GLOBAL INDUSTRIAL 436963 Fingertip Pulse Oximeter with OLED Display User Manual
- June 6, 2024
- GLOBAL INDUSTRIAL
Table of Contents
Fingertip Pulse Oximeter with OLED Display
Model 436963
USER MANUAL
Product Introduction and Operation Guide
Front View
Operation Method
A. Open the battery cover, and put insert 2 AAA batteries into the battery
compartment incorrect polarities, then replace the cover, (note: batteries not
included):
B. Press the bottom of the equipment and open the probe, then insert one
finger into the probe;
C. Press the power button to turn the equipment on, and the measured interface
will appear;
D. After about 8 seconds, the measurement result can be read directly from the
display screen;
E. Before reading the parameters, make sure that stable numbers of the pulse
oximeter interface have sustained more than 4 seconds;
F. The equipment will turn off automatically within 8 seconds when a finger is
left in the probe.
Battery Installation
A. Put the two AAA batteries into the battery compartment it correct
polarities (Figures).
B. Push the battery cover horizontally along the arrow shown as right.
WARNINGS:
- Battery polarities should be correctly installed, otherwise, damage may be caused to the equipment.
- Remove batteries if the equipment is not being used for a long period of time.
Lanyard installation
A. Pass the thinner end of the lanyard through the hanging hole; B. Pass the thicker end of the lanyard through the thinner end and tighten the lanyard (Figure6).
Attention for Operation
A. Before use check and confirm that the environment does not contain
combustible material, as well as to avoid high or low temperature and
humidity, but also need to pay attention to the following: a) To avoid glare
and direct sunlight exposure; b) To avoid radiation infrared or ultraviolet
radiation; c) Avoid contact with the organic solvent, mist, dust, corrosive
gases;
B. The equipment should not be used at a location or limb tied with arterial
canal or blood pressure cuff or receiving intravenous injection;
C. The equipment may not work normally on microcirculation barrier patients,
warm or rub the finger, or re-position the equipment could improve the
measurement.
D. The ray between the photodetector and the light-emitting diode should be
across the patient’s arteriole.
E. The patient should not use enamel or other makeup;
F. Do not insert a wet finger into the probe.
Notes:
A. The user should fully insert the finger into the probe,
B. It is recommended to let the LED light shine directly on the nail(Figure
7);
C. Don’t shake the finger and try to keep the patient still during the
measurement.
Functions and Menu Operation
The button operation rules
After turning on the oximeter, press and hold the power button for about 2 seconds. The oximeter will call up the parameter setting interface and set it by pressing the button. Defined here, long press indicates that the button hold time reaches 1-2s, short press indicates that the button hold time is less than 0.5s.
On parameter Interface
- Move’•• to the corresponding option, and hold the button to set Alm or Beep to on or off.
- When Alm is set to on and the measured Sp02 or PR Values go beyond the upper limit or lower limit, the Oximeter gives off an alert sound.
- When Alm is set to off and the measured values go beyond the limit, the Oximeter will not give any alert sound.
- When Beep is set to on,A ticking sound synchronized with the pulse is emitted during the measurement, and when Beep is set to off, no sound is output.
- While the “” symbol stays on the Restore option, hold the button to restore factory settings.
- Press the button to select a Brightness level ranging from I to 5. The greater the value, the greater the brightness of the screen.
On parameter Interface 2
- Press the button to switch between options. On this interface. you can set the upper limit and lower limit of Sp02 Alm and PR Alm.
- While the ••• symbol stays on the +/- option, hold the button to set the option to + or -. In.wmode, select the corresponding option and hold the button to 5 increments the upper or lower limit:in mode. hold the button to decrement the upper or lower limit.
- Move ••• to the Exit option, and hold the button to return to the monitoring interface.
Specifications
Classification
Type of Protection Against Electric Shock: II (Internally powered equipment)
Degree of Protection Against Electric Shock: Type BF-Applied part
Operating Mode: Spot checking
Degree of Protection Against Hazards of Explosion: IP22
Power Requirements
Specification of Battery: Two AAA (LR03)
Operating Current: 25-50mA
Physical Specifications
WidthHeightDepth : 57x30x31 mm
Weight: 28g (Bare machine)
Measurement Specifications
Sp02 Declared Accuracy 70%-100%: ±2digits
0% 69%: unspecified
Sp02 Display Range: 30%-99%
Sp02 Resolution : 1%.
PR Declared Accuracy : 25-250 bpm: ±3digits
PR Resolution: 1 bpm
Environmental Specifications
Temperature
Operating : +41-+104°F / +5° – +40°C
Storage/Transportation: -40-+140°F / -40° – +60°C
Humidity
Operating : 10-95%, noncondensing
Storage/Transportation : 10-95%, noncondensing
Atmosphere Pressure
Operating: 70-106kpa
Storage/Transportation: 50-107.4kpa
Display
Display Type: OLED Display;
Display Color: YM201: 0.96 inch. Yellow&Blue
YM301: 1.3 inch ,Blue
Display content: Sp02%, Pulse Rate, P1%, Bar Graph
Battery Indicator, Pulse Wave
Notes:
- The claim for oxygen saturation accuracy should be supported by clinical studies covering the entire claimed range, The fraction of inspired oxygen (Fi02) delivered to test subjects is varied to achieve a series of targeted steady-state saturation periods over the specified Sp02 accuracy range (e.g. 70 % to 100 % ), then the Sp02 accuracy is calculated by comparing Sp02 readings of the pulse oximeter to the values of Sa02 determined with a Co-Oximeter.
- The clinical trial included 11 subjects, including 6 males and 5 females, with an age range of 18 to 46 years, the subject’s skin color included dark black, medium black, light color, and white.
Maintenance, Cleanin, Disinfection
Maintenance
The equipment’s design life expectancy is about 2 years, keep your equipment
and accessories free of dust and dirt, and follow these rules:
A. Please clean the equipment before use according to chapter 6.2; Remove the
batteries inside the battery cassette if the equipment will not be operated
for a long time;
B. Replace the batteries in time when the battery voltage indicates lamps were
empty;
C. It is recommended that the equipment should be kept in a dry environment
with no corrosive gases and good ventilation anytime. The moisture and high-
light environments will affect its lifetime and even might damage the
equipment.
D. It is best to preserve the product in a place where the temperature is
between -20 to 60°C and the relative humidity is less than 95%.
E. The packed equipment can be transported by ordinary conveyance. The
equipment not be transported mixed with toxic, harmful, corrosive materials.
WARNING
- No modification of this equipment is allowed.
Cleaning
Your equipment should be cleaned on a regular basis. If there is heavy
pollution or lots of dust and sand in your place, the equipment should be
cleaned more frequently. Before cleaning the equipment, consult your
hospital’s regulations for cleaning the equipment. Recommended cleaning agents
are
a) Mild soap (diluted).
b) Ethanol (70%).
To clean your equipment, follow these rules:
a) Shut down the pulse oximeter;
b) Clean the display screen using a soft, clean cloth dampened with a glass
cleaner;
c) Clean the exterior surface of the equipment and probe using a soft cloth
dampened with the cleaner;
d) Wipe off all the cleaning solution with a dry cloth after cleaning if
necessary;
e) Dry your equipment in a ventilated, cool place. To avoid damage to the
equipment, follow these rules:
CAUTIONS
- Always dilute according to the manufacturer’s instructions or use the lowest possible concentration.
- Do not immerse part of the equipment in the liquid.
- Do not pour liquid onto the equipment or accessories.
- Never use abrasive materials (such as steel wool or silver polish), or erosive cleaners (such as acetone or acetone-based cleaners).
- If you spill liquid onto the equipment, contact us or your service personnel.
Disinfection
Clean the pulse oximeter before disinfecting it. The recommended disinfectant
is ethanol 70%. Disinfection steps are the same as cleaning.
CAUTION
- Never use ETO or formaldehyde for disinfection.
Disposal
Dispose of the pulse oximeter in accordance with the local environment and waste disposal laws and regulations.
Accessories
One lanyard.
Two AAA batteries(Optional).
One user manual.
One certificate card.
Troubleshooting
Trouble | Possible Reason | Solutions |
---|---|---|
The device can not be turned on | The batteries are drained away or almost | |
drained away. | Replace batteries. | |
The battery installation is incorrect. | Install the battery over again. | |
The device works abnormally. | Please contact the product distributor |
Trouble | Possible Reason | Solutions |
---|---|---|
The Spo2 and PR are not displayed normally | The finger size is too big or | |
small | Select the suitable size finger to measure | |
Excessive ambient light | Avoid the excessive ambient light irradiation | |
The user’s blood perfusion is very low | Warm the finger and try again | |
The display is off suddenly | The device was set to shut down automatically in | |
8 seconds when there is no correct physiological signals | Normal | |
The battery is almost drained away | Replace batteries | |
The Spo2 and are not PR displayed stably | The finger is not inserted deep | |
enough | Replace the finger and try again | |
The finger is shaking or the body is moving | Try to keep still | |
Not used in the work environment required by this manual | Please use in the |
normal working environment
The device works abnormally.| Please contact the product distributor
For additional questions or concerns, contact Global Industrial Customer Service at 1-800-645-2986 or visit www.globalindustrial.com
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>