GETInGE iCast Covered Stent System Instruction Manual
- June 10, 2024
- GETInGE
Table of Contents
- GETInGE iCast Covered Stent System
- iCast Covered Stent System
- Instructions For Use
- Indication For Use
- Directions for Use General Procedure
- Removal of Unexpanded Stent
- MRI Safety Information
- Summary of iCARUS Clinical Study
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
GETInGE iCast Covered Stent System
iCast Covered Stent System
The iCast Covered Stent System is a medical device used to improve luminal diameter in patients with symptomatic atherosclerotic disease of the native common and/or external iliac arteries up to 110 mm in length, with a reference vessel diameter of 5 to 10 mm.
Description
- The delivery catheter has a wire lumen that is used for flushing and guidewire introduction.
- The delivery system is compatible with a 0.035 guidewire and has a useable length of 80 cm or 120 cm.
- The secondary or inflation lumen(s) are used for inflation/deflation of the attached balloon to deploy the endoprosthesis.
- To facilitate accurate device placement, two radiopaque bands are attached to the catheter shaft marking the ends of the crimped device and dilatation length of the balloon.
- Available stent sizes: Diameter (mm) – 5, 6, 7, 8, 9, 10; Length (mm) unexpanded – 16, 22, 38, 59
- Note that the entire iCast covered stent matrix is available in both 80 cm and 120 cm catheter lengths.
- The device is provided sterile for one procedure only.
- Do not resterilize or reuse.
- Do not use the iCast covered stent system if the sterile package is compromised or the iCast covered stent is damaged.
- Reuse, reprocessing, or resterilization may compromise the structural integrity of the device and/or lead to device failure which, in turn, may result in patient injury, illness, or death.
Indication For Use
The iCast covered stent system is indicated for improving luminal diameter in patients with symptomatic atherosclerotic disease of the native common and/or external iliac arteries up to 110 mm in length, with a reference vessel diameter of 5 to 10 mm.
Contraindications
The iCast covered stent is contraindicated for use in:
- Patients with uncorrected bleeding disorders.
- Patients who cannot receive recommended antiplatelet and/or anticoagulation therapy.
- Patients who are judged to have a lesion that prevents full expansion of the implant.
- Lesions in which the lumen diameter post-balloon angioplasty is insufficient for the passage of the endovascular system.
- Lesion locations subject to external compression.
Warnings and Precautions
- The device is provided sterile for one procedure only. Do not resterilize or reuse. Reuse, reprocessing, or resterilization may compromise the structural integrity of the device and/or lead to device failure which, in turn, may result in patient injury, illness, or death.
- Caution: Federal (USA) Law restricts this device to sale by or on the order of a physician.
- Do not use the iCast covered stent system if the box or sterile package is damaged or the iCast covered stent is damaged.
- People with allergies to 316L stainless steel (in particular nickel) or PTFE may suffer an allergic response to this implant.
- Do not overlap stents manufactured with differing materials.
- Use of appropriate anticoagulant and/or antiplatelet therapy per standard of care is recommended for use with this system. Special consideration should be given to patients who have a history of coagulation disorders, contraindications to anticoagulation or antiplatelet therapy, or contraindications to medications necessary for the procedure.
- Do not cut the device. The device should only be placed and deployed using the supplied balloon catheter.
- Use of the system requires advanced technical skills. Physicians should be trained in interventional techniques such as PTA and placement of intravascular stents. The following instructions will give technical guidance but do not obviate formal training in the use of the device.
Instructions For Use
- Inspect the package for damage. Do not use if the package is damaged.
- Inspect the device for damage. Do not use if the device is damaged.
- Prepare the patient according to standard of care for endovascular procedures in the target vessel.
- Select appropriate catheter length and stent size based on vessel diameter and lesion location.
- Advance the delivery catheter to the lesion location under fluoroscopic guidance.
- Advance guidewire to cross the lesion, then remove dilator/sheath and guidewire together.
- Insert the balloon catheter through the delivery catheter and position it across the lesion.
- Inflate the balloon to the specified pressure for the specified time period.
- Deflate the balloon and remove the balloon catheter.
- Remove the delivery catheter.
- Inspect the device placement under fluoroscopy.
- Follow standard of care for post-procedure care and monitoring.
Instructions For Use
Description
The iCast® covered stent system is a balloon-expandable stent, made of 316L
stainless steel and encapsulated with expanded PTFE (ePTFE). The delivery
system comprises an over-the-wire (OTW) catheter with a non-compliant balloon.
The device is available in diameters of 5 mm to 10 mm and crimped lengths of
16 mm, 22 mm, 38 mm, and 59 mm depending on diameter. See Table 1.
The OTW delivery catheter has a wire lumen that is used for flushing and
guidewire intro-duction. The delivery system is compatible with a 0.035”
guidewire and has a useable length of 80 cm or 120 cm. The secondary or
inflation lumen(s) are used for inflation/ deflation of the attached balloon
to deploy the endoprosthesis. To facilitate accurate device placement, two
radiopaque bands are attached to the catheter shaft marking the ends of the
crimped device and dilata-tion length of the balloon.
The device is provided sterile for one procedure only. Do not resterilize or
reuse. Do not use the iCast covered stent system if the sterile package is
compromised or the iCast covered stent is damaged. Reuse, reprocessing or
desterilization may compromise the structural integrity of the device and/ or
lead to device failure which, in turn, may result in patient injury, illness
or death.
Table 1 Available Stent Sizes
Diameter (mm) | Length (mm) unexpanded |
---|---|
16 | 22 |
5 | √ |
6 | √ |
7 | √ |
8 | – |
9 | – |
10 | – |
Note that the entire iCast covered stent matrix is avail- able in both 80 cm and 120 cm catheter lengths.
Indication For Use
The iCast covered stent system is indicated for improving luminal diameter in patients with symptomatic atherosclerotic disease of the native common and/or external iliac arteries up to 110 mm in length, with a reference vessel diameter of 5 to 10 mm.
Contraindications
The iCast covered stent is contraindicated for use in:
- Patients with uncorrected bleeding disorders.
- Patients who cannot receive recommended antiplatelet and/or anticoagulation therapy.
- Patients who are judged to have a lesion that prevents full expansion of the implant.
- Lesions in which the lumen diameter post-balloon angioplasty is insufficient for the passage of the endovascular system. compression.
Warnings and Precautions
The device is provided sterile for one procedure only. Do not desterilize or
reuse. Reuse, reprocessing or reserialization may compromise the structural
integrity of the device and/or lead to device failure which, in turn, may
result in patient injury, illness or death.
- Caution: Federal (USA) Law restricts this device to sale by or on the order of a physician.
- Do not use the iCast covered stent system if the box or sterile package is damaged or the iCast covered stent is damaged.
- People with allergies to 316L stainless steel (in particular nickel) or PTFE may suffer an allergic response to this im-plant.
- Do not overlap stents manufactured with differing materials.
- Use of appropriate anticoagulant and/ or antiplatelet therapy per standard of care is recommended for use with this system. Special consideration should be given to patients who have a history of coagulation disorders, contraindications to anticoagulation or antiplatelet therapy, or contraindications to medications necessary for the procedure.
- Do not cut the device. The device should only be placed and deployed using the supplied balloon catheter.
- Clinical experience with the stent has shown that the Arrow International Ar-rowFlex® Polyurethane Sheath should not be used.
- Use of the system requires advanced technical skills. Physicians should be trained in interventional techniques such as PTA and placement of intravascular stents. The following instructions will give technical guidance but do not obviate formal training in the use of the device.
- Do not use air or gaseous medium to inflate the balloon. Significant amounts of air in the balloon may cause uneven expansion of the stent, difficulty in de-ployment of the stent, or balloon rupture (refer to 3.0 DIRECTIONS FOR USE). Do not preinflate the balloon prior to stent deployment.
- The 316L stainless steel stent may cause susceptibility artifacts in MRI scans due to distortion of the magnetic field (see 5.0 MRI INFORMATION).
- Follow the Directions for Use supplied by the manufacturer of any ancillary device used in conjunction with the iCast covered stent system.
- Do not attempt to reposition or remove a stent once balloon inflation/deployment has been initiated.
- Do not attempt to withdraw or reposition a balloon catheter within the lumen of the deployed device unless the bal-loon is completely deflated.
- Do not remove, reposition or hand crimp the stent.
- Do not handle or in any way disrupt the placement of the stent on the balloon.
- Do not expose this device to organic solvents, ionizing radiation or ultraviolet light.
- The delivery system is not designed for use with power injection systems.
- Do not force passage or withdrawal of the guidewire through the iCast covered stent system if resistance is encountered.
- Always consult the label to ensure a compatible sheath is selected for use. Do not force passage or withdrawal of the iCast covered stent and delivery system when interfacing with a sheath or introducer if resistance is encountered. Forcing passage can result in damage to the stent or balloon catheter, stent dislodgment from the balloon, or separation of the catheter components. If using a 10mm device with a Cook introducer sheath, it is recommended to size up to an 8 French sheath to minimize potential for resistance.
- Prior to balloon inflation, utilize fluoros-copy to verify that the stent has been properly positioned. Do not expand the stent if it is not properly positioned.
- Balloon pressures are indicated on the product label and should be monitored during inflation. To ensure full stent expansion, inflate to at least the nominal pressure. Do not exceed the rated burst pressure. Use of pressures higher than specified on product label may result in a ruptured balloon with possible vessel damage or perforation.
- Under-expansion of the stent may result in stent migration.
- Do not pull an unexpanded stent back through a guiding catheter or sheath (see directions 4.0 REMOVAL OF UNEX-PANDED STENT).
- Stenting across a bifurcation will eliminate or compromise flow and impact future diagnostic or therapeutic procedures.
- 25. Recrossing a partially or fully expanded stent with adjunct devices must be per-formed with extreme caution.
- Do not stent when it is not possible to access the site with standard placement technique.
- Do not use this device in heavily calcified lesions that are not amenable to PTA or stent placement.
- Special consideration whether to stent should be given to lesions that contain fresh thrombus, with aneurysms adjacent to site of stent implantation, or with stenosis distal to site of stent implantation.
- Oversizing of the iCast covered stent and use of higher than recommended inflation pressures may cause dissection.
- The iCast covered stent is intended to perform as a system. The stent should not be removed for use in conjunction with other dilation catheters, nor should the delivery system be used in conjunction with other stents.
- Device should not be implanted in patients who have a known active infection at the site of implant.
- Do not post dilate the iCast covered stent greater than the recommended post dilation diameter (see Table 2 below). Post dilation greater than the recommended diameter may result in the fracture of the stent causing serious injury or death
Table 2 Maximum Recommended Post-Dilation Diameter (mm).Labeled| Device Length
---|---
Diameter| 16 mm| 22 mm| 38 mm| 59 mm
5| 7.3| 7.3| 9.8| 9.8
6| 7.3| 7.3| 10.0| 10.0
7| 7.3| 7.3| 10.1| 10.1
8| –| –| 10.2| 10.2
9| –| –| 10.4| 10.4
10| –| –| 10.6| – - Use the stent system prior to the ‘Use By’ date specified on the package.
- Do not use in locations subject to exter-nal compression.
Potential Hazards and Adverse Events
The following anticipated adverse events (AEs) have been identified as
possible com-plications of endovascular stent implanta-tion in iliac arteries:
- Acute myocardial infarction
- Allergic reaction to stainless steel, PTFE, drugs or contrast agent
- Angina/coronary ischemia
- Arrhythmia
- Arterial aneurysm
- Arterial rupture
- Arteriovenous fistula
- Bleeding complications from anticoagulant or antiplatelet medication requiring transfusion or surgical intervention
- Death
- Detachment, dislodgement, and/or im-plantation of a component of the system
- Emboli (air, tissue, or thrombotic emboli)
- Emergency surgery to correct vascular complications
- Hematoma/hemorrhage
- Hypotension/hypertension
- Infection, local or systemic
- Intimal injury/dissection/perforation
- Pain at catheter insertion site or limb
- Pseudoaneurysm formation
- Pulmonary embolism
- Renal insufficiency or failure
- Restenosis of the stented artery/occlusion
- Short-term hemodynamic deterioration
- Stent malposition/stent migration
- Stent strut fracture
- Stent thrombosis/occlusion
- Stroke
- Target limb loss (amputation of toe, foot and/or leg)
- Tissue necrosis
- Vascular thrombosis
- Vessel spasm
- Worsening claudication/rest pain
Directions for Use General Procedure
-
Patient preparation should be the same as for any angioplasty procedure. Standard techniques for placement of a sheath and guidewire should be em-ployed when using this device.
-
Perform diagnostic angiography to confirm site of implantation and major branch vessels.
-
Evaluate and mark the lesion.
Note: Standard PTA techniques should be used if predilation is necessary.
Sizing -
Measure the length of the lesion to determine the length of the stent required. The iCast covered stent will foreshorten during deployment. Refer to the stent sizing table (Table 3) listed below and on the product label for the stent length dimensions of the deployed stent. It is recommended to have the stent extend at least 5 mm proximal and distal to the lesion after expansion when possible.
Note: Should more than one stent be required to cover the lesion, place the first stent in the distal segment of the lesion followed by placement of the proximal stent(s). Overlap the stents a minimum of 10 mm for best results. -
Measure the inner diameter of the vessel to determine the appropriate diameter stent. The iCast covered stent is labeled as the outer diameter. Refer to the stent sizing table (Table 3) listed below and on the product label for the stent OD dimensions of the deployed stent.
Table 3 Stent Sizing and Foreshort
Balloon Diameter (mm)
|
Stent Length (mm)
| Post Nominal Pressure
8 ATM (811 kPa)
(mm)
| Post Rated Burst Pressure
12 ATM (1216 kPa)
(mm)
---|---|---|---
OD| Length| OD| Length
5| 16| 4.9| 15.9| 5.2| 15.6
5| 22| 4.9| 21.3| 5.2| 21.0
5| 38| 5.1| 37.2| 5.3| 37.7
5| 59| 5.0| 58.6| 5.3| 60.0
6| 16| 5.7| 15.7| 6.2| 15.1
6| 22| 5.8| 20.8| 6.2| 20.2
6| 38| 6.0| 36.6| 6.3| 37.0
6| 59| 6.0| 57.8| 6.3| 58.7
7| 16| 6.9| 15.0| 7.3| 14.2
7| 22| 6.9| 20.1| 7.3| 19.4
7| 38| 6.9| 35.8| 7.3| 35.7
8| 38| 8.1| 34.7| 8.5| 34.7
8| 59| 8.0| 56.0| 8.4| 56.5
9| 38| 8.9| 33.7| 9.3| 32.7
9| 59| 8.9| 54.6| 9.3| 54.0
10| 38| 10.0| 30.8| 10.4| 30.9
6. Based upon Nominal and Burst Pressure
Preparation
Before use, the physician must inspect the iCast covered stent and delivery
system in order to ensure that it was not damaged in transport.
-
Carefully remove the device from the sterile package and ensure the crimped stent or catheter shaft does not come in contact with any non-sterile surfaces or instruments.
-
Visually inspect the stent to ensure it is centered within the RO markers which are viewable through the folded balloon.
-
Prefill a syringe with sterile saline.
-
Attach the prefilled syringe to the guidewire lumen port as shown (labeled W on the catheter hub) and flush the guidewire lumen until the fluid exits the guidewire exit at the distal tip of the device.
-
Prepare a 20 cc syringe or balloon inflation device with saline or diluted contrast mixture and attach to the balloon inflation port (labeled I on the catheter hub). Hold the delivery system vertical with the distal tip pointing down. Draw back on the syringe applying negative pressure to the catheter and hold for 30 seconds with the syringe in the vertical position.
-
Release the pressure to neutral by gently letting go of the syringe plunger (keeping syringe in an upright position) to allow contrast to fill the delivery catheter inflation lumen(s).
Important: Do not apply positive pressure to the balloon as this can loosen the stent. -
Repeat steps 11 and 12 at least 3 times until all of the air in the delivery system is expelled. Steps 11 & 12 remove the air from the inflation lumen(s) and balloon, to allow diluted contrast media to fill the inflation lumen(s) to provide more uniform inflation of the balloon during stent deployment.
-
Prepare a 20 cc inflation device equipped with a manometer by filling with 10 cc of diluted contrast media for 5-10 mm diamter balloons. Follow the inflation device instructions. Expel all of the air from the inflation device.
-
Attach the prefilled inflation devic to the inflation lumen of the catheter hub (labeled I), ensuring no air bubbles remain at the catheter connection.
-
After the inflation device is connected, confirm the covered stent is between the RO markers. This visual inspection step verifies all preparation steps did not create positive pressure (which could cause inadequate stent deployment) and the stent remains securely positioned within the RO markers prior to patient use.
Deployment -
Visually inspect the stent for adherence to the balloon and centered placement in relation to the RO markers on the bal-loon.
-
Confirm that the sheath selected is a compatible size with the selected iCast covered stent system identified on the product label.
Note: Increased resistance may be en-countered when inserting a 10mm device through a 7 Fr Cook sheath; therefore, it is recommended to upsize to an 8 Fr sheath when a Cook sheath is used with a 10mm device. -
Advance the device over a (0.035”) guidewire and through the sheath to the target location under direct fluoroscopic visualization.
-
Utilize fluoroscopy to verify the stent has not shifted or been damaged during positioning. Do not expand the stent if it is not properly positioned.
Steadily inflate the balloon to the nominal deployment pressure as stated on the product label. -
Maintain the inflation pressure for a minimum of 15 seconds at nominal pressure but not greater than burst pressure for full stent expansion.
-
Deflate balloon by pulling vacuum on the inflation device to its maximum volume and allow sufficient time for full deflation.
Note: Deflation times may vary based on balloon size, catheter length, and inflation media used. Deflation may take longer with larger devices and higher concentrations of contrast.
IMPORTANT: Visually verify full deflation of the balloon via fluoroscopy before proceeding to step 24 for withdrawal of the delivery system.
CAUTION: Do not force withdrawal of the delivery system if resistance is encountered. Forcing withdrawal may result in damage to the delivery system, including separation of the balloon or catheter hub from the delivery catheter. If unable to fully deflate the balloon or resistance is encountered, remove the delivery system and introducer sheath/guiding catheter as one unit.
Note: It is recommended that the guide-wire remain across the lesion until the procedure is completed.
- While maintaining guidewire position and negative pressure on the inflation device, slowly withdraw the delivery catheter.
- Confirm optimal stent apposition using standard angiographic procedures.
Further Dilation
If necessary, the stent can be post-dilated using standard angioplasty
techniques. Maximum iCast covered stent post dilation sizes are listed in
Table 2, located in section 2.2 of this IFU.
Removal of Unexpanded Stent
Extreme caution must be used when removal of an unexpanded or partially expanded stent is necessary. The stent/ delivery system should not be withdrawn until the proximal end of the stent is aligned with the distal tip of the introducer sheath. The sheath and the stent delivery system should then all be removed as one unit. After removal, the iCast covered stent system should not be reused.
MRI Safety Information
Non-clinical testing has demonstrated that the iCast covered stent is MR
Conditional. A patient with the iCast covered stent
can be scanned safely, immediately after placement, in an MR system meeting
the following conditions:
- Static magnetic field of 1.5 T or 3.0 T only.
- Maximum spatial gradient magnetic field of 3,000 gauss/cm (30 T/m).
- Maximum MR system-reported, whole-body averaged specific absorption rate (SAR) of 2 W/Kg (Normal Operating Mode).
Under the scan conditions defined above, the iCast covered stent is expected
to produce a maximum temperature rise of 2.2 °C after 15 minutes of continuous
scanning.
In non-clinical testing, the image artifact caused by the device extends
approximately 15 mm from the iCast covered stent when imaged with a gradient
echo pulse sequence and a 3.0 T MRI system. The lumen of the iCast covered
stent cannot be visualized on the gradient echo or T1-weighted, spin echo
pulse sequence.
Summary of iCARUS Clinical Study
The iCARUS (iCast covered stent Atrium Registry Ultrasound Study) study was a
prospective, multicenter, non-randomized, single-arm study performed to
evaluate the safety and effectiveness of the iCast covered stent system in the
treatment of patients with symptomatic claudication or rest pain and
angiographic confirmation of de novo or restenotic lesions in the common
and/or external iliac artery. Patients were treated between October 18, 2007
and Oc-tober 27, 2010. Data was collected through 31 October 2013 and included
165 subjects enrolled from 25 investigational sites. The primary endpoint of
the iCARUS study was a composite endpoint defined as the occur-rence of death
within 30 days, target site revascularization (TSR) within 9-months, or
restenosis at 9-months post-procedure. The determination of restenosis was
made by direct evaluation of the iliac arteries using duplex ultrasound (or
angiography done in lieu of ultrasound). Key secondary endpoints included
Major Adverse Vascular Event (MAVE) at each time point through 12 months and
Major Adverse Event (MAE) rate at 30 days; device success, acute procedural
success, clinical success at each time point through 36 months; patency at
each time point through 36 months; and a composite rate of 30-day death,
9-month TSR, and
9-month restenosis in subjects without total occlusions of the iliac.
Patient Population
Patients were required to be at least 18 years of age with lifestyle limiting
claudication or rest pain (Rutherford-Becker scale 2-4), and with presence of
de novo and/or restenotic lesions in the common and/or external iliac artery.
Single or multiple target lesions (ipsi-lateral and/or bilateral), with ≥50%
stenosis by visual estimate which could be crossed with a guide wire and
dilated, were allowed. Eligible target lesions were between 5 and 12 mm in
diameter and <110 mm in length; with evidence of patent profunda or
superficial femoral artery (SFA) in the target limb(s). Treatment of non-
target lesions in the SFA, meeting certain criteria, was permissible following
successful treatment of the target lesion.
To participate in the study, patients were required to give informed written
consent, be willing to adhere to the required follow-up visits and testing
through month 36 and the required medication regimen following the procedure.
The 165 subjects were enrolled at 24 US sites and 1 German site. Of the 165
subjects enrolled, 13 subjects had inclusion/exclusion violations and were not
included in the Intent to Treat (ITT) population.
Subject demographics, medical history and risk factors are summarized in Table
4. The demographics of the study population were typical for this type of
study performed in the United States. The iCARUS population was predominantly
male (61.8%) with a history of smoking (91.4%), hypertension (84.9%) and
hypercholesterolemia (85.5%).
Table 4 Subject Demographics, Medical History and Risk Factors – ITT
m=number of subjects in the category
n=number of subjects in the study group with sufficient data for analysis
*The safety and effectiveness of the iCast covered stent in pediatric populations were not studied in the iCARUS trial. As Peripheral Arterial Disease (PAD) is not typically found in the pediatric population, with the exception of rare homozygous lipid disorders, there is no plan to treat atherosclerotic disease of the common and/or external iliac arteries in this population.
Tables 5 and 6 summarize pre and post-procedure lesion characteristics as
assessed by the angiographic core laboratory. Two hun-dred twenty-three (223)
lesions were treated in 206 limbs in 152 subjects. The majority of lesions had
severe calcification (60.1%) pre-procedure. Among the 223 evaluable baseline
lesions, mean reference vessel diameter (RVD) was 8.3±2.1 mm, mean percent
stenosis was 69.3±16.7% and mean minimum lumen diameter (MLD) was 2.6±1.5 mm.
Table 5 Angiographic Morphology Data – ITT
m=number of subjects in the category
n=number of subjects in the study group with sufficient data for analysis
Table 6 Angiographic Quantitative Analysis – ITT
n = number of lesions in the study group with sufficient data for analysis
Methods
Prior to implantation, standard diagnostic angiography was performed to
confirm angiographic criteria for the study were met. Lesion length and native
vessel diameter were measured in order to select the appropriate length and
diameter of the stent(s). Subjects were considered enrolled once the
investigational device was entered into the body.
Standard procedures for percutaneous interventional procedures were followed.
Pre and post dilation of the lesions were per-formed at the discretion of the
investigator. Concomitant antiplatelet/anti-thrombin and anti-coagulant
therapy was administered per routine practice. After the procedure, sub- jects
were prescribed up to 325mg of daily aspirin and 75mg of clopidogrel for 6
months. Ticlopidine could be substituted for subjects with sensitivity to
clopidogrel.
After hospital discharge, patients were required to return to the study center
for clinical assessments on Day 30 (± 7 days), 6-months (± 14 days), 9-months
(± 14 days), 12-months (± 30 days), and annually thereafter through 36 months
(± 30 days). Assessments including, but not limited to ECG (30 day only),
Rutherford Becker score, ankle-brachial index (ABI), health status assessment,
and duplex scans were performed. Additionally, an angiogram was performed as
needed to confirm restenosis and assess the safety and effectiveness of the
iCast covered stent.
Results
The primary composite endpoint rate, based on available cases in the ITT
population, was 8.1% (10/123), with an exact one-sided upper 95% CI of 13.4%
(p=0.005). This was below the protocol-specified performance goal of 16.57%
and the study stent is considered to have met the performance goal (Table 7).
Table 7 Primary Composite Endpoint – ITT
Key secondary endpoints included Major Adverse Vascular Events (MAVE) and
Major Adverse Events (MAE) as adjudicated by the independent Clinical Events
Committee, de-vice success, acute procedural success (with and without
pressure gradient), clinical success, patency (primary, primary-assisted, and
secondary), and a composite rate consisting of 30-day death, 9-month TSR, and
9-month restenosis in subjects without iliac total occlusions. MAVE was
defined as a compos-ite rate of myocardial infarction at 30-days, stent
thrombosis, clinically apparent distal embolization, defined as causing end-
organ damage, arterial rupture, acute limb ischemia, target limb amputation or
a procedure related bleeding event requiring transfusion. MAVEs are summarized
through 12 months post procedure (Table 8). MAE was defined as a composite
rate of MAVE, any death, or stroke up to 30 days post-procedure.
Device success was defined as the successful delivery and deployment of the
study stent and intact retrieval of the delivery system.
Acute procedural success was defined as device success with <30% residual
stenosis immediately after stent placement, mean trans-stenotic pressure
gradient <5 mmHg, and without occurrence of in-hospital MAVE.
Early clinical success was defined as improvement of the Rutherford-Becker
clinical ≥ 1 category, and assessed between baseline and the 1 month follow-up
visit.
Late clinical success was defined as a maintained improvement in ABI, assessed
as either (1) Normalized (>0.90) or (2) An in-crease of 0.1 from the baseline
level and not decreased by >0.15 from the maximum result (observed post-
procedure). This endpoint was to be assessed during follow-up at the 6, 9, 12,
24, and 36 month visits.
Patency was assessed at each follow up time point, categorized as:
- Primary patency, defined as continuous flow without revascularization, bypass, or target limb amputation.
- Primary assisted patency, defined as continuous flow assisted when the target vessel has restenosed at any time post-procedure.
- Secondary patency, defined as reestablishment of flow to distal arteries after occlusion has occurred at the target vessel.
In the ITT population, 4.6% (7/151) of subjects experienced at least one (1)
MAVE to 30 days and 6.9% (10/144) of subjects experienced at least one (1)
MAVE to 6 months. The most common MAVE event was procedure-related bleeding
event that required transfusion. This type of event occurred among 4 of 151
(2.6%) ITT subjects to 30 days, and among 7 of 144 (4.9%) subjects to 6
months. There were no additional subjects experiencing MAVE events beyond 6
months post-procedure, for a total MAVE rate of 7.1% (10/140) at 9 months and
7.2% (10/139) at 1 year.
Eight (8) of 151 (5.3%) subjects experienced a MAE to 30 days (seven (7)
subjects experienced a MAVE and one (1) subject experienced a stroke). There
were no deaths in the ITT population reported to 30 days.
Acute procedural success was achieved in 92.7% (140/151) of subjects. In the
eleven (11) subjects that did not achieve acute procedural success, seven (7)
subjects experienced an in-hospital MAVE; three (3) subjects had residual
stenosis >30%; and two (2) subjects were documented as device non-success.
Note that the eleven (11) subjects includes one (1) subject that experienced
both device non-success as well as residual stenosis >30%. One (1) additional
subject was missing angiography and was not assessed for acute procedural
success, though the subject achieved device success and had no in-hospital
MAVE.
Primary patency was achieved in 100.0% of the ITT subjects at 1-month, 99.3%
of the subjects at 6-months and 96.4% of the subjects at 9-months. Primary
patency was 93.5% at 12 months, 87.6% at 24 months and 84.7% at 36 months.
Primary-assisted paten-cy was achieved in all (100.0%) of subjects at 1, 6, 9
and 12 months, in 99.2% of subjects at 24 months, and 99.1% of subjects at 36
months. Secondary patency was achieved in all (100%) subjects at all time
points through 36 months post-procedure (Table 8). Device, procedural and
clinical success results are listed in Table 8.
The composite rate of 30-day death 9-month TSR, and 9-month restenosis in
subjects without total occlusions of the iliac was 5% (5/100).
Table 8 Device, Procedural and Clinical Success – ITT
Safety Results
A total of 44 of 165 (26.7%) enrolled subjects experienced Serious Adverse
Events (SAEs) to 9-months. The most frequent SAEs were in the vascular
disorders MedDRA System Organ Class (SOC) (14/165) and the cardiac SOC
(13/165). A total of 86 of 165 (52.1%) enrolled subjects experienced SAEs to
36-months. The most frequent SAEs were in the vascular disorders SOC (30/165)
and the cardiac SOC (33/165).
There were 14 deaths total reported in
the study (six (6) deaths reported prior to 9-months). None of the deaths were
deemed to be related to the study procedure or the study device. There were no
Unanticipated Adverse Device Effects (UADEs) reported in the study. SAEs that
occurred throughout the study for all enrolled subjects are listed in Table 9.
Results were similar in the ITT population.
Table 9 Summary of Serious Adverse Events– All Enrolled Subjects (N=165)
*Preferred Term applies to events to 36 months only
Conclusion
This prospective, multi-center, single-arm clinical trial (iCARUS) has
demonstrated that the iCast covered stent is safe and effective for its
intended use as a treatment for iliac artery disease in the indicated
population.
This clinical trial compared the safety and effectiveness of the iCast covered
stent to a predetermined performance goal. The com-posite safety and
effectiveness endpoint was defined as the occurrence of death with-in 30 days,
target site revascularization (TSR) within 9 months, or restenosis at 9 months
post-procedure (as determined by DUS). The primary composite endpoint rate was
8.1% (10/123), with an exact one-sided upper 95% CI of 13.4% as compared to
the performance goal of 16.57% (p=0.005). As this is signifi-cantly below the
performance goal of 16.57%, the trial met its primary composite endpoint.
There were no deaths within 30 days, the rate of TSR within 9-months was 2.9%
and the restenosis rate at 9-months was 4.9%. Although no formal performance
metric was designated for other secondary safety and effectiveness endpoints,
analysis of key pre-specified endpoints is consistent with a low rate of MAVE,
and MAE, and a high rate of early and late clinical and angiographic suc-cess
with use of the iCast covered stent.
The benefits of stenting with the iCast cov-ered stent were maintained in the
majority of subjects during long-term follow-up. At 36 months, primary patency
was present in 84.7% of subjects, primary-assisted patency was present in
99.1%, and 100% of subjects in the ITT population had secondary patency.
Similarly, late clinical success was pres-ent in 72.4% of subjects at this
time point. These data are consistent with comparable long-term effectiveness
data available for other approved iliac stents and support the safety and
effectiveness of the iCast covered stent when used for revascularization of
iliac artery lesions. These results were consis-tent with clinical results
from the COBEST study. Altogether, these data support the safety and
effectiveness of the iCast covered stent when used for revascularization of
iliac lesions in the studied patient population.
Mwipatayi, B.P., et al., A comparison of covered vs bare expandable stents
for the treatment of aortoiliac occlu-sive disease. J Vasc Surg, 2011. 54(6):
p. 1561- 70.
Mwipatayi, B.P., et al., Durability of the balloon-expand-able covered versus
bare-metal stents in the Covered versus Balloon Expandable Stent Trial
(COBEST) for the treatment of aortoiliac occlusive disease. J Vasc Surg, 2016.
64(1): p. 83-94.e1.
Symbols Used On Product Labels
- KEEP AWAY FROM SUNLIGHT
- KEEP DRY
- DO NOT RESTERILIZE
- DO NOT RE-USE
- NON-PYROGENIC
- MR CONDITIONAL
- CAUTION
- USE-BY DATE
- MANUFACTURER
- DO NOT USE IF PACKAGE IS DAMAGED
- STERILIZED USING ETHYLENE OXIDE
- CATALOGUE NUMBER
- SERIAL NUMBER
- DIMENSIONS
-
- CONSULT INSTRUCTIONS FOR USE
- BOX QUANTITY
- CAUTION: FEDERAL LAW (U.S.A.) RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN
The iCast covered stent system is manufactured by Atrium Medical Corporation /
40 Continental Blvd, Merrimack, NH 03054 /
603-880-1433.
CAUTION: Federal Law (U.S.A.) restricts this device to sale by or on the order
of a physician.
GETINGE, and iCast are registered trademarks of Getinge AB or its subsidiaries
in the United States. Copyright 2023 Atrium Medical Corporation. All rights
reserved.
Atrium Medical Corporation • 40 Continental Blvd. • Merrimack • New Hampshire
• 03054 • U.S.A. 603-880-1433
www.getinge.com
Printed in USA • Part No. 009603
AW009603-EN Rev 11 • 2023-03