RESMED H5i Standard Water Tub Instruction Manual
- June 6, 2024
- ResMed
Table of Contents
RESMED H5i Standard Water Tub
**Instruction Manual**
You should regularly carry out cleaning and maintenance of the standard water tub as described in this section.
Disassembling the H5i water tub
Refer to illustration A,
- Slide the latch.
- Lift open the flip lid.
- Remove the water tub.
- Discard any excess water from the water tub.
- Unclip all four side latches.
- Pull apart the tub lid and base.
Daily cleaning
- Wash the disassembled tub lid and base in warm water using a mild detergent.
- Rinse thoroughly in clean water and allow them to dry away from direct sunlight.
Monthly cleaning
- Peel the flip lid seal from the flip lid and wash it in warm water using a mild detergent.
- Wipe the exterior of the H5i with a damp cloth and mild detergent.
Maintenance checklist
⇒Inspect the water tub and flip lid seal for wear and deterioration.
⇒Replace the water tub if any component is leaking or has become cracked,
cloudy or pitted.
⇒Replace the flip lid seal if cracked or torn.
⇒Clean white powder deposits in the water tub by using a solution of one part
household vinegar to 10 parts water.
Reassembling and filling the H5i water tub
Refer to illustration B
- Place the tub lid back onto the base.
- Clip all four side latches.
- Fill the water tub with distilled or deionized water up to the maximum water level mark.
- Return the water tub to the H5i.
- Close the flip lid ensuring that it clicks into place.
For more information on the H5i, see your Welcome and Information Guides. This
documentation is also available to download from
www.resmed.com on the Products page under Service &
Support.
WARNING
- Beware of electrocution. Do not immerse your device or power cord in water. Always unplug the device before cleaning and make sure it is dry before plugging it back in.
- Take care when handling your H5i. The water and the water tub base may both be hot. Allow 10 minutes for the heater plate
and any excess water to cool.
Manufacturer : ResMed Pty Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia. Distributed by: ResMed Corp 9001 Spectrum Center Boulevard San Diego CA 92123 USA. ResMed (UK) Ltd 96 Milton Park Abingdon Oxfordshire OX14 4RY UK. See www.resmed.com for other ResMed locations worldwide.
Protected by patents: AU 2002233025, CN ZL02804936.5, HK 1065483, JP 4420605, NZ 527088, US 6935337, US 7614398. Other patents pending. Protected by design registrations: AU 331676, AU 331677, CN 200930203243.5, EU 1720582, JP 1402844, JP 1402845, NZ 413752, NZ 413753, US D626642, US D627044. Other designs pending. H5i is a trademark of ResMed.
© 2011 ResMed. 368659/1 2011 10
Global leaders in sleep and respiratory medicine
www.resmed.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>