proceq PS8000 Paper Roll Hardness Tester Instruction Manual
- June 16, 2024
- proceq
Table of Contents
- proceq PS8000 Paper Roll Hardness Tester
- Product Information
- Product Usage Instructions
- Getting Started
- Care and maintenance
- Safety and Liability
- Measure the Roll Hardness Profile
- Technical Specifications
- Maintenance and Support
- Paperlink Software
- Viewing the data
- Further Functions
- 02-02A Morningstar Centre
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
proceq PS8000 Paper Roll Hardness Tester
Product Information
Specifications
- Product Name: Paper Roll Hardness Tester
- Origin: Switzerland
- Website: https://manual-hub.com/
Product Usage Instructions
Getting Started
Note: First time user: Complete the tutorial OR see a demo by a
qualified Proceq representative. If you only need to do simple measurements,
please go to chapter 3.3.
Charging the PaperSchmidt: The PaperSchmidt uses a built-in Lithium
Polymer accumulator. In case of a low battery alert, you can still record
approximately 100 to 200 impacts. When the display indicates a low battery,
connect the PaperSchmidt to the charger or to a PC. Approximately 90 minutes
of charging will bring your PaperSchmidt to its full capacity. The
PaperSchmidt can be recharged at any time.
Power ON: Press the SELECT button to power on.
Check the calibration of the hammer: Perform a calibration check on your
hammer as described in chapter 7.2.
Performing a reset: In case a reset is required, this can be done by
performing an impact with the select button pressed.
Note: This will remove any lower and upper limits that may have been set.
- Should your instrument be fully discharged, start with a reset after recharging.
Power OFF: The PaperSchmidt will shut down automatically after 5 minutes of inactivity.
Standards and Regulations Applied: The following standard has been applied in the PaperSchmidt. TAPPI T 834 om-07 (Determination of containerboard roll hardness). The following recommendation has been taken into consideration in the PaperSchmidt. TAPPI TIP 1004-01 (TAPPI Roll Number for inventory/tracking systems and bar codes). This standard recommends a 12 digit roll number. The PaperSchmidt is able to accept roll identifier names (Roll IDs) in this format.
Care and Maintenance
Battery Care
Avoid this problem by regularly charging your PaperSchmidt during prolonged
periods of inactivity. If the battery is damaged, it will not be possible to
turn the hammer on, and it will have to be returned to a Proceq Service Centre
for replacement.
FAQ
-
Q: How long does it take to fully charge the PaperSchmidt?
A: Approximately 90 minutes of charging will bring your PaperSchmidt to its full capacity. -
Q: Can the PaperSchmidt be recharged at any time?
A: Yes, the PaperSchmidt can be recharged at any time. -
Q: How long does the PaperSchmidt battery last before needing a recharge?
A: The PaperSchmidt can still record approximately 100 to 200 impacts when a low battery alert is displayed.
Getting Started
Note: First time user: Complete the tutorial OR see a demo by a qualified Proceq representative. If you only need to do simple measurements, please go to chapter 3.3.
Charging the PaperSchmidt
The PaperSchmidt uses a built-in Lithium Polymer accumulator. In case of a low
battery alert, you can still record approximately 100 to 200 impacts. When the
display indicates a low battery, connect the PaperSchmidt to the charger or to
a PC. Approximately 90 minutes of charging will bring your PaperSchmidt to its
full capacity. The PaperSchmidt can be recharged at any time.
Power ON
Press the SELECT button to power on.
Check the calibration of the hammer
Perform a calibration check on your hammer as described in chapter 7.2.
Congratulations! Your new PaperSchmidt is fully operational and you can now
continue with your measurements.
Performing a reset
In case a reset is required, this can be done by performing an impact with the
select button pressed.
Note: This will remove any lower and upper limits that may have been set.
- Should your instrument be fully discharged, start with a reset after recharging.
Power OFF
The PaperSchmidt will shut down automatically after 5 minutes of
inactivity.
Standards and Regulations Applied
-
The following standard has been applied in the PaperSchmidt.
TAPPI T 834 om-07 (Determination of containerboard roll hardness) -
The following recommendation has been taken into consideration in the PaperSchmidt.
TAPPI TIP 1004-01 (TAPPI Roll Number for inventory/tracking systems and bar codes)
This standard recommends a 12 digit roll number. The PaperSchmidt is able to accept roll identifier names (Roll IDs) in this format.
Care and maintenance
Battery care
The PaperSchmidt uses a built-in Lithium Polymer accumulator with a capacity
of 150 mAh. When the hammer shuts down, the PCB controller goes into a standby
mode. Thus, very low currents of 10 to 15 µA are constantly flowing. If fully
charged, the battery can last in standby mode for up to 2 years. However, be
aware that deep discharging as well as storage when empty is particularly
damaging the Lithium-Polymer batteries and should be avoided to prevent
permanent damage to the battery pack. Avoid this problem, by regularly
charging your PaperSchmidt during prolonged periods of inactivity. If the
battery is damaged, it will not be possible to turn the hammer on and it will
have to be returned to a Proceq Service Centre for replacement.
Plunger care and lubrication
Correct operation of the hammer can be affected by a sticking plunger.
Possible indications are:
- Low reading on the anvil.
- Intermittent failure to register an impact.
- Plunger does not fully return to starting position before the next impact.
This problem can be solved by applying one or two drops, (no more), of light
oil (e.g. sewing machine oil) on the lateral surface of the plunger. If the
plunger is dirty, wipe it with a soft cloth before applying the oil. After
applying the oil, turn the plunger a few times and depress it a few times to
spread the oil evenly. Wipe off any excess oil before use.
When removing the plunger to clean it, be careful not to lose the plunger
spring that is located within the barrel of the plunger!!
Removing plunger
Attention:
- Be careful that the plunger spring does not get lost.
- Procedure:
- Pull off the plunger from hammer guide bar.
- Remove plunger spring from plunger.
If the plunger spring is lost, the hammer will not register any impacts. If lost, contact a Proceq Service Centre to order a replacement plunger spring (part number 341 10 312).
Safety and Liability
General Information
This manual contains important information on the safety, use and maintenance
of the PaperSchmidt. Read through the manual carefully before the first use of
the instrument. Keep the manual in a safe place for future reference.
Liability
Our “General Terms and Conditions of Sales and Delivery” apply in all cases.
Warranty and liability claims arising from personal injury and damage to
property cannot be upheld if they are due to one or more of the following
causes:
- Failure to use the instrument in accordance with its designated use as described in this manual.
- Incorrect performance check for operation and maintenance of the instrument and its compo-nents.
- Failure to adhere to the sections of the manual dealing with the performance check, operation and maintenance of the instrument and its components.
- Unauthorised modifications to the instrument and its components.
- Serious damage resulting from the effects of foreign bodies, accidents, vandalism and force majeure.
All information contained in this documentation is presented in good faith and believed to be correct. Proceq SA makes no warranties and excludes all liability as to the completeness and/or accuracy of the information.
Safety Instructions
The equipment is not allowed to be operated by children or anyone under the
influence of alcohol, drugs or pharmaceutical preparations. Anyone who is not
familiar with this manual must be super-vised when using the equipment.
- Carry out the stipulated maintenance properly and at the correct time.
- Following completion of the maintenance tasks, perform a functional check.
Correct Usage
- The instrument is only to be used to determine the roll hardness profile of paper and film rolls.
- Replace faulty components only with original replacement parts from Proceq.
- Accessories should only be installed or connected to the instrument if they are expressly autho-rized by Proceq. If other accessories are installed or connected to the instrument then Proceq will accept no liability and the product guarantee is forfeit.
Real Tests
Preparation
If desired, define Roll IDs for the paper rolls under test using Paperlink and
download these onto the PaperSchmidt (See 8.8) By default PaperSchmidt will
ignore invalid impacts, (i.e. impacts that are too low to register a correct
reading). It may be that on rolls with very soft spots, the user may wish to
store such readings also. This is done in PaperLink “Device – Options – Store
invalid readings as zero”.
Define a Roll ID for the Test Series
The data list is the starting point for a new test series. Navigate within the
data list by rolling the PaperSchmidt either away from you or towards you.
There are 3 methods to define a Roll ID for a new test series.
Method 1 – Perform an impact
It is not necessary to define a specific Roll ID. Simply begin a measurement
series by performing an impact directly from the data list. Whatever
upper/lower limits were previously set will be adopted automatically for this
series. In this case the actual 4-digit value of the internal impact counter
is used as a Roll ID and will appear in the review list once the series has
been terminated.
Roll ID = Counter
Method 2 – Manual entry of the Roll ID
Scroll to “NEW” and press the SELECT button. Enter the Roll ID on the PaperSchmidt (see 4.5) Accept or modify the upper/lower limits (see 4.6) and begin measuring the roll hardness profile.
Method 3 – Select a pre-defined Roll ID
Scroll to the desired pre-defined Roll ID in the up-per section of the data
list and press the SELECT button. Accept or modify the upper/lower limits (see
4.6) and begin measuring the roll hardness profile.
Note: A single pre-defined name can be used for a whole batch of rolls.
In this case, each time you move on to the next roll, select the same name
again from the review list and the device will increase an index which is used
to identify that particular roll in the list.
Measure the Roll Hardness Profile
Each factory has its own recommendations for measuring the roll hardness. However, the method is very similar in all cases and some general principles can be applied. Begin at one end of the roll, not too close to the edge. To carry out an impact, place the hammer perpendicular to the paper roll. Push the hammer against the surface of the roll at a moderate speed until the impact is triggered. (For best results, impact at the same angle and apply the same pressure for each reading.) Repeat this procedure at a regular spacing to obtain a profile across the entire roll.
The rebound value is displayed on the screen after each impact along with the number of impacts in the current series. (A series may contain up to 99 impacts.)
“TIP” the plunger to display the running profile. The mini-mum and maximum measured values are displayed on the right. (TIP = Slightly depress the plunger, but not far enough to impact.)
Pressing the SELECT button gives you the option of ter-minating a series (SUM icon) or deleting the last impact if it was an obvious outlier (Dustbin icon). Scroll to the desired icon and press the SELECT button again.
A confirmation screen prevents terminating a series pre-maturely. Use cancel (X-symbol) or confirm (tick-symbol). If an outlier is deleted, the display returns to the previous value and the series counter is reduced by one.
Terminate the Series and review
At the end of the roll, terminate the series by pressing the SELECT button.
Then press on the “Sum” icon as shown above and confirm the action. This will
take you to the statistics screen where you can review all of the significant
data for the test series.
Tilt the PaperSchmidt to show the roll hardness profile. The screen can show a roll hardness profile of up to 50 impacts. However, a single series may contain up to 99 impacts for very large rolls. To see beyond the 50 impacts simply TILT the PaperSchmidt to show the impacts to the right. A marker on the scale at the bottom of the profile shows the “50th” impact position on both screens.
50th impact
50th impact
Exit the summary display mode by pressing the SELECT button if you want to
enter or select a new roll ID. Simply proceed to a next measurement series if
you don’t use specific roll IDs.
Entering a Roll ID manually
Having selected “NEW” from the data list, the user may define a Roll ID for
the test series manually. The Roll ID can be up to 12 characters long. The
screen begins with the last Roll ID that was entered, or with the impact
counter value following a reset.
The cursor position is shown inverted. The screen can show up to 10 characters. Arrows at the bottom indicate if there is more to the left or right. TILT to scroll to the character you wish to change and press the SELECT button to access the character select screen.
The character select screen is arranged in 3 rows to allow the user to locate the desired characters quickly.
- ROLL and TILT to the desired character and press the SELECT button to set it.
- Repeat this procedure to enter the complete Roll ID.
- Scroll to the “ENTER” symbol at the far right of the Roll ID and press the SELECT button.
Deleting a character
- TILT to scroll to the character you wish to delete and press the SELECT button to access the character select screen.
- TILT to the delete symbol and press the SELECT button. (See below)
- The character is deleted. Any characters to the right of the character deleted will shift left.
The character select screen has a special symbol (highlight-ed), that allows the user to delete the selected character from the Roll ID
Accept or modify limits
This function is a rapid way to check if the roll hardness profile is within a
pre-defined tolerance. After selecting a pre-defined Roll ID or after entering
a Roll ID manually, the actual upper and lower limit values are displayed,
e.g. 28R is the lower limit, 56R is the upper limit.
TILT to adjust the value and press SELECT.
(It may also be blank if you were previously working with no limits. To work
without limits, adjust them to 0R resp. 100R.) To accept the limits displayed
press the SELECT button. To modify the limit values, scroll the cursor to the
limit you wish to change and press SELECT.
Note: If the lower limit is higher than the upper limit, they will be
reversed automati-cally
Display warnings
When carrying out a measurement (See 4.3) the display will give a warning if any of the limits are exceeded. In this example, the limit sign next to the R indicates that the lower limit has been exceeded. If the upper limit has been exceeded, the arrow points upwards
In the roll profile view (after TIP function and in the sum-mary view), impacts which exceed the limits will blink.
Review list
A previous series can be reviewed directly from the data list. The lower
portion of the data list shows the last 20 completed measurement series
identified by name.
- This symbol identifies a file which contains data.
Scroll to the series you wish to review and press the select button. You are given the option to look at the summary data or to delete the series. If you choose to delete the series, the confirmation window will be displayed.
If you wish to look at the summary data, you will be taken directly to the
statistics screen. (See 4.4)
Note: Only the last 20 completed measurement series can be reviewed in
the data list. The PaperSchmidt stores many more. The remaining series can
only be seen once the data has been put into Paperlink.
Technical Specifications
Mechanical Data|
---|---
Impact energy| 0.735 Nm
Hammer mass| 115 g
Spring constant| 0.262 N/mm
Spring extension| 75 mm
Dimensions of housing| 55 x 55 x 250 mm (340 mm to tip of plunger)
Dimensions (plunger visible part)| 105 x ø15 mm (radius of spherical tip 25mm)
Weight| 570 g
Memory Data|
No. of impacts in a series| Maximum 99
No. of series names| 50 series names. Each name can be up to 12 characters
long.
Memory capacity| Dependent on length of test series, Example 1: 401 series
with 10 values per series, Example 2: 246 series with 20 values per series
Electrical Data|
Display| 17 x 71 pixel, graphic
Power consumption| ~13 mA measuring, ~4 mA set-up and review, ~0.02 mA idle
Accumulator duty| > 5000 impacts between charges
Charger connection| USB type B (5V, 100 mA)
Accumulator capacity| ~150 mAh
Environmental conditions|
Operating temperature| 0 to 50°C
Storage temperature| -10 to 70°C
IP Classification| IP54
Part Numbers and Accessories
Units
Part No. | Description |
---|---|
342 10 000 | PaperSchmidt consisting of battery charger with USB-cable, CD with |
Paperlink software, carrying strap, documentation, carrying bag
Parts and Accessories
341 10 113 | Cover for USB port |
---|---|
342 10 310 | Plunger complete with plunger spring |
342 10 400 | Testing Anvil PaperSchmidt |
341 80 105 | Carrying bag |
351 90 018 | USB cable 1.8m |
341 80 112 | USB charger, global |
341 80 203 | Carrying strap (loop) |
Please contact your Proceq business unit or authorized service centre for pricing and availability.
Maintenance and Support
Self test function
The PaperSchmidt performs a number of self checks and gives an alert when
there is a problem.
These three alerts show the battery status:
Low battery, Battery full, Battery charging
Low impact energy: The impact spring is out of tolerance and should be replaced. It is recommended to return the PaperSchmidt to an au-thorized service centre.
Checking the calibration on the anvil
The PaperSchmidt checks for correct impact energy with each and every impact.
Nevertheless, it is recommended to check the calibration on a regular basis.
- Perform a series of 10 impacts on the anvil (Proceq Part Nr. 342 10 400)
- Check that the average R-value lies within the tolerance shown on the label.
- If it is out of tolerance, it should be returned to an authorized service centre for recalibration. The EURO anvil p/n 310 09 040 can also be used.
Note: There is a small spring located inside the plunger. Take care not to lose this when removing the plunger for cleaning. Therefore hold the Paper Schmidt with the plunger downward. Pull out the plunger and remove the spring.
Support Concept
Proceq is committed to providing a complete support service for this
instrument by means of our global service and support facilities. It is
recommended that the user register the product on
www.proceq.com to obtain the latest on available
updates and other valuable information.
Standard Warranty and Extended Warranty
The standard warranty covers the electronic portion of the instrument for 24
months and the me-chanical portion of the instrument for 6 months. An extended
warranty for one, two or three years for the electronic portion of the
instrument may be purchased up to 90 days of date of purchase.
Paperlink Software
Starting Paperlink
Locate the file “Paperlink_Setup” on your computer or on the CD and click on it. Follow the instructions you see on the screen.
Application settings
The menu item “File – Application settings” allows the user to select the
language and the date and time format to be used.
Connecting to the PaperSchmidt
Connect the PaperSchmidt to a USB port, then click on the icon to bring up
below window:
Leave the settings as default or if you know the COM port you can enter it
manually.
Click on “Next >”
The USB driver installs a virtual COM port which is used to communicate with the PaperSchmidt. When a PaperSchmidt has been found, you will see another window. Click on the “Finish” button to establish the connection.
Viewing the data
The data stored on your PaperSchmidt will be displayed on the screen:
The test series is identified by the “Impact counter” value and by a “Roll ID” if assigned.
- The user may alter the Roll ID directly in the “Roll ID” column.
- The “Date & Time” when the measurement series was made.
- The “Mean value”.
- The “Total” number of impacts in this series.
- The “Lower limit” and “Upper limit” set for that series.
- The “Range” of the values in this series.
- The “Std dev.” Standard deviation of the measurement series.
Click on the double arrow icon in the impact counter column to see the profile.
If limits have been set, they are displayed as above with a blue band. It is
also possible to adjust the limits directly in this window by clicking on the
blue limit values.
Note: The user may also add a comment to the measurement series. To do
so, click on “Add”.
- The user may alter the order in which the measurements are shown. Click on “measurement order” to switch to “ordered by value”.
Summary window
In addition to the above “Series” view, Paperlink also offers a “Summary”
window. This is useful when comparing a batch of rolls of the same type. Click
on the respective tab to switch between views.
Note: To include or exclude a series from the summary, click on the summary symbol in the impact counter column. This symbol is either “black” or “greyed out”, which shows whether or not the particular series is included in the summary. The summary view can be adjusted in a similar way to the detailed view of a series.”.
Adjusting the max/min settings
Maximum and Minimum settings that were used in the PaperSchmidt at the time of
the measurement series can be adjusted subsequently in Paperlink. This can be
done either by right clicking directly on the item in the appropriate column,
or by clicking on the blue setting item in the detailed view of a measurement
series. In each case a selection box will appear with the choice of setting.
Adjusting the date and time
Right click in the date and time column. The time will be adjusted for the selected series only.
Exporting data
Paperlink allows you to export selected series or the entire project for use
in third party programs. To export selected series, click in the table of the
measurement series you wish to export. It will be highlighted as shown.
- Click on the “Copy as text” icon. The data for this measurement series is copied to the clipboard and can be pasted into another program such as Excel. If you wish to export the individual impact values of the series, you have to display them by clicking on the double arrow icon as described above before you “Copy as text”.
- Click on the “Copy as picture” icon – For exporting the selected items only into another document or report. This performs the same action as above, but the data is exported in the form of a picture and not as text data.
- “Export as text” icon – Allows you to export the entire project data as a text file that can then be imported into another program such as Excel. Click on the “Export as text” icon. This will open the “Save As” window where you can define the location in which you wish to store the *.txt file.
Note: Paperlink has two “tabs” with two display formats. “Series” and
“Summary”. When performing this operation, the project data will be exported
in the format de-fined by the active “Tab”, i.e. either in “series” or
“summary” format.
To open the file in Excel, locate the file and right click on it and “Open
with” – “Microsoft Excel”. The data will be opened in an Excel document for
further processing. Or drag and drop the file into an open Excel window.
Note: Measurement series may be copied between Paperlink project files by
select-ing with CTRL-C and pasting using CTRL-V.
Deleting and restoring data
The menu item “Edit – Delete” allows you to delete one or more selected series
from the downloaded data.
Note: This does not delete data from the PaperSchmidt, only data in the
current project.
The menu item “Edit – Select all”, allows the user to select all series in the
project for exporting etc.
Restoring original downloaded data
Select the menu item: “File – Restore all original data” to restore the
data to the original format as it was downloaded. This is a useful feature if
you have been manipulating the data, but wish to go back to the raw data once
again. A warning will be given to say that the original data is about to be
restored. Confirm to restore.
Note: Any names or comments that have been added to series will be lost.
Deleting data stored on the PaperSchmidt
Select the menu item “Device – Delete all Data on Device” to delete all data
stored on the Paper-Schmidt. A warning will be given to say that the data is
about to be deleted on the device. Confirm to delete.
Note: Please note, this will delete every measurement series and cannot
be undone. It is not possible to delete individual series.
Further Functions
The following menu items are available via the icons at the top of the screen:
- “Upgrade” icon – Allows you to upgrade your firmware via the Internet or from local files.
- “Open project” icon – Allows you to open a previously saved project. It is also possible to drop a *.pqr file onto Paperlink to open it.
- “Save project” icon – Allows you to save the current project. (Note this icon is greyed out if you have opened a previously saved project.
- “Print” icon – Allows you to print out the project. You may select in the printer dialog, if you want to print out all of the data or selected readings only.
Pre-define Roll IDs
Select the menu item “Device – Series names…” to create pre-defined roll names
which can be down-loaded onto the PaperSchmidt before testing a batch of
rolls.
This will bring up the following window.
Simply type in the names for the series and they are transferred automatically
onto the PaperSchmidt. Up to 50 series names can be defined in this way. Each
name can be up to 12 characters long.
Note: Please note, that also “copy and paste” works with data from Excel.
Technical Information Paperlink software
System requirements: Windows XP, Windows Vista, Windows 7, Windows 8,
USB-Connector. An Internet connection is necessary for automatic updates, if
available. An Internet connection is neces-sary for firmware updates (using
PqUpgrade), if available. PDF Reader is required to show the “Help Manual”.
Proceq Europe
Ringstrasse 2
8603 Schwerzenbach
Switzerland
- Phone +41-43-355-3800
- Fax +41-43-355-3812
- Email: info-europe@proceq.com
Proceq UK Ltd.
Bedford i-lab, Priory Business Park
Stannard Way
Bedford MK44 3RZ
United Kingdom
- Phone +44-12-3483-4515
- Email info-uk@proceq.com
Proceq USA, Inc.
117 Corporation Drive
Aliquippa, PA 15001
U.S.A.
- Phone +1-724-512-0330
- Fax +1-724-512-0331
- Email: info-usa@proceq.com
Proceq Asia Pte Ltd
12 New Industrial Road
02-02A Morningstar Centre
Singapore 536202
- Phone +65-6382-3966
- Fax +65-6382-3307
- Email info-asia@proceq.com
Proceq Rus LLC
Ul. Optikov 4
Korp. 2, lit. A, Office 410
197374 St. Petersburg
Russia
- Phone/Fax +7-812-448-3500
- Email info-russia@proceq.com
Proceq Middle East
P. O. Box 8365, SAIF Zone,
Sharjah, United Arab Emirates
- Phone +971-6-557-8505
- Fax +971-6-557-8606
- Email : info-middleeast@proceq.com
Proceq SAO Ltd.
Rua Paes Leme, 136, cj 610
Pinheiros, São Paulo
Brasil Cep. 05424-010
- Phone +55-11-3083-3889
- Email info-southamerica@proceq.com
Proceq China
Unit B, 19th Floor
Five Continent International Mansion, No. 807
Zhao Jia Bang Road
Shanghai 200032
China
- Phone +86-21-6317-7479
- Fax +86-21-6317-5015
- Email: info-china@proceq.com
- Website: www.proceq.com
Subject to change without notice.
Copyright © 2017 by Proceq SA, Schwerzenbach 82034201E ver 11 2017
References
- Screening Eagle Technologies - Protect The Built World - Protect The Built World
- Screening Eagle Technologies - Protect The Built World - Protect The Built World
- Manual-Hub.com – Free PDF manuals!
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>