Health o meter 2400 Series Digital Portable Wheelchair Scale Instruction Manual
- August 15, 2024
- Health o meter
Table of Contents
- CAUTIONS AND WARNINGS
- SPECIFICATIONS
- DEFINITION OF SYMBOLS
- CERTIFICATIONS / CONNECTIVITY / DISPOSAL
- ASSEMBLY INSTRUCTIONS
- DISPLAY HEAD WALL MOUNTING INSTRUCTIONS
- POWERING THE SCALE
- SET-UP INSTRUCTIONS
- EVERLOCK® INSTRUCTIONS
- OPERATING INSTRUCTIONS
- TRANSPORT INSTRUCTIONS
- MAINTENANCE AND CLEANING
- CALIBRATION
- TROUBLESHOOTING
- WARRANTY
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Model 2400KL / 2400KG
Digital Portable Wheelchair Scales
Equipped with
Everlock®
User Instructions
PELSTAR, LLC 9500 West 55th St. McCook, IL 60525-7110 USA
www.homscales.com © Pelstar, LLC 2018
Thank you for your purchase of this Health o meter® Professional product.
Please read this manual carefully, and keep it for easy reference or training.
Note: This scale has been factory calibrated, and does not require
calibration prior to use.
Please register your scale for warranty coverage at: www.homscales.com
/product-support/technical-support/product-warranty-
registration
For User Instructions updates and revisions, please go to:
www.homscales.com
CAUTIONS AND WARNINGS
INTENDED USE
This Health o meter® Professional wheelchair scale is intended to be used in a
professional medical environment by trained medical staff. This product was
designed to weigh patients who are safely seated in a wheelchair. Do not
modify the product or use it for anything other than its intended purpose.
To prevent patient/caregiver injury or damage to your scale, please follow the
instructions in this user manual very carefully.
- Do not exceed the weight capacity specified for this scale.
- For accurate weighing, this scale must be placed on a flat, stable surface.
- For accurate weighing, verify before each use the proper operation according to the procedure described in this manual.
- Do not use in the presence of flammable or explosive materials.
- If the scale becomes damaged, it should not be operated until properly serviced.
- Do not transport the scale with a patient or object on the scale.
- To prevent injury and damage to your scale, do not drop or mishandle scale, during use or transport.
- Ensure that the AC adapter and scale do not come in contact with liquids, excessive temperature, or excessive humidity.
Patient/caregiver safety
- This scale is designed for static weighing of patients only. No scale should be used for patient transfer.
- To prevent patient injury, the patient must be attended throughout the entire weighing event.
- Ensure that the wheel tracks are kept clean and dry. Exercise caution when rolling the wheelchair onto and off of the scale.
When operating the scale with the AC adapter:
- Operate this scale exclusively with the AC adapter that is provided (ADPT30). Use of an unspecified adapter will void the warranty, and can pose a serious safety hazard.
- Prior to using this scale, inspect the AC adapter cord for cracking / fraying, or for broken / bent plug prongs.
- Prior to using this scale, ensure that the AC adapter is plugged into an outlet with the rated voltage appropriate for operation.
- Ensure that the AC adapter outlet is wired to a circuit breaker or other protected power source.
- Unplug the AC adapter, and carefully store both the adapter and adapter cord before moving the scale.
- Operating this scale at voltages and frequencies other than specified can damage the equipment and will void the warranty.
When operating the scale with batteries:
- If the scale’s battery icon indicates that the batteries are low replace the batteries or switch to the AC adapter as soon as possible.
- When replacing / inserting batteries, be sure to use all new batteries.
This equipment has been tested and found to comply with the EMC limits for the Medical Device Directive 93/42/EEC (EN 55011 Class A and EN 60601-1-2). These limits are designed to provide reasonable protection against harmful interference in a typical medical installation. The equipment generates, uses and may radiate radio frequency energy and, if not installed and used in accordance with these instructions, may cause harmful interference to other devices in the vicinity. There is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference with other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to eliminate the interference by one or more of the following measures:
- Reorient or relocate the receiving device.
- Increase the separation between the equipment.
- Connect the equipment into an outlet on a circuit different from that to which the other device(s) is connected.
- Consult Customer Service or a field service technician for help.
In no event whatsoever shall Pelstar, LLC be liable for damages or injuries whatsoever arising out of or connected with the assembly, use, or misuse of its products.
SPECIFICATIONS
General
This scale uses highly sophisticated microprocessor technology. Each precision
scale is designed to provide accurate, reliable, and repeatable weight
measurements. In addition, each scale is designed to provide the user with
features that make the weighing process simple, fast, and convenient.
This scale uses motion-sensing weighing technology in order to determine the
actual weight of a moving patient.
The 2400KL displays weight in pounds (lb) or kilograms (kg). Refer to Page 12
to learn more about configuring the weight measuring unit setting. Note: The
2400KG displays weight in kilograms only.
The scale features a remote display head which can be placed on a flat surface
or mounted on a wall. For easy mobility the scale’s wheel tracks fold and the
scale can be carried using the built-in transport handle. The scale can be
powered using an AC adapter (ADPT30 included), or by 6 D batteries (not
included).
Specifications
Capacity and Resolution| 2400KL: 800 lb x 0.2 lb (363 kg x 0.1 kg) 2400KG: 363
kg x 0.1 kg
---|---
Power Requirements| 100-240V AC, 50-60Hz to 9V DC Adapter (Type: UE15WCP1 9
VDC @ 500 mA), Item # ADPT30 (included) or 6 D batteries (not included)
Environmental| Operating temperatures: 50°F to 95°F (10°C to 35°C) Storage
temperatures: 30°F to 125°F (0°C to 50°C) Maximum Humidity: 85% RH
Physical Dimensions| Folded for Storage I Transport Length: 39 7/8″ (1013 mm)
Width: 27″ (675 mm) Height: 6″ (150 mm) Weight: 35 lb (16 kg)| Assembled Size
and Weight Length: 39 7/8″ (1013 mm)
Width: 26″ to 50″ (660 mm to 1270 mm) Height: 3 3/8″ (86 mm)
Weight: 35 lb (16 kg)
Scale information can be found by pressing and selecting “Scale Info” in the submenu. This screen provides model number, capacity, calibration weight, the most recent field calibration date, number of field calibrations, and software version.
DEFINITION OF SYMBOLS
| Attention: Consult Accompanying Documents
---|---
| Caution
| Type B Applied Parts
| DC Power Supply
| Do not dispose of this device in the domestic waste stream.
Scale surface is applied part, Type B.
CERTIFICATIONS / CONNECTIVITY / DISPOSAL
Certification Descriptions
| Pelstar, LLC has been officially certified as the manufacturer of Health o
meter® Professional medical devices. The Pelstar, LLC quality assurance system
covers the development, production, sales and service of medical scales and
measuring systems
---|---
| Complies with:
ANSI/AAMI ES60601-1 (2005) + AMD 1 (2012), CAN/CSA- C22.2 No. 60601-1 (2014),
IEC 60601-1-6 (2010 ) + AMD 1
(2013)
| Products carrying this symbol meet the requirements of: Directive governing
medical devices 93/42/EEC
All Health o meter® Professional scales are manufactured in FDA-registered
factories.
Scale Connectivity Information
To enable reliable transmission of weight, height, and Body Mass Index (BMI)
data, this scale is designed to connect to a computer, monitor, or other
electronic device via USB port or Pelstar wireless technology. This
connectivity helps minimize a source of potential medical errors caused by
incorrectly copying and then recording patient data. For more information
regarding how to connect the scale to other electronic devices, please contact
Health o meter® Professional Scales Technical Support, available during normal
business hours at 1-800-638-3722.
Scale Disposal
Health o meter® Professional scales must be disposed of properly as electronic
waste. Follow the national, regional or local regulations which apply to you
for disposal of electronic waste or batteries. Do not dispose of this device
in the domestic waste stream.
ASSEMBLY INSTRUCTIONS
Before Assembly
Each 2400KL/2400KG scale is shipped disassembled in a single carton. Carefully
inspect the carton for shipping damage before unpacking. If damage is found,
contact your shipper or a Health o meter® Professional representative
immediately at 1-800-815-6615. Claims
must be filed with the shipper as soon as possible after receipt of the
package. The following information details what you will find inside the main
carton as you unpack the parts for assembly.
To prevent scratching any components, carefully remove each assembly from the
carton and unwrap the packing materials. Set the carton aside for storage. To
avoid damaging the scale parts when unpacking, do not use a box cutter, knife,
scissors, or any sharp object to open the protective inner packaging.
Parts List
(1) Wheelchair scale
(1) Small brown box
The small brown box contains:
(1) Display assembly
(1) Load cell cable
(1) User information
(1) Long white box
The long white box contains:
(2) Locking knobs
(1) AC adapter
(8) Feet
Note: In order to mount the scale’s display head onto the wall, a set of
wall anchors and screws will be required (not included). See page 9 for
display head wall mounting instructions.
Tools Required
Phillips head screwdriver (not included)
If mounting the display head on a wall, a drill or screwdriver is required.
(not included)
Health o meter® Professional Scales recommends that assembly be performed by
two people.
Warning: Do not drop the scale, as this can permanently damage the load
cells.
-
Remove the scale from the carton and place it upside down on a flat, level, and dry surface. Screw the (8) feet into the underside of the wheel tracks. Completely hand tighten the feet, then turn them back 1-½ full turns. This allows for proper load cell operation.
-
Carefully flip the scale over to its upright position on the floor.
Warning: Do not drop the scale, as this can permanently damage the load cells.
Open the scale to its operational position by folding open the wheel tracks with the feet securely placed onto the floor. Insert the two locking knobs into the carrying handle.
Note: For accurate weighing the 2 locking knobs must be completely hand tightened. -
Adjust the feet to level the wheel tracks. Make sure all scale parts are leveled at least ½” (12 mm) from the floor paying special attention to the two folding hinges in the center. If needed, raise the scale by adjusting all eight feet.
-
Bring the wheelchair close to the wheel tracks to determine the desired distance between the wheel tracks. Roll the wheelchair onto the scale and ensure both the front and rear wheels of the wheelchair are on the wheel tracks. To adjust the distance between the wheel tracks loosen the adjusting knobs and slide the ramp to the desired width. Tighten the knobs completely to lock the ramp positions.
-
Connect the load cell cable to the scale by inserting the load cell cable into the load cell port located on the inner side of the wheel track.
-
If mounting the display head onto a wall see the Display Head Wall Mounting Instructions on the next page. If using as a remote display head plug the load cell cable into the load cell port on the rear of the display head. Carefully position the display head on a desk, table or other flat surface. See Powering the Scale, page 11, for battery and AC adapter instructions. The scale is now ready to use.
Caution: Ensure the load cell cable is adequately secured to prevent a tripping hazard.
DISPLAY HEAD WALL MOUNTING INSTRUCTIONS
Required Parts
(1) Display head with mounting bracket
(2) #8 Anchoring screws (4.2 mm nominal OD)
(2) Mounting anchors, compatible with wall construction
Caution: When wall mounting the display head, ensure that the mounting
location is free of electrical lines, gas pipes, and other safety hazards.
Ensure that wall is of construction that can support the display head.
Note: Remote display head can be mounted on a wall up to 71” (1.8 m) from
the platform or placed on a table up to 71” (1.8 m) from the platform.
-
Place the display head with its attached mounting bracket on the wall in the desired location. Ensure the display head is level and that the load cell cable can be plugged into the display. Draw a line on the wall at the bottom center of the display head. Set the display head aside.
-
Draw a level horizontal line, 2 ½” (63mm) above the bottom line.
-
Draw a vertical line starting from the center of the bottom line to the line created in the previous step.
-
Locate the point where the upper horizontal line and vertical line intersect. From that point draw mounting anchor locations on the upper horizontal line 2 ¼” (57.5mm) to the left and right of the intersection. Drill holes at the marked locations.
-
Once the holes are drilled, use a hammer or Phillips head screwdriver to set the 2 drywall anchors into the holes. Use a screwdriver to insert the drywall anchors completely into the wall. Screw the two mounting screws into the anchors. Adjust the screw head depth so the screw heads protrude 1/8” (3.2 mm) from the wall.
Warning: Extending the screw heads beyond 1/8” (3.2 mm) can damage the display assembly. -
Obtain the display head and the load cell cable. Plug the load cell cable into the load cell port on the rear of the display head. See Powering the Scale, page 11, for battery and AC adapter instructions.
-
After inserting batteries or plugging in the AC adapter into the adapter jack, position the mounting bracket holes onto the mounting screws, and carefully pull down on the display head to secure. The scale is now ready to use.
POWERING THE SCALE
Powering the Scale
Plug the AC adapter into the power adapter jack (the jack is located on the
rear of the display head), then plug the scale’s AC adapter into the power
source.
Warning: To avoid a safety hazard, do not use an unspecified AC adapter.
Only use the Health o meter® Professional AC adapter model ADPT30 designed for
this scale. For battery operation, see Battery Installation below.
Battery Installation
- If connected, unplug the AC adapter from the power source. Use a Phillips head screwdriver to remove the battery cover on the side of the display head by unscrewing the battery cover screw.
- Remove the battery pack from the display head. Replace or install fresh batteries, making sure to install them in the proper direction. Health o meter® Professional recommends using an alkaline battery rated for 130°F (54°C) operation. Do not use Zinc-Carbon batteries.
- Re-insert the battery pack into the display head, ensuring the end with the silver contacts is inserted facing the inside of the display. Re-attach the battery cover to the display head by replacing the battery cover screw.
Warning: If the scale will not be used for some time, remove the batteries to avoid a safety hazard. Disposal of batteries must be performed in accordance with the prevailing national, regional or local regulations that apply to you.
SET-UP INSTRUCTIONS
The scale must be powered on before performing any of the following
operations.
Press to turn on the scale.
Unit of Measure Configuration / Unit Lock
(This feature is not available on KG models)
The scale ships from the factory set to display weight in kilograms. The
weight measuring unit can be changed by pressing . Follow this procedure to
lock the weight measuring unit into LB or KG only.
- Press . Follow this procedure to lock the weight measuring unit into LB or KG to access the Help menu options.
- While in the Help menu, press to select “Set Options”. Press to advance to the next menu.
- “Units Lock” is the first setting listed in the “Set Options” submenu. With “Units Lock” highlighted in red, press to change the “Units Lock” option. Choose “Unlocked”, “LBS Locked”, or “KG Locked”.
Note: You must press to save your selection.
Note: To permanently lock the weight measuring unit, see the Everlock®
section on page 15.
Enabling Unit View with Unit of Measure Locked
(This feature is not available on KG models or models with a date code lower
than 3118)
If the scale is locked per the instructions above, users can still view the
alternate weight measuring unit by enabling the “Unit View” option. Enabling
the “Unit View” option allows the user to press the Unit button to briefly
display the weight in the alternate weight measuring unit. After releasing the
Unit button the display will return to the locked weight measuring unit. The
“Unit View” default setting is “Disable”. Follow this procedure to enable
“Unit View” when the weight measuring unit is locked.
Note: If Everlock® is engaged this feature is not available and the
setting will show as “Everlocked”.
- Press to access the Help menu options.
- While in the Help menu, press to select “Set Options”. Press to advance to the next menu.
- While in the “Set Options” sub-menu, press to scroll through the options until “Unit View” is highlighted in red.
- Press to change the “Unit View” option. Choose “Enable” to allow the user to view the weight in the alternate weight measuring unit. Choose “Disable” to completely disable the Unit button when weight measuring unit is locked.
Note: You must press to save your selection.
Setting Sound Options
The display can be set to stay silent or beep while operating the scale.
Follow this procedure to set the preferred option.
- Press to access the Help menu options.
- While in the Help menu, press to select “Set Options”. Press to advance to the next menu.
- While in the “Set Options” sub-menu, press to scroll through the options until “Sounds” is highlighted in red.
- Press to change the “Sounds” option. Choose “Disable” for the display to stay silent while operating the scale. Choose “Enable” for the display to beep while operating the scale.
Note: You must press to save your selection.
Activating/Deactivating AC Auto Off
When powering the scale via the AC adapter, the display can be set to power
off after a period of inactivity. Follow this procedure to set the preferred
option. Note: When powering the scale via batteries, the scale will
automatically power off after a period of inactivity. See the next section to
set the preferred time of inactivity before power off.
- Press to access the Help menu options.
- While in the Help menu, press to select “Set Options”. Press to advance to the next menu.
- While in the “Set Options” sub-menu, press to scroll through the options until “AC Auto Off” is highlighted in red.
- Press to change the “AC Auto Off” option. Choose “Enable” for the display to turn off after a period of inactivity. Choose “Disable” for the display to always remain on.
Note: You must press to save your selection.
Setting Auto Off Timer
When powering the scale via batteries or if Auto Off is Enabled for the AC
adapter, the period of inactivity before the scale powers off can be set to a
specific amount of time. Follow this procedure to set the amount of seconds of
inactivity before power off.
- Press to access the Help menu options.
- While in the Help menu, press to select “Set Options”. Press to advance to the next menu.
- While in the “Set Options” sub-menu, press to scroll through the options until “Auto Off Seconds” is highlighted in red.
- Press to change the “Auto Off Seconds” option. Choose from 60, 90, 120, 150 or 180 seconds.
Note: You must press to save your selection.
Setting Auto Zero Option
When a scale is set to zero, it creates a reference point for weights that are
placed on it. There are two approaches to how this can take place. This option
allows the user to select whether the scale will reset itself to ZERO each
time it is powered up or hether it will use the zero reading from the last
time it was zeroed.
-
“Enable” means the scale will zero itself when it is powered on and this may take a few seconds. After each use, the scale will clear the weight of the patient and set itself back to the initial ZERO level it found on start-up. Health o meter® Professional recommends keeping this setting as Enable.
Note: Hitting a key to wake the scale from sleep mode will not prompt the scale to zero. -
Disable” means the scale does not attempt to find a zero when powered on, rather it will use the last zero setting it had from its most recent use. There will be no delay in using the scale if the Auto Zero is disabled. Health o meter® Professional recommends that if the Disable option is selected, users should manually zero the scale before each weight measurement by pressing . Users will see dashes appear on the screen as the scale is finding zero.
- Press to access the Help menu options.
- While in the Help menu, press to select “Set Options”. Press to advance to the next menu.
- While in the “Set Options” sub-menu, press to scroll through the options until “Auto Zero on Power Up” is highlighted in red.
- Press to change the “Auto Zero on Power Up” option. Choose from Enable or Disable.
Note: You must press to save your selection.
Setting Date and Time
- Press to access the Help menu options.
- While in the Help menu, press to select “Set Clock”.Press to advance to the next menu.
- Use and to set the time and date.
Note: You must press to save the information.
EVERLOCK® INSTRUCTIONS
Everlock®
(This feature is not available on KG models or models with a date code
lower than 3118)
The Everlock® feature allows the scale to be permanently locked in KG (metric)
or LB (imperial) units. When activating this feature, the Unit button will
become disabled. This lock is permanent and can only be reversed at the
factory.
This procedure engages Everlock® and will permanently lock the scale into LB
or KG.
Note: This feature is standard on models with a date code of 3118 or
higher.
Warning: This lock is permanent and can only be reversed at the factory.
If you need to change the units on your scale after Everlock® has been
engaged, contact Health o meter® Professional Scales Customer Service at
1-800-815-6615.
-
Press to access the Help menu options.
-
While in the Help menu, press to select “Set Options”. Press to advance to the next menu.
-
While in the “Set Options” sub-menu, press to scroll through the options until “Everlock” is highlighted in red.
-
Press to begin the Everlock® procedure.
To exit the procedure without engaging Everlock® press .
-
The screen will prompt you to enter the password. Use the arrows keys to enter the password as follows:
-
Use the up and down arrow keys to select “Everlock LB” or “Everlock KG”, press to advance to the next step.
-
The screen will display a warning. To proceed with locking the weight measuring unit, use the arrow keys to enter the password as follows:
-
The weight measuring unit is now locked. The Options screen will now show “Units Lock” and “Unit View” as “Everlocked” and those settings cannot be changed by any user.
OPERATING INSTRUCTIONS
Key definitions can also be found by pressing and selecting “Key Definitions” in the submenu. Keypad
Function | Description |
---|---|
Powers scale on and off. | |
Zeroes the scale prior to a weighing. Also used as a screen navigation key |
to exit a display menu.
| Holds the weight value on the display until the button is pressed again to
clear the value.
| Prompts entry of height data to calculate patient’s Body Mass Index (BMI).
| Recalls the previous weight.
| TARE-Deducts the weight of an object on the platform.
PRETARE-When held for 3 seconds, prompts entry of a weight value to be
deducted from the weight on the platform. Also releases tare weight (returns
display to zero).
| Toggles between pounds and kilograms. If weight measuring unit is locked and
Unit View setting is enabled, press and hold Unit button to briefly display
the
alternate weight measuring unit. Note: The Unit button is disabled on KG
models or
on KL models with Everlock® engaged or with the “Unit View” setting disabled.
|
|
Function| Description
---|---
| Powers scale on and off.
| Zeroes the scale prior to a weighing. Also used as a screen navigation key
to exit a display menu.
| Holds the weight value on the display until the button is pressed again to
clear the value.
| Prompts entry of height data to calculate patient’s Body Mass Index (BMI).
| Recalls the previous weight.
| TARE-Deducts the weight of an object on the platform.
PRETARE-When held for 3 seconds, prompts entry of a weight value to be
deducted from the weight on the platform. Also releases tare weight (returns
display to zero).
| Toggles between pounds and kilograms. If weight measuring unit is locked and
Unit View setting is enabled, press and hold Unit button to briefly display
the alternate weight measuring unit. Note: The Unit button is disabled on KG
models or on KL models with Everlock® engaged or with the “Unit View” setting
disabled.
| Allows repeated weighing of the patient without stepping off the scale.
| Used to accept height value in BMI mode, and tare value in PRETARE mode.
Also used to save selections in Set Options menu.
| Display navigation keys.
Tare Function
When using this scale, the weight of an object, such as a wheelchair, can be
subtracted from the total weight in order to determine the weight of the
patient alone. The Tare Function automatically performs this subtraction. The
scale allows a tared weight to be entered manually by using the keypad, or
automatically, as outlined in this section.
Weighing a Patient and Taring an Unknown Weight
- Wait until “0.0” appears on the display.
- Place the object to be tared on the scale. The display will show the object’s weight. The minimum weight that can be tared is 5 lb / 2.27 kg. Note: The scale must lock onto a weight before it can be tared.
- With the object remaining on the scale, press . The scale will zero, and the word “TARE” will appear at the bottom center of the display along with the tared weight. Remove the object from the scale.
- Place the patient and the object back on the scale. The scale will automatically deduct the tared weight of the object from the gross weight of the patient and object together. The display will show the patient’s weight.
- With no weight on the scale, the display will show the tare value as a negative amount. The tare value will remain on the display until changed, cleared, or until the scale is powered off.
- The weight may be held on the display by pressing .The scale will continue to display the patient’s weight after they are off the scale and will remain on the display until released or the scale is powered off. To release, press .
- To reweigh the patient without removing them from the scale, press .
- To see the patient’s weight in another unit of measurement, press and hold down . If the scale is in LB mode, it will convert the weight to KG. If the scale is in KG mode, it will convert the weight to LB. Note: This function is not available on KG models or on KL models with Everlock® engaged or with the “Unit View” setting disabled.
- To remove the Tare, press . The word “TARE” will disappear from the bottom of the display, and the tare value will be cleared from the display. The scale will resume normal operation.
Weighing a Patient and Taring a Known Weight / Pre-Tare
Pre-Tare can be performed within the following weight limits: 0.2 lb – 200 lb
/ 0.1 kg – 91 kg)
- Wait until “0.0” appears on the display.
- Press and hold for approximately 3 seconds. The display will change and prompt entry of the weight value to be tared. Use the arrow keys to enter the weight to be tared. Press to save the tare value.
- The scale will zero and the tare value will show as a negative weight on the display. The word “TARE” will also appear at the bottom center of the display along with the tare value.
- Place the patient on the scale. The scale will automatically deduct the tare value from the weight of the patient. The display will show the patient’s weight.
- The weight may be held on the display by pressing The scale will continue to display the patient’s weight after they are off the scale and will remain on the display until released or the scale is powered off. To release, press .
- To reweigh the patient without removing them from the scale, press .
- To see the patient’s weight in another unit of measurement, press and hold down . If the scale is in LB mode, it will convert the weight to KG. If the scale is in KG mode, it will convert the weight to LB. Note: This function is not available on KG models or on KL models with Everlock® engaged or with the “Unit View” setting disabled.
- To remove the Tare, press . The word “TARE” will disappear from the bottom of the display, and the tare value will be cleared from the display. The scale will resume normal operation.
Weight Recall
- To recall the last weight measured, press .
2. To clear the display of the recalled weight, press .
BMI Function
-
Make sure there is nothing on the weighing platform and the display shows “0.0”.
-
Follow weight measuring procedure. After obtaining the patient’s weight, and with the patient remaining on the scale, press .
-
The scale will prompt entry of the patient’s height. Use the arrow keys to enter the patient’s height.
Note: When weighing in pounds, the height is displayed in inches. When weighing in kilograms, the height is displayed in centimeters. -
Once height is keyed in, press .
-
The display will show the patient’s BMI.
-
Press to return to the normal weighing mode.
TRANSPORT INSTRUCTIONS
This scale can be easily folded for movement from one location to another. This scale includes mounting pins to secure the display assembly during transport (the pins are located below the transport handle). To prepare the scale for transport, follow these steps.
- If powering the scale via AC adapter, unplug the adapter from the display head, and remove the adapter from the outlet. Carefully store the AC adapter and cord to prevent damage during transport.
- If the display head is mounted on a wall, remove it from the wall by lifting it up and off of the mounting screws. Set it aside.
- Unplug the load cell cable from the scale platform and the display head. Carefully store the load cell cable during transport.
- Loosen the ramp adjusting knobs and fully expand the ramps to the widest position. Fully tighten the knobs to lock the ramps.
- Loosen the two locking knobs located on the carrying handle.
- Grasp the handle and lift the scale into its upright position.
- Position the display head onto the mounting pins located on the folded ramp.Slide the display head mounting bracket onto the raised mounting pins. The scale is now ready for transport.
MAINTENANCE AND CLEANING
Maintenance
The following pages provide instructions for maintenance, cleaning,
calibration, and troubleshooting for your scale. Maintenance operations other
than those described in this manual should be performed by qualified service
personnel.
Caution: Before first use, or after long periods of non-use, check the scale
for proper operation and function. If the scale does not operate correctly,
refer to qualified service personnel.
- Check the overall appearance of the scale for obvious damage, wear, and tear.
- Inspect the AC adapter cord for cracking, fraying, and for broken/bent prongs.
Cleaning and Disinfecting
Proper care and cleaning is essential to ensure a long life of accurate and
effective weighing.
Caution: Disconnect the scale from the AC adapter power source prior to
cleaning the unit.
-
Health o meter® Professional recommends using one of the following solutions on a soft cloth or disposable wipe:
● mild soap and water solution
● 70% isopropyl alcohol
● solution with 1-5% hydrogen peroxide concentration
After cleaning/disinfecting, wipe with a cloth dampened with water and then with a clean dry cloth. To avoid residue build-up or damp surfaces, ensure the screen and scale parts are completely dry after cleaning. -
Never use rough or abrasive materials to clean the scale, as these will damage the scale’s finish.
-
Do not submerge the scale in water or any other liquid.
-
Do not pour or spray fluids directly on the scale.
Replacement Parts
Parts may be available if a piece of the scale needs to be replaced. Contact
Health o meter Professional Scales Customer Service at
1-800-815-6615 to inquire about the
availability of these replacement parts.
Part # | Description |
---|---|
ADPT30 | Power adapter |
B2266701-0 | Replacement foot |
B4010298 | Strain relief bushing |
B3822801-0 | Replacement battery door cover |
10-00252 | Replacement battery holder |
CALIBRATION
This scale has been factory calibrated, and does not require calibration prior
to use.
There are no requirements for calibration in the field; users should adhere to
their institution’s calibration policies.
Calibration of your scale can be performed using calibrated kilogram (kg) or
pound (lb) weight standards. Minimum calibration is 200 lb / 100 kg. Maximum
calibration weight is 800 lb / 363 kg. The required calibration standard will
be the Unit of Measure (UOM) for which the scale is configured. KG models can
only be calibrated in kilograms. If Everlock® is engaged the scale can only be
calibrated in the locked unit of measure.
Note: Only weights certified and traceable to national standards should
be used for calibration.
Warning: Always ensure that the scale is returned to the correct UOM
after scale calibration.
- Press to access the Help menu options.
- While in the Help menu, press to select “Calibrate”. Press to advance to the next step.
- The display will prompt for entry of the password. Press the following keys in this sequence to proceed with the calibration procedure.
- The display will show a warning. If prepared and qualified to perform the calibration procedure, press to advance to the next step.
- Follow the screen prompt and ensure the scale is on a level surface with no weight on the platform. Press to advance to the next step.
- The display will show “Zeroing Scale” as it calculates the zero value. Do not touch the scale during the zeroing process.
- When the zeroing process is complete the display will prompt for entry of the calibration weight. Minimum calibration is 200 lb / 100 kg. Maximum calibration weight is 800 lb / 363 kg. Calibration weight can be adjusted in 10 lb or 10 kg increments. Press when the calibration weight is set.
- The display will prompt for the calibration weights to be placed on the platform. Press Enter after calibration weights are placed on the platform.
- The display will show “Calibrating” during the calibration process.
- If calibration is successful, the display will show “Calibration Complete” with a 4 digit calibration count.
- If calibration is unsuccessful, the display will show an error message that the entered calibration weight and the certified weights do not match. The display will then revert back to the set calibration weight screen for entry of calibration weight.
TROUBLESHOOTING
Troubleshooting
Before contacting service personnel, refer to the following instructions to
check and to correct any failures. For further assistance, contact Health o
meter® Professional Scales Technical Support at
1-800-638-3722.
Symptom | Possible Cause | Corrective Action |
---|---|---|
Scale does not power on | 1.Dead battery |
2.Faulty electrical outlet
3.Bad power supply| 1.Replace batteries
2.Use a different outlet
3.Replace AC adapter
Questionable weight or
the scale does not zero| 1.External object interfering with the scale
2.The display did not show “0.0” before weighing
3.Scale is not placed on a level floor or feet are not properly balanced
4.Scale is out of calibration| 1.Remove interfering object from the scale
2.Remove the patient from the scale, press to zero the
scale and begin weighing process again
3.Place the scale on a level floor. Ensure all 8 feet are properly balanced.
Completely hand tighten the feet, then turn them back 1-% full turns.
4.Check weight with knownweight value
Weighing is performed but the weighing process takes too long and the weight
does not lock on the display| The patient is not sitting still| Ask the
patient to sit still
The battery icon on the display indicates that the batteries are low| The
batteries are depleted| Replace batteries according to instructions on page 11
Scale is displaying “Load
Cell Cable Not Detected”| The load cell cable is not completely plugged into
the ports| Inspect the load cell cable to ensure it is completely plugged into
the ports on the back of the display head and the side of the wheel track. See
assembly instructions. If cable was plugged in, unplug and plug in again until
it clicks into place.
In the Help menu, display is not advancing to the next screen for the menu
selection and saying “invalid key”| Correct key is not being pressed| To make
a selection in the Help Menu and advance the screen, press .
WARRANTY
Limited Warranty
What does the Warranty Cover?
This Health o meter® Professional scale is warranted from date of purchase
against defects of materials or in workmanship for a period of two (2) years.
If product fails to function properly, return the product, freight prepaid and
properly packed to Pelstar, LLC (see “To Get Warranty Service”, below, for
instructions). If the manufacturer determines that a defect of material or in
workmanship exists, the customer’s sole remedy will be replacement of the
scale at no charge. Replacement will be made with a new or remanufactured
product or component. If the product is no longer available, replacement may
be made with a similar product of equal or greater value. All replaced parts
are covered only for the original warranty period.
Who is Covered?
The original purchaser of the product must have proof of purchase to receive
warranty service. Please save your invoice or receipt. Pelstar dealers or
retail stores selling Pelstar products do not have the right to alter, or
modify or any way change the terms and conditions of this warranty.
What is Excluded?
Your warranty does not cover normal wear of parts or damage resulting from any
of the following: negligent use or misuse of the product, use on improper
voltage or current, use contrary to the operating instructions, abuse
including tampering, damage in transit, or unauthorized repair or
alternations. Further, the warranty does not cover natural disasters, such as
fire, flood, hurricanes and tornadoes. This warranty gives you specific legal
rights, and you may also have other rights that vary from country to country,
state to state, province to province or jurisdiction to jurisdiction.
To get Warranty Service make sure you keep your sales receipt or document
showing proof of purchase. Call (+1)
800-638-3722 or (+1)
708-377-0600 to receive a return
authorization (RA) number, which must be included on the return label. Attach
your proof of purchase to your defective product along with your name,
address, daytime telephone number and description of the problem. Carefully
package the product and send with shipping and insurance prepaid to:
Pelstar, LLC
Attention R/A#_____
Return Department
9500 West 55th Street
McCook, IL 60525
Extended Warranty Available
This product is eligible for Health o meter® Professional ScaleSurance
Extended Warranty Program. ScaleSurance extends the warranty period for an
additional two years. This extension to the standard Limited Warranty can be
purchased with new products or for a facility’s existing product before its
current warranty has expired. To learn more, visit
www.homscales.com/scalesurance/ or
contact your medical supply distributor.
Typical product life is 5 years but may vary significantly based on daily
usage, treatment and other factors.
QNET BV KANTSTRAAT 19
NL-5076 NP HAAREN
THE NETHERLANDS
PELSTAR, LLC
9500 West 55th Street • McCook, IL 60525 • USA
1-800-638-3722 or
1-708-377-0600
PLEASE REGISTER YOUR SCALE FOR WARRANTY COVERAGE AT:
www.homscales.com
Health o meter® is a registered trademark of Sunbeam Products, Inc. used under
license. Health o meter® Professional
products are manufactured, designed, and owned by Pelstar, LLC. We reserve the
right to improve, enhance, or modify
Health o meter® Professional product features or specifications without
notice.
© Pelstar, LLC 2018