GLOBAL INDUSTRIAL 436963 Fingertip Pulse Oximeter with OLED Display User Manual

June 6, 2024
GLOBAL INDUSTRIAL

Fingertip Pulse Oximeter with OLED Display
Model 436963
USER MANUAL

Product Introduction and Operation Guide

Front View
Operation Method

A. Open the battery cover, and put insert 2 AAA batteries into the battery compartment incorrect polarities, then replace the cover, (note: batteries not included):
B. Press the bottom of the equipment and open the probe, then insert one finger into the probe;
C. Press the power button to turn the equipment on, and the measured interface will appear;
D. After about 8 seconds, the measurement result can be read directly from the display screen;
E. Before reading the parameters, make sure that stable numbers of the pulse oximeter interface have sustained more than 4 seconds;
F. The equipment will turn off automatically within 8 seconds when a finger is left in the probe.

Battery Installation

A. Put the two AAA batteries into the battery compartment it correct polarities (Figures).
B. Push the battery cover horizontally along the arrow shown as right.
WARNINGS:

  • Battery polarities should be correctly installed, otherwise, damage may be caused to the equipment.
  • Remove batteries if the equipment is not being used for a long period of time.
Lanyard installation

A. Pass the thinner end of the lanyard through the hanging hole; B. Pass the thicker end of the lanyard through the thinner end and tighten the lanyard (Figure6).

Attention for Operation

A. Before use check and confirm that the environment does not contain combustible material, as well as to avoid high or low temperature and humidity, but also need to pay attention to the following: a) To avoid glare and direct sunlight exposure; b) To avoid radiation infrared or ultraviolet radiation; c) Avoid contact with the organic solvent, mist, dust, corrosive gases;
B. The equipment should not be used at a location or limb tied with arterial canal or blood pressure cuff or receiving intravenous injection;
C. The equipment may not work normally on microcirculation barrier patients, warm or rub the finger, or re-position the equipment could improve the measurement.
D. The ray between the photodetector and the light-emitting diode should be across the patient’s arteriole.
E. The patient should not use enamel or other makeup;
F. Do not insert a wet finger into the probe.

Notes:
A. The user should fully insert the finger into the probe,
B. It is recommended to let the LED light shine directly on the nail(Figure 7);
C. Don’t shake the finger and try to keep the patient still during the measurement.

Functions and Menu Operation

The button operation rules

After turning on the oximeter, press and hold the power button for about 2 seconds. The oximeter will call up the parameter setting interface and set it by pressing the button. Defined here, long press indicates that the button hold time reaches 1-2s, short press indicates that the button hold time is less than 0.5s.

On parameter Interface

  •  Move’•• to the corresponding option, and hold the button to set Alm or Beep to on or off.
  • When Alm is set to on and the measured Sp02 or PR Values go beyond the upper limit or lower limit, the Oximeter gives off an alert sound.
  • When Alm is set to off and the measured values go beyond the limit, the Oximeter will not give any alert sound.
  • When Beep is set to on,A ticking sound synchronized with the pulse is emitted during the measurement, and when Beep is set to off, no sound is output.
  • While the “” symbol stays on the Restore option, hold the button to restore factory settings.
  • Press the button to select a Brightness level ranging from I to 5. The greater the value, the greater the brightness of the screen.

On parameter Interface 2

  • Press the button to switch between options. On this interface. you can set the upper limit and lower limit of Sp02 Alm and PR Alm.
  • While the ••• symbol stays on the +/- option, hold the button to set the option to + or -. In.wmode, select the corresponding option and hold the button to 5 increments the upper or lower limit:in mode. hold the button to decrement the upper or lower limit.
  • Move ••• to the Exit option, and hold the button to return to the monitoring interface.

Specifications

Classification

Type of Protection Against Electric Shock: II (Internally powered equipment)
Degree of Protection Against Electric Shock: Type BF-Applied part
Operating Mode: Spot checking
Degree of Protection Against Hazards of Explosion: IP22

Power Requirements

Specification of Battery: Two AAA (LR03)
Operating Current: 25-50mA

Physical Specifications

WidthHeightDepth : 57x30x31 mm
Weight: 28g (Bare machine)

Measurement Specifications

Sp02 Declared Accuracy 70%-100%: ±2digits
0% 69%: unspecified
Sp02 Display Range: 30%-99%
Sp02 Resolution : 1%.
PR Declared Accuracy : 25-250 bpm: ±3digits
PR Resolution: 1 bpm

Environmental Specifications

Temperature
Operating : +41-+104°F / +5° – +40°C
Storage/Transportation: -40-+140°F / -40° – +60°C
Humidity
Operating : 10-95%, noncondensing
Storage/Transportation : 10-95%, noncondensing
Atmosphere Pressure
Operating: 70-106kpa
Storage/Transportation: 50-107.4kpa

Display

Display Type: OLED Display;

Display Color: YM201: 0.96 inch. Yellow&Blue
YM301: 1.3 inch ,Blue
Display content: Sp02%, Pulse Rate, P1%, Bar Graph
Battery Indicator, Pulse Wave
Notes:

  1. The claim for oxygen saturation accuracy should be supported by clinical studies covering the entire claimed range, The fraction of inspired oxygen (Fi02) delivered to test subjects is varied to achieve a series of targeted steady-state saturation periods over the specified Sp02 accuracy range (e.g. 70 % to 100 % ), then the Sp02 accuracy is calculated by comparing Sp02 readings of the pulse oximeter to the values of Sa02 determined with a Co-Oximeter.
  2. The clinical trial included 11 subjects, including 6 males and 5 females, with an age range of 18 to 46 years, the subject’s skin color included dark black, medium black, light color, and white.

Maintenance, Cleanin, Disinfection

Maintenance

The equipment’s design life expectancy is about 2 years, keep your equipment and accessories free of dust and dirt, and follow these rules:
A. Please clean the equipment before use according to chapter 6.2; Remove the batteries inside the battery cassette if the equipment will not be operated for a long time;
B. Replace the batteries in time when the battery voltage indicates lamps were empty;
C. It is recommended that the equipment should be kept in a dry environment with no corrosive gases and good ventilation anytime. The moisture and high- light environments will affect its lifetime and even might damage the equipment.
D. It is best to preserve the product in a place where the temperature is between -20 to 60°C and the relative humidity is less than 95%.
E. The packed equipment can be transported by ordinary conveyance. The equipment not be transported mixed with toxic, harmful, corrosive materials.
WARNING

  •  No modification of this equipment is allowed.
Cleaning

Your equipment should be cleaned on a regular basis. If there is heavy pollution or lots of dust and sand in your place, the equipment should be cleaned more frequently. Before cleaning the equipment, consult your hospital’s regulations for cleaning the equipment. Recommended cleaning agents are
a) Mild soap (diluted).
b) Ethanol (70%).
To clean your equipment, follow these rules:
a) Shut down the pulse oximeter;
b) Clean the display screen using a soft, clean cloth dampened with a glass cleaner;
c) Clean the exterior surface of the equipment and probe using a soft cloth dampened with the cleaner;
d) Wipe off all the cleaning solution with a dry cloth after cleaning if necessary;
e) Dry your equipment in a ventilated, cool place. To avoid damage to the equipment, follow these rules:

CAUTIONS

  • Always dilute according to the manufacturer’s instructions or use the lowest possible concentration.
  • Do not immerse part of the equipment in the liquid.
  • Do not pour liquid onto the equipment or accessories.
  • Never use abrasive materials (such as steel wool or silver polish), or erosive cleaners (such as acetone or acetone-based cleaners).
  • If you spill liquid onto the equipment, contact us or your service personnel.
Disinfection

Clean the pulse oximeter before disinfecting it. The recommended disinfectant is ethanol 70%. Disinfection steps are the same as cleaning.
CAUTION

  • Never use ETO or formaldehyde for disinfection.
Disposal

Dispose of the pulse oximeter in accordance with the local environment and waste disposal laws and regulations.

Accessories

One lanyard.
Two AAA batteries(Optional).
One user manual.
One certificate card.

Troubleshooting

Trouble Possible Reason Solutions
The device can not be turned on The batteries are drained away or almost
drained away. Replace batteries.
The battery installation is incorrect. Install the battery over again.
The device works abnormally. Please contact the product distributor
Trouble Possible Reason Solutions
The Spo2 and PR are not displayed normally The finger size is too big  or
small Select the suitable size finger to measure
Excessive ambient light Avoid the excessive ambient light irradiation
The user’s blood perfusion is very low Warm the finger and try again
The display is off suddenly The device was set to shut down automatically in
8 seconds when there is no correct physiological signals Normal
The battery is almost drained away Replace batteries
The Spo2 and  are not PR displayed stably The finger is not inserted deep
enough Replace the finger and try again
The finger is shaking or the body is moving Try to keep still
Not used in the work environment required by this manual Please use in the

normal working environment
The device works abnormally.| Please contact the product distributor

For additional questions or concerns, contact Global Industrial Customer Service at 1-800-645-2986 or visit www.globalindustrial.com

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

GLOBAL INDUSTRIAL User Manuals

Related Manuals