senseca ECO122 Hand Soil Compost Thermometer Instruction Manual

August 19, 2024
senseca

Seneca ECO122 Hand Soil Compost Thermometer

About this documentation

Foreword
Read this document carefully and familiarize yourself with the operation of the device before you use it. Keep this document ready to hand and near the device so that it is available to the personnel/user for reference at all times in case of doubt. The user must have carefully read and understood the operating manual before beginning any work.

Legal notices
The liability and warranty of the manufacturer for damages and consequential damages are voided with misuse, disregarding this document, disregarding safety notices, assignment of inadequately qualified technical personnel, and arbitrary modifications of the
device. This document is entrusted to the recipient for personal use only. Any impermissible transfer, duplication, translation into other languages, or excerpts from this operating manual are prohibited. The manufacturer assumes no liability for print errors.

Further information
Software version of the device: V1.2 or later For the exact product name, refer to the type plate on the rear side of the device.

Note
For information about the software version, press and hold the ON button to switch on the device for longer than 5 seconds. The series is shown in the main display and the software version of the device is shown in the secondary display.

Safety

Explanation of safety symbols

Danger!
This symbol warns of imminent danger, which can result in death, severe bodily injury, or severe property damage in case of non-observance.

Caution!
This symbol warns of potential dangers or harmful situations, which can cause damage to the device or the environment in case of non-observance.

Note
This symbol indicates processes, which can have a direct influence on operation or can trigger an unforeseen reaction in case of non-observance.

Foreseeable misuse
The fault-free function and operational safety of the device can only be guaranteed if applicable safety precautions and the device-specific safety instructions for this document are observed. If these notices are disregarded, personal injury or death, as well as property damage can occur.

Danger! Incorrect area of application!
To prevent erratic behavior of the device, personal injury, and property damage, the device must be used exclusively as described under intended use.

  • Do not use it in safety. Emergency Stop devices!
  • The device is not suitable for use in explosion-prone areas!
  • The device must not be used for diagnostic or other medical purposes on patients!
  • Not suitable for use with requirements on functional safety, e.g. SIL!

Safety instructions

Caution! Risk of injury at the temperature sensor! Devices with insertion probes entail the risk of stab injuries due to the pointed probe design.

  • Handle insertion probes with care!
  • Fit a protective cap on the measuring probe if the probe is not in use!
  • Do not put the device down with the sensor tube pointing upwards. Empty batteries and batteries of inferior quality can leak more easily, which can destroy the device.
  • Please also observe the instructions in the chapter “Operation and Maintenance”.

Caution!
There is an immediate danger to life if live parts come into contact with the temperature sensor!

  • Therefore, do not conduct measurements on or in the direct vicinity of live parts or lines.
  • Never plug the temperature sensor into sockets, etc.

Note

  • This device does not belong in children’s hands!
  • The device and sensor are not designed for continuous contact with foods.

Intended use
The device is a water-protected thermometer for precise temperature measurements . The special design of the ECO 122 makes it a convenient insertion thermometer for:

  • soft soils
  • compost
  • agricultural windrows
  • bulk material

Qualified personnel
For commissioning, operation, and maintenance, the relevant personnel must have adequate knowledge of the measuring process and the significance of the measurements. The instruc-tions in this document must be understood, observed, and followed. To avoid any risks arising from the interpretation of the measurements in the concrete application, the user must have additional expertise. The user is solely liable for damages/danger resulting from misinterpretation due to inadequate expertise.

The device at a glance

Display elements

Connections

  • Probe Permanently connected sensor

Operating elements

Operation

Opening the configuration menu

  1. Press the Function key for 2 seconds to open the Configuration menu.
  2. (ONF appears in the display. Release the Function key.
Parameter Values Meaning
L, TE Backlight
  OFF Backlight deactivated
  15, 30, 60, 120, 240 Automatic shut-off of the backlight after a selected

time in seconds, during which no buttons have been pressed
 | ON| No automatic shut-off of the backlight
UN, T| Display unit
 | °(| Temperature display in °C
 | °F| Temperature display in °F
IN, T| Factory settings
 | NO| Use current configuration
 | YES| Reset the device to factory settings.

After confirming with the function button, the display shows: IN, T DONE

Adjustment of the measuring input
The temperature input can be adjusted with the zero point correction and the gradient correction. If an adjustment is made, you change the pre-adjusted factory settings. This is signaled with the display text T.OF or T.SL when switching on.

  1. Switch the device off.
  2. Hold the down button and press the On/Off button briefly to switch on the device and open the Adjustment menu.
  3. The display shows the first parameter. Release the down button.
Parameter Values Meaning
T.SL Gradient correction of the temperature
  0.00 No gradient correction of the temperature
  -5.00 … 5.00 Gradient correction in %

The formula used by the device

  • Temperature = °C: Display = (measured value – T.OF) (1 + T.SL / 100) Temperature = °F: Display = (meas. value – 32 °F – T.OF) (1 + T.SL / 100) + 32 °F
  • Adjustment example: Temperature unit= °C, the adjustment is carried out in 2 separate steps at 0 °C (e.g. ice water) and at a reference temperature (e.g. clinical thermometer with water bath 37 °C).
    • First, set the values of t.oF and t.SL in the adjustment menu to 0

Zero point

  • Set the temperature probe to 0 °C and let it adjust.
  • Start the adjustment menu and enter the display value at 0 °C for t.oF
  • After leaving the menu, the device should now display 0.0 °C.

Slope

  • Set the temperature sensor to the reference temperature and let it adjust.
  • Calculate slope correction:
  • Start the adjustment menu and enter the calculated value at t.SL
  • Exit the menu, the device should now display the reference temperature.

Measurement Basics

Sensor/device accuracy
Penetration into the medium to be measured should be done without exerting much force. Try to exert the force on the axis of the probe tube to avoid creating a lever effect on the housing. Never use tools (e.g. a hammer) to knock the sensor tube into the ground via the device housing! If you encounter strong resistance when piercing (high compaction, stones), please try to pierce at another suitable place in order not to damage the tip or the device. If this is not possible, you can pierce an appropriate hole with another rod (e.g. reinforcing iron).

Possible measuring errors

Immersion depth

  • Liquid, soil:
    • to a depth of at least 100 mm (and then stir). Otherwise, measuring errors can occur due to the heat transmission of the sensor tube if the immersion depth is too shallow.

Surface effects and poor heat transfer
Special measuring sensors are required for this purpose. Surface characteristics, de-sign of the measuring sensor, heat transfer, and environmental temperature influence the measurement result.

Note
Thermally conductive paste between the measuring sensor and surface can also in-crease measurement accuracy in some cases.

Cooling/evaporation
When measuring the air temperature, the probe should be dry, otherwise the temperature measured will be too low.

Response time
An adequate wait time must be observed for the measuring process before reading the measured value. The response time t90 describes the time in which the displayed measured value reached 90% of the end value.

Operation and maintenance

Operating and maintenance notices

Note

  • The device and temperature probe must be handled with care and used by the technical data. Do not throw or strike.
  • Plugs and sockets must be protected from soiling.
  • If the device is stored at a temperature above 50 °C or is not used for an extended period, the batteries must be removed. Leaks from the batteries are avoided as a result.

Battery

Battery indicator
If the empty frame in the battery display blinks, the batteries are depleted and must be replaced. However, the device will still operate for a certain length of time. If the BAT display text appears in the main display, the battery voltage is no longer adequate for the operation of the device. The battery is fully depleted.

Changing battery

Danger! The danger of explosion!
Using damaged or unsuitable batteries can generate heat, which can cause the batteries to crack and possibly explode!

  • Only use high-quality and suitable alkaline batteries!

Caution! Damage!
If the batteries have different charge levels, leaks and thus damage to the device can occur.

  • Only use high-quality and suitable alkaline batteries!
  • Do not use different types of batteries!
  • Remove depleted batteries immediately and dispose of them at a suitable collection point.

Note
Unnecessary unscrewing endangers the protection against moisture and should therefore be avoided.

  • Read the following handling instructions before replacing batteries and follow them step by step.
  • If disregarded, the device could be damaged or the protection from moisture could be diminished.
  1. Unscrews the Phillips screws (A)and remove the cover.
  2. Carefully replace the two Mignon AA batteries (B). Ensure that the polarity is correct! It must be possible to insert the batteries in the correct position without using force.
  3. The O-ring (C) must be undamaged, clean and positioned at the intended depth.
  4. Fit the cover (D) on evenly. The O-ring must remain at the intended depth!
  5. Tighten the Phillips screws (A).

Error and system messages

Display Meaning Possible causes Remedy
—- Measurement far outside of the measuring range

Probe defect

| •   Measurement outside of the measuring range

•   Measuring probe or device defect

|

  • Stay within allowable measurement range
  • Send in for repair

No display, unclear char- acters or no response when buttons are pressed| Battery depleted System error Device is defective| •   Battery depleted

•   Error in the device

|

  • Replace battery
  • Send in for repair

BAT| Battery depleted| •   Battery depleted|

  • Replace battery

ERR.1| Measuring range exceeded| •   Measurement too high

•   Measuring probe or device defect

|

  • Stay within allowable measurement range
  • Send in for repair

ERR.2| Measuring range is undercut| •   Measurement too low

•   Measuring probe or device defect

|

  • Stay within allowable measurement range
  • Send in for repair

SYS ERR| System error| •   Error in the device|

  • Switch device on/off
  • Replace batteries
  • Send in for repair

Technical data

Measuring range -70.0 .. +250.0 °C (-94.0 .. +482.0 °F)
Accuracy -20 .. +100 °C:      ± 0.1 K ± 1 Digit

-70 .. +250 °C:      ± 0.2 % of measured value ± 2 digits

Probe| Pt1000,

Probe tube firmly screwed under- neath, similar T-handle

|
 | Probe tube| Ø 8 x 920 mm, stainless steel
Probe tip|
Response time t90| approx.. 4 s (water (0,4 m/s))
Measuring cycle| approx. 2 measurements per second
Display| 3-line segment LCD, additional symbols, illuminated (white, luminous duration adjustable)
Standard functions| Min/max/hold,

alarm (optical and acoustic)

Adjustment| Offset and gradient correction
Housing| Break-proof ABS housing
 | Protection rating| IP54
Dimensions LWH| 108 54 28 mm, device without probe tube
Weight| ~ 285 g incl. batteries
Nominal temperature| 25 °C
Operating conditions| -20 to 50 °C; 0 to 95 %RH (temporarily condensing)
Storage temperature| -20 bis 70 °C
Current supply| 2 * AA batteries (mignon)
 | Current requirement| approx. 0.4 mA, approx. 2 mA with backlight
Battery life| Service life > 5000 hours with alkaline batteries (without backlighting)
Battery indicator| 4-stage battery status indicator,

Replacement indicator for depleted batteries: “BAT”

Auto-power-OFF function| The device switches off automatically if this is activated
Directives and standards| The devices conform to the following Directives of the Council for the harmonization of legal regulations of the Member States:

2014/30/EU EMC Directive 2011/65/EU RoHS

Applied harmonized standards:

EN 61326-1:2013     Emission limits: Class B

Immunity according to table A.1 Additional errors: < 1 % FS

EN IEC 63000:2018

The device is intended for mobile use and/or stationary operation in the scope of the specified operating conditions without further limitations.

---|---

Disposal

Separation by material and recycling of device components and packaging must take place at the time of disposal. The valid regional statutory regulations and directives applicable at the time must be observed.

Note
The device must not be disposed of with household waste. Return it to us, freight prepaid. We will then arrange for the proper and environmentally friendly disposal. Private end users in Germany have the possibility of dropping off the device at the municipal collection center.

  • Batteries must be removed beforehand!
  • Please dispose of empty batteries at the collection points intended for this purpose

Service

Manufacturer
If you have any questions, please do not hesitate to contact us.

Calibration and adjustment service
The purpose of the calibration is to verify the precision of the measuring device by comparing it with a traceable reference. Both ISO calibration certificates and DAkkS calibration certificates are available from Senseca.

Explanation

  • The ISO standard 9001 is applied for the iso-calibration certificates.
  • These certificates are affordable alternatives to the DAkkS calibration certificates and provide information on the traceable reference, a list of individual values, and documentation.
  • The DAkkS calibration is based on DIN EN ISO/17025, and the accreditation basis is recognized worldwide. These certificates offer high-quality calibration and consistently high quality. The DAkkS calibration includes any necessary adjustment to minimize a deviation of the measuring device.
  • The device is delivered with a test report. This confirms that the measuring device has been adjusted and tested, without making any statement about the accuracy of a temperature sensor.
  • Only the manufacturer can check the basic settings and make corrections if necessary.
  • senseca.com
  • Senseca Germany GmbH
  • Hans-Sachs-Straße 26
  • 93128 Regenstauf
  • GERMANY
  • INFO@SENSECA.COM
  • WEEE reg. no.: DE 93889386

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals