Omron BP5250 Arm Blood Pressure Monitor Instruction Manual

October 30, 2023
Omron

Omron BP5250 Arm Blood Pressure Monitor Instruction Manual

Introduction

Thank you for purchasing the OMRON BP5250 Blood Pressure Monitor. This monitor can be used by multiple users. This blood pressure monitor uses the oscillometric method of blood pressure measurement. This means this monitor detects your blood movement through your brachial artery and converts the movements into a digital reading.

Safety Instructions
This instruction manual provides you with important information about the OMRON BP5250 Blood Pressure Monitor. To ensure the safe and proper use of this monitor, READ and UNDERSTAND all of these instructions. If you do not understand these instructions or have any questions, contact 1-800-634-4350 before attempting to use this monitor. For specific information about your own blood pressure, consult with your physician.

Intended Use

This device is a digital monitor intended for use in measuring blood pressure and pulse rate in the adult patient population. The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.

Receiving and Inspection
Remove this monitor and other components from the packaging and inspect for damage. If this monitor or any other components is damaged, DO NOT USE and contact 1-800-634-4350.

Symbols Glossary

For symbol information, visit: OmronHealthcare.com/symbols- glossary

Contraindications

  • This monitor is contraindicated for use in ambulatory environments.
  • This monitor is contraindicated for use on aircraft.

Important Safety Information

Read the Important Safety Information in this instruction manual before using this monitor. Follow this instruction manual thoroughly for your safety. Keep it for future reference. For specific information about your own blood pressure, CONSULT WITH YOUR PHYSICIAN.

Warning

  • Indicates a potentially hazardous situation that, if not avoided, could result in death or serious injury.
  • DO NOT use this monitor on infants, toddlers, children, or persons who cannot express themselves.
  • DO NOT adjust medication based on readings from this blood pressure monitor. Take medication as prescribed by your physician. ONLY a physician is qualified to diagnose and treat high blood pressure.
  • DO NOT use this monitor on an injured arm or an arm under medical treatment.
  • DO NOT apply the arm cuff on your arm while on an intravenous drip or blood transfusion.
  • DO NOT use this monitor in areas containing high-frequency (HF) surgical equipment, magnetic resonance imaging (MRI) equipment, or computerized tomography (CT) scanners. This may result in incorrect operation of the monitor and/or cause an inaccurate reading.
  • DO NOT use this monitor in oxygen-rich environments or near flammable gas.
  • Consult with your physician before using this monitor if you have common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation; arterial sclerosis; poor perfusion; diabetes; pregnancy; pre-eclampsia or renal disease.
  • NOTE that any of these conditions in addition to patient motion, trembling, or shivering may affect the measurement reading.
  • NEVER diagnose or treat yourself based on your readings. ALWAYS consult with your physician.
  • To help avoid strangulation, keep the air tube and AC adapter cable away from infants, toddlers, and children.
  • This product contains small parts that may cause a choking hazard if swallowed by infants, toddlers, and children.

Data Transmission
This product emits radio frequencies (RF) in the 2.4 GHz band. DO NOT use this product in locations where RF is restricted, such as on an aircraft or in hospitals. Turn off the Bluetooth® feature in this monitor, remove batteries, and/or unplug the AC adapter when in RF-restricted areas. For further information on potential restrictions refer to documentation on Bluetooth usage by the FCC/ISED.

AC Adapter (optional accessory) Handling and Usage

  • DO NOT use the AC adapter if this monitor or the AC adapter cable is damaged. If this monitor or the cable is damaged, turn off the power and unplug the AC adapter immediately.
  • Plug the AC adapter into the appropriate voltage outlet. DO NOT use in a multi-outlet plug.
  • NEVER plug in or unplug the AC adapter from the electric outlet with wet hands.
  • DO NOT disassemble or attempt to repair the AC adapter.
  • Battery Handling and Usage
  • Keep batteries out of the reach of infants, toddlers, and children.

Caution
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient, or cause damage to the equipment or other property.

  • Stop using this monitor and consult with your physician if you experience skin irritation or discomfort.
  • Consult with your physician before using this monitor on an arm where intravascular access or therapy, or an arterio-venous (A-V) the shunt is present because of temporary interference with blood flow which could result in injury.
  • Consult with your physician before using this monitor if you have had a mastectomy.
  • Consult with your physician before using this monitor if you have severe blood flow problems or blood disorders as cuff inflation can cause bruising.
  • DO NOT take measurements more often than necessary because bruising, due to blood flow interference, may occur.
  • ONLY inflate the arm cuff when it is applied to your upper arm.
  • Remove the arm cuff if it does not start deflating during a measurement.
  • DO NOT use this monitor for any purpose other than measuring blood pressure.
  • During measurement, make sure that no mobile device or any other electrical device that emits electromagnetic fields is within 12 inches (30 cm) of this monitor. This may result in incorrect operation of the monitor and/or cause an inaccurate reading.
  • DO NOT disassemble or attempt to repair this monitor or other components. This may cause an inaccurate reading.
  • DO NOT use it in a location where there is moisture or a risk of water splashing this monitor. This may damage this monitor.
  • DO NOT use this monitor in a moving vehicle such as in a car.
  • DO NOT drop or subject this monitor to strong shocks or vibrations.
  • DO NOT use this monitor in places with high or low humidity or high or low temperatures. Refer to section 10.
  • During measurement, observe the arm to ensure that the monitor is not causing prolonged impairment to blood circulation.
  • DO NOT use this monitor in high-use environments such as medical clinics or physician offices.
  • DO NOT use this monitor with another medical electrical (ME) equipment simultaneously. This may result in incorrect operation of the monitor and/or cause an inaccurate reading.
  • Avoid bathing, drinking alcohol or caffeine, smoking, exercising, and eating for at least 30 minutes before taking a measurement.
  • Rest for at least 5 minutes before taking a measurement.
  • Remove tight-fitting or thick clothing from your arm while taking measurements.
  • Remain still and DO NOT talk while taking measurements.
  • ONLY use the arm cuff on persons whose arm circumference is within the specified range of the cuff.
  • Ensure that this monitor has acclimated to room temperature before taking a measurement. Taking a measurement after an extreme temperature change could lead to an inaccurate reading.
  • OMRON recommends waiting for approximately 2 hours for the monitor to warm up or cool down when the monitor is used in an environment within the temperature specified as operating conditions after it is stored either at the maximum or at the minimum storage temperature. For additional information on operating and storage/transport temperature, refer to section 10.
  • DO NOT use this monitor after the durable period has ended. Refer to section 10.
  • DO NOT crease the arm cuff or the air tube excessively.
  • DO NOT fold or kink the air tube while taking measurements. This may cause an injury by interrupting blood flow.
  • To unplug the air plug, pull on the plastic air plug at the base of the tube, not the tube itself.
  • ONLY use the AC adapter, arm cuff, batteries, and accessories specified for this monitor. Use of unsupported AC adapters, arm cuffs, and batteries may damage and/or may be hazardous to this monitor.
  • ONLY use the approved arm cuff for this monitor. Use of another arm cuffs may result in incorrect readings.
  •  Inflating to a higher pressure than necessary may result in bruising of the arm where the cuff is applied. NOTE: refer to “If your systolic pressure is more than 210 mmHg” in section 3 of this instruction manual for additional information. Data Transmission
  • DO NOT replace batteries or unplug the AC adapter while your readings are being transferred to your smart device. This may result in incorrect operation of this monitor and failure to transfer your blood pressure data.
  • AC Adapter (optional accessory) Handling and Usage
  • Fully insert the AC adapter into the outlet.
  • When unplugging the AC adapter from the outlet, be sure to safely pull it from the AC adapter. DO NOT pull from the AC adapter cable.
  • When handling the AC adapter cable:
    • DO NOT damage it.
    • DO NOT break it.
    • DO NOT tamper with it.
    • DO NOT forcibly bend or pull it.
    • DO NOT twist it.
    • DO NOT use it if it is gathered in a bundle.
    • DO NOT pinch it.
    • DO NOT place it under heavy objects.
  • Wipe any dust off of the AC adapter.
  • Unplug the AC adapter when not in use.
  • Unplug the AC adapter before cleaning this monitor.
  • Battery Handling and Usage
  • DO NOT insert batteries with their polarities incorrectly aligned.
  • ONLY use 4 “AA” alkaline or manganese batteries with this monitor.
  • DO NOT use other types of batteries. DO NOT use new and used batteries together. DO NOT use different brands of batteries together.
  • Remove batteries if this monitor will not be used for a long period of time.
  •  If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Consult with your physician immediately.
  •  If battery fluid should get on your skin, wash your skin immediately with plenty of clean, lukewarm water. If irritation, injury, or pain persists, consult with your physician.
  • DO NOT use batteries after their expiration date.
  • Periodically check batteries to ensure they are in good working condition.

Know Your Monitor

Contents/Product Includes
Monitor, Easy Wrap ComFit Cuff, 4 “AA” batteries, storage case, instruction manual, quick start guide

Monitor and Arm Cuff
Front of the monitor

  • Display

  • [Forward/Backward] buttons

  • [Connection] button

  • (Date/Time)

  • [Memory] button

  • [START/STOP] button
    Left side, back, and bottom of the monitor

  • Air jack

  • AC adapter jack (for optional AC adapter)

  • Battery compartment

Arm cuff
(Upper arm circumference: 9″ – 17″ (22 – 42 cm)

  • Air plug
  • Air tube

Display and Symbols

Classification of BP (Blood Pressure)
ACC/AHA 2017 Hypertension guidelines define categories of BP in adults as below.

Categories of BP in Adults

Individuals with Systolic BP and Diastolic BP in 2 categories should be designated to the higher BP category. BP indicates blood pressure (based on an average of ≥ 2 careful readings obtained on ≥ 2 occasions).
Source: ACC/AHA 2017 High Blood Pressure Clinical Practice Guideline

BP Category Systolic BP Diastolic BP
Normal < 120 mmHg    and    < 80 mmHg
Elevated 120 – 129 mmHg and     < 80 mmHg

Hypertension

Stage 1

Stage 2

| ****


130 – 139 mmHg  or  80 – 89 mmHg

≥ 140 mmHg    or       ≥ 90 mmHg

Warning
NEVER diagnose or treat yourself based on your readings. ALWAYS consult with your physician.

Preparation

Installing Batteries

  1. Push down on the hook of the battery cover and pull downward.
  2. Insert 4 “AA” batteries as indicated in the battery compartment.
  3. Close the battery cover. As soon as batteries are correctly installed, the year flashes on the display. Refer to sub-section 2.2 to set the date and time manually, or refer to sub-section 4.1 to pair your monitor to set the date and time automatically.

Note

  • When the “ ” symbol flashes on the display, it is recommended you replace the batteries.
  • To replace batteries, turn your monitor off and remove all batteries. Then replace with 4 new batteries at the same time.
  • Replacing batteries will not delete previous readings.
  • The supplied batteries may have a shorter life span than new batteries.
  • Disposal of used batteries should be carried out in accordance with local regulations.

Setting the Date and Time
Note

  • Set your monitor to the correct date and time before taking a measurement for the first time.
  • The date and time will automatically be set when the monitor is paired with the “OMRON connect US/CAN” app. Refer to sub-section 4.1.
  • As soon as batteries are correctly installed, the year flashes on the display of your monitor.
  1. If nothing is displayed, press the button to display the year. If the year is already displayed, skip this step.

  2. Press the or button to adjust the year. Press and hold the button to quickly advance the years. Press and hold the button to quickly go back through the years.

  3. Press the button to confirm the year, then the month will flash. Repeat the same steps to adjust the month, day, hour, and minute.

  4. Press the button to store the setting. Your monitor automatically turns off 3 seconds after the minutes have been set in this sequence.
    Note

    • If you need to reset the date and time, replace batteries, or press the button, then set the date and time as shown in the above steps.
    • If the date and time are not set, “ ” flashes during or after measurement.

Blood Pressure Measurement Tips

To help ensure an accurate reading, follow these directions:

  • Stress raises blood pressure. Avoid taking measurements during stressful times.

  • Measurements should be taken in a quiet place.

  • It is important to take measurements at the same time each day. Taking measurements in the morning and in the evening is recommended.

  • remember to have a record of your blood pressure and pulse readings for your physician. A single measurement does not provide an accurate indication of your true blood pressure. You need to take and record several readings over a period of time.
    Caution
    Avoid bathing, drinking alcohol or caffeine, smoking, exercising, and eating for at least 30 minutes before taking a measurement.

  • Rest for at least 5 minutes before taking a measurement.

Applying the Arm Cuff
Note

  • The following steps are for applying the arm cuff to your left arm. When you take a measurement on your right arm, follow the cuff wrap instructions for right arm use mentioned under “To take a measurement on your right arm” located at the end of this sub-section.
  • The blood pressure can differ between the right arm and the left arm, and the measured blood pressure readings can be different.
  • OMRON recommends always using the same arm for measurement. If the readings between both arms differ substantially, check with your physician to determine as to which arm to use for your measurement

Caution
Remove tight-fitting or thick clothing from your arm while taking measurements.

  1. Plug the arm cuff into your monitor by inserting the air plug into the air jack securely until it clicks.

  2.  Apply the arm cuff to your left upper arm.
    Note
    The bottom edge of the arm cuff should be 0.5 inches (1 to 2 cm) above the inside elbow. The air tube is on the inside of your arm and aligned with your middle finger.

  3.  Make sure that the air tube is positioned on the inside of your arm and wrap the cuff securely, so it can not move around your arm

To take a measurement on your right arm
When you take a measurement on your right arm, the air tube should run along the side of your elbow, along the bottom of your arm. Be careful not to rest your arm on the air tube.

Sitting Correctly

To take a measurement, you need to be relaxed and comfortably seated in a room with a comfortable temperature. Place your arm on the table.

  • Sit in a comfortable chair with your back and arm supported.
  • Keep your feet flat and your legs uncrossed.
  • The arm cuff should be placed on your arm at the same level as your heart, with the arm resting comfortably on a table.

Using Your Monitor without a Smart Device

Taking a Measurement
You can use your monitor without pairing it with a smart device. To pair your monitor with a smart device, refer to sub-section 4.1.
Note
To stop the measurement, press the [START/STOP] button once to deflate the arm cuff.

Caution

  • DO NOT use this monitor with another medical electrical (ME) equipment simultaneously. This may result in incorrect operation of the monitor and/ or cause an inaccurate reading.
  • Remain still and DO NOT talk while taking measurements.
  1. Press the [START/STOP] button. All symbols appear on the display before starting the measurement.

  2. Remain still and do not talk until the entire measurement process is complete. The “ ” symbol flashes at every heartbeat.

  3. After your monitor has detected your blood pressure and pulse rate, the cuff automatically deflates. Your blood pressure and pulse rate are displayed. If either the systolic or the diastolic reading is high (refer to subsection 1.3), the “ ” symbol appears.

  4. Press the [START/STOP] button to turn your monitor off.
    Note

    • Your monitor automatically turns off after 2 minutes.
    • Wait 2 – 3 minutes between measurements. The wait time allows the arteries to decompress and return to their pre-measurement state.
    • You may need to increase the wait time depending on your individual physiological characteristics.

Using the Guest Mode
The guest mode can be used to take a single measurement for another user.
Note
No readings are stored in the memory.

  1. Press and hold the [START/STOP] button for more than 2 seconds. The date/time display will disappear.
  2. Release the [START/STOP] button when the date/time display turns off. The arm cuff automatically starts to inflate.

If your systolic pressure is more than 210 mmHg
If your systolic pressure is more than 210 mmHg, take a measurement manually. After the arm cuff starts to inflate, press and hold the [START/STOP] button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure.

Note
Do not inflate above 299 mmHg.

Caution
Inflating to a higher pressure than necessary may result in bruising of the arm where the cuff is applied.

Pairing Your Monitor with a Smart Device
The date and time on your monitor will automatically be set when you pair your monitor with your smart device.

  1. Enable Bluetooth on your smart device.

  2. Download and install the free “OMRON connect US/CAN” app onto your smart device. If you already have the “OMRON connect US/CAN” app and have created your account, open the app and add your new monitor.

  3. Open the app and follow the pairing instructions shown on your smart device.

  4. Confirm that your monitor is connected successfully. When your monitor is connected successfully to your smart device, the OK symbol appears.

  5. Press the [START/STOP] button to turn your monitor off.
    Note

    •  If the “Err” symbol appears, follow the instructions in the “OMRON connect US/CAN” app.
    • Your monitor automatically turns off after 2 minutes.
    • Be aware that OMRON will not be responsible for the loss of data and/or information in the app.
    • “OMRON connects US/CAN” is the only app that we recommend using with your monitor to transfer data correctly.
  6. Using Your Monitor without a Smart Device START INFLATING COMPLETED

  7. Using Your Monitor with a Smart Device

  8. Review the list of compatible smart devices at OmronHealthcare.com

Taking a Measurement

Refer to sub-section 3.1 to start a measurement. Open the app on your smart device and follow the instructions.

Note
Make sure that your monitor is within 16 ft. (5 m) of your smart device.

Transferring Data Automatically/Manually
The readings will be transferred to your smart device within 1 hour after taking a measurement. To transfer the readings manually, press the button.

Disabling/Enabling Bluetooth
Warning
This product emits radio frequencies (RF) in the 2.4 GHz band. DO NOT use this product in locations where RF is restricted, such as on an aircraft or in hospitals. Turn off the Bluetooth feature in this monitor and remove batteries when in RF-restricted areas. For further information on potential restrictions refer to the documentation on Bluetooth usage by the FCC/ISED. Bluetooth is enabled as default.

  1. When your monitor is off, press and hold the button for more than 10 seconds. “oFF” flashes 3 times.
  2. Press the [START/STOP] button to turn your monitor off.
    Note
    To enable Bluetooth, press and hold the button for more than 2 seconds. “on” appears on the display. Your monitor automatically turns off 2 minutes after the completion. When Bluetooth is disabled, the “ ” symbol appears.

Using Memory Function

Your monitor automatically stores up to 80 readings.
Note
This monitor can be used by multiple users. When you use the memory function with other users, track (or manage) your own readings appropriately.

Viewing the Readings Stored in Memory

Press the button.
The Memory number appears for 1.5 seconds before the pulse rate is displayed. The most recent reading stored in the memory will be numbered as “1”.

Note

  • If the reading is high (refer to sub-section 1.3), the “ ” symbol appears.
  • If there are no readings stored in the memory, the screen to the right is displayed.
  • If the memory is full, new readings will replace the old readings, starting with the oldest reading.
  • Press the or button repeatedly to scroll through the previous readings stored in the memory.
  • To view the older readings
  • To view the more recent readings

Viewing the Average Reading
Your monitor can calculate an average reading based on the latest 2 or 3 readings taken within the most recent 10-minute timeframe.

Note
Set your monitor to the correct date and time before taking a measurement to view the average readings. When your monitor is off, press and hold the button for more than 2 seconds. The “ ” symbol flashes.

Note
If there are only 2 readings in the memory for that period, the average will be based on these 2readings.  If the average reading is high (refer to sub-section  1.3), the “ ” symbol appears.

Other Settings

Deleting All Readings Stored in Memory

  1. Press the button. After a memory number appears, the latest reading will appear.
  2. While holding the button down, press and hold the [START/STOP] button for more than 3 seconds. The display to the right flashes.
  3. While holding the button down again, press and hold the [START/STOP] button for more than 3 seconds. The display stops flashing, then all readings are deleted.
  4. Press the [START/STOP] button to turn your monitor off.
    Note
    All readings will be deleted. You cannot partially delete the readings stored in the memory. our monitor automatically turns off after 2 minutes.

Restoring Your Monitor to the Default Settings
To delete all the information stored in your monitor, follow the instructions below. Make sure that your monitor is turned off.

  1. While holding the button down, press and hold the [START/ STOP] button for more than 3 seconds. The display to the right flashes.

  2. While holding the button down again, press and hold the [START/STOP] button for more than 3 seconds. The display stops flashing, then your monitor is restored to the default settings.

  3. Press the [START/STOP] button to turn your monitor off.
    Note
    After the completion, the year flashes on the display. Refer to subsection 2.2 to set the date and time manually, or refer to sub-section 4.1 to pair your monitor to set the date and time automatically.

  4. Reverting to the default setting of your monitor does not delete the information in the app.

Other Settings

Error Messages and Troubleshooting

If any of the below problems occur during measurement, check to make sure that no other electrical device is within 12 inches (30 cm). If the problem persists, refer to the table below.

Maintenance

To protect your monitor from damage, follow the directions below:
Caution
DO NOT disassemble or attempt to repair this monitor or other components. This may cause an inaccurate reading.

Storage
Keep your monitor and other components in the storage case when not in use. Store your monitor and other components in a clean, safe location. Remove the arm cuff from the monitor.

Caution

  1. To unplug the air plug, pull on the plastic air plug at the base of the tube, not the tube itself.

  2. Gently fold the air tube into the arm cuff.
    Note
    Do not bend or crease the air tube excessively.

  3. Place your monitor and other components in the storage case

Cleaning

  • Do not use any abrasive or volatile cleaners.
  • Use a soft dry cloth or a soft cloth moistened with mild (neutral) detergent to clean your monitor and arm cuff and then wipe them with a dry cloth.
  • Do not wash or immerse your monitor and arm cuff or other components in water.
  • Do not use gasoline, thinners or similar solvents to clean your monitor and arm cuff or other components.

Disposal
Dispose of your monitor, other components, and optional accessories according to applicable local regulations. Unlawful disposal may cause environmental pollution.

Optional Accessories

Note

  • Do not throw the air plug away. The air plug can be applicable to the optional cuff.
    For further information, visit our website at OmronHealthcare.com
    Using the AC Adapter (optional accessory)
    Note

  • Make sure not to place your monitor in a location where it is difficult to plug and unplug the AC adapter.

  • We recommend keeping batteries in your monitor at all times, even if you choose to use the AC adapter. If only the AC adapter is used without keeping batteries in your monitor, you may need to reset the date and time each time you unplug and plug back the AC adapter. The readings will not be deleted.

  • Insert the AC adapter plug into the AC adapter jack on the rear side of your monitor.

  • Plug the AC adapter into an electrical outlet.

  • To unplug the AC adapter, unplug the AC adapter from the electrical outlet and then remove the AC adapter plug from the monitor.

Caution
ONLY use the AC adapter, arm cuff, batteries, and accessories specified for this monitor. Use of unsupported AC adapters, arm cuffs, and batteries may damage and/or may be hazardous to this monitor

Specifications

Note
These specifications are subject to change without notice. This monitor complies with the requirements of ISO 81060-2:2013. In the clinical validation study, K5 was used on 85 subjects for the determination of diastolic blood pressure.
IP classification is the degree of protection provided by enclosures in accordance with IEC 60529. This monitor and optional AC adapter are protected against solid foreign objects of 0.5 inches (12.5 mm) diameter and greater such as a finger. The optional AC adapter is protected against vertically falling water drops which may cause issues during a normal operation

About a wireless communication interference
This product operates in an unlicensed ISM band at 2.4 GHz. In the event this product is used near other wireless devices such as microwaves and wireless LAN, which operate on the same frequency band as this product, there is a possibility that interference may occur. If interference occurs, stop the operation of the other devices or relocate this product away from other wireless devices before attempting to use it.

The pictogram on the Product

FCC/ISED Statement

FCC CAUTION
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. This device complies with part 15 of FCC Rules and Innovation, Science and Economic Development Canada’s license-exempt RSS(s). Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.

Note
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

  • Reorient or relocate the receiving antenna.
  •  Increase the separation between the equipment and the receiver.
  • Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
  • Consult the dealer or an experienced radio/TV technician for help.

This transmitter must not be co-located or operated in conjunction with any other antenna or transmitter. This equipment complies with FCC/ISED radiation exposure limits set forth for an uncontrolled environment and meets the FCC radio frequency (RF) Exposure Guidelines and RSS-102 of the ISED radio frequency (RF) Exposure rules as this equipment has very low levels of RF energy.

Limited Warranty

Your BP5250 Blood Pressure Monitor, excluding the arm cuff and batteries, is warranted to be free from defects in materials and workmanship appearing within 3 years from the date of purchase, when used in accordance with the instructions provided with the monitor. The arm cuff is warranted to be free from defects in materials and workmanship appearing within 1 year from the date of purchase when the monitor is used in accordance with the instructions provided by the monitor. The above warranty extends only to the original retail purchaser. We will, at our option, replace without charge any monitor or arm cuff covered by the above warranty. Replacement is our only responsibility and
your only remedy under the above warranty. To obtain warranty service contact Customer Service by calling 1-800-634-4350 for the address of the inspection center and the return shipping and handling fee that may apply. Enclose the original printed receipt. Include a letter, with your name, address, phone number, and description of the specific problem. Pack the product carefully to prevent damage in transit. Because of possible loss in transit, we recommend insuring the product with the return receipt requested.

THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY DISCLAIMS ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. IMPLIED WARRANTIES AND OTHER TERMS THAT MAY BE IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION TO THE PERIOD OF THE ABOVE EXPRESS WARRANTY.

SOME STATES DO NOT ALLOW LIMITATIONS ON HOW LONG AN IMPLIED WARRANTY LASTS, SO THE ABOVE LIMITATION MAY NOT APPLY TO YOU. OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER
SPECIAL, INCIDENTAL, CONSEQUENTIAL, OR INDIRECT COSTS, EXPENSES, OR DAMAGES. SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGES, SO THE ABOVE LIMITATION OR EXCLUSION MAY NOT APPLY TO YOU. This warranty provides you with specific legal rights, and you may have other rights that vary by jurisdiction. Because of special local requirements, some of the above limitations and exclusions may not apply to you.

FOR CUSTOMER SERVICE
Visit our website at: For USA – OmronHealthcare.com
For Canada – OmronHealthcare.ca
Call toll-free: 1-800-634-4350

Important information regarding Electromagnetic Compatibility
(EMC)

BP5250 conforms to IEC60601-1-2:2014 Electromagnetic Compatibility (EMC) standard. Further documentation in accordance with this EMC standard is available at OmronHealthcare.com/emc. Refer to the EMC information for BP5250 on the website The Bluetooth® word mark and logos are registered trademarks owned by the Bluetooth SIG, Inc. and any use of such marks by OMRON HEALTHCARE Co., Ltd. is under license. Other trademarks and trade names are those of their respective owners. App Store and App Store logo are service marks of Apple Inc., registered in the U.S. and other countries. Google Play and the Google Play logo are trademarks of Google LLC.

OMRON HEALTHCARE Co., Ltd

  • 53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN
  • Distributed by: OMRON HEALTHCARE, INC.
  • 1925 West Field Court, Lake Forest, IL 60045 U.S.A.
  • For USA – OmronHealthcare.com
  • For Canada – OmronHealthcare.ca
  • 2019 OMRON HEALTHCARE, INC. Made in Vietnam

Download PDF:Omron BP5250 Arm Blood Pressure Monitor Instruction Manual

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals