aspire LSS390580 Patient Mover User Manual

June 5, 2024
aspire

aspire logo.jpg

aspire LSS390580 Patient Mover User Manual

aspire LSS390580 Patient Mover.jpg

IMPORTANT:

Do not use the Aspire Star Patient Mover without having first read this manual.

1. OVERVIEW

The Aspire Star Patient Mover (the Star Mover), provides a simple solution for transferring a user from a bed or chair in comfort and safety.

Offering a compact and versatile design, the Star Mover is suitable for Homecare, Acute Care and Aged Care environments.

FIG 1 OVERVIEW.jpg

2. SAFETY PRECAUTIONS

INTENDED USE OF THE PRODUCT
The Star Mover has been designed to transfer human patients with the assistance of one carer or attendant.

The Star Mover should only be used with Aspire approved standing slings, which have been chosen and assessed by a competent trained health professional.

For a complete list of all compliant slings please contact Aidacare on 1300 133 120.
IMPORTANT – only approved slings and accessories are compatible with this product. If you are unsure please contact Aidacare on 1300 133 120.

The correct and appropriate use of the Star Mover reduces the risks associated with manual patient handling. The lifting and transferring of patients should always be assessed and planned by a trained health professional.

SAFETY CONSIDERATIONS

  • ALWAYS ensure that attendants have read this guide before operating the patient mover.
  • ALWAYS use slings that are assessed and considered suitable by a trained healthcare professional.
  • ALWAYS ensure the area around the patient mover and patient is clear of obstructions.
  • ALWAYS minimise the distance over which the patient is being transferred by positioning support surfaces as close as possible.
  • ALWAYS ensure braking castors are locked when performing transfers in and out of the Star Mover.
  • DO NOT operate the patient mover without correct manual handling training. IF IN DOUBT, DO NOT USE.
  • DO NOT use the patient mover where any doubt exists regarding the safety of the patient mover, including but not limited to wear or damage with particular importance to be placed on the standing sling attachment points.
  • DO NOT use the patient mover with slings showing any signs of wear or damage, including fraying.
  • DO NOT exceed the 185kg safe working load of the Star Mover.
  • DO NOT leave user unattended at any time.

3. PARTS GLOSSARY

FIG 2 PARTS GLOSSARY.JPG

FIG 3 PARTS GLOSSARY.JPG

4. ASSEMBLY

FIG 4 Tools Required.JPG

Carton Contains:

  1. Base Unit
  2. Right Upright Frame, M10 bolt and washer
  3. Left Upright Frame, M10 bolt and washer
  4. Knee pad with 2 x spacers, 2 x steel washers, 2 x nylon washers and 2 x M8 bolts
  5. Right Seat Pad
  6. Left Seat Pad
  7. 1 x Grab bar with 2 x Sling anchor bolts
  8. 2 x Seat Pad mounting hardware
    • 2 x Narrow retaining ring
    • 2 x Deep retaining ring (already mounted in each Upright Frame)
    • 2 x Mounting Plate
    • 6 x Mounting Bolts
    • 2 x Cover Caps

FIG 5 Carton Contains.JPG

FIG 6 ASSEMBLY.JPG

FIG 7 ASSEMBLY.JPG

FIG 8 ASSEMBLY.JPG

FIG 9 ASSEMBLY.JPG

FIG 10 ASSEMBLY.JPG

FIG 11 ASSEMBLY.JPG

5. PRELIMINARY CHECKS

After assembling the Star Mover, complete the following checks prior to first use.

  1. Check the seat pads are secure and rotate back and forth smoothly throughout the full range of movement.
  2. Check that locking castors engage and disengage.
  3. Check that the leg opening and closing mechanism works correctly using the Foot Pedal.
  4. If using a sling, ensure the clip attachment is securely mounted to the sling anchor bolts on the frame.

6. GENERAL OPERATION

Transferring to the Star Mover

FIG 12 Transferring to the Star Mover.JPG

FIG 13 Transferring to the Star Mover.JPG

FIG 14 Transferring to the Star Mover.JPG

FIG 15 Transferring From the Star Mover.JPG

FIG 16 Transferring From the Star Mover.JPG

FIG 17 Sling Attachment.JPG

FIG 18 Knee Pad Adjustment.JPG

7. TRANSFER GUIDE

Using the Star Mover with a Bed

The Star Mover can be used to aid a user getting out of bed.

FIG 19 Using the Star Mover with a Bed.JPG

FIG 20 Using the Star Mover with a Bed.JPG

FIG 21 Using the Star Mover with a Chair.JPG

FIG 22 Using the Star Mover with a Chair.JPG

8. SLING COMPATIBILITY

The Aspire Standing Sling range is designed specifically for safe user with the Star Patient Mover. Always consult with Aidacare if wanting to use a different brand of clip sling to ensure compatibility.

FIG 23 SLING COMPATIBILITY.jpg

DELUXE STANDING SLING

FIG 24 DELUXE STANDING SLING.JPG

9. CARE & MAINTENANCE

For trouble-free operation and prolonged working life of the Star Mover, we recommended periodic inspection by a person suitably qualified and acquainted with the design, use and care of patient movers.

If inspection reveals that the safety of the Star Mover has been jeopardized in any way, the unit should be IMMEDIATELY WITHDRAWN FROM SERVICE, tagged as such, and reported to the owner.

MAINTENANCE GUIDE & CHECK LIST

  • Check Castors are fastened securely to the frame
  • Castors should run and swivel smoothly – periodically remove any hair or other fibres that have built up within the Castor.
  • Check that the brakes on the rear Castors function normally.
  • Clean surfaces with a pH-neutral detergent – chlorine and phenol based detergents should not be used as these may damage patient mover components and surfaces.
  • All moving parts and pivot points should be routinely checked for fatigue and lubricated if necessary.
  • Sling attachment points should be checked for general wear and tear and replaced if necessary.

10. WARRANTY & SERVICE

AIDACARE warrants its products to be free from material- and workmanship- related defects under normal use and service. From the date of purchase, within the period stated below, we shall repair or replace without cost to the original customer, any part assembly or portion thereof which shall be returned to AIDACARE and deemed to be defective by our own inspection.

1. Star Mover Frame and all Parts – 2 years

WARRANTY IS CONTINGENT UPON ANNUAL SERVICING BEING COMPLETED FOR THE LENGTH OF THE WARRANTY PERIOD.

AIDACARE cannot be held responsible for any personal injury, damage to the Star Mover or damage to property resulting from improper or unsafe use of the product. No warranty claim shall apply where the product or any part thereof has been modified, varied, altered or damaged either accidentally or through improper or negligent use.

No warranty claim shall apply where the patient mover has been repaired or serviced by parties other than those accredited and/or approved by Aidacare.

ALL WARRANTIES MADE ARE BACK TO BASE WARRANTIES MEANING THAT ANY PARTS OR WHOLE STAR MOVERS REQUIRING REPAIR NEED TO BE RETURNED TO THE NEAREST AIDACARE SERVICE CENTRE AT THE CUSTOMER’S EXPENSE

– PLEASE CONTACT AIDACARE FOR A CURRENT LIST OF SERVICE CENTRE LOCATIONS.

SPARE PARTS
SPARE PARTS LIST AVAILABLE UPON REQUEST FROM
AIDACARE ON 1300 130 122 or www.aspirecare.com.au

11. TROUBLESHOOTING

WARNING: TROUBLESHOOTING SHOULD ONLY BE CARRIED OUT BY AIDACARE AUTHORISED SERVICE TECHNICIANS.

The following is a guideline only to general troubleshooting.

FIG 25 TROUBLESHOOTING.JPG

12. SPECIFICATIONS

FIG 26 SPECIFICATIONS.JPG

FIG 27 SPECIFICATIONS.JPG

*Range of motion dimensions are accurate to a +/-10mm tolerance

13. SERVICE LOG BOOK

FIG 28 SERVICE LOG BOOK.JPG

FIG 29 SERVICE LOG BOOK.JPG

FIG 30 SERVICE LOG BOOK.JPG

FIG 31 SERVICE LOG BOOK.JPG

FIG 32 SERVICE LOG BOOK.JPG

FIG 33 SERVICE LOG BOOK.JPG

AIDACARE PTY LTD
Building 3A, 1 Moorebank Avenue,
Moorebank NSW Australia 2170
1300 133 120 | www.aspirecare.com.au

Read More About This Manual & Download PDF:

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

aspire User Manuals

Related Manuals