Health o meter Professional 500KG Eye Level Digital Scale Instruction Manual

June 9, 2024
Health o meter Professional

Health o meter Professional 500KG Eye Level Digital Scale

Model 500KL/500KG

Find the date code on your scale’s product label and follow this chart to select the version specific calibration instructions for your scale.

Date Codes Link/Page #
0220, 0120, 0720, 1820, 1920,
2120, 2420, 3520, 3620, 3720, Click here or skip to page 2
3920, 4420, 4620
0121, 0221, 0421, 0521, 0821,
1021, 22212621, 2721, 2821, 3121,
3321, 3521, 3721, 4121, 4421,
4621, 4721
0222, 0322, 1022, 1122, 2522,
2622, 2722, 3322, 3522, 4522
1321, 1421, 1721 Click here or skip to page 5

For all other date codes not shown above, click here or skip to page 8

Date code can be found on the product label on the back of the display head.
Health-o-meter-Professional-500KG-Eye-Level-Digital-Scale-
fig-1

This scale has been factory calibrated and does not require calibration prior to use. There are no requirements for calibration in the field; users should adhere to their institution’s calibration policies. If required, the scale can be calibrated. Calibration of your scale is performed using 80 kilograms (KG) or 200 pounds (LB), according to the units of measure used upon entering into the calibration path. THE 500KG CAN ONLY BE CALIBRATED IN KILOGRAMS. If Everlock® is engaged the scale can only be calibrated in the locked unit of measure.

Note: Only weights certified and traceable to national standards should be used for calibration procedures.

  1. Press and hold the UNIT and ZERO/TARE buttons at the same time, and press and release the button to turn on the scale. Remain pressing the UNIT and ZERO/TARE buttons until “CAL” appears on the display. Release all buttons. Note: The display will show “CAL” until the button combination is released.
  2. Once the buttons are released the following message will scroll horizontally across the screen : “– 2Ero SCALE – ”. With no weight on the platform, press the ZERO/TARE button.
  3. The scale will zero and the following message will scroll horizontally across the screen: “- PLACE 200 ON SCALE PrESS EnTEr–” (if in pounds) or “– PLACE 80 ON SCALE PrESS EnTEr– ” (if in KG). You may toggle between pounds and kilograms using the UNIT button. If Everlock® is engaged you will only be able to calibrate in the locked weight measurement unit.
  4. Once the proper weight is placed on the platform, press the ENTER button. The message “– CAL –” will flash on the screen while the scale is calibrating.
    • If the weight is within calibration range, a four digit number will display on the screen. Remove the weight from the scale and the scale will restart. Note: the 4 digit calibration count will stay until the weight is removed.
    • If the weight placed on the scale does not match the correct calibration standard (200 lb or 80 kg), the following message will be displayed: “– Out OF RANGE –”. Calibration is not performed. Press the CLEAR/REWEIGH button to exit the calibration mode. The scale will restart and go through the power up sequence to show 0.0. Power off the scale and begin the calibration process again.

This scale has been factory calibrated, and does not require calibration prior to use. If required, the scale can be calibrated. Calibration of your scale is performed using kilograms (KG) or pounds (LB), according to the units of measure used upon entering into the calibration path. When calibrating in pounds, a total of 300 lbs is required applied in increments of 100 lb, 200 lb, and 300 lb. If calibrating in kilograms, a total of 120 kg is required applied in increments of 40 kg, 80 kg, and 120 kg. THE 500KG CAN ONLY BE CALIBRATED IN KILOGRAMS. Note: Only weights certified and traceable to national standards are to be used for calibration procedures.

ACTION DISPLAY SHOWS

With the scale off, press and hold the UNIT button (500KL) or the blank blue button (500KG) and ZERO buttons at the same time, then press the ON/OFF button to turn on the scale. Release all three buttons when “CAL” appears on the display. Numbers will appear indicating that you can begin calibration.| ****


“CAL” followed by numbers

When calibration numbers stabilize, press the ZERO button. “C-100” (for LB mode) or “C-40” (for KG mode) will flash on the display, followed by “0” or “1”.| Numbers followed by “C-100” or

“C-40” then “0” or “1”

Place 100 lbs or 40 kg on the scale. Numbers will appear on the display. When calibration numbers stabilize, press the ZERO button. “C-200” (for LB mode) or “C-80” (for KG mode) will appear on the display, followed by “0” or “1”.| Numbers followed by “C-200” or

“C-80” then “0” or “1”

Place an additional 100 lbs or 40 kg on the scale, for a total of 200 lbs or 80 kg. Numbers will appear on the display. When calibration numbers stabilize, press the ZERO button. “C-300” (for LB mode) or “C-120” (for KG mode) will appear on the display, followed by “0” or “1”.| Numbers followed by “C-300” or

“C-120” then “0” or “1”

Place an additional 100 lbs or 40 kg on the scale, for a total of 300 lbs or 120 kg. Numbers will appear on the display. When calibration numbers stabilize, press the ZERO button. “END” will appear on the display,

followed by “OFF”. The scale will automatically shut off.

| Numbers followed by “END” then “OFF”

This scale has been factory calibrated, and does not require calibration prior to use. If required, the scale can be calibrated. Calibration of your scale is performed using kilograms (KG) or pounds (LB), according to the units of measure used upon entering into the calibration path. When calibrating in pounds, a total of 300 lbs is required (models with date codes listed below require 500 lb), applied in increments of 100 lb and 200 lb. If calibrating in kilograms, a total of 120 kg is required (or 200 kg for models with date codes listed below), applied in increments of 40 kg, and 80 kg. THE 500KG

CAN ONLY BE CALIBRATED IN KILOGRAMS.

Note: Only weights certified and traceable to national standards are to be used for calibration procedures.For scales with these date codes 500 lb (200kg) of weights are required for calibration. 2215, 2315, 2715, 2815, 3015, 3115, 3215, 3415, 3515, 3715, 3815, 3915, 4115, 4215, 4315, 4515, 4615, 4715, 4915, 5015, 5215, 0316, 0916, 1016, 1116, 1216, 1316, 1516, 1616, 1816

ACTION DISPLAY SHOWS

With the scale off, press and hold the LB/KG button (500KL) or the blank blue button (500KG) and ZERO buttons at the same time, then press the ON/OFF button to turn on the scale. Release all three buttons when “CAL” appears on the display. Numbers will appear indicating that you can begin calibration.| ****


“CAL” followed by numbers

When calibration numbers stabilize, press the ZERO button. “C-100” (for LB mode) or “C-40” (for KG mode) will flash on the display, followed by “0” or “1”.| Numbers followed by “C-100” or

“C-40” then “0” or “1”

Place 100 lbs or 40 kg on the scale. Numbers will appear on the display. When calibration numbers stabilize, press the ZERO button. “C-200” (for LB mode) or “C-80” (for KG mode) will appear on the display, followed by “0” or “1”.| Numbers followed by “C-200” or

“C-80” then “0” or “1”

Place an additional 100 lbs or 40 kg on the scale, for a total of 200 lbs or 80 kg. Numbers will appear on the display. When calibration numbers stabilize, press the ZERO button. “C-300” (for LB mode) or “C-120” (for KG mode) will appear on the display, followed by “0” or “1”.| Numbers followed by “C-300” or

“C-120” then “0” or “1”

Place an additional 100 lbs or 40 kg on the scale, for a total of 300 lbs or 120 kg. Numbers will appear on the display. When calibration numbers stabilize, press the ZERO button. Calibration procedure is now complete and “END” will appear on the display. For models with date codes listed above the display will show “C-500” (for LB mode) or “C-200” (for KG mode) will appear on the display, followed by “0” or “1”.| ****

“END”

or numbers

followed by “C-500” or “C-200” then “0” or “1”

Place an additional 200 lbs or 80 kg on the scale, for a total of 500 lbs or 200 kg. Numbers will appear on the displaly. When calibration numbers stabilize, press the ZERO button. “END” will appear on the display,

followed by “OFF”. The scale will automatically shut off.

| “XXXX” followed by “END” then “OFF”

  • P/N UM500KL 500KG Rev 20190408_Web

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals