Joytech Healthcare DBP-6275B Arm-type Fully Automatic Digital Blood Pressure Monitor Owner’s Manual
- June 6, 2024
- Joytech Healthcare
Table of Contents
- Joytech Healthcare DBP-6275B Arm-type Fully Automatic Digital Blood
- Safety Notice
- Unit Illustration
- Important Testing Guidelines
- Quick Start
- Unit Operation
- Power On
- Power Off
- Transfer your readings
- Troubleshooting
- Health Reminder
- Maintenance
- Specifications
- Warranty
- Electromagnetic Compatibility Information
- Additional Notes
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Joytech Healthcare DBP-6275B Arm-type Fully Automatic Digital Blood
Pressure Monitor
Safety Notice
Thank you for purchasing the DBP-6275B Blood Pressure Monitor. The unit has
been constructed using reliable circuitry and durable materials. Used
properly, this unit will provide years of satisfactory use.
Measure blood pressure(systolic and diastolic)and pulse rate of adults and
adolecents age 12 through 21years of age. All functions can be used safely and
values can be read out in one LCD DISPLAY. Measurement position is on adult
upper arm only. The PATIENT is an intended OPERATOR Blood pressure measurement
determined with this device are equivalent to those obtained by a trained
observer using the cuff/ stethoscope auscultation method, within the limits
prescribed by the Recognized Consensus Standard (IEC 81060-2-30) for
electronic sphygmomanometers.
Precautions to Ensure Safe, Reliable Operation
- Do not drop the unit. Protect it from sudden jars or shocks.
- Do not insert foreign objects into any openings.
- Do not attempt to disassemble the unit.
- Do not crush the pressure cuff.
- If the unit has been stored at temperatures below 0 oC, leave it in a warm place for about 15 minutes before using it. Otherwise, the cuff may not inflate properly.
- If the unit has been stored at temperatures above 40 oC, leave it in a cool place for about 15 minutes before using it. Otherwise, the cuff may not inflate properly.
- Do not store the unit in direct sunlight, high humidity or dust.
- To avoid any possibility of accidental strangulation, keep this unit away from children and do not drape tubing around your neck.
- Ensure that children do not use the instrument unsupervised; some parts are small enough to be swallowed.
- . Some may get a skin irritation from the cuff taking frequent readings over the course of the day, but this irritation typically goes away on its own after the monitor is removed.
Important Instructions Before Use
- Do not confuse self-monitoring with self-diagnosis. Blood pressure measurements should only be interpreted by a health professional who is familiar with your medical history.
- Contact your physician if test results regularly indicate abnormal readings.
- If you are taking medication, consult with your physician to determine the most appropriate time to measure your blood pressure. NEVER change a prescribed medication without first consulting with your physician.
- Individuals with serious circulation problems may experience discomfort. Consult your physician prior to use.
- For persons with irregular or unstable circulation resulting from diabetes, liver disease, arteriosclerosis or other medical conditions, there may be variations in blood pressure values measured at the wrist versus at the upper arm. monitoring the trends in your blood pressure taken at either the arm or the wrist is nevertheless useful and important.
- People suffering from vascular constriction, liver disorders or diabetes, people with cardiac pacemakers or a weak pulse, and women who are pregnant should consult their physician before measuring their blood pressure emselves. Different values may be obtained due to their condition.
- People suffering from arrhythmias such as atrial or ventricular premature beats or atrial fibrillation only use this blood pressure monitor in consultation with your doctor. In certain cases oscillometric measurement method can oduce incorrect readings.
- Too frequent measurements can cause injury to the patient due to blood flow interference.
- . The cuff should not be applied over a wound as this can cause further injury.
- DO NOT attach the cuff to a limb being used for IV infusions or any other intravascular access, therapy or an arterio-venous (A-V) shunt. The cuff inflation can temporarily block blood flow, potentially causing harm to the patient.
- The cuff should not be placed on the arm on the side of a mastectomy. In the case of a double mastectomy use the side of the least dominant arm.
- Pressurization of the cuff can temporarily cause loss of function of simultaneously used monitoring equipment on the same limb.
- A compressed or kinked connection hose may cause continuous cuff pressure resulting in blood flow interference and potentially harmful injury to the patient.
- Check that operation of the unit does not result in prolonged impairment of the circulation of the patient.
- Product is designed for its intended use only. Do not misuse in any way.
- Product is not intended for infants or individuals who cannot express their intentions.
- Prolonged over-inflation of the bladder may cause ecchymoma of your arm.
- Do not disassemble the unit or arm cuff. Do not attempt to repair.
- Use only the approved arm cuff for this unit. Use of other arm cuffs may result in incorrect measurement results.
- The system might produce incorrect readings if stored or used outside the manufacturer’s specified temperature and humidity ranges. Make sure to store the blood pressure monitor, children, pets and pests are outside of accessible range.
- Do not use the device near strong electrical or electromagnetic fields generated by cell phones or other devices, they may cause incorrect readings and interference or become interference source to the device.
- Do not mix new and old batteries simultaneously.
- Replace batteries when Low Battery Indicator“ ”appears on screen. Replace both batteries at the same time.
- Do not mix battery types. Long-life alkaline batteries are recommended.
- Remove batteries from device when not in operation for more than 3 months.
- Dispose batteries properly; observe local laws and regulations.
- Only use a recommended AC adaptor double-insulated complying with EN 60601-1 and EN 60601-1-2. An unauthorized adapter may cause fire and electric shock.
- Advising operator that Instruction manual/ Booklet must be consulted.
- . Do not use the device during transport vehicles for influencing measurement accuracy, such as patient transport in an ambulance or helicopter.
- Contains small parts that may cause a chocking hazard if swallowed by infants.
- Please align the polarities of each battery with the +ve and -ve signs imprinted on the battery housing when you replace the batteries .
- Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
- Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the unit, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Unit Illustration
Monitor Unit
Arm Cuf f Medium size cuff (fits arm circumference: 22.0 cm -36.0 cm).
If air is leaking from the arm cuff, replace the arm cuff with a new one. It is generally recomm ended to have the cuff replaced timely to ensure correct functioning and accuracy. Please consult your local authorized Sejoy distributor or dealer.
Display
Contents
Important Testing Guidelines
- Avoid eating, exercising, and bathing for 30 minutes prior to testing.
- Sit in a calm environment for at least 5 minutes prior to testing.
- Do not stand while testing. Sit in a relaxed position while keeping your arm level with your heart.
- Avoid speaking or moving body parts while testing.
- While testing, avoid strong electromagnetic interference such as microwave ovens and cell phones.
- Wait 3 minutes or longer before re-testing.
- Try to measure your blood pressure at the same time each day for consistency.
- Test comparisons should only be made when monitor is used on the same arm, in the same position, and at the same time of day.
- This blood pressure monitor is not recommended for people with severe arrhythmia.
- Do not use this blood pressure monitor if the device is damaged.
Any blood pressure recording can be affected by the following factors:
- The position of the subject, his or her physiologic condition;
- The performance and accuracy of the device;
- Cuff size: too small cuff (bladder) will produce a higher blood pressure value than usual, too big cuff (bladder) will produce a lower blood pressure value;
- Measuring position does not keep level with your heart;
- Speaking or moving body parts while testing;
- Not relaxing for about 5 minutes before taking the measurement.
Quick Start
- Install batteries
- Insert cuff air plug into the behind side of monitor unit. (See Figure B)
- Remove thick clothing from the arm area.
- Rest for several minutes prior to testing. Sit down in a quiet place comfortably, back and arm support on a desk or table,with your legs uncrossed,your arm resting on a firm and your feet flat on the floor. (See Figure C )
- Apply cuff to your left arm and middle of the cuff at the level of your heart. Bottom of cuff should be placed approximately 1-2cm (1/2”) above elbow joint. (See Figures D&E)
- Press ” START/STOP ” Button to start testing.
Unit Operation
Battery Installation
Slide battery cover off as indicated by arrow. Install 4 new AA alkaline
batteries according to polarity. Close battery cover.
AC Adapter jack is on the back side of the monitor. Medical AC adapter(DC 5.0 V,1000mA) can be used with the device (recommended, not provided). The adapter connect pin should be positive inside and negtive outside with a 2.1mm coaxial joint. Do not use any other type of AC adaptor as it may harm the unit.
Note :Power supply is specified as part of ME EQUIPMENT.
System Settings
With power off, press “SET” button to activate System Settings. The Memory Group icon flashes.
-
Select Memory Group While in the System Setting mode, you may accumulate test results into 2 different groups. This allows multiple users to save individual test results (up to 150 memories per group.) Press ” MEM ” button to choose a group setting. Test results will automatically store in each selected group.
-
Time/Date setting Press“SET” button again to set the Time/Date mode. Set the year first by adjusting the “MEM” button.
Press“SET” button again to confirm current month. Continue setting the date, hour and minute in the same way. Every time the “SET” button is pressed, it will lock in your selection and continue in succession ( month,day,hour, minute,12/24 hours) -
Time Format setting Press the “ SET ”button again to set the time fornat setting mode.Set the time format by adjusting the“MEM” button. EU means European Time US means U.S Time.
-
Unit Setting Press “SET” button to enter unit setting mode. Set format bypressing the “MEM” button.
-
Voice Setting Press “SET” button to enter voice setting mode. Set voice format ON or OFF by pressing the “MEM” button.
-
Volume Setting Press “SET” button to enter volume setting mode. Set the voice volume by adjusting the “MEM” button . There are six volume levels.
-
Saved Settings While in any setting mode, press ” START/STOP ” button to turn the unit off. All information will be saved.
Note : If unit is left on and not in use for 3 minutes, it will automatically save all information and shut off.
Applying the Arm Cuff
- Firmly insert air plug into opening located on behind side of monitor unit.
- With sticky nylon section facing outward, insert end of cuff underneath metal ring of cuff.
- Fasten cuff about 1-2cm (0.4-0.8”)above the elbow joint. For best results apply cuff to bare arm and keep level with heart while testing
Note: Do not insert air plug into opening located on right side of monitor unit. This opening is designed for an optional power supply only.
Testing
Power On
**** Press and hold “START/STOP” button to turn the unit on. The LCD screen will appear for one second as unit performs a quick diagnosis. A voice tone will indicate when unit is ready for testing.
Note : Unit will not function if residual air from previous testing is present in cuff. The LCD will flash ” ” until pressure is stabilized.
Testing After cuff inflation, air will slowly rise as indicated by the corresponding cuff pressure value. A flashing ” ” will appear simultaneously on screen signaling heart beat detection.
Result Display The screen will display measurements for systolic and diastolic blood pressure with voice broadcast. A indicator representing the current measurement will appear next to the corresponding WHO Classification.
Note : Refer to Page 23~24 for detail WHO Blood Pressure Classification Information.
Irregular Heartbeat IndicatorIf the monitor detects an irregular heart rhythm two or more times during the measuring process, the Irregular Heartbeat Symbo Il ” “HB appears on screen along with measurement results. Irregular heartbeat rhythm is defined as rhythm that is either 25% slower or faster than the average rhythm detected while measuring systolic blood pressure and diastolic blood pressure. Consult your physician if the Irregular Heartbeat Symbol IHB” ” frequently appears with your test results.
Power Off
The ” START/STOP ” button can be pressed to turn off the unit in any mode. The unit can turn off the power itself about 3 minutes no operation in any mode.
Arm Shake Indicator
If there is arm movement during the measurement, ” ” may be shown. Indicates
that it may lead to abnormal accurate measurement results. At this time,the
LCD will display ” Err”.
Cuff loose Indicator
When starting the measurement, ” When the cuff is too loose, ” and start measuring again. OK “will be displayed when the cuff is properly wound. “will be displayed. At this time, please wear the cuff correctly
Last 3 Tests Average
With power off,press the “MEM” button to activate screen display. After the unit performs a self-diagnosis ,the screen will display the average test results from the last 3 readingsof the last group used. The “AVG” symbol will appear along with the corresponding WHO Blood Pressure Indicator. The Memory Check mode can be accessed by pressing “MEM” button. To check the average results from other groups, select the desired group first prior to activating “SET” button in the off position.(See “Select Memory Group” on Page 11)
Memory Check
You may check past test results by using the “MEM” button. The most recent test result and oldest test result in memory can be viewed by pressing and holding the “MEM” button. Upon activating test results. you can press the “MEM” button to scroll through all test results stored in memory.
Memory Deletion
Memory for a selected group may be deleted while in Memory Check mode. Press
and hold the ” SET ” button for approximately 3 seconds to delete all memory
records from theselected group. with voice broadcast “Memory Clear” and then
transfer into testing mode. Press the “START/STOP ” button to turn the unit
off.
Note: Memory cannot be recovered once it has been deleted.
Low Battery Indicator
The unit will broadcast “Low Battery” when battery life is depleting and
unable to inflate cuff for testing. The ” ” appears simultaneously for
approximately 5 seconds prior to shutting off. Replace batteries at this time.
No memory loss will occur throughout this process.
Static Pressure Measurement
In the power down state, press and hold the ” START/STOP” button, and theninstall the batteries. Until the LCD screen is full, release the ” START/STOP” button. When the LCD screen displays the double zero, the bloodpressure meter is in static state. Software version is displayed:10 is a software version in the figure.
Note : Only Service personnel permitted to access to this mode, the mode unavailable in normal use.
Bluetooth requirements
The monitor requires a device with: .
- Bluetooth 4.0 or later
- Android 5.0 or later
- IOS 9.0 or later And works with:
- iphone , iPod, iPad
- Android Phones and Tablets
Bluetooth connection
-Using for the first time
- Download the free “JoyHealth” App: On your mobile phone or tablego to www.sejoy.com.
- Open the App on your phone or tablet. If requested, you should enable Bluetooth on your device. You can enable Bluetooth under the Settings menu on your smart phone or table.
- Create a new user login, or login with your existing user name and password.
- Selection device “Blood pressure monitor”.
Pairing your monitor with a Smart Device
Open the “blood pressure monitor” and follow the pairing instructions shown on your smart phone. The date and time on your monitor will automatically be set when you pair it with your smart device.
Confirm that your monitor is connected successfully.
Transfer your readings
As soon as your measurement is complete, open the app on your smart phone to
transfer your reedings.
Note : On the paired smartphone, Bluetooth must be enabled. 2. You can
view your blood pressure readings on the app.
Troubleshooting
Problem | Possible Cause | Solution |
---|
Blood pressure results are not within typical range
|
Cuff is too tight or not properly positioned on the arm
| Firmly reposition cuff approximately1-2cm (1/2″) above the elbow joint
( See Page 12)
Inaccurate test results due to body movement or monitor movement
|
Sit in a relaxed position with arm placed near heart. Avoid speaking or moving body parts while testing. Make sure the monitor unit is placed in a stationary position throughout the testing period.
( See Page 7)
” “displayed
|
Cuff fails to inflate properly
| Make sure hose is properly fastened to cuff and monitor unit
Improper operation
|
Read user manual carefully and re-test properly.
Pressurization is over cuff rated pressure 300mmHg|
Read user manual carefully and re-test properly.
Connection failure./ Data is not being transmitted
| The blood pressure monitor might not be porperly placed within the smart
device’s tranmission range and is too far from the smart device.| If there are
no causes of data transmission interference found near the blood pressure
monitor, move the blood pressure monitor within 16ft.(5m) of the smart device
and try again
The blood pressure did not pair successfully to the smart device| Try to pair
the devices once again.
The application on the smart device is not ready.| Check the application then
try sending the data again.
Blood Pressure
Blood pressure is the force of blood pushing against the walls of arteries. It
is typically measured in millimeters of mercury (mmHg.) Systolic blood
pressure is the maximum force exerted against blood vessel walls each time the
heart beats. Diastolic blood pressure is the force exerted on blood vessels
when the heart is resting between beats.
An individual’s blood pressure frequently changes throughout the course of a
day. Excitement and tension can cause blood pressure to rise, while drinking
alcohol and bathing can lower blood pressure. Certain hormones like adrenaline
(which your body releases under stress) can cause blood vessels to constrict,
leading to a rise in blood pressure.
If these measuring numbers become too high, it means the heart is working
harder than it should.
WHO Blood Pressure Classification Indicator
The DBP-6275B is equipped with a classification indicator based on established guidelines from the World Health Organization. The chart below (color coded on monitor unit) indicates test results.
Health Reminder
Hypertension is a dangerous disease that can affect the quality of life. It can lead to a lot of problems including heart failure, kidney failure, and cerebral hemorrhaging. By maintaining a healthy lifestyle and visiting your physician on a regular basis, hypertension and relative diseases are much easier to control when diagnosed in their early stages.
Note: Do not be alarmed if an abnormal reading occurs. blood pressure occurs after 2-3 readings are taken at the same time extended period of time. Consult your physician if test results remain abnormal.
Maintenance
Avoid dropping, slamming, or throwing the unit.
Avoid extreme temperatures. Do not expose unit directly under sunshine.
When cleaning the unit, use a soft fabric and lightly wipe with mild detergent. Use a damp cloth to remove dirt and excess detergent.
Cuff Cleaning and Disinfection:
- A) Spread the cuff (skin-contact surface) upwards onto a clean table. Use a damp clean cloth (water-based) to wipe the skin-contact surface with a force.
- B) Soak the cloth clean with drinking water and wring it dry. Repeat A) with the damp cloth (water-based) for 3 times.
- C) Apply 70%-80% alcohol to a new cloth (or 75% alcohol cotton-ball), use it to wipe the skin-contact surface with a force. Then soak the cloth with the alcohol again (or change a new 75% alcohol cotton-ball), repeat the disinfection procedure for 3 times.
- D) When the disinfection towards the skin-contactsurface is finished,wipe thenon-skin contact surface with a cloth (alcohol-based) or alcohol cotton-ball thoroughly for3 times.
- E) Leave the cuff naturally dry,then it is ready for reuse.
Notice: Do not soak in water or splash water on it. Do not use petrol, thinners or similar solvents. Remove batteries when not in operation for an extended period of time. Do not disassemble product
It is recommended the performance should be checked every 2 years. Expected service life: Approximately three years at 10 tests per day. No service and maintenance while it is in use and maintenance only be performed by service personnel. Service and maintenance require parts, repair, technical support will be provided.
Specifications
Product Description | Arm-type Fully Automatic Blood Pressure Monitor |
---|---|
Model | DBP-6275B |
Display | LCD Digital Display Size:118mm×65mm (4.65″ x 2.56″) |
Measurement Method | Oscillometric Method |
Measurement Range
| Systolic Pressure| 60mmHg~260mmHg
Diastolic Pressure| 30mmHg~200mmHg
Pressure| 0mmHg~299mmHg
Pressure| ±3mmHg
Pulse| 30 ~ 180 Beats/Minute
Pulse| ±5%
Pressurization| Automatic Pressurization
Memory| 2×60 Memories in Two Groups with Date and Time
Function
| Irregular Heartbeat Detection
WHO Classification Indicator
Last 3 Tests Average
Low Battery Detection
Automatic Power-Off
Voice
Backlight
Bluetooth
Power Source| 3 AAA batteries or Medical AC Adapter(DC5.0V, 1000mA)
(recommended, not provided)
Battery Life| Approximately 2 months at 3 tests per day
Unit Weight| Approx.259g (9.14 oz.) (excluding battery)
Unit Dimensions| Approx.128 x 130 x 44mm (5.04″ x 5.12″ x 1.73″ )(L x W x H)
Cuff Circumference| Medium cuff: Fits arm circumference 22-36 cm
Operating Environment
| Temperature| 10℃ ~ 40℃ (50℉~104℉)
Humidity| 15% ~ 93% RH
Operating Environment| Pressure| 800hPa~1060hPa
---|---|---
Storage Environment
| Temperature| -25℃~55℃ (-13℉~131℉)
Humidity| ≤93% RH
Transport Environment
| Temperature| -25℃~55℃ (-13℉~131℉)
Humidity| ≤93% RH
Bluetooth
| Modulation Type| GFSK
Version| 5.0.1 BT Signal mode
Operation frequency| 2. 4GHz( 2400~ 2483. 5MHz)
Antenna gain| 0.5 dBi
Bandwidth| 2.0 MHz
Classification|
Internal Powered Equipment,Type BF , Cuff is the Applied Part
Ingress Protection rating|
IP 20,Indoor Used Only
Battery Shelf life| 60 months
Battery Storage Temperature| -25℃~55℃ (-13℉~131℉)
Specifications are subject to change without notice.
This Blood Pressure Monitor complies with the European regulations and bears
the
CE mark“CE 0197”.This blood pressure monitor also complies with mainly
following standards
(included but not limited):
Safety standard:
EN 60601-1 Medical electrical equipment part 1: General requirements for
safety
EMC standard:
EN 60601-1-2 Medical Electrical Equipment — Part 1-2: General Requirements For
Basic Safety And Essential Performance — Collateral Standard: Electromagnetic
Disturbances – Requirements And Tests.
Performance standards:
IEC80601-2-30, Medical electrical equipment – Part 2-30: Particular
requirements for the basic safety and essential performance of automated non-
invasive sphygmomanometers. EN 1060-3 Non-invasive sphygmomanometers –
Supplementary requirements for electromechanical blood pressure measuring
systems.
ISO 81060-2, non-invasive sphygmomanometers – part 2: clinical validation of
automated measurement type.
Warranty
The Blood Pressure Monitor is guaranteed for 2-year from the date of purchase. If the Blood Pressure Monitor does not function properly due to defective components or poor workmanship, we will repair or replace it freely. The warranty does not cover damages to your Blood Pressure Monitor due to improper handling. Please contact local retailer for details.
The device satisfies the EMC requirements of the international standard IEC 60601-1-2. The requirements are satisfied under the conditions described in the table below. The device is an electrical medical product and is subject to special precautionary measures with regard to EMC which must be published in the instructions for use. Portable and mobile HF communications equipment can affect the device. Use of the unit in conjunction with non-approved accessories can affect the device negatively and alter the electromagnetic compatibility. The device should not be used directly adjacent to or between other electrical equipment.
Electromagnetic Compatibility Information
Table 1
Guidance and declaration of manufacturer-electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environmen
Emissions test| Compliance| Electromagnetic environment
-guidance
Radiated emission CISPR 11
|
Group 1, class B.
| The device uses RF energy only for its internal function. Therefore, its emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Conducted emission CISPR 11
|
Group 1, class B.
| The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2|
Class A
Voltage fluctuations/ flicker emissions IEC 61000-3-3|
Complies
Table 2
Guidance and declaration of manufacturer-electromagnetic immunity
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in
such an environm- ent.
IMMUNITY test| IEC 60601
test level
| Compliance level| Electromagnetic environment
-guidance
Electrostatic discharge (ESD) IEC 61000-4-2
| ± 8 kV
contact
±2 kV,±4 kV,
±8 kV,
±15 kV air
| ± 8 kV
contact
±2 kV,±4 kV,
±8 kV,
±15 kV air
| Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Electrostatic transient/burst IEC 61000-4-4
|
± 2 kV ,
100kHz, for AC power port
|
± 2 kV ,
100kHz, for AC power port
|
Mains power quality should be that of a typical commercial or hospital environment.
Surge
IEC 61000-4-5
|
±0.5kV, ±1kV
(differential mode)
|
±0.5kV, ±1kV
(differential mode)
|
Mains power quality should be that of a typical commercial or hospital environment.
| 0 % UT;
0,5 cycle
At 0°, 45°, 90°,
135°, 180°, 225
, 270° and 315°
| 0 % UT;
0,5 cycle
At 0°, 45°, 90°,
° 135°, 180°, 225
, 270° and 315°
|
° Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interrupti- ons and voltage variations on p- ower supply in- put lines| 0 % UT; 1 cycle and
70 % UT;
25/30 cycles Single phase: at 0°
| 0 % UT; 1 cycle and
70 % UT;
25/30 cycles Single phase: at 0°
IEC 61000-4-11| 0 % UT;
250/300 cycle
| 0 % UT;
250/300 cycle
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
|
30 A/m; 50Hz or 60Hz
|
30 A/m; 50Hz or 60Hz
| Power frequency magnetic fields should be at levels charactertic of a
typical location in a typical comme- rcial or hospital environment.
Guidance and declaration of manufacturer-electromagnetic immunity
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in
such an environm- ent.
IMMUNITY test| IEC 60601
test level
| Compliance level| Electromagnetic environment
-guidance
Radiated RF EM fields
IEC 61000-4-3
|
3V/m or 10 V/m 80MHz-2.7
Ghz 80%AM at 1kHz
|
3V/m or 10 V/ 80MHz-2.7
Ghz 80%AM at 1kHz
|
Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
mRecommended seperation distance 80 MHz to 800 MHz 800 MHz to 2.7 Ghz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the
following symbol:
Conducted disturbances Induced by RF fields
IEC 61000-4-6
|
3 V in 0.15
MHz- 80 MHz 6 V in ISM
and/or amateur radio bands between 0.15
MHz and 80
MHz 80 %
AM at 1kHz
|
3 V in 0.15
MHz- 80 MHz 6 V in ISM
and/or amateur radio bands between 0.15
MHz and 80
MHz 80 %
AM at 1kHz
|
Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than th recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended seperation distance 80 MHz to 800 MHz 800 MHz to 2.7 Ghz where P is the maximum output power rating of the transmitter in watts (W) according to the transm- itter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be les
than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:
TABLE 3
Guidance and declaration of manufacturer-electromagnetic immunity
Nowadays, many RF wireless equipments have being used in various healthcare
locations where medical equipment and/or systems are used. When they are used
in close proximity to medical equipment and/or systems, the medical equipment
and/or systems’ basic safety and essential performance may be affected. Arm-
type Fully Automatic Digital Blood Pressure Monitor has been tested with the
immunity test level in the below table and meet the related requirements of
IEC 60601-1-2:2014. The customer and/or user should help keep a minimum
distance between RF wireless communications equipment and this medical
equipment and/or systems as recommended below.
Test frequency (MHz)| Band (MHz)|
Service
|
Modulation
| Maximum power (W)| Distance (m)| Immunity test level (V/m)
385| 380-390| TETRA 400| Pulse
modulation 18Hz
| 1.8| 0.3| 27
450| 430-470| GMRS 460
FRS 460
| FM
± 5 kHz deviation 1 kHz sine
| 2| 0.3| 28
710|
704-787
|
LTE
Band 13, 17
|
Pulse modulation
217Hz
|
0.2
|
0.3
|
9
745
780
810|
800-960
| GSM 800/900, TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
|
Pulse modulation
18Hz
|
2
|
0.3
|
28
870
930
1720|
1700-1990
|
GSM 1800;
CDMA 1900;
GSM 1900; DECT;
LTE Band 1, 3,
4, 25; UMTS
|
Pulse modulation
217Hz
|
2
|
0.3
|
28
1845
1970
2450| 2400-2570| Bluetooth,WLAN,
802.11 b/g/n,RFID 2450,LTE Band 7
| Pulse modulation 217Hz| 2| 0.3| 28
5240|
5100-5800
|
WLAN 802.11
a/n
|
Pulse modulation
217Hz
|
0.2
|
0.3
|
9
5500
5785
TABLE 4
Recommended separation distances between portable and mobile RF
communications equipment and the device
The device is intended for use in an electromagnetic environment in which
radiated therefore disturbances are controlled. The customer or the user of
the device can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment
(transmitters) and the device as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum output power of transmitter
W
| Separation distance according to frequency of transmitter m
80 MHz to 800 MHz| 800 MHz to 2.7 GHz
0.01| 0.12| 0.23
0.1| 0.38| 0.73
1| 1.2| 2.3
10| 3.8| 7.3
100| 12| 23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Additional Notes
Important Instructions Before Use
1. WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
2. WARNING: PORTABLE RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of Arm- type Fully Automatic Digital Blood Pressure Monitor, including cables specified by the MANUFACTURER. Otherwise, degradation of the performance of this equipment could result. 3.The software identifier refer to the software evaluation report , and the file code is JYRJ200921006.
4. verify manometer pressure accuracy:
In the power down state, press and hold the ” START/STOP” button, and theninstall the batteries. Until the LCD screen is full, release the ” START/STOP” button.
When the LCD screen displays the double zero, the bloodpressure meter is in static state.
At this point, 500ml gas capacity, calibrated standard pressure gauge and manual pressure device can be connected to the sphygmomanometer through the sleeve interface of the sphygmomanometer, and manual pressure can be applied to the effective display range of the sphygmomanometer, and then the difference between the reading of the sphygmomanometer and that of the standard pressure gauge can be compared.This mode can be used to verify manometer pressure accuracy.
5. Contraindications:
Product is not intended for infants or individuals who cannot express their intentions. 6.Intended Use
The digital blood pressure monitor are reusable for clinical and home use and are non-invasive blood pressure measurement systems designed to measure the systolic and diastolic blood pressure and pulse rate of adolescents and adults individual by using a non-invasive technique, which is a well-known technique in the market called the “oscillometric method”.
it can measure the systolic blood pressure, diastolic blood pressure and pulse rated on up-arm,and the device is reusable for clinical or home use.
7.The patient is the operator:
the PATIENT is an intended OPERATOR.
the PATIENT Do not carry out other maintenance operations except to replace the battery. 8.WARNING:
Do not modify this equipment without authorization of the manufacturer.
9. ESSENTIAL PERFORMANCE Maintenance advice:
Pressure calibration will be carried out when this product leaves the factory. Patients can use the method described in the section “Verify Manometer Pressure Accuracy” to verify the accuracy. If the accuracy deviation is large, please contact the manufacturer to recalibration.
10. Mechanical strength and resistance to heatThe resistance to heat will be retained by device during the EXPECTED SERVICE LIFE of the ME EQUIPMENT.
| 11. Do not place the blood pressure monitor and cuff at will. It will cause asphyxiation if the child swallows or twine around his neck.
12. The cuff and the case of the blood pressure monitor have been tested for biocompatibility and do not contain allergenic or harmful materials.Please stop using it if allergy occurs during use. 13.Warning:
Non-professionals do not modify the equipment, otherwise it will make the equipment measurement is not accurate.
14. Warning:
Do not expose the equipment for a long time, otherwise it will reduce the performance of the equipment.
15. Warning:
This device is not used for children and pets 16.Clean:
The equipment can be cleaned by lay operator according to the cleaning procedures in the instructions
17.Warning:
Do not use a damaged cuff for blood pressure measurement. 18.Warning:
When measuring with the cuff, if the tester feels seriously uncomfortable, press the button of the blood pressure monitor to deflate the cuff, or remove the cuff directly from the arm.
19. Warning:
If an unexpected reading occurs, the operator can take several more measurements and consult a doctor.
20. Warning:
This equipment is used outside the specified environment, may damage the equipment, and may be inaccurate measurement.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product indicates that it should not be disposed with other household waste at the end of its life. To prevent potential harm to the environment or to human health, please separate this product from other types of wastes and recycle it responsibly. When disposing this type of product, contact the retailer where product was purchased or contact your local government office for details regarding how this item can be disposed in an environmentally safe recycling center.
Business users should contact their supplier and check the terms and conditions of the purchasing agreement. This product should not be mixed with other commercial wastes for disposal. This product is free of hazardous materials.
---|---