FOX INSTRUCTIONS FOR USE
The Food Xplorer (FOX) is an in-vitro diagnostic test based on an Enzyme- Linked Immunosorbent Assay (ELISA) for the semiquantitative determination of specific immunoglobulin G (IgG).
INTENDED USE
The Food Xplorer (FOX) is a semi-quantitative in vitro diagnostic test for the determination of specific immunoglobulin G (IgG) against 286 food antigens. The Food Xplorer is suitable for use with serum or plasma. FOX can be used to support dietary clarification and to monitor dietary compliance. Furthermore, FOX can be used in case of reasonable suspicion of immunologically caused food intolerance. Before starting a diet based on FOX results, it is important to discuss the results with a physician or a dietician. The processing of FOX should always be carried out by professionally trained personnel and/or medical professionals.
SUMMARY AND EXPLANATION OF THE TEST
Type III pattern Food Allergy is a delayed hypersensitivity reaction
presenting a whole-body systemic response involving several organ systems.
Gastrointestinal tract dysfunction is central to the pathogenesis of a variety
of systemic manifestations. Due to the high permeability of the intestinal
wall, food proteins may cross the intestine barrier unnaturally, exposing them
to the immune system and ultimately triggering inflammatory reactions. The
associated symptoms occur in a delayed manner and range from gastrointestinal
symptoms and joint pain to migraines. Delayed-type allergic reactions to food
is mediated by IgG antibodies. By detecting the food-specific IgG antibodies
in human serum, the triggering antigens can be easily identified and the
symptoms alleviated by avoiding the corresponding food.
All major type III food sources are covered by FOX. A complete list of FOX
food extracts can be found at the bottom of this instruction. Link
Important information for the user!
For the correct use of FOX, it is necessary for the user to carefully read and
follow these instructions for use. The manufacturer assumes no liability for
any use of this test system that is not described in this document or for
modifications by the user of the test system.
PRINCIPLE OF THE PROCEDURE
FOX is a solid-phase immunoassay test based on an Enzyme-Linked Immunosorbent Assay (ELISA). Food extracts, which are coupled to nanoparticles, are deposited in a systematic fashion onto a solid phase forming a macroscopic array. First, the particle-bound proteins react with specific IgG that is present in the patient’s sample. After incubation, non-specific IgG is washed off. The procedure continues by adding an enzyme-labeled anti-human IgG detection antibody which forms a complex with the particle-bound specific IgG. After a second washing step, the substrate is added which is converted to an insoluble, colored precipitate by the antibody-bound enzyme. Finally, the enzyme-substrate reaction is stopped by adding a blocking reagent. The amount of precipitate is proportional to the concentration of specific IgG in the patient sample. The lab test procedure is followed by image acquisition and analysis using the ImageXplorer device. The test results are analyzed with MADx’s RAPTOR SERVER Analysis Software and reported in IgG classes.
SHIPMENT AND STORAGE
The shipment of FOX takes place at ambient temperature conditions.
Nevertheless, the kit must be stored immediately upon delivery at 2-8°C.
Stored correctly, FOX and its components can be used until the indicated
expiration date.
Kit reagents are stable for 12 months after opening (at the indicated
storage conditions).
WASTE DISPOSAL
Dispose of the used FOX cartridge and unused kit components with laboratory
chemical waste. Follow all national, state, and local regulations regarding
disposal.
GLOSSARY OF SYMBOLS
| The meaning of the symbols stays the same, regardless of color.
---|---
| Warning (GHS pictogram) Refer to the safety data sheet for details
| Catalogue number Indicates the manufacturer’s catalog number so that the
medical device can be identified.
| Contains sufficient for
| Do not use if the package is damaged Indicates a medical device that should
not be used if the package has been damaged or opened.
| CE mark
| Batch code Indicates the manufacturer’s batch code so that the batch or lot
can be identified.
| Consult instructions for use Indicates the need for the user to consult the
instructions for use.
| Manufacturer Indicates the medical device manufacturer, as defined in EU
Directives 90/385/ EEC, 93/42/EEC and 98/79/EC.
| In vitro, diagnostic medical device Indicates a medical device that is
intended to be used as an in vitro diagnostic medical device.
| Do not re-use Indicates a test that is intended for one use, or for use on a
single patient during a single procedure.
| Temperature limit Indicates the temperature limits to which the test
(reagents) can be safely exposed to.
| Use-by date Indicates the date after which the test (reagents) is not to be
used.
| Important Note
FOX KIT COMPONENTS
Each component (reagent) is stable until the date stated on each individual component’s label. It is not recommended to pool any reagents from different kit lots.
FOX kit components REF 80- 5001-01 | Content | Properties |
---|---|---|
FOX Cartridge | 5 Blisters a 10 FOX for 50 analyses in total. | Ready for use. |
Store at 2-8°C until expiry date.Opened blister is stable for 12 months at
2-8°C.
FOX Sample Diluent| 1 bottle a 30 mL| Ready for use. Store at 2-8°C until
expiry date. Allow reagent to reach room temperature before use. Opened
reagent is stable for 12 months at 2-8°C.
FOX
Washing Solution| 4 x conc. 1 bottle a 250 mL| Store at 2-8°C until expiry
date. Dilute 1 to 4 with purifed water before use (250 ml Washing Solution +
750 ml purified water). Allow reagent to reach room temperature before use.
Opened reagent is stable for 12 months at 2-8°C.
FOX
Detection Antibody| 1 bottle a 30 mL| Ready for use. Store at 2-8°C until
expiry date. Allow reagent to reach room temperature before use. Opened
reagent is stable for 12 months at 2-8°C.
FOX Substrate
Solution| 1 bottle a 30 mL| Ready for use. Store at 2-8°C until the expiry
date. Allow the reagent to reach room temperature before use. The opened
reagent is stable for 12 months at 2-8°C.
| 1 bottle a 10 mL| Ready for use. Store at 2-8°C until the expiry date. Allow
the stop solution to reach room temperature before use. The opened reagent is
stable for 12 months at 2-8°C. May appear as a turbid solution after prolonged
storage. This has no effect on results.
FOX Stop Solution
REQUIRED EQUIPMENT FOR PROCESSING AND ANALYSING FOX
- Arrayholder (optional)
- Lab Rocker (inclination angle 8°, required speed 8 rpm)
- ImageXplorer or MAX 45k analyser
- Incubation chamber (WxDxH 35x25x2 cm)
- RAPTOR SERVER Analysis
- Software PC/Laptop For a fully automated test procedure with FOX, a MAX 45k instrument and a PC/Laptop with an Internet connection is required.
Please refer to 16-IFU-01-EN-xx for details about the automated FOX test procedure using the MAX 45k instrument.
Required equipment, not provided by MADx:
- Purified Water
- Pipettes & tips (100 µL & 1000 µL). Dilution tubes (e.g. 1.5ml Eppendorf Tubes)
Maintenance services according to manufacturer’s instructions.
HANDLING OF FOX ARRAYS
Do not touch the array surface. Any surface defects caused by blunt or sharp objects can interfere with the correct readout of the results. Do not acquire FOX images before the array is completely dry (dry at room temperature).
WARNINGS AND PRECAUTIONS
- It is recommended to wear hand and eye protection as well as lab coats and follow good laboratory practices when preparing and handling reagents and samples.
- In accordance with good laboratory practice, all human source material should be considered potentially infectious and handled with the same precautions as Patient samples. The sample diluent is partially prepared from human blood sources.
- The product was tested non-reactive for Hepatitis B Surface Antigen (HBsAg), antibodies to Hepatitis C (HCV), and antibodies to HIV-1/HIV-2.
- Sample diluent and washing solution contain sodium azide as a preservative and must be handled with care. The safety data sheet is available upon request.
- For in vitro diagnostic use only. Not for internal or external use in humans or animals.
- Only personnel trained in laboratory practice should use this kit.
- Upon arrival, check the kit components for damage. If one of the components is damaged (e.g. buffer bottles), contact MODx (support@macroarraydx.com). Do not use damaged kit components, as their use may lead to poor kit performance.
- Do not use reagents beyond their expiry dates.
- Do not mix reagents from different batches.
ELISA ASSAY PROCEDURE
Preparation:
Preparation of samples : Serum or plasma (heparin, citrate, EDTA) samples
from capillary or venous blood can be used. Blood samples can be collected
using standard procedures. Store samples at 28°C for up to one week. Keep
serum and plasma samples at -20°C for prolonged storage. Shipment of
serum/plasma samples at room temperature is applicable. Always allow samples
to reach room temperature before use.
Preparation of Washing Solution: Pure the content of 1 vial of Washing
Solution into the washing container of the instrument. Fill distilled water up
to the red mark and carefully mix the container several times without
generating foam. Store at 2-8°C until the expiry date if not in use.
Incubation chamber: Close lid for all assay steps to prevent a drop in
humidity.
Parameters of Procedure
- Predilute 10 µL sample with 490 µL FOX Sample Diluent in a separate tube and add 500 µl of the mixture onto the FOX Cartridge
- 500 µL FOX Detection Antibody
- 500 µL FOX Substrate Solution
- 100 µL FOX Stop Solution
- 4500 µL FOX Washing Solution
Assay time is approximately 3 h 30 min.
It is not recommended to run more assays than can be pipetted in 8 min. All
incubations are performed at room temperature, 2026°C.
FOX ELISA ASSAY PROCEDURE
All reagents are to be used at room temperature (20-26°C). The essay must not
be performed in direct sunlight.
Prepare the incubation chamber
Open the incubation chamber and place paper towels on the bottom part. Soak
paper towels with purified water until no dry parts of the paper towels are
visible.
1. Sample incubation
Predilute 10 µL sample with 490 µL FOX Sample Diluent in a separate tube. Take
out the needed number of FOX cartridges and place them into the array
holder(s). Add 500 µl of the prediluted sample to the cartridges. Ensure that
the resulting solution is spread evenly. Place the cartridges onto the lab
rocker and start the serum incubation at 8 rpm for 2 hours. Close the
incubation chamber before starting the lab rocker. After 2 hours, discharge
the patient samples into a collection container. Carefully wipe off droplets
from the cartridge using a paper towel.
Avoid touching the array surface with the paper towel! Avoid any carryover or
cross-contamination of patient samples between individual FOX cartridges!
1a. Washing I
Add 500 µL FOX Washing Solution to each cartridge and incubate on the lab
rocker (at 8 rpm) for 5 minutes. Discharge the washing solution into a
collection container and vigorously tap the cartridges on a stack of dry paper
towels.
Repeat this step 2 more times.
2. Add detection antibody
Add 500 µL of FOX Detection Antibody to each cartridge.
Make sure that the complete array surface is covered by the FOX Detection
Antibody solution.
Place the cartridges into the incubation chamber on the lab rocker and
incubate at 8 rpm for 30 minutes. Discharge the detection antibody solution
into a collection container. Carefully wipe off the remaining droplets from
the cartridges using a paper towel.
2a. Washing II
Add 500 µL FOX Washing Solution to each cartridge and incubate on the lab
rocker at 8 rpm for 5 minutes. Discharge the washing solution into a
collection container and vigorously tap the cartridges on a stack of dry paper
towels.
Repeat this step 4 more times.
3+4. Add substrate solution and stop substrate reaction
Add 500 µL of FOX Substrate Solution to each cartridge. Start a timer with
filling the first cartridge and proceed with the filling of the remaining
cartridges. Make sure that the complete array surface is covered by the
substrate solution and incubate the arrays for 8 minutes without shaking (lab
rocker at 0 rpm and in horizontal position).
After exactly 8 minutes, add 100 µL of the FOX Stop Solution to all
cartridges, starting with the first cartridge to assure that all arrays are
incubated for the same time with the Substrate solution. Carefully agitate to
evenly distribute the stop solution in the array cartridges, after the stop
solution was pipetted onto all arrays. Afterwards discharge substrate/stop
solution from the cartridges and wipe off any remaining droplets from the
cartridges using a paper towel.
Do NOT SHAKE during substrate incubation!!
4a. Washing III
Add 500 µL FOX Washing Solution to each cartridge and incubate on the lab
rocker at 8 rpm for 30 seconds. Discharge the washing solution into a
collection container and vigorously tap the cartridges on a stack of dry paper
towels.
5. Image analysis
After finishing the assay procedure, air dries the arrays at room temperature
until they are completely dry (can take up to 45 min).
Complete drying is essential for the sensitivity of the test. Only completely
dried arrays provide an optimal signal-to-noise ratio.
Finally, the dried arrays are scanned with the ImageXplorer and analyzed with
RAPTOR SERVER software (see details in the RAPTOR SERVER software handbook).
The RAPTOR SERVER Analysis Software is only verified in combination with the
ImageXplorer instrument and with the MAX 45k analyzer, therefore MADx does not
take any responsibility for results, which have been obtained with any other
image capture device (like scanners).
Assay Calibration
Semi-quantitative evaluation of the test results is based on a heterologous
calibration against an IgG reference curve. Curve parameters are determined
from the standard features of the test panel. A curve fit is calculated, and
the resulting equation is applied to transform arbitrary intensity units into
quantitative units. The standard curve for each lot is adjusted by in-house
reference testing against serum samples tested on a reference test system for
specific IgG against several antigens. Lot-specific calibration parameters are
encoded in the barcode. FOX Sigg test results are expressed as µg/ml.
Measuring Range
Specific IgG 5.0 – 50.0 µg/ml
QUALITY CONTROL
Record keeping for each assay
According to good laboratory practice, it is recommended to record the lot
numbers of all reagents used.
Control Specimens
According to good laboratory practice, it is recommended that quality control
samples are included within defined intervals.
DATA ANALYSIS
For the image analysis of processed arrays, the ImageXplorer or MAX 45k
analyzer has to be used. FOX images are automatically analyzed using MADx’s
RAPTOR SERVER Analysis Software and a report is generated summarising the
results for the user.
RESULTS
FOX is a semi-quantitative ELISA method for specific IgG. Specific IgG
antibodies are expressed as IgG response units (µg/ml). Marx’s RAPTOR SERVER
Analysis Software automatically calculates and reports Sigg results semi-
quantitatively as classes (low, intermediate, and highly elevated).
LIMITATIONS OF THE PROCEDURE
The IgG concentrations determined with the FOX test system show the degree of
sensitization of the patient to the food antigens investigated. A correlation
between the level of the determined IgG concentrations and the occurrence or
severity of clinical symptoms cannot be deduced.
The results obtained must always be interpreted in combination with the full
clinical symptoms. A definitive clinical diagnosis should only be made in
conjunction with all available clinical findings by medical professionals and
shall not be based on the results of a single diagnostic method only.
False-positive test results may occur due to the cross-reactivity of the
tested antigen with epitopes of other antigens. Antigenic epitopes may be lost
during the production of food extracts, which may lead to false negative
results. IgG antibodies to food antigens that occur only during industrial
preparation, food preparation or digestion may not be detectable in the
patient sample.
EXPECTED VALUES
The close association between antigen-specific IgG antibody levels and type
III food allergy symptoms is well known and is described in the literature
(see section “References” below). Each sensitized patient will show an
individual IgG profile when tested with FOX. The IgG response with samples
from healthy individuals will be below 10.0 µg/ml for single-food antigens
when tested with FOX. Good laboratory practice recommends that each laboratory
establishes its own range of expected values.
PERFORMANCE CHARACTERISTICS
All performance characteristics of the FOX test were determined according to
the current CLSI guidelines.
Precision (lot-lot variation)
The lot-to-lot variation was determined on three cartridge lots in three
separate runs. Multi-sensitized samples were included in the study. The study
comprised 867 antigen/sample combinations covering 121 individual antigens
over the entire measuring range.
Concentration µg/ml | Intra CV% | Inter CV% | Total CV% |
---|---|---|---|
10.0 – 19.9 | 6.9 | 11.2 | 9.1 |
≥20.0 | 3.1 | 5.5 | 4.3 |
≥10.0 | 4.8 | 7.9 | 6.3 |
Repeatability (within-run precision)
In the repeatability study, multi-sensitized samples were tested 10 times by
the same operator on different days. The study comprised 862 antigen/sample
combinations covering 115 individual antigens over the entire measuring range.
Concentration – µg/ml | Total CV% |
---|---|
10.0 – 19.9 | 11.3 |
≥20.0 | 5.4 |
≥10.0 | 7.2 |
Analytical Sensitivity
Analytical sensitivity is a measurement of the accuracy of a test at low
concentrations of the analyte. The limit of detection (LOD) of the FOX test is
below 5.0 µg/ml for all food extracts.
Interference
There is no detectable interference with bilirubin, triglycerides and
hemoglobin at normal physiological concentrations.
WARRANTY
The herein presented performance data were obtained using the procedure outlined in this Instructions for Use. Any change or modification in the procedure may affect the results and MacroArray Diagnostics disclaims all warranties expressed (including the implied warranty of merchantability and fitness for use) in such an event. Consequently, MacroArray Diagnostics and its authorized distributors shall not be liable for damages indirect or consequential in such an event.
Antigen list FOX
Vegetables: Artichoke, Arugula, Avocado, Bamboo sprouts, Broccoli,
Brussels sprouts, Cabbage, Caper, Carrot, Cauliflower, Celery bulb, Celery
stalk, Chard, Chicorèe, Chinese cabbage, Chives, Cucumber, Eggplant, Endive,
Fennel, Garlic, Green cabbage, Horseradish, Kiwano, Kohlrabi, Lamb’s lettuce,
Leek, Nettle leaves, Olive, Onion, Parsnip, Pok-Choi, Potato, Pumpkin
(Butternut), Pumpkin (Hokkaido), Radicchio, Radish, Red beet, Red cabbage,
Romanesco, Savoy, Shallot, Spinach, Sweet potato, Tomato, Turnip, Watercress,
White Asparagus, White cabbage, Wild garlic, Zucchini
Fish & Seafood: Abalone, Atlantic cod, Atlantic herring, Atlantic
redfish, Carp, Caviar, Cockle, Common mussel, Crab, Eel, European anchovy,
European pilchard, European plaice, Gilt-head bream, Haddock, Hake, Lobster,
Mackerel, Monkfish, Noble crayfish, Northern pike, Northern prawn, Octopus,
Oyster, Razor shell, Salmon, Scallop, Sepia, Shrimp mix, Sole, Squid,
Swordfish, Thornback Ray, Trout, Tuna, Turbot, Venus clam
Fruits: Apple, Apricot, Banana, Blackberry, Blueberry, Cherry, Cranberry,
Date, Elderberry, Fig, Gooseberry, Grape, Grapefruit, Kiwi, Lemon, Lime,
Lychee, Mango, Melon, Mulberry, Nectarine, Orange, Papaya, Passion fruit,
Peach, Pear, Physalis, Pineapple, Plum, Pomegranate, Raisin, Raspberry, Red
currant, Strawberry, Tangerine, Watermelon
Spices: Anise, Basil, Bay leaf, Caraway, Cardamom, Cayenne pepper, Chili
(red), Cinnamon, Clove, Coriander, Cumin, Curry, Dill, Fenugreek, Ginger,
Juniper berry, Lemongrass, Majoram, Mint, Mustard, Nutmeg, Oregano, Paprika,
Parsley, Pepper (black/white/green/red/yellow), Rosemary, Sage, Tarragon,
Thyme, Turmeric, Vanilla
Cereals & Seeds: Amaranth, Barley, Buckwheat, Chickpea, Corn, Durum,
Einkorn, Emmer, Gluten, Hempseed, Linseed, Lupineseed, Malt (barley), Millet,
Oat, Pine nut, Polish wheat, Poppyseed, Pumpkin seed, Quinoa, Rapeseed, Rice,
Rye, Sesame, Spelt, Sunflower, Wheat, Wheat bran, Wheat gliadin, Wheatgrass
Novel Foods: Almond milk, Aloe, Aronia, Baobab, Chia seed, Chlorella,
Dandelion root, Ginkgo, Ginseng, Greater burdock root, Guarana, House cricket,
Maca root, Mealworm, Migratory locust, Nori, Safflower oil, Spirulina,
Tapioca, Wakame, Yacòn root
Egg & Milk: Buffalo milk, Buttermilk, Camel’s milk, Camembert, Cottage
cheese, Cow’s milk, Egg white, Egg yolk, Emmental, Goat cheese, Goat milk,
Gouda, Mozzarella, Parmesan, Quail egg, Sheep cheese, Sheep milk
Meat: Beef, Boar, Chicken, Duck, Goat, Horse, Lamb, Ostrich, Pork,
Rabbit, Stag, Turkey, Veal, Venison
Nuts: Almond, Brazil nut, Cashew, Coconut, Coconut milk, Hazelnut, Kola
nut, Macadamia, Pecan nut, Pistachio, Sweet chestnut, Tigernut, Walnut
Coffee & Tea: Chamomile, Cocoa, Coffee, Hibiscus, Jasmine, Moringa,
Peppermint, Tee (black), Tee (green)
Legumes: Green bean, Lentil, Mung bean, Pea, Peanut, Soy, Sugar pea,
Tamarind, White bean
Edible Mushrooms: Boletus, Chanterelle, Enoki, French horn mushroom,
Oyster mushroom, White mushroom
Others : Agar agar, Aspergillus niger, Baker’s yeast, Brewer’s yeast,
Elderflower, Honey, Hops, M-Transglutaminase (meat glue), Cane sugar, Cross-
reactive Carbohydrate Determinants
References
- Clinical relevance of IgG antibodies against food antigens in Crohn’s disease: a double-blind cross-over diet intervention study. Bentz, S et al. 2010; Zurich Open Repository and Archive
- Food Exclusion Based on IgG Antibodies Alleviates Symptoms in Ulcerative Colitis: A Prospective Study. L Jian et al. 2018; Inflammatory Bowel Diseases
- Serological investigation of IgG and IgE antibodies against food antigens in patients with inflammatory bowel disease. HY Wang et al. 2019; World Journal of Clinical Cases
- The Value of Eliminating Foods According to Food-specific Immunoglobulin G Antibodies in Irritable Bowel Syndrome with Diarrhoea. H Guo et al. 2012; The Journal of International Medical Research
- Prevalence of IgG-mediated food intolerance among patients with allergic symptoms. Shakoor et al. 2016; Annals of Saudi medicine
- The Clinical Application Value of Multiple Combination Food Intolerance Testing. S Lin et al. 2018; Iranian Journal of Public Health
- Chronic Food Antigen-specific IgG-mediated Hypersensitivity Reaction as A Risk Factor for Adolescent Depressive Disorder. R Tao et al. 2019; Genomics, Proteomics & Bioinformatics
- Diet restriction in migraine, based on IgG against foods: A clinical double-blind, randomized, cross-over trial. Alpay et al. 2010; Cephalalgia
The connection between food intake, elevated IgG levels, and chronic disorders has been described in peer-reviewed publications and case studies. Nonetheless, this connection is still debated in the scientific community and a consensus has not been reached so far.
© Copyright by MacroArray Diagnostics
MacroArray Diagnostics GmbH (MADx)
Lemböckgasse 59/Top 4
1230 Wien Austria +43 (0)1 865 2573
support@macroarraydx.com
Version number: 80-IFU-01-EN-03
Released: 05-2021
11.05.21, 10:04