Micsig CP2100 Series DC Current Probe User Guide
- July 15, 2024
- Micsig
Table of Contents
Micsig CP2100 Series DC Current Probe
Operation Instructions:
- Connect BNC interface to the oscilloscope (or other instruments), then plug the USB cable to power the probe.
- Select appropriate range according to current range, the corresponding button light will turn Green.
- Adjust the oscilloscope settings: Input impedance 1M; select probe to Current or display as A; Set attenuation ratio on corresponding channel, 100A0.01V/A to 100X, range 10A0.1V/A to 10X.
- Press Zero button to do zero calibration, after success, the buzzer will beep one time; if beep three times meaning zero calibration has failed; can also go Manual to adjustment. The external magnetic field may have slight influence on the DC zero position, do not move it in a large range after zero adjustment is completed.
- Clamp the conductor under test according to the direction indicated by the jaws. Note: If the measured current flows in the opposite direction, the output will be negative.
- Adjust the oscilloscope to get the best waveform.
Frequently Asked Questions
Q: What should I do if the measured current exceeds the range?
A: When the current exceeds the range, the buzzer will beep for a long time, and the button light will flash.
PRODUCT IDENTIFICATION
Product Introduction
The CP2100 series current probe is a BNC type that can measure both DC and AC. With split design and compact look, it also can be connected to multimeters via connector. Automatic and manual zero adjustment, powered by USB, no need additional power supply, making measurement more convenient, often used in motor drives, industrial frequency, inverters, power supplies, avionics and other fields.
Safety Precautions
-
The measurable circuit should be CAT II 600V or below
o not measure bare conductors -
o not touch the measured conductor and sensor head during measurement
➢Do not use in a humid environment
➢Do not touch the instrument or the measured object with wet hands
➢Please ground this product through the USB power cord
➢Please use this product as required
Specifications
- Model: CP2100A / CP2100B
- Bandwidth: DC~800KHz / DC~2.5MHz
- Rise time: 437.5ns / 140ns
- Range: 10A/100A
- Output sensitivity: 0.1V/A (10A) / 0.01V/A (100A)
- DC accuracy (typical): 100ns
- Signal delay: 50mA~10Apk (10A) / 1A~100Apk (100A)
- Measuring range: 100Apk-70.7Arms DC+AC (10A) / 200Apk-pk-70.7Arms AC (100A)
- Max measurable current: 13mm
- Max working voltage: CAT III 300V
- Max float voltage: CAT II 600V
- Max conductor diameter: 13mm
- Overload indication: Buzzer beeps and Button light flashes
- Supplied power: DC 5V
- Net weight: 290g
- Package weight: 1000g
- Operating temperature: 0~40°C
- Operating humidity: 5%~65% (40~50, No condensation) 5%~95% (0~40, No condensation)
Operation Instructions
- Connect BNC interface to the oscilloscope (or other instruments), then plug the USB cable to power the probe;
- Select appropriate range according to current range, the corresponding button light will turn Green;
- Adjust the oscilloscope settings: Input impedance 1MΩ; select probe to Current or display as A; Set attenuation ratio on corresponding channel, 100A(0.01V/A)to 100X, range 10A(0.1V/A)to 10X;
- Press Zero button to do zero calibration, after success, the buzzer will “beep” one time; if “beep” three times, meaning zero calibration has failed; can also go Manual to adjustment. The external magnetic field may have slight influence on the DC zero position, do not move it in a large range after zero adjustment is completed;
- Clamp the conductor under test according to the direction indicated by the jaws. Note: If the measured current flows in the opposite direction, the output will be negative;
- Adjust the oscilloscope to get the best waveform;
Note: When the current exceeds the range, the buzzer will beep for a long time and the button light will flash.
References
Maintenance
During the warranty period of the product (one-year) and under normal use, the company will be responsible for free repairs due to fault caused by the quality of the product itself, and the product must not be disassembled or repaired without Micsig permission. Please keep the product dry, clean and tidy. If there is dirt, use a soft cloth or sponge with alcohol to wipe off. Do not use water. In order to ensure the performance of the product, it is recommended to check or calibrate once a year.
Statement
The information provided in this document is subject to change in future versions without notice. In addition, to the maximum extent permitted by applicable laws, Micsig does not provide any express or implied warranty for this manual and any information contained in it.
Shenzhen Micsig Technology Co., Ltd.
Tel: +86 (0)755 88600880
Email: sales@micsig.com
Add: A106, Huafeng International Robot Industrial Park, Bao’An district,
Shenzhen, 518126, Guangdong, China
website: www.micsig.com