senseca ECO 230 Barometer Instruction Manual
- August 7, 2024
- senseca
Table of Contents
ECO 230 Barometer
“`html
Specifications:
- Model: ECO 230 barometer | altimeter
- Model Number: B-H88.0.01.DK2-2.0
Product Information:
About this documentation
The operating manual for the ECO 230 barometer | altimeter
provides detailed information on the purpose of the document, legal
notices, and further information.
Safety
The manual explains safety symbols, foreseeable misuse, safety
instructions, intended use, and the need for qualified personnel to
operate the device safely.
The product at a glance
Provides an overview of display elements and operating elements
of the ECO 230 barometer | altimeter.
Operation
Instructions on opening the configuration menu and adjusting
settings through the adjustment menu.
Measurement Basics
General information about absolute pressure measurement,
altimeter function, and special functions like NVLL Tare
function.
Product Usage Instructions:
Safety Instructions:
- Read the manual carefully before use.
- Ensure proper handling to prevent damage.
- Do not use the device for purposes other than specified.
Operating Elements:
Refer to the manual for specific details on each operating
element and its function.
Configuration Menu:
-
Press and hold the designated button to open the configuration
menu. -
Use the navigation keys to scroll through options.
-
Select desired settings by pressing the confirmation
button.
Adjustment Menu:
-
To access the adjustment menu, follow the steps outlined in the
manual. -
Make necessary adjustments using the provided controls.
-
Confirm changes before exiting the menu.
Altimeter Function:
For altitude measurements, ensure proper calibration as per the
manual instructions.
FAQ:
Q: How do I calibrate the altimeter?
A: Refer to section 5.2 in the manual for detailed instructions
on calibrating the altimeter function.
“`
OPERATING MANUAL
ECO 230 barometer | altimeter
B-H88.0.01.DK2-2.0
ECO 230 barometer | altimeter
Table of contents
Table of contents
1
About this documentation ………………………………………………………………………………………….4
1.1 Purpose of the document………………………………………………………………………………………………….. 4
1.2 Legal notices………………………………………………………………………………………………………………………… 4
1.3 Further information …………………………………………………………………………………………………………… 4
2
Safety …………………………………………………………………………………………………………………………….5
2.1 Explanation of safety symbols …………………………………………………………………………………………. 5
2.2 Foreseeable misuse……………………………………………………………………………………………………………. 5
2.3 Safety instructions……………………………………………………………………………………………………………… 6
2.4 Intended use ……………………………………………………………………………………………………………………….. 6
2.5 Qualified personnel……………………………………………………………………………………………………………. 6
3
The product at a glance ……………………………………………………………………………………………….7
3.1 Display elements ………………………………………………………………………………………………………………… 7
3.2 Operating elements …………………………………………………………………………………………………………… 8
4
Operation ……………………………………………………………………………………………………………………..9
4.1 Opening the configuration menu …………………………………………………………………………………….. 9
4.2 Open the adjustment menu …………………………………………………………………………………………… 11
5
Measurement Basics …………………………………………………………………………………………………12
5.1 General information about absolute pressure measurement …………………………………. 12
5.2 Altimeter …………………………………………………………………………………………………………………………… 12
5.3 Special functions ……………………………………………………………………………………………………………… 12 5.3.1 NVLL Tare function ……………………………………………………………………………………………. 12
5.4 Use of the trend indicator………………………………………………………………………………………………. 13 5.4.1 Meteorology: weather forecasting……………………………………………………………….. 13 5.4.2 Hiking, cycling, flying, motor sports: use as a variometer…………………………. 14
6
Operation and maintenance…………………………………………………………………………………….15
6.1 Battery ………………………………………………………………………………………………………………………………. 15 6.1.1 Battery indicator ……………………………………………………………………………………………… 15 6.1.2 Changing battery……………………………………………………………………………………………… 15
B-H88.0.01.DK2-2.0
Page 2 of 20
ECO 230 barometer | altimeter
Table of contents
7
Disposal ………………………………………………………………………………………………………………………16
8
Error and system messages………………………………………………………………………………………17
9
Technical data ……………………………………………………………………………………………………………18
B-H88.0.01.DK2-2.0
Page 3 of 20
ECO 230 barometer | altimeter
1 About this documentation
About this documentation
1.1 Purpose of the document
Read this document carefully and familiarize yourself with the operation of
the product before you use it. Keep this document ready to hand and in the
immediate vicinity of the product so that it is available to the
personnel/user for reference at all times in case of doubt. The user must have
carefully read and understood the operating manual before beginning any work.
1.2 Legal notices
The liability and warranty of the manufacturer for damages and consequential
damages are voided with misuse, disregarding this document, disregarding
safety notices, assignment of inadequately qualified technical personnel and
arbitrary modifications of the product. This document is entrusted to the
recipient for personal use only. Any impermissible transfer, duplication,
translation into other languages or excerpts from this operating manual are
prohibited. The manufacturer assumes no liability for print errors.
1.3 Further information Software version of the product: · V1.7 or later For
the exact product name, refer to the type plate on the rear side of the
product.
Note For information about the software version, press and hold the ON button
to switch on for longer than 5 seconds. The series is shown in the main
display and the software version of the product is shown in the secondary
display.
B-H88.0.01.DK2-2.0
Page 4 of 20
ECO 230 barometer | altimeter
2 Safety
Safety
2.1 Explanation of safety symbols
Danger! This symbol warns of imminent danger, which can result in death,
severe bodily injury, or severe property damage in case of non-observance.
Caution!
This symbol warns of potential dangers or harmful situations, which can cause
damage to the device or to the environment in case of non-observance.
Note Blue underlining indicates processes, which can have a direct influence
on operation or can trigger an unforeseen reaction in case of non-observance.
2.2 Foreseeable misuse
The fault-free function and operational safety of the product can only be
guaranteed if applicable safety precautions and the device-specific safety
instructions for this document are observed. If these notices are disregarded,
personal injury or death, as well as property damage can occur.
Danger Incorrect area of application!
In order to prevent erratic behavior of the product, personal injury and
property damage, the product must be used exclusively as described in the
chapter De-scription in the operating manual. The product is not suitable for
use in explosion-prone areas! The product must not be used for diagnostic or
other medical purposes on patients! For measurements requiring devices that
are subject to authorization or special
approvals, this product is not a substitute for such products and can only be
used as an aid in preparatory or comparison measurements! Do not use in safety
/ emergency stop devices!
B-H88.0.01.DK2-2.0
Page 5 of 20
ECO 230 barometer | altimeter
2.3 Safety instructions Note This product does not belong in children’s hands!
Safety
2.4 Intended use
The device measures the absolute pressure in the air. The ambient pressure is
measured directly via the integrated sensor. Pressure equalization between the
unit and the environment takes place via a water-impermeable membrane on the
front side. Applications include:
· Barometric measurements (e.g. weather)
· Elevation determination
The device must only be used under the conditions and for the purposes for
which it was designed. It must be handled with care and used according to the
technical data (do not throw, strike, etc.). Suitable measures must be used to
protect the pressure equalization opening and be protected from dirt.
2.5 Qualified personnel
For commissioning, operation and maintenance, the relevant personnel must have
adequate knowledge of the measuring process and the significance of the
measurements. The instructions in this document must be understood, observed
and followed. In order to avoid any risks arising from interpretation of the
measurements in the concrete application, the user must have additional
expertise. The user is solely liable for damages/danger resulting from
misinterpretation due to inadequate expertise.
B-H88.0.01.DK2-2.0
Page 6 of 20
ECO 230 barometer | altimeter
3 The product at a glance
The product at a glance
LCD Display
Top view with pressure equalization
opening
Front view
3.1 Display elements
Display
Battery indicator Unit display Main display
Auxiliary display
Bar display
Evaluation of the battery status
Display of the units or Min/Max/Hold information text
Measurement of the current pressure or value for min/max/hold
Measurement of the current pressure in Min/Max/Hold mode
Trend display in 7 steps
/ / falling (strong / medium / slight)
stable
/ / rising (slight / medium / strong)
B-H88.0.01.DK2-2.0
Page 7 of 20
ECO 230 barometer | altimeter
3.2 Operating elements On / Off button Press briefly Long press
The product at a glance
Switch on the device Activate / deactivate lighting Switch off the device
Reject changes in a menu
Up / Down button
Press briefly Long press Both simultaneously
Display of the min/max value Change value of the selected parameter Reset the
min/max value of the current measurement
Rotate display, overhead display
Function button
Press briefly
Freeze measurement (Hold) Call up next parameter
Long press, 2s
Start menu “configuration”, (ONF appears in the display
Long press, 4s
Activation of the Tare function NVLL / altitude input
Operating status
device is in measured value display device is in the configuration menu
B-H88.0.01.DK2-2.0
Page 8 of 20
ECO 230 barometer | altimeter
4 Operation
4.1 Opening the configuration menu 1 Press the Function key for 2 seconds to open the Configuration menu. 2 (ONF appears in the display. Release the Function key.
Parameter Values
Meaning
Operation
UN,T
RATE SEA.L ALT, TEND
Display unit
HPa
Barometer in [hPa]
mBaR
Barometer in [mbar]
PSI
Barometer in [psi]
mmKG
Barometer in [mmHg]
m
Altimeter/Elevation display in [m]
FT
Altimeter/Elevation display in [feet]
Measuring rate
SLO
Slow measurement speed
FAST
Fast measurement speed
(not recommended for altimeter display)
SeaLevel-correction (not available at altimeter/elevation display)
NO
Inactive, display measured air pressure directly
YES
Active, display air pressure compensated to sea level
High above sea level
-500 .. 9000
Height above sea level in m for correction At altimeter: default value for altitude input
Trend display OFF
Bar display and tendency value display deactivated
B-H88.0.01.DK2-2.0
Page 9 of 20
ECO 230 barometer | altimeter
Operation
Parameter Values
Meaning
I_S / 60_S
Time base for bar display 1 second (0.2 hPa / bar level corresponds to ~1.7 m / 5 ft at sea level) or 60 seconds (12 hPa / bar level corresponds to ~100 m / 328 ft at sea level)
I_H / 3_H
Time base for bar display 1 hour (1 hPa / bar level) or 3 hours (3 hPa / bar
level) for usual meteorological assessments.
The value will be renew every minute.
LCD.2
Additional information
OFF
No additional information in auxiliary display
°(
Temperature in °C
°F
Temperature in °F
With activated TEND display additionally:
TEND
Trend value
°(.TE °F.TE
Temperature in °C and trend value Temperature in °F and trend value
°(.P.T
Temperature in °C, trend value and air pressure (only available at diameter display units)
°F.P.T
Temperature in °F, trend value and air pressure (only available at diameter display units)
POFF
Shut-off time
OFF
No automatic shut-off
0:15, 0:30, 1:00, 4:00, 12:00
Automatic shut-off after a selected time in hours:minutes, during which no
buttons have been
pressed
L,TE
Backlight
OFF
Backlight deactivated
0:15, 0:30, 1:00, 2:00, 4:00 Automatic shut-off of the backlight after a
select-ed
time in minutes:seconds, during which no buttons have been pressed
ON
No automatic shut off of the backlight
B-H88.0.01.DK2-2.0
Page 10 of 20
ECO 230 barometer | altimeter
Parameter Values
IN,T
Factory settings
NO
YES
Meaning
Operation
Use current configuration
Reset device to factory settings. After confirming with the function-button,
the display shows: IN,T DONE
4.2 Open the adjustment menu
The sensor adjustment can be adjusted with the zero point correction and the
gradient correction. If an adjustment is made, you change the pre-adjusted
factory settings. This is signaled with the display text PR.OF or PR.SL when
switching on.
1 Switch the device off. 2 Hold the down button and press the On/Off button to
switch on the device and open the
Adjustment menu.
3 The display shows the first parameter. Release the down button.
Parameter Values
Meaning
PR.OF
Zero point correction
0.0
No zero point correction
-5.0 … 5.0
Zero point correction [in selected display unit] (e.g. ± 5.0 hPa)
PR.SL
Gradient correction
0.00
No gradient correction of the temperature
-5.00 … 5.00 Gradient correction in %
Formula: Zero point correction: Displayed value = measured value – PR.OF Gradient correction: Display = (measured value – PR.OF) * (1 + PR.SL / 100)
B-H88.0.01.DK2-2.0
Page 11 of 20
ECO 230 barometer | altimeter
5 Measurement Basics
Measurement Basics
5.1 General information about absolute pressure measurement
The device measures absolute pressure. However, this should not be confused
with the “sea level air pressure” indicated by weather stations. The altitude-
based air pressure decrease is calculated for these pressure specifications.
The device is capable of correcting this air pressure altitude correction. To
do this, set the parameter value SEA.L in the Configuration menu to YES and
enter the current altitude above sea level in the parameter value ALT,. The
nautical norm correction must be deactivated for vacuum measurements.
5.2 Altimeter
If meters or feet are selected as the display unit in the Configuration menu, the device is in altimeter mode. The unit calculates the current altitude from the current air pressure: At higher altitudes, the air pressure is lower. It is important to note that not only changes in altitude but also the weather influence the measurement. To correct this weather influence, the displayed altitude can be corrected via the keys.
Altimeter display: In the Altimeter operating mode, you will be prompted to enter the
current altitude after pressing the Function key for 4 s.
Note:
The first time you are prompted after switching on the device, the value is set to the alti value in the Configuration menu. (default value for starting point)
The current value is retained for all further calls.
Note on min/max display: in the secondary display, the sum of the altitude meters covered is also shown for min: dESC – descent, for max: ASC – ascent
5.3 Special functions
5.3.1 NVLL Tare function
If a pressure unit is selected as the display unit in the Configuration menu,
the device is in barometer mode. After it is switched on, the device starts up
in standard measuring mode, the special function “tare” is started by pressing
and holding the Function key for 4 s..
Barometer display:
The display can be zeroed by pressing the Function key for 4 s. If the tare
function is activated, NVLL blinks in the lower dis-play. The tare
function can be reset by pressing the Function key again for 4 s.
B-H88.0.01.DK2-2.0
Page 12 of 20
ECO 230 barometer | altimeter
Measurement Basics
Note The tare function is independent of the zero point correction accessible via the adjustment menu.
5.4 Use of the trend indicator
Note For the trend display, it is necessary that the device has collected a
sufficiently large number of measured values. This is the case for the time
bases 1h and 3h only after this time has elapsed. The value is therefore not
shown in the display before this time has elapsed.
5.4.1 Meteorology: weather forecasting
Observation of variable weather conditions by assessing the rate of change of air pressure. To do this, set the device to:
UN,T
HPa
RATE
SLO
(international standard) (power saving during continuous operation)
SEA.L
YES
ALT,
altitude of the location above sea level
TEND
1_H or 3_H
1 hour (1 hPa / bar level) or 3 hours (3 hPa / bar level)
LCD.2
°(.TE
POFF
OFF
continuous operation
This results in the following display, for example: display unit hPa current pressure at sea level trend: + 0.4 hPa per hour stable (trend lower 1 hPa)
B-H88.0.01.DK2-2.0
Page 13 of 20
ECO 230 barometer | altimeter
Measurement Basics
5.4.2 Hiking, cycling, flying, motor sports: use as a variometer
A variometer or inclinometer indicates the change in altitude per unit of
time, i.e. the rate of ascent or descent. Common units are [ft/min] or [m/s]
for flight and motor sports, or [m/h] for example in hiking/running.
Note The output value of the tendency display in the setting 1_S or 60_S shows
the current value for the tendency based on the measurements of the last 5
seconds, scaled to the selected setting. In the 60_S setting, the resolution
is reduced to 1 m or 5 ft.
To do this, set the device to:
Hiking, cycling
flying,
motor sports
UN,T m RATE SLO
FT or m FAST
(Altimeter)
(power saving during continuous operation) (fast measured value response) *1
ALT, TEND 1_H
default value for input start
1_S or 60_S (time base 1 second for m/s or 60 seconds for ft/min)
LCD.2 TEND
TEND
or for more information: °(.P.T or °F.P.T
POFF OFF
OFF
continuous operation
This results in the following display, for example: display unit m or m/s
current altitude above sea level
trend: – 2.8 m per s
slightly falling * = Note: brief influence on the measured value possible when
pressing buttons or housing
disc
B-H88.0.01.DK2-2.0
Page 14 of 20
ECO 230 barometer | altimeter
6 Operation and maintenance
Operation and maintenance
6.1 Battery
6.1.1 Battery indicator If the empty frame in the battery display blinks, the
batteries are depleted and must be replaced. However, the device will still
operate for a certain length of time. If the BAT display text appears in the
main display, the battery voltage is no longer adequate for operation of the
product. The battery is fully depleted.
6.1.2 Changing battery
Danger Danger of explosion! Using damaged or unsuitable batteries can generate
heat, which can cause the batteries to crack and possibly explode! Only use
high-quality and suitable alkaline batteries!
Caution Damage! If the batteries have different charge levels, leaks and thus
damage to the device can occur. Only use high-quality and suitable alkaline
batteries! Do not use different types of batteries! Remove depleted batteries
immediately and dispose of them at a suitable collection
point.
Note Unnecessary unscrewing endangers the protection against moisture and
should therefore be avoided.
Note Read the following handling instructions before replacing batteries and
follow them step by step. If disregarded, the device could be damaged or the
protection from moisture could be diminished.
B-H88.0.01.DK2-2.0
Page 15 of 20
ECO 230 barometer | altimeter
Disposal
1 Unscrews the Phillips screws (A)and remove the cover.
2 Carefully replace the two Mignon AA batteries (B). Ensure that the polarity
is correct! It must be possible to insert the batteries in the correct
position without using force.
3 The O-ring (C) must be undamaged, clean and positioned at the intended
depth.
4 Fit the cover (D) on evenly. The O-ring must remain at the intended depth!
5 Tighten the Phillips screws (A).
7 Disposal
Separation by material and recycling of device components and packaging must
take place at the time of disposal. The valid regional statutory regulations
and directives applicable at the time must be observed.
Note The device must not be disposed of with household waste. Return it to us,
freight prepaid. We will then arrange for the proper and environmentally
friendly disposal.
Private end users in Germany have the possibility of dropping off the device
at the municipal collection center. Batteries must be removed beforehand!
Please dispose of empty batteries at the collection points intended for this
purpose.
B-H88.0.01.DK2-2.0
Page 16 of 20
ECO 230 barometer | altimeter
Error and system messages
8 Error and system messages
Display
Meaning
Possible causes
Remedy
—-
Calculation not
· Measurement data Waiting for data
possible
acquisition is running collection
No display, unclear Battery depleted ·
characters or no System error
·
response when buttons are pressed
Product is defective
·
Battery depleted Error in the product Product is defective
Replace battery Send in for repair
BAT
Battery depleted · Battery depleted
Replace battery
ERR.1
Measuring range exceeded
· Measurement too high
· Product is defective
Stay within allowable measurement range
Send in for repair
ERR.2
Measuring range is · Measurement too
undercut
low
· Product is defective
Stay within allowable measurement range
Send in for repair
ERR.3
Display range has been exceeded
· Incorrect display unit Correct setting · Value not displayable
ERR.4
Display range has been undercut
· Incorrect display unit Correct setting · Value not displayable
SYS ERR
System error
· Error in the product
Switch product on/off
Replace batteries
Send in for repair
B-H88.0.01.DK2-2.0
Page 17 of 20
ECO 230 barometer | altimeter
9 Technical data
Measuring Barometer range
Altimeter Temperature Accuracy Barometer Altimeter Temperature Overload
Measuring medium Pressure connection Measuring cycle Display Standard function
Additional functions Adjustment
Technical data
300.0 .. 1100.0 hPa (mbar) abs. 4.350 .. 15.950 PSI abs. 225.0 .. 825.0 mmHg
(Torr) abs. -500.0 .. 9000.0 m -1640 .. 19999 ft. -20.0 .. 50.0 °C -4.0 ..
122.0 °F ± 1 hPa typical (at T: 0 .. 30 °C) ± 0.25 % FSS max. corresponds ± 2
hPa typ. ± 1 m relative (over a short period at constant ambient pressure @
25°C) typ. ± 0.5 °C @ 25°C 4000 hPa abs. Air No connection, integrated sensor
Pressure equalization via diaphragm-protected opening FAST: approx. 2.5
measurements per second SLO: approx. 4 seconds 3-line segment LCD, additional
symbols, illuminated (adjustable white, permanent illumination) Min/Max/Hold
Trend display, time base selectable (1s, 60s, 1h, 3h) corresponds to Altimeter
Variometer display with units m/s, ft/min, m/h, ft/h Only with altimeter
function: The altitude meters covered are calculated (ascent AS(, descent
DES(, resolution 1m)
NVLL: Tare function
Zero point and gradient adjustment
B-H88.0.01.DK2-2.0
Page 18 of 20
ECO 230 barometer | altimeter
Technical data
Housing
Break-proof ABS housing
Protection rating
IP67
Measurements
108 54 28 mm
Weight
140 g, incl. batteries
Nominal temperature
25 °C
Operating conditions
-20 bis 50 °C; 0 to 95 %RH (short-term condensation possible)
Storage temperature
-20 bis 70 °C
Current supply
2 * AA batteries (mignon)
Current requirement
approx. 0.4 mA, approx. 2.4 mA with backlight
(with slow measuring range)
battery life
Service life with alkaline batteries: approx. 6000 hours (without backlighting and with measuring rate = Slo)
Battery indicator
4-stage battery status indicator, Replacement indicator for depleted batteries: “BAT”
Auto-power-OFF function The device switches off automatically if this is activated
Directives and standards
The devices conform to the following Directives of the Council for the
harmonization of legal regulations of the Member States:
2014/30/EU EMC Directive 2011/65/EU RoHS
Applied harmonized standards: EN 61326-1:2013 Emission limits: Class B
Immunity acc. to Table 1 Additional errors: < 1 % FS EN IEC 63000:2018
The device is intended for mobile use and/or stationary operation in the scope
of the specified operating conditions without further limitations.
B-H88.0.01.DK2-2.0
Page 19 of 20
senseca.com
Senseca Germany GmbH Hans-Sachs-Straße 26
93128 Regenstauf GERMANY
INFO@SENSECA.COM WEEE reg. no.: DE 93889386