midmark 002-11029-00 Manifold Kit Installation Guide
- May 15, 2024
- Midmark
Table of Contents
midmark 002-11029-00 Manifold Kit
Kit Components and Compatibility
Kit Number | Components |
---|---|
002-11029-00 | Syringe |
002-11058-00 | Low-Speed Handpiece |
002-11059-00 | High-Speed Handpiece |
002-11060-00 | Scaler |
Applies to Models: All
Special Tools: None
Note: Some components will vary depending on the manifold kit number.
Installation Overview
- Turn power switch to off. Unplug from power cord.
- Release pressure from pressure tank by holding cloth on tube and opening valve.
- Remove mat from lid. Open top lid by pulling lever back and lifting lid all the way up.
- (High Speed Manifold Kit shown) Remove tubing from desired manifold block and label as needed.
- (High Speed Manifold Kit shown) Remove screws (2) form manifold mounting plate.
- (High Speed Manifold Kit shown) Remove screws (2) and old manifold. Replace with new manifold and secure with screws (2) provided.
- Install manifold mounting plate with screws (2).
- For the four handpiece delivery system tubing connections of manifolds, refer to the detailed diagram on page 4 of the manual.
- For the five handpiece manifold tubing connections, refer to the detailed diagram on page 5 of the manual.
- Close lid and push down on middle until a click is heard.
- Install mat onto lid.
- Plug in dental delivery system and turn on.
Tubing Connection Diagrams
4 Handpiece Manifold Tubing Connections
Connect tubing with sleeve clamps and uni-clamp to manifold components in delivery system as shown in the diagram on page 4.
5th Handpiece Manifold Tubing Connections
Connect tubing with sleeve clamps and uni-clamp to manifold components in delivery system as shown in the diagram on page 5.
Specifications
Property | Specification |
---|---|
Document Revision | TP200 20-42-FO-00012 Rev A2 ECO088522 |
Release Date | 1/11/2024 |
Style | G |
FAQ
- Does this installation kit apply to all models?
- Yes, this kit is compatible with all models.
- Are there any special tools required for installation?
- No special tools are required for the installation of this kit.
- How do I know which components are included in my manifold kit?
- The components included may vary depending on the manifold kit number. Refer to the kit components and compatibility section for details.
Product Overview
Installation Instruction
- Step 1:
- Turn power switch to off. Unplug from power cord
- Step 2:
- Release pressure from the pressure tank by holding a cloth on the tube and opening valve
- Step 3:
- Remove mat from the lid. Open top lid by pulling the lever back and lifting lid all the way up.
- Step 4: (High-Speed Manifold Kit shown)
- Remove tubing from desired manifold block and label as needed.
- Step 5: (High-Speed Manifold Kit shown)
- Remove screws (2) form the manifold mounting plate.
- Step 6: (High-Speed Manifold Kit shown)
- Remove screws (2) and old manifold. Replace with new manifold and secure with screws (2) provided
- Step 7:
- Install manifold mounting plate with screws (2).
- Step 8a:
- Refer to page 4 for the four handpiece delivery system tubing connections of manifolds.
- Step 8b:
- Refer to page 5 for the five handpiece manifold tubing connections.
- Step 9:
- Close lid and push down on middle until a click is heard
- Step 10:
- Install mat onto lid.
- Step 11:
- Plug in dental delivery system and turn on.
- Step 8a: 4 Handpiece Manifold Tubing Connections
- Connect tubing with sleeve clamps and uni-clamp to manifold components in delivery system as shown.
- Step 8b: 5th Handpiece Manifold Tubing Connections
- Connect tubing with sleeve clamps and uni-clamp to manifold components in delivery system as shown.
2024 Midmark Corp. |