MEAN WELL GSM220B Series 220W AC and DC Reliable Green Medical Adaptor Owner’s Manual
- June 1, 2024
- MEAN WELL
Table of Contents
MEAN WELL GSM220B Series 220W AC and DC Reliable Green Medical Adaptor
Features
- 2 pole AC inlet IEC320-C8, Class Il power unit
- Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1/-1-11 and IEC/BS EN/EN60601-1/-1-11
- Extremely low leakage current
- No load power consumption<0.15W
- Energy efficiency level VI and meet CoC Version 5
- -30~+70°C wide range working temperature
- Protections: Short circuit / Overload / Over voltage / Over temperature
- LED indicator for power on
- Lifetime > 70K hours
- Various DC plug quick adapter accessories available (Plug kit sold separately, please refer to: https://www.meanwell.com/upload/pdf/DC_plug.pdf)
- 3 years warranty
Applications
- Mobile clinical workstation
- Oral irrigator
- Portable hemodialysis machine
- Breath Machine
- Medical computer monitor
GTIN CODE
MW Search: https://www.meanwell.com/serviceGTIN.aspx
Description
GSM220B is a highly reliable, 220W desktop-style single-output green medical adaptor series. This product is equipped with a 2-pin (no FG) standard IEC320-C8 power plug, adopting the input range from 80 VAC to 264 VAC. The entire series supplies different output voltages between 12VDC and 48VDC that can satisfythe demands for various kinds of medical electrical devices. The circuitry design meets the international medical standards (2*MOPP), having an ultra-low leakage current (<100 MA), and fitting the medical devices in direct electrical contact with the patients. With an efficiency up to 94.5% and extremely low no-load power consumption below 0.15W, GSM220B is compliant with USA EISA 2007/DoE, Canada NRCan, Australia and New Zealand MEPS, EU ErP,and meets Code of Conduct (CoC) Version 5. The supreme feature allows the adaptor to save energy when it is either in operating mode or the standby mode. The entire series utilizes the 94V-0 flame retardant plastic case, providing the double insulation that effectively prevents electrical shock. GSM220B is approved with international medical safety certificates.
Model Encoding
SPECIFICATION
ORDER NO.| GSM220B12-R7B| GSM220B15-R7B| GSM220B20-R7B|
GSM220B24-R7B| GSM220B48-R7B
---|---|---|---|---|---
OUTPUT
| SAFETY MODEL NO.| GSM220B12| GSM220B15| GSM220B20| GSM220B24|
GSM220B48
DC VOLTAGE Note.2| 12V| 15V| 20V| 24V| 48V
RATED CURRENT| 15A| 13.4A| 11A| 9.2A| 4.6A
CURRENT RANGE| 0 ~ 15A| 0 ~ 13.4A| 0 ~ 11A| 0 ~ 9.2A| 0 ~ 4.6A
RATED POWER (max.)| 180W| 201W| 220W| 221W| 221W
RIPPLE & NOISE (max.) **Note.3| 80mVp-p| 80mVp-p| 120mVp-p|
120mVp-p| 150mVp-p
VOLTAGE TOLERANCE Note.4| ±5.0%| ±5.0%| ±4.0%| ±3.0%| ±2.0%
LINE REGULATION Note.5| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%
LOAD REGULATION| ±5.0%| ±5.0%| ±4.0%| ±3.0%| ±2.0%
SETUP, RISE TIME Note.6| 2000ms, 50ms / 230VAC 2000ms, 50ms
/ 115VAC at full load
HOLD UP TIME (Typ.)**| 24ms / 230VAC 24ms / 115VAC at full load
INPUT
| VOLTAGE RANGE Note.7| 80 ~ 264VAC 113 ~ 370VDC
FREQUENCY RANGE| 47 ~ 63Hz
POWER FACTOR (Typ.)| PF>0.91 / 230VAC PF>0.98 / 115VAC at full
load
EFFICIENCY (Typ.)| 90%| 90%| 92%| 93.5% 94.5%
AC CURRENT (Typ.)| 4A / 115VAC 2A / 230VAC
INRUSH CURRENT (max.)| Cold start 90A / 115VAC 110A / 230VAC
LEAKAGE CURRENT(max.)| Touch current < 100 A/264VAC
PROTECTION
| OVERLOAD| 105 ~ 135% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is
removed
OVER VOLTAGE| 105 ~ 135% rated output voltage
Protection type : Shut down o/p voltage, re-power on to recover
OVER TEMPERATURE| Shut down o/p voltage, recovers automatically after
temperature goes down
ENVIRONMENT
| WORKING TEMP.| -30 ~ +70℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20% ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03% / ℃ (0~40℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X,
Y, Z axes
OPERATING ALTITUDE Note.8| 3000 meters
SAFETY & **EMC**
(Note. 9)
| SAFETY STANDARDS| IEC60601-1/IEC60601-1-11, BS EN/EN60601-1/ BS
EN/EN60601-1-11, ANSI/AAMI ES60601-1 / ES60601-1-11(3.1 version),
CAN/CSA-C22.2 No. 60601-1:14 – Edition 3, EAC TP TC 004, KC K60950-1 (for
GSM220B24 only) approved
ISOLATION LEVEL| Primary-Secondary: 2xMOPP
WITHSTAND VOLTAGE| I/P-O/P: 4KVAC
ISOLATION RESISTANCE| I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION
| Parameter| Standard| Test Level / Note
Conducted emission| BS EN/EN55011 (CISPR11), FCC PART 15 / CISPR22, CAN
ICES-3(B)/NMB-3(B)| Class B
Radiated emission| BS EN/EN55011 (CISPR11), FCC PART 15 / CISPR22, CAN
ICES-3(B)/NMB-3(B)| Class B
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–
EMC IMMUNITY
| BS EN/EN60601-1-2, BS EN/EN61204-3
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 3, 1KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption
| ****
BS EN/EN61000-4-11
| 100% dip 1 periods, 30% dip 25 periods,
100% interruptions 250 periods
OTHERS
| MTBF| 1953.3K hrs min. Telcordia SR-332 (Bellcore) ; 208.7K hrs
min. MIL-HDBK-217F (25℃)
DIMENSION| 2108546mm (LWH)
PACKING| 1.1Kg; 12pcs/14.2Kg/0.81CUFT
CONNECTOR
| PLUG| See page 4~5; Other type available by customer requested
CABLE| See page 4~5; Other type available by customer requested
NOTE
| 1. All parameters are specified at 230VAC input, rated load, 25℃ 70% RH
ambient.
2. DC voltage: The output voltage set at point measure by plug terminal & 50%
load.
3. Ripple & noise are measured at 20MHz by using a 12″ twisted pair
terminated with a 0.1μF & 47μF capacitor.
4. Tolerance: includes set up tolerance, line regulation, load regulation.
5. Line regulation is measured from low line to high line at rated load.
6. Length of set up time is measured at first cold start. Turning ON/OFF the
power supply may lead to increase of the set up time.
7. Derating may be needed under low input voltage. Please check the derating
curve for more details.
8. The ambient temperature derating of 3.5℃/1000m with fanless models and of
5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).
9. The power supply is considered as an independent unit, but the final
equipment still need to re-confirm that the whole system complies with the EMC
directives. For guidance on how to perform these EMC tests, please refer to
“EMI testing of component power supplies.”
(as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )
※ Product Liability Disclaimer:For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx
Derating Curve
Static Characteristics
Mechanical Specification
DC output plug
- Standard plug: R7B
DC plug changeable through:
- Customization of the standard part with an optional DC plug according to the table (MOQ applicable)
- Quick adapter accessory (sold separately without MOQ) Please refer to the below table and online selection guide :
Example quick adapter accessory:
Optional DC plug: (Available in customized cable or quick adapter)
Installation Manual
Please refer to : http://www.meanwell.com/manual.html
- Downloaded from Arrow.com.