Step 1: Software installation

  1. Scan the QR code below or go to https://www.aandd.jp/products/medical/professional/abpm.html to fill out the registration and download the software..https://www.aandd.jp/products/medical/professional/abpm.html

  2. When the .zip file is done downloading (this may take a few minutes):
    A D TM-2440 Professional Ambulatory Blood Pressure Monitor User Guide - When
the .zip file is done downloading

  3. Follow the guided instructions. This may take a few minutes.

  4. To open the application, navigate to your desktop and double click the A&D Medical ABPM Data Manager icon. If it is not on your desktop, search for AND Client in recently installed programs.

Step 2: Inserting the batteries

  1. Open the battery cover.
  2. Refer to the symbol inside the battery compartment. Insert two new AA batteries into the proper “+” and “-” direction.
  3. Close the battery cover.

Battery Tips

  1. New AA batteries will last approximately 200 measurements. This varies depending on the quality of batteries used and operating temperatures.
  2. There is an internal coin cell battery that will retain measurement results and clock settings if the AA batteries die. This is charged by new AA batteries.
  3. During first use or if monitor is stored for a month or longer, replace AA batteries and leave in for at least 48 hours. This will recharge the internal coin cell battery. The monitor CAN be used during this time.

Step 3: Connecting the monitor

  1. Connect the micro USB cable between the monitor and computer.
    A D TM-2440 Professional Ambulatory Blood Pressure Monitor User Guide -
Connect the micro USB cable between the monitor and
computer

  2. The buzzer will sound and OFF appears on the OLED display. The monitor enters its standby mode.

Step 4: Programming the monitor

  1. The first time the software launches, you will be prompted to enter your license code that you received by email when you registered. Then press the Register button.

  2. Click on Recorder icon.

  3. Select settings for a new ABPM study. Once complete, select Save. The device is now programmed.
    A D TM-2440 Professional Ambulatory Blood Pressure Monitor User Guide -
Select settings for a new ABPM study

  4. Disconnect the micro USB cable from the device.

  5. Turn the ABPM function ON. For the TM-2440 only, press and hold Event for at least three seconds. The icon should appear on the display. For the TM-2441 only, slide the Auto Switch to ON, then lock it.

A D TM-2440 Professional Ambulatory Blood Pressure Monitor User Guide - Turn
the ABPM function ON

Step 5: Prepare the patient

  1. Wrap the cuff around the patient’s upper arm (refer to below cuff guide).
    A D TM-2440 Professional Ambulatory Blood Pressure Monitor User Guide - Wrap
the cuff around the patient’s upper arm

  2. Position tube around the back of the neck and attach monitor to the belt loop. Clip the tube to the undershirt. If user does not have undershirt, tape to skin. Wrap the belt loop around waist. Fasten the air plug from the air hose into the air socket. Secure monitor into the carrying case.
    A D TM-2440 Professional Ambulatory Blood Pressure Monitor User Guide -
Position tube around the back of the neck and attach
monitor

  3. Press the START/STOP button on the top of the device to start the first measurement. The programmed intervals will automatically begin.

Patient Tips
DO

  1. When inflation begins, relax arm, stay still and quiet.
  2. Ensure the tube is not compressed or kinked.
  3. Track your bed time and awake time then report to clinician.

DO NOT

  1. Operate heavy machinery. Vibrations may disrupt measurements.
  2. Remove batteries during session.
  3. Drop or shock the device.
  4. Expose the device to rain or water.
  5. Arrange the cuff hose around your neck during sleep.

Step 6: Downloading the data

A D TM-2440 Professional Ambulatory Blood Pressure Monitor User Guide -
Downloading the data A D TM-2440 Professional
Ambulatory Blood Pressure Monitor User Guide - Downloading the
data

Step 7: Reviewing data and printing report

  1. The data list for all measurements appear.
    A D TM-2440 Professional Ambulatory Blood Pressure Monitor User Guide - The
datalist for all measurements appear

  2. To generate report, click the Layout icon on the top menu.

  3. Customize your report by selecting and deselecting items in the report layout. Click Apply and Save to set layout for future report generation.

4A. Click on the Print icon to print results.
4B. To save as .PDF, choose Views. Then, click on the Export icon.
Choose the file location and file name. Click Save As. It will automatically generate a .CSV and .PDF file and save it in the selected file location.

A D TM-2440 Professional Ambulatory Blood Pressure Monitor User Guide -
Click on the Print icon to print results

andmedical.com
©2022 A&D Medical. All rights reserved. TM2440/41-QSG 1WDEK4011217