midmark 8012-00-001 Veterinary Folding Exam Table Installation Guide

July 16, 2024
Midmark

midmark 8012-00-001 Veterinary Folding Exam Table Installation Guide

Equipment Alert
The wall structure must be capable of supporting the table load. Weight: 90 lb. (41 kg) Combined table and patient torque (max) 1660 ft. lb.
(2250 N-m) For wall construction types not shown here (i.e. concrete, block, brick, etc.), it is the responsibility of the contractor/installer to supply appropriate hardware.

Loose Parts Sent with Table

Shaft Support|
---|---
E-Ring|
Spacer Bearing|
3/8″ x 4″ Hex HD Lag Bolt|
3/8″ Washer|
1/4-20 x 7/8″ Hex Screw|
Upper Shaft|
Lower Shaft|

Note
Shafts come installed on wall base and folding section. Must remove before installing the folding section to the wall base.

Step 1: Determine mounting location.
A) Use dimension information shown to determine the mounting location.

Install Table A Minimum of 36″ From Floor For Clearance when Table is in Drop Position

Step 2a: Install wall base for Wood Studs.
A) Mark top two mounting holes on stud centers, 33.25″ from floor.
B) Drill two 1/4″ holes 4″ deep at mark hole locations.
C) Secure wall base to wall with two 3/8″ lag bolts (for wood) and washers.
D) Drill other four mounting holes for wall base and install remaining bolts and washers.
E) As you tighten all bolts, use level to make sure wall base is flat.

WARNING
Serious injury possible if instructions are not followed

Equipment Alert
Table top must be 36″ minimum from floor.

Note
Refer to chase installation manual for mounting hole locations in chase cabinet. To install table to chase, use hardware included with chase cabinet. (3/8″ bolts, washers and nuts)

Step 2b: Install wall base for Steel Stud (wood blocking design).
If mounting table to a steel stud wall, there are two options. See option 2 on next page.

  1. Install wood blocking design into steel studs as described below. Install wall base directly into wood

blocking with lag bolts. In existing facilities cut hole in drywall large enough to install blocking design.
A) Make relief cut on 2 – 2×12’s. Starting with right side insert first 2×12 on steel stud stiffening lip and use 4 – 2” #14 flat head wood screws to affix 2×12 to steel web at 34.5″ high. Screw plywood spacer to first 2×12. Insert second 2×12 onto stiffening lip and screw into place. Repeat for left side.
B) Screw 4 – 3″ #14 flat head wood screws onto front of stud flange using a total of 8 to secure blocking.
C) Repair drywall and mark top two mounting holes (on new blocking design), 33.25″ from floor.
D) Drill two 1/4″ holes 4″ deep at mark hole locations.
E) Secure wall base to wall with two 3/8″ lag bolts (3/8 bolt for backer plate) and washers.
F) Drill four mounting holes for wall base (wood block design). Install remaining bolts and washers.
G) As you tighten all bolts, use level to make sure wall base is flat.

Step 2c: Install wall base for Steel Stud (backer plate kit).

  1. Use backer plate kit, 002-1692-00 (ordered separately), install per instructions included with kit.

A) Secure wall base to wall with two 3/8″ lag bolts (3/8 bolt for backer plate) and washers.
B) Drill four mounting holes for wall base (wood block design). Install remaining bolts and washers.
C) As you tighten all bolts, use level to make sure wall base is flat.

Step 3: Install folding section.
A) Install first e-ring on one end of upper shaft.
B) Position folding section to wall base and slide top shaft through mounting holes.
C) Install second e-ring to other end of top shaft.

Step 4: Install gas spring.
A) Remove protective packaging and zip tie from gas spring.
B) Snap rod end of gas spring into ball stud on wall base.
C) If gas spring is not aligning with the ball stud on wall base, do the following:

  • Loosen four table top screws under folding section.
  • Pivot folding section higher to align gas spring to ball stud.
  • Reposition table top to have approximately 1/8″ seam gap and tighten screws.

Note: Mismark® recommends the gas spring be installed rod end down.

Step 5: Install locking arm.
A) Install spacer bearing onto one side of locking arm
B) Slide locking arm through opening in folding section bracket.
C) Install spacer bearing onto other side of locking arm.
D) Slide lower shaft through spacer bearings on locking arm.
E) Align shaft in locking arm with holes in wall base.
F) Install shaft supports using 1/4-20 screws and a 5/32 allen wrench.

Equipment Alert
Install Locking Arm with Gas Spring Cut Out facing top. Will damage Gas Spring if installed incorrectly

Step 6: Level table.
A) Place level on table folding section to verify it is flat.
B) Adjust table if needed.
C) Lift one side of the locking arm handle to allow table to go down.
Lift up the table top to raise and verify it is working properly.

To adjust table (if needed)..
A) Loosen all four bolts under end of table.

  • Turn two adjustment screws clockwise to raise table.
  • Turn two adjustment screws counterclockwise to lower table.
    B) Tighten all four bolts.

Note: Both adjustment screws need to be loosened to properly level.\

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals