SIGLENT SCP5000 Series Current Probes Instructions
- May 15, 2024
- SIGLENT
Table of Contents
- SIGLENT SCP5000 Series Current Probes
- Safety Notices
- Features and Applications
- Description of Products
- Specifications
- Mechanical Specifications
- Environmental Characteristics
- Operating Method
- Methods to Deal With Abnormal Situations
- Q&A
- Care and Maintenance
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
SIGLENT SCP5000 Series Current Probes
Safety Notices
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating
procedure, practice, or the like that, if not correctly performed or adhered
to, could result in personal injury or death. Do not proceed beyond a WARNING
notice until the indicated conditions are fully understood and met.
Safe Probing
This device is designed to comply with Safety Standards and has been
thoroughly tested for safety prior to shipment. However, mishandling during
use could result in injury or death, as well as damage to the device. Be
certain that you understand the instructions and precautions in the manual
before use. We disclaim any responsibility for accidents or injuries not
resulting directly from device defects.
WARNING
- Never attach the clamp to a circuit that operates over the maximum rated voltage to earth.
- Please avoid clamping around bare conductors during measurement.
- While clamping and measuring, do not touch the clamp in front of the barrier or the conductor being measured.
- Be careful to avoid damaging the insulation surface while taking measurements.
- Make sure that the waveform measuring equipment connected to this device’s output terminal (SAPBUS) is equipped with a protective earthling with double-insulation construction.
- Do not allow the device to get wet, and do not take measurements with wet hands. This may cause an electric shock.
- If the waveform measuring instrument being connected to the output terminal (SAPBUS) on this device is equipped with any other measurement terminals, take the following precautions to ensure that the other instrument does not form a bridge between the probe and any hazardous live part of a part.
- Isolate the terminal to which the probe is connected from other terminals on the measuring instrument using basic insulation conforming to the measurement category, working voltage, and pollution degree requirements of the circuit being tested.
- If basic insulation requirements cannot be met between the terminal to which this device is connected and other terminals of the measuring instrument, make sure that the voltage input to the measurement terminal does not exceed the Separated Extra-Low Voltage Earthed.
- Read and observe all warnings and precautions relating to electrical safety for the measuring instrument being connected to the probe.
CAUTION
- To avoid damage to the device, protect it from vibration or shock during transport and handling, and be especially careful to avoid dropping.
- Do not store or use the device where it could be exposed to direct sunlight, high temperature, humidity, or condensation. Under such conditions, the device may be damaged and insulation may deteriorate so that it no longer meets specifications.
- Before using the device the first time, verify that it operates normally to ensure that the no damage occurred during storage or shipping. If you find any damage, contact your dealer or SIGLENT representative.
- This device is not designed to be entirely water- or dust- proof. To avoid damage, do not use it in a wet or dusty environment.
- The sensor head is a precision assembly including a molded component, a ferrite core, and a Hall Effect element. It may be damaged if subjected to sudden changes in ambient temperature, or mechanical strain or shock, and therefore great care should be exercised in handling it.
- The matching surfaces of the sensor head are precision ground, and should be treated with care. If these surfaces are scratched, performance may be impaired.
- Foreign substances such as dust on the contact surfaces of the sensor head can cause acoustic resonance and degrade measurement, so it should be cleaned by gently wiping with a soft cloth.
- To avoid damaging the sensor cable and power supply cable, do not bend or pull the cables.
- When the power is on, keep closed, except when clamping them onto the conductor to be measured. The facing surface of the core section can be scratched while it is open.
- Do not place any un-clamped conductor with an electric current of a frequency of 10 kHz or more near the sensor head. Current flowing in the conductor nearby may heat up the sensor head and cause its temperature to rise, leading to damage to the sensor. For example, when one side of a go-and-return conductor is clamped and the other side is also placed near the sensor head even if the electric current is lower than the consecutive maximum current, electric currents in both sides will heat up the wires and raise the temperature, thereby causing damage to the sensor.
- The maximum continuous input range is based on heat that is internally generated during measurement. Never input current more than this level. Exceeding the rated level may result in damage to the probe.
- The maximum continuous input range varies according to the frequency of the current being measured.
- If excess current is input, generated heat activates a built-in safety function that blocks normal output. If this happens, remove the input immediately (remove the sensor from the conductor being measured or reduce the input current to zero). Wait until the sensor has had sufficient time to cool before resuming operation.
- Even if the input current does not exceed the rated continuous maximum, continuous input for an extended period may result in activation of the safety circuit to prevent damage resulting from heating of the sensor.
- At high ambient temperatures, the built-in safety circuit may activate at current input levels below the rated continuous maximum.
- Continuous input of current exceeding the rated maximum or repeated activation of the safety function may result in damage to the unit.
- The probe is rated for maximum input under two conditions in addition to the input maximums shown in the Specifications. These are (1) 30 A peak for non-continuous input and (2) 50 A peak for pulse widths 10 µs. (1) indicates an upper waveform response limit of 30 A peak. Use the sensor at RMS current input levels that are within the rated continuous maximums. (2) Indicates the upper response limit for a single input pulse.
- When opening the sensor head of the probe, be sure to operate with the opening lever. If an upper core is forced to open when the sensor head is locked, the open-close mechanism can be damaged.
NOTE
- The output of this unit is terminated internally. Use an oscilloscope with an input impedance of at least 1 MΩ.
- Immediately after powering on the probe, the probe may be subject to an appreciable offset drift due to the effect of self-heating. To counteract this, allow the probe to warm up for about 30 minutes before carrying out measurement.
- When performing continuous measurements, it is necessary to be aware that the offset voltage drifts, depending on factors such as the ambient temperature.
- Under certain circumstances, oscillation may occur if the probe is connected to the power supply while the power supply is on. This does not indicate a malfunction. Oscillation can be stopped and operation restored to normal by opening and closing the sensor head.
- Depending on the measured current frequency, some sound maybe produced by resonance, but does not affect measurements.
- The reading may be affected by the position within the clamp aperture of the conductor being measured. The conductor should be in the center of the clamp aperture.
- When carrying out a measurement, push the opening lever of the sensor to lock the probe and check that the sensor head is properly closed. If the sensor head is not properly closed, an accurate measurement is not possible.
- Accurate measurement may be impossible in locations subject to strong external magnetic fields, such as transformers and high-current conductors, or in locations subject to strong external electric fields, such as radio transmission equipment.
- At high frequencies, common mode noise may affect measurements taken on the high-voltage side of circuits. If this occurs, reduce the frequency range of the waveform measuring instrument or clamp onto the low- voltage side of the circuit.
Features and Applications
The SCP5000 series current probes are wide band width DC / AC active current probes, featuring high bandwidth, fast and accurate capture the current wave, accuracy up to 1% and low circuit insertion loss. This probe can be used with any oscilloscope having a high-impedance SAPBUS input.
The key features include:
- Highly accurate current measurements
- Wide bandwidth
- Accurate and easy current measurements
- DC/AC measuring capabilities
- Over-current protection with popup indicator information
- High and low range selection
- Low current measurements
- Degaussing and automatic zero setting
- UI visualization operation design, more convenient and faster to use
Applications
- Switching and linear power design
- LED lighting design
- New energy resources
- Frequency conversion household appliances
- Experiment of electronic engineering
- Semiconductor devices design
- Inverters/transformer design
- Electronic ballast design
- Industrial control/consumer electronic design
- Engine driven design
- Power electronic and electrical drive experiment
- Electric vehicle transportation design
Description of Products
SCP5030, SCP5030A:
SCP5150, SCP5500:
Sensor Head
The core component to measure conductor current. The component contains a
precise semi-conductor that could be damaged by drastic change of
environmental temperature, external pressure and shock. Please be careful
during measurement.
Opening Lever
The operating lever used to open the sensor head. Move the lever to open the
sensor jaw, put in the cable under test, and push the lever to lock the sensor
head to measure the current.
Output Connector
This probe can be used with any oscilloscope having a SAPBUS input.
Description of Operation Interface
Connect the SCP5000 with oscilloscope, click on the channel parameter area in
the bottom left corner of the oscilloscope user interface to open the channel
parameter setting dialog box. Click “Probe” to pop up the probe information
interface, where you can set the probe.
Toggle lock
Use to lock probe to the BNC connector of the oscilloscope or unlock from the
scope.
The information includes the following contents:
- Probe Type : The type of the currently connected probe.
- Probe SN : The serial number of the currently connected probe.
- Bandwidth (-3dB)
- Attenuation: Attenuation ratio.
- Resistance: Oscilloscope input resistance value.
- RatedCurrent: Rated current value of the current range.
- CurrentRange: Current range state. Auto: Automatic mode; Lock to High: The probe range is locked to the maximum attenuation ratio.
- Degauss / AutoZero: Frequent usage of the device will generate a residual magnetic field. Please degauss and zero set before measurement for better measurement precision. Click the “Degauss” and “AutoZero” to trigger the process.
- Coupling: DC coupling: passing through all components of the signal; AC coupling: suppressing the DC component of the signal; GND: Grounding.
Specifications
Model | SCP5030 ( A ) | SCP5150 | SCP5500 |
---|
Bandwidth (-3 dB)
| SCP5030| DC-50 MHz
(picture1.a)
| ****
DC-12 MHz
(Figure 4)
| ****
DC-2 MHz
(Figure 7)
SCP5030A| DC-100 MHz
(picture1.b)
Rise Time
| SCP5030| ≤ 7 ns| ****
≤29 ns
| ****
≤175 ns
SCP5030A| ≤ 3.5 ns
Continuous maximum input range
| SCP5030| 30 Arms (picture2.a)| ****
150 Arms
(Figure 5)
| ****
500 Arms
(Figure 8)
SCP5030A| 30 Arms (picture2.b)
Max peak current value| 50 Apk| 300 Apk| 750 Apk
Range
| 5 A| 1X| 30 A| 10X| 75 A| 10X
30 A| 10X| 150 A| 100X| 500 A| 100X
Overload
| 5 A| ≥5 A| 30 A| ≥30 A| 75 A| ≥75 A
30 A| ≥50 A| 150 A| ≥300 A| 500 A| ≥750 A
Current transfer ratio
| 5 A| 1 V/A| 30 A| 0.1 V/A| 75 A| 0.1 V/A
30 A| 0.1 V/A| 150 A| 0.01 V/A| 500 A| 0.01 V/A
Resolution
| 5 A| 1 mA| 30 A| 10 mA| 75 A| 10 mA
30 A| 10 mA| 150 A| 100 mA| 500 A| 100 mA
Accuracy
(DC, 45-66 Hz)
| 5 A| ±1%±1 mA| 30 A| ±1%±10
mA
| 75 A| ±1%±10
mA
30 A| ±1%±10 mA| 150 A| ±1%±100
mA
| 500 A| ±1%±100
mA
Input resistance
| SCP5030| Reference (Figure 3.a)| ****
Reference (Figure 6)
| ****
Reference (Figure 9)
SCP5030A| Reference (Figure 3.b)
Delay
| Probe| 14 ns| 36 ns| 42 ns
BUS cable (1 .5m)| 7.5 ns
Terminal load requirements| ≥100 kΩ
Power Supply| Directly powered by the oscilloscope through SAPBUS
Voltage of insulated wire| 300 V CAT I| 600 V CATII 300 V CATIII
Safety compliance| EN61010-1: 2010+A1:2019 EN 61010-2-032:2019
EMC standard| EN61326-1:2013 EN61000-3-2:2014 EN61000-3-3:2013
Mechanical Specifications
Model
| ****
SCP5030/SCP5030A
| ****
SCP5150 / SCP5500
---|---|---
Measurement conductor diameter max| 5 mm| 20 mm
Clamp dimensions ( LWH )| 17639.518 mm| 17467.530 mm
Termination unit ( LWH )| 923924.5 mm
Probe weight| 270 g| 475 g
Environmental Characteristics
Operating temperature and humidity | 0-40℃, 80% or less |
---|---|
Storage temperature and humidity | -10-50℃, 80% or less |
Operating altitude | 2000 m |
Storage altitude | 12000 m |
Operating Method
Note
- The output interface of this machine is set inside. When using the oscilloscope, it is automatically configured with high input resistance (1 MΩ).
- Please make sure the current measured doesn’t surpass the maximum current. The magnetic core will saturate. The saturated magnetic core will neutralize the generate waveform during saturation. The overcharged inrush might cause mistaken degaussing and need to be zero set again.
- When power is connected, offset might occur because of the heat generated by the machine. But it will be stabilized after about 30 min.
- Strong magnetic field like transformer, large circuit, high electricity like wireless will cause deviation.
- The voltage might deviate because of the surrounding temperature, so please be careful when testing sequentially.
- The frequency of the current under test may cause resonance, but this won’t influence the testing.
- The position of conductor under test in the sensor will influence the result, so please move the conductor under test into the center of the sensor.
- Before measuring, push the opening lever to the lock state and confirm that the opening lever is locked and the entire part is indeed closed. If the entire part is not truly closed, it will not be able to measure normally.
Attention
- When disconnecting the output terminal, please pull out the connector after unlocking. The output terminal will be damaged if you force to drag the cable out before unlocking.
- The continuous maximum input range is the fixed value caused by the machine’s operating heat. Please do not put in current higher than this value, or the device will be damaged.
- The continuous maximum input range will change according to the frequency of the current under test. The probe will be damaged when operated under an overcharged current.
- When the input current continuously surpasses the maximum input range, the self-protection will be activated by the heating of the sensor and cause the wrong output. Please stop the current input and wait for full cool down before next operation.
- The protection circuit will be mistakenly activated by the high temperature even when the continuous current under test is below the max input.
- When the connect input surpass the max input range current and activates the protection function too often, the device may be damaged.
- You must open the entire part through the switch controller.
- At the lock state, please do not press the entire part as shown below.
Preparation before testing
- Prepare the high frequency current probe SCP5000 series and oscilloscope.
- Set the oscilloscope: Zero set the oscilloscope and turn the oscilloscope mode to DC mode.
- Choose the proper range according to the current under test.
Degaussing and Zero setting
- Connect the SCP5000 with oscilloscope (Make sure the input impedance of the oscilloscope is 1 MΩ).
- Lock the probe and confirm that the opening lever is in lock state.
- In the probe setting interface of the oscilloscope, click “Degauss / AutoZero” to degauss and zero set.
Measurement methods:
- Confirm the previous steps.
- Move the switch control pole of the sensor, open the head of the sensor and make the current direction mark on the front end of the sensor consistent with the flow direction of the measured current, and clamp the measured conductor in the middle of the sensor head.
- Move the switch control pole of the sensor to lock state, make sure the entire part is closed.
- Configure appropriate vertical sensitivity and observe the waveform under test.
Methods to Deal With Abnormal Situations
Situation | Possible reason | Dealing method |
---|
Can’t measure DC, or the value obtained is comparatively low in the frequency range
| Abnormal power supply| Reconnect SAPBUS
Oscilloscope set to AC coupling| Set to DC coupling
Sensor is not locked| Please lock the sensor
Auto degaussing or zero setting unsuccessful| The probe is on the operating
circuit under test when degaussing or zerosetting is applied| ****
Turn off the circuit under test andzero set again.
The amplitude is comparatively
low in the frequency range
| The input resistance of the test equipment like oscilloscope is 50 Ω| ****
Set the resistance over 1 MΩ
Q&A
Does the SCP5000 series fit the oscilloscope of any brand?
A: SCP5000 series has a standard SAPBUS interface can be applied to the
oscilloscope equipped with SAPBUS, and the probe is powered directly by the
oscilloscope.
Can the SCP5000 series product measure small current?
A: Yes. For now, the SCP5000 series current probe has two optional ranges, and
one is for small currents. The current resolution of the SCP5030(A) is 1 mA.
When measuring small current, please accurately zero set and degaussing the
probe, and do not change the position of the probe hand grip. To observe the
waveform please set the bandwidth restriction of the oscilloscope to 20 MHz to
eliminate the interference of noise. When measuring an extremely small current
(a few mA for example), one could make a few more loop of cable around the
probe and divide the result with some loops to obtain the actual current
value.
Any more tips?
A:
- When measuring high-frequency current, please do not let the current surpass the value shown by the curve of max peak current vs frequency. The max continuous current over the curve will burn the probe.
- To measure accurately, please degauss and zero set the probe, and make sure the probe is locked during the process.
- Set the input impedance of the oscilloscope to 1 MΩ (default)
- Make sure the probe is locked during testing.
- The probe should be away from the interference source like the transformer. The method to judge if the probe is interfered is to put the probe close to circuit under test. IF there’s any output, there could be interference in the testing environment because the probe is not on the circuit yet.
- The current under test should not surpass the limit value of the probe.
- Please always maintain your probe and do not use it in a humid environment
- If there’s anything wrong with the probe, please set it back for repair. If you dismantle the device on your own, we won’t guarantee repair.
Packing List
Packing list
Item| Quantity
Probe| 1
Instruction manual| 1
Test report| 1
Care and Maintenance
- Keep the probe clean and dry.
- Please wipe with soft dry cloth when needed, must not use chemicals to clean.
- Please put the probe in the package provided, and put it in cool, clean and dry places.
- Please put the probe in the package provided to prevent shock.
- Do not forcefully pull the input and output lead to prevent bending, twisting, and folding.
Contact SIGLENT
SIGLENT TECHNOLOGIES CO., LTD
- Address: 3/F, Building 4, Antongda Industrial Zone, Liuxian Road, Baoan District, Shenzhen, P.R. CHINA
- Service Tel: 0086 755 3688-7876
- Post Code: 518101
- E-mail:sales@siglent.com
- http://www.siglent.com
Headquarters:
SIGLENT Technologies Co., Ltd Add: Bldg No.4 & No.5, Antongda Industrial Zone,
3rd Liuxian Road, Bao’an District, Shenzhen, 518101, China
- Tel: + 86 755 3688 7876
- Fax: + 86 755 3359 1582
- Email: sales@siglent.com
- Website: int.siglent.com
North America:
SIGLENT Technologies America, Inc 6557 Cochran Rd Solon, Ohio 44139 Tel:
440-398-5800
- Toll Free: 877-515-5551
- Fax: 440-399-1211
- Email: info@siglentna.com Website: www.siglentna.com
- Europe: SIGLENT Technologies Germany GmbH
- Add: Staetzlinger Str. 70 86165 Augsburg, Germany
- Tel: +49(0)-821-666 0 111 0
- Fax: +49(0)-821-666 0 111 22 Email: info-eu@siglent.com Website: www.siglenteu.com
About SIGLENT
SIGLENT is an international high-tech company, concentrating on R&D, sales,
production and services of electronic test & measurement instruments. SIGLENT
first began developing digital oscilloscopes independently in 2002. After more
than a decade of continuous development, SIGLENT has extended its product line
to include digital oscilloscopes, isolated handheld oscilloscopes,
function/arbitrary waveform generators, RF/MW signal generators, spectrum
analyzers, vector network analyzers, digital multimeters, DC power supplies,
electronic loads and other general purpose test instrumentation. Since its
first oscilloscope was launched in 2005, SIGLENT has become the fastest-
growing manufacturer of digital oscilloscopes. We firmly believe that today
SIGLENT is the best value in electronic test & measurement.
Follow us on Facebook: SiglentTech
References
- Expert in Test & measurement Solution - SIGLENT Technologies
- 数字示波器_射频信号发生器_频谱分析仪-鼎阳官方网站
- 数字示波器_射频信号发生器_频谱分析仪-鼎阳官方网站
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>