rotronic HC2A-S Filter and Filter Carriers Owner’s Manual
- June 15, 2024
- ROTRONIC
Table of Contents
rotronic HC2A-S Filter and Filter Carriers
DESCRIPTION
Filter carriers protect the humidity and temperature sensors against mechanical damage. Filters act as a protective barrier against contaminants/pollutants that can influence the sensor. When choosing the correct combination of filter carrier and filter there are many factors to consider. Specific conditions such as high air velocities, pollutants in the air, disinfection and cleaning routines, mechanical impacts, high bioactivity, condensation, airborne chemical contaminants and required response time are some of the many considerations.
Plastic filter carrier | Metal filter carrier |
---|
• Maximum temperature 100 °C
• Mechanical protection
| • Maximum temperature 200 °C
• Mechanical protection
Overview filters
| Teflon filters| Polyethylene filters| Sintered steel filters (stainless
steel)| Wire mesh filters (metal)
Maximum temperature
(consider range of application of filter carrier)
| 200 °C| 100 °C| 200 °C| 200 °C
Protection against particulates| ✓ ✓| ✓ ✓| ✓| ✓
Protection against abrasives in the air| | | ✓ ✓ ✓| ✓
Pore size| 10 µm| 40 µm| 25 µm| 10 µm
Max. air velocity [m/s] (continuous load)| 50| 50| 70| 50
Filters & carriers| Overview probes
---|---
Range of application (temp.)| -50…100 °C| -40…85°C| –
100…200°C
Probe
| | HC2A-S| HC2A-S3| HC2A-SH| HC2A-S3H| HC2A-S-HH| HC2A-SM| ****
XD33A-S3X
| HC2A-ICxxx| HC2A-Icxxx-A| HC2A-IMxxx-M| HC2A-IExxx
-50…100 °C| SPA-PCB| ✓| | ✓|
SPA-PCW| ✓| | ✓|
-100…200°C| SPA-SS| ✓| ✓| ✓| ✓
-50…100 °C
| SPA-PCB-PE| ✓| | ✓|
SPA-PCB-PTFE| ✓| | ✓|
SPA-PCB-WM| ✓| | ✓|
SPA-PCW-PE| ✓| | ✓|
SPA-PCW-PTFE| ✓| | ✓|
SPA-PCW-WM| ✓| | ✓|
SPA-SS-PE| ✓| ✓| ✓|
SPA-PE| ✓| ✓| ✓|
-100…200°C
| SPA-SS-PTFE| ✓| ✓| ✓| ✓
SPA-SS-WM| ✓| ✓| ✓| ✓
SPA-SSS| ✓| ✓| ✓| ✓
SPA-WM| ✓| ✓| ✓| ✓
SPA-PTFE| ✓| ✓| ✓| ✓
HC2A-S / HC2A-S3 / HC2A-SH / HC2A-S3H / HC2A-S-HH / HC2A-SM
Order code| Filter carrier| Filter element| Pore
size| Range of application|
---|---|---|---|---|---
SPA-PCB| Polycarbonate, black| No filter, only filter carrier| -50…100 °C|
SPA-PCB-PE| Polyethylene, white| 40-50 µm|
SPA-PCB-PTFE| PTFE, white| 10 µm|
SPA-PCB-WM| Wire mesh 1.4401| 10 µm| ****
SPA-PCW| Polycarbonate, white| No filter, only filter carrier| -50…100 °C|
SPA-PCW-PE| Polyethylene, white| 40-50 µm| **
SPA-PCW-PTFE| PTFE, white| 10 µm|
SPA-PCW-WM| Wire mesh 1.4401| 10 µm|
SPA-PE| No filter carrier, only filter| Polyethylene, white| 40-50 µm|
SPA-PTFE| No filter carrier, only filter| PTFE, white| 10 µm| -100…200 °C|
Particulate filter / Waterproof
SPA-POM-FD2| POM white| Teflon| 2 µm| -50…100 °C| **
SPA-PE-FD2| POM white| Polyethylene| 2 µm| ****
HC2A-IC / HC2A-IM / HC2A-IE
HC2A-SM
Note
The range of application depends on the component with the smallest
temperature range.
References
- Measurement Devices & Monitoring Solutions Humidity Dewpoint etc. | PST
- Measurement Solutions - Humidity, Temperature, CO2 and Differential Pressure Measurement (EN_US)