biosan 6003 S-BT Smart BioTherm Compact CO2 Incubator Instruction Manual
- June 6, 2024
- bioSan
Table of Contents
- biosan 6003 S-BT Smart BioTherm Compact CO2 Incubator
- Manufacturer
- About this edition of user instructions
- Safety precautions
- General information
- Application areas
- Getting started
- Operation
- Fault diagnostics
- Ordering information
- Care and maintenance
- Warranty
- EU Declaration of conformity
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
biosan 6003 S-BT Smart BioTherm Compact CO2 Incubator
If you have any feedback on our products or services, we would like to hear from you. Please send all feedback to:
Manufacturer
SIA Biosan
Ratsupites iela 7 k-2, Riga, LV-1067, Latvia
Phone: +371 674 261 37
Fax: +371 674 281 01
www.biosan.lv
Marketing: marketing@biosan.lv
Service: service@biosan.lv
About this edition of user instructions
The current edition of the user instructions applies to the following models and ver-sions of compact CO2 Incubator:
S-Bt Smart BioTherm V.1AW, V.1AE
Safety precautions
- Caution! Make sure you have fully read and understood the present Manual before using the equipment. Please pay special attention to sections marked by this symbol.
- Caution! Biohazard. Incubator operations should comply criteria of biosafety in microbiological and biomedical laboratories, according to working mate-rial biohazard level.
- Caution! Incubator is equipped with an audible signal to draw attention to sensor data and warnings. Please follow the information on display. Error and alarm descriptions can be found in the Fault Diagnostics section.
- Caution! The incubator contains a source of UV radiation. Do not start the unit without the cover of the UV recirculator. Exposure to UV light is harmful and can cause damage to unprotected eyes and skin.
General safety
- The protection provided can be ineffective if the operation of the appliance does not comply with the manufacturer’s requirements.
- Save the unit from shocks and falling.
- Do not use the unit if it has visible mechanical damage.
- Store and transport the unit at ambient temperatures between -20°C and +60°C and maximum relative humidity of 80%.
- After transportation or storage and before connecting it to the electric circuit, keep the unit under room temperature for 2-3 hrs.
- Before using any cleaning or decontamination methods except those recommended by the manufacturer, check with the manufacturer that the proposed method will not damage the equipment.
- Do not make modifications in design of the unit.
- When handling hazardous substances, follow the instructions contained in the safety data sheets for the individual substances that are used and observe the relevant ac-cident prevention regulations.
- Follow respective safety datasheet when installing and handling the CO2 reservoir.
Electrical safety
- Connect only to the mains with voltage corresponding to that on the serial number label.
- Do not plug the unit into an ungrounded power socket, and do not use an ungrounded extension lead.
- Ensure that the power plug is easily accessible during use.
- Disconnect the unit from the mains before moving.
- If liquid penetrates into the unit, disconnect it from the mains and have it checked by a repair and maintenance technician.
- Do not operate the unit in premises where condensation can form. Operating condi-tions of the unit are defined in the Specifications section.
During operation
- Do not operate the unit in environments with aggressive or explosive chemical mix-tures. Please contact manufacturer for possible operation of the unit in specific at-mospheres.
- Do not operate the unit if it is faulty or has been installed incorrectly.
- Do not use outside laboratory rooms.
Biological safety
- The user is responsible to carry out appropriate decontamination if hazardous mate-rial spills on or penetrates into the equipment.
- The user is responsible for decontamination of the unit before its decommissioning and utilization.
General information
S-Bt Smart BioTherm is designed for work in the areas of cell biology
(operations with animal cell cultures and tissues), molecular biology (DNA/RNA
reaction analysis, hy-bridization reactions), biotechnology (synthesis of
target proteins and other molecules) and immunology (synthesis of antibodies
and other proteins of immune system).
Unit provides a six-sided heating: the heating elements are located on the
walls and on the door, thus providing excellent uniform temperature
distribution, regardless of external factors, such as ambient temperature and
positioning of the device.
Built-in infrared CO2 sensor allows accurate control of the CO2 level. The
sensor makes measurement non-sensitive to changes in temperature and humidity
inside the incu-bator.
The chamber is made of stainless steel with smoothed seams to minimize
contami-nation and to facilitate cleaning.
S-Bt is equipped with a UV air recirculation system — 1 UV lamp and a fan are
mounted behind the rear wall, providing decontamination of the working volume.
A convenient access port is built in the wall of the incubator for easy output
of wire sensors or devices’ installed inside. The access port is heated
independently to prevent for-mation of condensate.
Unit is equipped with error tracing and alarm systems, which significantly
lower po-tential risks during operation.
Unit is equipped with a “black box” system that records temperature, humidity
and CO2 levels as well as statuses for door opening, UV lamp, fan and errors
to the inner memory.
Bluetooth connection to PC is available.
Application areas
- Cell biology: operations with animal cell cultures and tissues
- Molecular biology: DNA/RNA reaction analysis, hybridization reactions
- Biotechnology: synthesis of target proteins and other molecules
- Immunology: synthesis of antibodies and other proteins of immune system Features:
- Six-sided heating provides uniform distribution of the temperature inside the chamber
- Infrared CO2 sensor, non-sensitive to temperature and humidity changes
- UV recirculation system for decontamination cycles
- Bluetooth data transfer to PC
- “Black box” parameter logging system
- Error tracing and alarm system
- Separately heated sealable port for chamber access for cables.
Getting started
Unpacking. Remove packing materials carefully and retain them for future
shipment or storage of the unit. Examine the unit carefully for any damage
incurred during transit. The warranty does not cover in-transit damage.
Warranty covers only the units transported in the original package.
Caution! Due to the high weight of the unit, its unpacking and installing must
be
carried out by two persons.
Complete set. Package contents:
Standard set:
- S-Bt Smart Biotherm, Compact CO2 Incubator ………………………………………. 1 pce.
- Power cable ………………………………………………………………………………………. 1 pce.
- Hydrophobic filter ………………………………………………………………………………… 2 pcs.
- Shelves ……………………………………………………………………………………………… 3 pcs.
- Shelf rack …………………………………………………………………………………………… 1 pce.
- Water pan with lid ………………………………………………………………………………….. 1 set
- Access port seal (foamed PE) ……………………………………………………………… 10 pcs.
- Replacement humidity sensor ……………………………………………………………….. 1 pce.
- USB drive with control/logging software for PC ………………………………………… 1 pce.
- Calibration protocol …………………………………………………………………………….. 1 copy
- Operating instructions, declaration of conformity ……………………………………. 1 copy
Optional accessories:
- Bluetooth adapter ………………………………………………………………………….. on request
- Additional shelves …………………………………………………………………………. on request
- Incubator stacking device ……………………………………………………………….. on request
Setup.
- Position the incubator on a stable platform that can support the weight of the incubator and materials inside the incubator. Ensure access to an appropriate mains outlet.
- Connect the power cable to the socket on the rear panel of the unit. Check that power switches and power plug are easily accessible.
- Locate the unit away from doors and windows and heating and air conditioning ducts. Drafts or direct sunshine can affect thermal processes and cause condensation.
Before starting operations for the first time.
- Attach the hydrophobic filter to inlets on the top side of the rear panel. Pay attention that “IN” marking must be facing right side when viewing the incubator from the front.
- Use 75% alcohol solution or any other decontaminating agent to clean the inner surfaces of the incubator and the water pan with lid.
- Switch on the recirculator for half an hour in an empty incubator to decontaminate air in the incubator. To switch on the UV recirculator, see 5.9.2.1.
CO2 tank requirements and setup.
- This incubator is designed for CO2 gas only. Connecting any other gas can result in a hazardous condition. The incubator works with food grade CO2 or better rated. Do not use a tank with siphon tube, the incubator can be damaged by liquid CO2.
- A two-stage reductor installation is mandatory. Pressure in the output must not be higher than 1.0 bar. Follow gas supplier instructions on installing reductors and valves. Gas inlet on the incubator is a 3/8” barb fitting, marked Supply port. Secure the tube with appropriate clamps and check all connections for leaks. Gas flow should be set at 6 L/min.
- Caution! Pressure higher than 1.0 bar can damage the unit!
Operation
Connect the incubator to a grounded mains outlet. Turn the power switch on the
rear of the incubator to position I (on).
After initialization, display (fig. 1) shows:
- Top line: current date and time, relative humidity inside, UV lamp icon (when the lamp is on);
- Middle left: temperature panel, current temperature in larger font, set temperature in smaller;
- Middle right: CO2 control panel, current percentage in larger font, set percentage in smaller;
- Bottom line: Status bar, UV recirculator timer (when the lamp is on);
- Bottom: six buttons, Menu, t°C ▼, t°C ▲, CO2 ▼, CO2 ▲, Lock.
Installing shelves. Open the front door and the inner glass door. Position the shelves according to your specification.
Humidity delivery. Pull out the water pan located on the bottom of the inner
camera and fill with distilled water (~80% of volume). With the set
temperature of +37°C, incubator reaches 90% of relative humidity in 8-12
hours.
Note. When operating the incubator at 90% of relative humidity for longer
than 48 hours, humidity sensor values can have a measurement error.
Using access port. In order to use the access port, unscrew the cap of the access port located on the right side of the incubator. Insert necessary cable through the port and into the supplied PE seal. Install the seal into the access port. The access port is individually heated to compensate for temperature loss and condensate formation.
Setting temperature. Using t°C ▼ and t°C ▲ buttons, set the necessary temperature.
Setting CO2 levels. Using CO2 ▼ and CO2 ▲ buttons, set the necessary level of
CO2.
Caution! Please make sure that the CO2 tank is attached correctly, see
4.4.
Note. After reaching temperatures higher than 50°C, CO2 level measurement
accuracy does not conform to specifications.
Locking the buttons. Lock button locks the use of all the buttons under the display until Lock button is pressed again.
Incubator menu (fig. 2). Menu button opens incubator’s menu. Use Down and Up buttons to navigate and Enter button to select. While browsing in the menu, select the topmost Quit option in order to return to the previous level. Option GetID shows the serial ID and name of the incubator.
Timer (fig. 3). In this submenu user can set UV decontamination timer and Auto UV timer. UV Decontamination Timer. Timer parameters are set in HH:MM format. After setting the timer, UV lamp automatically switches on, main working screen displays an UV indicator (fig. 4, top right). After set time has passed, the UV lamp automatically switches off. A timer is seen on the main working screen (fig. 4, bottom right). To manually switch the lamp off, select the Turn Off UV option from the submenu. Auto UV Timer. Each time after closing the door, UV lamp automatically switches on. After time set under this menu option has passed or the door was opened, the UV lamp automatically switches off.
- Note. A turned on UV lamp can rise the temperature above the set point: the longer UV lamp stays on, the faster temperature jumps above the set point. Use this feature at your own discretion.
- Caution! UV recirculation should be used for decontamination cycle while setting up and after operation, see 9.3.
- Note. In case of manual UV lamp switch off, timer continues working despite the UV lamp is off.
Date & Time (fig. 5). To set date and time in this submenu, press Enter button to select the parameter, then Down and Up buttons to change the value and Enter button to confirm changes. Service (fig. 6). This submenu has available options Current Errors and Settings. 5.9.3.1 Current Errors (fig. 7). This option allows user to view and reset the list of active errors. See 6.3, for a list of possible errors. Settings (fig. 8). User can access the Alarm Settings and Auto Lock Delay options.
-
Alarm Settings (fig. 9). User can set the alarm settings based on values of temper-ature, CO2 level, and relative humidity. All alarms are working on the following princi-ple, temperature taken as an example:
If Tact < Tthr.low or Tact > Tthr.high for tdelay minutes, then an alarm will be triggered. Where Tact is the actual temperature inside the incubator, Tthr.low is the lower threshold limit value, Tthr.high is the higher threshold limit value and tdelay is the time delay in minutes. Time delay is used in order to avoid false triggering of the alarms. CO2 level and relative humidity alarms are operated using the same principle. -
Auto Lock Delay (fig. 10). After being idle for several minutes, the incubator locks the buttons (except the Lock button). This timer allows changing the idle time.
Active Alarms (fig. 11). This option allows user to view and reset the list of
active alarms. See 6.5, for a list of possible alarms.
Data Logger (fig. 12). Data logger features following options: View Plot T(t),
View Plot CO2%(t), View Plot RH%(t), Clear log, Logger on/off.
All View Plot options use following interface. Figure 13 shows the temperature
plot in relation to time.
- Use Left and Right buttons to move the plot along the x axis
- Use Up and Down buttons to move the plot along the y axis.
- Use <-xx-> button to choose the x axis navigational step. Available scales are 15 minutes, 1 hour, 1 day, 1 week and 1 month.
Clear Log button clears the logger’s data.
Logger On/Off button triggers the automatic logging of the data.
Memory write location indicator. When the data logger uses up all allocated
memory, it begins overwriting data from beginning. 100% of memory is used up
in approximately 1 month.
After stabilization of temperature and CO2 level, open the doors and load the materials in the incubator according to methodological recommendations.
Note.
If the door is left open for more than a minute, a short signal is sounded
every 20 seconds until the door is closed.
Incubator air sampling. Sampling can be performed through the capped opening marked as “air sample”, on the rear access panel.
To turn off the incubator, set the power switch into position О (off) and detach the mains plug from the mains outlet.
Operation recommendations.
- The incubator saves settings even when switched off.
- During decontamination treat every visible surface and door. When working with po-tentially pathogenic material, use more potent decontamination method. See 9.3.
Fault diagnostics
- The incubator is equipped with an alarm and error system. Active alarms and errors are appropriately indicated on the display. Alarms and errors are followed by sound signal that can be turned off by pressing any button.
- Error text can be found in Menu > Service > Settings > Current Errors, see 5.9.4. Choose Errors Reset and confirm to reset the error list.
- Error list can be found in the table below:
CO2 Sens. Err.| CO2 sensor error
---|---
Back| Back temperature sensor error
Bott.| Bottom temperature sensor error
Left| Left wall temperature sensor error
Right| Right wall temperature sensor error
Up| Ceiling temperature sensor error
Door| Door temperature sensor error
Chmb.| Temperature sensor in the incubator error
Perimeter T.Sens. Err.| Perimeter temperature sensor error
Recirc. Lamp Err.| UV lamp error
Communication Err.| Error connecting to power board
CO2.Sens.Comm.Err.| Error connecting to CO2 sensor board - Alarm text can be found in Menu > Active Alarms. Choose Alarms Reset and confirm to reset the alarm list.
- Alarms list can be found in the table below. Alarms are adjustable, see 5.9.3.2.
Temp. High Alarm|
Adjustable limit is exceeded.
---|---
Temp. Low Alarm
CO2 High Alarm
CO2 Low Alarm
Low Humidity
UV Lamp Hours > 9000h| Lamp reaches end of lifetime. Consider a replacement.
Specifications
The unit is designed for operation in cold rooms and closed laboratory rooms
at am-bient temperature from +4°C to +40°C in a non-condensing atmosphere and
maximum rel-ative humidity 80% for temperatures up to 31°C decreasing linearly
to 50% relative humidity at 40°C.
Biosan is committed to a continuous programme of improvement and reserves the
right to alter design and specifications of the equipment without additional
notice.
-
Chamber Material ……………………………………………………………………. Stainless Steel
-
Temperature setting range ……………………………………………………….+25°C … +60°C
-
Temperature control range ………………………………. 2.5°C above room temp…+60°C
-
Temperature resolution …………………………………………………………………………. 0.1°C
-
Temperature stability …………………………………………………………………………… ±0.1°C
-
Temperature uniformity @ +37°C …………………………………………………………. ±0.3°C
-
Working volume ……………………………………………………………………………….. 46 litres
-
Number of shelves ……………………………………………………………. 6 max. (3 provided)
-
Outer door ……………………………………………………………………. Heated, made of steel
-
Inner door ……………………………… Lockable glass with UV protection film, 4 mil clear
-
Relative humidity ……………………………………………………………………… >90% @ 37°C
-
Humidity delivery ………………………………………………………… water pan by ventilation
-
CO2 control range1 …………………………………………………………………. 0 – 20% ± 0.2%
-
CO2 sensor ……………………………………………………………………… Infrared CO2 sensor
-
Display ………………………………………………………………………………………………TFT 5’’
-
Recirculation system ………………………………………….. 1 × 6 W UV lamp (TUV G6T5)
-
Recirculation system radiation type ………………………………………………. UV (254 nm)
-
Average UV lamp life time ……………………………………………………………………. 9000 h
-
Data transfer …………………………………………………………………… Wireless, Bluetooth®
-
Internal data memory ………………………………………………1 month of nonstop logging
-
Access port ………………………………………………………………1 (ø 26 mm with PE seal)
-
Dimensions (W×D×H) …………………………………………………………. 500×560×550 mm
-
Inner chamber dimensions (W×D×H) …………………………………….. 350×330×390 mm
-
Alarms ……………………………………………………… Over-heating/cooling, CO2 leakage,
…………………………………………………………………………. Humidity loss, UV lamp error -
Electrical mains parameters (see 8.1) ………………………………… 230/115 V, 50/60 Hz
-
Weight2 …………………………………………………………………………………………….. 37.7 kg
Ordering information
Models and versions available:
Model | Version | Description | Catalogue number |
---|---|---|---|
S-Bt Smart Biotherm, Compact CO2 Incubator | V.1AW | 230 V, 50/60 Hz |
BS-010425-A01
V.1AE| 120 V, 50/60 Hz
To inquire about or order the optional accessories or the replacement parts,
contact Biosan or your local Biosan representative.
Optional accessories:
Optional accessories | Catalogue number |
---|---|
Shelf | BS-010425-AK |
Shelf rack | on request |
Bluetooth dongle and software for PC | BS-010425-BK |
Incubator stacking device | BS-010425-CK |
Replacement parts:
Replacement parts | Catalogue number |
---|---|
Hydrophobic filter with tubing | BS-010425-S01 |
Water pan and lid | BS-010425-S02 |
Access port seal, foamed PE | BS-010425-S03 |
Care and maintenance
- Service.
- If the unit is disabled (e.g., no heating or CO2 control, no reaction on key and switch presses, etc) or requires maintenance, disconnect the unit from the mains and con-tact Biosan or your local Biosan representative.
- All maintenance and repair operations (except listed below) must be performed only by qualified and specially trained personnel.
- Operating integrity check. If the unit follow the procedure described in section Opeation, then no additional checks are required.
- Cleaning and disinfection. Regular service schedule.
- After operation. Switch on the recirculator running for 1 to 3 hours in an empty incubator to decontaminate air in the incubator. To switch on the UV recirculator, see 5.9.1.1. After decontamination, slightly open the incubator doors – UV lamp has been known to produce reactive oxygen species in minor concentrations.
- Monthly (more frequently, if necessary):
* Rub the inner surfaces, water pan with lid and shelves of the incubator with 75%ethanol solution or other acceptably disinfectant such as sodium hypochlorite solu-tion, and allow them to dry.
* Inner side of the glass door is coated with 4 mil protective film. To clean the door, use mild soap and water with a soft cloth or sponge for cleaning the panels. Rinse remaining washing solution with distilled water. Wipe dry the excess water with clean soft cloth or sponge.
* Remove dust with moist cloth piece from outer incubator surface, especially front screen and upper part of the incubator.
* Visually check the hydrophobic filter (see 9.7) and access port seal (see 9.8) and replace if necessary.
* Check the accuracy of CO2 sensor system by measuring the actual CO2 concentration with a reference CO2 analyser.
- Expendable materials:**
Name| Description**
---|---
UV lamp| TUV 6W, T5, G6
Fuse| M 3.15 A
Hydrophobic filter| CO2 filter
Hydrophobic filter set| Filter and tubing
Access port seal| Made from foamed PE
.
-
Fuse replacement. Disconnect incubator from the mains. Disconnect power plug from power socket. Open fuse holder. Check and replace if necessary, 230 V – М 3.15 А (type М – time lag: Medium).
-
UV lamp replacement. Average lifetime of the supplied UV lamps is 9000 h. Re-placement is necessary after lamp stops functioning or at the end of manufacturer specified lifetime. Only the persons who have completed special training are allowed to perform lamp replacement. UV lamp operation lifetime is checked and tracked by the incubator.
-
Hydrophobic filter replacement. The filter should be replaced when it becomes glazed with dirt or with sucked liquid. Maximum time of use is 30 days. We recommend replacing the filter immediately after full cycle of maintenance.
Close the CO2 valve on the gas reservoir and turn off the unit. Remove the filter from the tubing. Insert a new filter into the socket and tighten it. Pay attention that “IN” marking must be facing right side when viewing the incubator from the front. -
Access port seal replacement. To prevent possible contamination of access port seal (made from foamed polyethylene) and to ensure complete sealing of the port, the seal should be replaced after 30 days from the first operation.
Humidity sensor replacement. One replacement sensor is included in the standard set. Replace as necessary, replacement can be done only by qualified and specially trained personnel. -
Calibration. Recalibrate the unit after the date on calibration certificate. To complete the process, contact Biosan or your local Biosan representative.
Warranty
-
The Manufacturer guarantees the compliance of the unit with the requirements of Specifications, provided the Customer follows the operation, storage and transportation instructions.
-
The warranted service life of the unit from the date of its delivery to the Customer is 24 months. For extended warranty, see 10.5.
-
Warranty covers only the units transported in the original package.
-
If any manufacturing defects are discovered by the Customer, an unsatisfactory equipment report shall be compiled, certified and sent to the local distributor address. To obtain the claim form, visit Technical support page on our website at link below.
-
Extended warranty. For S-Bt Smart Biotherm, the Smart class model, extended warranty is a paid service. Contact your local Biosan representative or our service department through the Technical support section on our website at the link below.
-
Description of the classes of our products is available in the Product class description section on our website at the link below.
Technical supportbiosan.lv/en/support
Product class description
biosan.lv/classes- en -
The following information will be required in the event that warranty or post-warranty service comes necessary. Complete the table below and retain for your records.
Model| Serial number| Date of sale
---|---|---
S-Bt Smart Biotherm , Compact CO2 Incubator| |
.
- The Bluetooth® logo and name are registered trademarks of Bluetooth SIG, Inc. andany use by Biosan is under licence.
EU Declaration of conformity
- Unit type: Compact CO2 incubator
- Models: S-Bt Smart BioTherm
- Serial number: 14 digits styled XXXXXXYYMMZZZZ, where XXXXXX is model code, YY and MM – year and month of production, ZZZZ – unit number.
- Manufacturer: SIA BIOSAN Latvia, LV-1067, Riga, Ratsupites str. 7/2
The objects of the declaration described above is in conformity with the following relevant Union harmonization legislations:
LVD 2014/35/EU
| LVS EN 61010-1:2011 Safety requirements for electrical equipment for measurement, control, and laboratory use. General requirements.
LVS EN 61010-2-010:2015 Particular requirements for laboratory equipment for the heating of materials.
---|---
EMC 2014/30/EU
| LVS EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use. EMC requirements.
General requirements.
RoHS3 2015/863/EU
| Directive on the restriction of the use of certain hazardous substances in
electrical and electronic equipment.
WEEE 2012/19/EU| Directive on waste electrical and electronic equipment.
I declare that the Declaration of Conformity is issued under sole responsibility of the manufacturer and belongs to the above-mentioned objects of the declaration.
Biosan SIA
Ratsupites 7 k-2, Riga, LV-1067, Latvia Phone: +371 67426137 Fax: +371
67428101 http://www.biosan.lv
Edition 1.03 – April of 2021
References
- Product classes | Biosan
- Technical support | Biosan
- 📧marketing@biosan.lv
- 📧service@biosan.lv
- Home | Biosan
- Home | Biosan
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>