DeltaOHM LPPHOT02 Photometric Probe Instruction Manual
- May 15, 2024
- DeltaOHM
Table of Contents
Operating manual
Photometric probe
LPPHOT02www.deltaohm.com
Keep for future reference.
INTRODUCTION
The LPPHOT02 probe measures illuminance (lux), defined as the ratio between
the luminous flux (lumen) through a surface and the surface area itself (m2).
The spectral response curve of a photometric probe is equal to the human eye,
known as standard photopic curve V(λ). The difference in spectral response
between the LPPHOT02 and the standard photopic curve V(λ) is evaluated through
the calculation of the error f’1.
The probe has been designed for outdoor installation for long periods, in
particular for the measurement of daylight in climatological and
meteorological applications.
Available in the following versions:
- LPPHOT02: PASSIVE.
- LPPHOT02AC: ACTIVE with 4..20 mA CURRENT output.
- LPPHOT02AV: ACTIVE with 0..1 or 0..5 or 0..10 V VOLTAGE output to be defined when ordering.
Each luxmeter is individually calibrated at the factory and is distinguished by its own calibration factor. The calibration is performed by comparison with the reference luxmeter of the Delta OHM metrology laboratory, using an illuminant “A” as a source, as required by CIE publication N°69 “Methods of characterizing illuminance meters and luminance meters: Performance, characteristics and specifications”.
WORKING PRINCIPLE
LPPHOT02 probe is based on a solid state sensor, whose spectral response has been adapted to that of the human eye using a special filter. The relative spectral response is reported on figure 2.1.
The luxmeter is provided with a 50 mm external diameter dome in order to
supply a suitable protection of the sensor to the weather agents.
The response in accordance with the cosine law has been obtained thanks to the
particular shape of the diffuser and of the housing. The deviation between the
theoretical response and the measured one is shown in the Fig. 2.2.
The excellent relation between the response of the luxmeter and the cosine law allows using the instrument also when the sun has a very low raising.
INSTALLATION
Before installing the sensor, refill the cartridge containing silica-gel
crystals. Silica gel absorbs humidity in the dome chamber and prevents, in
particular climatic conditions, condensation on the internal walls of the
domes and measurement alteration.
Do not touch the silica gel crystals with your hands while refilling the
cartridge. Carry out the following instructions in an environment as drier as
possible:
- Loosen the three screws that fix the white shade disk.
- Unscrew the silica gel cartridge using a coin.
- Remove the cartridge perforated cap.
- Open the sachet containing silica gel (supplied with the instrument).
- Fill the cartridge with the silica gel crystals.
- Close the cartridge with its own cap, paying attention that the sealing O-ring be properly positioned.
- Screw the cartridge to the sensor body using a coin.
- Check that the cartridge is screwed tightly (if not, silica gel life will be reduced).
- Position the shade disk and screw it with the screws.
- The sensor is ready for use.
The figure below shows the operations necessary to fill the cartridge with the silica gel crystals.
- The sensor must be mounted in an easy-to-reach location in order to clean the dome regularly and carry out maintenance. At the same time, make sure that no buildings, constructions, trees or obstructions exceed the horizontal plane where the sensor lies. If this is not possible, select a site where obstructions in the path of the sun from sunrise to sunset do not exceed 5 degrees of elevation.
- The sensor must be located far from any kind of obstruction, which might reflect sunlight (or sun shadow) onto the sensor itself.
- For fixing, use the holes on the sensor body (remove the shade disk to access the holes and reposition it after mounting) or the suitable accessories (see the figures below). In order to allow an accurate horizontal positioning, the sensor is equipped with a levelling device: the adjustment is made by means of the two levelling screws that allow adjusting the sensor inclination. The mast height does not exceed the sensor plane to avoid measurement errors caused by any reflection or shadow of the mast itself.
- It is preferably to thermally insulate the sensor from its mounting bracket ensuring, at the same time, a good electrical contact to ground.
ELECTRICAL CONNECTIONS
LPPHOT02… has a 4-pole connector and uses the CPM12AA4… optional cables.
The metallic housing of the sensor should preferably be grounded ( ****)
locally. In this case, do not connect the wire of the cable corresponding to
the housing to prevent ground loops.
Only if it is not possible to ground locally the metallic case of the sensor,
connect the wire of the cable corresponding to the housing to ground. Note: in
LPPHOT02AV the housing is not connected to the connector.
4.1 LPPHOT02 CONNECTIONS
The luxmeter LPPHOT02 is passive and does not require power supply. It is to
be connected either to a millivoltmeter or to a data acquisition system.
Typically, the luxmeter output signal does not exceed 150 mV. In order to
better exploit the luxmeter features, the readout instrument should have 1 µV
resolution.
Connector | Function | Color |
---|---|---|
1 | +Vout | Red |
2 | -Vout | Blue |
3 | Housing (C) | White |
4 | Cable shield (SH) | Black |
4.2 LPPHOT02AC CONNECTIONS
The LPPHOT02AC luxmeter has 4…20 mA output and requires 10…30 Vdc external
power supply. It is to be connected to a power supply and an instrument with
4…20 mA input as shown in fig. 4.2. The load resistance of the instrument
reading the signal must be ≤ 500 Ω.
Connector | Function | Color |
---|---|---|
1 | Positive (Iin) | Red |
2 | Negative (Iout) | Blue |
3 | Housing (C) | White |
4 | Cable shield (SH) | Black |
4.3 LPPHOT02AV CONNECTIONS
The LPPHOT02AV luxmeter has 0…1 V, 0…5 V or 0…10 V output (depending on the
ordered output) and requires external power supply: 10…30 Vdc for 0…1 V and
0…5 V outputs, 15…30 Vdc for 0…10 V output. It is to be connected to a power
supply and an instrument with voltage input as shown in fig. 4.3. The load
resistance of the instrument reading the signal must be ≥ 100 kΩ.
Connector | Function | Color |
---|---|---|
1 | Output positive (+Vout) | Red |
2 | Output negative Power supply negative (GND) | Blue |
3 | Power supply positive (+Vdc) | White |
4 | Cable shield (SH) | Black |
MEASUREMENT
Below are the ways to calculate the illuminance.
5.1 LPPHOT02
Each luxmeter is distinguished by its own sensitivity (or calibration factor)
S expressed in mV/klux and shown in the label on the sensor (and in the
calibration report).
The irradiance Ee is obtained by measuring with a multimeter the difference of
potential DDP at the ends of the sensor and applying the following formula:
Ee = DDP / S
where:
Ee is the illuminance expressed in klux;
DDP is the difference of potential expressed in mV measured by the multimeter;
S is the sensitivity of the luxmeter expressed in mV/klux.
5.2 LPPHOT02AC
The 4…20 mA output signal corresponds to 0…150 klux illuminance range.
The illuminance Ee is obtained by measuring with a multimeter the current Iout
absorbed by the sensor and applying the following formula:
Ee = 9.375 • (Iout – 4mA)
where:
Ee is the illuminance expressed in klux;
Iout is the current expressed in mA absorbed by the luxmeter.
5.3 LPPHOT02AV
The output signal (0…1 V, 0…5 V or 0…10 V depending on the version)
corresponds to 0…150 klux illuminance range.
The illuminance Ee is obtained by measuring with a multimeter the output
voltage Vout of the sensor and applying the following formula:
Ee = 150 • Vout for the version 0…1 V
Ee = 30 • Vout for the version 0…5 V
Ee = 15 • Vout for the version 0…10 V
where:
Ee is the illuminance expressed in klux;
Vout is the output voltage expressed in V measured by the multimeter.
MAINTENANCE
In order to grant measurements high accuracy, it is important to keep the
glass dome clean. Consequently, the more the dome will be kept clean, the more
measurements will be accurate.
You can wash it using water and standard papers for lens. If necessary, use
pure ETHYL alcohol. After using alcohol, clean again the dome with water only.
Because of the high temperature changes between day and night, some
condensation might appear on the sensor dome. In this case the performed
reading is highly overestimated. To minimize the condensation, the sensor is
provided with a cartridge containing dessicant material (silica-gel). The
efficiency of the silica-gel crystals decreases over time while absorbing
humidity. Silica-gel crystals are efficient when their colour is yellow, while
they turn white/translucent as soon as they lose their efficiency. Read
instructions at chapter 3 about how to replace the silica-gel crystals.
Silicagel typical lifetime goes from 2 to 6 months depending on the
environment where the sensor works.
To exploit all the sensor features, it is highly recommended that the
calibration be checked annually.
TECHNICAL SPECIFICATIONS
Measuring range | 0…150 klux |
---|---|
Typical sensitivity | 0.5…2 mV/klux |
Viewing range | 2p sr |
Impedance | 0.5…1 KW |
Spectral range | Standard photopic curve |
Response time | <0.5 s (95%) |
Output | LPPHOT02 = mV/klux LPPHOT02 AC = 4…20 mA |
LPPHOT02 AV = 0…1, 0…5, 0…10 V (depending on the model)
Power supply| 10…30 Vdc (15…30 Vdc only for 0…10 V output)
Operating temperature| -40 °C…+80 °C
Response according to the cosine law| < 8 % (0…80°)
Long term instability (1 year)| <|±3| %
Error f’ 1| <9 %
Non linearity| <÷±1÷ %
Temperature response| < 0.1 %/°C
Weight| 900 g approx.
SAFETY INSTRUCTIONS
General safety instructions
The instrument has been manufactured and tested in accordance with the safety
standard EN61010-1:2010 “Safety requirements for electrical equipment for
measurement, control and laboratory use” and has left the factory in perfect
safety technical conditions.
The instrument proper operation and operating safety can be ensured only if
all standard safety measures as well as the specific measures described in
this manual are followed.
Do not use the instruments in places where there are:
- Corrosive or flammable gases.
- Direct vibrations or shocks to the instrument.
- High-intensity electromagnetic fields, static electricity.
User obligations
The instrument operator shall follow the directives and regulations below that
refer to the treatment of dangerous materials:
- EEC directives on workplace safety.
- National law regulations on workplace safety.
- Accident prevention regulations.
ACCESSORIES ORDERING CODES
The luxmeter is supplied with shade disk, sylica-gel cartridge, 2 spare
sachets, levelling device, M12 connector and Calibration Report.
Accessories
LPSP1 | UV-resistant shade disk (spare part). |
---|---|
LPS1 | Only attachment bracket, suitable for mast with diameter 40…50 mm. |
Installation on horizontal or vertical mast.
LPRING02| Base with levelling device and adjustable holder for mounting the
pyranometer in an inclined position (specify upon ordering on which luxmeter
model has to be mounted).
HD2003.79K| Kit to mount the luxmeter on ∅ 40 mm clamping. To install the
luxmeter on a transverse mast.
LPS6| Kit for the installation of the luxmeter. The kit includes: 750 mm mast,
base fitting, graduated support plate, bracket for luxmeter.
CPM12AA4…| 4-pole cable. 4-pole M12 connector on one end, open wires on the
other end. Available length 2 m (CPM12AA4.2), 5 m (CPM12AA4.5) or 10 m
(CPM12AA4.10).
LPSG| Cartridge to contain desiccant silica-gel crystals, complete with O-ring
and cap (spare part).
LPG| Pack of 5 sachets of silica-gel crystals.
DELTA OHM metrology laboratories LAT N° 124 are ISO/IEC 17025 accredited by ACCREDIA for Temperature, Humidity, Pressure, Photometry / Radiometry, Acoustics and Air Velocity. They can supply calibration certificates for the accredited quantities.
NOTES
……………..
WARRANTY
The manufacturer is required to respond to the “factory warranty” only in
those cases provided by Legislative Decree 6 September 2005 – n. 206. Each
instrument is sold after rigorous inspections; if any manufacturing defect is
found, it is necessary to contact the distributor where the instrument was
purchased from. During the warranty period (24 months from the date of
invoice) any manufacturing defects found will be repaired free of charge.
Misuse, wear, neglect, lack or inefficient maintenance as well as theft and
damage during transport are excluded. Warranty does not apply if changes,
tampering or unauthorized repairs are made on the product. Solutions, probes,
electrodes and microphones are not guaranteed as the improper use, even for a
few minutes, may cause irreparable damages.
The manufacturer repairs the products that show defects of construction in
accordance with the terms and conditions of warranty included in the manual of
the product. For any dispute, the competent court is the Court of Padua. The
Italian law and the “Convention on Contracts for the International Sales of
Goods” apply.
TECHNICAL INFORMATION
The quality level of our instruments is the result of the continuous product
development. This may lead to differences between the information reported in
the manual and the instrument you have purchased.
We reserves the right to change technical specifications and dimensions to fit
the product requirements without prior notice.
DISPOSAL INFORMATION
**** Electrical and electronic equipment marked with specific symbol in
compliance with 2012/19/EU Directive must be disposed of separately from
household waste. European users can hand them over to the dealer or to the
manufacturer when purchasing a new electrical and electronic equipment, or to
a WEEE collection point designated by local authorities. Illegal disposal is
punished by law.
Disposing of electrical and electronic equipment separately from normal waste
helps to preserve natural resources and allows materials to be recycled in an
environmentally friendly way without risks to human health.
Please note our new name:
Senseca Italy Srl
Via Marconi 5, 35030 Padua, Italy
Documents are in the process of being changed. Members of GHM GROUP:
Documents / Resources
|
DeltaOHM LPPHOT02 Photometric
Probe
[pdf] Instruction Manual
LPPHOT02 Photometric Probe, LPPHOT02, Photometric Probe, Probe
---|---
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>