Woodpecker Uds-K LED Ultrasonic Piezo Scaler Instruction Manual

June 1, 2024
WOODPECKER

Woodpecker Uds-K LED Ultrasonic Piezo Scaler

Woodpecker-Uds-K-LED-Ultrasonic-Piezo-Scaler-product

Specifications

  • Product Name: UDS-K LED Ultrasonic Piezo Scaler
  • Brand: Woodpecker
  • Model: UDS-K
  • Manufacturer: Guilin Woodpecker Medical Instrument Co., Ltd.
  • Website: www.glwoodpecker.com

Product Usage Instructions

1. Safety Precautions

Before operating the UDS-K LED Ultrasonic Piezo Scaler, please read the manual carefully to ensure safe usage. Follow all safety precautions provided by the manufacturer.

2. Setup

Connect the scaler to the appropriate power source as per the manufacturer’s instructions. Make sure all connections are secure before use.

3. Operating Instructions

Turn on the scaler using the designated power button. Adjust the settings according to the required intensity for the specific dental procedure.

4. Cleaning and Maintenance

After each use, clean the scaler following the recommended cleaning procedures in the manual. Regular maintenance will ensure optimal performance and longevity of the device.

FAQ

  • Q: How do I adjust the intensity of the scaler?
    • A: The intensity of the scaler can be adjusted using the settings provided on the device. Refer to the manual for detailed instructions on adjusting the intensity levels.
  • Q: Can the scaler be used for all dental procedures?
    • A: The UDS-K LED Ultrasonic Piezo Scaler is designed for a variety of dental procedures, but it is recommended to consult with a dental professional for specific treatment recommendations.

The installation and components of the equipment

Instruction

Guilin Woodpecker Medical Instrument Co., Ltd. is a professional manufacturer in researching, developing and producing ultrasonic piezo scalers. The product is mainly used for teeth cleaning and also indispensable equipment for tooth disease prevention and treatment. The product UDS-K LED ultrasonic piezo scaler mainly used for tooth disease prevention and treatment. It contains the following features:

  • Optical handpiece, more convenient for clinical operation.
  • Automatic frequency tracking ensures that the machine always works on the best frequency and performs more steadily.
  • The handpiece is detachable and can be autoclaved to the high temperature of 135°C and pressure of 0.22MPa.
  • Digitally controlled, easy operation and more efficient for scaling.

Components

  • The components of the machine are listed in the packing list.
  • Product performance and structure The Ultrasonic piezo scaler is composed of an electro circuit, water way and an ultrasonic transducer.
  • Scope of application

Ultrasonic piezo scaler UDS-K LED is used for dental calculus elimination.

The main technical specifications

  • Power source input: 220V to 240V- 50Hz/60Hz 150mA
  • Main unit input: 24V~ 50Hz/60Hz 1.3A
  • Output primary tip Vibration excursion: ≤100um
  • Output half_excursion force: <2N
  • Output tip Vibration frequency: 28kHz‡3kHz
  • Output power: 3W to 20W
  • Main unit fuse: TI.6AL 250V
  • Power source fuse: TO.SAL 250V
  • Water pressure: 0.1bar to 5bar (0.01MPa to 0.5MPa)
  • Weight of main unit: 0.75kg
  • Weight of power source: 1.2kg
  • Operating mode: Continuous operation
  • Type of protection against electric shock: Class 1I
  • Degree of protection against electric shock: Type BF equipment
  • Degree of protection against harmful ingress of water: Ordinary equipment (IPX0), protection degree against water (used on the foot switch): IPX1
  • Degree of safety of application in the presence of a Flammable Anaesthetic Mixture with air or with Oxygen or Nitrous Oxide: Equipment not suitable for being used in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous oxide

Working condition:

  • Environment temperature: 5°C to 40°C
  • Relative humidity: ≤ 80%
  • Atmospheric pressure: 70kPa to 106kPa

Installation of the equipment

  • The components of the equipment are shown in picture 1:

Woodpecker-Uds-K-LED-Ultrasonic-Piezo-Scaler-fig \(1\)

The components of the detachable handpiece is shown in picture 2:Woodpecker-
Uds-K-LED-Ultrasonic-Piezo-Scaler-fig \(2\)

Instruction for main components of detachable handpiece:

  • Nipple: the nipple can be removed. You can screw out the nipple and clean the pole with alcohol termly.
  • Seal: the seal can be disassembled and cleaned with alcohol regularly.
  • Handpiece: the main part of the ultrasonic scaler, can be autoclaved under the high temperature and pressure.
  • Symbol: autoclaved (135°C, 0.22MPa)
  • The connector of the cable: connect the handpiece with the water source and power supply of the main unit.
  • Note: Keep the connector dry.

Torque Wrench Instruction

  • a) Brief introduction and illustration (see picture 3) The torque wrench’s structure is designed in special way which can control the strength of the scaling tip installation properly and correctly It also can guarantee the operator screw or unscrew the scaling tip effectively and keep their hands away from being scratched.

  • b) Sterilization condition Sterilize in steam with a temperature of 135°C and pressure 0.22MPa

  • c) Operation

    1. Take the wrench as shown in picture 4;Woodpecker-Uds-K-LED-Ultrasonic-Piezo-Scaler-fig \(4\)
    2. Tip installation: Hold the handpiece,turn the tip toward the clockwise direction with the torque wrench. Turn two more circles after the tip stops, then the tip is installed ;
    3. Uninstall: Hold the handpiece, turn the torque wrench toward a counter-clockwise direction;
    4. Sterilize after each operation;
    5. The torque wrench must be cooled naturally after sterilization to avoid scalding when using next time ; Keep in a cool ,dry and ventilative place and keep it clean .
  • d) Notice
    Forbidden sterilization way as follows:

    1. Braise in liquor;
    2. Dip in iodine, alcohol or glutaraldehyde;
    3. Torrefy in oven or microwave oven.
      Notice: We are not responsible for the damage of the torque wrench for any cases listed in above.

Product function and operation

Operation

  • Open the packing box, make sure that all the parts and accessories are complete according to the packing list.
  • Take the main unit out of the box and put it on a stable plane.
  • Turn the water control knob towards clockwise direction to the max and turn the vibrating intensity control knob towards clockwise direction to a suitable position.
  • Insert the plug of the foot switch to its socket.
  • Connect one end of the water pipe to the water entrance, and the other end to the pure water source.
  • Connect the handpiece: screw on the scaling tip to the handpiece by the torque wrench, then insert the connector of the cable to the handpiece correctly.
  • Get through to the power.
  • Press the power switch of the main unit, then the power indicator shines.

Operation method and function instruction

  • Make the scaler straight to the operator. Before turning on, please turn the vibrating intensity control knob to the minimum and the water control switch to the maximum. (turn three circles towards a clockwise direction from the minimum to the maximum)
  • The normal frequency is as high as 28kHz+ 3kHz. Under normal working conditions, light touch and a certain to-and-fro motion will eliminate the tartar without obvious heating. Overexertion and overstaying are forbidden.
  • Select a suitable scaling tip according to your request, and screw it onto the handpiece tightly by the torque wrench.(as shown in picture 4)
  • Step on the foot switch, the tip begins to vibrate, and the LED lamp on the top of the handpiece shines. Release the foot switch, the LED lamp keeps shining for 10 seconds.
  • Vibrating intensity: Adjust the vibrating intensity as you need, generally turn the knob to the middle grade. Because different patients has different sensitivity and the rigidity of the gingival tartar is not alike too, the vibrating intensity should be adjusted during the clinical treatment.
  • Water adjustment: Step on the foot swich, the tip begins to vibrate, then turn the water control switch to form a fine spray to cool down the handpiece and clean the teeth.
  • The handpiece can be handled in the same gesture as a pen in hand.
  • Make the tip touch the surface of the teeth lightly, don’t use too much pressure, or else the teeth will be hurt and the scaling tip will be damaged.
  • After finishing operation, keep the machine working for 30 seconds with the wate supply, so that the handpiece and the scaling tips can be cleaned.
  • Pull out the handpiece and unscrew the scaling tip, make them be sterilized.
    Notice: Don’t pull out the handpiece when the foot switch is stepped on and the machine is working.

Sterilization and maintenance

  • Sterilization of handpiece
  • The optional autoclave temperature/pressure:
    • a) Sterilization at 121°C/1 bar (0.1MPa).
    • b) Sterilization at 135°C/2.2 bar (0.22MPa).
  • Pull out the handpiece and unscrew the scaling tip after every operation.
  • Pack the handpiece with sterile gauze or bag.
  • Reuse the handpiece after it cools naturally, in case of a scalding hand.
  • Notice:
    • a) Clear the cleaning liquid in the handpiece by compressed air before sterilization.
    • b) Be sure that the scaling tip has been unscrewed from the handpiece, and it can not be sterilized with others.
    • c) Please pay attention to whether the outer of the handpieces is damaged during the treatment or sterilization, don’t smear any protective oil on the surface of handpiece.
    • d) There are two waterproof “O” rings at the end of the handpiece. Please lubricate them with dental lube frequently, as sterilization and repeating pulling and inserting will reduce their life-span.
    • Once they are damaged or worn excessively, please change to new ones.
    • e) The following sterilizing methods are forbidden:
    1. Put the handpiece into any liquid for boiling.
    2. Put the handpiece in disinfectors such as alcohol, iodine or glutaraldehyde.
    3. Put the handpiece in oven or microwave oven for baking.
  • Sterilization of the scaling tip and endo chuck
    • All the scaling tips can be autoclaved to 135°C.
  • The sterilization of the torque wrench
    • The wrench can be sterilized under high temperatures and pressure
  • The cleaning of the torque wrench and scaling tip
    • Both of them can be cleaned with an ultrasonic cleaner.

Troubleshooting and notes

Fault Possible causes Solutions

The scaling tip doesn’t vibrate and there is no water flowing out when stepping on the foot switch.

| The power pipe plug is in loose contact.| Make the plug insert to the socket well.
The foot switch is in loose

contact.

| Insert the foot switch to its

socket tightly.

The fuse of the transformer is broken.| Contact our dealers or us.
The fuse in the main unit

is broken.

| Contact our dealers or us.

The scaling tip doesn’t vibrate but there is water flowing out when stepping on the foot switch.

|

The tip is in loose contact.

| Screw the tip on the handpiece tightly (picture 4).
The connect plug between the handpiece and the circuit board is in loose

contact.

| Contact our dealers or us.

Fault

|

Possible causes

|

Solutions

---|---|---

The scaling tip doesn’t vibrate but there is water flowing out when stepping on the foot switch.

|

Something wrong with the handpiece.

|

Send it to our company to repair.

Something wrong with the cable.|

Contact our dealers or us.

The scaling tip vibrates but there is no water flowing out when stepping on the switch.

| The water control switch is turned off.| Tum on the water control switch [note 1].
There is impurity in the

solenoid valve.

| Contact our dealers or us.
The solenoid valve is abnormal.| Percuss the solenoid valve

by some hard things [note 2].

There is still water flowing out after the power is off.| There is impurity in the solenoid valve.|

Contact our dealers or us.

The handpiece generates heat.| The water control switch is in a low grade.| Tum the water control switch to a higher grade [note 1].

The amount of spouting water is too little.

| The water control switch is in a low grade.| Tum the water control

switch to a higher grade [note 1].

The water pressure is not high enough.| Make the water pressure higher.
The water pipe is blocked.| Clean the water line by multi-function syringe [note2].

Fault

|

Possible causes

|

Solutions

---|---|---

The vibration of the tip becomes weak.

| The tip hasn’t been screwed on to the handpiece tightly.| Screw on the tip to the handpiece tightly (as showed in picture 4).
The tip is loose by because

of vibration.

| Screw on the tip tightly (as

shown in picture 4).

The coupling between the handpiece and the cable

isn’t dry.

|

Dry it by the heated wind.

The tip is damaged [note3

].

| Change a new one.
The vibration is too strong and the vibrating intensity control knob is

malfunction.

|

The vibrating intensity control knob is damaged.

|

Contact our dealers or us.

There is water seeping from the coupling between the handpiece

and the cable.

|

The waterproof “O” ring was damaged.

|

Change a new “O” ring.

If the problem still can’t be solved yet, please contact the local distributors or our company.

  • [Notel] The water control knob can adjust the water volume according to the symbol.
  • [Note2] To clean the water pipe with the multi-function syringe of the dental unit. (as shown in the picture 5) :Woodpecker-Uds-K-LED-Ultrasonic-Piezo-Scaler-fig \(5\)
  1. Snip the water pipe with scissors at a distance of 10cm-15cm from the water entrance
  2. Turn on the power switch, get through to the power.
  3. Connect the multi-function syringe of dental unit to the water pipe.
  4. Step on the foot switch.
  5. Turn on the switch of the multi-function syringe, press the air or water into the water line in the machine, then the impurity blocked in the water line can be eliminated.

[Note3] If the scaling tip has been screwed on tightly and there is fine spray too, the following phenomena show that the scaling tip was damaged:
The vibrating intensity and the water atomization degree become weak obviously. When operating, there is some buzz when the scaling tip is working.

Precaution

  • Notice when using equipment
  • Keep the scaler clean before and after opreation.
  • The handpiece, scaling tip and torque wrench must be sterilized before every treatment.
  • Don’t screw the scaling tip when stepping on the foot switch.
  • The scaling tip must be fastened and there must be fine spray coming out from the tip when operating. the tip or rub the tip.
  • Change a new one when the tip is damaged or worn excessively.
  • Don’t make the tip twist or rub it.
  • Don’t use impure water source. Never replace the distilled water with physidogical saline.
  • If use water source without hydraulic pressure, the water surface should be one meter higher than the head of the patient.
  • Be sure that the connecting end of handpiece and the socket of the connector of cable are complete dried before handpiece installation.
  • Don’t pull the cable emphatically in operating.
  • Please don’t rub or knock the handpiece.
  • After operation, turn off the power, then pull out the plug.
  • As a professional company producing medical instruments’ we are responsible for the safety only when maintenance’ repair and change are done by “Woodpecker” company or our authorized distributors the replacing spare parts belong to ours and operating by the manual.
  • The screw thread of the scaling tips that produced by some other manufacturers may be coarse, rusty and collapsed, this will damage the screw thread of the handpiece irretrievably. Please use “Woodpecker” brand scaling tip.

Contraindication

  • The patient who has hemophilia is not allowed to use this equipment.
  • The patient or doctor who uses heart pacemaker is forbidden to use this equipment.
  • Heart disease patient, pregnant woman and children should be cautious to use the equipment.

Storage and maintenance

  • The equipment should be handled carefully and lightly, be sure that it is far from the vibration, and is installed or keep it in a cool, dry and ventilated place.
  • Don’t put the machine together with the articles that is combustible poisonous,caustic, or explosive.
  • This equipment should be reserved in a room where the relative humidity is 580%, atmospheric pressure is 50kPa to 106kPa, the temperature is – 10°C to +50°C.
  • Please turn oft the power switch and pull out the power line plug when the equipment is not used. If the machine is not used for a long time, please make it get through to the power and water once per month for five minutes.

Transportation

  • Excessive impact and shake should be forbidden in transportation. Lay it carefully and lightly and don’t invert it.
  • Don t put it together with dangerous goods.
  • Avoid solarization and getting wet in rain and snow during transportation.

After service

From the date this equipment has been sold, base on the warranty card, we will repair this equipment free of charge if it has quality problems, please refer to the warranty card for the warranty period.

Symbol instruction

Woodpecker-Uds-K-LED-Ultrasonic-Piezo-Scaler-fig
\(6\) Woodpecker-Uds-K-LED-Ultrasonic-Piezo-
Scaler-fig \(7\)

Environmental protection

There are no harmful factors in our product. You can deal with it based on the local law.

Manufacturer’s right

We reserve the right to change the design of the equipment, the technique, fittings, the instruction manual and the content of the original packing list at any time without notice. If there are some differences between blueprint and real equipment, take the real equipment as the norm.

For technical data, please contact

EC REP

Wellkang Ltd (www.CE-Marking.eu) 29 Harley St.,LONDON,W1G 9QR, UK

Declaration of conformity

The product conforms to the following standards

  • EN 60601-1:2006
  • EN 60601-1-2:2007
  • EN 61000-3-2:2006
  • EN 61000-3-3:2008
  • EN 60601-1-4:1996
  • EN 60601-1-6:2007
  • EN 61205:1994
  • EN ISO 22374:2005
  • EN 62304:2006
  • EN 980:2008
  • EN ISO 9687:1995
  • EN 1041:2008
  • EN ISO 14971:2009
  • EN ISO 7405:2008
  • EN ISO 17664:2004
  • EN ISO 17665-1:2006
  • EN ISO 10993-1:2009
  • EN ISO 10993-5:2009
  • EN ISO 10993-10:2010

EMC – Declaration of conformity

Guidance and manufacturer’s declarationElectromagnetic emissions

The model UDS-K LED is intended for use in the electromagnetic environment specified below. The customer or the user of the model UDS-K LED should ensure that it is used in such an environment.
Emissions test| Compliance| Electromagnetic environmentaudience
RF emissions

CISPR11

|

Group 1

| The model UDS-K LED uses RF energy only for its internal

function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

RF emissions

CISPR11

| Class B| The model UDS-KLED is suitable for used in domestic establishments! and in establishments directly connected to a low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2| Class A
Voltage fluctuations

/ flicker emissions IEC 61000-3-3

|

Complies

Guidance & Declarationelectromagnetic immunity

The models UDS-K LED is intended for use in the electromagnetic environment specified below. The customer or the user of the models UDS-K LED should assure that It is used in such an environment.
Immunity test| IEC 60601

test level

| Compliance

level

| Electromagnetic environment •

11uidance

Electrostatic discharge (ESD)

IEC 61000-4-2

|

±6 kV contact

±8 kV air

|

±6 kV contact

±8 kV air

| Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative

humidity should be at least 30 %.

Electrical fast transient/burst

IEC 61000-4-4

| ±2kV for power supply lines

±1 kV for Input/output lines

| ±2kV for power supply lines

±1kVfor interconnecting

cable

| Mains power quality should be that of a typical commercial or hospital environment.
Surge

IEC 61000-4-5

| ±1 kV line to line

±2 kV line to earth

| ±1 kV line to

line

| Mains power quality should be that of a typical commercial or hospital

environment.

Voltage dips, short interruptions and voltage variations on power supply input lines

IEC 61000-4-11.

|

<5% Ur

(>95% dip in Ur.)

for 0.5 cycle

40%Ur

(60% dip in Ur)

for 5 cycles

70% Ur

(30% dip in Ur)

for 25 cycles

| <5% Ur

(>95% dip in

Ur.)

for 0.5 cycle

40%Ur

(60% dip in Ur) for5 cycles 70% Ur

(30% dip in Ur)

for 25 cycles

| Mains power quality should be that of a typical commercial or hospital environment. If the user of the models UDS-K LED require continued operation during power mains interruptions, it is recommended that the models UDS-K LED be powered from an uninterruptible power supply or a battery.
| <5% Ur

(>95 % dip in Ur)

for5 sec

| <5% Ur

(>95 % dip in

Ur)

for5 sec

|

Power frequency (50/60 Hz)

magnetic field IEC 61000-4-8

|

3Nm

|

3Nm

|

Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

NOTE Ur is the a.c. mains voltage prior to application of the test level.
Guidance & Declaration • Electromagnetic immunity

The models UDS-KLED is intended for use in.the electromagnetic environment specified below. The customer or the user of the models UDS-K LED should assure that it is used in such an environment.
Immunity test| IEC 60601 test level| Compliance level| Electromagnetic environment guidance
|

Vrms

150 kHz to 80 MHz

aVo /m

MHz to 2.5 GHz

| |

Portable and mobile RF communications equipment

| | should be used no closer to any part of the models
| | UDS-K LED including cables, than the recommended
| | separation distance calculated from the equation
| | applicable to the frequency of the transmitter.
| | Recommended separation distance
Conducted RF

IEC 61000-4-6

Radiated RF IEC 61000-4-3

| 3v

3 Vim

| 3V

d=1.2xP 112 BO MHz to 800 MHz

d=2.3xP 112 BOO MHz to 2.5 GHz

| | where Pis the maximum output power rating of the
| | transmitter In watts (JV) according to the transmitter

manufacturer and d Is the recommended separation

| | distance in meters (m).
| | Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,• should
| | be less than the compliance level in each frequency range.b
| | Interference may occur In the vicinity of equipment
| | marked with the following symbol:
| |
NOTE I At 80 MHz end 800 MHz. the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

•     Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the models UDS-K LED is used exceeds the applicable RF compliance level above, the model UDS-K LED should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the models UDS-KLED.

b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.

Recommended separation distances between

portable and mobile RF communications equipment and the models UDS-K LED


The model’s UDS-K LED is intended for use in an electromagnetic environment in which radiated RF disturbances is controlled. The customer or the user of the model UDS-K LED can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the models UDS-KLED as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power

of transmitter

w

| Separation distance according to frequency of transmitter m
150kHz to 80MHz

d=1.2xP 112

| B0MHzto 800MHz d=1.2xP 112| 800MHz to 2,5GHz

d=2.3xP 112

0,01| 0.12| 0.12| 0.23
0,1| 0.38| 0.38| 0.73
1| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where Pis the maximum output power rating of the transmitter in watts rt’/) accordable to the transmitter manufacturer.

NOTE I At 80 MHz and 800 MHz. the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

The device has been tested and homologated in accordance with EN 60601-1-2 for EMC. This does not guarantee in any way that this device will not be effected by electromagnetic interference.Avoid using the device in high electromagnetic environment.

Statement

All rights of modifying the product are reserved to the manufacturer without further notice. The pictures are only for reference. The final interpretation rights belong to GUILIN WOODPECKER MEDICAL INSTRUMENT CO., LTD. The industrial design, inner structure, etc, have claimed for several patents by WOODPECKER, any copy or fake product must take legal responsibility.

Guilin Woodpecker Medical Instrument Co., Ltd.
Information Industrial Park, National High-Tech Zone, Guilin, Guangxi, 541004 P. R. China

  • Tel:
  • Europe Sales Dept.: +86-773-5873196,
    • +86-773-2125222
  • North America, South America &
  • Oceania Sales Dept.:+86-773-5873198,
    • +86-773-2125123
  • Asia & Africa Sales Dept.: +86-773-5855350,
    • +86-773-2125896
  • Fax: +86-773-5822450
  • E-mail: woodpecker@mailgl.cn,
    • sales@g/woodpecker.cn
  • Website: http://www.glwoodpecker.com

EC REP

Wellkang Ltd (www.CE-Marking.eu) 29 Harley St., LONDON,W1G 9QR, UK

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals