MEAN WELL MFM-10-5 10W High Reliable Green Medical On Board Type Owner’s Manual
- June 1, 2024
- MEAN WELL
Table of Contents
MEAN WELL MFM-10-5 10W High Reliable Green Medical On Board Type
Specifications :
- Model: MFM-10 series
- Power Output: 10W
- Size: 1.65 x 0.88 inches
- Safety Approval: ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
- Medical Safety Standard: 2 x MOPP
- Application: Suitable for BF application with appropriate system consideration
- No load power consumption
Product Usage Instructions:
-
Safety Precautions:
Before using the MFM-10 series product, please read the user manual carefully and follow all safety instructions provided. -
Installation:
-
Ensure that the power source matches the product specifications.
-
Connect the product to the appropriate system following the guidelines provided in the manual.
-
Powering On:
-
Once installed correctly, turn on the power source.
-
The product should now be ready for operation.
-
Product Care:
-
Keep the product away from water or any liquid exposure.
-
Do not attempt to open or modify the product casing.
Frequently Asked Questions (FAQ):
-
Q: What is the size of the MFM-10 series product?
A: The product has a compact size of 1.65 x 0.88 inches. -
Q: Is the MFM-10 series suitable for medical applications?
A: Yes, the MFM-10 series is Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1, making it suitable for medical applications with appropriate system consideration.
10W High Reliable Green Medical On Board Type
Features
- 1.65″x0.88″ compact size
- Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
- Suitable for BF application with appropriate system consideration
- No load power consumption<0.075W
- Extremely low leakage current
- Wide operating temp. range -30 ~ +85°C
- EMI class B for class II configuration
- Protections: Short circuit / Overload / Over voltage / Over temperature
- No minimum load required
- 3 years warranty
Applications
- Portable medical device
- Mobile clinical workstation
- Medical computer monitor
- Medical examination instrument
GTIN CODE
MW Search: https://www.meanwell.com/serviceGTIN.aspx
Description
MFM-10 is a 10W high density and small size (4222.320.5mm) AC/DC on board type medical grade power supply series. It features the operation for 80~264 VAC, a low no load power consumption less than 0.075W, a high efficiency up to 84%, Class II (no FG) double insulation, outstanding dissipation, 5G anti- vibration, high EMC performance, 4KVAC isolation, etc. The design observes IEC/BS EN/EN60601-1 and ANSI/AAMI ES60601-1 version three with 2xMOPP level and ultra-low leakage current (<80 m A). It is very suitable for BF (patient contact) type medical device or relevant equipment.
Model Encoding
SPECIFICATION
MODEL| MFM-10-3.3| MFM-10-5| MFM-10-12| MFM-10-15|
MFM-10-24
---|---|---|---|---|---
OUTPUT| DC VOLTAGE| 3.3V| 5V| 12V| 15V| 24V
RATED CURRENT| 2.5A| 2A| 0.85A| 0.67A| 0.42A
CURRENT RANGE Note.2| 0 ~ 2.5A| 0 ~ 2A| 0 ~ 0.85A| 0 ~ 0.67A| 0 ~
0.42A
PEAK CURRENT| 2.75A| 2.2A| 0.94A| 0.74A| 0.46A
RATED POWER| 8.3W| 10W| 10.2W| 10W| 10W
PEAK LOAD(10sec.) **Note.3| 9W| 11W| 11.3W| 11.1W| 11W
RIPPLE & NOISE (max.) **Note.4| 120mVp-p| 100mVp-p| 180mVp-p|
180mVp-p| 200mVp-p
VOLTAGE TOLERANCE Note.5| ±2.5%| ±2.5%| ±2.5%| ±2.5%| ±2.5%
LINE REGULATION| ±0.3%| ±0.3%| ±0.3%| ±0.3%| ±0.3%
LOAD REGULATION| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%
SETUP, RISE TIME| 1000ms, 30ms/230VAC 1000ms, 30ms/115VAC at
full load
HOLD UP TIME (Typ.)| 40ms/230VAC 8ms/115VAC at full load
INPUT| VOLTAGE RANGE Note.6| 80 ~ 264VAC
FREQUENCY RANGE| 47 ~ 440Hz
EFFICIENCY (Typ.)| 78%| 81%| 83%| 83%| 84%
AC CURRENT (Typ.)| 0.3A/115VAC 0.2A/230VAC
INRUSH CURRENT (Typ.)| COLD START 25A/115VAC 45A/230VAC
LEAKAGE CURRENT (max.) Note.7| Touch current <80μA/264VAC
PROTECTION| OVERLOAD| 110% ~ 180% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is
removed
OVER VOLTAGE| 3.8 ~ 5V| 5.75 ~ 6.8V| 13.8 ~ 16.2V| 17.3 ~ 20.3V| 27.6 ~
32.4V
Protection type : Shut off o/p voltage, clamping by zener diode
OVER TEMPERATURE| Protection type : Shut down o/p voltage, recovers
automatically after temperature goes down
ENVIRONMENT| WORKING TEMP.| -30 ~ +85℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +100℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03%/℃ (0 ~ 50℃)
SOLDERING TEMPERATURE| Wave soldering: 265℃,5s (max.); Manual soldering:
390℃,3s (max.)
VIBRATION| 10 ~ 500Hz, 5G 10min./1cycle, period for 60min. each along X,
Y, Z axes
OPERATING ALTITUDE Note.8| 5000 meters
SAFETY & EMC
(Note 9)
| ****
SAFETY STANDARDS
| IEC60601-1, BS EN/EN60601-1, EAC TP TC 004,UL ANSI/AAMI ES60601-1(3.1 version), CAN/CSA-C22 3rd Edition approved ;
Design refer to BS EN/EN60335-1(by request)
ISOLATION LEVEL| Primary-Secondary: 2xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC
ISOLATION RESISTANCE| I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION| Parameter| Standard| Test Level / Note
Conducted| BS EN/EN55011 (CISPR11)| Class B
Radiated| BS EN/EN55011 (CISPR11)| Class B
Harmonic Current| BS EN/EN61000-3-2| Class A
Voltage Flicker| BS EN/EN61000-3-3| —–
EMC IMMUNITY| BS EN/EN55035, BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 3, 1KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| BS EN/EN61000-4-11| 100% dip 1 periods, 30% dip 25
periods,
100% interruptions 250 periods
OTHERS
| MTBF| 9314.1K hrs min. Telcordia SR-332 (Bellcore) ; 1756.2K hrs
min. MIL-HDBK-217F (25℃)
DIMENSION| 4222.320.5mm (LWH) or 1.65″*0.88″0.80″ inch
PACKING| 0.018Kg; 270pcs/5.8Kg/0.94CUFT
NOTE|
- All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
- No minimum load required.
- 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power
- Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1μF & 47μF parallel capacitor.
- Tolerance : includes set up tolerance, line regulation and load regulation.
- Derating may be needed under low input voltages. Please check the derating curve for more details.
- Touch current was measured from primary input to DC output.
- The ambient temperature derating of 3.5℃/1000m with fan less models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).
- The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.” (as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )
※ Product Liability Disclaimer:For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx
Block Diagram
Derating Curve
Output Derating VS Input Voltage
Mechanical Specification
Installation Manual
Please refer to : http://www.meanwell.com/manual.html
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>