OMRON PM500 TENS Therapy Pain Relief
INTRODUCTION
- Thank you for purchasing the OMRON Max Power Relief.
- The Max Power Relief is a portable TENS (Transcutaneous Electrical Nerve Stimulation) device that provides relief of chronic, acute* and arthritic pain. It is designed to reduce and relieve muscle and joint pain, stiffness, and numbness in the back, arms, legs, shoulder, and foot by applying electrical nerve stimulation to the surface of the skin near the site of the pain. The Max Power Relief uses comfortable, self-adhesive Long Life Pads™ that can be used up to 150 times.
- TENS is a safe and drug-free pain relief solution that has been used for over 40 years by medical professionals such as physical therapists and chiropractors.
Acute pain refers to sore or achy muscles due to strain from exercise or normal household and work activities.
- Register your product online at www.register-omron.com.
- Keep your purchase receipt as proof of purchase for warranty coverage.
INTENDED USE
This device is intended for the relief of pain associated with sore or aching muscles of the lower back, arms, legs, shoulder, or foot due to strain from exercise or normal household and work activities. When used for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis, use the Shoulder mode of stimulation.
- Environments of Use: Clinics, hospital and home environments
- Patient Population: Adult
SAVE THESE INSTRUCTIONS
-
Symbols Glossary:
For symbol information, visit: OmronHealthcare.com/symbols- glossary -
Contraindication
Do not use this device if you have a cardiac pacemaker, implanted defibrillator or other implanted metallic or electronic device. Such use could cause electric shock, burns, electrical interference or death.
SAFETY PRECAUTIONS
IMPORTANT SAFETY PRECAUTIONS AND WARNINGS
WARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
DO NOT USE THIS UNIT WITH THESE OTHER DEVICES:
- Do not use this device while using another TENS device.
- With a life-supporting medical electronic device such as an artificial heart lung or respirator.
- For Hospitals and Clinics: In the presence of electronic monitoring equipment such as cardiac monitors and ECG alarms the equipment may not operate properly when the electrical stimulation device is in use.
- For Hospitals and Clinics: Simultaneous connection of a patient to high-frequency surgical ME equipment may result in burns at the site of the stimulator electrodes and possible damage to the stimulator.
- For Hospitals and Clinics: Operation in close proximity (e.g. 1 m) to a shortwave or microwave therapy ME equipment may produce instability to the stimulator output.
DO NOT USE THIS UNIT UNDER THESE CONDITIONS
- Consult with your physician before using this unit. The unit may cause lethal rhythm disturbances in certain susceptible individuals. If you have had a recent surgical procedure, the stimulation may disrupt the healing process.
Before using your unit, ensure that:
- The cord is not broken.
- The gel pad is not damaged.
- The electrode cord connection is not broken.
- The unit is intact and working.
- There is no battery leakage.
DO NOT USE ON THESE INDIVIDUALS
- Pregnant women.
- Do not use on children or infants because this device has not been evaluated for pediatric use.
- Keep out of the reach of young children because the electrode cord could cause strangulation.
- Persons incapable of expressing their thoughts or intentions.
- Persons incapable of operating the unit by themselves.
- Use caution if you have a tendency to bleed internally, such as following an injury or fracture.
- If you have suspected or diagnosed heart disease, you should follow precautions recommended by your physician.
- If you have suspected or diagnosed epilepsy, you should follow precautions recommended by your physician.
- Use caution if stimulation is applied over the menstruating uterus.
DO NOT APPLY THE PADS TO THESE BODY AREAS:
|
The head, the mouth, or any area of the face.
---|---
**| The neck or any area of the throat because this could cause severe
muscle spasms resulting in closure of the airway, difficulty in breathing, or
adverse effects on heart rhythm or blood pressure.
**| ****
Near the heart, or on the genital area.
|
Both sides of the chest cavity simultaneously (lateral or front and back), or across your chest because the introduction of electrical current may cause rhythm disturbances which could be lethal.
---|---
|
On the calves of both legs at the same time because this may cause cardiac disturbance.
|
On the bottom of both feet at the same time because this may cause cardiac disturbance.
Open wounds or rashes or over swollen, red, infected or inflamed areas or skin eruptions (such as varicose veins, phlebitis, thrombophlebitis and thrombosis), or on top of or close to cancerous lesions, or over ares of skin that lack normal sensation.
DO NOT USE THIS UNIT DURING THESE ACTIVITIES
- When in the bath or shower;
- While sleeping;
- While driving, operating machinery, or during any activity in which electrical stimulation can put you at risk of injury.
PAIN MANAGEMENT WARNINGS
- If you have had medical or physical treatment for your pain, consult with your physician before using this unit.
- If your pain does not improve, becomes seriously chronic or severe, or continues for more than five days, stop using the unit and consult with your physician.
- The mere existence of pain functions as a very important warning telling us that something is wrong. Therefore, if you suffer from any serious illness, consult your physician in order to confirm that it is advisable for you to use this unit.
DO NOT ALTER THE UNIT
- Do not plug this cord into anything other than this unit.
- No modification of this unit is allowed.
- Use this unit only with the leads, electrodes, pads and accessories recommended by the manufacturer to avoid damage to the unit.
WARNINGS REGARDING THE PADS
- Apply pads to normal, healthy, dry, clean skin (of adult patients).
- If you experience any skin irritation or redness after a session, do not continue stimulation in that area of the skin.
Possible Adverse Reactions
- You should stop using the unit and consult with your physician if you experience adverse reactions from the unit.
- You may experience skin irritation and burns beneath the stimulation electrodes applied to your skin.
- You may experience headache and other painful sensations during or following the application of electrical stimulation near your eyes and to your head and face.
CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient, or cause damage to the equipment or other property.
PRECAUTIONS REGARDING THE PADS
- Do not move the pads to another location while the unit is on.
- Therapy won’t work with just one pad. You MUST USE TWO PADS at the same time.
- Make sure the components are connected well and the pads are fixed on the part of the body you wish to treat or the therapy may not be effective.
- Pad should not touch any metal object, such as a belt buckle, necklace, or other metal worn under clothing.
- Do not overlap pads or put pads on top of each other.
- Gel pads may also stick together and cause gel to be removed when separating.
- Do not share pads with another person. This may cause a skin irritation or infection. Pads are intended for use by one person.
- Do not leave pads attached to the skin after treatment.
- Do not bend or fold the pads because the gel may get damaged and it won’t stick or function properly.
- To avoid damage to the adhesive surface of the pads, put the pads only on the skin or on the plastic pad holder provided.
- Always place clean pads in accordance with illustrations provided (Refer to pages 19-22, Pad Placement).
- Do not apply ointment or any solvent to the pads or to your skin because it will disrupt the pads from functioning properly. The self-adhesive pads will adhere to your skin.
- Do not add spray, lotions or creams to skin or pads.
- Place pads on each side of the pain, not directly on the pain.
- Place pads at least 1 inch (3 cm) apart for optimal results.
CAUTION WHILE USING UNIT
MAIN UNIT
- If the unit is not functioning properly or you feel discomfort, immediately stop using it.
- Do not use for any other purpose except for what it is intended.
- Do not place in a room with high humidity, such as a bathroom. This will damage the unit. Ideal temperature for using 50 °F – 104 °F (10 °C – 40 °C), 30 – 80 % relative humidity.
- Do not use the unit without proper lighting. You may not be able to operate unit successfully.
- While using this device, make sure that no mobile phone or any other electrical devices that emit electromagnetic fields is within 12 inches (30 cm). This may result in poor performance of the device.
- Do NOT turn unit on until pads are on your skin.
CORD
- Do not insert the electrode plug into any place other than the jack on the unit.
- Do not pull on the electrode cord during treatment.
- Do not bend or pull the end of the cord.
- When pulling out the cord from the unit, hold the plug and pull.
- Replace the cord when broken or damaged.
BATTERY
- Do not throw the batteries into a fire. The batteries may explode.
- Dispose of the unit, batteries, and components according to applicable legal regulations. Unlawful disposal may cause environmental pollution.
- Do not mix alkaline and manganese batteries as this will shorten the battery life.
- During therapy, do not remove the battery cover and do not touch the battery terminals.
- Do not use batteries after their expiration date.
PADS
- The size, shape and type of pads may affect the safety and effectiveness of electrical stimulation.
- The electrical performance characteristics of pads may affect the safety and effectiveness of electrical stimulation.
- Using pads that are too small or incorrectly applied, could result in discomfort or skin burns.
- Detach the pads before replacing the batteries.
General Precautions
- The long-term effects of electrical stimulation are unknown.
- Apply stimulation to only normal, intact, clean, dry and healthy skin.
- TENS is not effective in treating the original source or cause of the pain, including headache.
- TENS is not a substitute for pain medications and other pain management therapies.
- TENS devices do not cure disease or injuries.
- TENS is a symptomatic treatment and, as such, suppresses the sensation of pain that would otherwise serve as a protective mechanism.
- Effectiveness is highly dependent upon patient selection by a practitioner qualified in the management of pain patients.
- You may experience skin irritation or hypersensitivity due to the electrical stimulation or electrical conductive medium (gel).
- Keep unit away from young children. The unit contains small pieces that may be swallowed. Immediately contact your physician.
- Ensure that this device has acclimated to room temperature before using. Using this device after an extreme temperature change could lead to an improper operation. OMRON recommends waiting for approximately 2 hours for the device to warm up or cool down when the device is used in an environment within the temperature specified as operating conditions after it is stored either at the maximum or at the minimum storage temperature. For operating, storage and transport conditions, refer to page 34.
- The clip may fall off if attached to soft or thin clothing.
Possible Adverse Reactions
- Do not use to treat one region for extended periods of time (more than 30 minutes a session, up to three times/day) or muscles in that region may become exhausted and sore.
HOW MAX POWER RELIEF WORKS
How does it work?
- The OMRON Max Power Relief is a powerful TENS device that provides relief of chronic, acute* and arthritic pain.
Scientific theory suggests that TENS therapy may work in several ways:
- Gentle electrical pulses move through the skin to nearby nerves to block or shut out the pain message from ever reaching the brain from the source of the pain.
- Gentle electrical pulses increase the production of the body’s natural pain killers, such as endorphins.
- Blood circulation improves as muscles contract and relax with the flow of the electrical stimulation.
- Acute pain refers to sore or achy muscles due to strain from exercise or normal household and work activities.
KNOW YOUR UNIT
Instruction Manual
Quick Start Guide/Pad Placement Guide
FEATURES
- Nine preset modes (Arm, Lower Back, Leg, Foot, Shoulder, Joint, Tap, Knead, Rub).
- Fifteen intensity levels (1 low to 15 high).
- Pair of pads (durable, reusable, washable, up to 150 uses). (Only use OMRON manufactured pads and cords with this unit).
- Automatic 15-minute shut off.
- Big screen lets you clearly control your therapy with mode, intensity level and minutes left displayed.
- Low battery icon displays to remind you when to replace the batteries.
- Pad icon displays if pad falls off.
BUTTONS AND THEIR FUNCTIONS
ASSEMBLY STEPS
Before using your unit, ensure that:
- The cord is not broken.
- The gel pad is not damaged.
- The electrode cord connection is not broken.
- The unit is intact and working.
- There is no battery leakage.
STEP 1 – INSERT BATTERIES
- Remove the battery cover on the back using a coin.
- Insert batteries. Make sure the signs correspond when inserting batteries.
- Reinstall the battery cover. Tighten with a coin.
STEP 2 – ATTACH ELECTRODE CORD TO THE MAIN UNIT
- Attach the electrode cord plug to the bottom of the main unit.
STEP 3 – SNAP EITHER ELECTRODE CORD TO EACH OF THE PADS
For the first time, take the pads out of the sealed package.
CAUTION
- Do NOT turn unit on until pads are on your skin.
- Therapy won’t work with just one pad. You MUST USE TWO PADS at the same time.
NOTE: Pads will not stick if the skin has too much hair. We recommend shaving the area for effective treatment.
STEP 4 – REMOVE AND DISCARD PLASTIC FILM FROM PADS
- Remove the clear plastic film from the back of the pad.
- Discard the plastic film backing as well as the clear packaging.
STORING PADS ON PAD HOLDER Remove plastic film and put sticky side of pads on either side of the pad holder.
STEP 5 – ATTACH THE UNIT TO THE BELT CLIP
CAUTION The clip may fall off if attached to soft or thin clothing.
GET STARTED WITH YOUR THERAPY
(USE FOR A MAXIMUM OF 30 MINUTES PER SESSION)
STEP 1 – PAD PLACEMENT
CAUTION
For optimal therapy:
- Place pads on each side of the pain, not directly on the pain.
- Place pads at least 1 inch (3 cm) apart for optimal results.
- Therapy won’t work with just one pad. You MUST USE TWO PADS at the same time.
- Do not overlap pads or put pads on top of each other.
- Do not add spray, lotions or creams to skin or pads.
- Do not share pads with another person.
Before starting your therapy, rate your pain from 1 low to 10 high. This mental check gives you a basis you can compare to once the session is complete.
LOWER BACK
- Attach both pads on the lower back according to your pain.
- Place pads on muscle of back, not on spine.
GET STARTED WITH YOUR THERAPY
STEP 1 – PAD PLACEMENT (continued)
- LOWER BACK Attach one pad below and above the region in pain, both on same side.
- ARM Attach pads on each side of the region where you feel pain.
- JOINT (ELBOW) Attach pads on each side of the joint with the pain.
- LEG (HIP & THIGH) Attach pads on each side of the area with pain.
- JOINT (KNEE) Attach both pads above the knee or above and below the joint with pain.
- LEG (CALF) Attach both pads on the calf where you feel pain.
WARNING Do not apply the pads on the calves of both legs at the same time.
- FOOT (ANKLE) Attach pads on the left for pain on the outside of your ankle/foot. Attach the pads on the right for pain on the inside of your ankle/foot.
WARNING Do not apply the pads on the bottom of both feet at the same time.
- SHOULDER Attach both pads on the shoulders according to your pain.
- SHOULDER Attach one pad on the front and on the back of your shoulder.
WARNING
- Do not apply the pads on both sides of the chest cavity simultaneously (lateral or front and back), or across your chest because the introduction of electrical current may cause rhythm disturbances which could be lethal.
- Do not apply the pads near the heart.
STEP 2 – SELECT 1 OF 9 MODES
- Push “ON/OFF”.
- Choose 1 of the 9 modes. Modes cannot be combined.
Choose a massage-like mode
- Tap
- Knead
- Rub Or choose a pain mode:
- Arm
- Lower Back
- Leg
- Foot
- Joint (Knee/elbow/wrist)
- Shoulder
How to switch modes
- The unit automatically defaults to the last mode selected. Each time you push the mode button, it switches to the next mode at the lowest intensity. You can only use ONE MODE at a time.
How to select the right mode
- Any of the modes can be used on body parts or pains described in this manual or Quick Start Guide/Pad Placement Guide.
- Select the mode that feels right for your unique pain.
Therapies designed for | Arm | Lower Back | Leg | Foot |
---|---|---|---|---|
What does the therapy deliver? | Series of low to medium rate |
tapping, tingling and pulsing sensations.
| Series of high rate to low tingling sensations, followed by tapping. With higher intensity, you may feel kneading or massage-like sensations.| Series of low to medium tapping and rubbing sensations.| Series of low rate tapping, pulsing sensations.
Therapies designed for| Joint| Shoulder| Tap| Knead|
Rub
---|---|---|---|---|---
What does the therapy deliver?| Series of medium to high rate tapping, pulsing
sensations.| Series of low to high rate
tapping, pulsing, kneading and massage-like sensations.
| Series of low rate tapping sensations.| Series of medium rate pulsing
sensations to mimic massage.
| Series of high rate pulsing sensations to mimic hands rubbing.
STEP 3 – SELECT THE CORRECT INTENSITY LEVEL (1 LOW – 15 HIGH)
Start at the lowest intensity level and slowly increase it by pushing the “▲(Up)” arrow button. You should feel a gentle pulsing sensation.
How do I pick the right intensity level for my pain?
- Each time you push the “▲(Up)” or “▼(Down)” arrow, it moves to another level. If the stimulation sensation becomes weaker or disappears, increase the intensity until it is restored. But, if the sensation is at all uncomfortable, press the down arrow to decrease the intensity.
- Press ▲ for higher intensity.
- Press ▼ for lower intensity.
How long is the therapy?
- The unit will continue automatically for 15 minutes before it shuts off. We recommend a total of two 15-minutes therapy in one sitting, up to three times/day.
- The screen shows you how many minutes are remaining.
HOW TO CONTROL AND REDUCE YOUR PAIN
-
When should I start therapy?
Use as soon as your pain begins. Start with one session (unit automatically turns off at 15 minutes). -
Get to your pain early
If you get to your pain early, it may prevent the pain from becoming worse, or even chronic. It’s better for you to get it under control sooner so that it does not reach a high pain threshold where it limits your daily activities. -
How long should you use it?
Start with one 15-minute session. Rate your pain to check your progress, 1 low to 10 high. Stop therapy session if pain has reduced or stopped. Press the “On” button to continue therapy for another 15-minute session.- 1 session 15-minute automatic shut-off
- Max minutes/session 30 minutes
- Max times/day 3 times per day
CAUTION
Possible Adverse Reactions Do not use to treat one region for extended periods of time (more than 30 minutes a session, up to 3 times/day) or muscles in that region may become exhausted and sore.
When to stop using the unit?
- If you experience an adverse reaction (skin irritation/redness/burns, headache or other painful sensation, or if you feel any unusual discomfort).
- If your pain does not improve, becomes seriously chronic and severe, or continues for more than five days.
What type of pain is it best for?
- This therapy works best on acute pain because it is localized. Acute pain is pain in one area for less than 3 months. If you have chronic pain, you may have pain in more than one area and for longer than 6 months. Chronic pain may be compounded by other issues that this unit cannot address.
- Remember, this unit does not cure your pain or the original cause of the pain. It provides temporary relief or reduction of pain so that you can control your life and activities better.
CLEANING AND STORAGE
The unit is designed for repeated use over time. The pads will last up to 150 uses, or 5 months (based on use one time per day). Here are important cleaning and storage instructions:
Cleaning the pads
- Turn the power off and remove the electrode cord from the pads.
- Wash the pads when the adhesive surface becomes dirty and/or the pads are difficult to adhere.
- Wash the pad softly with your fingertips under slow running cold water for several seconds (do not use a sponge/cloth/sharp object like a nail on adhesive side, do not use detergents, chemicals or soap).
- Pads can be washed after 15 uses, approximately ten times for up to 150 uses. Do not wash the pads too long or too frequently.
- Dry the non-sticky surface of the pads and let the adhesive surface air-dry completely. Do not wipe with a tissue paper or cloth.
- Pads are replaceable and can be purchased when needed by calling 1-800-634-4350 or going to OmronHealthcare.com
The life of the pads may vary by how often the pads are washed, the skin condition, and how the pads are stored.
When should you replace your pads?
- Replace the pads when they no longer stick to your skin or if more than 25 % of the pad’s surface is not in contact with your skin.
Cleaning the unit
- Turn unit off and disconnect the electrode cords from the pads.
- Clean with a lightly moistened cloth (or a cloth soaked in a neutral cleaning solution) and wipe gently.
- Do not use chemicals (like thinner, benzene).
- Do not let water get into the internal area.
Storing the pads
-
Turn the unit off and remove the cord from the bottom of the unit.
-
Remove the pads from your body.
-
Leave the electrode cords connected to the pads.
Place the pads on the pad holder, one pad on each side with the sticky side of each pad on the pad holder. -
Wrap the electrode cords around the pad holder.
Storing the unit and pads
- Store the unit with the belt clip on. Store the pads with the electrode cords on the pad holder, and put into the pouch.
- Do not keep in areas subject to direct sunlight, high or low temperatures, humid areas, near a fire, vibration, or shock.
- Operating and storage temperature, +50 °F to +104 °F (+10 °C to +40 °C), 30 to 80 % relative humidity.
- Do not keep in places that can be easily reached by children.
- When not in use for a long period, remove the batteries before storage to avoid liquid discharge from batteries.
- Do not wrap the electrode cords around the unit because it may damage the cord.
TROUBLESHOOTING
If any of the below problems occur, check to make sure that no other electrical device is within 12 inches (30 cm). If the problem persists, refer to the table below.
If this happens… | Possible causes… | Try this solution… |
---|---|---|
The intensity is not felt. Very weak intensity level. | Are you using | |
only one pad? | Put the other pad on your skin. You must use BOTH PADS for |
therapy to work.
Have you removed the transparent
film from the pad?
| Peel off film on the adhesive
surface of pads.
Are the pads stacked together or do pads overlap?| Check placement of pads.
Refer to Pad Placement Guide.
Is the cord properly connected to the unit?| Connect cord plug correctly into
the jack at bottom of this unit.
Is the intensity setting getting weak?| Press the ▲ up button.
Is the gel damaged?| Replace pad.
Are the batteries weak?| Replace both AAA batteries.
Is the intensity “1”?| Press the ▲ up button.
The skin turns red or the skin feels irritated.| Is the adhesive surface
of pads dirty or dry?| Wash adhesive surface of pads softly with your
fingertips for about 3 seconds under slow running water.
Is therapy time too long?| Use less than 15 minutes.
Are the two pads attached properly to the body?| Refer to the Pad Placement
Guide and attach correctly.
Is the pad surface worn out?| Replace both pads at the same time.
No power source.| Are the polarities of battery (+ and -) aligned in the
wrong direction? Are the batteries depleted?| Check batteries for correct
alignment. Replace batteries.
Power cut off during use.| Are the batteries weak?| Replace both
batteries at the same time.
Is the cord broken?| Replace cord.
If this happens…| Possible causes…| Try this solution…
---|---|---
Battery icon is empty.| Are the batteries weak?| Replace both batteries
at the same time.
Gel pad does not stick to skin.| Have you removed the transparent
film from the pad?
| Peel off film from the adhesive
surface of pads.
Is the pad wet? Is your skin too wet?| Dry the pad. Dry the skin.
The gel pad may be damaged.| Replace the pad.
Is there too much hair on your skin?| Shave the immediate area for proper pad
adhesion.
Are you using pad while perspiring?| Dry the pad placement area.
Have the pads been washed too long and/or too frequently?| Leave the pad in
freezer for overnight.
Were the pads stored under high temperature, high humidity, or direct
sunshine?| Replace both pads.
Pad icon is displayed.
| Only one pad is attached, or both pads are not attached.| Reattach detached pad(s) onto
the skin firmly.
Have you removed the transparent
film from the pad?
| Peel off film from the adhesive
surface of pads.
Is the cord properly connected to the main unit?| Connect cord plug correctly
into the jack at the bottom of the main unit.
Is the adhesive surface of pads dirty or dry?| Wash adhesive surface of pads
softly with your fingertips for about 3 seconds under slow running water.
If the above measures are not effective, contact us at
1-800-634-4350.
LIMITED WARRANTY
- Your OMRON Max Power Relief unit, excluding the batteries and pads, is warranted to be free from defects in materials and workmanship appearing within 1 year from the date of purchase, when used in accordance with the instructions provided with the device. The above warranty extends only to the original retail purchaser.
- The pads are warranted to be free from defects in materials and workmanship appearing within 30 days from the date of purchase when the device is used in accordance with the instructions provided with the device. The above warranty extends only to the original retail purchaser.
- We will, at our option, replace without charge any unit covered by the above warranty. Replacement is our only responsibility and your only remedy under the above warranty.
- To obtain warranty service, contact Customer Service by calling 1-800-634-4350 for the address of the inspection center and the return shipping and handling fee that may apply. Enclose the proof of purchase. Include a letter, with your name, address, phone number, and description of the specific problem. Pack the product carefully to prevent damage in transit. Because of possible loss in transit, we recommend insuring the product with return receipt requested.
- THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY DISCLAIMS ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. IMPLIED WARRANTIES AND OTHER TERMS THAT MAY BE IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION TO THE PERIOD OF THE ABOVE EXPRESS WARRANTY.
- SOME STATES DO NOT ALLOW LIMITATIONS ON HOW LONG AN IMPLIED WARRANTY LASTS, SO THE ABOVE LIMITATION MAY NOT APPLY TO YOU.
- OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER SPECIAL, INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS, EXPENSES OR DAMAGES.
- SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGES, SO THE ABOVE LIMITATION OR EXCLUSION MAY NOT APPLY TO YOU.
- This warranty provides you with specific legal rights, and you may have other rights that vary by jurisdiction.
FOR CUSTOMER SERVICE
- Visit our web site at:OmronHealthcare.com
- Call toll free: 1-800-634-4350
SPECIFICATIONS
Product Name | OMRON Max Power Relief |
---|---|
Model # | PM500 |
Power Source | DC 3 V |
(2 AAA alkaline batteries or 2 AAA manganese batteries)
Battery Life| New batteries (2 AAA alkaline batteries) will last for approx. 4
months (when used for 15 minutes a day, Lower Back Mode, max. intensity).
Frequency| Approx. 1 to 238 Hz
PULSE Duration| 150 μsec
Maximum Output Voltage| 70 V (during 500 Ω load)
Power Control| 15 intensity levels
Operating and Storage Conditions (When using product)| +50 °F to +104 °F (+10
°C to +40 °C),
30 to 80 % RH (non-condensing), 700 to 1060 hPa
Transportation Conditions| -4 °F to +140 °F (-20 °C to +60 °C),
10 to 95 % RH, 700 to 1060 hPa
Weight| Approx. 3.5 oz (100 g) (incl. batteries)
Outer Dimension| Approx. 2” (W) × 4.4” (H) × 1” (D)
(52 mm × 112 mm × 25 mm)
Classification of ME equipment| Internally powered
IP classification| IP 22
Operation Mode| Continuous operation
Applied Part| Type BF (Pads)
Durable period (Service Life)| Device: 5 years
Pad: 150 uses
NOTE: These specifications are subject to change without notice.
This OMRON product is produced under the strict quality system of OMRON HEALTHCARE Co. Ltd., Japan. Designed for a minimum of 5 years life expectancy. IP classification is degrees of protection provided by enclosures in accordance with IEC 60529. This device is protected against solid foreign objects of diameter 12.5 mm such as a finger and greater, and against oblique falling water drops which gives trouble to normal operation.
Optional Accessories
- PMLLPAD-L (3”x4.5” (75 mm × 114 mm))
- PMLLPAD (2.5”x4” (63 mm × 102 mm))
To order: OmronHealthcare.com
FCC STATEMENT
FCC CAUTION Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
GUIDANCE AND MANUFACTURER’S DECLARATION
- OMRON Max Power Relief Information for Accompanying Documents in the Scope of IEC60601-1-2:2014
- Important information regarding Electromagnetic Compatibility (EMC) PM500 conforms to IEC60601-1-2:2014 Electromagnetic Compatibility (EMC) standard. Further documentation in accordance with this EMC standard is available at OmronHealthcare.com/emc. Refer to the EMC information for PM500 on the website.
OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, Kyoto 617-0002 JAPAN
Distributed by: OMRON HEALTHCARE, INC. 1925 West Field Court, Lake Forest, IL 60045 U.S.A. OmronHealthcare.com
© 2019 OMRON HEALTHCARE, INC. Made in China.
Omron Healthcare helps you make positive lifestyle changes with innovative medical solutions that you can use safely and accurately at home. OmronHealthcare.com
- For questions: 1-800-634-4350
- Get general pain info: OmronPainRelief.com/blog
FREQUENTLY ASKED QUESTIONS
What is the brand and model of the TENS therapy device?
The TENS therapy device is the OMRON PM500.
What are the product dimensions and weight of the OMRON PM500?
The OMRON PM500 has dimensions of 4.4 x 1 x 2 inches and weighs 6.4 ounces.
How many AAA batteries are required to operate the OMRON PM500?
The OMRON PM500 requires 2 AAA batteries for operation.
What is the power control feature of the OMRON PM500?
The OMRON PM500 offers 15 intensity levels for power control, allowing users to adjust the electrical impulses to their preferred level of comfort.
What is the classification of ME (Medical Electrical) equipment for the OMRON PM500?
The OMRON PM500 is classified as internally powered ME equipment.
What is the IP classification of the OMRON PM500?
The OMRON PM500 has an IP classification of IP22, indicating its resistance to water and dust ingress.
What is the warranty period for the OMRON PM500?
The OMRON PM500 comes with a 1-year warranty provided by Omron.
How much does the OMRON PM500 cost?
The price of the OMRON PM500 is $60.00.
Can the OMRON PM500 be used without an external power source?
Yes, the OMRON PM500 is internally powered, meaning it does not require an external power source for operation.
What is the brand of the manufacturer of the OMRON PM500?
The manufacturer of the OMRON PM500 is Omron.
What should I do if my OMRON PM500 TENS therapy device isn't turning on?
First, ensure that the batteries are correctly inserted and have sufficient charge. If the device still doesn't turn on, try replacing the batteries with new ones.
How can I troubleshoot if the intensity levels on my OMRON PM500 are not adjusting properly?
Check the electrode pads to ensure they are securely connected to the device and placed correctly on your skin. If the intensity levels still aren't adjusting, try cleaning the electrode pads and the skin area where they are applied.
What should I do if my OMRON PM500 TENS therapy device is producing weak or inconsistent electrical impulses?
First, check the battery level to ensure it's not low. If the batteries are fine, inspect the electrode pads for any signs of wear or damage.
How can I troubleshoot if the OMRON PM500 is causing discomfort or irritation during use?
Ensure that the electrode pads are clean and positioned correctly on your skin. If discomfort persists, try adjusting the intensity level or using a different set of electrode pads. If the issue persists, discontinue use and consult with a healthcare professional.
What should I do if the OMRON PM500 TENS therapy device is emitting an unusual smell or overheating?
Immediately turn off the device and remove the batteries. Allow it to cool down completely before inspecting it for any signs of damage.