Micsig CP2100 Series AC DC Current Probe User Manual

June 16, 2024
Micsig

Micsig-CP2100-Series-AC-DC-Current-Probe-logo

Micsig CP2100 Series AC DC Current Probe

Micsig-CP2100-Series-AC-DC-Current-Probe-producyt

Specifications

  • Model: CP2100A, CP2100B
  • Bandwidth: DC ~ 800KHz (CP2100A), DC ~ 2.5MHz (CP2100B)
  • Rise time : 437.5ns (CP2100A), 140ns (CP2100B)
  • Range: 10A/100A
  • Output sensitivity: 0.1V/A (10A), 0.01V/A (100A)
  • DC accuracy (typical): 150ns (10A), 200ns (100A)
  • Signal delay: 50mA ~ 10Apk (10A), 1A ~ 100Apk (100A)
  • Measuring range : 100Apk, 70.7Arms DC+AC pk, 200Apk-pk, 70.7Arms AC
  • Max measurable curren t: CAT III 300V (10A), CAT II 600V  (100A)
  • Max working voltage: CAT III 300V
  • Max float voltage : CAT II 600V
  • Max conductor diameter: 13mm
  • Overload indication: Buzzer beeps and button light flashes
  • Supplied power: DC 5V
  • Net weight: 290g
  • Package weight: 1000g
  • Operating temperature: 5% ~ 95% (0 ~ 40, No condensation)
  • Operating humidity : 5% ~ 65% (40 ~ 50, No condensation)

Operation Instructions

  1. Connect the BNC interface to the oscilloscope (or other instruments), then plug the USB cable to power the probe.

  2. Select the appropriate range according to the current range.
    The corresponding button light will turn Green.

  3. Adjust the oscilloscope settings:

    • Input impedance: 1M
    • Select probe to Current or display as A
    • Set attenuation ratio on the corresponding channel:
    • 100A -> 0.01V/A to 100X
    • 10A -> 0.1V/A to 10X
  4. Press the Zero button to do zero calibration. After a successful calibration, the buzzer will beep one time. If it beeps three times, it means zero calibration has failed. Manual adjustment is also possible. Note: Do not move it in a large range after zero adjustment is completed, as the external magnetic field may have a slight influence on the DC zero position.

  5. Clamp the conductor under test according to the direction indicated by the jaws. Note: If the measured current flows in the opposite direction, the output will be negative.

  6. Adjust the oscilloscope to get the best waveform.

Product Introduction

The CP2100 series current probe is a BNC type that can measure both DC and AC. With a split design and compact look,  it also can be connected to multimeters via a connector. Automatic and manual zero adjustments, powered by USB, no need for additional power supply, making measurement more convenient, often used in motor drives, industrial frequency, inverters, power supplies, avionics, and other fields.

Safety Precautions

  • The measurable circuit should be CAT II 600V or below
  • Do not measure bare conductors
  • Do not touch the measured conductor and sensor head during the measurement
  • Do not use in a humid environment
  • Do not touch the instrument or the measured object with wet hands
  • Please ground this product through the USB power cord
  • Please use this product as required

Specifications

Model CP2100A CP2100B
Bandwidth DC~800KHz DC~2.5MHz
Rise time ≤437.5ns ≤140ns
Range 10A/100A
Output sensitivity 0.1V/A(10A);     0.01V/A(100A)
DC accuracy (typical) 3%±50mA(10A);     4%±50mA(100A,500mA~40Apk)

15%(100A,40Apk~100Apk)

Signal delay| <150ns(10A); <200ns(100A)
Measuring range| 50mA~10Apk(10A);  1A~100Apk(100A)
Max measurable current| 100Apk,70.7Arms(DC+AC pk)

200Apk-pk,70.7Arms(AC)

Max working voltage| CAT III 300V CAT II 600V
Max float voltage| CAT III 300V CAT II 600V
Max conductor diameter| 13mm
Overload indication| Buzzer beeps and Button light flashes
Supplied power| DC 5V
Net weight| 290g
Package weight| 1000g
Operating temperature| 0~40°C
Operating humidity| 5%~95% (0~40℃, No condensation)

5%~65% (40℃~50℃, No condensation)

Operation Instructions

  1. Connect the BNC interface to the oscilloscope (or other instruments), then plug the USB cable to power the probe;
  2. Select the appropriate range according to the current range, the corresponding button light will turn Green;
  3. Adjust the oscilloscope settings: Input impedance 1MΩ; select probe to Current or display as A; Set attenuation  ratio on corresponding channel, 100A(0.01V/A)to 100X, range 10A(0.1V/A)to 10X;
  4. Press the Zero button to do zero calibration, after success, the buzzer will “beep” one time; if “beep” three times, meaning zero calibration has failed; can also go Manual to adjustment. The external magnetic field may have a slight influence on the DC zero position, do not move it in a large range after zero adjustment is completed;
  5. Clamp the conductor under test according to the direction indicated by the jaws. Note: If the measured current flows in the opposite direction, the output will be negative;
  6. Adjust the oscilloscope to get the best waveform;
    Note: When the current exceeds the range, the buzzer will beep for a long time and the button light will flash

References

Micsig-CP2100-Series-AC-DC-Current-Probe-fig-2

Maintenance

During the warranty period of the product (one year) and under normal use, the company will be responsible for free repairs due to faults caused by the quality of the product itself, and the product must not be disassembled or repaired without Micsig’s permission. Please keep the product dry, clean, and tidy. If there is dirt, use a soft cloth or sponge with alcohol to wipe it off. Do not use water. To ensure the performance of the product, it is recommended to check or calibrate once a year.

Statement

The information provided in this document is subject to change in future versions without notice. In addition, to the maximum extent permitted by applicable laws, Micsig does not provide any express or implied warranty for this manual and any information contained in it.

FAQs

  • Q: What is the maximum current vs frequency curve?
    • A: Please refer to F1 in the product manual for the maximum current vs frequency curve.
  • Q: Where can I find the amplitude-frequency characteristic curves?
    • A: The amplitude-frequency characteristic curves for CP2100A and CP2100B can be found in F2 and F3 respectively.
  • Q: Is there a DC signal linearity graph available?
    • A: Yes, the DC signal linearity graph (0.01V/A) is provided in  F4 of the user manual.
  • Q: How can I contact Shenzhen Micsig Technology Co., Ltd.?
    • A: You can reach Shenzhen Micsig Technology Co., Ltd. through the following contact details:

Contact

  • Tel: +86 (0)755 88600880
  • Email: sales@micsig.com
  • Address: A106, Huafeng International Robot Industrial Park, Bao’An District, Shenzhen, 518126, Guangdong, China
  • Website : www.micsig.com

Shenzhen Micsig Technology Co., Ltd.

  • Tel: +86 (0)755 88600880
  • Email: sales@micsig.com
  • Add: A106, Huafeng International Robot Industrial Park,  Bao’An district, Shenzhen, 518126, Guangdong, China
    www.micsig.com