JOYTECH HEALTHCARE DET-306b Infrared Forehead Thermometer
Introduction
The DET-306b infrared forehead thermometer is specifically designed for safe use on the forehead. The Infrared Forehead Thermometer is a device capable of measuring people’s body temperature by detecting the intensity of infrared light emitted from the forehead. It converts the measured heat into a temperature reading displayed on the LCD. The infrared forehead thermometer is intended for the intermittent measurement of human body temperature from the skin surface of forehead by people of all ages. When properly used, it will quickly assess your temperature in an accurate manner.
Please read all instructions carefully and thoroughly before using this product.
Warning:
- There is no gender or age limitation for using infrared forehead
- Do not touch the temperature probe with hands.
- Use of this Forehead thermometer is not intended as a substitute for consultation with your physician.
- Do not allow children to take their temperatures unsupervised, some parts are small enough to be
- Never immerse this device in water or other liquids(not waterproof).
- Do not modify this equipment without the authorization of the manufacturer.
- Do not expose the thermometer to temperature extremes (below -25℃/-13℉ or over 55℃/131℉) nor excessive humidity (>95%RH).
- Keep the battery away from children.
- Remove the battery from the device when not in operation for a long time.
Product Description
LCD Display Introduction
Basic Functions
Real Time Clock
| The real time clock will be recorded with the memory function and help you to recognize each measurement result.
→ Please see the Real time clock setting section to learn how to
setup the time in the first use.
---|---
Forehead Mode
| The thermometer has been designed for practical use. It’s not meant to replace a visit to the doctor. Please also remember to compare the measurement result to your regular body temperature.
→ Please see the Illustration For Use section to learn how to
measure the body temperature.
Object Mode
| The object mode shows the actual, unadjusted surface temperatures, which is different from the body temperature. It can help you to monitor if the object temperature is suitable for the baby or patient, for example the baby’s milk.
→ Please see the Illustration For Use section to learn how to
measure the object temperature.
Beep Alarm
| If thermometer detects a body temperature ≥ 37.8℃(100.0℉) under forehead mode, there will be a long beep sound followed by
three short beep sound to warn the user.
Memory Mode| There are each 10 sets memories for forehead and object measurements. Each memory also records the measurement
date/time/mode icon.
℃/ ℉ Switch| Please see the Selecting A Temperature Scale section to learn
how to change between Celsius and Fahrenheit.
Distance Detection| If the object is too far away from the probe, the temperature will not
be measured until the probe is moved within the measured range.
Forehead Thermometer Advantages
Infrared Forehead Thermometer measures core body temperature, which is the temperature of a body’s vital organs. (See Figure 1) This thermometer is designed to measure the temperature of the skin surface over the temporal artery, a major artery of the head. The temporal artery is connected to the heart via the carotid artery, directly leading from the aorta, the main trunk of the arterial system. It offers constant blood flow. Therefore, body temperature changes are reflected sooner in the forehead than they are in other parts of the body such as oral, rectal and underarm.
Real-Time Clock Setting
When using a thermometer for the first time, please set the parameters of the thermometer. With the thermometer off, press and hold BUTTON to enter into setting mode for 3 seconds.
Set the time format
The device can display the time in either an AM/PM (12-hour) or a
24:00(24-hour) format. Press and release START BUTTON to select the format.
With the preferred time format on the display, press
BUTTON, the Hour figure is flashing automatically.
Set the hour
Press and releas e the START BUTTON to advance one hour until the correct hour appears. After the hour is set, press BUTTON, the Minute figure is flashing automatically.
Set the minute
Press and release the START BUTTON to advance one minute until the correct minute appears. After the minute is set, press BUTTON, the Year figure is flashing automatically.
Set the year
Press and release the START BUTTON to advance one year until the correct year appears. After the year is set, press BUTTON, the Month figure will appear.
Set the month
Press and release the START BUTTON to advance one month until the correct month appears. After the mont.h is set, press BUTTON, the Date figure is flashing automatically.
Set the date
Press and release the START BUTTON to advance one day until the correct month appears. After the day is set, press BUTTON to exit the setting mode.
Temperature Taking Hints
To ensure that the reading always reflects the body temperature accurately, you need to take account of the following factors which may affect an accurate reading.
- It is important to know each individual’s normal temperature when they are well. This is the only way to accurately diagnose a fever. To determine normal temperature, take multiple readings when healthy. Re-measure with a standard digital thermometer for confirmation.
- Users must be inside for 30 minutes before taking a measurement. Note: Users and the thermometer should be in the same ambient temperature for at least 10 minutes before taking a reading.
- Users should not drink, eat, or be physically active such as bathing, showering, shampooing and hair drying before/while taking the measurement. Remove hat and hair and wait 10 minutes before taking a reading.
- . Oils or cosmetics on the forehead may give a lower temperature reading than the actual one. Remove dirt from the forehead before taking a measurement. Wait at least 10 minutes after washing the forehead area before taking a reading.
- Holding a hand on the forehead for any length of time will affect the temperature reading.
- Do not take temperature over scar tissue, open sores or abrasions.
- Do not use the thermometer on a perspiring or sweating forehead, as this may affect the reading.
- Don’t take a measurement while or immediately after nursing a baby.
- Do not take temperatures with this thermometer near places that are very hot, such as fireplaces and stoves.
- The probe window of the thermometer is the most delicate part of the device. Do not touch the probe window. The accuracy of the reading may be affected if the probe window is damaged or dirty.
- If the thermometer is stored in a significantly different environment than testing location, place it in the testing location for approximately 30 minutes prior to use.
- It is not intended for use in the oxygen rich environment and presence of flammable anesthetic mixture with air, oxygen or nitrous oxide.
Illustration For Use
To measure forehead temperature:
- Press the BUTTON, The display is activated to show all segments. Ater self-checking Figure 2 appears on the display screen with beeps, so you can start a new measurement.
- Aim the thermometer at the center of the forehead with a distance less than 5cm (See figure 3) and then press the START BUTTON. Note1: Do not remove the thermometer from the forehead before hearing the final beep. A waiting period of 1 seconds between testing is recommended to ensure complete accuracy. Note2:If the object to be tested is too far from the probe, a prompt will appear(See figure 4).
- Read the temperature on the display.
- Press the BUTTON to turn off.
How to change the forehead mode and object mode:
You can press / BUTTON to switch the mode between Forehead mode and Object
mode.
To measure object temperature:
- Press the BUTTON to turn on the thermometer, you can take the object temperature after hearing two beep sounds.(see figure 5)
- Aim the thermometer at the center of the object you want to measure with a distance less than 5cm.
- Press the START BUTTON and then read the temperature on the display.
- Press the BUTTON to turn off.
After measurement:
- Power off: Device will automatically shut off if left idle for more than 1 minute to extend battery life.
- Clean the probe after each use to ensure an accurate reading and avoid cross contamination. (See the section of Care and Cleaning for details.)
BackLight :
In Forehead mode:
- The display will be lighted GREEN for 3 seconds with a happy face when the unit is ready for measurement and measurement is completed with reading less than 37.3℃(99.1℉).
- The display will be lighted YELLOW for 3 seconds with a happy face when a measurement is completed with reading less than 37.8℃(100.0℉).
- The display will be lighted RED for 3 seconds with a bad face when a measurement is completed with a reading equal to or higher than 37.8℃(100.0℉).
In Object mode:
The display will only be lighted GREEN for 3 seconds when the unit is ready for measurement and measurement is completed.
Bluetooth requirements
The thermometer requires a bluetooth device with:
- Bluetooth 4.0 or later.
- Android 6.0 or later.
- IOS 8.0 or later
And works with:
- iphone , iPod, iPad
- Android Phones and Tablets
Using for the first time
- Download the “JoyHeath” App from Website or APP Store (Such as Apple Store).
- Open the App on your phone or tablet. If requested, you should enable Bluetooth on your device. You can enable Bluetooth under the Settings menu on your smart phone or tablet.
- Create a new user login, or login with your existing user name and password.
- Selection device “Thermometer”.
Match your thermometer with a Smart Device
1. If this is your first time using it, bind first.Open ” SETTING”
menu,choose“Bind and unbind device” and select the
appropriate model.
The date and time on your thermometer will automatically be
updated when it’s connected with your phone.
2. Confirm that your thermometer is connected successfully.
When your thermometer is connected successfully to your smart
phone, the “ ” symbol stop flashing and keep showing.
Transfer your readings
1. As soon as your measurement is finished, open the app on
your smart phone to transfer the reedings.
Note: On the matched smart phone, Bluetooth must be enabled.
2. You can view your temperature readings in the app.
Memory Mode
- The Memory Mode can be accessed ether in forehead mode or object mode:
When the thermometer has been turned on and followed by Figure 2/4 or finished testing, press and hold the BUTTON for three seconds. The letter M will appear in the center-right corner of the display. (See Figure 6)
- The thermometer will automatically memorize the last 30 temperature readings. Each memory also records the measurement date/time/mode Each time the BUTTON is pressed, the screen displays past readings that correspond with a number 1-30. The number 1 reflects the most recent reading, while the number 30 reveals the oldest reading stored in memory.(See Figure 7)
- In the memory mode, mark or mark will not The user can press the START BUTTON to take new measurements.
Selecting A Temperature Scale
- Temperature readings are available in the Celsius (℃) or Fahrenheit (℉) scale.
- With the thermometer off, press and hold the START BUTTON for 3 seconds to enter into unit changing mode.
- Press and release START BUTTON to select the unit.
- When the preferred unit on the display, press BUTTON to exit the unit changing mode.
Care And Cleaning
- The probe window must be kept clean, dry, and undamaged at all times to ensure accurate readings. The accuracy of temperature readings can be affected by damage to the probe window, or the presence of dirt, fingerprints, earwax, dust, and other soiling compounds on the probe window. Degraded sensors can degrade performance or cause other problems.
- The probe window is the most delicate part of the thermometer. Use a soft cloth slightly moistened with a 75% isopropyl alcohol solution to disinfect probe window and the thermometer. Do not use abrasive cleaners. After cleaning, allow at least 10 minutes of drying time before taking temperatures. Note : Do not use any chemical other than isopropyl alcohol to clean the probe window.
- Use a soft, dry cloth to clean the thermometer display and exterior.
- Do not put the thermometer into the water directly.
- Store the thermometer in a dry location, free from dust and contamination and away from direct sunlight.
- Strong electromagnetic fields may interfere with the proper operation of the thermometer. The device needs special precautions regarding EMC according to the EMC information.
- Put the thermometer back to the original packaging after use.
Battery Replacement
- Replace battery when appears in the upper right corner of LCD display. (See Figure 9)
- Slide battery cover down as shown in Figure 10.
- Remove battery and install 2 new AAA alkaline batteries as shown in Figure 11 .
- Slide battery cover back on.
Specifications
Measuring range| Forehead mode: 34.0℃~43.0℃(93.2℉~109.4℉) Object mode:
0℃~100℃(32℉~212℉)
---|---
Measuring site| Forehead(Forehead Mode)
Reference body site| Oral (This thermometer converts the forehead temperature
to display its “oral equivalent.”)
Operation mode| Forehead mode(Adjust mode)
Laboratory accuracy
| Forehead mode:
±0.2℃ (0.4℉) during 35.5℃~42.0℃ (95.9℉~107.6℉)
at 15℃~35℃ (59.0℉~95.0℉) operating temperature range
±0.3℃ (0.5℉) for other measuring and operating temperature range
Object mode: ±4% or ±2℃(4℉) whichever is greater
Display resolution| 0.1℃ or 0.1℉
Measure time| Approximately 1 second
Operating temperature range:| 10℃~40℃(50℉~104℉), 15%~85%RH, non-condensing
Atmospheric Pressure : 700hPa ~ 1060hPa
Storage and transport temperature range| -25℃~ 55℃ (-13℉~131℉), 15%~95%RH,
non-condensing Atmospheric Pressure : 700hPa ~ 1060hPa
Clinical accuracy
| Clinical bias: -0.09℃(- 0.16℉) Clinical repeatability: 0.13℃( 0.23℉) Limits
of agreement: 0.87℃( 1.57℉)
Shock| withstands drop of 3 feet
Dimension| 1534144mm
Weight| Approx.84grams(with batteries)
Battery| DC3V(2×AAA battery)
Battery life| Approx. 1 year/6000 readings
Expected service life| Three years
Ingress protecting rating| IP22
Troubleshooting
Error message | Problem | Solution |
---|
|
The thermometer is not functioning properly.
| Unload the battery, wait for 1 minute and repower it. If the message reappears, contact the retailer for service.
| The ambient temperature is not within the range between 10℃ and 40℃
(50℉~104℉).| Place the thermometer in a room for at least 30 minutes at room
temperature between 10℃ and 40℃ (50℉~104℉)
Error message| Problem| Solution
---|---|---
| In Forehead mode: Temperature taken is higher than 43.0 ℃ (109.4℉).
In Object mode: Temperature taken is higher than 100 ℃
(212℉).
|
Read Temperature Taking Hints Thoroughly, then take a new temperature measurement.
| In Forehead mode: Temperature taken is lower than 34.0 ℃ (93.2℉).
In Object mode: Temperature taken is
lower than 0℃ (32℉).
|
Read Temperature Taking Hints thoroughly , then make sure the lens filter are clean, then take a new temperature measurement.
|
The thermometer could not work due to low battery.
|
Replace two new alkaline batteries size AAA.
Calibration
The thermometer is initially calibrated at the time of manufacture. If the thermometer is used according to the use instruction, periodic readjustment is not required. However, We recommends checking calibration every two years or whenever the clinical accuracy of the thermometer is in question. Please send the complete device to the dealers or manufacturer. The above recommendations do not supersede the legal requirements. The user must always comply with legal requirements for the control of the measurement, functionality, and accuracy of the device which are required by the scope of relevant laws, directives or ordinances where the device is used. A clinical summary and procedures for checking calibration are available upon request.(Turn on the thermometer and press the power button a long time until entering into calibrate mode, the software version will be displayed.)
- Type B F applied part
- Direct Current
- Disposal of this product and used batteries should be carried out in accordance with the national regulations for the disposal of electronic products
- Consult Accompanying Documents
- Batch Code
- Storage and Transportation Temperature Limit: -13℉~131℉(-25℃~55℃)
Service
The thermometer has a limited one year warranty. Do not attempt to disassemble or repair the thermometer by yourself. Should service be required during or after the warranty period you must contact the manufacturer. Repackage the thermometer carefully in its original packaging or securely pack to avoid damage during shipping. Include the original sales slip indicating the date of purchase, a note describing the problem, and your return address. Send the thermometer prepaid and insured. The lay operator or lay responsible organization should contact the manufacturer or the manufacturer’s representative:
- For assistance, if needed, in setting up, using or maintaining the thermometer; or
- To report unexpected operation or events.
Manufacturer:
JOYTECH HEALTHCARE CO. LTD.
Area C, Building 2, No.365, Wuzhou Road, Yuhang Economic Development Zone,
Hangzhou City, 311100 Zhejiang, China
Telephone: +86-571-81957767
Fax: +86-571-81957750
The product is in compliance with the requirements of MDD 93/42/EEC, “0197” is the identification number of notify body.
European Authorized Representative:
Shanghai International Holding Corp. GmbH (Europe) Eiffestrasse 80, 20537
Hamburg, Germany.
Warranty
This appliance conforms to the following standards: ISO 80601-2-56 Medical
electrical equipment —Part 2-56: Particular requirements for basic safety and
essential performance of clinical thermometers for body temperature
measurement, IEC 60601-1-11 Medical electrical equipment —Part 1-11: General
requirements for basic safety and essential performance –Collateral Standard:
Requirements for medical electrical equipment and medical electrical systems
used in the home healthcare environment and complies with the requirements of
IEC 60601-1-2(EMC) ,
IEC 60601-1(Safety) standards. And the manufacturer is ISO 13485 certified.
The thermometer is warranted by manufacturer to be free from defects in
material and workmanship under normal use and service for a period of one year
from the date of delivery to the first user who purchases the instrument. This
warranty does not cover batteries, damage to the probe window, or damage to
the instrument caused by misuse, negligence or accident, and extends to only
to the first purchaser of the product.
FCC Information
Caution: Changes or modifications to this unit not expressly approved by the
party responsible for compliance could void the user authority to operate the
equipment.
Note:
This equipment has been tested and found to comply with the limits for a Class
B digital device, pursuant to Part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses, and can radiate
radio frequency energy. If this equipment does cause harmful interference to
radio or television reception, which can be determined by turning the
equipment off and on, the user is encouraged to try and correct the
interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the distance between the equipment and the receiver.
- Connect the equipment to an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
- This device may not cause harmful interference, and
- This device must accept any interference received, including interference that may cause undesired operation.
The device has been evaluated to meet general RF exposure requirements. The device can be used in portable exposure conditions without restriction.
Electromagnetic Compatibility Information
The device satisfies the EMC requirements of the international standard IEC 60601-1-2.The requirements are satisfied under the conditions described in the table below.The device is an electrical medical product and is subject to special precautionary measures with regard to EMC which must be published in the instructions for use.Portable and mobile HF communications equipment can affect the device.Use of the unit in conjunction with non-approved accessories can affect the device negatively and alter the electromagnetic compatibility.The device should not be used directly adjacent to or between other electrical equipment.
TABLE 1
TABLE 2
TABLE 3
TABLE 4
TABLE 5
WARNINGS!
- This device should not be used in the vicinity or on top of other electronic equipment such as cell phones, transceivers, or radio control products. If you have to do so, the device should be observed to verify normal operation.
- The use of accessories and power cord other than those specified, with the exception of cables sold by the manufacturer of the equipment or system as replacement parts for internal components, may result in increased emissions or decreased immunity of the equipment or system.