Specifications :

  • Model: MFM-10 series
  • Power Output: 10W
  • Size: 1.65 x 0.88 inches
  • Safety Approval: ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
  • Medical Safety Standard: 2 x MOPP
  • Application: Suitable for BF application with appropriate system consideration
  • No load power consumption

Product Usage Instructions:

  1. Safety Precautions:
    Before using the MFM-10 series product, please read the user manual carefully and follow all safety instructions provided.

  2. Installation:

  3. Ensure that the power source matches the product specifications.

  4. Connect the product to the appropriate system following the guidelines provided in the manual.

  5. Powering On:

  6. Once installed correctly, turn on the power source.

  7. The product should now be ready for operation.

  8. Product Care:

  9. Keep the product away from water or any liquid exposure.

  10. Do not attempt to open or modify the product casing.

Frequently Asked Questions (FAQ):

  • Q: What is the size of the MFM-10 series product?
    A: The product has a compact size of 1.65 x 0.88 inches.

  • Q: Is the MFM-10 series suitable for medical applications?
    A: Yes, the MFM-10 series is Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1, making it suitable for medical applications with appropriate system consideration.

10W High Reliable Green Medical On Board Type

MEAN-WELL-MFM-10-5-10W-High-Reliable-Green-Medical-On-Board-Type-01
\(1\)

Features

  • 1.65″x0.88″ compact size
  • Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
  • Suitable for BF application with appropriate system consideration
  • No load power consumption<0.075W
  • Extremely low leakage current
  • Wide operating temp. range -30 ~ +85°C
  • EMI class B for class II configuration
  • Protections: Short circuit / Overload / Over voltage / Over temperature
  • No minimum load required
  • 3 years warranty

Applications

  • Portable medical device
  • Mobile clinical workstation
  • Medical computer monitor
  • Medical examination instrument

GTIN CODE
MW Search: https://www.meanwell.com/serviceGTIN.aspx

Description

MFM-10 is a 10W high density and small size (4222.320.5mm) AC/DC on board type medical grade power supply series. It features the operation for 80~264 VAC, a low no load power consumption less than 0.075W, a high efficiency up to 84%, Class II (no FG) double insulation, outstanding dissipation, 5G anti- vibration, high EMC performance, 4KVAC isolation, etc. The design observes IEC/BS EN/EN60601-1 and ANSI/AAMI ES60601-1 version three with 2xMOPP level and ultra-low leakage current (<80 m A). It is very suitable for BF (patient contact) type medical device or relevant equipment.

Model Encoding

SPECIFICATION

MODEL| MFM-10-3.3| MFM-10-5| MFM-10-12| MFM-10-15| MFM-10-24
---|---|---|---|---|---
OUTPUT| DC VOLTAGE| 3.3V| 5V| 12V| 15V| 24V
RATED CURRENT| 2.5A| 2A| 0.85A| 0.67A| 0.42A
CURRENT RANGE Note.2| 0 ~ 2.5A| 0 ~ 2A| 0 ~ 0.85A| 0 ~ 0.67A| 0 ~ 0.42A
PEAK CURRENT| 2.75A| 2.2A| 0.94A| 0.74A| 0.46A
RATED POWER| 8.3W| 10W| 10.2W| 10W| 10W
PEAK LOAD(10sec.) **Note.3| 9W| 11W| 11.3W| 11.1W| 11W
RIPPLE & NOISE (max.) **Note.4| 120mVp-p| 100mVp-p| 180mVp-p| 180mVp-p| 200mVp-p
VOLTAGE TOLERANCE Note.5| ±2.5%| ±2.5%| ±2.5%| ±2.5%| ±2.5%
LINE REGULATION| ±0.3%| ±0.3%| ±0.3%| ±0.3%| ±0.3%
LOAD REGULATION| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%
SETUP, RISE TIME| 1000ms, 30ms/230VAC            1000ms, 30ms/115VAC at full load
HOLD UP TIME (Typ.)| 40ms/230VAC            8ms/115VAC at full load
INPUT| VOLTAGE RANGE Note.6| 80 ~ 264VAC
FREQUENCY RANGE| 47 ~ 440Hz
EFFICIENCY (Typ.)| 78%| 81%| 83%| 83%| 84%
AC CURRENT (Typ.)| 0.3A/115VAC            0.2A/230VAC
INRUSH CURRENT (Typ.)| COLD START  25A/115VAC            45A/230VAC
LEAKAGE CURRENT (max.) Note.7| Touch current <80μA/264VAC
PROTECTION| OVERLOAD| 110% ~ 180% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
OVER VOLTAGE| 3.8 ~ 5V| 5.75 ~ 6.8V| 13.8 ~ 16.2V| 17.3 ~ 20.3V| 27.6 ~ 32.4V
Protection type : Shut off o/p voltage, clamping by zener diode
OVER TEMPERATURE| Protection type : Shut down o/p voltage, recovers automatically after temperature goes down
ENVIRONMENT| WORKING TEMP.| -30 ~ +85℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +100℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03%/℃ (0 ~ 50℃)
SOLDERING TEMPERATURE| Wave soldering: 265℃,5s (max.); Manual soldering: 390℃,3s (max.)
VIBRATION| 10 ~ 500Hz, 5G 10min./1cycle, period for 60min. each along X, Y, Z axes
OPERATING ALTITUDE Note.8| 5000 meters
SAFETY & EMC

(Note 9)

| ****

SAFETY STANDARDS

| IEC60601-1, BS EN/EN60601-1, EAC TP TC 004,UL ANSI/AAMI ES60601-1(3.1 version), CAN/CSA-C22 3rd Edition approved ;

Design refer to BS EN/EN60335-1(by request)

ISOLATION LEVEL| Primary-Secondary: 2xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC
ISOLATION RESISTANCE| I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION| Parameter| Standard| Test Level / Note
Conducted| BS EN/EN55011 (CISPR11)| Class B
Radiated| BS EN/EN55011 (CISPR11)| Class B
Harmonic Current| BS EN/EN61000-3-2| Class A
Voltage Flicker| BS EN/EN61000-3-3| —–
EMC IMMUNITY| BS EN/EN55035, BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz ) Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 3, 1KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| BS EN/EN61000-4-11| 100% dip 1 periods, 30% dip 25 periods,

100% interruptions 250 periods


OTHERS

| MTBF| 9314.1K hrs min.       Telcordia SR-332 (Bellcore) ; 1756.2K hrs min.        MIL-HDBK-217F (25℃)
DIMENSION| 4222.320.5mm (LWH) or 1.65″*0.88″0.80″ inch
PACKING| 0.018Kg; 270pcs/5.8Kg/0.94CUFT
NOTE|

  1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
  2. No minimum load required.
  3. 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power
  4. Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1μF & 47μF parallel capacitor.
  5. Tolerance : includes set up tolerance, line regulation and load regulation.
  6. Derating may be needed under low input voltages. Please check the derating curve for more details.
  7. Touch current was measured from primary input to DC output.
  8. The ambient temperature derating of 3.5℃/1000m with fan less models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).
  9. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.” (as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )

Product Liability Disclaimer:For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx

Block Diagram

MEAN-WELL-MFM-10-5-10W-High-Reliable-Green-Medical-On-Board-Type-01
\(3\) Derating Curve

Output Derating VS Input Voltage

Mechanical Specification

Installation Manual
Please refer to : http://www.meanwell.com/manual.html

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

MEAN WELL User Manuals

Related Manuals