Joytech Healthcare BM 92 Arm Type Fully Automatic Digital Blood Pressure Monitor Owner’s Manual
- June 4, 2024
- Joytech Healthcare
Table of Contents
- Joytech Healthcare BM 92 Arm Type Fully Automatic Digital Blood Pressure
- Safety Notice
- Unit Illustration
- Display
- Contents
- Important Testing Guidelines
- Quick Start
- Unit Operation
- Troubleshooting
- Blood Pressure Information
- Maintenance
- Specifications
- Warranty
- Electromagnetic Compatibility Information
- Additional Notes
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Joytech Healthcare BM 92 Arm Type Fully Automatic Digital Blood Pressure
Monitor
Contact Information
The lay operator or lay responsible or ganization should contact the
manufacturer or the representative of manufacturer.
for assistance,if needed,in setting up,using or maintaining the product,or to
report unexpected operation or events.
Manufactured by JOYTECH Healthcare Co.,Ltd.
No.365, Wuzhou Road,Yuhang Economic Development Zone, Hangzhou City,311100
Zhejiang,China
Email: info@sejoy.com
Telephone: +86-571-81957767
Fax: +86-571-81957750
Safety Notice
Thank you for purchasing the BM 92 Blood Pressure Monitor. The unit has been constructed using reliable circuitry and durable materials. Used properly, this unit will provide years of satisfactory use.
Indications for use: The Arm-type Fully Automatic Digital Blood Pressure
monitors are intended to measure blood pressure(systolic and diastolic) and
pulse rate of adults and adolescents over 12 years of age.
All functions can be used safely and values can be read out in one LCD
DISPLAY. Measurement position is on adult upper arm only.
The PATIENT is an intended OPERATOR.
Blood pressure measurement determined with this device are equivalent to those
obtained
by a trained observer using the cuff/ stethoscope auscultation method, within
the limits prescribed by the Recognized Consensus Standard (IEC 80601-2-30)
for electronic sphygmomanometers.
Precautions to Ensure Safe, Reliable Operation
- Do not drop the unit. Protect it from sudden jars or shocks.
- Do not insert foreign objects into any openings.
- Do not attempt to disassemble the unit.
- Do not crush the pressure cuff.
- If the unit has been stored at temperatures below 0 ℃, leave it in a warm place for about 15 minutes before using it. Otherwise, the cuff may not inflate properly.
- If the unit has been stored at temperatures above 40 ℃, leave it in a cool place for about 15 minutes before using it. Otherwise, the cuff may not inflate properly.
- Do not store the unit in direct sunlight, high humidity or dust.
- To avoid any possibility of accidental strangulation, keep this unit away from children and do not drape tubing around your neck.
- Ensure that children do not use the instrument unsupervised; some parts are small enough to be swallowed.
- Some may get a skin irritation from the cuff taking frequent readings over the course of the day, but this irritation typically goes away on its own after the monitor is removed.
Important Instructions Before Use
-
Do not confuse self-monitoring with self-diagnosis. Blood pressure measurements should only be interpreted by a health professional who is familiar with your medical history.
-
Contact your physician if test results regularly indicate abnormal readings.
-
If you are taking medication, consult with your physician to determine the most appropriate time to measure your blood pressure. NEVER change a prescribed medication without first consulting with your physician.
-
Individuals with serious circulation problems may experience discomfort. Consult your physician prior to use.
-
For persons with irregular or unstable circulation resulting from diabetes, liver disease, arteriosclerosis or other medical conditions, there may be variations in blood pressure values measured at the wrist versus at the upper arm. Monitoring the trends in your blood pressure taken at either the arm or the wrist is nevertheless useful and important.
-
People suffering from vascular constriction, liver disorders or diabetes, people with cardiac pacemakers or a weak pulse, and women who are pregnant should consult their physician before measuring their blood pressure themselves. Different values may be obtained due to their condition.
-
People suffering from arrhythmias such as atrial or ventricular premature beats or atrial fibrillation only use this blood pressure monitor in consultation with your doctor. In certain cases oscillometric measurement method can produce incorrect readings.
-
Too frequent measurements can cause injury to the patient due to blood flow interference.
-
The cuff should not be applied over a wound as this can cause further injury.
-
DO NOT attach the cuff to a limb being used for IV infusions or any other intravascular access, therapy or an arterio-venous (A-V) shunt. The cuff inflation can temporarily block blood flow, potentially causing harm to the patient.
-
The cuff should not be placed on the arm on the side of a mastectomy. In the case of a double mastectomy use the side of the least dominant arm.
-
Pressurization of the cuff can temporarily cause loss of function of simultaneously used monitoring equipment on the same limb.
-
A compressed or kinked connection hose may cause continuous cuff pressure resulting in blood flow interference and potentially harmful injury to the patient.
-
Check that operation of the unit does not result in prolonged impairment of the circulation of the patient.
-
Product is designed for its intended use only. Do not misuse in any way.
-
Product is not intended for infants or individuals who cannot express their intentions.
-
Prolonged over-inflation of the bladder may cause ecchymoma of your arm.
-
Do not disassemble the unit or arm cuff. Do not attempt to repair.
-
Use only the approved arm cuff for this unit. Use of other arm cuffs may result in incorrect measurement results.
-
The system might produce incorrect readings if stored or used outside the manufacturer’s specified temperature and humidity ranges. Make sure to store the blood pressure monitor, children, pets and pests are outside of accessible range.
-
Do not use the device near strong electrical or electromagnetic fields generated by cell phones or other devices, they may cause incorrect readings and interference or become interference source to the device.
-
Do not mix new and old batteries simultaneously.
-
Replace batteries when Low Battery Indicator appears on screen. Replace both batteries at the same time.
-
Do not mix battery types. Long-life alkaline batteries are recommended.
-
Remove batteries from device when not in operation for more than 3 months.
-
Dispose batteries properly; observe local laws and regulations.
-
Only use a recommended class II AC Medical approved adaptor which comply with 2MOPP, and pass IEC 60601-1, IEC 60601-1-2(or FCC) standard . An unauthorized adapter may cause fire and electric shock.
-
Advising operator that Instruction manual/ Booklet must be consulted.
-
Do not use the device during transport vehicles for influencing measurement accuracy such as patient transport in an ambulance or helicopter.
-
Contains small parts that may cause a chocking hazard if swallowed by infants.
-
Please align the polarities of each battery with the +ve and -ve signs imprinted on the battery housing when you replace the batteries .
-
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation.
If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally. -
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the unit, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
-
the blood pressure monitors equipped with bluetooth(BT) module is only transmit historical blood pressure (BP) from the subject device to a user-supplied digital device, and is not intended for active patient monitoring.
-
It cannot be maintained or repaired during normal use.
-
The service life of the product is about 5 years, the service life of the cuff is about 2 years, and the battery is about 3 months.
WARNING SIGNS AND SYMBOLS USED
Federla Commulcation Commission (FCC) Interference Statement
- This device complies with part 15 of the FCC Rules. Operation is subject to the condition that this device does not cause harmful interference.
- This device is verified to comply with part 15 of the FCC Rules for use with cable television service.
- This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
- This device may not cause harmful interference, and
- this device must accept any interference received, including interference that may cause undesired operation. Please note that changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
- This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
- This equipment complies with radio frequency exposure limits set forth by the FCC for an uncontrolled environment.
- This device must not be co-located or operating in conjunction with any other antenna or transmitter.
- Essential performance:
Electrosurgery interference recovery | Refer 202.6.2.101 | IEC 80601-2-30 |
---|---|---|
Limits of the error of the manometer | Refer 202.12.1.102 | IEC 80601-2-30 |
Reproducibility of the BLOOD PRESSURE DETERMINATION | Refer 201.12.1.107 | IEC |
80601-2-30
Unit Illustration
Monitor Unit
Arm Cuff Medium size cuff (fits arm circumference: 22.0 cm -36.0 cm).
If air is leaking from the arm cuff, replace the arm cuff with a new one. It is generally recomm ended to have the cuff replaced timely to ensure correct functioning and accuracy. Please consult your local authorized Sejoy distributor or dealer.
Display
Contents
Important Testing Guidelines
- Avoid eating, exercising, and bathing for 30 minutes prior to testing.
- Sit in a calm environment for at least 5 minutes prior to testing.
- Do not stand while testing. Sit in a relaxed position while keeping your arm level with your heart.
- Avoid speaking or moving body parts while testing.
- While testing, avoid strong electromagnetic interference such as microwave ovens and cell phones.
- Wait 3 minutes or longer before re-testing.
- Try to measure your blood pressure at the same time each day for consistency.
- Test comparisons should only be made when monitor is used on the same arm, in the same position, and at the same time of day.
- This blood pressure monitor is not recommended for people with severe arrhythmia.
- Do not use this blood pressure monitor if the device is damaged.
Any blood pressure recording can be affected by the following factors
- The position of the subject, his or her physiologic condition;
- The performance and accuracy of the device;
- Cuff size: too small cuff (bladder) will produce a higher blood pressure value than usual, too big cuff (bladder) will produce a lower blood pressure value;
- Measuring position does not keep level with your heart;
- Speaking or moving body parts while testing;
- Not relaxing for about 5 minutes before taking the measurement.
Quick Start
- Before first use, please carefully check the packaging is complete, if not complete please contact the seller for replacement. Every time before use, please check the equipment for obvious damage, if there is equipment has been damaged, please use caution. In turn open the color box, asylum package (or receive bag), remove the blood pressure monitor, check whether the parts is complete, and install the batteries (as shown in figure A) battery installed, normal blood pressure monitor for power self-checking, if the battery power is insufficient, the screen shows “LO” error code, if the battery is normal, the device automatically after about 3 seconds time date set into the interface. After the setting, can enter the blood pressure measurement mode. (See Figure A )
- Insert cuff air plug into the behind side of monitor unit. (See Figure B)
- Remove thick clothing from the arm area.
- Rest for several minutes prior to testing. Sit down in a quietplace comfortably , back and arm support on a desk or table ,withyour legs uncrossed ,your arm resting on a firm and your feet flat on the floor. (See Figure C )
- Apply cuff to your left arm and middle of the cuff at the level of your heart. Bottom of cuff should be placed approximately 1-2cm (1/2”) above elbow joint. (See Figures D&E)
- Press ” START/STOP ” Button to start testing.
Unit Operation
Battery Installation
Slide battery cover off as indicated by arrow.
Install 4 new AAA alkaline batteries according to polarity. Close battery
cover.
System Settings
Time/Date setting
First to electricity, battery installed, normal blood pressure monitor for
power self-checking, if the battery power is insufficient, the screen shows
“LO” error code and flashing.
If battery power is normal, the device automatically after about 3 seconds
time date set into the interface. Not for the first time on electricity, under
the main interface , long press “on/off” button for 3 seconds can be time date
set into the interface. The first is for 24 h and 12 h mode Settings. Can
through the key and V mode conversion, press the key to determine the time
pattern, the system default for 24 h mode.
Mode after a certain time period, into the year set pattern, in the meantime, a flashing icon of the year is 0.5 seconds, press the button eight year of increase, press the button V decreased year, long press the key into the Λ or V can be adjusted quickly. Press the open key to determine the current Settings year, the system default for year 2022.
Determine the year after, enter in setting mode, in the meantime, icon will
display month scintillation a 0.5 seconds, press the keys into the increase in
plus, V decrease in press buttons,
long press buttons or keys in V can be adjusted quickly. Press the open key to
determine the current Settings in the system default for 1 month in September.
After the month of determining, into the day set patterns, in the meantime,
display, 0.5 seconds icon will flash once, press the keys into the plus the
number of days, press the button V reduce
the number of days, long press key into or V can quickly adjust the number of
days. Press the open key to determine the date of the current setting, the
system default date for no. 1.
After determine the date, time set into the interface, in the meantime, according to time of 0.5 seconds icon will flash smelting time, first set the hours, according to the first step in setting model to determine the time, if the first set is 12 h mode, the default time is 12: 00 am, if the first step is 24 h mode, set the default time for 0:00, more than two kinds of mode, press the button in the add hours, press the button V reduced hours, long press the key into the or V can quickly adjust hour, press the key to determine the current setting hours.
Determine the hours, minutes set into the interface, in the meantime,
according to minutes 0.5 seconds icon will flash once, press the button add
minutes, press the button V reduced minutes,
long press the key into the or V can quickly adjust minutes, press the key to
determine the current set of minutes.
Voice Setting
Time date set up is completed, enter the language setting interface, the
system default is L1 language, press the keys into the into the next L2
language, press the button V into L0, L0 closed for language.
Volume Setting
After the completion of the language Settings, enter the sound Settings
interface, the system default for V1 volume, press the button enter into the
next volume L2, press the button V into
V0, V0 for voice closed, with a total of four standard and increasing the
volume of successive sound.
Saved Settings
While in any setting mode, press ” START/STOP ” button to turn the unit
off. All information will be saved.
Note: If unit is left on and not in use for 1 minutes, it will automatically save all information and shut off.
User memory group Settings
This device has two memory group, each group has 120 memory value
respectively. In sleep mode, press the memory button, the device into the main
screen display, 2 seconds, the screen shows the current time, date and user
memory block. At the same time open the blood pressure monitor screen
backlight.By pressing the buttons on the blood pressure cuff body 1 and 2 keys
to select the corresponding memory block, press the button 1 on behalf of the
current group 1 sphygmomanometer measurement data is stored in the user
memory, memory speech group 1 at the same time; Press the button 2 represent
the current blood pressure measurement data is stored in theuser memory in
group 2, speech user memory group 2 at the same time. Memory group selection
is completed, presses the button, switch machine, the measurement of blood
pressure. When blood pressure within 30 seconds without operation, backlit
closed, no operation within 60 seconds, enter a dormant state.
Applying the Arm Cuff
- Firmly insert air plug into opening located on behind side of monitor unit.
- With sticky nylon section facing outward, insert end of cuff underneath metal ring of cuff .
- Fasten cuff about 1-2cm (0.4-0.8”) above the elbow joint. For best results apply cuff to bare arm and keep level with heart while testing
Testing
Power On
Under the main interface display, press the switch machine buttons, backlight
automatically open, 3 seconds after the show the most recent measurement
values, continuous press switch machine button can be skipped if the above two
delay directly into the measurement.
Testing
Blood pressure cuff pressure automatically, automatic stop, please keep
quiet when measuring.
If the cuff is not connected or cuff loose connection causes such as cuff
pressure rise is unusual, the blood pressure cuff at start-up pump inflatable
after a certain period of time, will stop the pump work, turn to wait for
pressure, the screen shown above Err3. Please test cuff pneumatic connection
state, can press “memory” button to enter to view memory or press the
“start/stop” button to turn it off.
Note: in the process of the pressure must not be arbitrary shaking body, otherwise, the system will also to determine this for high blood pressure, and pressure again;
Blood pressure cuff pressure in the process of blood pressure measurement can
be made. Armband internal pressure increases gradually, LCD real-time display
the current pressure.
When detected pulse, LCD screen flashing symbols, and display the current
pressure corresponding to the heartbeat of cuff pressure value
Note: Do not speak, in the process of measuring twist arm, shaking body, etc., please keep quiet state.
Result Display
The screen will display measurements for systolic and diastolic blood
pressure with voice broadcast. A indicator representing the current
measurement will appear next to the corresponding WHO Classification.
Note: Refer to Page 23~24 for detail WHO Blood Pressure Classification Information.
Irregular Heartbeat Indicator
If the monitor detects an irregular heart rhythm two or more times during the
measuring process, the Irregular Heartbeat Symbol ” “appears on screen along
with measurement results. Irregular heartbeat rhythm is defined as rhythm that
is either 25% slower or faster than the average rhythm detected while
measuring systolic blood pressure and diastolic blood pressure. Consult your
physician if the Irregular Heartbeat Symbol ” ” frequently appears with your
test results.
Power Off
The ” START/STOP ” button can be pressed to turn off the unit in any mode.
The unit can turn off the power itself about 1 minutes no operation in any
mode.
Safety Precaution: If pressure in arm cuff becomes too extreme while testing, press the ” START/STOP ” button to turn power off. The cuff pressure will rapidly dissipate once the unit is off.
HSD
HSD: A detection algorithm of human are strained, when measuring blood
pressure, it can be through the human bodyMeasure blood flow rate to determine
whether the body is in nervous condition, suitability for blood pressure
measurement. If the body is in nervous condition, sphygmomanometer can remind
and terminate the blood pressure measurement, the user should be measured
again after at least 5 minutes. If the normal leisure of the human body, blood
pressure will also remind and normal work and measured the blood pressure.
Cuff loose Indicator
At the beginning of the measurement, when the cuff is properly worn, the ”
” icon will be displayed, and when thecuff is loose, ” ” The icon will be
displayed. When this happens, please turn off the machine, adjust the cuff and
measure again. If the ” ” icon is still displayed, please check the cuff plug
connect with device correctly and measure again.
If it still cannot be resolved, please contact the seller or manufacturer.
Memory view and average view
In the main interface LCD screen all show self-checking, memory according
to [memory] key value view, see the first is the average of the last seven
days, and then the seven days before the measurement of mean and average in
the afternoon in the morning, the last is the latest measurement data, and
blood pressure by pressing enter key memory value view. According to the
average switching, press V return on a liquid crystal display interface. On
average, LCD shows average prompt “AVG”, and display the corresponding average
blood pressure classification information and classification identification on
the morning of the afternoon.
If you want to see other memory within the group, please user memory group set up in front of the boot, specific operation as shown in the “system Settings” in ” User memory group Settings “.
Memory Deletion
In the condition of main interface, press the [memory] key, enter the
average interface. Through user button to select the user, long press A button
and V 15 seconds for user delete memory value. Press A button to enter the
average interface in the morning, in the morning the average interface, can be
long press A button Λ and V 15 seconds to delete all data.
Note: Memory cannot be recovered once it has been deleted.
If you want to delete the other set of memories, please user memory group set up in front of the boot, specific operation as shown in the “system Settings” in ” User memory group Settings “;
Power Off
In any state, the boot can be fixed by press the “start/stop” button for 2
seconds to turn it off. If the device idle for 30 seconds, backlit closed, 60
seconds after the device has reached the “sleep” mode. Press any button
except/key can be reactivated backlit.In the condition of error and low
electricity, according to the key, open/equipment into “sleep” state
immediately.
Low Battery Indicator
When measuring screen does not appear when the battery symbol, said the
battery voltage is normal, sphygmomanometer can work normally.
When the screen appear when measuring marks, said the battery voltage is low,
please replace the battery, prevent blood pressure monitor can not work
normally.
Screen appears when measuring the flashing, and LCD screen will appear low voltage symbol “LO”, sphygmomanometer automatic shutdown after about 5 s please immediately replace the battery, ensure the sphygmomanometer can normal use.
Static Pressure Measurement
In the press key insert battery into the test mode at the same time,
according to an inferior version available algorithm and software version
number, alternating press Λ and V may continue to see test.
Note: Only Service personnel permitted to access to this mode, the mode unavailable in normal use.
Bluetooth requirements
The monitor requires a device with: . Bluetooth 4.0 or later
-
Android 5.0 or later
-
IOS 9.0 or later
And works with: -
iphone , iPod, iPad
-
Android Phones and Tablets
Wireless communication
Frequency range : 2.4 Ghz (2400-2483.5 MHz)
Modulation : GFSK
Antenna gain:0.5dBi
Bluetooth connection
Using for the first time
- Download the free “JoyHealth” App: On your mobile phone or table go to www.sejoy.com
- Open the App on your phone or tablet. If requested, you should enable Bluetooth on your device. You can enable Bluetooth under the Settings menu on your smart phone or table.
- Create a new user login, or login with your existing user name and password.
- Selection device “Blood pressure monitor”.
Pairing your monitor with a Smart Device
- Long press the ” ” button on your monitor until the word “bE” flashes on your monitor .
- Open the APP and click “pair device” button, a popup window will appear. Click “Search device” button to display all sphygmomanometer devices(The display information includes the device model and MAC address), then you can find the sphygmomanometer through the “device Name”, click the pictureA film or text can be used to pair the device.
- After the pairing is complete, the device interface switches from boo interface, indicating that the pairing is successful.
Transfer your readings
-
As soon as your measurement is complete, open the app on your smart phone to transfer your reedings.
Note: On the paired smartphone, Bluetooth must be enabled. -
You can view your blood pressure readings on the app.
Troubleshooting
Abnormal phenomenon | Cause analysis | Processing method |
---|---|---|
Abnormal sphygmomanometer | The armband is tied too tight or too loose, Or the | |
arm strap is tied incorrectly; | Roll the armband correctly | |
Move the arm during measurement or Electronic sphygmomanometer | Stay quiet, |
keep your arm steady, and do not move the monitor
Speaking, nervous or emotional during measurement| Instead of talking, take
deep breaths to calm your mood and relax your body
Incorrect measurement posture| Adjust posture, see
“Blood pressure gauge Wearing”
There is interference in charging process or improper operation in measuring process| See operation Instructions.
The following table shows the error signs that may occur during measurement, possible causes and handling methods. Please measure again using the correct method
Error display | The cause of the problem | The solution |
---|---|---|
Er1 | Can’t detect high and low |
pressure
| Please fasten the cuff before
measuring
Er2|
Cuff too loose or loose
| Please fasten the cuff before measuring
Er3| The pressure exceeds 15mmHg for 3 minutes| Check whether the cuff is
knotted or the vent valve is blocked. If the problem persists, contact the
manufacturer
Er4| Improper compression caused by arm or body movement| Hold the arm or body
still and measure again
Er5| The pressure exceeds 300mmHg| Please fasten the cuff before measuring
LO| Battery dead| Replace the battery or connect the power adapter (if any).
Note : If you cannot solve the abnormal situation by yourself, you can consult the manufacturer or the manufacturer’s designated unit by phone. It is forbidden to disassemble and repair without permission. If necessary, professional maintenance personnel can ask the manufacturer for the list of components and circuit schematic diagram.
Blood Pressure Information
Blood Pressure
Blood pressure is the force of blood pushing against the walls of arteries. It
is typically measured in millimeters of mercury (mmHg.) Systolic blood
pressure is the maximum force exerted against blood vessel walls each time the
heart beats. Diastolic blood pressure is the force exerted on blood vessels
when the heart is resting between beats.
An individual’s blood pressure frequently changes throughout the course of a
day. Excitement and tension can cause blood pressure to rise, while drinking
alcohol and bathing can lower blood pressure. Certain hormones like adrenaline
(which your body releases under stress) can cause blood vessels to constrict,
leading to a rise in blood pressure.
If these measuring numbers become too high, it means the heart is working
harder than it should.
WHO Blood Pressure Classification Indicator
The BM 92 is equipped with a classification indicator based on established
guidelines from the World Health Organization. The chart below (color coded on
monitor unit) indicates test results.
Health Reminder
Hypertension is a dangerous disease that can affect the quality of life. It
can lead to a lot of problems including heart failure, kidney failure, and
cerebral hemorrhaging. By maintaining a healthy lifestyle and visiting your
physician on a regular basis, hypertension and relative diseases are much
easier to control when diagnosed in their early stages.
Note: Do not be alarmed if an abnormal reading occurs. A better indication of an individual’s blood pressure occurs after 2-3 readings are taken at the same time each day over an extended period of time. Consult your physician if test results remain abnormal.
Q: What is the difference between measuring blood pressure at home or at a
professional healthcare clinic?
A: Blood pressure readings taken at home are now seen to give a more accurate
account as they better reflect your daily life. Readings can be elevated when
taken in a clinical or medical environment. This is known as White Coat
Hypertension and may be caused by feeling anxious or nervous.
Note: Abnormal test results may be caused by:
-
Improper cuff placement
Make sure cuff is snug-not too tight or too loose.
Make sure bottom of the cuff is approximately 1-2cm (1/2”) above the elbow joint. -
Improper body position
Make sure to keep your body in an upright position. -
Feeling anxious or nervous
Take 2-3 deep breaths, wait a few minutes and resume testing.
Q: What causes different readings?
A: Blood pressure varies throughout the course of a day. Many factors
including diet, stress, cuff placement, etc. may affect an individual’s blood
pressure.
Q: Should I apply the cuff to the left or right arm? What is the
difference?
A: Either arm can be used when testing, however, when comparing results, the
same arm should be used. Testing on your left arm may provide more accurate
results as it is located closer to your heart.
Q: What is the best time of day for testing?
A: Morning time or any time you feel relaxed and stress free.
Maintenance
-
Avoid dropping, slamming, or throwing the unit.
-
Avoid extreme temperatures. Do not expose unit directly under sunshine.
-
When cleaning the unit, use a soft fabric and lightly wipe with mild detergent.
Use a damp cloth to remove dirt and excess detergent. If the previous cleaning instructions are not successful, repeat several times until the visible dirt is removed. -
Cuff Cleaning and Disinfection:
- Spread the cuff (skin-contact surface) upwards onto a clean table. Use a damp clean cloth (water-based) to wipe the skin-contact surface with a force.
- Soak the cloth clean with drinking water and wring it dry. Repeat A) with the damp cloth (water-based) for 3 times.
- Apply 70%-80% alcohol to a new cloth (or 75% alcohol cotton-ball), use it to wipe the skin-contact surface with a force. Then soak the cloth with the alcohol again (or change a new 75% alcohol cotton-ball), repeat the disinfection procedure for 3 times.
- When the disinfection towards the skin-contactsurface is finished,wipe thenon-skin contact surface with a cloth (alcohol-based) or alcohol cotton-ball thoroughly for 3 times.
- Leave the cuff naturally dry,then it is ready for reuse.
Notice: Do not soak in water or splash water on it.
-
Do not use petrol, thinners or similar solvents.
-
Remove batteries when not in operation for an extended period of time.
-
Do not disassemble product.
-
It is recommended the performance should be checked every 2 years.
-
Expected service life: Approximately three years at 10 tests per day.
-
No service and maintenance while it is in use and maintenance only be performed by service personnel. Service and maintenance require parts, repair, technical support will be provided.
Specifications
Product Description | Arm-type Fully Automatic Blood Pressure Monitor |
---|---|
Model | BM 92 |
Display | LCD Digital Display Size:71mm x 76mm(2.80″ x 2.99″) |
Measurement Method | Oscillometric Method |
Measurement Range
| Systolic Pressure| 60mmHg~260mmHg
Diastolic Pressure| 40mmHg~200mmHg
Pressure| 0mmHg~299mmHg
Pressure| ±3mmHg
Pulse| 30 ~ 180 Beats/Minute
Pulse| ±5%
Pressurization| Automatic Pressurization
Memory| 2×120 Memories in Two Groups with Date and Time
Function
| Irregular Heartbeat Detection
WHO Classification Indicator
Last 7 days average
Average for the last 7 days
Afternoon average for the last 7 days
Low Battery Detection
Automatic Power-Off
Voice
Backlighting
Bluetooth
HSD
Power Source
|
4 AAA batteries
Battery Life| Approximately 2 months at 3 tests per day
Unit Weight| Approx.180g (6.35 oz.) (excluding battery)
Unit Dimensions| Approx.125mm X 95mm X 44mm(4.92″ x 3.74″ x 1.73″)
Cuff Circumference| Medium cuff: Fits arm circumference 22-36 cm
Operating Environment
| Temperature| 10℃ ~ 40℃ (50 ℉~104 ℉)
Humidity| 15% ~ 93% RH
Operating Environment| Pressure| 800hPa~1060hPa
---|---|---
Storage Environment
| Temperature:| -25 ℃~55 ℃ (-13 ℉~131 ℉)
Humidity| ≤93% RH
Ingress Protection rating:| IP 21,Indoor Used Only
Battery Shelf life:| 60 months
Battery Storage Temperature:| -25℃~55℃ (-13℉~131℉)
Specifications are subject to change without notice.
Safety Standard(included but not limited):
- IEC 80601-2-30, medical electrical equipment – part 2-30: particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers. (Cardiovascular)
- ISO 81060-2, non-invasive sphygmomanometers – part 2: clinical validation of automated measurement type. (Cardiovascular)
- AAMI / ANSI ES 60601-1:2005/(R)2012 and C1:2009/(R)2012 and, a2:2010/(r)2012 (consolidated text) medical electrical equipment — part 1: general requirements for basic safety and essential performance
- AAMI/ANSI/IEC 60601-1-2, Medical Electrical Equipment — Part 1-2: General Requirements For Basic Safety And Essential Performance — Collateral Standard: Electromagnetic Disturbances — Requirements And Tests (General II (ES/EMC))
- IEC 60601-1-11, medical electrical equipment – part 1-11: general requirements for basic safety and essential performance – collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product indicates that it should not be disposed
with other household waste at the end of its life. To prevent potential harm
to the environment or to human health, please separate this product from other
types of wastes and recycle it responsibly. When disposing this type of
product, contact the retailer where product was purchased or contact your
local government office for details regarding how this item can be disposed in
an environmentally safe recycling center.
Business users should contact their supplier and check the terms and
conditions of the purchasing agreement. This product should not be mixed with
other commercial wastes for disposal. This product is free of hazardous
materials.
Warranty
The Blood Pressure Monitor is guaranteed for 2-year from the date of purchase. If the Blood Pressure Monitor does not function properly due to defective components or poor workmanship, we will repair or replace it freely. The warranty does not cover damages to your Blood Pressure Monitor due to improper handling. Please contact local retailer for details.
Electromagnetic Compatibility Information
The device satisfies the EMC requirements of the international standard IEC 60601-1-2. The requirements are satisfied under the conditions described in the table below. The device is an electrical medical product and is subject to special precautionary measures with regard to EMC which must be published in the instructions for use. Portable and mobile HF communications equipment can affect the device. Use of the unit in conjunction with non-approved accessories can affect the device negatively and alter the electromagnetic compatibility. The device should not be used directly adjacent to or between other electrical equipment.
Table 1
Guidance and declaration of manufacturer-electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
Emissions test| Compliance| Electromagnetic environment
guidance
Radiated emission CISPR 11| Group 1, class B.| The device uses RF energy only
for its internal function. Therefore, its emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
Conducted emission CISPR 11| Group 1, class B.| The device is suitable for use
in all establishments, including domestic establishments and those directly
connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2| Class A
Voltage fluctuations/ flicker emissions IEC 61000-3-3| Complies
Table 2
Guidance and declaration of manufacturer-electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment- ent.
IMMUNITY test| IEC 60601
test level
| Compliance level| Electromagnetic environment
-guidance
Electrostatic discharge (ESD) IEC 61000-4-2| ± 8 kV
contact
±2 kV,±4 kV,
±8 kV,
±15 kV air
| ± 8 kV
contact
±2 kV,±4 kV,
±8 kV,
±15 kV air
| Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Electrostatic transient/burst IEC 61000-4-4
| ****
± 2 kV ,
100kHz, for AC power port
| ****
± 2 kV ,
100kHz, for AC power port
| ****
Mains power quality should be that of a typical commercial or hospital environment.
Surge
IEC 61000-4-5
| ****
±0.5kV, ±1kV
(differential mode)
| ****
±0.5kV, ±1kV
(differential mode)
| ****
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interrupti- ons and voltage variations on p- ower supply in- put lines| 0 % UT;
0,5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°
, 270° and 315°
0 % UT; 1 cycle and
70 % UT;
25/30 cycles Single phase: at 0°
| 0 % UT;
0,5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°
, 270° and 315°
0 % UT; 1 cycle and
70 % UT;
25/30 cycles Single phase: at 0°
| Mains power quality should be that of a typical commercial or hospital
environment.
IEC 61000-4-11| 0 % UT;
250/300 cycle
| 0 % UT;
250/300 cycle
|
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
| ****
30 A/m; 50Hz or 60Hz
| ****
30 A/m; 50Hz or 60Hz
| Power frequency magnetic fields should be at levels charactertic of a
typical location in a typical comme- rcial or hospital environment.
Guidance and declaration of manufacturer-electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
IMMUNITY test| IEC 60601 test level| Compliance level| Electromagnetic environment
-guidance
Radiated RF EM fields
IEC 61000-4-3
| 10 V/m80MHz-
2.7Ghz 80%AM
at 1kHz
| 10 V/m80MHz-
2.7Ghz 80%AM
at 1kHz
| Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended seperation distance 80 MHz to 800 MHz 800 MHz to 2.7 Ghz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:
Conducted disturbances Induced by RF fields
IEC 61000-4-6
| 3 V in 0.15
MHz- 80 MHz 6 V in ISM
and/or amateur radio bands between 0.15
MHz and 80
MHz 80 %
AM at 1kHz
| 3 V in 0.15
MHz- 80 MHz 6 V in ISM
and/or amateur radio bands between 0.15
MHz and 80
MHz 80 %
AM at 1kHz
| Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended seperation distance 80 MHz to 800 MHz 800 MHz to 2.7 Ghz where P is the maximum output power rating of the transmitter in watts (W) according to the transm- itter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:
Table 3
Guidance and declaration of manufacturer-electromagnetic immunity|
---|---
Nowadays, many RF wireless equipments have being used in various healthcare
locations where medical equipment and/or systems are used. When they are used
in close proximity to medical equipment and/or systems, the medical equipment
and/or systems’ basic safety and essential performance may be affected. Arm-
type Fully Automatic Digital Blood Pressure Monitorhas been tested with the
immunity test level in the below table and meet the related requirements of
IEC 60601-1-2:2014. The customer and/or user should help keep a minimum
distance between RF wireless communications equipment and this medical
equipment and/or systems as recommended below.|
Test frequency
(MHz)
| Band (MHz)| ****
Service
| ****
Modulation
| Maximum power (W)| Distance (m)| Immunity test level (V/m)|
385| 380-390| TETRA 400| Pulse
modulation 18Hz
| 1.8| 0.3| 27|
450| 430-470| GMRS 460
FRS 460
| FM
± ٥ kHz deviation 1 kHz sine
| 2| 0.3| 28|
710| ****
704-787
| ****
LTE
Band 13, 17
| ****
Pulse modulation 217Hz
| ****
0.2
| ****
0.3
| ****
9
|
745|
780|
810| | GSM 800/900,| ****
Pulse modulation
18Hz
| | | |
800-960| TETRA 800,
iDEN 820,
CDMA 850,
| 2| 0.3| 28|
870|
930|
| LTE Band 5| | | |
1720| | ****
GSM 1800;
| ****
Pulse modulation 217Hz
| | | |
| CDMA 1900;| | | |
1845|
1700-1990| GSM 1900;
DECT;
| 2| 0.3| 28|
| LTE Band 1, 3,| | | |
|
| | 4, 25; UMTS| | | |
1970| | | | | |
2450| 2400-2570| Bluetooth,WLAN,
802.11 b/g/n,RFID 2450,LTE Band 7
| Pulse modulation 217Hz| 2| 0.3| 28|
5240| 5100-5800| WLAN 802.11
a/n
| Pulse modulation
217Hz
| 0.2| 0.3| 9|
5500|
5785|
Table 4
Recommended separation distances between portable and mobile RF communications equipment and the device
The device is intended for use in an electromagnetic environment in which
radiated therefore disturbances are controlled. The customer or the user of
the device can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment
(transmitters) and the device as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum output power of transmitter
W
| Separation distance according to frequency of transmitter
m
80 MHz to 800 MHz| 800 MHz to 2.7 GHz
0.01| 0.12| 0.23
0.1| 0.38| 0.73
1| 1.2| 2.3
10| 3.8| 7.3
100| 12| 23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Additional Notes
Important Instructions Before Use
-
WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
-
WARNING: PORTABLE RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of Arm-type Fully Automatic Digital Blood Pressure Monitor, including cables specified by the MANUFACTURER. Otherwise, degradation of the performance of this equipment could result.
-
The software identifier refer to the software verification and validation report, and the file code is JYRJ201009001.
-
verify manometer pressure accuracy:
In the power down state, press and hold the ” START/STOP” button, and then install
the batteries. Until the LCD screen is full, release the ” START/STOP” button.
When the LCD screen displays the double zero, the blood pressure meter is in static state.
At this point, 500ml gas capacity, calibrated standard pressure gauge and manual pressure device can be connected to the sphygmomanometer through the sleeve interface of the sphygmomanometer, and manual pressure can be applied to the effective display range of the sphygmomanometer, and then the difference between the reading of the sphygmomanometer and that of the standard pressure gauge can be compared.This mode can be used to verify manometer pressure accuracy. -
Contraindications:
The product is not intended for infants or individuals who cannot express their intentions. -
Intended Use
The digital blood pressure monitor are reusable for clinical and home use and are non-invasive blood pressure measurement systems designed to measure the systolic and diastolic blood pressure and pulse rate of adolescents and adults individual by using a non-invasive technique, which is a well-known technique in the market called the “oscillometric method”.
it can measure systolic blood pressure, diastolic blood pressure and pulse rated on
up-arm,and the device is reusable for clinical or home use. -
The patient is the operator: the PATIENT is an intended OPERATOR. the PATIENT Do not carry out other maintenance operations except to replace the battery.
-
WARNING:
Do not modify this equipment without the authorization of the manufacturer. -
ESSENTIAL PERFORMANCE Maintenance advice:
Pressure calibration will be carried out when this product leaves the factory. Patients can use the method described in the section “Verify Manometer Pressure Accuracy” to verify the accuracy. If the accuracy deviation is large, please contact the manufacturer to recalibration. -
Mechanical strength and resistance to heatThe resistance to heat will be retained by device during the EXPECTED SERVICE LIFE of the ME EQUIPMENT.
-
Do not place the blood pressure monitor and cuff at will. It will cause asphyxiation if the child swallows or twine around his neck.
-
The cuff and the case of the blood pressure monitor have been tested for biocompatibility and do not contain allergenic or harmful materials.Please stop using it if allergy occurs during use.
-
Warning:
Non-professionals do not modify the equipment, otherwise, it will make the equipment measurement is not accurate. -
Warning:
Do not expose the equipment for a long time, otherwise, it will reduce its performance of the equipment. -
Warning:
This device is not used for children and pets -
Clean:
The equipment can be cleaned by lay operator according to the cleaning procedures in the instructions -
Warning:
Do not use a damaged cuff for blood pressure measurement. -
Warning:
When measuring with the cuff, if the tester feels seriously uncomfortable, press the button of the blood pressure monitor to deflate the cuff, or remove the cuff directly from the arm. -
Warning:
If an unexpected reading occurs, the operator can take several more measurements and consult a doctor. -
Warning:
This equipment is used outside the specified environment, may damage the equipment, and may be an inaccurate measurement. -
ME equipment not intended for use in conjunction with flammable agents “ME equipment not intended for use in the oxygen-rich environment”
-
Materials, accessories and detachable parts other than those specified in the manual may pose unacceptable risks.
-
Do not modify. Do not connect to other devices.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product indicates that it should not be disposed
with other household waste at the end of its life. To prevent potential harm
to the environment or to human health, please separate this product from other
types of wastes and recycle it responsibly. When disposing this type of
product, contact the retailer where product was purchased or contact your
local government office for details regarding how this item can be disposed in
an environmentally safe recycling center.
Business users should contact their supplier and check the terms and
conditions of the purchasing agreement. This product should not be mixed with
other commercial wastes
for disposal. This product is free of hazardous materials.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>