ResMed HFT Lumis High Flow Therapy Device for COPD User Guide

June 6, 2024
ResMed

Quick start guide

Read your entire Lumis™ HFT User Guide before use.

Setup

With the device on a stable level surface below head height, connect one end of the power cord into the power supply unit and the other end into the power outlet.

Fit the tubing wrap around the air tubing.

Connect the air tubing firmly to the air outlet located on the rear of the device.

Open the humidifier and fill it with water up to the maximum water level mark.

Close the humidifier and insert it into the side of the device.

Connect the free end of the air tubing firmly onto the high-flow nasal cannula.

Note: It is recommended to place the device on a heat-resistant pad to avoid possible heat damage to surfaces.

Connecting/disconnecting supplemental oxygen
Your Lumis HFT device is designed to be compatible with levels of supplemental oxygen up to 15 L/min.
Always make sure that the device is turned on and airflow generated before the oxygen supply is turned on. Always turn the oxygen supply off before the device, so that unused oxygen does not accumulate within the device enclosure and create a risk of fire.
Note: Do not connect supplemental oxygen input directly to a medical gas pipeline system, oxygen bottle, or another high-pressure oxygen source. A flow regulator must be fitted to ensure the delivered oxygen remains at or below 15 L/min.
For advice on connecting to an oxygen source, contact your home care provider.

Refer to the Lumis HFT User Guide for warnings and cautions and details on how to perform a functional check.

Setting therapy
Note: Your clinician should have applied your prescribed settings to your device.

  1. Select Run Warmup.1. Select Run Warmup

  2. Warmup will take between 10-20 mins. Therapy can be started at any time during Warmup.

  3. When Warmup is complete, Ready will be displayed.

  4. Fit your high-flow nasal cannula.

  5. Press the Start/Stop button to start therapy.

  6. If you have supplemental oxygen, ensure the source is now turned ON.

  7. Ensure oxygen source is turned OFF.

  8. Remove your High Flow Nasal Cannula.

  9. Press the Start/Stop button and then press the dial to confirm Yes to stop therapy

  10. Drying mode will commence.

Warming mode pre-heats the water before starting therapy so that the air is not cold or dry at the start of therapy.
Drying mode runs for 90 mins and dries the air tubing and cannula so that it is ready for use the next time. Ensure that the air tubing and cannula remain connected to the device and airflow unimpeded during the Drying mode.
Caring for your device
Refer to the Lumis HFT User Guide for detailed cleaning and maintenance information.
Cleaning
You should clean the device weekly as described. Refer to the nasal cannula user guide for detailed instructions on cleaning your nasal cannula.

  1. Wash the humidifier and air tubing in warm water using mild detergent.
  2. Wash the tubing wrap in cold to warm water using a mild detergent. Rinse well and drip dry.
  3. Rinse the humidifier and air tubing thoroughly and allow to dry out of direct sunlight and/or heat.
  4. Wipe the exterior of the device with a dry cloth.

Checking – tubing and tubing wrap
Remove the tubing wrap and check the air tubing is clean. Replace it if there are any holes, tears, or cracks.

MANUFACTURER
ResMed Pty Ltd
1 Elizabeth Macarthur Drive
Bella Vista NSW 2153 Australia
See ResMed.com for other ResMed locations worldwide. Lumis is a trademark and/or registered trademark of the ResMed family
of companies. For patent and other intellectual property information, see www.resmed.com/ip.
© 2021 ResMed. 288389/2 2021-06
ResMed.com

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

ResMed User Manuals

Related Manuals