RMS-LOG-L-D
Short instruction manual
GENERAL DESCRIPTION
Congratulations on your new RMS data logger. The data logger has an internal
data memory of 44,000 measured-value pairs and transmits these values
continuously to the RMS software by Ethernet. These short instructions
describe the main functions of the device.
Please read these short instructions and the instruction manual at
https://service.rotronic.com/manual/ carefully scan the QR code to open the
instruction manual directly.
https://rotronic.live/RMS-LOG-L-D
COMMISSIONING
The device is supplied with power as soon as the data logger is supplied with
24 V (terminal block: V+ / V-) or PoE. Only then can the data be transmitted.
The data logger can be mounted easily with the wall bracket. Choose a suitable
position for measurement. Avoid disruptive influences such as sunlight,
heating elements, etc. The device is connected to the RMS software by pairing.
Cloud integration
Integration of LAN devices into the Rotronic Public Cloud requires for the
local network Port 80 to be enabled and a DHCP server must assign an IP
address to the LAN device. For all other integrations, please check the online
manual.
INTEGRATION OF A DATA LOGGER (PAIRING) IN 6 STEPS
-
If you do not want to connect the LAN device to the Rotronic Cloud, the server must be configured in the device.
a. Connect the device to the local network and start the RMS configuration software.
b. Search for the device under Device > Search > Network Device. The software finds all RMS devices in the local network.
c. Enter the host (server address) and the URL of the software services under Settings. d. Finish configuration by clicking “Write”. Close the software. -
Log into the RMS software/Cloud. Select Tools > Setup > Device > New Wireless device or LAN device.
-
LAN device — Enter the serial number of the device.
-
Wait until the device flashes orange. Briefly press the button on the device as shown in the picture of the RMS software. The LED flashes green when the connection is successful.
-
Configure the device.
-
Finish configuration.
LED INDICATORS
State
| LED function|
Meaning
---|---|---
Connected| Flashes green| Status OK, data transmitted
Flashes orange| Device not connected to the Internet
Flashes red|
- time: low battery, replace soonest
- times: no probe connected
Not connected| Flashes orange| Device waiting for integration into the software
ACCESSORIES
- RMS-PS: Power supply, 24 VDC, 15 W
- E2-OXA :Extension cable, various lengths
TECHNICAL DATA
General specifications
Measurement interval| 10 s to 300 s
Startup time| < 10 s
Software compatibility| V1.3.0, from V2.1 all functions
Application range| -20…70°C, non-condensing
Storage conditions| -20…30 °C, non-condensing
Maximum altitude| 2000 m ASL
Power supply| 24 VDC ±10 %/ Battery: RMS-BAT (2xAA, LiSocl2)
Max. current consumption| 50 mA
AC adapter requirements| 24 VDC ±10 %, 4 W minimum, ) 5W Limited Power Source
PoE| 802.3af-2003, Class 1
Device data
Order code| RMS-LOG-L-D
Ethernet cable requirement| Min. Cat 5, SFTP, max. 30 m
Interface| Ethernet
Protocols| HTTP / ModbusTCP
Number of measuring points| 2
Battery life (@60 s & 600 s interval)| HCD-S / HCD-IC: 7 d
CCD-S-XXX: 2.4 d
PCD-S-XXX: 15 d
Storage capacity| 44,000 data points
Conformity with standards
Soldering material| Lead free / RoHS conformity
FDA/GAMP directives| 21 CFR Part 11 / GAMP 5
Housing / Mechanics
Housing material| PC. ABS
Dimensions| 105 x 113 x 38 mm
IP protection class| IP65
Fire protection class| UL94-V2
Weight| 240 g
CONNECTIONS
Marking
|
Function
---|---
Ethernet| PoE / Ethernet interface
V+| Power supply +
V-| Power supply –
DIMENSIONS
DELIVERY PACKAGE
- Data logger, with clamps
- Short instruction manual
- 2 batteries
- Certificate
- Velcro strips