Disinfection guide
This guide is intended for multi-patient re-use of only the Sleep Lab Mask
(SLM) variant of the AirFit F40 mask in a sleep lab, clinic or hospital. If
you use the mask as a single user in the home, refer to the User Guide for
cleaning instructions. The processor is responsible for ensuring that
reprocessing is completed in accordance with ResMed’s validated procedures.
Potable water with hardness
≤150 mg/L is recommended for all cleaning, rinsing and disinfection steps.
Mask component1
| Cleaning|
Validated number of cycles
---|---|---
Alconox™ (Mild alkaline, anionic detergent)
Thermal Disinfection
Manual
167°F (75°C) for 30 minutes
194-199°F (90-93°C) for 1 to 10 minutes
Cushion| |
30
Frame|
Multi-hole vent ring|
Elbow and short tube|
Headgear|
QuietAir™ vent ring| –| Not intended for multi-patient use
1 May not be available in all regions. For full details regarding the correct use of these masks, please refer to the User Guide. For a list of available replacement parts, check the Product Guide on ResMed.com
A | Elbow and short tube |
---|---|
B | Frame |
C | Cushion |
D | Headgear |
1 | Anti-asphyxia valve (AAV) |
2 | Side button |
3 | Frame stud hole |
4 | Frame magnet |
5 | QuietAir vent ring |
6 | Multi-hole vent ring |
7 | Cushion stud |
--- | --- |
8 | Magnetic clip |
9 | Lower headgear strap |
10 | Upper headgear strap |
11 | Top headgear strap |
Manual cleaning and disinfection procedure
GENERAL WARNINGS AND CAUTIONS
- Always follow cleaning and disinfection instructions. Some cleaning products may damage the mask, its parts and their function, or leave harmful residual vapors. Any deviations from the procedures listed in this guide, including exceeding the specified number of maximum cycles, can have an adverse effect on the mask and consequently the safety or the quality of
- Failure to thoroughly clean and disinfect the mask component as indicated may result in inadequate
- When using detergents, disinfectants or equipment, always follow the manufacturer’s In the event of conflict, this guide takes precedence.
- Always follow safe operating practices, including the use of appropriate personal protective equipment (PPE), as Refer to the manufacturer’s safety instructions for more details.