GETINGE Sanizyme Neutral Solution For Sterile Processing Instructions

July 17, 2024
GETInGE

Sanizyme Neutral Solution For Sterile Processing

Product Information

Specifications:

  • Product Name: Sanizyme+
  • Chemistries: pH-neutral triple-enzyme liquid detergent
  • Enzymes: Protease, Lipase, Amylase
  • Formulation: Translucent at all doses and temperatures
  • Order Numbers: 6036005201, 6036005301, 6036005401

Product Usage Instructions:

Directions for Use in Ultrasonics:

  1. Fill the ultrasonic cleaner with Sanizyme+ solution.

  2. Place the items to be cleaned in the ultrasonic cleaner.

  3. Run the ultrasonic cleaning cycle for sufficient time for
    desired cleaning.

  4. For heavily soiled items, longer cleaning times may be
    required.

  5. After cleaning, rinse all items thoroughly, ensuring any lumens
    are also flushed.

  6. Dispose of the solution according to facility policy; do not
    reuse visibly soiled solutions.

Storage Information:

Sanizyme+ is compatible with a variety of materials including
but not limited to Aluminum, Stainless Steel, Glass, Ceramics,
Rubber, Plastic, and TeflonTM/PTFE. Pay special attention to
guidelines from instrument/device manufacturers.

Caution:

Dispose of contents and containers following local, regional,
national, and international regulations. Sanizyme+ contains
triethanolamine, protease, amylase, and lipase enzymes.

Manufacturer Information:

Manufacturer: Quadralene Ltd, United Kingdom. Sales Offices in
US and Canada. Visit www.getinge.com for more details.

FAQ:

Q: Can Sanizyme+ be used for cleaning flexible scopes?

A: Yes, Sanizyme+ is suitable for use with flexible scopes along
with other listed instruments.

Q: Is it safe to reuse the cleaning solution?

A: Solutions should not be reused if visibly soiled to ensure
effective cleaning and hygiene standards.

Sanizyme+

Chemistries for the ultimate peace of mind

Instructions for Use

Orthopedics

Robotic Instruments

Flexible Scopes

Rigid Scopes

Ultrasonic

XEN Sanizyme+ is a pH-neutral triple-enzyme liquid detergent for use in ultrasonic cleaning. Sanizyme+ is low foaming to prevent clogging of instruments or machines and interference with ultrasonics. Its three enzymes — protease, lipase, and amylase — break down blood and proteins, fats and oils (lipids), and carbohydrates, respectively.
Following Getinge’s integration of Ultra Clean Systems, Sanizyme+ is a recharged formulation of the well-known Sanizyme chemistry. Sanizyme+ is translucent at all doses and temperatures for enhanced visibility.

Order no. 6036005201 6036005301 6036005401

Description Sanizyme+ 1 x 5 L. Sanizyme+ 1 x 5 gal. Sanizyme+ 1 x 15 gal.

Directions for Use in Ultrasonics
1. Fill the ultrasonic bath to an appropriate level with warm water (for optimal results, wash temperature should be set at 77°F­113°F/25°C­45°C, or in accordance with your equipment manufacturer IFU), ensuring the quantity of water used allows the correct dose to be calculated. Dosing is 1.02 oz./gal. (8 ml/l), depending on soil load and water quality. Higher doses will be required for higher soil loads and for poor water quality.

2. Run the ultrasonic cleaning cycle, ensuring sufficient time to obtain the desired cleaning. Longer times cleaning may be required for heavily soiled items.
3. After cleaning, rinse all items thoroughly, ensuring any lumens are also flushed thoroughly.
4. Discard the solution in accordance with facility policy. Solutions should not be reused if they are visibly soiled.

Caution
· Do not mix with other chemicals.
· Wear correct PPE and follow applicable guidelines. Additional information on SDS.
· Check material compatibility before use.
Dispose of contents and container in accordance with local, regional, national, and international regulations.
Contains triethanolamine, protease, amylase, and lipase enzymes. For more ingredient information, visit www.getinge.com

Storage Information
Store in a cool, dry place protected from frost and away from acids and strong oxidizing agents. Store upright in original containers. Optimum storage temperature 41°F­77°F/5°C­25°C.

Aluminum Anodized Aluminum Titanium Copper Brass Gold Nickel Cobalt Chrome Matte chrome plated Stainless steel Glass Ceramics Rubber Plastic TeflonTM/ PTFE Polycarbonate

Sanizyme+

················

· = The product is suitable for the material
= Pay special attention to the guidelines from the instruments/device manufacturer
TeflonTM is a trademark of Chemours.

Getinge is a registered trademark of Getinge AB or its subsidiaries or affiliates in the United States or other countries. Copyright 2024 Getinge AB or its subsidiaries or affiliates. · All rights reserved.
Manufacturer: Quadralene Ltd, Unit 3, St. Modwen Park, Andressey Way, Derby, DE21 6YH, United Kingdom. Sales Office, US: 1 Geoffrey Way, Wayne, NJ 07470, USA Sales Office, Canada: 90 Matheson Blvd West, Suite 300, Mississauga, Ontario L5R 3R3, Canada
www.getinge.com

MCV00110915 REVA

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals