………………………………………………………………………… 5 pcs.
    • Spare hydrophobic filter ……………………………………………………………………….. 1 pce.
    • External power supply …………………………………………………………………………. 1 pce.
    • Operating instructions, certificate …………………………………………………………. 1 copy
    • Extended tubing (2 m), with fittings and MA-U adapter ………………. 1 set, on request
    • MA-8, 8-channel aspiration tip with holder ……………………………….. 1 set, on request
    • HAS-1, hand operator set with SH-6 stand ………………………………. 1 set, on request
FTA-1 setup
  • Place the unit base on a horizontal even working surface.
  • Install the trapping flask (fig. 1/8) on the unit.
  • Check that the insert (fig. 1/7) is in the lid (fig.1/6). Secure the lid onto the trapping flask. Insert the hydrophobic filter set (fig. 1/5) into the lid.

Caution: Position the filter with the IN marking pointing towards the unit.

  • Check the tube connections: filter to bottle (fig. 1/2 and 1/4), bottle to aspiration tip (fig. 1/7 and fig. 2).
  • Place the microtubes for aspiration tip into the holder (fig. 1/3).
  • Connect the external power supply unit into the socket at the rear side of the unit and position the unit for an easy access to the external power supply and the power switch.

Optional MA-8 aspiration tip installation

  • Remove microtubes and aspiration tip from the holders on the base (fig. 1/3).
  • Place MA-8 storing holder into the unit base. Replace the aspiration tip with the MA-8 aspiration tip. Store the aspiration tip in the new holder.

FTA-2i setup

  • Place the unit on horizontal even working surface.
  • Install the trapping flask (fig. 3/2) on the unit.
  • Screw on the lid with the sensor needles and valves for tubing (fig. 3/1) to trapping flask.
  • Assemble the filter set (figure 6) and connect it to the port on the rear panel of the unit (fig. 7/1).

Caution: Position the filter with the IN marking pointing away from the unit (fig. 6/1).

  • Attach the fitting (fig. 4/1) of the tubing to the filter (fig. 4/2) to the valve with white ring on the lid.
  • Ensure that the filter set is tightly fixed in the slot (fig. 7/1).
  • Assemble the MA-U aspiration tip set (figure 5).biosan-FTA-Series-Aspirator-with-Trap-Flask-FIG-5
  • Attach the fitting (fig. 4/3) of the tubing for liquids (fig. 4/4 and 5/3) to the valve with the blue ring on the lid.
  • Connect the sensor cable (fig. 4/5) to the lid of the unit and to the port on the rear panel of the unit (fig. 7/4).
  • Place the microtubes for the aspiration tip into the holder (fig. 3/3).
  • Connect the external power supply unit to the socket at the rear side of the unit (fig. 7/5) and position the unit for an easy access to the external power supply and the power switch.

Optional MA-8 aspiration tip installation

  • Remove microtubes and aspiration tip from the holder (fig. 3/3). Remove the aspiration tip (fig. 5/1) from the MA-U adapter and connect the MA-8 aspiration tip (fig. 8/1A).

Optional HAS-1 hand operator installation

biosan-FTA-Series-Aspirator-with-Trap-Flask-FIG-9+

  • Unscrew the MA-U adapter from the aspiration tube (fig. 5/3) and connect the HAS-1 hand operator (fig. 8/5) by screwing it to the fitting.

Operation

  1. Connect the external power supply to the mains.
  2. Turn the Power switch (fig. 1/1 or 7/2) to position I(on). The pump starts working.
  3. (For model FTA-2i) Regulate the aspiration intensity with the regulator knob on the front panel of the unit (fig. 3/4). Clockwise turning direction increases power, counter clockwise – decreases. Current intensity is indicated by a light ring around the regulator knob.
  4. Touch the surface of the liquid with the aspiration tip (fig. 2, fig. 5/1 or MA-8) to remove the liquid.
    • Caution: During operation, do not allow the collected liquid to exceed the maximum level. Empty the trapping flask when the liquid level reaches a maximum allowed level, according to 5.6.
    • Note: In model FTA-2i, when collected liquid exceeds the maximum level, the sensor turns off the pump and produces sound and light warning signals.
  5. (For model FTA-2i) When using the optional HAS-1 hand regulator, touch the liquid with the aspiration tip, then press and hold the blue button on the regulator to collect.
  6. After finishing the operation, switch off the Power switch and disconnect the external power supply from the mains.
  7. Emptying the trapping flask. Turn off the unit. Remove the tubings and cables from the lid. Remove the closed flask from its place. Unscrew the lid and empty the trap-ping flask. Screw the lid back on and return the trapping flask on the unit. Replace the tubings and the cables.

Caution: Take necessary precautions in utilizing waste liquid in accordance with general laboratory standards. For information regarding autoclaving procedures, see 9.2.4.

Specifications

  1. Biosan is committed to a continuous program of improvement and reserves the right to alter the design and specifications of the equipment without additional notice.

Aspiration specifications

Model FTA-1 FTA-2i
Aspiration speed up to 1.7 mL/s up to 27.8 mL/s
Vacuum regulation Yes
Working vacuum pressure (at sea level) -500 mbar -200 … -800 mbar
Overflow protection sensor Yes, invasive
Overflow protection actions Pump stop, sound and light signals
Trapping flask 1 L, glass 2 L, polypropylene (PP)
Air filtration Hydrophobic microbiologic air filter 2200/021
Filter pore diameter 0.027 μm

General specifications

Model FTA-1 FTA-2i
Operating voltage 12 V=
Operating current 300 mA 1 A
Power consumption 3.6 W 10.8 W
External power supply input 100–240 V~, 50/60 Hz; output 12 V=
Dimensions (W×D×H) 186×213×327 mm 185x290x390 mm
Weight, accurate within ±10% 1.80 kg 1.85 kg

Workroom requirements

Workroom description Cold rooms and closed laboratory rooms
Temperature range +4 °C … +40 °C
Humidity requirements Maximum of 80% RH at 31 °C, decreasing linearly to 50%

RH at 40 °C. Non-condensing atmosphere.
Operating height, maximum| 2000 m ASL

  1. Eliminates risk of contamination from the trapping flask with bacteria, viruses, and infected particles.

Ordering information

Models and versions available

Model Version Catalogue number
FTA-1, aspirator with trap flask V.5AW BS-040108-AAG
FTA-2i, aspirator with trap flask V.1A01 BS-040120-A02
  1. To inquire about or order the optional accessories or the replacement parts, contact Biosan or your local Biosan representative.

Optional accessories

Description Catalogue number
МА-8 , 8-channel aspiration tip (fig. 8/1A) with holder (fig. 8/1B)

BS-040108-BK
HAS-1 , hand operator set (only for FTA-2i ). Set includes: Handheld vacuum controller with adapter (fig. 8/5),

8-channel adapter (with ejector) for 200 µL tips (fig. 8/6),

1-channel adapter with 125 mm stainless steel pin (fig. 8/7), 1-channel adapter with 40 mm stainless steel pin (fig. 8/8), 8-channel adapter with 35 mm stainless steel pin (fig. 8/9), 1-channel adapter (with ejector) for 200 µL tips (fig. 8/10C), Stand SH-6 for hand operator parts (fig. 8/10)

|

BS-040118-PK

Extended tubing (2 m), with fittings and MA-U adapter| BS-040120-DK

Replacement parts for FTA-1

Description Catalogue number
Hydrophobic filter BS-040108-S25
Aspiration tip, 200 µL, 1000 pcs. (fig. 8/2) BS-000001-S04
Aspiration tip, 8 channels, for MA-8 (fig. 8/1A) BS-040108-S24
Trapping flask, 1 L, glass, with lid, inserts and sealing ring BS-040108-S11
Tubing and fitting set BS-040108-S22

Replacement parts for FTA-2i

Description Catalogue number
MA-U , universal adapter for 200/1000 µL single use tips (fig. 8/4)

BS-040118-AK
Hydrophobic filter set| BS-040120-S10
Aspiration tip, 200 µL, 1000 pcs. (fig. 8/2)| BS-000001-S04
Aspiration tip, 1000 µL, 1000 pcs. (fig. 8/3)| BS-000001-S05
Aspiration tip, 8 channels, for MA-8 (fig. 8/1A)| BS-040108-S24
Trapping flask, 2 L, polypropylene, with lid, inserts and sealing ring| BS-040120-S11
Tubing and fitting set| BS-040120-S14
Sensor cable| BS-040120-S12

Chemical compatibility

Different aerosols and vapours from volatile and hazardous agents can have a damaging effect on the specific parts of the units. Please see below for FTA-1 and FTA-2i material compatibility tables 1 and 2 with different common laboratory chemicals. It is required to use the FTA-2i pump’s exhaust (fig. 7/3) by connecting it to the means of required filtration through a tube or, alternatively, by placing the unit in a fume hood when working with volatile and hazardous agents. FTA-1 does not have pump exhaust connector available; therefore, it is obligatory to place the unit in a fume hood. It is suggested to test the compatibility of related parts to a specific chemical prior to extensive use. Biosan does not warrant that the data in the tables is correct or comprehensive and that any material is fit for any purpose.

Resistance grades used in tables

  • 1 (Excellent) – at ambient temperature and pressure, the material should not be affected.
  • 2 (Good) – the material is slightly affected but not to the point of being unsuitable.
  • 3 (Fair) – a degree of reaction that is generally considered unsuitable and should not be used.
  • 4 (Severe) indicates that the material should not be used under any circumstances.

Table 1. FTA-1 chemical resistance chart



Parts

| ****


Material

| Formalde- hyde CH2O| Acetic acid C2H4O2| Hydrochloric acid HCl| Sodium hydroxide

NaOH

| Ethanol C2H5OH| Acetone (CH3)2CO
---|---|---|---|---|---|---|---
Tubing| Silicone| 2| 2| 3| 3| 1| 3
Bottle 1L| Glass| 1| 1| 1| 1| 1| 1
Lid| Polypropylene| 1| 1| 1| 1| 1| 1
Lid seal| NBR| 2| 3| 2| 2| 2| 4
Fittings| Polypropylene| 1| 1| 1| 1| 1| 1
Quick couplings| Ertacetal POM| 1| 1| 3| 1| 1| 1
Housing| ABS| 2| 4| 1| 1| 2| 4
Pump head housing| Polycarbonate| 1| 2| 4| 4| 2| 4
Pump head membrane| EPDM| 2| 1| 2| 1| 1| 1
Pump valve| EPDM| 2| 1| 2| 1| 1| 1
Filter membrane| PTFE| 1| 1| 1| 1| 1| 1

Table 2. FTA-2i chemical resistance chart



Parts

| ****


Material

| Formalde- hyde CH2O| Acetic acid C2H4O2| Hydrochloric acid HCl| Sodium hydroxide

NaOH

| Ethanol C2H5OH| Acetone (CH3)2CO
---|---|---|---|---|---|---|---
Tubing| Silicone| 2| 2| 3| 3| 1| 3
Bottle 2L| Polypropylene| 1| 1| 1| 1| 1| 1
Lid| Polypropylene| 1| 1| 1| 1| 1| 1
Lid seal| TPE| 1| 3| 1| 1| 1| 1
Level detection pins| Stainless steel| 3| 3| 3| 2| 1| 1
Fittings| Polypropylene| 1| 1| 1| 1| 1| 1
Quick couplings| POM| 1| 1| 3| 1| 1| 1
Housing| ABS| 2| 4| 1| 1| 2| 4
Pump head housing| Nylon| 4| 4| 4| 3| 1| 1
Pump head membrane| EPDM| 2| 1| 2| 1| 1| 1
Pump valve| EPDM| 2| 1| 2| 1| 1| 1
Filter membrane| PTFE| 1| 1| 1| 1| 1| 1

Care and maintenance

Service

  1. If the unit is disabled (e.g., no aspiration, no reaction to key presses, etc.) or requires maintenance, consult the troubleshooting table in 9.5. If the problem is not covered, disconnect the unit from the mains and contact Biosan or your local Biosan representative.
  2. All maintenance and repair operations (except listed below) must be performed only by qualified and specially trained personnel.
  3. Operating integrity check. If the unit follows the procedure described in section Op-eration, then no additional checks are required.

Cleaning and disinfection

  1. Use mild soap and water with a soft cloth or sponge for cleaning the exterior. Rinse remaining washing solution with distilled water. Wipe dry the excess water with clean, soft cloth or sponge.
  2. To disinfect the plastic and glass parts, use 75% ethanol or DNA/RNA removing solution (e.g., Biosan PDS-250). After disinfecting, wipe the surfaces dry.
  3. Silicone and fittings. Detach and disassemble the tubing. Use only distilled water, no less than 0.5 L, and air dry after rinsing.

Autoclaving procedures

| Lid, fittings, tubing| Flask| Hydrophobic filter, cables, base pump
---|---|---|---
FTA-1| 121°C and 1 atm., 15 min.| 121°C and 1 atm., 55 min.| Do not autoclave
FTA-2i| 121°C and 1 atm., 15 min.

Caution: Do not put a closed flask in the autoclave, as it might irreversibly deform. Completely disengage the thread on the cap.

Note: Hydrophobic filter and sensor cable are sensitive to moisture and are not autoclavable. Tubing is made of silicone and deteriorates after frequent autoclaving and should be replaced.

Tubing replacement

  1. Change the silicone tubing every 6 months.

Filter replacement

  1. Replace the filter as soon as it accumulates visible traces of dirt, or when the liquid is sucked into it. Maximum working time for the filter is 30 days. When working with infected materials, replace the filter immediately after finishing the procedure to prevent the spread of infection.
  2. Changing procedure. Remove the filter (fig. 1/5 or 6/1) by detaching it from adapters. Insert a new filter and reconnect the adapters.

Caution: Observe the orientation of the filter:

  • For model FTA-1, the IN marking should be facing towards the unit (figure 1).
  • For model FTA-2i, the IN marking should be facing away from the unit (figure 6).

Troubleshooting

Problem Possible reason Troubleshooting

Unit not working

| No power| Check external power supply and connector
( FTA-2i ) Trapping flask full| Sensor shut down the motor. Empty the trapping flask according to 5.6.
( FTA-2i ) Sensor not connected properly| Check sensor cable and attachments

Aspiration slow or absent

| ( FTA-2i ) Aspiration intensity set to small| Turn the regulator knob clockwise to in- crease aspiration.
Breach in the system| Check all the tubing and fittings. Tightly close the lid of the unit
Dirt in the tubing| Clean the tubing, see 9 .2.3.
Liquid in the filter| Replace the filter
( FTA-2i ) Sensor activated, but the flask is not full| Foam in the trapping flask activated the sensor| Do not use the unit with a foaming liquid or use an antifoaming agent.

Disposal

Disposal of the appliance requires special precautions and must be carried out at an appropriate disposal site, separate from normal household waste. To prevent pollution of the environment, all waste resulting from the disposal of the product must be collected and disposed of in the country of use, in accordance with the applicable requirements for the handling of electronic waste.

Storage and transportation

  1. Store and transport the unit in a horizontal position (see package label) at ambient temperatures between -20°C and +60°C and maximum relative humidity of 80%.
  2. After transportation or storage and before connecting it to the electric circuit, keep the unit under room temperature for 2-3 hrs.
  3. For extended storage, disassemble and clean according to 9.3. Empty the trap flask. The unit does not require special procedures.

Warranty

  1. The Manufacturer guarantees the compliance of the unit with the requirements of Specifications, provided the Customer follows the operation, storage and transportation instructions.
  2. The warranted service life of the unit from the date of its delivery to the Customer is 24 months. For an extended warranty, see 11.6.
  3. Warranty for the unit does not cover optional accessories or replacement parts listed in 7.2.
  4. Warranty covers only the units transported in the original package.
  5. If any manufacturing defects are discovered by the Customer, an unsatisfactory equipment report shall be compiled, certified and sent to the local distributor address. To obtain the claim form, visit Technical support page on our website at link below.
  6. Extended warranty.
    • For FTA-2i, the Premium class model, one year of extended warranty is available free of charge after registration, during 6 months from the date of sale. Online registration form can be found in section Warranty registration on our website at the link below.
    • or FTA-1, the Basic Plus class model, extended warranty is a paid service. Contact your local Biosan representative or our service department through the Technical support section on our website at the link below.
  7. Description of the classes of our products is available in the Product class description section on our website at the link below.biosan-FTA-Series-Aspirator-with-Trap-Flask-FIG-10
  8. The following information will be required in the event that warranty or post-warranty service comes necessary. Complete the table below and retain for your records.
Model Serial number Date of sale

FTA-1, FTA-2i ,

Aspirators with trap flask

| |

Production date

The production date is placed in the serial number, on the label of the unit. Serial number consists of 14 digits styled XXXXXXYYMMZZZZ, where XXXXXX is model code, YY and MM – year and month of production, ZZZZ – unit number.

EU Declaration of Conformity

Aspirators with trap flasks FTA-1 and FTA-2i are in conformity with the following relevant Union legislations:

LVD 2014/35/EU

| LVS EN 61010-1:2011 Safety requirements for electrical equipment for measurement, control, and laboratory use. General requirements.
---|---

EMC 2014/30/EU

| LVS EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use. EMC requirements. General requirements.

RoHS3 2015/863/EU

| Directive on the restriction of the use of certain hazardous substances in electrical and electronic equipment.
WEEE 2012/19/EU| Directive on waste electrical and electronic equipment.

  1. Declaration of Conformity is available for download on the page for the relevant model on our website by links below, in the Downloads section:

CONTACT

SIA Biosan

  • Ratsupites 7 k-2, Riga, LV-1067, Latvia
  • Phone: +371 67426137
  • Fax: +371 67428101
  • https://biosan.lv/

Edition: 1.-5.01 – March of 2023

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

bioSan User Manuals

Related Manuals