senseca ECO 123-AIR Precision Ambient Thermometer with Alarm
Product Information
Specifications
- Product Name: ECO 123-AIR Precision Ambient Thermometer with Alarm
- Model Number: B-H86.0.09.DK2-1.0
- Display: Digital display
- Alarm Function: Yes
- Power Source: Battery
Product Usage Instructions
Safety
Before using the ECO 123 Ambient Thermometer, please read and understand
the safety instructions provided in the manual.
Operating Elements
The device features a digital display for temperature readings and
operating buttons for configuration and adjustments.
Operation
To open the configuration menu, follow the steps outlined in section 4.1 of
the manual. Adjust the measuring input as needed by referring to section 4.2.
Measurement Basics
For accurate temperature measurements, follow the measuring instructions
provided in section 5.1.
Operation and Maintenance
Refer to section 6 for operating and maintenance notices. The device is
powered by a battery, and instructions for battery replacement can be found in
section 6.2.
FAQ (Frequently Asked Questions)
-
Q: How do I change the battery?
- A: To change the battery, please follow the steps outlined in section 6.2.2 of the manual.
-
Q: What should I do if I encounter an error message?
- A: Refer to section 7 of the manual for information on error and system messages and troubleshooting steps.
B-H86.0.09.DK2-1.0
About this documentation
-
Foreword
Read this document carefully and familiarize yourself with the operation of the device before you use it.
Keep this document ready to hand and in the immediate vicinity of the device so that it is available to the personnel/user for reference at all times in case of doubt.
The user must have carefully read and understood the operating manual before beginning any work. -
Legal notices
The liability and warranty of the manufacturer for damages and consequential damages are voided with misuse, disregarding this document, disregarding safety notices, assignment of inadequately qualified technical personnel and arbitrary modifications of the device.
This document is entrusted to the recipient for personal use only. Any impermissible trans-fer, duplication, translation into other languages or excerpts from this operating manual are prohibited.
The manufacturer assumes no liability for print errors. -
Further information
Software version of the device: V1.2 or later
For the exact product name, refer to the type plate on the rear side of the device.
Note
For information about the software version, press and hold the ON button to
switch on the device for longer than 5 seconds. The series is shown in the
main display and the software version of the device is shown in the secondary
display.
Safety
Explanation of safety symbols
Danger!
This symbol warns of imminent danger, which can result in death, severe bodily
injury, or severe property damage in case of non-observance.
Caution!
This symbol warns of potential dangers or harmful situations, which can cause
damage to the device or to the environment in case of non-observance.
Note
This symbol indicates processes, which can have a direct influence on
operation or can trigger an unforeseen reaction in case of non-observance.
Foreseeable misuse
The fault-free function and operational safety of the device can only be
guaran-teed if applicable safety precautions and the device-specific safety
instructions for this document are observed.
If these notices are disregarded, personal injury or death, as well as
property damage can occur.
Danger! Incorrect area of application!
In order to prevent erratic behavior of the device, personal injury and
property damage, the device must be used exclusively as described under
intended use.
- Do not use in safety. Emergency Stop devices!
- The device is not suitable for use in explosion-prone areas!
- The device must not be used for diagnostic or other medical purposes on patients!
- Not suitable for use with requirements on functional safety, e.g. SIL!
Safety instructions
Caution! Risk of injury at the temperature sensor!
Empty batteries and batteries of inferior quality can leak more easily, which
can destroy the device.
Please also observe the instructions in the chapter “Operation and maintenance”.
Note
- This device does not belong in children’s hands!
- Make sure that no dirt gets into the openings of the sensor head. If this is the case, carefully unscrew the cap. Improper handling can damage the temperature sensor!
Intended use
The device measures the ambient or air temperature; it can be held in the hand
(pivoting movement increases the response speed) or placed vertically or
horizontally on a flat surface. Protect the sensor from moisture and dirt!
Qualified personnel
For commissioning, operation and maintenance, the relevant personnel must have
adequate knowledge of the measuring process and the significance of the
measurements. The instructions in this document must be understood, observed
and followed.
In order to avoid any risks arising from interpretation of the measurements in
the concrete application, the user must have additional expertise. The user is
solely liable for dam-
ages/danger resulting from misinterpretation due to inadequate expertise.
The device at a glance
Display elements
Operating elements
On / Off button Switch on the device
Press briefly Activate / deactivate lighting
Long press Switch off the device
Reject changes in a menu
Up / Down button Display of the min/max value
Press briefly Change value of the selected parameter
Long press Reset the min/max value of the current
measurement
Both simultaneously Rotate display, overhead display
Function button Press briefly Freeze measurement (Hold)
Call up next parameter
Long press, 2s ****Start menu “configuration”, (ONF
appears in the display
Operating status ****device is in measured value display
device is in a menu
Operation
Opening the configuration menu
- Press the Function key for 2 seconds to open the Configuration menu.
- (ONF appears in the display. Release the Function key.
Parameter | Values | Meaning |
---|---|---|
AL. | Alarm | |
OFF | No alarm active | |
ON | Alarm via text insertion, acoustic signal and flashing of the background |
lighting
| BEEP| Alarm via text insertion and acoustic signal
| L,TE| Alarm via text insertion and flashing of the back- ground lighting
AL.LO| Min. alarm limit (only available if AL <> off)
| -25.0 .. AL.X,| a min. alert is triggered if the value falls below this
value. (at °F: -13.0 .. AL.Hi)
AL.XI| Max. alarm limit (only available if AL <> off)
| AL.LO .. 70.0| If the value is exceeded, a max. alarm is triggered. (at °F:
AL.Hi .. 158.0)
POFF| Shut-off time
| OFF| No automatic shut-off
| 15, 30, 60, 120, 240| Automatic shut-off after a selected time in minutes,
during which no buttons have been pressed
L,TE| Backlight
| OFF| Backlight deactivated
| 15, 30, 60, 120, 240| Automatic shut-off of the backlight after a selected
time in seconds, during which no buttons have been pressed
| ON| No automatic shut off of the backlight
Parameter| Values| Meaning
---|---|---
UN,T| Display unit
| °(| Temperature display in °C
| °F| Temperature display in °F
IN,T| Factory settings
| NO| Use current configuration
| YES| Reset device to factory settings.
After confirming with the function-button, the display shows: IN,T DONE
Adjustment of the measuring input
The temperature input can be adjusted with the zero point correction and the
gradient correction. If an adjustment is made, you change the pre-adjusted
factory settings.
This is signaled with the display text T.OF or T.SL when switching on.
- Switch the device off.
- Hold the down button and press the On/Off button briefly to switch on the device and open the Adjustment menu.
- The display shows the first parameter. Release the down button.
Parameter | Values | Meaning |
---|---|---|
T.OF | Zero point correction | |
0.00 | No zero point correction | |
-5.00 … 5.00 | Zero point correction in °C. (at °F -9.00 .. 9.00) | |
T.SL | Gradient correction of the temperature | |
0.00 | No gradient correction of the temperature | |
-5.00 … 5.00 | Gradient correction in % |
Formula used by device
Temperature = °C: Display = (measured value – T.OF) (1 + T.SL / 100)
Temperature = °F: Display = (meas. value – 32 °F – T.OF) (1 + T.SL /
100) + 32 °F
Measurement Basics
Measuring instructions
For fast measurements of ambient air: Hold device by the outstretched arm and
sway back and forth (to fan) to accelerate the air exchange and the
temperature matching.
As soon as the measured value is quite stable, it can be read off. It may make
sense to press the hold key to “freeze” all values and therefore read them off
easily.
Possible measuring errors
Cooling / evaporation
When measuring the air temperature, the probe should be dry, otherwise the
temperature measured will be too low.
Response time
An adequate wait time must be observed for the measuring process before
reading the measured value. The response time t90 describes the time in which
the displayed measured value reached 90% of the end value.
Heat radiation (e.g. breathing air, body heat, sun)
If the device is held in the hand during the measurement, both temperature and
humidity are altered by the body heat. To minimize this influences the device
should be held as far away from the sensor as possible and a direct contact
with the exhaled air should be avoided. Most accurate measurements are
achieved, if the device is set down and the measured value (as soon as it is
stable) is read at adequate distance.
When setting the device down, ensure that it is not exposed to any other heat source or, for example, in the sun.
Operation and maintenance
Operating and maintenance notices
Note
- The device and temperature probe must be handled with care and used in accordance with the technical data. Do not throw or strike.
- Plugs and sockets must be protected from soiling.
- If the device is stored at a temperature above 50 °C, or is not used for an extended period of time, the batteries must be removed. Leaks from the batteries are avoid-ed as a result.
Battery
Battery indicator
If the empty frame in the battery display blinks, the batteries are depleted
and must be replaced. However, the device will still operate for a certain
length of time.
If the BAT display text appears in the main display, the battery voltage is no
longer adequate for operation of the device. The battery is fully depleted.
Changing battery
Danger! Danger of explosion!
Using damaged or unsuitable batteries can generate heat, which can cause the
batteries to crack and possibly explode!
Only use high-quality and suitable alkaline batteries!
Caution! Damage!
If the batteries have different charge levels, leaks and thus damage to the
device can occur.
- Only use high-quality and suitable alkaline batteries!
- Do not use different types of batteries!
- Remove depleted batteries immediately and dispose of them at a suitable collection point.
Note
Unnecessary unscrewing endangers the protection against moisture and should
therefore be avoided.
- Read the following handling instructions before replacing batteries and follow them step by step.
- If disregarded, the device could be damaged or the protection from moisture could be diminished.
- Unscrews the Phillips screws (A)and remove the cover.
- Carefully replace the two Mignon AA batteries (B). Ensure that the polarity is correct! It must be possible to insert the batteries in the correct position without using force.
- The O-ring (C) must be undamaged, clean and positioned at the intended depth.
- Fit the cover (D) on evenly. The O-ring must remain at the intended depth!
- Tighten the Phillips screws (A).
Error and system messages
Display | Meaning | Possible causes | Remedy |
---|---|---|---|
—- | Measurement far outside of the measuring range |
Temperature sensor defect
|
- Measurement outside of the measuring range
- Temperature sensor or device defect
|
- Stay within allowable measurement range
- Send in for repair
No display, unclear char- acters or no response when buttons are pressed| Battery depleted System error Device is defective|
- Battery depleted
- Error in the device
|
- Replace battery
- Send in for repair
BAT| Battery depleted| • Battery depleted| u Replace battery
ERR.1| Measuring range exceeded|
- Measurement too high
- Temperature sensor or device defect
|
- Stay within allowable measurement range
- Send in for repair
ERR.2| Measuring range is undercut|
- Measurement too low
- Temperature sensor or device defect
|
- Stay within allowable measurement range
- Send in for repair
SYS ERR| System error| • Error in the device|
- Switch device on/off
- Replace batteries
- Send in for repair
Technical data
Measuring range | -25.0 .. +70.0 °C (-13.0 .. +158.0 °F). |
---|---|
Accuracy | ± 0.5 % of measured value ± 0.5 °C |
Response time t90 | approx. 15 sec. (at air 2 m/s) |
Measuring cycle | approx. 2 measurements per second |
--- | --- |
Display | 3-line segment LCD, additional symbols, illuminated (white, luminous |
duration adjustable)
Standard functions| Min/max/hold,
alarm (optical and acoustic)
Adjustment| Offset and gradient correction
Housing| Break-proof ABS housing
| Protection rating| IP20
Dimensions LWH| 108 (142) 54 28 mm, without (with) sensor head
Weight| ~ 126 g incl. batteries
Nominal temperature| 25 °C
Operating conditions| -20 to 50 °C; 0 to 95 %RH (temporarily condensing)
Storage temperature| -20 bis 70 °C
Current supply| 2 * AA batteries (mignon)
| Current requirement| approx. 0.4 mA, approx. 2 mA with backlight
Battery life| Service life > 5000 hours with alkaline batteries (without
backlighting)
Battery indicator| 4-stage battery status indicator,
Replacement indicator for depleted batteries: “BAT”
Auto-power-OFF function| The device switches off automatically if this is
activated
Directives and standards| The devices conform to the following Directives of
the Council for the harmonization of legal regulations of the Member States:
2014/30/EU EMC Directive 2011/65/EU RoHS
Applied harmonized standards:
EN 61326-1:2013 Emission limits: Class B
Immunity according to table A.1 Additional errors: < 1 % FS
EN IEC 63000:2018
The device is intended for mobile use and/or stationary operation in the scope of the specified operating conditions without further limitations.
---|---
Disposal
Separation by material and recycling of device components and packaging must
take place at the time of disposal. The valid regional statutory regulations
and directives applicable at the time must be observed.
Note The device must not be disposed of with household waste. Return it
to us, freight prepaid. We will then arrange for the proper and
environmentally-friendly disposal.
Private end users in Germany have the possibility of dropping off the device
at the municipal collection center.
- Batteries must be removed beforehand!
- Please dispose of empty batteries at the collection points intended for this purpose
Service
Manufacturer
If you have any questions, please do not hesitate to contact us.
Calibration and adjustment service
The purpose of the calibration is to verify the precision of the measuring
device by comparing it with a traceable reference.
Both ISO calibration certificates and DAkkS calibration certificates are
available from Senseca.
Explanation
-
The ISO standard 9001 is applied for the isocalibration certificates.
-
These certificates area affordable alternative to the DAkkS calibration certificates and provide information of the traceable reference, a list of individual values and documentation.
-
The DAkkS calibration is based on DIN EN ISO/17025, the accreditation basis is recognized worldwide. These certificates offer high-quality calibration and consistently high quality. The DAkkS calibration includes any necessary adjustment with the purpose of minimizing a deviation of the measuring device.
-
The device is delivered with a test report.
This confirms that the measuring device has been adjusted and tested, without making any statement about the accuracy of a temperature sensor. -
Only the manufacturer can check the basic settings and make corrections if necessary.
- Senseca Germany GmbH Hans-Sachs-Straße 26 93128 Regenstauf GERMANY
- INFO@SENSECA.COM
- WEEE reg. no.: DE 93889386