KROHNE 5200 C Optiswitch Instruction Manual
- June 16, 2024
- KROHNE
Table of Contents
KROHNE 5200 C Optiswitch
About this document
Function
This operating instructions manual provides all the information you need for
mounting, connection and setup as well as important instructions for
maintenance and fault rectification. Please read this information before
putting the instrument into operation and keep this manual accessible near the
device.
Target group
This operating instructions manual is directed to trained specialist
personnel. The contents of this manual should be made available to these
personnel and put into practice by them.
Symbols used
-
Information, tip, note
This symbol indicates helpful additional information. -
Caution: If this warning is ignored, faults or malfunctions can result.
-
Warning: If this warning is ignored, injury to persons and/or serious damage to the instrument can result.
-
Danger: If this warning is ignored, serious injury to persons and/or destruction of the instrument can result.
-
Ex applications
This symbol indicates special instructions for Ex applications. -
SIL applications
This symbol indicates instructions for functional safety which must be taken into account, particularly for safety-relevant applications. -
List
The dot set in front indicates a list with no implied sequence. -
Action
This arrow indicates a single action. -
Sequence of actions
Numbers set in front indicate successive steps in a procedure. -
Battery disposal
This symbol indicates special information about the disposal of batteries and accumulators.
For your safety
Authorised personnel
All operations described in this operating instructions manual must be carried
out only by trained specialist personnel authorised by the plant operator.
During work on and with the device, the required personal protective equipment
must always be worn.
Appropriate use
The OPTISWITCH 5200 C, 5250 C is a sensor for point-level detection. You can
find detailed information about the area of application in the chapter
“Product description”.
Operational reliability is ensured only if the instrument is properly used
according to the specifications in the operating instructions manual as well
as possible supplementary instructions.
For safety and warranty reasons, any invasive work on the device beyond that
described in the operating instructions manual may be carried out only by
personnel authorised by the manufacturer. Arbitrary conversions or
modifications are explicitly forbidden.
Warning about incorrect use
Inappropriate or incorrect use of the instrument can give rise to application-
specific hazards, e.g. vessel overfill or damage to system components through
incorrect mounting or adjustment.
General safety instructions
This is a state-of-the-art instrument complying with all prevailing
regulations and guidelines. The instrument must only be operated in a
technically flawless and reliable condition. The operator is responsible for
the trouble-free operation of the instrument.
During the entire duration of use, the user is obliged to determine the
compliance of the necessary occupational safety measures with the currently
valid rules and regulations and also take note of new regulations. The safety
instructions in this operating instructions manual, the national installation
standards as well as the valid safety regulations and accident prevention
rules must be observed by the user. For safety and warranty reasons, any
invasive work on the device beyond that described in the operating
instructions manual may be carried out only by personnel authorised by the
manufacturer. Arbitrary conversions or modifications are explicitly forbidden.
The safety approval markings and safety tips on the device must also be
observed.
Safety label on the instrument
The safety approval markings and safety tips on the device must be observed.
CE conformity
This device fulfils the legal requirements of the applicable EC guidelines.
By attaching the CE mark, we confirm successful testing.
SIL conformity
OPTISWITCH 5200 C, 5250 C fulfills the requirements on functional safety
according to IEC 61508 resp. IEC 61511. You can find further information in
the Safety Manual “OPTISWITCH 5XXX”.
Safety instructions for Ex areas
Please note the Ex-specific safety information for installation and operation
in Ex areas. These safety instructions are part of the operating instructions
manual and come with the Ex-approved instruments.
Product Description
Configuration
Scope of delivery
The scope of delivery encompasses:
- OPTIC SWITCH 5200 C, 5250 C point level switch
- Documentation
- this operating instructions manual
- Safety Manual “Functional safety (SIL)” (optional)
- Ex-specific “Safety instructions” (with Ex versions)
Constituent parts
The OPTIC SWITCH 5200 C, 5250 C consists of the components:
- Housing lid
- Housing with electronics
- Process fitting with a tuning fork
Type label
The type label contains the most important data for the identification and use of the instrument:
- Article number
- Serial number
- Technical data
- Article numbers, documentation
- SIL identification (with SIL rating ex works)
In addition to the type label outside the instrument, you find the serial number also inside the instrument.
Principle of operation
Application area
OPTISWITCH 5200 C, 5250 C is a point-level sensor with a tuning fork for
point-level detection. It is designed for industrial use in all areas of
process technology and can be used in liquids.
Typical applications are overfill and dry run protection. With its tuning fork
of only 40 mm in length, OPTISWITCH 5200 C, 5250 C can also be mounted e.g. in
pipelines from DN 32. The small tuning fork allows use in vessels, tanks and
pipes. Thanks to its simple and robust measuring system, OPTISWITCH 5200 C,
5250 C is virtually unaffected by the chemical and physical properties of the
liquid. It functions even under difficult conditions such as turbulence, air
bubbles, foam generation, buildup, strong external vibration or changing
products.
Function monitoring
The electronics module of OPTIC SWITCH 5200 C, 5250 C continuously monitors
the following criteria via frequency evaluation:
- Strong corrosion or damage on the tuning fork
- Loss of vibration
- Line break to the piezo drive
If a malfunction is detected or in case of power failure, the electronics take on a defined switching condition, i.e. the relay deenergises (safe state).
Functional principle
The tuning fork is piezoelectrically energised and vibrates at its mechanical resonance frequency of approx. 1200 Hz. The piezos are fixed mechanically and are hence not subject to temperature shock limitations. The frequency changes when the tuning fork is covered by the medium. This change is detected by the integrated electronics module and converted into a switching command.
Voltage supply
OPTISWITCH 5200 C, 5250 C is a compact instrument, i.e. it can be operated
without an external evaluation system. The integrated electronics evaluates
the level signal and outputs a switching signal. With this switching signal, a
connected device can be operated directly (e.g. a warning system, a pump
etc.).
The data for power supply are specified in the chapter “Technical data”.
Operation
The switching condition of OPTISWITCH 5200 C, 5250 C with plastic housing can
be checked when the housing is closed (signal lamp). With the basic setting,
products with a density ≥ 0.7 g/cm³ (0.025 lbs/ in³) can be detected. The
instrument can be adapted if products with lower density are to be measured.
On the electronics module, you will find the following display and adjustment
elements:
- Signal lamp for indication of the switching condition (green/red)
- DIL switch for sensitivity adjustment
- Mode adjustment for selection of the switching condition (A/B)
Storage and transport
Packaging
Your instrument was protected by packaging during transport. Its capacity to
handle normal loads during transport is assured by a test based on ISO 4180.
The packaging of standard instruments consists of environment-friendly,
recyclable carton material. The sensing element is additionally protected with
a cardboard cover. For special versions, PE foam or PE foil is also used.
Please dispose of the packaging material through specialised recycling
companies.
Transport
Transport must be carried out in due consideration of the notes on the transport packaging. Nonobservance of these instructions can cause damage to the device.
Transport inspection
The delivery must be checked for completeness and possible transit damage immediately at receipt. Ascertained transit damage or concealed defects must be appropriately dealt with.
Storage
Up to the time of installation, the packages must be left closed and stored
according to the orientation and storage markings on the outside.
Unless otherwise indicated, the packages must be stored only under the
following conditions:
- Not in the open
- Dry and dust-free
- Not exposed to corrosive media
- Protected against solar radiation
- Avoiding mechanical shock and vibration
Storage and transport temperature
- • Storage and transport temperature see chapter “Supplement – Technical data – Ambient conditions”
- Relative humidity 20 … 85 %
Mounting
General instructions
Suitability for the process conditions
- Make sure that all parts of the instrument coming in direct contact with the process, especially the sensor element, process seal and process fitting, are suitable for the existing process conditions, such as process pressure, process temperature as well as the chemical properties of the medium.
- You can find the specifications in the chapter “Technical data” and on the nameplate.
Switching point
In general, OPTISWITCH 5200 C, and 5250 C can be installed in any position.
The instrument only has to be mounted in such a way that the tuning fork is at
the height of the desired switching point.
The tuning fork has lateral markings (notches) that indicate the switching
point with vertical mounting. The switching point applies to water in
conjunction with the basic setting of the density switch ≥ 0.7 g/cm³ (0.025
lbs/in³). When mounting OPTISWITCH 5200 C, 5250 C, make sure that this marking
is at the height of the requested switching point. Keep in mind that the
switching point of the instrument will shift if the medium has a density other
than water – water is 1 g/cm³ (0.036 lbs/in³). For products ≤ 0.7 g/cm³ (0.025
lbs/in³) and ≥ 0.5 g/cm³ (0.018 lbs/in³) the density switch must be set to ≥
0.5 g/ cm³.
Keep in mind that foams with a density ≥ 0.45 g/cm³ (0.016 lbs/in³) are
detected by the sensor. This can lead to erroneous switching, particularly
when the sensor is used for dry-run protection.
-
Switching point approx. 13 mm (0.51 in)
-
Switching point with lower density
-
Switching point with higher density
-
Switching point approx. 27 mm (1.06 in)
-
Switching point
-
Switching point
-
Marking on top with threaded versions, marking aligned to flange holes with flange versions
With flange versions, the fork is aligned to the flange holes as follows.
Moisture
Use the recommended cables (see chapter “Connecting to power supply”) and
tighten the cable gland.
You can give your OPTISWITCH 5200 C, and 5250 C additional protection against
moisture penetration by leading the connection cable downward in front of the
cable entry. Rain and condensation water can thus drain off. This applies
mainly to outdoor mounting as well as installation in areas where high
humidity is expected (e.g. through cleaning processes) or on cooled or heated
vessels.
Transport
Caution
Do not hold OPTISWITCH 5200 C, 5250 C on the tuning fork. Particularly with
flange or tube versions, the tuning fork can be damaged just by the weight of
the instrument. Transport coated instruments very carefully and avoid touching
the tuning fork.
Remove the packaging or the protective cover just before installation.
Pressure/Vacuum
The process fitting must be sealed if there is a gauge or low pressure in the
vessel. Before use, check if the seal material is resistant to the measured
product and the process temperature.
The max. permissible pressure is specified in the chapter “Technical data” or
on the type label of the sensor.
Handling
The vibrating level switch is a measuring instrument and must be treated accordingly. Bending the vibrating element will destroy the instrument.
Warning
The housing must not be used to screw the instrument in! Applying tightening
force can damage internal parts of the housing.
Use the hexagon above the thread for screwing in.
Mounting instructions
Welding socket
OPTISWITCH 5200 C, 5250 C has a defined thread starting point. This means that
every OPTISWITCH 5200 C, 5250 C is in the same fork position after being
screwed in. Remove therefore the supplied seal from the thread of OPTISWITCH
5200 C, 5250 C. This seal is not required when using a welded socket with an
O-ring in front. Keep in mind that this welded socket is not suitable for
coated instrument versions.
Screw OPTIC SWITCH 5200 C, 5250 C completely into the welded socket. The
latter position can be determined already before welding. Mark the appropriate
position of the welded socket. Before welding, unscrew OPTISWITCH 5200 C, and
5250 C and remove the rubber ring from the welded socket. The welded socket
has a marking (notch). Weld the socket with the notch facing upward, or in the
case of pipelines (DN 32 up to DN 50), aligned with the direction of flow.
Adhesive products
In the case of horizontal mounting in adhesive and viscous products, the
surfaces of the tuning fork should be vertical to reduce buildup on the tuning
fork. On the screwed version you will find a marking on the hexagon. With
this, you can check the position of the tuning fork when screwing it in. When
the hexagon touches the seal, the thread can still be turned by approx. half a
turn. This is sufficient to reach the recommended installation position.
In the case of flange versions, the fork is aligned with the flange holes.
When used in adhesive and viscous products, the tuning fork should protrude
into the vessel to avoid buildup. For that reason, sockets for flanges and
mounting bosses should be avoided when mounting horizontally.
Inflowing medium
If OPTISWITCH 5200 C, 5250 C is mounted in the filling stream, unwanted false
measurement signals can be generated. For this reason, mount OPTISWITCH 5200
C, 5250 C at a position in the vessel where no disturbances, e.g. from filling
openings, agitators, etc., can occur.
This applies particularly to instrument types with long extension tubes.
Product flow
To make sure the tuning fork of OPTIC SWITCH 5200 C, 5250 C generates as little resistance as possible to product flow, mount the sensor so that the surfaces are parallel to the product movement.
Agitators
-
Due to agitators, vibrations or similar, the level switch can be subjected to strong lateral forces. For this reason, do not use an overly long extension tube for OPTISWITCH 5200 C, 5250 C, but check if you can mount an OPTISWITCH 5100 C, 5150 C level switch on the side of the vessel in the horizontal position.
-
Extreme vibration caused by the process or the equipment, e.g.
agitators or turbulence in the vessel can cause the extension tube of OPTISWITCH 5200 C, 5250 C to vibrate in resonance. This leads to increased stress on the upper weld joint. Should a longer tube version be necessary, you can provide suitable support directly above the tuning fork to secure the extension tube. -
This measure applies mainly to applications in Ex areas category 1G or WHG. Make sure that the tube is not subject to bending stress due to this measure.
Enamel coating
Instruments with enamel coating should be treated very carefully and shocks should be avoided. Unpack OPTISWITCH 5200 C, 5250 C directly before installation. Insert OPTISWITCH 5200 C, 5250 C carefully into the vessel opening and avoid touching any sharp vessel parts.
Gas-tight leadthrough
The second seal of the gas-tight leadthrough (option) prevents an uncontrolled leakage of the medium. The service life of the gas-tight leadthrough depends on the chemical resistance of the materials. See “Technical data”.
Caution
If it is determined (e.g. via an error message from OPTISWITCH 5200 C, 5250 C)
that medium has already penetrated into the vibrating element, the instrument
must be exchanged immediately.
Connecting to power supply
Preparing the connection
Note safety instructions
Always keep in mind the following safety instructions:
Warning
Connect only in the complete absence of line voltage.
- The electrical connection must only be carried out by trained personnel authorised by the plant operator.
- Always switch off power supply, before connecting or disconnecting the instrument.
Take note of safety instructions for Ex applications
- In hazardous areas you must take note of the respective regulations, conformity and type approval certificates of the sensors and power supply units.
Voltage supply
- Connect the voltage supply according to the connection diagrams.
- The electronics module with relay output is designed in protection class I. To maintain this protection class, it is absolutely necessary that the earth conductor be connected to the inner earth conductor terminal. Keep the general installation regulations in mind. Take note of the corresponding installation regulations for hazardous areas with Ex applications.
- The data for power supply are specified in chapter “Technical data”.
Connection cable
- The instrument is connected with standard three-wire cable without screen. If electromagnetic interference is expected which is above the test values of EN 61326 for industrial areas, screened cable should be used.
- Use cable with round cross-section. A cable outer diameter of 5 … 9 mm (0.2 … 0.35 in) ensures the seal effect of the cable gland.
- If you are using cable with a different diameter or cross-section, exchange the seal or use a suitable cable gland.
- In hazardous areas, use only approved cable connections for OPTISWITCH 5200 C, 5250 C.
Connection cable for Ex applications
- Take note of the corresponding installation regulations for Ex applications.
- Cover all housing openings conforming to standard according to EN 60079-1.
Connection procedure
- With Ex instruments, the housing cover may only be opened if there is no explosive atmosphere present.
Proceed as follows
-
Unscrew the housing cover
-
Loosen compression nut of the cable entry gland
-
Remove approx. 10 cm (4 in) of the cable mantle, strip approx.
1 cm (0.4 in) of insulation from the ends of the individual wires -
Insert the cable into the sensor through the cable entry
-
Open the terminals with a screwdriver
-
Insert the wire ends into the open terminals according to the wiring plan
-
Tighten the terminals with a screwdriver
-
Check the hold of the wires in the terminals by lightly pulling on them
-
Tighten the compression nut of the cable entry gland. The seal ring must completely encircle the cable
-
Screw the housing lid back on The electrical connection is finished.
Wiring plan, single chamber housing
- The following illustrations apply to the non-Ex as well as to the Ex-d version.
Housing overview
- Plastic (not with EEx d)
- Aluminium
- Stainless steel (not with EEx d)
- Filter element for pressure compensation (not with EEx d)
Electronics and terminal compartment
- Control lamp
- DIL switch for mode adjustment
- DIL switch for switching point adaptation
- Ground terminal
- Connection terminals
Wiring plan
We recommend connecting OPTISWITCH 5200 C, 5250 C in such a way that the switching circuit is open when there is a level signal, line break or failure (safe state).
Information
The relays are always shown in non-operative condition.
- Relay output
- Relay output
- Voltage supply
Setup
General information
The figures in brackets refer to the following illustrations.
Function/Configuration
With plastic housings, the switching condition of the electronics can be
checked when the housing cover is closed (control lamp). With the basic
setting, products with a density ≥ 0.7 g/cm³ (0.025 lbs/in³) can be detected.
For products with lower density, the switch must be set to ≥ 0.5 g/cm³ (0.018
lbs/in³).
On the electronics module you will find the following display and adjustment
elements:
- Signal lamp (1)
- DIL switch for mode adjustment – A/B (2)
- DIL switch for sensitivity adjustment (3)
Note
Always immerse the tuning fork of OPTISWITCH 5200 C, 5250 C in a liquid to
test its function. Do not test the function of OPTISWITCH 5200 C, 5250 C with
your hand. This can damage the sensor.
Adjustment elements
- Control lamp (LED)
- DIL switch for mode adjustment
- DIL switch for sensitivity adjustment
Signal lamp (1)
Control lamp for indication of the switching status
- green = relay energized
- red = relay deenergized
- red (flashing) = failure
Mode adjustment (2)
With the mode adjustment (A/B) you can change the switching condition of the relay. You can set the required mode according to the “Function chart” (A – max. detection or overflow protection, B – min. detection or dry run protection).
Sensitivity adjustment (3)
With this DIL switch (3) you can set the switching point to liquids having a density between 0.5 and 0.7 g/cm³ (0.018 and 0.025 lbs/ in³). With the basic setting, liquids with a density of ≥ 0.7 g/cm³ (0.025 lbs/in³) can be detected. In liquids with lower density, you must set the switch to ≥ 0.5 g/cm³ (0.018 lbs/in³). The specifications for the position of the switching point relate to water – density value 1 g/cm³ (0.036 lbs/in³). In products with a different density, the switching point will shift in the direction of the housing or tuning fork end depending on the density and type of installation.
Note
Keep in mind that foams with a density ≥ 0.45 g/cm³ (0.016 lbs/in³) are
detected by the sensor. This can lead to erroneous switchings, particulary
when the sensor is used for dry run protection.
Function chart
The following chart provides an overview of the switching conditions depending
on the set mode and the level.
Maintenance and fault rectification
Maintenance
If the instrument is used properly, no special maintenance is required in normal operation.
Rectify faults
Reaction when malfunctions occur
- The operator of the system is responsible for taking suitable measures to rectify faults.
Causes of malfunction
OPTISWITCH 5200 C, 5250 C offers maximum reliability. Nevertheless, faults can occur during operation. These may be caused by the following, e.g.:
- Sensor
- Process
- Voltage supply
- Signal processing
Fault rectification
The first measure to take is to check the output signal. In many cases, the causes can be determined this way and the faults quickly rectified.
Checking the switching signal
Error | Cause | Rectification |
---|
OPTISWITCH 5200 C,
5250 C signals “cov- ered” without being submerged (overfill pro- tection)
OPTISWITCH 5200 C,
5250 C signals “un- covered” when being submerged (dry run pro- tection)
| Operating voltage too low| Check operating voltage
Electronics defective| Press the mode switch. If the instrument then changes
the mode, the
vibrating element may be covered with buildup or mechanically damaged. Should the switching function in the correct mode still be faulty, re- turn the instrument for repair.
| | Press the mode switch. If the instrument then does not change the mode,
the electronics module may be defective. Ex- change the electronics module.
| Unfavourable installation location| Mount the instrument at a location in
the vessel where no dead zones or air bubbles can form.
| Buildup on the vibrating element| Check the vibrating ele- ment and the
sensor for buildup and remove the buildup if there is any.
| Wrong mode selected| Set the correct mode with the mode switch (overflow
protection, dry run protection). Wir-
ing should be carried out according to the closed- circuit principle.
Signal lamp flashes red| Error on the vibrating el- ement| Check if the
vibrating el- ement is damaged or extremely corroded.
Interference on the elec- tronics module| Exchanging the electron- ics module
instrument defective| Exchange the instrument or send it in for repair
Reaction after fault rectification
- Depending on the reason for the fault and the measures taken, the steps described in chapter “Set up” may have to be carried out again.
Exchanging the electronics
- If the electronics module is defective, it can be replaced by the user.
- In Ex applications only an electronics module with respective Ex approval may be used.
- You can find all the information you need to carry out an electronics exchange in the handbook of the new electronics module.
- In general, all electronics modules of series SW60 can be interchanged. If you want to use an electronics module with a different signal output, you carry out the complete setup. You find the necessary, suitable operating instruction on our homepage.
Note
Keep in mind that enamelled instrument versions need special electronics
modules. These electronics modules are called SW60E or SW60E1.
Instrument repair
If a repair is necessary, please proceed as follows:
On our homepage in the Internet under http://www.krohne-mar.com/fileadmin
/media-lounge/PDF-Download/Specimen_e.pdf you can download a return form.
By doing this you help us carry out the repair quickly and without having to
call back for needed information.
- Print and fill out one form per instrument
- Clean the instrument and pack it damage-proof
- Attach the completed form and possibly also a safety data sheet to the instrument
Dismount
Dismounting steps
Warning
- Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel, high temperatures, corrosive or toxic products etc.
- Take note of chapters “Mounting” and “Connecting to power supply” and carry out the listed steps in reverse order.
- With Ex instruments, the housing cover may only be opened if there is no explosive atmosphere present.
Disposal
The instrument consists of materials which can be recycled by specialised
recycling companies. We use recyclable materials and have designed the parts
to be easily separable.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/EG and the
respective national laws. Pass the instrument directly on to a specialised
recycling company and do not use the municipal collecting points. These may be
used only for privately used products according to the WEEE directive.
Correct disposal avoids negative effects on humans and the environment and
ensures recycling of useful raw materials. Materials: see chapter “Technical
data”
If you have no way to dispose of the old instrument properly, please contact
us concerning return and disposal.
Supplement
Technical data
General data
Output variable
Accuracy (according to DIN EN 60770-1)
Measuring accuracy
Deviation ± 1 mm (0.04 in)
Influence of the process temperature on the switching point
- Shifting of the switching point in mm (in)
- Process temperature in °C (°F)
- Switching point at reference conditions (notch)
- Tuning fork
Influence of the product density on the switching point
- Shifting of the switching point in mm (in)
- Product density in g/cm³ (lb/in³)
- Switch position ≥ 0.5 g/cm³ (0.018 lb/in³)
- Switch position ≥ 0.7 g/cm³ (0.025 lb/in³)
- Switching point at reference conditions (notch)
- Tuning fork
Influence of the process pressure to the switching point
- Shifting of the switching point in mm (in)
- Process pressure in bar (psig)
- Switching point at reference conditions (notch)
- Tuning fork
- Repeatability 0.1 mm (0.004 in)
- Hysteresis approx. 2 mm (0.08 in) with vertical installation
- Switching delay approx. 500 ms (on/off)
- Measuring frequency approx. 1200 Hz
Ambient conditions
- Ambient temperature on the housing -40 … +70 °C (-40 … +158 °F)
- Storage and transport temperature -40 … +80 °C (-40 … +176 °F)
Process conditions
-
Process temperature in °C (°F)
-
Ambient temperature in °C (°F)
-
Temperature range with temperature adapter
-
Process pressure in bar (psig)
-
Process temperature in °C (°F)
-
Process pressure in bar (psig)
-
Process temperature in °C (°F)
Electromechanical data
Dimensions
OPTISWITCH 5200 C, 5250 C
- Plastic housing
- Stainless steel housing
- Aluminium housing
OPTISWITCH 5200 C, 5250 C
- Thread
- Clamp
- Cone DN 25
- Slotted nut DN 40
- Flange
- Tuchenhagen Varivent
L = Sensor length, see chapter “Technical data”
OPTISWITCH 5200 C, 5250 C, options
- Gas-tight leadthrough
- Temperature adapter