OMRON HBP-1300 Professional Blood Pressure Monitor Instruction Manual
- May 15, 2024
- Omron
Table of Contents
- HBP-1300 Professional Blood Pressure Monitor
- Specifications:
- Product Usage Instructions:
- 1. Components of the Product:
- 2. Features of the Product:
- 3. Installing the Optional Battery Pack:
- 4. Connecting the AC Adapter:
- 5. Non-Invasive Blood Pressure Measurement:
- 6. Maintenance:
- Q: Can this blood pressure monitor be used for personal home
- Q: How often should the cuff be replaced?
HBP-1300 Professional Blood Pressure Monitor
“`html
Specifications:
-
Model: HBP-1300
-
Intended Use: Professional Blood Pressure Monitor
-
Power Source: Optional Battery Pack or AC Adapter
-
Measurement Method: Non-Invasive Blood Pressure (NIBP)
Measurement -
Manufacturer: OMRON HEALTHCARE
Product Usage Instructions:
1. Components of the Product:
The blood pressure monitor includes the main unit, cuff,
optional accessories, battery pack, and AC adapter.
2. Features of the Product:
The product is designed for professional use and offers
non-invasive blood pressure measurement. It comes with various
features and functions for accurate readings.
3. Installing the Optional Battery Pack:
To install the optional battery pack, follow the instructions
provided in the manual to ensure proper functioning of the monitor
even when not connected to a power source.
4. Connecting the AC Adapter:
If using the AC adapter, connect it to the designated port on
the monitor to power the device during use.
5. Non-Invasive Blood Pressure Measurement:
Follow these steps for accurate blood pressure measurement:
-
Cuff Selection and Connection: Choose the appropriate cuff size
and connect it to the monitor. -
Applying the Cuff to Patient: Place the cuff on the patient’s
arm as instructed. -
Taking the Measurement: Follow the manual to take measurements
in manual or auscultation mode. -
Stopping the Measurement: End the measurement process when
completed. -
Displaying Last Reading: View and record the last reading
displayed on the monitor.
6. Maintenance:
Refer to the maintenance section of the manual for guidelines on
cleaning, storage, and upkeep of the blood pressure monitor to
ensure longevity and accuracy of readings.
FAQ:
Q: Can this blood pressure monitor be used for personal home
use?
A: No, this blood pressure monitor is intended for professional
use only and should be operated by trained healthcare
practitioners.
Q: How often should the cuff be replaced?
A: It is recommended to replace the cuff according to the
manufacturer’s guidelines, typically when signs of wear or damage
are visible.
“`
Professional Blood Pressure Monitor
HBP-1300
Instruction Manual
Thank you for purchasing this OMRON Professional Blood Pressure Monitor.
Please competely read this Instruction Manual before using the monitor for the
first time. Read this manual to ensure the safe and accurate use of the
monitor.
For Professional Use
KEEP THIS INSTRUCTION MANUAL NEAR THE BLOOD PRESSURE MONITOR FOR FUTURE
REFERENCE
Table of Contents
Introduction
Intended Use………………………………………. 1 Exemptions ………………………………………… 1 Notes on Safety
………………………………….. 2 Warnings and Cautions………………………… 3
Using the Unit
Components of the Product ………………… 10 Optional Accessories …………………………. 10
Features of the Product ……………………….11 Features / Functions of Unit………………… 12
Installing the Optional Battery Pack ……… 14 Connecting the AC Adapter………………….
15
System Settings
Menu Mode ………………………………………. 16 Utility Mode ………………………………………. 17
Non-Invasive Blood Pressure (NIBP) Measurement
Non-Invasive Pressure Measurement Principles …..18 Cuff Selection and
Connection ……………. 19 Applying the Cuff to Patient ………………… 20 Taking the
Measurement in “Manual Mode” ….21 Taking the Measurement in “Auscultation
Mode” …22 Stopping the Measurement…………………. 23 Displaying Last Reading
…………………….. 23
Maintenance
Maintenance Inspection and Safety Management …24 Managing Supplies
……………………………. 24 Device Maintenance ………………………….. 24 Accessory
Care…………………………………. 24 Check before Use ……………………………… 25 Troubleshooting
………………………………… 26 List of Error Codes …………………………….. 28 Disposal
…………………………………………… 29 Limited Warranty……………………………….. 30
Specifications
Factory Default Settings……………………… 31 Specifications……………………………………. 31
Manufacturer’s Declaration …………………. 33
Caution: Federal (USA) law restricts this device to sale by or on the order of a physician or licensed healthcare practitioner
Introduction
Introduction
Intended Use
Medical Purpose The device is a digital monitor intended for use in measuring
blood pressure and pulse rate in adult and pediatric patient population with
arm circumference ranging from 5 inches to 20 inches (from 12 cm to 50 cm).
Intended User This device should be used by a medical professional. Patient
Population This device is intended for use on adults and children of age
3-years and older. Environment The instrument is designed for use in
physicians’ offices, hospitals, clinics and other medical facilities.
Measurement Parameter Non-Invasive Blood Pressure Pulse rate Precautions for
Use Warnings and cautions described in the instruction manual should be
observed at all times.
Exemptions
Liability and Warranty OMRON HEALTHCARE does not accept liability and the
warranty becomes void under the following circumstances: 1. Failures, damage,
or injuries due to maintenance or repair work performed by persons or
a repair facility not authorized by OMRON HEALTHCARE. 2. Failures, damage, or
injuries when use and / or operation of this device is adversely
affected by a device not manufactured by OMRON HEALTHCARE. 3. Failures,
damage, or injuries caused by failure to odserve the safety instructions and
operational procedures provided in this manual. 4. Use and / or operation of
this unit in conditions that do not conform to the product usage
conditions indicated in this manual. 5. Failures, damage, or injuries due to
natural catastrophes, such as fire, earthquake,
flooding or other natural disasters. HBP-1300 Instruction Manual 1. The
contents of this manual are subject to change without notice. 2. Considerable
care has been taken in the preparation of this manual. If you suspect there
may be an error in the instruction manual, please contact OMRON HEALTHCARE by
calling 1-877-216-1336. 3. Unauthorized reproduction or copying any or this
entire instruction manual without the permission of OMRON HEALTHCARE is
prohibited by copyright. 4. Product brand names shown in this maual are
trademarks or registered trademarks of OMRON HEALTHCARE.
1
Notes on Safety
The warning signs and symbol examples indicated below are intended to ensure
safe use of the product and prevent damage and injury to you and others. The
signs and symbols are explained below.
Safety Symbols used in this Instruction Manual
Warning Caution
Indicates the matters in which death or severe bodily damage may arise as a
result of incorrect handling.
Indicates the matters in which bodily harm or material damage may arise as a
result of incorrect handling.
General Information Note: Indicates general information to keep in mind when using the unit and other useful information.
2
Introduction
Warnings and Cautions Usage warnings and cautions
Setup
Warning
Do not use the cuff or AC adapter to lift the unit, it can also cause the unit
to malfunction. If the unit has broken down, contact OMRON HEALTHCARE Customer
Service. Do not use in combination with a hyperbaric oxygen therapy device, or
in an environment where combustible gas may be generated. Do not use in
combination with magnetic resonance imaging (MRI) equipment. If MRI is to be
performed, remove cuff connected to the unit from the patient. Do not install
the unit in the following locations: – Locations subject to vibration such as
ambulances and emergency helicopters. – A location where there is gas or
flame. – A location where there is water or steam. – A location where
chemicals are stored. Do not use at extremely high temperature, high humidity,
or high altitude. Use only within the required ambient conditions. Do not
subject the unit to intense shock. Do not place heavy objects on the AC
adapter cable, or allow the unit to sit on the cord. Clinical testing has not
been conducted on newborn infants and pregnant women. Do not use on newborn
infants and pregnant women. Do not plug in or unplug the AC adapter with wet
hands.
Caution
Do not install the unit in the following locations: – Locations with dust,
salt, or sulfur. – Locations directly exposed to sunlight for extended periods
of time (in particular, do not leave in
direct sunlight or near a source of ultraviolet light for extended periods, as
ultraviolet light will cause deterioration of the LCD). – Locations subject to
vibration or shock. – Near heaters. Do not use the unit near large equipment
that uses a switching relay for power ON/OFF.
3
Before use / during use
Warning
The unit complies with the EMC standard (IEC60601-1-2). As such, it can be
used simultaneously with multiple medical instruments. However, if instruments
that generate noise such as an electric scalpel or a microwave therapy device
are near the unit, check the operation of the unit during and after use of
these instruments. If an error occurs or a measurement result is questionable,
check the vital signs of the patient by auscultation or palpation. Avoid
relying solely on the measurement results of the unit when judging the
patient’s condition. Only trained healthcare providers should use this device.
Do not allow patients to operate this device. Properly connect the connectors
and AC adapter cable. Do not place objects or liquids on top of this unit.
Check the following before using the unit: – Make sure the AC adapter cable is
not damaged (wires are not exposed or broken), and the
connections are firm. For the AC adapter connected to the unit, supplies, and
optional devices, use only the standard accessories or OMRON-specified
products. Do not use in a location with moisture, or a location where water
may splash on the unit. This unit is intended for use in a physician’s office.
Do not use the unit if it emits smoke, an abnormal odor, or abnormal noise. Do
not bring cellular telephones or transceivers into the room where the unit is
installed or being used. Do not connect multiple monitors to the same patient.
Do not connect the unit to a power outlet that is controlled by a wall switch.
4
Introduction
Caution
Before using the unit, verify that none of the following apply to the patient:
– Poor peripheral circulation, noticeably low blood pressure, or low body
temperature (there will be
low blood flow to the measurement position) – The patient uses an artificial
heart and lung (there will be no pulse) – An SpO2 sensor and the cuff are
attached to the same arm – The patient has an aneurysm – The patient has
frequent arrhythmia – Body motions such as convulsions, arterial pulsations,
or trembling (cardiac massage in progress,
minute continuous vibrations, rheumatism, etc.) Before use, visually inspect
the unit to make sure there are no deformations due to falling, and that there
is no dirt or moisture on the unit. When the unit has not been used for an
extended period of time, always verify that it operates normally and safely
before use. Do not use in a location where the unit may easily fall. In the
event that the unit falls, verify that it operates normally and safely.
5
Cleaning
Warning
When cleaning the unit, turn off the power and disconnect the AC adapter from
the unit. After cleaning the unit, make sure it is completely dry before
connecting to a power outlet. Do not spray, pour, or spill liquids into or
onto the unit, accessories, connectors, buttons, or openings in the housing.
Caution
Do not use thinner, benzene, or other solvents to clean the unit. Do not
sterilize by autoclave or gas sterilization (EOG, formaldehyde gas, high-
concentration ozone, etc.). If using an antiseptic solution for cleaning,
follow the instructions of the manufacturer. Clean the unit regularly.
Maintenance and inspection
Warning
To use the unity safely and correctly, always inspect the unit when starting
work. Unauthorized modification is prohibited by law. Do not attempt to
disassemble or modify the unit.
6
Introduction
Warnings and cautions for safe measurement Optional rechargeable battery
Warning
If battery fluid comes in contact with the eye, immediately flush with copious
amounts of water. Do not rub. Seek medical attention immediately. Do not use
the optional battery pack in any other device besides this unit. Do not throw
into flame, disassemble, or heat. Always disconnect the AC adapter from the
unit before removing or installing a battery. If the unit will not be used for
a month or longer, remove the battery from the unit and store. Charge the
battery once every 6 months. (Storage conditions for the battery are a
temperature of -20 to 30°C (-4 to 86°F) and a humidity of 65 ±20%.) Before
use, always charge the battery. Do not attempt to disassemble or modify the
battery. Do not apply pressure to and deform the battery. Do not throw, pound,
drop, bend, or hit the battery. The battery has positive/negative polarity. If
the battery does not connect well to the unit, do not forcibly connect it. Do
not connect the positive and negative terminals of the battery with a wire or
other metal object. Do not carry or store a battery with metal necklaces,
hairpins, or other metal objects. Use only the specified type of battery.
Caution
Do not touch the positive and negative terminals of the optional battery pack
with a wire or other metal object. If battery fluid comes into contact with
the skin or clothes, immediately rinse with water.
7
Non-Invasive Blood Pressure (NIBP) measurement
Warning
If a cuff is used on a patient with an infection, treat the cuff as medical
waste, or sterilize before reuse. If frequently performing NIBP measurement
using a cuff over an extended period of time, periodically check the patient’s
circulation. In addition, wrap the cuff as indicated in the cautionary points
in this manual. Do not connect the NIBP cuff or cuff joint to a luer lock
adapter. Do not bend cuff tube during inflation and deflation, particularly
after a change of body position. Do not wrap the cuff on the following parts:
– An upper arm on which intravenous drip or a blood transfusion is being
performed. – An upper arm on which an SpO2 sensor, IBP catheter, or other
instrument is attached. – An upper arm with a shunt for hemodialysis If
measuring blood pressure with the cuff wrapped on the arm on the side of the
body where a mastectomy was performed, check the patient’s condition.
Caution
NIBP measurement should be performed on the upper arm. During NIBP, stop
excessive body movement by the patient and minimize trembling. If a doctor has
indicated that the patient has hemorrhagic diathesis or hypercoagulability,
check the condition of the arm after measurement. Use the appropriate cuff
size to ensure correct measurements. If too large a cuff is used, the measured
blood pressure value tends to be lower than the actual blood pressure. If too
small a cuff is used, the measured blood pressure value tends to be higher.
Before and during measurement, verify that none of the following apply to the
patient: – An inappropriate cuff size is used. – The part where the cuff is
wrapped is at a different height than the heart.
(A difference of 4 inches (10 cm) in height may cause a variation in the blood
pressure value of up to 7 or 8 mmHg.) – Body movement or conversing during
measurement. – Cuff wrapped over thick clothing. – Pressure on the arm due to
a rolled up sleeve. In the case of a cuff for adults, the cuff should be
wrapped to a tightness that allows two fingers to be inserted in between the
cuff and the arm. The accuracy of a flashing measurement value that is out of
the measurement range cannot be guaranteed. Always check the patient’s
condition before deciding what steps to take. Do not use the cuff if it is
damaged or has holes. Only OMRON GS CUFF can be used with this device.
8
Introduction
Note:
Setup · Read and understand the manual for each optional accessory. This
manual does not contain
cautionary information for optional accessory. · Exercise caution with the
cables and arrange so that the patient does not become entangled or
bound. Before use / during use · Check the following after turning on the
power: – No smoke, abnormal odor, or abnormal sound is emitted. – Press each
button and verify that it operates. – For functions that cause icons to light
or flash, verify that the icons light or flash. – Measurement can be performed
normally, and measurement error is within the tolerance value. · If the screen
is not displayed normally, do not use the unit. · When recycling or disposing
of parts (including batteries) of the unit, follow local government rules
and regulations. Cleaning · For cleaning, see page 24. Optional rechargeable
battery · To prevent accidents, keep batteries out of reach of infants and
small children. · If you sense that something is wrong with a battery,
immediately move it to a safe location and contact
the administrator responsible for the unit or OMRON HEALTHCARE Customer
Service. · If the battery voltage is low, operation by battery may not be
possible. Non-Invasive Blood Pressure (NIBP) measurement · If the patient has
acute inflammation, a pyogenic ailment, or an external wound at the location
where
the cuff is to be wrapped, follow the instructions of the doctor. · Non-
Invasive Blood Pressure measurement (NIBP) is performed by compressing the
upper arm.
Some people may experience intense pain, or transient spotting caused by
subcutaneous hemorrhaging (bruising) may appear. The spotting will disappear
with time; however, it may be appropriate to inform patients for whom this may
be a concern that spotting sometimes occurs, and if necessary, refrain from
measurement. · To measure correctly, it is recommended that the patient relax
and not talk during measurement. · To measure correctly, it is recommended
that the patient rest quietly for 5 minutes before measurement.
9
Using the Unit
Components of the Product
Before using the unit, make sure that no accessories are missing and that the
unit and accessories are not damaged. If an accessory is missing or there is
damage, please contact OMRON HEALTHCARE Customer Service.
Main unit
Standard accessories
AC adapter
GS CUFF L and M Instruction Manual
Optional Accessories
Wall mount kit
WMKIT-1300 (HXA-WMKIT-1300)
Battery pack
BAT-2000 (HXA-BAT-2000)
AC adapter
AC-1300Z (60120H1040SW-Z)
10
GS CUFF XL
GCUFF-XL (HXA-GCUFF-XLLB)
GS CUFF L
GCUFF-L (HXA-GCUFF-LLB)
GS CUFF M
GCUFF-M (HXA-GCUFF-MLB)
GS CUFF S
GCUFF-S (HXA-GCUFF-SLB)
Using the Unit
GS CUFF SS
GCUFF-SS (HXA-GCUFF-SSLB)
CUFF HOSE (3.5 m)
HOSE15ZZ (A015ZZ)
CUFF HOSE (1.5m)
HOSE16ZZ (A016ZZ)
Caution
Only OMRON GS CUFF can be used with this device.
Features of the Product
The OMRON HBP-1300 is an affordable professional blood pressure unit that is
clinically proven accurate and provides fast, reliable results and is easy to
use.
Key features, benefits, look
· 5 cuffs available – (SS: 5 to 7 inches, S: 7 to 9 inches, M: 9 to 13 inches,
L: 13 to 17 inches, XL: 17 to 20 inches) (5 to 20 inches arm circumference
range)
· Designed to be used on a table or mounted on the wall · Motion stop function
(When body movement is detected, the device stops deflation for 5 seconds.) ·
Irregular pulse Indicator – Helps identify changes in heart rate, rhythm, or
pulse which may be caused
by heart disease or other serious health issues. · Inflation Pressure Setting
– 4 options: Auto, 220 mmHg, 250 mmHg, 280 mmHg · Last reading display
function · Auto-Off · Custom Rechargeable battery · Large, easy-to-read
backlight display
11
Features / Functions of Unit Front and back of unit
4
6 5
23
1
7
Power ON / Measure blood pressure
Press when the power is off to turn on the power and start
[START/STOP] button blood pressure measurement. During inflation or
1
(Power ON/OFF)
measurement, press to stop.
Power OFF
Hold down at least 3 seconds to turn off the power.
Press to enter “Menu Mode” and configure various settings.
2
[MODE] button
If held down at least 3 seconds when a measurement result is displayed, the measured data is cleared without being
stored in memory.
Displays the system settings (page 16) and the last reading
3
[ ] [ ] button
(page 23). When using “Auscultation Mode”, inflates and
deflates the cuff.
4
Alarm lamp
Lights up or flashes when an alarm occurs (page 28).
5
NIBP connector
Connects the cuff tube.
6
DC jack
Connects the AC adapter.
7
Battery cover
Open to install or replace the battery.
Meaning of the Symbols
Symbol Description
Symbol Description
This shows the type BF device.
Refer to manual.
MEDICAL EQUIPMENT WITH RESPECT TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL60601-1 and CAN/CSA C22.2 No.601.1
12
Using the Unit
LCD Display
1
2
4
5
3
6
7
8
9
10
1
SYS
Displays systolic blood pressure.
2
DIA
Displays diastolic blood pressure.
3
Pulse
Displays the pulse rate.
4
Pulse
Flashes in synchronization with the pulse during
synchronization icon measurement.
Lights in the measurement result display and memory
5
Irregular pulse wave icon
display if the pulse wave interval was irregular or there
was body movement during measurement (page 23).
6
Memory icon
Lights while the previous data are being displayed (page 23).
7
Charge icon*
Flashes during charging. Solid colored light displays
when charging is finished.
8
Battery charge level icon*
Displays how much charge the battery has (page 14).
9
Auscultation icon
Lights when “Auscultation Mode” is ON (page 16).
10
Inflation setting icon The set initial Inflation pressure value lights (page 16).
- Only when the optional battery is installed.
13
Installing the Optional Battery Pack
Warning
· If battery fluid comes in contact with the eye, immediately flush with
copious amounts of water. Do not rub. Seek medical attention immediately.
· Do not use the optional battery pack in any other device besides this unit.
Do not throw into flame, disassemble, or heat.
Caution
Do not short the positive and negative terminals of the optional battery pack
with a wire or other metal object. If battery fluid comes into contact with
the skin or clothes, immediately rinse with water.
1. Make sure the AC adapter has been disconnected.
2. Remove the two screws from the top of the rear unit cover, and remove the
battery cover.
3. Connect the battery pack to the connector on the monitor, and insert the
battery pack into the compartment so that the triangle mark on the battery
pack is aligned with the triangle mark on the monitor.
4. Replace the battery cover and secure with screws.
Take care not to pinch the wires when replacing the cover.
5. Connect the AC adapter to the unit and charge the battery pack.
When using the battery pack for the first time, be sure to fully charge (about
4 hours) before using.
Battery life
· About 300 measurements are possible with one charge. · A general guideline
for replacement of the battery pack is about one year, however, the usage time
per charge may grow shorter depending on usage conditions. If you find that
the usage time after
each charge has grown shorter and the
icon frequently appears, replace the battery pack.
Charging time
· Charging automatically starts after the AC adapter is connected. When a new battery pack or a battery pack that has not been used for a long time is used, some time may elapse before charging starts.
· The
icon flashes during charging.
· Charging is completed in about 4 hours.
Low battery
When the
icon starts flashing, promptly charge the battery pack.
When the icon changes to
, the battery is too low for blood pressure measurement. Charge the
battery.
Auto Power Off
· When the unit is powered by the optional battery pack, the power
automatically turns off after the set time elapses if you forget to turn off
the power.
· When the unit is used with the AC adapter connected, the “Auto Power Off”
setting does not operate.
14
Battery charging
State
LCD contents and operations
Charging
The icon is flashed.
Fully charged (Charging completed)
Problem with the battery
The icon is lit. Error message is displayed.
Battery level
Battery level Fully charged Battery level is 20%.
Battery level is 5%.
LCD contents and operations
The icon is lit. The unit can be used.
The icon is flashed (E40 error not displayed). The unit can be used.
The icon is flashed (E40 error displayed). The unit cannot be used. If the
unit is being continuously used, the power will be automatically turned off in
30 seconds.
Icon –
Icon
Connecting the AC Adapter
AC power
Verify that the power outlet supplies the specified voltage and frequency (100
– 240 V AC, 50/60 Hz).
Using the Unit
Connect the AC adapter to the DC jack on the unit and the power outlet.
Note:
When the optional battery pack is installed
If there is no problem with the following.
– AC adapter
– DC jack
– Power outlet
– Battery
And the
icon does not flash, contact OMRON HEALTHCARE Customer Service.
15
System Settings
The system settings are divided into two modes, “Menu Mode” and “Utility Mode”.
Menu Mode
“Menu Mode” allows you to configure the “Initial Inflation Pressure Value” and “Auscultation Mode” settings.
Auscultation Mode
Initial Inflation Pressure Value
1. Press the [MODE] button.
The “Initial Inflation Pressure Value” setting screen appears.
2. Press the [ ] or [ ] button to change the setting.
· Press the [ ] button to change the setting in the order “280”, “250”, “220”, “AUTO”, “280”…
· Press the [ ] button to change the setting in the order “220”, “250”, “280”, “AUTO”, “220”…
When the “Initial Inflation Pressure Value” is set, inflation takes place at a fixed speed to the set value and thus is quicker.
“AUTO” estimates the systolic blood pressure during inflation and automatically inflates the cuff to a suitable value.
When using “220”, “250”, or “280” mmHg, select the value that is 30 to 40 mmHg higher than the estimated systolic blood pressure.
3. When you have completed the “Initial Inflation Pressure Value” setting,
press the [MODE] button.
The “Auscultation Mode” settings screen appears.
“ON” or “OFF” appears.
4. Press the [ ] or [ ] button to set to “ON” or “OFF”.
When set to “ON”, SYS and DIA can be recorded using auscultation measurement.
For information on auscultation measurement, see page 22.
5. When you have completed the “Auscultation Mode” setting, press the [MODE]
button.
“0” appears. To start blood pressure measurement, press the [START/STOP]
button.
Note:
· The body movement detection function is disabled while “Auscultation Mode”
is in use. · If the power is turned off, the settings revert to the factory
settings.
“Initial Inflation Pressure Value” changes to “AUTO”, and “Auscultation Mode”
changes to “OFF”.
16
Utility Mode
“Utility Mode” allows you to configure the “Auto Power Off” and “Pressure
Accuracy Confirmation” setting.
1. Confirm that the device is switched off.
2. Hold down the [MODE] button until the
“Auto Power Off” setting screen appears.
The “Initial Inflation Pressure Value” setting screen appears, and changes to the “Auto Power Off” setting screen appears.
“5” or “10” appears.
3. Press the [ ] or [ ] button to change the auto power off setting.
The “5 min.” or “10 min.” setting is entered.
When the unit is powered by the battery pack and is not used for the set time (“5 min.” or “10 min.”), the power automatically turns off to save battery power.
If an alarm has occurred, the power does not turn off automatically.
4. When you have completed the “Auto Power Off” setting, press the [MODE] button.
The “Pressure Accuracy Confirmation” screen appears. “0” appears.
5. Check the Pressure Accuracy.
Change the pressure externally. Compare with the displayed value and verify that there is no problem.
“0” appears.
6. When you have completed “Pressure Accuracy Confirmation”, press the
[START/STOP] button.
The power turns off.
Example:
System Settings
1. Connect the blood pressure monitor, the standard pressure gauge, and the
cuff and rubber ball.
For details on the standard pressure gauge, contact OMRON HEALTHCARE Customer
Service.
2. Check the blood pressure value of the blood pressure monitor and the blood
pressure value of the standard pressure gauge.
Note:
· The standard is within ± 3 mmHg or 2 % of manometer reading. · To rapidly
deflate the cuff, press the [ ] button.
To repeat “Pressure Accuracy Confirmation”, turn off the power and repeat the
procedure from step 1 of “Utility Mode”.
17
Non-Invasive Blood Pressure (NIBP) Measurement
Non-Invasive Pressure Measurement Principles
Oscillometric method The beat in the pulsation generated by the contraction of
the heart is captured as the pressure inside the cuff to measure the blood
pressure. If the cuff wrapped around the upper arm is pressurized
sufficiently, the blood flow stops, but the beat of the pulsation is present
and the pressure inside the cuff receives this and oscillates. Next, as the
pressure inside the cuff gradually decreases, the oscillation of the pressure
within the cuff gradually increases and reaches a peak. As the pressure within
the cuff decreases further, the oscillation decreases from its peak. The
pressure within the cuff and the relationship with the increase and decrease
of the oscillation within the cuff in this series of processes are stored into
memory, calculations are carried out, and the blood pressure value is
determined. The pressure within the cuff when the oscillation increases
drastically is the systolic pressure and the pressure within the cuff when the
oscillation decreases drastically is the diastolic pressure. Also, the
pressure within the cuff when the oscillation peaks is taken as the average
pulsation pressure. The oscillometric method does not determine the blood
pressure value instantly like a microphone type automatic blood pressure gauge
with the auscultation method, but rather determines it from the series of
change curves as explained above. Therefore, it is not easily affected by
external noise, an electric scalpel or other electro surgical instruments.
KOROTKOV SOUNDS
RADIAL PULSE
18
Cuff Selection and Connection
Selecting the cuff
Warning
Treat a cuff used on a patient with an infection as medical waste, or
sufficiently sterilize the cuff before reuse.
Caution
· Do not use the cuff if it is damaged or has holes. · Use the appropriate
cuff size to ensure correct measurements. If too large a cuff is used, the
measured blood pressure value tends to be lower than the actual blood
pressure. If too small a cuff is used, the measured blood pressure value tends
to be higher.
Note:
It is important to use the correct sized cuff for a patient in order to get an
accurate reading.
Measure the circumference of the patient’s arm and select the cuff size that
is appropriate for the circumference.
Select the cuff that is suitable for the patient from the cuffs below.
Cuff name
GS CUFF XL (GCUFF-XL) GS CUFF L (GCUFF-L) GS CUFF M (GCUFF-M) GS CUFF S
(GCUFF-S) GS CUFF SS (GCUFF-SS)*
Arm circumference
(inch)
(cm)
17-20
42 – 50
13-17
32 – 42
9-13
22 – 32
7-9
17 – 22
5-7
12 – 18
- Available as an optional accessory.
Connecting the cuff Connect the cuff tube to the NIBP connector on the unit and turn clockwise to lock.
Non-Invasive Blood Pressure (NIBP) Measurement
Caution
Only OMRON GS CUFF can be used with this device.
Note:
· If the cuff tube is too short, the optional 3.5 m or 1.5 m CUFF HOSE can be
connected to lengthen the cuff tube. Do not make the hose longer than 3.5 m,
as this will affect measurement accuracy.
· Make sure the connections are tight.
19
Applying the Cuff to the Patient
1. Wrap on a bare arm or over thin clothing.
Wrap the cuff on a bare arm or over thin clothing. Thick clothing or a rolled
up sleeve will cause inaccurate blood pressure measurements. The device can be
used on either the right or left arm.
2. Align the artery mark “INDEX ARTERY” with the brachial artery.
Run the cuff tube out the peripheral side with no bends (the brachial artery is on the inner side of the patient’s upper arm). Make sure that “INDEX ARTERY” is within the “RANGE” and the lower edge of the cuff is 1/2 -1 inches from the inner side of the elbow joint. If “INDEX ARTERY” is outside the “RANGE”, error in the blood pressure value will increase. In this case, use a different cuff size. As a guideline for the tightness of the cuff, It should be possible to insert about two fingers under the cuff.
1/2″-1″
3. During measurement, keep the brachial artery on which the cuff is wrapped
at the same height as the right atrium of the heart.
20
Caution
Make sure the cuff is wrapped in the correct arm position and is at the same
height as the heart. A difference of 4 inches (10 cm) in height may cause a
variation in the blood pressure value of up to 7 – 8 mmHg.
Note:
· If measurement is difficult due to arrhythmia, use a different blood
pressure measurement method. · If the patient has acute inflammation, a
pyogenic ailment, or an external wound at the location where
the cuff is to be wrapped, follow the instructions of the doctor. · Non-
Invasive Blood Pressure (NIBP) measurement is performed by compressing the
upper arm.
Some people may experience intense pain, or transient spotting caused by
subcutaneous hemorrhaging may appear. The spotting will disappear over time,
however, if it is possible that this will disturb the patient, try the
following technique: – Wrap a thin towel or cloth (one layer) under the cuff.
If the towel or cloth is too thick, there will be insufficient cuff
compression and the blood pressure value will measure high. · If the patient
moves or the cuff is touched, this may be falsely detected as a pulse and
over-inflation will occur. · Do not inflate the cuff when it is not wrapped on
the upper arm. This may damage the cuff.
Taking the Measurement in “Manual Mode”
1. Press the [START/STOP] button.
Blood pressure measurement is performed once.
2. The measurement results are displayed.
If a measurement value is outside the corresponding range below, the value will flash. SYS: 59 mmHg or less, or 251 mmHg or higher. DIA: 39 mmHg or less, or 201 mmHg or higher. PULSE: 39 bpm or less, or 201 bpm or higher.
Normal measurement
Measurement error / failure
Non-Invasive Blood Pressure (NIBP) Measurement
Caution
The accuracy of a flashing measurement value that is outside the measurement
range is not guaranteed. Always check the patient’s condition before deciding
what steps to take.
Note:
If inflation is insufficient, inflation may restart while measurement is in
progress.
21
Irregular pulse wave detection function
If the pulse wave interval becomes irregular during measurement, the irregular
pulse wave detection icon will light to notify you.
Body movement detection function
If body movement is detected during measurement, deflation stops for 5
seconds. The irregular pulse wave icon appears in the measurement result
display. Deflation stopped
After 5 seconds, measurement resumes, and an attempt is made to complete
measurement in one cycle.
Taking the Measurement in “Auscultation Mode”
Use a stethoscope to perform measurement. When set to “ON”, SYS and DIA can be
recorded using auscultation measurement. To determine SYS and DIA during
measurement, press the [MODE] button. SYS is registered the first time and DIA
is registered the second time you press the [MODE] button during deflation.
After DIA is determined, the cuff rapidly deflates and SYS and DIA are
displayed as the measurement results. The pulse rate does not appear in the
measurement result display. During deflation, the [ ] button can be held down
to re-inflate, or the [ ] button can be held down to deflate faster. Stored
data that was measured in “Auscultation Mode” is displayed when the
auscultation icon is lit. During re-inflation
Note:
The body movement detection function is disabled while “Auscultation Mode” is
in use. For Auscultation Mode settings, see page 16.
22
Stopping the Measurement
To stop measurement while measurement is in progress, press [START/STOP]
button.
Displaying Last Reading
The previous measurement value (systolic blood pressure, diastolic blood
pressure, and pulse rate) and whether or not an irregular pulse wave was
detected can be displayed. In the case of auscultation measurement, the
auscultation icon appears. Press the [ ] or [ ] button to display the previous
data. This function is also available when the device is switched off.
If an irregular pulse wave or body movement was detected, the irregular pulse
wave icon appears.
When the data was measured in “Auscultation Mode”, the auscultation icon
appears.
Lights while the previous data are being displayed.
If held down at least 3 seconds when a measurement result is displayed, the
measured data is cleared without being stored.
Note:
When the monitor is left idle for one minute, the backlight will disappear.
Non-Invasive Blood Pressure (NIBP) Measurement
23
Maintenance
Maintenance Inspection and Safety Management
The HBP-1300 must be maintained to ensure functionality and to secure the
safety of patients and operators. Daily checks and maintenance should be
performed by the operator. (page 25) In addition, qualified personnel are
necessary to maintain the performance and the safety, and to conduct periodic
inspections. We recommend that the verification test be performed at least
once a year. (page 17)
Managing Supplies
Keep the following supplies on hand. · GS CUFF – All sizes
Device Maintenance
Cleaning and disinfecting should be performed in accordance with your
facility’s infection control practice and OSHA regulations. Surface cleaning
Wipe with a cloth that has been moistened with isopropyl alcohol diluted to 50
v/v%, or ethyl alcohol (disinfection alcohol) diluted to 76.9 – 81.4 v/v% and
wrung out. Do not wipe the DC jack or allow it to become wet. Removing dust
Use a moistened cotton bud to remove dust that has accumulated on the vent
ports. Service The device requires no routine service other than cleaning, and
visually checking the cuffs, tubing, etc.
Caution
· Do not sterilize by autoclave or gas sterilization (EOG, formaldehyde gas,
high-concentration ozone, etc.).
· If using an antiseptic solution for cleaning, follow the instructions of the
manufacturer.
Accessory Care
Non-Invasive Blood Pressure Measurement (NIBP) Cuff Wipe clean on the surface
of the cuff with a cloth moistened with a 70 v/v% dilution of isopropyl
alcohol, or a 76.9 to 81.4 v/v% dilution of disinfection ethanol (ethyl
alcohol). Do not allow any liquids inside the cuff. If a liquid gets in the
cuff, dry the inside well.
24
Check before Use
Before conducting safety checks, be sure to implement the items in the “Device
Maintenance” and “Accessory Care” sections. (page 24)
Before turning on the power Before turning on the power, check for the
following External appearance The device or accessories are not deformed due
to falling or other impact. The device is not dirty. The device is not wet. AC
adapter The AC adapter is firmly connected to the connector on the device.
There are no heavy objects lying on the AC adapter cable. The AC adapter cable
is not damaged (core-wire exposure, breaks, etc.).
When turning on the power When turning on the power, check the following
Display/lamp When the [START/STOP] button is pressed to turn on the power, the
screen below appears and the
alarm lamp lights.
The alarm lamp is lit up.
After turning on the power After turning on the power, check for the following
External appearance There is no smoke or odor coming from the device. The
device is not making any unusual noises. Buttons Press each button and check
that it works. Non-invasive blood pressure (NIBP) Make sure that a suitable
OMRON GS CUFF is attached (one that fits the circumference of the
patient’s arm). The cuff tube is firmly connected. The person checking the
cuff should wrap the cuff around arm, perform cuff measurement and check
to see that blood pressure is in the vicinity of normal measurements. While
measurement is in progress, bend the relevant arm and move body to halt
discharge and
during this halt check that cuff pressure does not drop.
25
Maintenance
Troubleshooting
The power does not turn on
Cause
Solution
If the unit is being powered by the battery, the battery is not installed or the charge is depleted.
Insert battery or replace with a new battery (page 14).
Internal part failure
Disconnect the AC adapter, remove the battery, and contact OMRON HEALTHCARE Customer Service.
· Check if the AC adapter is disconnected or the connection is loose. · Check if the AC adapter or battery has failed.
The unit display does not operate Cause / solution Stop using the unit and contact OMRON HEALTHCARE Customer Service.
The unit becomes hot
Cause
Solution
An object is on top of the unit or right next to the unit. Keep the area around the unit free of objects.
If the unit becomes too hot to be touched, there may be a problem in the unit. Turn off the unit power, disconnect the AC adapter, remove the battery, and contact OMRON HEALTHCARE Customer Service at 1-877-216-1336.
The unit is connected to a power outlet, but it runs on the the battery pack
Cause / solution
If AC power cannot be supplied, the unit will operate by battery only. · Check
if the AC adapter is properly connected to the unit. · Check if AC adapter is
connected to a power outlet. · Check that the electrical outlet is working by
connecting a different device to the same power outlet.
The cuff does not inflate when the [START/STOP] button is pressed
Cause
Solution
Loose cuff tube connection.
Check the connection.
There is an air leak in the cuff.
Replace the cuff.
If pressure is displayed, the cuff tube is bent.
Make sure no part of the cuff tube is bent.
Measurement was not possible
Cause / solution
Check the patient by palpation or other method. After checking the patient,
check the error code and see “List of Error Codes” (page 28) for Non Invasive
Blood Pressure (NIBP) measurement.
26
Abnormal measurement value
Cause / solution
The causes below are possible. Check the patient by palpation and then repeat
measurement. · Body movement (chills or other trembling) · Arrhythmia. · Noise
in the cuff
– A nearby person touched the patient. – Cardiac massage was being performed.
The measurement value is questionable
Cause
Solution
Deflates quickly
Check for a loose cuff connection.
Stethoscope
Simultaneously perform measurement with a stethoscope. Place the stethoscope and listen while viewing the pressure display of the manometer.
Blood pressure may vary widely due to physiological effects. The causes below
are possible. · Emotional excitement or agitation
– Pain due to cuff wrapping – White coat hypertension · Cuff size or wrapping
method not correct · Cuff wrapping position on upper arm not at the same
height as the heart · Patient’s blood pressure not stable due to pulsus
alternans, respiratory changes, or other reason
Incorrect cuff size used.
Measure circumference or patient’s arm and ensure correct sized cuff is used.
Cuff wrapped over thick clothing.
Ensure cuff is applied to a bare arm, or very thin clothing.
Patient not seated properly.
Ensure patient is seated, feet flat on the floor, cuff at heart level.
Patient ate, drank, or exerted themselves recently.
Ensure before measurement taken, patient has not had food, caffeinated or alcohol beverages, or exerted/exercised in the last 30 minutes.
Maintenance
27
List of Error Codes
The alarm lamp flashes when a medium-priority alarm occurs, and lights
steadily when a low-priority alarm occurs. To clear an alarm, press any
button. If a low-priority alarm and a medium-priority alarm occur at the same
time, the medium-priority alarm
is displayed. If the alarms are the same priority level, the error code of the
alarm that occurred first is displayed. However, to prevent battery
consumption when powered by battery, a low battery error (E40 error) is always
given priority. Example: E2
SYSTEM
Error code
E9
Priority
Description
Points to check
Medium
Sensor failure: Sensor range not stable Outside of sensor range Pressure gain
lower limit value / upper limit value Pressure sensor count upper limit /
lower limit error
EEPROM failure: Checksum error Read error Protect code error
Contact OMRON HEALTHCARE Customer Service.
Switch error: Error in switch on state
NIBP
Error code
E1
E2
Priority
Description
Points to check
The cuff tube is not connected
Firmly connect the cuff tube.
Low Air is leaking from the cuff.
Replace with an OMRON GS CUFF that does not leak.
Did not inflate properly because the arm Have the patient not move the arm or body,
or body moved during measurement.
and repeat measurement.
Moved body or arm during measurement, or talked.
Have the patient not talk or move, and repeat measurement.
The cuff is not applied correctly.
Correctly apply the cuff.
The sleeve is rolled up and is
Low
compressing the arm.
Remove the garment and re-wrap the cuff.
Measurement time has exceeded specified time. Specified time: 165 seconds
The measurement time exceeds the expected time, so the measurement was ended in order to avoid patient discomfort. There is a possibility that measurement is being repeated over and over due to air leaking from the cuff.
28
Other problems
Error Priority
code
Cause
Inflated the cuff to 300 mmHg or more during inflation in “Auscultation Mode”.
E3 Low
Over-inflation occurs
E40 Medium The battery is depleted. E41 Medium Battery failed to charge. E42 Medium Battery voltage error
Solution
When inflating in “Auscultation Mode”, release the button when the pressure
reaches the desired value.
If this occurs during measurement, repeat measurement. If this occurs when not
performing measurement, contact OMRON HEALTHCARE Customer Service.
Recharge the battery, or replace with a new battery. (page 14)
Try charging again. If continues to fail, replace with a new battery. (page
14)
Replace the battery with a new battery. If the error continues, contact OMRON
HEALTHCARE Customer Service.
Disposal
Description
As there is a risk of environmental pollution, follow your applicable Federal,
state and local legal regulations regarding disposal or recycling of this
equipment and batteries. The main constituents of each part are listed in the
table below. As there is a risk of infection, do not recycle patient
attachments such as cuffs, but dispose of them as instructed by your
facility’s procedures and applicable regulations.
Item Package
Main unit and accessories
Battery pack
Parts Box Cushion Bag Enclosure Internal parts Outer tube Cell batteries Internal parts
Material Cardboard Cardboard PE ABS, PC , SR General electronic components PVC Nickel-hydrogen battery General electronic components
Maintenance
29
Limited Warranty
Your OMRON HBP-1300 Professional Blood Pressure Monitor, excluding the arm
cuffs is warranted to be free from defects in materials and workmanship
appearing within 2 years from the date of purchase, when used in accordance
with the instructions provided with the unit. The arm cuffs are warranted to
be free from defects in materials and workmanship appearing within one year
from the date of purchase when used in accordance with the instructions
provided with the unit. The above warranties extend only to the original
retail purchaser. We will, at our option, repair or replace without charge any
unit or arm cuffs covered by the above warranties. Repair or replacement is
our only responsibility and your only remedy under the above warranties. To
obtain warranty service contact Customer Service by calling 1-877-216-1336 for
the address of the repair location and the return shipping and handling fee.
Enclose the Proof of Purchase. Include a letter, with your name, address,
phone number, and description of the specific problem. Pack the product
carefully to prevent damage in transit. Because of possible loss in transit,
we recommend insuring the product with return receipt requested. THE FOREGOING
IS THE SOLE WARRANTY PROVIDED BY OMRON IN CONNECTION WITH THIS PRODUCT, AND
OMRON HEREBY DISCLAIMS ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
IMPLIED WARRANTIES AND OTHER TERMS THAT MAY BE IMPOSED BY LAW, IF ANY, ARE
LIMITED IN DURATION TO THE PERIOD OF THE ABOVE EXPRESS WARRANTY. OMRON SHALL
NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER SPECIAL, INCIDENTAL, CONSEQUENTIAL
OR INDIRECT COSTS, EXPENSES OR DAMAGES. This warranty provides you with
specific legal rights, and you may have other rights that vary by
jurisdiction. Because of special local requirements, some of the above
limitations and exclusions may not apply to you. FOR CUSTOMER SERVICE Call
toll free: 1-877-216-1336
30
Specifications
Factory Default Settings
Factory default settings and backup are as shown below. Backup
: Setting is retained even if the power is interrupted. : Reverts to factory
default setting if the power is turned off.
Setting Pressure Value Initial Inflation Pressure Value Auscultation Mode Auto Power Off
Settings AUTO, 220, 250, 280 ON, OFF 5 min, 10 min
Factory Setting AUTO OFF 5 min
Specifications Main unit
Backup
Measurement Parameter Dimension
Weight
Display UL Classfield Safety Standards
Protection Class Degree of Protection MDD Classification
NIBP Main unit: 4.84 x 7.91 x 3.90 (inch) 123 x 201 x 99 (mm) (W x H x D) AC
adapter: 1.81 x 2.60 x 1.46 (inch) 46 x 66 x 37 (mm) (W x H x D) Battery: 2.13
x 1.71 x 0.61 (inch) 54 x 43.5 x 15.4 (mm) (W x H x D) Main unit: Approx.
18.34 oz (0.52 kg) (not including accessories and options) AC adapter: Approx.
7.05 oz (0.2 kg) Battery: Approx. 3.53 oz (0.1 kg) (optional accessory) 7
segment LCD WITH RESPECT TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS
ONLY. IEC 60601-1:1988+A1:1993+A2:1995 UL60601-1:First Edition Medical
electrical equipment-Part1:General requirements for safety Class II Internal
powered equipment
Type BF
Class II a
Specifications
31
Power supply
AC adapter
Optional rechargeable battery
Input voltage range: AC 100 V to 240 V Rated Current: 1.6 A Frequency: 50/60
Hz Output voltage range: DC 6 V ±5%
Type: 3.6 V, 1900 mAh Number of operation cycles when fully charged: 300
Measurement conditions · New battery fully charged · Ambient temperature of
23°C (73.4°F) · Using M-size cuff · SYS 120 / DIA 80 / PR 60 (Inflation
setting: AUTO) · One 5-minute cycle consisting of “cuff measurement time +
wait time”
Environmental Conditions
Operational temperature and humidity Storage and transportation
EMC: Reference standard
Temperature range: 5 to 40°C (41 to 104°F) Humidity range: 15 to 85%RH (not
condensed) Atmospheric pressure: 700 to 1060hPa
Temperature range: -20 to 60°C (-4 to 140°F) Humidity range: 10 to 95%RH (not
condensed) Atmospheric pressure: 500 to 1060hPa
IEC60601-1-2:2007 Medical electrical equipment Part 1-2: General requirements
for basic safety and essential performance – Collateral standard:
Electromagnetic compatibility – Requirements and tests.
Non-Invasive Blood Pressure (NIBP)
Measurement technology
Oscillometric
Measurement method
Dynamic Linear Deflation method
Pressure display range
0 to 300 mmHg
Pressure display accuracy
Within ±3 mmHg or 2 %
NIBP measurement range
SYS 60 to 250 mmHg DIA 40 to 200 mmHg Pulse rate 40 to 200 /min
NIBP accuracy*
Maximum mean error within ±5mmHg Maximum standard deviation within ±8mmHg
Pulse rate accuracy
Within ±5 % of reading
Reference standard:
IEC 80601-2-30:2009 ANSI/AAMI SP-10:2002+A1:2003+A2: 2006/(R)2008 ISO81060-2:2009
- Comparison with auscultation method performed by a trained professional. DIA determined by the auscultation method is “K5”.
32
Manufacturer’s Declaration
The HBP-1300 is intended for use in the electromagnetic environment specified
below. The customer or the user of the HBP-1300 should assure that it is used
in such an environment.
Electromagnetic Emissions: (IEC60601-1-2)
Emission Test
RF emission CISPR 11
Compliance
Electromagnetic Environment
Group 1
The HBP-1300 uses RF energy only for internal functions. Therefore, this RF emission is extremely weak and there is little chance of it creating any kind of interference whatsoever with nearby electronic equipment.
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/flicker IEC 61000-3-3
Class B Class A Complies
The HBP-1300 is suitable for use in all establishments, including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes.
Electromagnetic Immunity: (IEC60601-1-2)
Immunity test
IEC60601-1-2 test level
Compliance level
Electromagnetic environment – guidance
Electrostatic discharge (ESD) IEC 61000-4-2
±6 kV contact ±8 kV air
±6 kV contact ±8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Electric fast transient/ burst IEC 61000-4-4
±2 kV for power supply lines ±1 kV for input/output lines
±2 kV for power supply lines ±1 kV for input/output lines
Mains power quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5
±1 kV differential mode ±2 kV common mode
±1 kV differential mode ±2 kV common mode
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
<5 % UT for 0.5 cycle
40 % UT for 5 cycles
70 % UT for 25 cycles
<5 % UT for 5 sec.
<5 % UT for 0.5 cycle
40 % UT for 5 cycles
70 % UT for 25 cycles
<5 % UT for 5 sec.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the HBP1300 requires continued operation during power mains interruptions, it is recommended that the HBP-1300 be powered from an uninterruptible power supply or a battery.
Power frequency (50/ 60 Hz) magnetic field IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Note: UT is the a.c. mains voltage prior to application of the test level.
Specifications
33
Immunity test
IEC60601-1-2 test level
Conducted RF IEC 61000-4-6
3 Vrms 150 kHz to 80 MHz 80% AM (2Hz)
Radiated RF IEC 61000-4-3
3 Vrms 80 MHz to 2.5 GHz 80% AM (2Hz)
IEC60601-1-2 test level
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should be used no closer to any part of the HBP-1300, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommend separation distance
3 Vrms
d = 1.2 P d = 1.2 P 80 MHz to 800 MHz d = 2.3 P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to he transmitter manufacturer and d is the recommended separation distance in meters (m).
3 V/m
Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya, should be less than the compliance level in each frequency rangeb.
Interference may occur in the vicinity of equipment marked with the following symbol:
Note1: At 80 MHz and 800 MHz, the higher frequency range applies. Note2: These
guidelines may not apply in all situations. Electromagnetic propagation is
affected by
absorption and reflection from structures, objects, and people.
a Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the HBP-1300 is used exceeds
the applicable RF compliance level above, the HBP-1300 should be observed to
verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the HBP-1300.
b Over the frequency range 150 kHz to 80MHz, field strengths should be less
than 3 V/m.
34
Note:
POTENTIAL FOR RADIO / TELEVISION INTERFERENCE (for U.S.A. only) This product
has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC rules. These limits are designed to
provide reasonable protection against harmful interference in a residential
installation. The product generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance with the instructions, may
cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If
the product does cause harmful interference to radio or television reception,
which can be determined by turning the product on and off, the user is
encouraged to try to correct the interference by one or more of the following
measures: · Reorient or relocate the receiving antenna. · Increase the
separation between the product and the receiver. · Connect the product into an
outlet on a circuit different from that to which the receiver is connected. ·
Consult the dealer or an experienced radio / TV technician for help. POTENTIAL
FOR RADIO / TELEVISION INTERFERENCE (for Canada only) This digital apparatus
does not exceed the Class B limits for radio noise emissions from digital
apparatus as set out in the interference-causing equipment standard entitled
“Digital Apparatus”, ICES003 of the Canadian Department of Communications. Cet
appareil numérique respecte les limites de bruits radioeléctriques applicables
aux appareils numériques de Clase B prescrites dans la norme sur le materiel
brouilleur: “Appareils Numériques”, ICES-003 édictée par le minister des
communications. Changes or modifications not expressly approved by the party
responsible for compliance could void the user’s authority to operate the
equipment.
35
Specifications
Recommended Separation Distances:
Recommended separation distance between portable and mobile RF communications equipment and the HBP-1300
The HBP-1300 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the HBP-1300 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the HBP-1300 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter (W)
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
80 MHz to 800 MHz 800 MHz to 2.5 GHz
d = 1.2 P
d = 1.2 P
d = 2.3 P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note1:At 80MHz and 800MHz, the separation distance for the higher frequency
range applies Note2:These guidelines may not apply in all situations.
Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
36
For Customer Service Call Toll Free
1-877-216-1336
Manufactured by OMRON HEALTHCARE Co., Ltd.
Distributed by
OMRON HEALTHCARE, INC. 1925 West Field Court, Lake Forest, IL 60045, USA
Made in China.
© 2012 OMRON HEALTHCARE, INC.
Read User Manual Online (PDF format)