deltaohm HD52 Ultrasonic anemometers Instruction Manual

June 5, 2024
DeltaOHM

deltaohm HD52 Ultrasonic anemometers

deltaohm-HD52-Ultrasonic-anemometers-PRODUCT

INTRODUCTION

The instruments of the series HD52.3D… are 2-axis ultrasonic static anemometers. The available measurement options combine in a single instrument the main quantities of meteorological interest, making the instrument like a compact and light meteorological station.
Measured quantities:

  • Wind speed and direction, U-V Cartesian components of wind speed
  • Wind Gust
  • Relative Humidity and temperature (optional)
  • Global solar radiation (optional)
  • Barometric pressure (optional)
  • Rainfall (optional)

The “Rainfall” and “Global solar radiation” options are alternatives (they cannot be both pre-sent in the same instrument).
The average of wind speed and direction over a period configurable up to 10 minutes is calcu-lated.
All models are equipped with a magnetic compass.
Wind speed and direction are determined by measuring the transit time of ultrasonic pulses between two pairs of ultrasonic transducers.
RS232, RS485, RS422 and SDI-12 serial interfaces are available with NMEA, MODBUS-RTU and SDI-12 communication protocols.
All versions have two analog outputs, for wind speed and direction, which are factory-configurable within 4÷20 mA (standard), 0÷1 V, 0÷5 V or 0÷10 V (to be specified when or-dering).
The heater option prevents the accumulation of snow and ice formation, allowing precise measurements in all environmental conditions.
Mounting on ∅ 40 mm mast. The electric connection is performed through a M23 19-pole con-nector situated on the bottom of the instrument.
The low power consumption of the instrument allows installation in remote sites, with power supplied by photovoltaic panel and backup battery.
The absence of moving parts minimizes instrument maintenance.

MODELS

The table below shows the optional features included in the various models of the series.

| OPTIONS INCLUDED
---|---

Model

|

Atmospheric pressure

| Relative humidity

+

Temperature

|

Global solar radiation

|

Rainfall

|

Heating

HD52.3D [R]| | | | |

With option R

in the code

HD52.3D4 [R]| | | |
HD52.3DP [R]| | | |
HD52.3DP4 [R]| | | |
HD52.3D17 [R]| | | |
HD52.3D147 [R]| | | |
HD52.3DP17 [R]| | | |
HD52.3DP147 [R]| | | |
HD52.3DT147| | | | | Not available

TECHNICAL SPECIFICATIONS

Wind speed

  • Sensor;Ultrasounds
  • Measuring range;0…60 m/s (0…50 m/s with rain gauge option)
  • Resolution ; 0.01 m/s
  • Accuracy;± 0.2 m/s or ± 2%, the greatest (0…35 m/s), ± 3% (> 35 m/s)

Wind direction

  • Sensor ;Ultrasounds
  • Measuring range;0…359.9°
  • Resolution ;;0.1°
  • Accuracy ;± 2° RMSE from 1.0 m/s

Compass

  • Sensor;Magnetic
  • Measuring range;0…360°
  • Resolution;0.1°
  • Accuracy;± 1°

Air temperature ( needs option 17 )

  • Sensor; Pt100
  • Measuring range;-40…+70 °C
  • Resolution;0.1 °C
  • Accuracy;± 0.15 °C ± 0.1% of measurement

Relative humidity ( needs option 17 )

  • Sensor; Capacitive
  • Measuring range;0…100%RH
  • Resolution;0.1%
  • Accuracy (@ T = 15…35 °C);± 1.5%RH (0…90%RH), ± 2%RH (remaining range)
  • Accuracy (@ T = -40…+70 °C) ;± (1.5 + 1.5% of measurement )%RH

Barometric Pressure ( needs option 4 )

  • Sensor ;Piezoresistive
  • Measuring range;300…1100 hPa
  • Resolution;0.1 hPa
  • Accuracy;± 0.5 hPa @ 20 °C

Solar radiation ( needs option P )

  • Sensor ; Thermopile
  • Measuring range;0…2000 W/m2
  • Resolution;1 W/m2
  • Accuracy;2nd Class Pyranometer

Rainfall ( needs option T )

  • Sensor; Tipping bucket
  • Resolution;0.2 mm
  • Accuracy;99% up to 120 mm/h
  • Maximum rainfall rate;2000 mm/h
  • Collector area;127 cm2

General features

  • Power Supply;10…30 Vdc
  • Power consumption;26 mA @ 24 Vdc without heater 8 W @ 24 Vdc with heater
  • Serial outputs; RS232, RS485 (¼ Unit Load), RS422 and SDI-12
  • Communication protocol; NMEA, MODBUS-RTU, SDI-12, proprietary RS232, and RS485
  • Analog outputs;2 analog outputs, for wind speed and direction.
  • Output at choice among 4…20 mA (standard), 0…1 V, 0…5 V, and 0…10 V (option 0…10 V needs 15…30 Vdc power supply)
  • Wind speed averaging interval; Configurable from 1 s to 10 min
  • Electrical connection;19-pole M23 male connector
  • Operating temperature; -40…+70 °C

Minimum temperature for the rainfall sensor 1 °C

  • Protection degree; IP 66
  • Survival speed;90 m/s (60 m/s with rain gauge option)
  • Weight ;About 1 kg (version HD52.3DP147) About 1.5 kg (version HD52.3DT147)
  • Case; Plastic material. Metal parts: AISI 316

Dimensions (mm)

deltaohm-HD52-Ultrasonic-
anemometers-2 deltaohm-HD52-Ultrasonic-
anemometers-3DESCRIPTION deltaohm-HD52-Ultrasonic-
anemometers-4

  1. Pyranometer
  2. Ultrasonic sensors for the measurement of wind speed and direction
  3. Protective shield from solar radiation for relative humidity and temperature sensors
  4. Fixing clamp to Ø 40 mm mast

Note: in the models measuring barometric pressure, the pressure sensor is inside the instrument.

  1. Rain gauge
  2. Ultrasonic sensors for the measurement of wind speed and direction
  3. Protective shield from solar radiation for relative humidity and temperature sensors
  4. Fixing clamp to Ø 40 mm mast

MEASURING THE PRINCIPLE OF WIND SPEED AND DIRECTION

Wind speed and direction are determined by measuring the time taken by ultrasonic pulses to cover the distance from the transducer that generates the pulse to the receiving transducer.
The instrument uses 2 pairs of transducers oriented along two orthogonal axes. Detecting the wind speed along two axes allows determining not only the intensity but also the wind direc-tion.
The instrument measures the travel time of the ultrasonic pulse between the two transducers of the same pair in both directions. The travel times in the two opposed directions are defined as tA (forward direction time) and tR (reverse direction time).
If wind speed is zero, tA and tR values are the same. In the presence of wind, one of the two-time values is greater than the other and the comparison between the two time values allows to determine the direction and the intensity of the wind.
Measuring the travel time in both directions allows to cancel the dependence of the transmission speed of ultrasounds in the air from the environmental conditions of temperature, humidity and barometric pressure. deltaohm-HD52
-Ultrasonic-anemometers-6 Where:
D = Distance between the two transducers of the same pair
C = Sound speed
VW = Component of wind speed along the measurement axis
Measuring the two travel times allows determining the wind speed component

INSTALLATION

To install the instrument, pass the connection cable inside the support mast and connect the 19-pole M23 female connector of the cable to the 19-pole M23 male connector situated at the bottom of the instrument. Ensure connection stability by tightening the connector external nut.
Align the instrument to the desired direction (see par. 5.1 “Alignment”), then fix it on the sup-port mast by tightening the cable tie at the bottom of the instrument. The support mast, having 40 mm maximum outer diameter and 36 mm minimum inner diame-ter, should be positioned on a stable surface.
The instrument should be installed vertically and in an open area, far from obstructions located in the vicinity that might alter the natural air flow. Any close objects (such as buildings, trees, pylons, etc.) should be at a distance equal to at least ten times their height. deltaohm-HD52-Ultrasonic-
anemometers-9In the presence of close objects, it is advisable to place the instrument at a height of 10 m (except version with rain gauge option). For version with rain gauge option it is recommended to place the instrument at a height not exceeding 3 m (the precipitation measurement is standardized at ground level, as it is very af-fected by the presence of wind). It is very important to place the instrument on a rigid mast, as the oscillation of the mast could cause false detections of the tipping bucket of the rain gauge.
For open-space installations, the instrument can be installed using the tripod HD2005.20 (2 m) or HD2005.20.1 (3 m).
If the instrument is installed on a building, the height of the instrument should be at least 1.5 times the minimum value between the height of the building and the roof longest diagonal.
Mount the instrument far from magnetic materials and devices that generate magnetic fields (electric motors, electric power cables, electric transformers, radars, radio transmitters etc.) in order to prevent faulty indications from the magnetic compass.
In case of mobile installations (for example on a boat), take into account that the instrument measures the relative (apparent) wind speed with respect to the instrument. To determine the absolute (real) wind speed one should consider the movement velocity of the instrument.
All instrument sensors are factory-calibrated and do not require additional interventions of the user.
Unless otherwise requested, with factory settings the instrument starts in configuration mode at power up and stands in waiting to receive the commands for the setting of the operating pa-rameters through a RS232 serial connection. To know the available configuration parameters, the related factory settings, the commands to modify the parameters and to select the operat-ing mode, see Chapter 6 “CONFIGURATION”. If, however, the instrument is already set to op-erate in one of the available operating modes (SDI-12, NMEA, MODBUS-RTU, proprietary RS232, proprietary RS485), the set mode will be active 10 seconds after power up.

ALIGNMENT OF THE INSTRUMENT

The instrument is equipped with a magnetic compass, and wind speed and direction measurements are automatically compensated and referred to as magnetic North, even if alignment to North is not performed. This allows for obtaining accurate measurements even in case of mobile installations.
It is possible to disable the compass compensation of the wind speed and direction measurements. In this case it is necessary to align the instrument during installation. The arrows on the case will facilitate the alignment. For an accurate alignment, connect the instrument to a PC (see following chapters for communication protocols), then rotate the instrument on its vertical axis until the compass measures 0.0° ± 0.1°.
When evaluating the wind direction, one should take into account that the geographic North differs from the magnetic North indicated by the compass. The difference, named magnetic declination, depends on the area where the instrument was installed (for example, about 15° in North America and less than 3° in Europe).
If wind speed and direction values are given in polar coordinates, 0° angle corresponds to a wind coming from North.

ELECTRICAL CONNECTIONS

All connections are performed through a 19-pole M23 male connector situated at the bottom of the instrument. The figure and the table below show the numbers and function of the connector contacts and the corresponding wires of the optional CP52.x cable:deltaohm-HD52-Ultrasonic-
anemometers-12

TAB. 5.A – Electrical connections

Connector pin N o| CP52.x cable wire N o /color| Symbol| Description
---|---|---|---
1| | | Not used
2| 2 / Blue| RX SDI| Data line for SDI-12 connection
3| 3 / Yellow| RX +| Serial receive (input) positive
4| | HEAT –| Heater power supply negative
5| | HEAT +| Heater power supply positive (10…30 Vdc)
6| 6 / Pink| HEAT –| Heater power supply negative
7| 7 / Violet| HEAT +| Heater power supply positive (10…30 Vdc)
8| 8 / Grey| GND| Serial ground
9| 9 / White| TX –| Serial transmission (output) negative “DATA –” RS485 output
10| | | Not used
11| | | Not used
12| 12 / Black| V –| Instrument power supply negative
13| 4 / Grey-Pink| RX –| Serial receive (input) negative
14| 10 / Brown| OUT 1| Analog output 1 positive
15| | GND| Analog ground (isolated from V –)
16| 11 / Green| OUT 2| Analog output 2 positive
17| 5 / Red-Blue| TX +| Serial transmission (output) positive “DATA +” RS485 output
18| | V +| Instrument power supply positive (10…30 Vdc)
19| 1 / Red| V +| Instrument power supply positive (10…30 Vdc)
—| Black (thick wire)| SHIELD| Cable shield

ATTENTION: The optional 12-pole CP52.x cable wire number does not always coin- cide with the M23 connector pin number.
RS232 SERIAL CONNECTIONdeltaohm-HD52-Ultrasonic-
anemometers-13

For RS232 connection, TX-, RX+ and serial GND signals (pin 9, 3 and 8 of M23 connector) are used, to be connected respectively to RX, TX and GND signals of RS232 port on PC (pin 2, 3 and 5 of 9-pole SubD connector). The length of RS232 cables should not exceed 15 m.
With RS232 connection, NMEA, MODBUS-RTU, and proprietary RS232 protocols can be used.
If the PC is not equipped with RS232 serial ports, the RS52 adapting cable (with built-in USB/RS232 converter) can be inserted between the PC and the instrument (see paragraph 5.2.6).
RS485 SERIAL CONNECTIONdeltaohm-HD52-Ultrasonic-
anemometers-14 Thanks to RS485 connection, multiple instruments can be connected to a multi-point network. Instruments are connected in sequence through a twisted-pair shielded cable for signals and a third wire for ground.
Line terminations should be placed at the two ends of the network. The cable shield should be connected to both ends of the line.
The maximum number of devices that can be connected to the RS485 line (Bus) depends on the load characteristics of the devices to be connected. The RS485 standard requires that the total load does not exceed 32 unit loads. The load of an HD52.3D… anemometer is equal to ¼ of unit load. If the total load is greater than 32 unit loads, divide the network into segments and add a signal repeater between a segment and the successive one. Line termination should be applied at both ends of each segment.
The maximum length of the cable depends on the transmission speed and the cable character-istics. Typically, the maximum length is 1200 m. The data line should be kept separated from any power lines in order to prevent interference with the transmitted signal.
Each instrument on the network is univocally identified by an address. Multiple transmitters with the same address should not be placed on the network.
If the instrument is connected to a data logger, the possibility to connect multiple sensors to the network depends on the data logger capacity to manage multiple sensors.
NMEA, MODBUS-RTU and proprietary RS485 protocols can be used with RS485 connection.
Before connecting the instrument to the network, configure address and Baud Rate (see chap-ter 6 “CONFIGURATION”).
RS422 SERIAL CONNECTIONdeltaohm-HD52-Ultrasonic-
anemometers-15

RS422 standard is used for point-to-point connection on long distances. The instrument is con-nected to a PC/data logger through a shielded cable with two twisted pairs for signals and an additional wire for ground. The extremities of the connections should end with line terminators.
The maximum length of the cable depends on the transmission speed and the cable character-istics. Typically, the maximum length is 1200 m. The data lines should be kept separated from any power lines in order to prevent interference with the transmitted signal.
NMEA, MODBUS-RTU and proprietary RS485 protocols can be used with RS422 connection.
Before connecting the instrument to the network, configure the address and Baud Rate (see chap-ter 6 “CONFIGURATION”).
SDI-12 SERIAL CONNECTIONdeltaohm-HD52-Ultrasonic-
anemometers-16

SDI-12 standard has three connection wires: power supply +12V, data line and ground.
Up to 10 sensors can be connected in parallel, each of them identified by its own address. Communication between sensors and PC/datalogger is performed at 1200 baud. Connection cables should not exceed 60 m.
The instrument should be configured to operate in SDI-12 mode. Set the address (see chapter 6 “CONFIGURATION”) before connecting the instrument to the network.
ANALOG OUTPUTS CONNECTION
Connect the outputs according to one of the two schemes below, depending on the output type, current (standard) or voltage (on request), available in the instrument. OUT 1 and OUT 2 are associated to wind speed and direction respectively. To change the type of speed and direction measurements associated to the outputs, see chapter 6 “CONFIGURATION”.deltaohm-HD52
-Ultrasonic-anemometers-17 CONNECTION OF RS52 CABLE
If the PC is not equipped with RS232 serial ports, but with USB ports only, the RS52 adapting cable (with built-in USB/RS232 converter) should be interposed between the PC and the instrument. deltaohm-HD52-Ultrasonic-
anemometers-18To use the RS52 cable, the drivers included in the HD52.3D-S software package should be installed in the PC (see the guide to USB drivers installation included in the software package). The instrument is powered directly from the USB port of the PC.
CONNECTION OF HEATING SYSTEM
Models with heat option (R option) have an integrated device that heats sonic transducers in order to prevent ice forming and ensure correct operation even in the presence of snow.
The heater power supply is separated from the instrument’s main power supply .deltaohm-HD52-Ultrasonic-anemometers-19Heating needs 10…30 Vdc power supply and 6 W power. The heat circuit is switched on below +4 °C. After switching on, the heating is switched off when the temperature exceeds +8 °C.

CONFIGURATION

Configuration mode allows to read the instrument’s general info (firmware version, calibration date, serial number), to set the operation mode, to read and set the instrument operation pa-rameters.
To configure the instrument, a RS232 serial connection to a PC should be performed (see paragraph 5.2.1). If your PC is not equipped with RS232 serial ports, the RS52 adapting cable can be placed between the PC and the instrument (see paragraph 5.2.6).
Communication parameters should be set in the PC as follows

  • Baud rate: 115200
  • Data Bits: 8
  • Parity: None
  • Stop Bit: 2

The instrument configuration can be realized with the aid of the HD52.3D-S application soft-ware (see the instructions of the software) or by sending serial commands via standard com-munication programs.
The configuration procedure by sending serial commands slightly differs depending on whether the instrument is set in configuration mode (factory setting, unless otherwise requested), or whether it is set in one of the available operation modes (SDI-12, NMEA, MODBUS-RTU, pro-prietary RS232, proprietary RS485).
INSTRUMENT SET IN CONFIGURATION MODE (factory setting)
If the instrument is set in configuration mode, at power-up it remains waiting to receive the configuration serial commands described in the following tables. To configure and make the in-strument operational, the following should be performed:

  • Send command for operation mode selection (see table Operation Mode described at paragraph 6.1 “Serial Commands”).
  • Send commands to set the parameters of the selected operation mode (see tables de-scribed at paragraph 6.1 “Serial Commands”).
  • Send commands to set operational general parameters (parameters independent from operation mode)
  • Turn the instrument off and on. The selected operation mode will become active after 10 seconds from restart.
    Note: configuration commands can be sent in any order; it is not necessary to respect the indicated sequence.

INSTRUMENT SET IN ONE OF THE AVAILABLE OPERATION MODES
If the instrument is set in one of the available operation modes, when started up it remains waiting for a serial command for 10 seconds. For instrument configuration, the following should be performed:

  • Send the following command to the instrument before 10 seconds have elapsed:
    @
    with = ASCII character Carriage Return.
    If the instrument doesn’t receive the over-mentioned command after 10 seconds from start up, the operation mode set in the instrument is immediately activated.

  • Send commands to change the desired parameters (see tables described at paragraph 6.1 “Serial Commands”).

  • Turn the instrument off and on. The selected operation mode will become active after 10 seconds from restart with the new parameter values.

SERIAL COMMANDS

The tables below describe the serial commands that allow to read the current instrument con-figuration and change the operation parameters.
Measuring units:

Command Reply Description
CGUVn & Sets measuring unit of wind speed:

§  m/s if n=1

§  cm/s if n=2

§  km/h if n=3

§  knot if n=4

§  mph if n=5

Default : m/s (n=1)

RGUV| n|| Reads the wind speed measuring unit set in the instrument
CGUTn| &|| Sets temperature measuring unit:

§  °C if n=1

§  °F if n=2

Default : °C (n=1)

RGUT| n|| Reads the temperature measuring unit set in the instrument
CGUPn| &|| Sets pressure measuring unit:

§  mbar if n=1      [Note:1 mbar=1 hPa]

§  mmHg if n=2

§  inchHg if n=3

§  mmH2O if n=4

§  inchH2O if n=5

§  atm if n=6

Default : mbar (n=1)

RGUP| n|| Reads the pressure measuring unit set in the instrument
CGURn| &|| Sets rainfall measuring unit:

§  mm if n=1

§  inch if n=2

Default : mm (n=1)

RGUR| n|| Reads the Rainfall measuring unit set in the instrument

Operation Mode:

Command Reply Description
CUMn & Sets instrument in mode:

§  Configuration if n=0

§  proprietary RS485 if n=1

§  proprietary RS232 if n=2

§  SDI-12 if n=3

§  NMEA if n=4

§  MODBUS-RTU if n=5

Default : Configuration (n=0)

RUM| & n|| Reads mode set in the instrument

Note 1: after sending the command for the selection of the operation mode, the instrument remains in configuration mode. The selected mode will become active at the following restart of the instrument.
Parameters for proprietary RS232 and RS485 modes:

Command Reply Description
CU1Ac & Sets the address for proprietary RS485 mode to c value

The address is an alphanumeric character ranging within 0…9, a…z, A…Z

Default : 0

RU1A| & c|| Reads the address for proprietary RS485 mode set in the instrument
CU1Bn| &|| Sets Baud Rate for proprietary RS485 mode to:

§  9600 if n=3

§  19200 if n=4

§  38400 if n=5

§  57600 if n=6

§  115200 if n=7

Default : 115200 (n=7)

RU1B| & n|| Reads Baud Rate setting for proprietary RS485 mode
CU2Bn| &|| Sets Baud Rate for proprietary RS232 mode to:

§  9600 if n=3

§  19200 if n=4

§  38400 if n=5

§         57600 if n=6

Default : 57600 (n=6)

RU2B| & n|| Reads Baud Rate setting for proprietary RS232 mode
CU1Dccccccccccc| &|| Sets measurements order in the string sent in proprietary RS232 and RS485 modes

In the sequence ccccccccccc (max. 11 characters) each character identifies a measurement according to the following correspondence:

0  Þ Barometric Pressure

1  Þ Temperature (sensor Pt100) 2 Þ Relative Humidity

3 Þ Pyranometer

6 Þ Wind speed (coordinates U,V) 7 Þ Wind speed (intensity)

8 Þ Wind Direction (Azimuth) T Þ Sonic Temperature

C Þ Compass E Þ Errors

note: the Wind Gust measurement and the rainfall measurement are not available in proprietary RS232 and RS485 modes

Default : 78

(see Note 2 )

RU1D| & ccccccccccc|| Reads measurements order in the string sent in proprietary RS232 and RS485 modes
Command| Reply| Description
---|---|---
CU2Rnnnn| &|| Sets transmission interval of string with measurements in proprietary RS232 mode to nnnn seconds

The interval should range within 1 and 3600 seconds

Default : 1 second

RU2R| & nnnn|| Reads transmission interval of string with measurements in proprietary RS232 mode

NOTE 2 : MEASUREMENTS ORDER
In the measurement string sent by the instrument (constantly in proprietary RS232 mode, on request in proprietary RS485 mode), measurements can be set in arbitrary order: one should simply indicate the desired order in the sequence of characters “ccc ccc ccc” sent by the command C1UD. The sequence of characters “ccc ccc ccc” can have a variable length up to a maximum of 11 characters.
Example: if the sequence of characters is set to 78012, wind speed, wind direction, barometric pressure, temperature and relative humidity measurements will appear from left to right in the data string sent by the instrument.
If the information on error conditions is requested (E character), three numbers with the following meaning will appear in the data string sent by the instrument:

  1. First number = error code identifying the transducers and the type of error.
    The number is composed by two digits. The first digit indicates the path (that is the pair of the transducers) that presents the error, according to the numeration indicated in the figure:deltaohm-HD52-Ultrasonic-
anemometers-20Digit 7 indicates an error in the compass. Digit 0 indicates that no errors are present in the ultrasonic transducers or in the compass.
    The second digit of the error code indicates the type of error: 0 = no errors; 1 = transducer broken, electric interruption, path obstruction; Other = codes re-served to technical service.

  2. Second number = state of activation of the ultrasonic transducers heating.
    0 = heating turned off, 1 = heating turned on

  3. Third number = number of invalid measurements.
    For example: if 21 0 2 appears in correspondence to the error condition in the data string sent by the instrument, it means that an error occurred (broken transducer or path obstruction) in the path number 2, that the heating is turned off and that two measurements have been rejected due to the error occurrence.

Parameters for NMEA mode:
Command Reply Description
CU4Bn & Sets Baud Rate for NMEA mode to:

§ 2400 if n=1

§ 4800 if n=2

§ 9600 if n=3

§ 19200 if n=4

§ 38400 if n=5

§ 57600 if n=6

§ 115200 if n=7

Default : 4800 (n=2)

RU4B| & n|| Reads Baud Rate setting for NMEA mode
CU4In| &|| Sets interface for NMEA mode to:

§ RS232 if n=0

§ RS485 if n=1

§ RS422 if n=2

Default : RS485 (n=1)

RU4I| & n|| Reads interface setting for NMEA mode
CU4Mn| &|| Sets parity and stop bits for NMEA mode to:

§  8N1 if n=0 [No parity, 1 stop bit]

§  8N2 if n=1 [No parity, 2 stop bits]

§  8E1 if n=2 [Even parity, 1 stop bit]

§  8E2 if n=3 [Even parity, 2 stop bits]

§  8O1 if n=4 [Odd parity, 1 stop bit]

§  8O2 if n=5 [Odd parity, 2 stop bits] The number of data bits is fixed to 8

Default : 8N1 (n=0)

RU4M| & n|| Reads current setting of parity and stop bits for NMEA mode
CU4Rnnn| &|| Sets transmission interval of a string with measurements in NMEA mode to nnn seconds

The interval should range between 1 and 255 seconds

Default : 1 second

RU4R| & nnn|| Reads setting of the transmission interval of string with measurements in NMEA mode

Parameters for MODBUS-RTU mode:

Command Reply Description
CU5Annn & Sets MODBUS address to nnn

The address should range between 1 and 247

Default : 1

RU5A| & nnn|| Reads the MODBUS address setting
CU5Bn| &|| Sets the Baud Rate for MODBUS mode to:

§ 9600 if n=3

§ 19200 if n=4

§ 38400 if n=5 (from firmware vers. 2.21)

§ 57600 if n=6 (from firmware vers. 2.21)

§ 115200 if n=7 (from firmware vers. 2.21)

Default : 19200 (n=4)

RU5B| & n|| Reads Baud Rate setting for MODBUS mode
CU5In| &|| Sets interface for MODBUS mode to:

§ RS232 if n=0

§ RS485 if n=1

§ RS422 if n=2

Default : RS485 (n=1)

Note: with RS232 option you can connect to PC or datalogger 1 instrument only; option useful to do tests without RS232/RS485 conversion.

RU5I| & n|| Reads interface setting for MODBUS mode
CU5Mn| &|| Sets parity and stop bits for MODBUS mode to:

§ 8N1 if n=0 [No parity, 1 stop bit]

§ 8N2 if n=1 [No parity, 2 stop bits]

§ 8E1 if n=2 [Even parity, 1 stop bit]

§ 8E2 if n=3 [Even parity, 2 stop bits]

§ 8O1 if n=4 [Odd parity, 1 stop bit]

§ 8O2 if n=5 [Odd parity, 2 stop bits] The number of bits is fixed to 8

Default : 8E1 (n=2)

RU5M| & n|| Reads the setting of parity and stop bits for MODBUS mode
CU5Wn| &|| Sets waiting time after transmission in MODBUS mode to:

§ Immediate reception if n=0 (violates protocol)

§ Waiting 3.5 characters if n=1 (respects protocol)

Default : Waiting 3.5 characters (n=1)

RU5W| & n|| Reads the setting of waiting time after transmission in MODBUS mode

Parameters for SDI-12 mode:

Command Reply Description
CU3Ac & Sets the SDI-12 address to c value

The address is an alphanumeric character ranging within 0…9, a…z, A…Z. Default : 0

RU3A| & c|| Reads the SDI-12 address set in the instrument

General parameters:

Command Reply Description
CGHn & Enables/disables heating:

– Disables if n=0                   – Enables if n=1

Default : Enabled (n=1)

RGH| n|| Reads heating enabling state set in the in- strument
CWCnnnn| &|| Sets wind speed threshold to nnnn value (in hundredths of m/s)

Value should range within 0 and 100 hun- dredths of m/s (= 0…1 m/s)

Default : 20 (= 0.2 m/s)

(see Note 3 )

RWC| & nnnn|| Reads the wind speed threshold value set in the instrument (in hundredths of m/s)
CWaLnnn| &|| Sets time interval for the calculation of aver- age speed and average direction to nnn value

Value should range within 1 and 600 s.

If the value is greater than 10 s, it must be an integer multiple of 10.

note:the value can be greater than or equal to 10 s starting from firmware version 2.20

Default : 1 s

RWaL| & nnn|| Reads the time interval for the calculation of average speed and average direction set in the instrument
CWaMn| &|| Sets the method for the calculation of average speed and average direction:

§  If n=0: scalar mean. The average inten- sity is calculated as average of intensities without consideration of direction. The average direction is calculated as average of directions, and expressed according to the extended characteristic (see Note 4 ).

§  If n=1: vector mean. The average of the coordinates along each measurement axis is calculated. The average intensity and the average direction are those deter- mined by the two average coordinates.

Default : vector mean (n=1)

Command Reply Description
RWaM & n Reads the method for the calculation of the average speed and

average direction set in the instrument
CCn

note: command available from firmware version 2.06

| &|| Enables/disables the compass compensation of wind speed and direction:

– Disables if n=N                   – Enables if n=Y

Default : Enabled (n=Y)

CRTnnnn| &|| Sets the resolution of the rain gauge tipping bucket to nnnn value (in µm)

Value should range within 50 and 1599 µm

Default : 200 (= 0.200 mm)

RRT| & nnnn|| Reads the rain gauge resolution value set in the instrument (in µm)

NOTE 3 : WIND SPEED THRESHOLD VALUE
If the wind speed is very low, the determination of the direction can result inaccurate. The instrument allows setting the threshold value of speed below which the direction value is frozen on the last acquired value.
NOTE 4 : WIND DIRECTION EXTENDED CHARACTERISTIC
With 0÷359.9° wind direction measuring range, the analog output continues to oscillate between maximum and minimum scale if the direction continues to slightly fluctuate around 0°:deltaohm-HD52-Ultrasonic-
anemometers-21This effect can be reduced through the extended (“wrap-around”) characteristic of wind direction. In this mode, the wind direction is considered as corresponding to a 0÷539.9° range instead of 0÷359.9°. The wide output fluctuation occurs the first time that the wind direction goes from 0 to 359.9°; if later the “physical” direction goes back to 0°, the analog output will always remain around 360°. Using the extended characteristic, the behavior of the above graph changes into the following :deltaohm-HD52-Ultrasonic-anemometers-22If 539.9° value is exceeded in extended mode, the output goes to the value corresponding to 180°.
The table below shows the correspondence between the value of the analog output and the di-rection of the wind in the two modes.

Wind direction| 4…20mA output| 0…1V output| 0…5V output| 0…10V output
---|---|---|---|---
standard| extended| standard| extended| standard| extended| standard| extended
0°| 4.00| 4.00| 0.00| 0.00| 0.00| 0.00| 0.00| 0.00
180°| 12.00| 9.33| 0.50| 0.33| 2.50| 1.67| 5.00| 3.33
360°| 20.00| 14.67| 1.00| 0.67| 5.00| 3.33| 10.00| 6.67
540°| —| 20.00| —| 1.00| —| 5.00| —| 10.00

Analog outputs:

Command Reply Description
CAFxnn & Sets offset and direction of the analog output x (x=1 or 2) to:

§  Standard if nn=00

[ ex. 4…20 mA , 0…1 V , 0…5 V , 0…10 V ]

§  Without offset if nn=01

[ ex. 0…20 mA ]

§  With offset if nn=02

[ ex. 0.2…1 V , 1…5 V , 2…10 V ]

§  Inverted if nn=04

[ ex. 20…4 mA , 1…0 V , 5…0 V , 10…0 V ]

§  Inverted without offset if nn=05

[ ex. 20…0 mA ]

§  Inverted with offset if nn=06

[ ex. 1…0.2 V , 5…1 V , 10…2 V ]

Default : Standard (nn=00)

RAFx| & nn|| Reads offset and direction setting of the analog output x (x=1 or 2)
CAMn| &|| Association of the analog outputs:

§  If n= 0:

Output 1 = Mean wind speed Output 2 = Mean wind direction

§  If n= 1 (see Note 5 ):

Output 1 = Instant wind speed component along V-axis

Output 2 = Instant wind speed component along U-axis

§  If n= 2 ( Tunnel mode, see Note 6 ):

Output 1 = Instant wind speed component along the direction indicated by the arrow on the instrument case

Output 2 = Instant wind direction referred to the direction indicated by the arrow on the instrument case

Default : n=0

RAM| & n|| Reads the association of analog outputs
CAHn| &|| Associates full scale of wind speed analog output to:

§  5 m/s if n=0                 § 50 m/s if n=9

§ 10 m/s if n=1                 § 55 m/s if n=10

§ 15 m/s if n=2                 § 60 m/s if n=11

§ 20 m/s if n=3                 § 65 m/s if n=12

§ 25 m/s if n=4                 § 70 m/s if n=13

§ 30 m/s if n=5                 § 75 m/s if n=14

§ 35 m/s if n=6                 § 80 m/s if n=15

§ 40 m/s if n=7                 § 85 m/s if n=16

§ 45 m/s if n=8                 § 90 m/s if n=17

Default : 60 m/s (n=11)

Command Reply Description
RAH & n Reads the value corresponding to the wind speed analog output full

scale

By selecting the U and V components, the speed value associated to the initial scale of the two analog outputs is equal to the opposite of the speed value associated to the full scale of the outputs.
For example, if the speed full-scale value is set to 60 m/s, the speed range associated to the analog outputs is -60…+60 m/s.
NOTA 6: TUNNEL MODE
In tunnel mode, the wind direction measurement is not compensated by the magnetic compass, but the measurement is referred to the direction indicated by the arrow on the instrument case.
Output 2 is set to full-scale value if the wind blows in the direction of the arrow, and to initial scale value if the wind blows in the opposite direction.
The initial scale of output 1 is associated to the speed value opposite to that associated to the output full scale.
The arrow on the instrument case should be aligned with the direction of the tunnel.

Instrument information

Command Reply Description
G1 &Vnn.nn yyyy/mm/dd Version and date of firmware
RGD &yyyy/mm/dd hh.mm.ss Date and time of calibration
RGS &nnnnnnnn Serial number of instrument
RGI &ccc…ccc User code
CGIccc…ccc & Sets user code to ccc…ccc (max. 34 characters)

PROPRIETARY RS232 MODE

In proprietary RS232 mode, the instrument sends automatically the acquired measurements at regular intervals. The interval is factory-set to 1 second and is configurable from 1 to 3600 seconds. To change the interval, you should enter in configuration mode and send the com-mand CU2Rnnnn, where nnnn indicates the interval value in seconds (see chapter 6 “CON-FIGURATION” for details regarding the setting of operation parameters).
To use this mode, you must connect to a RS232 serial port. Communication parameters should be set in your PC as follows:

  • Baud rate: 9600 to 57600 (same as the setting in the instrument)
  • Data bits: 8
  • Parity: None
  • Stop bits: 2

The instrument sends measurements in the following format: with …. = values of the first, second,…., nth measurement = ASCII character Carriage Return = ASCII character Line Feed
Fields …. consist of 8 characters each. Measurement values are justified right; spaces can be added to the left margin of the values to obtain the 8 character length re-quested by the fields.
The sequence of the measurement values …. is configurable (see chapter 6 “CONFIGURATION”).
EXAMPLE
Supposing that the instrument measures the following values (the measuring unit is not considered, being not output by the instrument): M1=2.23, M2=-28.34, M3=0.34, M4=28.30, M5=359.3, M6=-1.3, the data string sent by the instrument takes the form:
2.23 -28.34 c 0.34 28.30 359.3 -1.3

PROPRIETARY RS485 MODE

In proprietary RS485 mode, the instrument sends the acquired measurements only if requested by the PC.
To use this mode, you must connect to an RS485 or RS422 serial port. Communication parameters should be set in the PC or data logger as follows:

  • Baud rate: 9600 to 115200 (same as the setting in the instrument)
  • Data bits: 8
  • Parity: None
  • Stop bits: 2

The instrument is requested to send measurements by generating a Break Signal (*) on the serial line for at least 2 ms, and then sending the following command, consisting of 4 ASCII characters:
M


with
= address of the instrument measurements are requested to = any ASCII character
EXAMPLE
To ask the instrument with address 2 to send the acquired measurements, do the following:

  1. Break Signal for at least 2 ms;
  2. Send command: M2aa.

The instrument answers with the following string:
IIIIM

I&….&AAAM

with
= address of the instrument sending measurements

…. = values of the first, second,…., nth measurement = space = checksum (hex value of the 8-bit checksum of all the preceding characters) = ASCII character Carriage Return ** **Fields …. consist of 8 characters each. Measurement values are justified right; spaces can be added to the left margin of the values to obtain the 8 characters length requested by the fields. The sequence of the measurement values …. is con-figurable (see chapter 6 “CONFIGURATION”). **EXAMPLE** Supposing that the instrument with address 2 measures the following values (the measuring unit is not considered, being not output by the instrument): M1=2.23, M2=-28.34, M3=0.34, M4=28.30, M5=359.3, M6=-1.3, the instrument reply takes the following form: A minimum time interval should elapse between two commands, depending on the Baud Rate setting:
Baud Rate Minimum interval between two commands
9600 200 ms
19200 100 ms
38400 70 ms
57600 40 ms
115200 25 ms

*() Break Signal** means interruption of the serial communication for a given time interval. It is used to inform the devices connected to the network that a command is going to be sent.

NMEA MODE

NMEA protocol, mainly used in the nautical field and in satellite-based navigation systems, specifies that only one of the devices connected can send data, while the others can only act as recipients.
In NMEA mode, the instrument sends automatically the acquired measurements at regular in-tervals. The interval is factory-set to 1 second and can be configured within 1 and 255 sec-onds. To change the interval, you should access the configuration mode and send CU4Rnnn command, where nnn indicates the interval value in seconds (see chapter 6 “CONFIGURATION” for details regarding the setting of operation parameters).
This mode is available with RS232, RS485 and RS422 serial connections. Communication pa-rameters should be set in the PC or data logger as follows:

  • Baud rate:same as the setting in the instrument (default = 4800)
  • Data bits:8
  • Parity:same as the setting in the instrument (default = None)
  • Stop bits: same as the setting in the instrument (default = 1)

The instrument is compatible with NMEA 0183 V4.00 protocol. The protocol establishes that data are sent in the following format: $,  with = field consisting in 5 alphanumeric characters: the first two indicate the type  with = field consisting in 5 alphanumeric characters: the first two indicate the type
The checksum is calculated by performing the exclusive OR of all characters ranging within $ and
symbols. The 4 most significant bits and the 4 less significant bits of the result are con-verted in hexadecimal. The hexadecimal value corresponding to the most significant bits is transmitted as the first.
The instrument regularly sends a string in the following general format requested by the proto-col: deltaohm-HD52-Ultrasonic-
anemometers-27If the instrument model is not equipped to measure some of the quantities indicated in the general format, the relevant fields will be empty and multiple consecutive commas will appear to indicate the missing fields.
The previous string doesn’t include the measurement of solar radiation or amount of rainfall. Models equipped with a pyranometer or a rain gauge provide to send the measurement with a second string continuously alternating to the former: deltaohm-HD52-Ultrasonic-anemometers-28The total amount of rainfall is the amount measured from when the instrument is powered. Note: the Wind Gust measurement is not available in NMEA mode.
EXAMPLE
Suppose that there are the following environmental conditions:

  • Wind speed = 5.60 m/s (=10.88 knot)
  • Wind direction with respect to magnetic North = 38.7°
  • Barometric pressure = 1014.9 hPa (= 30.0 inHg)
  • Relative humidity = 64.2 %
  • Air temperature = 26.8 °C
  • Solar radiation = 846 W/m2

Based on the above values, the following can be calculated:

  • Absolute humidity = 16.4 g/m3
  • Dew Point = 19.5 °C

The strings sent by the instrument in three different cases are indicated below:

  1. Case 1 – instrument measuring only wind speed and direction: $IIMDA,,I,,B,,C,,C,,,,C,,T,38.7,M,10.88,N,5.60,M*3A

  2. Case 2 – instrument measuring wind speed and direction, temperature, relative humidity and barometric pressure:
    $IIMDA,30.0,I,1.0149,B,26.8,C,,C,64.2,16.4,19.5,C,,T,38.7,M,10.88,N,5.60,M*36

  3. Case 3 – instrument measuring wind speed and direction, solar radiation, temperature, rela-tive humidity and barometric pressure:
    $IIMDA,30.0,I,1.0149,B,26.8,C,,C,64.2,16.4,19.5,C,,T,38.7,M,10.88,N,5.60,M36 alternated to: $IIXDR,G,846,,PYRA29
    For additional information regarding the protocol, visit the site “www.nmea.org”.

 MODBUS-RTU MODE

In MODBUS-RTU mode, the instrument sends the acquired measurements only if specifically requested by the PC, PLC or data logger.
The mode is available with RS232, RS485 and RS422 serial connections.
Communication parameters should be set in the PC or data logger as follows:

  • Baud rate:same as the setting in the instrument (default =19200)
  • Data bits:8
  • Parity:same as the setting in the instrument (default = even)
  • Stop bits:same as the setting in the instrument (default = 1)

MEASUREMENT READING (function 04h)
The Modbus function code 04h allows reading the values measured by the instrument. The ta-ble below lists the available quantities with the corresponding register number:
TAB. 10.A – Input Registers

Register number| Quantity| Format| From FW

version

| Note
---|---|---|---|---
1| Wind speed (x100)| unsigned 16 bits| 1.00| (1)
2| Wind direction in degrees (x10)| unsigned 16 bits| 1.00|
3| Sonic temperature measured by the first transducers pair (x10)| 16 bits| 1.00| (1)
4| Sonic temperature measured by the second transducers pair (x10)| 16 bits| 1.00| (1)
5| Average of the two sonic temperatures measured by the two transducers pairs (x10)| 16 bits| 1.00| (1)
6| Temperature measured by Pt100 sen- sor (x10)| 16 bits| 1.00| (1)
7| Relative humidity in %RH (x10)| unsigned 16 bits| 1.00|
8| Barometric pressure (x1000 if the unit of measurement is atm, x10 in the other cases)| unsigned 16 bits| 1.00| (1)
9| Compass angle in degrees (x10)| unsigned 16 bits| 1.00|
10| Solar radiation in W/m2| unsigned 16 bits| 1.00|
11| Mean wind speed (x100)| unsigned 16 bits| 1.00| (1)
12| Mean wind direction in degrees (x10)| unsigned 16 bits| 1.00|
13| Absolute humidity in g/m3 (x100)| unsigned 16 bits| 1.00|
14| Dew point temperature (x10)| 16 bits| 1.00| (1)
15| Wind direction in degrees (x10) with extended feature (see page 26)| unsigned 16 bits| 1.00|
16| Wind speed (x100) along V-axis| unsigned 16 bits| 2.00|
17| Wind speed (x100) along U-axis| unsigned 16 bits| 2.00|
Register number| Quantity| Format| From FW

version

| Note
---|---|---|---|---
18| Status register

bit0=1 Þ speed measurement error bit1=1 Þ compass measurement error bit2=1 Þ temperature meas. error bit3=1 Þ humidity measurement error bit4=1 Þ pressure measurement error bit5=1 Þ solar rad. measurement error

| unsigned 16 bits| 2.00|
19| Wind speed unit of measurement 0 Þ m/s                  3 Þ knot

1  Þ cm/s              4 Þ mph

2  Þ km/h

| unsigned 16 bits| 2.00|
20| Temperature unit of measurement 0 Þ °C                     1 Þ °F| unsigned 16 bits| 2.00|
21| Atm. pressure unit of measurement 0 Þ mbar (=hPa)           3 Þ mmH2O

1  Þ mmHg                        4 Þ inchH2O

2  Þ inchHg                        5 Þ atm

| unsigned 16 bits| 2.00|
22| Wind Gust intensity (x100)| unsigned 16 bits| 2.20| (1),(2)
23| Wind Gust direction in degrees (x10)| unsigned 16 bits| 2.20| (2)
24,25| Total amount of rainfall (x1000 if the unit of measurement is mm, x10000 if the unit of measurement is inch)| unsigned 16 bits| 2.22| (3)
26,27| Partial amount of rainfall (x1000 if the unit of measurement is mm, x10000 if the unit of measurement is inch)| unsigned 16 bits| 2.22| (3)
28| Rainfall rate (x10 if the unit of measure- ment is mm/h, x100 if the unit of meas- urement is inch/h)| unsigned 16 bits| 2.22|
29| Rainfall unit of measurement 0 Þ mm                  1 Þ inch| unsigned 16 bits| 2.22|

  1. For quantities with configurable measurement unit, the measurement value is expressed in the unit set in the instrument.
  2. The Wind Gust measurement is determined by continuously calculating the wind speed av-erages in a time interval equal to 3 seconds, and detecting the maximum value of the calculat-ed averages over the time elapsed between the current reading command and the previous reading command (the Wind Gust measurement is reset after each reading command).
  3. The total amount of rainfall is the amount measured from when the instrument is powered. The partial amount of rainfall is the amount measured from the last reading command. The amount of rainfall measurements are 32-bit integer values. Two consecutive 16-bit registers must be accessed to read a measurement. The register with lower address (for example the register with address 24 for the total amount of rainfall) contains the most significant bits.

INSTRUMENT ERROR CONDITIONS (function 07h)
The Modbus function code 07h allows reading the 8-bit register containing information about error conditions of the instrument.
Each register bit corresponds to an error condition:

  • Bit 0: Wind speed measurement error;
  • Bit 1: Compass measurement error;
  • Bit 2: Temperature measurement error;
  • Bit 3: Relative humidity measurement error;
  • Bit 4: Barometric pressure measurement error;
  • Bit 5: Solar radiation measurement error;
  • Bit 6: Not assigned;
  • Bit 7: Not assigned.

There is an error condition if the corresponding bit value is 1.
Note: the register does not include the precipitation measurement error.
READING OF INSTRUMENT GENERAL INFORMATION (function 2Bh / 0Eh)
The Modbus function code 2Bh / 0Eh allows reading the basic general information of the instrument, consisting of:

  • Manufacturer;
  • Instrument model;
  • Firmware version.

For additional information regarding the protocol, visit the site “www.modbus.org”.

sDI-12 MODE

In SDI-12 mode, the instrument sends the acquired measurements only if specifically request-ed by the PC.
To use this mode, an SDI-12 serial connection should be performed. The communication pa-rameters of the protocol are:

  • Baud rate:1200
  • Data bits:7
  • Parity: Even
  • Stop bit:1

Communication with the instrument is performed by sending a command in the following format:

!
with
= address of the instrument the command is sent to. = type of operation requested to the instrument.
The instrument reply is as follows:

with
= address of the replying instrument = information sent by the instrument = ASCII character Carriage Return = ASCII character Line Feed The instrument is compatible with V1.3 version of the protocol. The table below shows the available SDI-12 commands. To comply with the SDI-12 standard, the instrument address is indicated in the table with the letter a. The instrument comes with a factory address preset to 0. The address can be modified by using the proper SDI-12 com-mand reported in the table. **SDI-12 COMMANDS**
Command Instrument reply Description
a! a Checks for the presence of the instrument.
aI! allccccccccmmmmmmvvvx…x

where:

a = address of the instrument (1 character) ll = compatible SDI-12 version (2 characters) cccccccc = manufacturer (8 characters)

mmmmmm = instrument model (6 characters) vvv = firmware version (3 characters)

x…x = instrument version (up to 13 characters)

Ÿ     Ÿ     Ÿ

Þ Example of reply: 113DeltaOhmHD523D103P147R

with:

1 = address of the instrument

13 = compatible SDI-12 version 1.3 DeltaOhm = name of manufacturer HD523D = model of HD52.3D series 103 = firmware version 1.0.3

P147R = instrument version HD52.3DP147R

| Request for instrument information.
Command| Instrument reply| Description
---|---|---
aAb!

where:

b = new

address

| b

Note: if character b is not an acceptable address, the instrument answers with a in the place of b.

| Change of instrument address.
?!| a| Request for instrument address. If more than a sensor is connected to the bus, a conflict will occur.
aM!| atttn

where:

a = address of the instrument (1 character)

ttt = number of seconds needed by the instrument to make measurements available (3 charac- ters)

n = number of detected quantities (1 character)

Note: ttt is always equal to 000 because the meas- uring process is continuous. Measurements can be directly requested (command aD0!) without send- ing this command before.

| Request to perform the measurement.
aD0! aD1! aD2! aD3! aD4! aD5!| a a a

a a a

where:

a = address of the instrument (1 character)

    = wind speed   = wind direction in degrees          = temperature (sensor Pt100)       = relative humidity in %RH       = absolute humidity in g/m3       = dew point temperature

         = barometric pressure          = solar radiation in W/m2          = compass angle in degrees = mean wind speed = mean wind direction in degrees = wind Gust intensity = wind Gust direction in degrees = total amount of rainfall = partial amount of rainfall       = rainfall rate


Notes:

The positive values are always preceded by a + sign to identify the beginning of the measurement value.

| Request for measured values.
Command| Instrument reply| Description
---|---|---
| As to the quantities with configurable measuring unit (wind speed, temperature, barometric pres- sure and rainfall), the measurement value is ex- pressed in the unit set in the instrument.

measurement has the same format as tem- perature.

If a quantity measurement is in error condition, a negative value consisting in a series of 9 is sent.

The fields corresponding to quantities not meas- ured by that particular model are always present: a negative value consisting in a series of 9 is sent.

The Wind Gust measurement is available starting from firmware version 2.20.

The Wind Gust measurement is determined by continuously calculating the wind speed averages in a time interval equal to 3 seconds, and detect- ing the maximum value of the calculated averages over the time elapsed between the current reading command and the previous reading command (the Wind Gust measurement is reset after each read- ing command).

The total amount of rainfall is the amount meas- ured from when the instrument is powered. The partial amount of rainfall is the amount measured from the last reading command.

|

For additional information regarding the protocol, visit the site.

INSTRUMENT STORAGE

Instrument storage conditions:

  • Temperature: -40…+70 °C.
  • Humidity: less than 90 %RH no condensation.
  • In storing, avoid locations where:
  • There is a high humidity level.
  • The instrument is exposed to direct sunlight.
  • The instrument is exposed to a high-temperature source.
  • There are high vibration levels.
  • There is the presence of vapor, salt and/or corrosive gas.

SAFETY INSTRUCTIONS

General safety instructions
The instrument has been manufactured and tested in accordance with the safety standard EN61010-1:2010 “Safety requirements for electrical equipment for measurement, control and laboratory use” and has left the factory in perfect safety technical conditions.
The instrument’s proper operation and operating safety can be ensured only if all standard safety measures, as well as the specific measures described in this manual, are followed.
The instrument’s proper operation and operating safety can be ensured only in the climatic conditions specified in this manual.
Do not use the instrument in places where there are:

  • Rapid ambient temperature variations that may cause condensation.
  • Corrosive or flammable gases.
  • Direct vibrations or shocks to the instrument.
  • High-intensity electromagnetic fields, static electricity.

If the instrument is moved from a cold place to a hot one, or vice versa, condensation for-mation can cause an instrument malfunction. In this case, you will have to wait for the instru-ment temperature to reach ambient temperature before turning the instrument on.
User obligations
The instrument operator shall follow the directives and regulations below that refer to the treatment of dangerous materials:

  • EEC directives on workplace safety
  • National low regulations on workplace safety
  • Accident prevention regulations

ACCESSORIES ORDERING CODES

  • CP52… Connecting cable with 19-pole M23 female free connector on one end, open wires on the other. Available lengths: 5 m (CP52.5), 10 m (CP52.10), 15 m (CP52.15), 20 m (CP52.20), 30 m (CP52.30), 50 m (CP52.50) and 75 m (CP52.75).
  • RS52 Serial connection cable with built-in USB/RS232 converter. USB connector for the PC and screw terminals on the instrument side.
  • HD2005.20 Tripod with adjustable legs for installing environmental sensors. Material: anodized aluminum. Max. height 225 cm. It can be fixed on a flat base with screws or to the ground with pegs. Foldable legs for transport.
  • HD2005.20.1; Tripod with adjustable legs for installing environmental sensors. Material: anodized aluminum. Max. height 335 cm. It can be fixed on a flat base with screws or to the ground with pegs. Foldable legs for transport.
  • CP52.C ; Additional 19-pole M23 female free connector.

Additional 19-pole M23 female free connector.

WARRANTY

Delta OHM is required to respond to the factory warranty” only in those cases provided byY Legislative Decree b September 2005 – n. 206. Each instrument is sold after ngorous inspections; if any manufacturing defect is found, instrument was purchased. During the warranty period (24 months from the date of invoice) any manufacturing defects fOund will be repaired free or charge. MISuse, wear, neglect, lack or inefficient maintenance as well as theft and damage duing transport are excluded. warranty does not apply Iir changes, tampering or unauthorized repairs are made on the product. Solutions, probes, electrodes and microphones are not guaranteed as the improper use, even for a few minutes, may cause irreparable damages. Delta OHM repairs the products that show defects of construction in accordance with the terms and conditions of the warranty included in the manual of the product. For any dispute, the competent court is the Court of Padua. The Italian law and the “Convention on Contracts for the International sales Or Goods appiy. is necessary to contact the distributor where the
TECHNICAL INFORMATION
The quality level of our instruments is the result of the continuOus product development. This may ead to differences between the information reported in the manual and the instrument you have purchased. In case of discrepancies and/or inconsistencies. Delta OHM reserves the right to change technical specifications and dimensions to fit the product requirements without prior notice.
DISPOSAL INFORMATION Electrical and electronic equipment marked with a specific symbol in compliance with 2012/19/EU Directive must be disposed or separately from household waste. European users can hand them over to the dealer or to the manufacturer wnen purchasing a new electrical and electronic equipment, or to a WEEE COllection point designated by local authorities. legal disposal is punished by law. Disposing of electrical and electronic equipment separately from normal waste helps to preserve natural resources and allows materials to be recycled in an environmentally friendly way without risks to human health.

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals