ResMed AirSense 10 AutoSet Adjusting CPAP Device User Guide
- June 1, 2024
- ResMed
Table of Contents
AirSense 10 AutoSet ELITE
User guide
AirSense 10 AutoSet Adjusting CPAP Device
Welcome
The AirSense™ 10 AutoSet™ and AirSense 10 Elite are ResMed’s premium auto-
adjusting pressure and Continuous Positive Airway Pressure (CPAP) devices.
WARNING
- Read this entire guide before using the device.
- Use the device according to the intended use provided in this guide.
- The advice provided by your prescribing doctor should be followed ahead of the information provided in this guide.
Indications for use
AirSense 10 AutoSet
The AirSense 10 AutoSet self-adjusting device is indicated for the treatment
of obstructive sleep apnoea (OSA) in patients weighing more than 30 kg. It is
intended for home and hospital use.
The humidifier is intended for single patient use in the home environment and
re-use in a hospital/institutional environment.
AirSense 10 Elite
The AirSense 10 Elite device is indicated for the treatment of obstructive
sleep apnoea (OSA) in patients weighing more than 30 kg. It is intended for
home and hospital use.
The humidifier is intended for single patient use in the home environment and
re-use in a hospital/institutional environment.
Clinical benefits
The clinical benefit of CPAP therapy is a reduction in apnoeas, hypopnoeas and
sleepiness, as well as improved quality of life.
The clinical benefit of humidification is the reduction of positive airway
pressure related side effects.
Contraindications
Positive airway pressure therapy may be contraindicated in some patients with
the following pre-existing conditions:
- severe bullous lung disease
- pneumothorax
- pathologically low blood pressure
- dehydration
- cerebrospinal fluid leak, recent cranial surgery, or trauma.
Adverse effects
You should report unusual chest pain, severe headache, or increased
breathlessness to your prescribing physician. An acute upper respiratory tract
infection may require temporary discontinuation of treatment.
The following side effects may arise during the course of therapy with the
device:
- drying of the nose, mouth, or throat
- nosebleed
- bloating
- ear or sinus discomfort
- eye irritation
- skin rashes.
At a glance
The AirSense 10 includes the following:
- Device
- HumidAir™ humidifier (if supplied)
- Air tubing
- Power supply unit
- Travel bag
- SD card (not available in all devices).
Contact your care provider for a range of accessories available for use with the device including:
- Air tubing (heated and non-heated): ClimateLineAir™ , SlimLine™ , ClimateLineAir Oxy, Standard
- Side cover for use without the humidifier
- Filter: Hypoallergenic filter, standard filter
- Air10™ DC/DC converter
- SD card reader
- Air10 oximeter adapter
- Air10 USB adapter
- Power Station II
- Air10 tubing elbow.
About your device
- Air outlet
- Air filter cover
- Power inlet
- Serial number and device number
- HumidAir humidifier
- Screen
- Adapter cover
- SD card cover
About the control panel
| Start/Stop button| Press to start/stop therapy.
Press and hold for three seconds to enter power save mode.
---|---|---
| Dial| Turn to navigate the menu and press to select an option.
Turn to adjust a selected option and press to save your change.
| Home button| Press to return to the Home screen.
Different icons may be displayed on the screen at different times including:
| Ramp Time
---|---
| Ramp Time Auto
| Humidity
| Humidifier warming
| Humidifier cooling
| Wireless signal strength (green)
| Wireless transfer not enabled (grey)
| No wireless connection
| Airplane Mode
Setup
CAUTION
Do not overfill the humidifier as water may enter the device and air tubing.
-
Place the device on a stable level surface.
-
Plug the power connector into the rear of the device. Connect one end of the power cord into the power supply unit and the other end into the power outlet.
-
Connect the air tubing firmly to the air outlet located on the rear of the device.
-
Open the humidifier and fill it with water up to the maximum water level mark.
Do not fill the humidifier with hot water. -
Close the humidifier and insert it into the side of the device.
-
Connect the free end of the air tubing firmly onto the assembled mask.
See the mask user guide for detailed information.
Recommended masks are available on www.resmed.com.
Starting therapy
- Fit your mask.
- Press Start/Stop or breathe normally if SmartStart™ is enabled.
You will know that therapy is on when the Sleep Report screen is displayed.
The current treatment pressure is shown in green.
During ramp time the pressure is gradually increasing and you will see a
spinning circle. Once the prescribed treatment pressure is reached, the entire
circle will be green. The screen will go black automatically after a short
period of time. You can press Home or the dial to turn it back on. If power is
interrupted during therapy, the device will automatically restart therapy when
power is restored.
The AirSense 10 device has a light sensor that adjusts the screen brightness
based on the light in the room.
Stopping therapy
- Remove your mask.
- Press Start/Stop or if SmartStart is enabled, therapy will stop automatically after a few seconds.
The Sleep Report now gives you a summary of your therapy session.
Usage hours–Indicates the number of hours of therapy you received last
session.
Mask Seal–Indicates how well your mask sealed:
Good mask seal.
Needs adjusting, see Mask Fit. Humidifier–Indicates if your humidifier is
working properly:
Humidifier working.
Humidifier might be faulty, contact your care provider.
If set by your care provider, you will also see:
Events per hour–Indicates the number of apnoeas and hypopnoeas experienced per
hour.
More Info–Turn the dial to scroll down to view more detailed usage data.
Power save mode
Your AirSense 10 device records your therapy data. In order to allow it to
transmit the data to your care provider, you should not unplug the device.
However, you can put it into power save mode to save electricity.
To enter power save mode:
-
Press and hold Start/Stop for three seconds.
The screen goes black.
To exit power save mode: -
Press Start/Stop once.
The Home screen is displayed.
My Options
Your AirSense 10 device has been set up for your needs by your care provider,
but you may find you want to make small adjustments to make your therapy more
comfortable.
Highlight My Options and press the dial to see yourcurrent settings. From
here, you can personalise yoptions. Ramp Time
Designed to make the beginning of therapy more comfortable, Ramp Time is the
period during which the pressure increases from a low start pressure to the
prescribed treatment pressure.
You can set your Ramp Time to Off, 5 to 45 minutes or Auto. When Ramp Time is
set to Auto, the device will detect when you have fallen asleep and then
automatically rise to the prescribed treatment pressure.
To adjust Ramp Time:
- In My Options, turn the dial to highlight Ramp Time and then press the dial.
- Turn the dial to adjust the ramp time to your preferred setting and press the dial to save the change.
Humidity Level
The humidifier moistens the air and is designed to make therapy more
comfortable. If you are getting a dry nose or mouth, turn up the humidity. If
you are getting any moisture in your mask, turn down the humidity.
You can set the Humidity Level to Off or between 1 and 8, where 1 is the
lowest humidity setting and 8 is the highest humidity setting.
To adjust the Humidity Level:
-
In My Options, turn the dial to highlight Humidity
Level and then press the dial. -
Turn the dial to adjust the humidity level and press the dial to save the change.
If you continue to get a dry nose or mouth, or moisture in your mask, consider
using ClimateLineAir heated air tubing. ClimateLineAir together with Climate
Control delivers more comfortable therapy.
Mask Fit
Mask Fit is designed to help you assess and identify possible air leaks
around your mask.
To check Mask Fit:
-
Fit the mask as described in the mask user guide.
-
In My Options, turn the dial to highlight Run Mask Fit and then press the dial.
The device starts blowing air. -
Adjust the mask, mask cushion and headgear until you get a Good result.
To stop Mask Fit, press the dial or Start/Stop. If you are unable to get a
good mask seal, talk to your care provider.
More options
There are some more options on your device which you can personalise.
Pressure Relief
When Pressure Relief is enabled, you may find it easier to breathe out. This
can help you get used to therapy.
SmartStart
When SmartStart is enabled, therapy starts automatically when you breathe into
your mask. When you remove your mask, it stops automatically after a few
seconds.
*When enabled by your care provider.
Caring for your device
It is important that you regularly clean your AirSense 10 device to make sure
you receive optimaltherapy. The following sections will help you with
disassembling, cleaning, checking and reassembling your device.
WARNING
Regularly clean your tubing assembly, humidifier and mask to receive
optimal therapy and to prevent the growth of germs that can adversely affect
your health.
Disassembling
- Hold the humidifier at the top and bottom, press it gently and pull it away from the device.
- Open the humidifier and discard any remaining water.
- Hold the cuff of the air tubing and gently pull it away from the device.
- Hold both the cuff of the air tubing and the swivel of the mask, then gently pull apart.
Cleaning
You should clean the device weekly as described. Refer to the mask user guide
for detailed instructions on cleaning your mask.
- Wash the humidifier and air tubing in warm water using mild detergent.
- Rinse the humidifier and air tubing thoroughly and allow to dry out of direct sunlight and/or heat.
- Wipe the exterior of the device with a dry cloth.
Notes:
-
The humidifier may be washed in a dishwasher on the delicate or glassware cycle (top shelf only).
It should not be washed at temperatures higher than 65ºC. -
Do not wash the air tubing in a dishwasher or washing machine.
-
Empty the humidifier daily and wipe it thoroughly with a clean, disposable cloth. Allow to dry out of direct sunlight and/or heat.
Checking
You should regularly check the humidifier, air tubing and the air filter for
any damage.
-
Check the humidifier:
• Replace it if it is leaking or has become cracked, cloudy or pitted.
• Replace it if the seal is cracked or torn.
• Remove any white powder deposits using a solution of one part household vinegar to 10 parts water. -
Check the air tubing and replace it if there are any holes, tears or cracks.
-
Check the air filter and replace it at least every six months. Replace more often if there are any holes or blockages by dirt or dust.
To replace the air filter: ![ResMed AirSense 10 AutoSet Adjusting CPAP Device
-
Checking](https://manuals.plus/wp-content/uploads/2024/04/ResMed-AirSense-10 -AutoSet-Adjusting-CPAP-Device-Checking.jpg)
-
Open the air filter cover and remove the old air filter.
The air filter is not washable or reusable. -
Place a new air filter onto the air filter cover and then close it.
Make sure the air filter is fitted at all times to prevent water and dust from entering the device.
-
Reassembling
When the humidifier and air tubing are dry, you can reassemble the parts.
- Connect the air tubing firmly to the air outlet located on the rear of the device.
- Open the humidifier and fill it with room temperature water up to the maximum water level mark.
- Close the humidifier and insert it into the side of the device.
- Connect the free end of the air tubing firmly onto the assembled mask.
Therapy data
Your AirSense 10 device records your therapy data for you and your care
provider so they can view and make changes to your therapy if required. The
data is recorded and then transferred to your care provider wirelessly, if a
wireless network is available, or via an SD card.
Data transmission
Your AirSense 10 device has the capability of wireless communication so
that your therapy data can be transmitted to your care provider to improve the
quality of your treatment. This is an optional feature that will only be
available if you choose to benefit from it and if a wireless network is
available. It also allows your care provider to update your therapy settings
in a more timely manner or upgrade your device software to ensure you receive
the best therapy possible.
The data is usually transmitted after therapy has stopped. In order to make
sure that your data is transferred, leave your device connected to the mains
power at all times and make sure that it is not in Airplane Mode.
Notes:
- Therapy data might not be transmitted if you use it outside of the country or region of purchase.
- Wireless communication depends on network availability.
- Devices with wireless communication might not be available in all regions.
SD card
An alternative way for your therapy data to be transferred to your care
provider is via the SD card.
Your care provider may ask you to send the SD card by mail or to bring it in.
When instructed by your care provider, remove the SD card.
Do not remove the SD card from the device when the SD light is flashing,
because data is being written to the card.
To remove the SD card: ![ResMed AirSense 10 AutoSet Adjusting CPAP Device
-
SD card](https://manuals.plus/wp-content/uploads/2024/04/ResMed-AirSense-10 -AutoSet-Adjusting-CPAP-Device-SD-card.jpg)
- Open the SD card cover.
- Push in the SD card to release it. Remove the SD card from the device.
Place the SD card in the protective folder and send it back to your care provider.
For more information on the SD card refer to the SD card protective folder
provided with your device.
Note: The SD card should not be used for any other purpose.
Travelling
You can take your AirSense 10 device with you wherever you go. Just keep the following in mind:
- Use the travel bag provided to prevent damage to the device.
- Empty the humidifier and pack it separately in the travel bag.
- Make sure you have the appropriate power cord for the region you are travelling to. For information on purchasing, contact your care provider.
- If you are using an external battery, you should turn off the humidifier in order to maximise the life of your battery. Do this by turning the Humidity Level to Off.
Travelling by plane
Your AirSense 10 device may be taken on board as carry-on luggage. Medical
devices do not count toward your carry-on luggage limit.
You can use your AirSense 10 device on a plane as it meets the Federal
Aviation Administration (FAA) requirements. Air travel compliance letters can
be downloaded and printed from www.resmed.com.
When using the device on a plane:
- Make sure the humidifier is completely empty and inserted into your device. The device will not work without the humidifier inserted.
- Turn on Airplane Mode.
To turn on Airplane Mode:
- In My Options, turn the dial to highlight Airplane Mode and then press the dial.
- Turn the dial to select On and then press the dial to save the change.
The Airplane Mode icon is displayed at the top right of the screen.
CAUTION
Do not use the device with water in the humidifier on a plane due to the risk
of inhalation of water during turbulence.
Troubleshooting
If you have any problems, have a look at the following troubleshooting topics.
If you are not able to fix the problem, contact your care provider or ResMed.
Do not try to open the device.
General troubleshooting
Problem/possible cause | Solution |
---|
Air is leaking from around my mask
Mask may be fitted incorrectly.| Make sure your mask is fitted correctly. See
your mask user guide for fitting instructions or use the Mask Fit function to
check your mask fit and seal.
I am getting a dry or blocked nose
Humidity level may be set too low.| Adjust the Humidity Level.
If you have ClimateLineAir heated air tubing, see the ClimateLineAir user
guide.
I am getting droplets of water on my nose, in the mask and air tubing
Humidity level may be set too high.| Adjust the Humidity Level.
If you have ClimateLineAir heated air tubing, see the ClimateLineAir user
guide.
My mouth is very dry and uncomfortable
Air may be escaping through your mouth.| Increase the Humidity Level.
You may need a chin strap to keep your mouth closed or a full face mask.
Air pressure in my mask seems too high (it feels like I am getting too much
air)
Ramp may be turned off.| Use the Ramp Time option.
Air pressure in my mask seems too low (it feels like I am not getting enough
air)
Ramp may be in progress.| Wait for air pressure to build up or turn Ramp Time
off.
My screen is black
Backlight on the screen may have turned off. It turns off automatically after
a short period of time.
Power may not be connected.| Press Home or the dial to turn it back on.
Connect the power supply and make sure the plug is fully inserted.
I have stopped therapy, but the device is still blowing air
Device is cooling down.| Device blows a small amount of air in order to avoid
condensation in the air tubing. It will stop automatically after 30 minutes.
My humidifier is leaking
Humidifier may not be assembled correctly.
Humidifier may be damaged or cracked.| Check for damage and reassemble the
humidifier correctly.
Contact your care provider for a replacement.
My therapy data has not been sent to my care provider
Wireless coverage may be poor.
The No wireless connection icon is displayed on the top right of the screen.
no wireless network available.
Device may be in Airplane Mode.
Data transfer is not enabled for your device.| Make sure that the device is
placed where there is coverage (ie, on your bedside table, not in a drawer or
on the floor).
The Wireless signal strength icon indicates good coverage when all bars are
displayed, and poor coverage when fewer bars are displayed.
Make sure that the device is placed where there is coverage (ie, on your
bedside table, not in a drawer or on the floor).
If instructed to do so, send the SD card to your care provider.
The SD card also contains your therapy data.
Turn off Airplane Mode, see Travelling by plane.
Talk to your care provider about your settings.
My screen and buttons are flashing
Software upgrade is in progress.| Software upgrade takes approximately 10
minutes to complete.
Device messages
Device message/possible cause | Solution |
---|
High leak detected, check your water tub, tub seal or side cover
Humidifier may not be inserted properly.
Humidifier seal may not be inserted properly.| Make sure the humidifier is
correctly inserted.
Open the humidifier and make sure that the seal is correctly inserted.
High leak detected, connect your tubing
Air tubing may not be connected properly.
Mask may be fitted incorrectly.| Make sure the air tubing is firmly connected
at both ends.
Make sure your mask is fitted correctly. See your mask user guide for fitting
instructions or use the Mask Fit function to check your mask fit and seal.
Tubing blocked, check your tubing
Air tubing may be blocked.| Check the air tubing and remove any blockages.
Press the dial to clear the message and then press Start/Stop to restart the
device.
SD card error, remove your card and press Start to begin therapy
SD card may not be inserted correctly.| Remove and reinsert the SD card.
Read only card, please remove, unlock and re-insert SD card
SD card switch may be in the lock (read-only) position.| Move the switch on
the SD Card from the lock position to the unlock position and then re-insert
it.
System fault, refer to user guide, Error 004
Device may have been left in a hot environment.
Air filter may be blocked.
Air tubing may be blocked.
There may be water in the air tubing.| Allow to cool before re-use. Disconnect
the power supply and then reconnect it to restart the device.
Check the air filter and replace it if there are any blockages.
Disconnect the power supply and then reconnect it to restart the device.
Check the air tubing and remove any blockages. Press the dial to clear the
message and then press Start/Stop to restart the device.
Empty the water from the air tubing. Disconnect the power supply and then
reconnect it to restart the device.
All other error messages, for example, System fault, refer to user guide,
Error 0XX
An unrecoverable error has occurred on the device.| Contact your care
provider. Do not open the device.
Reassembling parts
Some parts of your device are designed to easily come off in order to avoid
damage to the parts or the device. You can easily reassemble them as described
below.
To insert the humidifier seal:
- Place the seal into the lid.
- Press down along all edges of the seal until it is firmly in place.
To reassemble the humidifier lid:
- Insert one side of the lid into the pivot hole of the base.
- Slide the other side down the ridge until it clicks into place.
WARNING
-
Make sure that you arrange the air tubing so that it will not twist around the head or neck.
-
Regularly inspect power cords, cables, and power supply for damage or signs of wear.
Discontinue use and replace if damaged. -
Keep the power cord away from hot surfaces.
-
If you notice any unexplained changes in the performance of the device, if it is making unusual sounds, if the device or the power supply are dropped or mishandled, or if the enclosure is broken, discontinue use and contact your care provider or your ResMed Service Centre.
-
Do not open or modify the device. There are no user serviceable parts inside. Repairs and servicing should only be performed by an authorised ResMed service agent.
-
Beware of electrocution. Do not immerse the device, power supply or power cord in water.
If liquids are spilled into or onto the device, unplug the device and let the parts dry. Always unplug the device before cleaning and make sure that all parts are dry before plugging it back in. -
Supplemental oxygen must not be used while smoking or in the presence of an open flame.
-
Always make sure that the device is turned on and airflow generated before the oxygen supply is turned on. Always turn the oxygen supply off before the device is turned off, so that unused oxygen does not accumulate within the device enclosure and create a risk of fire.
-
Do not perform any maintenance tasks while the device is in operation.
-
The device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the device should be observed to verify normal operation in the configuration in which it will be used.
-
The use of accessories other than those specified for the device is not recommended. They may result in increased emissions or decreased immunity of the device.
-
Regularly check the antibacterial filter for signs of moisture or other contaminants, particularly during nebulization or humidification. Failure to do so could result in increased breathing system resistance.
CAUTION
- Use only ResMed parts and accessories with the device. Non-ResMed parts may reduce the effectiveness of the treatment and/or damage the device.
- Use only vented masks recommended by ResMed or by the prescribing doctor with this device. Fitting the mask without the device blowing air can result in rebreathing of exhaled air. Make sure that the mask vent holes are kept clear and unblocked to maintain the flow of the fresh air into the mask.
- Be careful not to place the device where it can be bumped or where someone is likely to trip over the power cord.
- Blocking the air tubing and/or air inlet of the device while in operation could lead to overheating of the device.
- Keep the area around the device dry, clean and clear of anything (eg, clothes or bedding) that could block the air inlet or cover the power supply unit.
- Do not place the device on its side as water might get into the device.
- Incorrect system setup may result in incorrect mask pressure reading. Ensure the system is correctly set up.
- Do not use bleach, chlorine, alcohol, or aromatic-based solutions, moisturising or antibacterial soaps or scented oils to clean the device, the humidifier or air tubing. These solutions may cause damage or affect the humidifier performance and reduce the life of the products. Exposure to smoke, including cigarette, cigar or pipe smoke, as well as ozone or other gases, may damage the device. Damage caused by any of the foregoing, will not be covered by ResMed’s limited warranty.
- If you use the humidifier, always place the device on a level surface lower than your head to prevent the mask and air tubing from filling with water.
- Leave the humidifier to cool for ten minutes before handling to allow the water to cool and to make sure that the humidifier is not too hot to touch.
- Make sure that the humidifier is empty before transporting the device.
Note: For any serious incidents that occur in relation to this device, these should be reported to ResMed and the competent authority in your country.
Technical specifications
Units are expressed in cm H2O and hPa. 1 cm H2O is equal to 0.98 hPa.
90W power supply unit
AC input range:
DC output:
Typical power consumption:
Peak power consumption:| 100–240V, 50–60Hz 1.0–1.5A, Class II
115V, 400Hz 1.5A, Class II (nominal for aircraft use)
24V 3.75A
53W (57VA)
104W (108VA)
Environmental conditions
Operating temperature:
Operating humidity:
Operating altitude:
Storage and transport temperature:
Storage and transport humidity:| +5°C to +35°C
Note: The air flow for breathing produced by this therapy device can be
higher than the temperature of the room.
Under extreme ambient temperature conditions (40°C) the device remains safe.
10 to 95% relative humidity, non-condensing
Sea level to 2,591 m; air pressure range 1013 hPa to 738 hPa
-20°C to +60°C
5 to 95% relative humidity, non-condensing
ElectroElectromagnetic compatibility
The AirSense 10 complies with all applicable electromagnetic compatibility
requirements (EMC) according to IEC 60601-1-2:2014, for residential,
commercial and light industry environments. It is recommended that mobile
communication devices are kept at least 1 m away from the device.
Information regarding the electromagnetic emissions and immunity of this
ResMed device can be found on
www.resmed.com/downloads/devices
Classification: EN60601-1:2006/A1:2013| Class II (double insulation), Type BF,
Ingress protection IP22.
Sensors
Pressure sensor:
Flow sensor:| Internally located at device outlet, analog gauge pressure type,
0 to 40 cm H2O (0 to 40 hPa)
Internally located at device inlet, digital mass flow type, -70 to +180 L/min
Maximum single fault steady pressure| Device will shut down in the presence of
a single fault if the steady state pressure exceeds:
30 cm H2O (30 hPa) for more than 6 sec or 40 cm H2O (40 hPa) for more than 1
sec.
Sound
Pressure level measured according to ISO 80601-2-70:2015 (CPAP mode):
SlimLine:
Standard:
SlimLine or Standard and humidification:
Power level measured according to ISO 80601-2-70:2015 (CPAP mode):
SlimLine:
Standard:
SlimLine or Standard and humidification:
Declared dual-number noise emission values in accordance with ISO 4871:| 25
dBA with uncertainty of 2 dBA
25 dBA with uncertainty of 2 dBA
27 dBA with uncertainty of 2 dBA
33 dBA with uncertainty of 2 dBA
33 dBA with uncertainty of 2 dBA
35 dBA with uncertainty of 2 dBA
1996.
Physical – device and humidifier
Dimensions (H x W x D):
Air outlet (complies with ISO 5356-1:2015): Weight (device and cleanable
humidifier):
Housing construction:
Water capacity:
Cleanable humidifier – material:| 116 mm x 255 mm x 150 mm
22 mm
1248 g
Flame retardant engineering thermoplastic
To maximum fill line 380 mL
Injection moulded plastic, stainless steel and silicone seal
Temperature
Maximum heater plate:
Cut-out:
Maximum gas temperature:| 68°C
74°C
≤ 41°C
Air filter
Standard:
Hypoallergenic:| Material: Polyester non woven fibre
Average arrestance: >75% for ~7 micron dust
Material: Acrylic and polypropylene fibres in a polypropylene carrier
Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron dust
Aircraft use| ResMed confirms that device meets the Federal Aviation
Administration (FAA) requirements (RTCA/DO-160, section 21, category M) for
all phases of air travel.
Wireless module
Technology used: 4G, 3G, 2G
It is recommended that the device is a minimum distance of 0.8″ (2 cm) from
the body during operation. Not applicable to masks, tubes or accessories.
Technology may not be available in all regions.
Declaration of Conformity (DoC to the Radio Equipment Directive)
ResMed declares that the AirSense 10 device (models 370xx or 371xx) is in
compliance with the essential requirements and other relevant provisions of
Directive 2014/53/EU (RED). A copy of the Declaration of Conformity (DoC) can
be found on Resmed.com/productsupport
The 2G radio equipment operates with the following frequency bands and maximum
radio-frequency power:
GSM 850/900: 35dBm
GSM 1800/1900: 32dBm
The 4G device can be used in all European countries without any restrictions.
All ResMed devices are classified as medical devices under the Medical Device
Directive. Any labelling of the product and
printed material, showing 0123, relates to the Council Directive 93/42/EEC
including the Medical Device Directive amendment (2007/47/EC).
Operating pressure range
AutoSet, CPAP:| 4 to 20 cm H2O (4 to 20 hPa)
Supplemental oxygen
Pneumatic flow path| 4 L/min
| 1. Flow sensor
2. Blower
3. Pressure sensor
4. Mask
5. Air tubing
6. Humidifier
7. Device
8. Inlet filter
---|---
Design life
Device, power supply unit:
Cleanable humidifier: Air tubing:| 5 years
2.5 years
6 months
General
The patient is an intended operator.|
Humidifier performance
Mask Pressure cm H2O (hPa)| RH output % at 17°Cambient temperature| RH output
% at 22°C ambient
temperature| Nominal system output AH1, BTPS2
---|---|---|---
| Setting 4| Setting 8| Setting 4| Setting 8
4
10
20| 85
85
85| 100
100
90| 6
6
6| >10
10
10
1AH – Absolute Humidity in mg/L
2BTPS – Body Temperature Pressure Saturated
Air tubing
Air tubing | Material | Length | Inner diameter |
---|
ClimateLineAir
ClimateLineAir Oxy
SlimLine
Standard
Heated air tubing temperature cut-out:| Flexible plastic and electrical
components
Flexible plastic and electrical components Flexible plastic Flexible plastic ≤
41°C| 2 m
1.9 m
1.8 m
2 m| 15 mm
19 mm
15 mm
19 mm
Notes:
- The manufacturer reserves the right to change these specifications without notice.
- The electrical connector end of the heated air tubing is only compatible with the air outlet at the device end and should not be fitted to the mask.
- Do not use electrically conductive or antistatic air tubing.
- The temperature and relative humidity settings displayed are not measured values.
Displayed values
Value | Range | Display resolution |
---|
Pressure sensor at air outlet:
Mask pressure Flow derived values:
Leak| 4–20 cm H20 (4–20 hPa) 0–120 L/min| 0.1 cm H2O (0.1 hPa)
1 L/min
Value| Accuracy
---|---
Pressure measurement¹:
Mask pressure²
Flow and flow derived values¹:
Flow
Leak²| ±[0.5 cm H20 (0.5 hPa) + 4% of measured value] ±6 L/min or 10% of
reading, whichever is greater, at 0 to 150 L/min positive flow
±12 L/min or 20% of reading, whichever is greater, 0 to 60 L/min
¹Results are expressed as STPD (Standard Temperature and Pressure, Dry).
²Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal
volumes <100 mL or minute ventilation <3 L/min.
³Measurement accuracy verified as per EN ISO 10651-6:2009 for Home Care
Ventilatory Support Devices (Figure 101 and Table 101) using nominal ResMed
mask vent flows.
Measurement system uncertainties
In accordance with ISO 80601-2-70:2015 the measurement uncertainty of the
manufacturer’s test equipment is:
For measures of flow
For measures of volume (< 100 mL)
For measures of volume (≥ 100 mL)
For measures of static pressure
For measures of dynamic pressure
For measures of time| ± 1.5 L/min or ± 2.7% of reading (whichever is greater)
± 5 mL or 6% of reading (whichever is greater)
± 20 mL or 3% of reading (whichever is greater)
± 0.15 cm H2O (hPa)
± 0.27 cm H2O (hPa)
± 10 ms
---|---
Note: ISO 80601-2-70:2015 stated accuracies and test results provided in this manual for these items already include the relevant measurement uncertainty from the table above.
Pressure accuracy
Maximum static pressure variation at 10 cm H2O (10 hPa) according to ISO
80601-2-70:2015
Without humidification
With humidification| Standard air tubing
± 0.5 cm H2O (± 0.5 hPa)
± 0.5 cm H2O (± 0.5 hPa)| SlimLine air tubing
± 0.5 cm H2O (± 0.5 hPa)
± 0.5 cm H2O (± 0.5 hPa)
---|---|---
Maximum dynamic pressure variation according to ISO 80601-2-70:2015
Device without humidification and Standard air tubing / Device with
humidification and Standard air tubing
Pressure [cm H2O (hPa)] 4
8
12
16
20| 10 BPM
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5| 15 BPM
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5| 20 BPM
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
---|---|---|---
Device without humidification and SlimLine air tubing / Device with humidification and SlimLine air tubing
Pressure [cm H2O (hPa)] 4
8
12
16
20| 10 BPM
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5| 15 BPM
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5| 20 BPM
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
---|---|---|---
Flow (maximum) at set pressures
The following are measured accordingly to ISO 80601-2-70:2015 at the end of
the specified air tubing:
Pressure
cm H2O (hPa)| AirSense 10 and Standard
L/min| AirSense 10, humidification and Standard
L/min| AirSense 10 and
SlimLine
L/min| AirSense 10,
humidification and
ClimateLineAir
L/min
---|---|---|---|---
4
8
12
16
20| 180
168
157
144
131| 143
135
136
134
123| 162
151
140
128
117| 151
142
135
121
109
Resistance to flow
The table illustrates the resistance to flow of the air tubing:
Air tubing| At flow (L/min) with
pressure of 20 cm H
2
O| Resistance to flow (cm H
2
O/L/min)
---|---|---
Air tubing with tubing
elbow| Air tubing only
Standard
SlimLine
ClimateLineAir
ClimateLineAir Oxy| 30
15
30
15
30
15
30
15| 0.006
0.005
0.008
0.006
–
–
–
–| 0.005
0.004
0.007
0.006
0.011
0.008
0.004
0.002
Compliance
The table illustrates the compliance of the air tubing:
Air tubing | Compliance (cm H2O/L/min) with pressure of 60 cm H2O |
---|---|
Air tubing with tubing elbow | Air tubing only |
Standard
SlimLine
ClimateLineAir
ClimateLineAir Oxy| 1.074
0.467
–
–| 1.056
0.454
0.482
0.729
Symbols
The following symbols may appear on the product or packaging.
Read instructions before use. Indicates a warning or caution. Follow
instructions before use. Manufacturer. European Authorised Representative.
Batch code. Catalogue number. Serial number. Device number. On / Off.
Device weight. IP22 Protected against finger sized objects and against
dripping water when tilted up to 15 degrees from specified orientation. Direct
current. type BF applied part. Class II equipment. Humidity limitation.
Temperature limitation. Non-ionising radiation. China pollution control logo
- China pollution control logo 2. Prescription only (In the US, Federal law
restricts these devices to sale by or on the order of a physician). Maximum
water level. Use distilled water only. Operating altitude. Atmospheric
pressure limitation. Complies with RTCA DO-160 section 21, category M. Date of
manufacture. Alarm inhibit. Importer. Medical device.
See symbols glossary at ResMed.com/symbols.
Environmental information
This device should be disposed of separately, not as unsorted municipal waste. To dispose of your device, you should use appropriate collection, reuse and recycling systems available in your region.The use of these collection, reuse and recycling systems is designed to reduce pressure on naturresources and prevent hazardous substances from damaging the environment.
If you need information on these disposal systems, please contact your local waste administration.The crossed-bin symbol invites you to use these disposal systems. If you require information on collection and disposal of your ResMed device please contact your ResMed office, local distribut or go to ResMed.com/environment.
Servicing
The AirSense 10 device is intended to provide safe and reliable operation when
operated in accordance with the instructions provided by ResMed. ResMed
recommends that the AirSense 10 device be inspected and serviced by an
authorised ResMed Service Centre if there is any sign of wear or concern with
device function. Otherwise, service and inspection of the products generally
should not be required during their design life.
Limited warranty
ResMed Pty Ltd (hereafter ‘ResMed’) warrants that your ResMed product shall be
free from defects in material and workmanship from the date of purchase for
the period specified below.
Product | Warranty period |
---|
• Mask systems (including mask frame, cushion, headgear and tubing)—excluding
single-use devices
• Accessories—excluding single-use devices
• Flex-type finger pulse sensors
• Humidifier water tubs| 90 days
• Batteries for use in ResMed internal and external battery systems| 6 months
• Clip-type finger pulse sensors
• CPAP and bilevel device data modules
• Oximeters and CPAP and bilevel device oximeter adapters
• Humidifiers and humidifier cleanable water tubs
• Titration control devices| 1 year
• CPAP, bilevel and ventilation devices (including external power supply
units)
• Battery accessories
• Portable diagnostic/screening devices| 2 years
This warranty is only available to the initial consumer. It is not
transferable.
During the warranty period, if the product fails under conditions of normal
use, ResMed will repair or replace, at its option, the defective product or
any of its components.
This limited warranty does not cover: a) any damage caused as a result of
improper use, abuse, modification or alteration of the product; b) repairs
carried out by any service organisation that has not been expressly authorised
by ResMed to perform such repairs; c) any damage or contamination due to
cigarette, pipe, cigar or other smoke; d) any damage caused by exposure to
ozone, activated oxygen or other gases; and e) any damage caused by water
being spilled on or into an electronic device.
Warranty is void on product sold, or resold, outside the region of original
purchase. For product purchased in a country in the European Union (“EU”) or
European Free Trade Association (“EFTA”), “region” means the EU and EFTA.
Warranty claims on defective product must be made by the initial consumer at
the point of purchase.
This warranty replaces all other expressed or implied warranties, including
any implied warranty of merchantability or fitness for a particular purpose.
Some regions or states do not allow limitations on how long an implied
warranty lasts, so the above limitation may not apply to you.
ResMed shall not be responsible for any incidental or consequential damages
claimed to have resulted from the sale, installation or use of any ResMed
product. Some regions or states do not allow the exclusion or limitation of
incidental or consequential damages, so the above limitation may not apply to
you.
This warranty gives you specific legal rights, and you may also have other
rights which vary from region to region. For further information on your
warranty rights, contact your local ResMed dealer or ResMed office.
Visit ResMed.com for the latest information on ResMed’s Limited Warranty.
Further information
If you have any questions or require additional information on how to use the
device, contact your care provider.
ResMed Pty Ltd
1 Elizabeth Macarthur Drive
Bella Vista NSW 2153 Australia
See ResMed.com for other ResMed locations worldwide.
Air10, AirSense, AutoSet, ClimateLine,
HumidAir and SlimLine are trademarks and/or
registered trademarks of the ResMed family of companies.
For patent and other intellectual property information,
see ResMed.com/ip. SD Logo
is a trademark of SD-3C, LLC.
© 2020 ResMed. 378970/1 2020-06
ResMed.com
References
- Sleep apnea and COPD - learn about symptoms and treatment | ResMed
- Legal disclaimers and intellectual property - ResMed
- Resmed.com/productsupport
- ResMed.com/symbols
- Sleep apnea and COPD - learn about symptoms and treatment | ResMed
- resmed.com/downloads/devices
- Environmental policy - ResMed