Invacare IRC5PO2 Oxygen Concentrator Operation User Manual

June 5, 2024
INVACARE

Invacare IRC5PO2 Oxygen Concentrator Operation User Manual
Invacare IRC5PO2 Oxygen Concentrator Operation

WARNING

DO NOT USE THIS PRODUCT OR ANY AVAILABLE OPTIONAL EQUIPMENT WITHOUT FIRSTCOMPLETELY READING AND UNDERSTANDING THESE INSTRUCTIONS AND ANY ADDITIONAL INSTRUCTIONAL MATERIAL SUCH AS OWNER’S MANUALS, SERVICE MANUALS OR INSTRUCTION SHEETS SUPPLIED WITH THIS PRODUCT OR OPTIONAL EQUIPMENT. IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS OR  INSTRUCTIONS, CONTACT A HEALTHCAREPROFESSIONAL, DEALER OR TECHNICAL PERSONNEL BEFORE ATTEMPTING TO USE THIS EQUIPMENT – OTHERWISE, INJURY OR DAMAGE MAY OCCUR.

ACCESSORIES WARNING

Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories. Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products.

ACCESSORIES

There are many different types of humidifiers, oxygen tubing, cannulas and masks that can be used with this device. You should contact your local home care provider for recommendations on which of these devices will be best for you. They should also give you advice on the proper usage, maintenance, and cleaning.

The supply accessories (nasal cannula, mask, tubing, humidifier, etc.) used to deliver oxygen to the patient may include a means to reduce the propagation of fire in the accessories for the safety of the patient and others. If a commercially available, fire-activated flow stop device is used in the accessories setup, it should be placed as close to the patient as practicable.

NOTE: Updated versions of this manual can be found at www.invacare.com.

SPECIAL NOTES

Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for definitions of the signal words.

SIGNAL WORD MEANING
DANGER Danger indicates an imminently hazardous situation which, if not

avoided, will result in death or serious injury.
WARNING| Warning indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION| Caution indicates a potentially hazardous situation which, if not avoided, may result in property damage or minor injury or both.

NOTICE
The information contained in this document is subject to  change without notice.

DANGER
DO NOT SMOKE while using this device. Keep all matches, lighted cigarettes or other sources of ignition out of the room in which this product is located and away from where oxygen is being delivered.
NO SMOKING signs should be prominently displayed.
Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air. Failure to observe this warning can result in severe fire, property damage and cause physical injury or death.

CAUTION
“Caution: Federal law restricts this device to sale or rental by or on order of a physician, or any other practitioner licensed by the law of the State in which he/she practices to use or order the use of this device.”
Invacare recommends an alternate source of supplemental oxygen in the event of a power outage, alarm condition or mechanical failure. Consult your physician or equipment provider for the type of reserve system required.
This equipment is to be used as an oxygen supplement and is not considered life supporting or life sustaining.

LABEL LOCATION

LABEL LOCATION

GENERAL GUIDELINES

In order to ensure the safe installation, assembly and operation of the Perfecto2 concentrator these instructions MUST be followed.

WARNING
This section contains important information for the safe operation and use of this product.

To Reduce the Risk of Burns, Electrocution, Fire or Injury to Persons

DANGER
Risk of electric shock. DO NOT disassemble. Refer servicing to qualified service personnel. No user serviceable parts.
Avoid using while bathing. If continuous usage is required by the physician’s prescription, the concentrator MUST be located in another room at least 7 ft from the bath.
DO NOT come in contact with the concentrator while wet.
DO NOT place or store product where it can drop into water or other liquid.
DO NOT reach for product that has fallen into water.
Unplug IMMEDIATELY.
If the concentrator has a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or dropped into water, call qualified technician for examination and repair.
A spontaneous and violent ignition may occur if oil, grease or greasy substances come in contact with oxygen under pressure. These substances MUST be kept away from the oxygen concentrator, tubing and connections, and all other oxygen equipment. DO NOT use any lubricants unless recommended by Invacare.
Avoid creation of any spark near medical oxygen equipment. This includes sparks from static electricity created by any type of friction.

Operating Information

For optimum performance, Invacare recommends that each concentrator be on and running for a minimum of 30 minutes at a time. Shorter periods of operation may reduce maximum product life.

Keep the oxygen tubing, cord, and unit out from under such items as blankets, bed coverings, chair cushions, clothing and away from heated or hot surfaces, including space heaters, stoves and similar electrical appliances.
DO NOT move or relocate concentrator by pulling on the power cord.

NEVER drop or insert any object or liquid into any opening.
Invacare recommends that Crush-Proof oxygen tubing be used with this product and not exceed 50 ft in length.

There are no user serviceable parts. This does not include normal maintenance items. See maintenance section for user maintenance items.

A product should NEVER be left unattended when plugged in. Make sure the Perfecto2 is Off when not in use.

Close supervision is necessary when this product is used near children or physically-challenged individuals.

Additional monitoring or attention may be required for patients using this device who are unable to hear or see alarms or communicate discomfort.

DO NOT connect the concentrator in parallel or series with other oxygen concentrators or oxygen therapy devices.

Radio Frequency Interference

This equipment has been tested and found to comply with EMC limits specified by IEC/EN 60601-1-2. These limits are designed to provide a reasonable protection against electromagnetic interference in a typical medical installation.

Other devices may experience interference from even the low levels of electromagnetic emissions permitted by the above standards. To determine if the emissions from the Perfecto2 is causing the interference, turn the Perfecto2 Off. If the interference with the other device(s) stops, then the Perfecto2 is causing the interference. In such rare cases, interference may be reduced or corrected by one of the following measures:

  • Reposition, relocate, or increase the separation between the equipment.
  • Connect the equipment into an outlet on a circuit different from that to which the other device(s) is connected.

FEATURES

FEATURES

ACCESSORIES (NOT SHOWN):
Precise RX™ Pediatric Humidifier/Flowmeter
Accessory – IRCPF16 (Perfecto2 5 Only)
Home Fill home oxygen compressor – IOH200

REAR VIEW
REAR VIEW

NOTE: This outlet fitting is to be used only for filling oxygen cylinders with the HomeFill home oxygen compressor. The outlet fitting does not affect concentrator performance.
Refer to the HomeFill owner’s manual, part number 1100873, for connection and operating instructions. When not in use, the plug provided with the concentrator should be inserted into the outlet fitting. For more information about the HomeFill, contact your Invacare dealer.

HANDLING

The concentrator should ALWAYS be kept in the upright position to prevent cabinet damage while being transported.

If the concentrator is to be reshipped by common carrier, additional cartons are available from Invacare.

Unpacking

NOTE: For this procedure, refer to FIGURE 3.1.

  1. Check for any obvious damage to the carton or its contents. If damage is evident, notify the carrier, or your local dealer.
  2. Remove all loose packing from the carton.
  3. Carefully remove all the components from the carton.
    NOTE: Unless the oxygen concentrator is to be used immediately, retain containers and packing materials for storage until use of the concentrator is required.

Inspection

Inspect/examine exterior of the oxygen concentrator for nicks, dents, scratches or other damages. Inspect all components.

Storage
  1. Store the repackaged oxygen concentrator in a dry area.
  2. DO NOT place objects on top of repackaged concentrator.

TYPICAL PRODUCT PARAMETERS

Electrical Requirements: 120 VAC ± 10% (132 VAC/108 VAC), 60 Hz
Rated Current Input: 3 A
Sound Level: 43 ± 2 dBA Average for Perfecto2 Models and Perfecto2V Model 40

± 2 dBA Average for Perfecto2W Models
Altitude:| All Perfecto 2 Models: Up to 8,000 ft (2438 meters) above sea level without degredation of concentration levels.

All Perfecto 2 Models while filling a HomeFill System: Up to 6,000 ft (1828 meters) above sea level without degradation of concentration levels.

*Oxygen Output

Concentration Levels:

| 87% to 95.6% for flows ranging at 0.5 to 5 L/min
Maximum Outlet Pressure:| 5 psi ± 0.5 psi (34.5 kPa ± 3.45 kPa)
Flow Range:| 0.5 to 5 L/min

For flowrates less than 1 L/min, we recommend the use of the Invacare Pediatric Flowmeter IRCPF16 Accessory.

Potential Obstruction Alert:| 0 L/min to 0.5 L/min

The concentrator detects a condition that may indicate a potential obstruction of the output oxygen.

Rapid audible beeping alert (this alert is deactivated when accessories are connected). May be associated with flow setting of 0.5 L/min or less.

Average Power Consumption:| Perfecto2 and Perfecto2W Models: 280 W

(Operating @ 3 L/min) **** Perfecto2V Model: 325 W

Pressure Relief Mechanism Operational at:| 35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)
Change in maximum recommended flow :| 0.7 L/min (Back Pressure of 7kPa applied)
Filters:| Cabinet, Outlet HEPA and Compressor Inlet
Safety System:| Current overload or line surge shutdown. High temperature compressor shutdown. High Pressure Alarm w/compressor shutdown. Low Pressure Alarm w/compressor shutdown.

Battery Free Power Loss Alarm. SensO2

Oxygen System Possible Obstruction Alert.

Width:| 15 in ± 3/8 in (38.1 cm ± 1 cm)
Height:| 23 in ± 3/8 in (58.4 cm ± 1 cm)
Depth:| 12 in ± 3/8 in (30.5 cm ± 1 cm)
Weight:| 43 lbs ± 2 lbs (19.5 kg ± 1 kg) Perfecto2 Models 45 lbs ± 2 lbs (20.5 kg ± 1 kg) Perfecto2W Models 40 lbs ± 2 lbs (18.1 kg ± 1 kg) Perfecto2V Model
Shipping Weight:| 48 lbs ± 2 lbs (21.8 kg ± 1 kg) Perfecto2 Models 50 lbs ± 2 lbs (22.8 kg ± 1 kg) Perfecto2W Models 45 lbs ± 2 lbs (20.4 kg ± 1 kg) Perfecto2V Model
Operating Ambient Temperature:| 50°F – 95°F (10°C – 35°C) at 20-60% relative humidity Perfecto2 Models and Perfecto2V Model 50°F – 90°F (10°C – 32°C) at 20-60% relative humidity Perfecto2W Models
Exhaust:| Less than Ambient + 35°F (+19°C)
Oxygen Output Temperature:| Less than Ambient + 6°F (+3°C)
Cabinet:| Impact Resistant flame-retardant plastic cabinet that conforms to UL94V-0.
Regulatory Listing:| Double Insulated Product ETL certified complying with UL1431 and UL1097 ETL certified to CSA C22.2 No. 68.
Electrical:| No extension cords.
Placement:| No closer than twelve inches from any wall, furniture, draperies, or similar surfaces.
Tubing:| 7 ft cannula with a maximum 50 ft of

Crush-Proof Tubing (DO NOT pinch).

Relative Humidity:| 20 to 60%
Time of Operation:| Up to 24 hours per day.
Recommended Storage and Shipping Temperature:| -20°F to 150°F (-29°C to 65°C), at 15-95% relative humidity.

NOTE: Stated concentration levels achieved after initial warm-up period (approximately 30 minutes).

OPERATING INSTRUCTIONS

Introduction

Your oxygen concentrator is intended for individual use indoors. It is an electronically operated device that separates oxygen from room air. It provides high concentration of oxygen directly to you through a nasal cannula. Clinical studies have documented that oxygen concentrators are therapeutically equivalent to other types of oxygen delivery systems.

Your provider will show you how to use your oxygen concentrator. He/She should be contacted with any questions or problems regarding your oxygen concentrator. This owner’s manual will tell you about your concentrator and will serve as a reference as you use your concentrator.

Select a Location

WARNING
NEVER block the air openings of the product or place it on a soft surface, such as a bed or couch, where the air opening may be blocked. Keep the openings free from lint, hair and the like.
Move oxygen concentrator at least 12 inches away from walls, draperies or furniture.

You may select a room where using your oxygen concentrator would be most convenient. Your concentrator can be easily rolled from room to room on its casters.

Your oxygen concentrator will perform best when operated under the following conditions. Refer to Typical Product Parameters on page 12. Usage in environments other than those described may result in the need for increased equipment maintenance. The air intake of the unit should be located in a well ventilated area to avoid smoke, airborne pollutants and/or fumes. DO NOT place in a closet.

Set Up

Plug in Power Cord
Plug in power cord to an electrical outlet.

Connect Humidifier (If So Prescribed)

WARNING
DO NOT overfill humidifier.
DO NOT reverse the oxygen input and output connections.
Water from the humidifier bottle will travel through the cannula back to the patient.

NOTE: For this procedure, refer to FIGURE 5.1, FIGURE 5.2 on page 16, FIGURE 5.3 on page 17

  1. Remove cap from bottle.

  2. Fill humidifier with distilled water to the level indicated by the manufacturer. Replace the humidifier cap and securely tighten.
    Filling the Humidifier
    FIGURE 5.1 Filling the Humidifier

  3. Insert a flathead screwdriver in the plate groove on the top edge of the filter access door and gently pry the filter access door off (FIGURE 5.2).

  4. Pull up and remove the humidifier bottle adapter (FIGURE 5.2).
    Humidifier Bottle Adapter
    FIGURE 5.2 Humidifier Bottle Adapter

  5. Attach the humidifier bottle adapter to the humidifier bottle by turning the wing nut on the humidifier bottle counterclockwise until it is securely fastened. See Detail “A” in FIGURE 5.3.

  6. Place the humidifier bottle/adapter assembly in the humidifier compartment on the concentrator. See Detail “B” in FIGURE 5.3.

  7. Attach oxygen tubing from the humidifier bottle/adapter assembly to oxygen outlet connector on the concentrator. See Detail “B” in FIGURE 5.3.

  8. Attach the cannula/patient supply tubing to the humidifier bottle outlet. See Detail “B” in FIGURE 5.3.

  9. After assembly, ensure that oxygen is flowing through the cannula.
    Humidifier Compartment
    FIGURE 5.3 Humidifier Compartment

Power Switch

NOTE: For this procedure, refer to FIGURE 5.4.

  1. Press power switch to on position. All the panel lights and the audible alarm will come on for one second, indicating that the unit is functioning properly.

FIGURE 5. 4 Power Switch

Flowrate

NOTE: For this procedure, refer to FIGURE 5.5 on page 19.

  1. Turn the flowrate knob to the setting prescribed by your physician or therapist
    CAUTION
    Never set the flow greater than 5 L/min.
    WARNING
    DO NOT change the L/min setting on the flowmeter unless a change has been prescribed by your physician or therapist.
    NOTE: To properly read the flowmeter, locate the prescribed flowrate line on the flowmeter. Next, turn the flow knob until the ball rises to the line. Now, center the ball on the L/min line prescribed.
    NOTE: A Potential Obstruction Alert indicates a condition that may be associated with a partial or complete obstruction of oxygen output.

  2. If the flowrate on the flowmeter ever falls below 0.5 L/min more than about one minute, the Potential Obstruction
    Alert will be triggered. This is a rapid beeping of the audible alarm. Check your tubing or accessories for blocked or kinked tubing or a defective humidifier bottle.
    After rated flow is restored between 0.5 L/min and 0.75 L/min, the Potential Obstruction Alert will turn off.
    NOTE: The use of some accessories such as the PreciseRx™ pediatric flowmeter and the HomeFill compressor will deactivate the Potential Obstruction Alert.
    Flowrate
    FIGURE 5.5
    Flowrate

Front Panel Indicators – Units with SensO2

The SensO2 feature monitors the purity of oxygen generated by the oxygen concentrator. If purity falls below factory preset standards, indicator lights on the control panel will illuminate.
Front Panel Indicators
FIGURE 5.6 Front Panel Indicators – Units with SensO2

Initial Startup of the Concentrator

NOTE: Concentrator may be used during the initial start warm-up time (approximately 30 min.) while waiting for the O2 purity to reach maximum.

LABEL SYMBOL O 2 PURITY INDICATOR LIGHTS (LED)
O2 SYSTEM OKAY

O2 over 85%

| GREEN Indicator Light
LABEL SYMBOL| O2 Between 73% to 85%| YELLOW Indicator light

A.   YELLOW Solid

B.  YELLOW Flashing Sensor Failure

Call a qualified technician. Have a back-up supply of oxygen ready.

| SYSTEM FAILURE

O2 Below 73%

| RED Indicator Light

Continuous Audible Alarm

Sieve-GARD™ Compressor Shut- down. Call a qualified technician. Switch to a back-up supply of oxy- gen immediately.

Front Panel Indicators – Units without SensO2

NOTE: For this procedure, refer to FIGURE 5.7.
Front Panel Indicators
FIGURE 5.7 Front Panel Indicators – Units without SensO2

Initial Startup of the Concentrator

NOTE: Concentrator may be used during the initial start warm-up time (approximately 30 min). O2 purity will build to a maximum during this period.

LABEL SYMBOL STATUS INDICATOR LIGHTS (LED)
I/O SYSTEM OKAY GREEN Indicator Light

__

| SYSTEM FAILURE| RED Indicator light Continuous Audible Alarm Sieve-GARD™ Compressor Shutdown

Call a qualified technician. Switch to a back-up supply of oxygen immediately.

MAINTENANCE

WARNING
The Invacare concentrators are specifically designed to minimize routine preventive maintenance. Only qualified personnel should perform preventive maintenance on the concentrator.

Unplug the concentrator when cleaning. To avoid electrical shock, DO NOT remove cabinet.

NOTE: At a minimum, preventive maintenance MUST be performed according to the maintenance record guidelines. In places with high dust or soot levels, maintenance may need to be performed more often. Refer to Preventive Maintenance Record on page 24.

Routine Maintenance

Cleaning the Cabinet Filter

CAUTION
DO NOT operate the concentrator without the filter installed.
NOTE: For this procedure, refer to FIGURE 6.1 on page 23.
NOTE: There is one cabinet filter located on the back of the cabinet.

  1. Remove the filter and clean as needed.
    NOTE: Environmental conditions that may require more frequent cleaning of the filters include but are not limited to: high dust, air pollutants, etc.

  2. Clean the cabinet filter with a vacuum cleaner or wash in warm soapy water and rinse thoroughly.

  3. Dry the filter thoroughly before reinstallation.
    Cleaning the Cabinet Filter
    FIGURE 6.1 Cleaning the Cabinet Filter

Cleaning the Cabinet

Clean the cabinet with a mild household cleaner and non-abrasive cloth or sponge.

Cleaning the Humidifier

NOTE: To clean the oxygen humidifier, follow the instructions provided by the manufacturer. If none are provided, follow these STEPS:

  1. Clean the humidifier every day.
  2. Wash it in soapy water and rinse it with a solution of ten parts water and one part vinegar.
  3. Rinse thoroughly with hot water and refill with distilled water to the level shown on the humidifier.

PREVENTIVE MAINTENANCE RECORD

ON EACH INSPECTION|  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---
Record Date of Service|  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
Record Elapsed Hours On Hour Meter|  |  |  |  |  |  |  |  |  |  |  |  |  |  | |  |  |
Clean Cabinet Filter(s)|  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
Check Prescribed L/min. Flow Rate|  |  |  |  |  |  |  |  |  |  |  |  |  |  | |  |  |
DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS
UNITS WITHOUT SensO 2 **– EVERY 4,**|

380 HOURS

Check Oxygen Concentration|  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
Clean/Replace Cabinet Filter(s)|  |  |  |  |  |  |  |  |  |  |  |  |  |  |  | |  |
Check Outlet HEPA Filter|  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
Check Compressor Inlet Filter
|  |  |  |  |  |  |  |  |  |  |  |  |  |  |  | |  |
Check Power Loss Alarm|  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
UNITS WITH SensO 2 – EVERY 26,280| HOURS OR 3 YEARS, WHICHEVER COMES FIRST
Check Oxygen Concentration|  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
Clean/Replace Cabinet Filter(s)|  |  |  |  |  |  |  |  |  |  |  |  |  |  |  | |  |
Check Outlet HEPA Filter|  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
Check Compressor Inlet Filter
|  |  |  |  |  |  |  |  |  |  |  |  |  |  |  | |  |
Check Power Loss Alarm|  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |

FIGURE 6.2 Preventive Maintenance Record

NOTE: Refer to Preventive Maintenance section of Service Manual.
NOTE: 4,380 hours are equivalent to usage 24 hours per day, 7 days per week, for 6 months. 26,280 hours are equivalent to usage 24 hours per day, 7 days per week, for 3 years.

TROUBLESHOOTING GUIDE

SYMPTOM : PROBABLE CAUSE: SOLUTION:

Alarm:

Short beeps, long pause

Concentrator not operating,

power switch On. Beep……Beep……

| Main Power Loss:

1. Power cord not plugged in.

| 1. Insert plug into outlet.
2. No power at outlet.| 2. Inspect house circuit breakers or fuses. If problem recurs, use a different outlet.
3. Tripped circuit breaker.| 3. Call Service Provider.
Alarm:

Continuous

Concentrator not operating, Power Switch On.

Beeeeeeeep….

| System Failure:

1. Unit overheating due

to blocked air intake.

| 1a. Remove and clean cabinet filters.
1b. Move oxygen concentrator at least twelve inches away from walls, draperies or furniture.
2. Insufficient power at outlet.| 2. DO NOT use extension cords. Move to another electrical outlet or circuit.
3. Internal repairs required.| 3. Call Service Provider.
Alarm:

Rapid

Beep….Beep… Beep…Beep

| Potential Obstruction Alert:

1. Possible internal obstruction in the oxygen path.Kinked or blocked tubing, cannula or humidifier.

| 1. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power Off for 60 seconds and then turn power back ON.
 | 2. Flowmeter set at 0.5L/min or less.| 2. Reset flowmeter to

prescribed flowrate. If prescribed flowrate is below

0.5 L/min, use the Pediatric Flowmeter IRCPF16 (Only with Perfecto2 5).

NOTE: The use of the pediatric flowmeter will deactivate the Potential Obstruction Alert.

---|---|---
YELLOW or RED light Illuminated.

Alarm: Continuous On RED light only.

  • Only applies to IRC5PO2.

| 1. Low oxygen purity.| 1. Clean or Replace filters.
2. Kinked or blocked tubing, cannula or humidifier.
| 2. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power Off for 60 seconds and then turn power back On.
3. Flowmeter set at 0.5 L/min or less.*| 3. Check flowmeter is set to

1.0 L/min or more.

4. Unit overheating due to blocked air intake.| 4a. Remove and clean cabinet filters.
4b. Move concentrator at least twelve inches from walls, draperies, and furniture.
5. Insufficient power at outlet.| 5. DO NOT use extension cords. Move to another electrical outlet or circuit.
6. Internal repairs required.| 6. Call Service Provider.
GREEN light with YELLOW Light Flashing.| 1. Internal repairs required.| 1. Call Service Provider.

LIMITED WARRANTY

NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975.

This warranty is extended only to the original purchaser who purchases this product when new and unused from Invacare Corporation or a dealer. This warranty is not extended to any other person or entity and it is not transferable or assignable to any subsequent purchaser or owner.

Coverage under this warranty will end upon any such subsequent sale or other transfer of title to any other person. This warranty gives you specific legal rights and you may also have other legal rights which may vary from state to state.

Invacare Corporation warrants its Perfecto2 Oxygen Concentrators when purchased new and unused to be free from defects in materials and workmanship for a period of three years from date of purchase from Invacare or a dealer, with a copy of the seller’s invoice required for coverage under this warranty.

DO NOT OPEN OR ATTEMPT TO SERVICE THE 4-WAY VALVE BODY. This will void any and all warranty applicable to the valve body. If within such warranty periods any such product shall be proven to Invacare Corporation’s satisfaction to be defective, such product shall be repaired or replaced, at Invacare Corporation’s option. This warranty only applies to the labor for repairs performed by the Invacare Service Department or Invacare Authorized Service Centers. It does not apply to the labor performed by the purchaser or user. This warranty does not include normal wear and tear or shipping charges. Invacare Corporation’s sole obligation and your exclusive remedy under this warranty shall be limited to such repair or replacement. Routine maintenance items, such as filters, are excluded from this warranty.

For warranty service, please contact Invacare Corporation’s Service Department at the toll free number on the back cover. Upon receiving notice of an alleged defect in a product, Invacare Corporation will issue a serialized return authorization. It shall then be the responsibility of the purchaser to return the entire unit or remove, at purchaser’s cost, the defective component part(s) identified, pack the component part(s) in a manner to avoid shipping damage and to ship the component part(s) to either Invacare Corporation’s plant or service center as specified by Invacare Corporation in advance. Defective component part(s) MUST be returned for warranty inspection using the serial number as identification within thirty days of return authorization date. DO NOT return products to our factory without prior consent. C.O.D. shipments will be refused; please prepay shipping charges.

LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY SHALL NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE, SOOT OR SMOKE-FILLED ENVIRONMENTS, OR OTHER THAN NORMAL APPLICATION, USE OR SERVICE, OR TO PRODUCTS MODIFIED WITHOUT INVACARE CORPORATION’S EXPRESS WRITTEN CONSENT (INCLUDING, BUT NOT LIMITED TO, MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS) OR TO PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE CORPORATION OR TO PRODUCTS DAMAGED BY CIRCUMSTANCES BEYOND INVACARE CORPORATION’S CONTROL.

THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER WARRANTIES WHATSOEVER, WHETHER EXPRESS OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER, SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN. THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN. INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER.

SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGE, OR LIMITATION OF HOW LONG AN IMPLIED WARRANTY LASTS, SO THE ABOVE EXCLUSION AND LIMITATION MAY NOT APPLY TO YOU.

THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE/PROVINCIAL LAWS AND REQUIREMENTS.

Invacare Corporation
www.invacare.com
USA One Invacare Way Elyria, Ohio USA
44036-2125
800-333-6900

Technical Services
800-832-4707

© 2013 Invacare Corporation. All rights reserved. Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. Trademarks are identified by ™ and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.