Congratulations on your new RMS gateway. The gateway transmits the data of the wireless data loggers continuously to the RMS software by Ethernet. These short instructions describe the main functions of the device.

https://rotronic.live/RMS-GW-868

Please read this short manual and the instruction manual on www.rotronic.com/rms carefully.

COMMISSIONING

The device must be supplied with 24 V (terminals: V+ / V-) or via PoE to be able to transmit data. The gateway can be mounted easily with the wall bracket. The device is connected to the RMS software by pairing.
Cloud integration
Integration of LAN devices into the Rotronic Public Cloud requires for the local network Port 80 to be enabled and a DHCP server must assign an IP address to the LAN device. For all other integrations, please check the online manual.

INTEGRATION OF THE GATEWAY (PAIRING) IN 6 STEPS

  1. If you do not want to connect the device to the Rotronic Cloud, the server must be configured in the device.
    a. Connect the device to the local network and start the RMS configuration software.
    b. Search for the device under Device > Search > Network Device. The software finds all RMS devices in the local network.
    c. Enter the host (server address) and the URL of the software services under Settings.
    d. Finish configuration with Write.

  2. Log into the RMS software/cloud.
    Select Tools > Setup > Devices > New > LAN device.rotronic RMS-GW-868
Monitoring System -e RMS software

  3. Enter the serial number of the device.rotronic RMS-GW-868 Monitoring System -serial number

  4. Wait until the device is flashing orange. Press the button on the device briefly, as shown in the picture on the RMS software. The LED flashes green,  when the connection is successful. rotronic RMS-GW-868 Monitoring System -flashing orange

  5. Enter the device name, channel, and group.rotronic RMS-GW-868 Monitoring System - device name

  6. Finish configuration.

rotronic RMS-GW-868 Monitoring System -configuration

LED INDICATORS

Status LED Function Meaning
Connected Flashes green Status OK
Flashes red 2 times: no connection to the server
Not connected Flashes orange Device in pairing mode, push the button for

integration into the software

ACCESSORIES

AC1321 Mounting kit with Allen key and mounting cone

TECHNICAL DATA

Power supply: 24 VDC ±10 % / <100 mA
Power supply requirements: 24 VDC ±10 % / >4 W / limited power source
Range of measurement/application: -40…70 °C
IP protection: IP65
Software:  RMS Monitoring Software
Weight: 200 g

DIMENSIONS

rotronic RMS-GW-868 Monitoring System -DIMENSIONS

DELIVERY PACKAGE

  • Gateway
  • Wall bracket
  • Short instruction manual

This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates,  uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.

Any changes or modifications to this device not explicitly approved by the manufacturer could void your authority to operate this equipment.

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

ROTRONIC User Manuals

Related Manuals