sauermann Tracklog Data Loggers User Manual

June 16, 2024
sauermann

sauermann Tracklog Data Loggers

Product Information

Specifications

  • Device Reference: KT TrackLog, KCC TrackLog, KP TrackLog
  • Display: Yes
  • Internal Sensors: Temperature
  • External Sensors: Temperature, hygrometry, atmospheric pressure, CO2
  • Number of Recording Points: 1 for KT TrackLog, 2 for KP TrackLog, 4 for KCC TrackLog
  • Inputs for Interchangeable Probes: Available for KP TrackLog

Product Usage Instructions

  • Safety Instruction
    Before using the device, please read carefully this user manual. It provides important information about the device operating, maintenance and reprocessing.

  • Precautions for Use
    Please follow the procedure described in this user manual to ensure your safety and avoid any damage to the device. Read the user manual before any use.

  • Warning
    This symbol indicates that the precautions for use must be respected during the device use.

  • Environmental Protection
    Do not throw your electronic device with domestic waste. Please send it back at the end of its duration of use. Once returned, required waste collection will be assured in the respect of the environment in accordance with European guidelines relating to WEEE. Batteries that have reached end-of-life will have to be properly disposed of and only in suitable containers.

  • Symbols Used
    The following symbol will be used in this user manual: Please read carefully the information notes indicated after this symbol.

Device Presentation

Use

  • Communication between the device and PC is carried out with a USB cable with a micro-USB female connector. This communication type allows the devices complete configuration and update.
  • The low-energy wireless connection allows communication with smartphones and tablets with Android and IOS. This communication type allows the devices complete configuration and the data download.

Applications
The TrackLog Data Loggers range is ideal for monitoring different parameters such as temperature, hygrometry, atmospheric pressure, differential pressure, CO2, etc. TrackLog devices ensure traceability in the food industry environment as well as validate the proper functioning of industrial installations. According to the protection index (see details on page 11), the device is adapted for external or internal use.

References

Device Reference| Display| Internal Sensors| External Sensors| Number of Recording Points
---|---|---|---|---
KT TrackLog| Yes| Temperature| N/A| 1
KCC TrackLog| Yes| N/A| Temperature, hygrometry, atmospheric pressure, CO2| 4
KP TrackLog| N/A| Temperature, hygrometry| N/A| 2

FAQ

  • Q: What should I do with the device at the end of its duration of use?
    A: Do not throw your electronic device with domestic waste. Please send it back at the end of its duration of use. Once returned, required waste collection will be assured in the respect of the environment in accordance with European guidelines relating to WEEE.

  • Q: How should I dispose of the batteries that have reached end-of-life?
    A: Batteries that have reached end-of-life will have to be properly disposed of and only in suitable containers.

  • Q: Can I use the device externally or internally?
    A: According to the protection index (see details on page 11), the device is adapted for external or internal use.

USER MANUAL
TRACKLOG DATA LOGGERSsauermann-Tracklog-Data-Loggers-01
\(1\)

SAFETY INSTRUCTION

Before using the device, please read carefully this user manual. It provides important information about the device operating, maintenance and reprocessing.

Precautions for use

  • The devices are developed, produced and sold exclusively to trained and qualified experts in the field of HVACR. Appropriate training is necessary in order to guarantee a riskless use of this instrument. Kimo, a member of Sauermann, is not responsible for any possible accident during its use.
  • Please always use the device in accordance with its intended use and within parameters described in the technical features in order not to compromise the protection ensured by the devices.
  • During the device installation, the whole system safety including the device is on the system assembler responsibility.
  • Only the accessories supplied with the device or available in option must be used.
  • Do not use the device if it is damaged or if it operates abnormally. Inspect the device before every use. In case of doubt, please contact the after-sales service.
  • Any use unspecified by the manufacturer can compromise the device protection.

Warning

  • The device must not be expose to rain or used in moist environments.
  • Do not use the device close to explosive gases, vapour or dust.
  • Do not use this device on systems containing chemical leak sealing. Leak sealing could break into the device and lead to permanent damages.
  • Do not perform contact measurements on non-insulated or undervoltage parts.
  • Do not store the device with solvents. Do not use desiccants.
  • The device does not contain internal part repairable by the user. Do not open the device, except for the batteries replacement. Only have maintenance with Kimo, a member of Sauermann.
  • This device allows to plug different inputs including electrical and mechanical ones. A particular attention must be paid in order to avoid any electric shock. Do not use the device if you are not able to identify electrical failures.
  • This device can pose a risk for wearers of pacemakers. Respect a distance of at least 10 cm between the device and the wearer.

Environmental protection
Send back the device at the end of its duration of use in an electrical and electronic components separate collection centre life (according to local regulations) or send it back for a waste collection assured in the respect of the environment.

Symbols used

  • For your safety and in order to avoid any damage of the device, please follow the procedure described in this user manual and read carefully the notes preceded by the following symbol:

  • The following symbol will also be used in this user manual:

  • Please read carefully the information notes indicated after this symbol.

Logos
On the back of the device, several logos are present:

  • Do not throw your electronic device with domestic waste. Please send it back at the end of its duration of use. Once returned, required waste collection will be assured in the respect of the environment in accordance with European guidelines relating to WEEE. Batteries that have reached end-of-life will have to be properly disposed and only in suitable containers.
  • This marking ensures that the product complies with the requirements of the European related directives.
  • This symbol indicates that the precautions for use must be respected during the device use.
  • Read the user manual before any use.

DEVICE PRESENTATION

Use

  • The TrackLog data loggers allow the measurement of several parameters:
    • KT TrackLog: internal measurement of temperature with two universal inputs for probe
    • KCC TrackLog: internal measurement of temperature, humidity, atmospheric pressure and CO2
    • KP TrackLog: internal measurement of differential pressure
    • KTT TrackLog: model with four thermocouple inputs
  • Communication between device and PC is carried out with a USB cable with a micro-USB female connector. This communication type allows the devices complete configuration and update.
  • The radio LoRa® communication is carried out between the Gateway and the TrackLog. It allows communication with smartphones and tablets running with Android & iOS via the mobile application and with computers via the web application. This communication type allows a partial configuration of devices and the data download.sauermann-Tracklog-Data-Loggers-01 \(7\)
  • The low-energy wireless connection allows to communicate with smartphones and tablets with Android and IOS. This communication type allows the devices complete configuration and the data download.

Applications
The TrackLog Data Loggers range is ideal for different parameters monitoring (temperature, hygrometry, atmospheric pressure, differential pressure, CO2,…). TrackLog devices ensure the traceability in the food industry environment as well as they validate the proper functioning of industrial installations. According to the protection index (see details on page 11), the device is adapted for an external or internal use.sauermann-Tracklog-Data-Loggers-01
\(8\)

References

Device reference| Display| Internal sensors| External sensors| Parameters| Number of recording points
---|---|---|---|---|---
Number| Type| Number| Type
KT TrackLog| Yes| 1| Temperature| 2| Inputs for interchangeable probes*| Temperature, hygrometry| 20,000
KCC TrackLog| 4| Temperature, hygrometry, atmospheric pressure,

CO2

| __

__

| Temperature, hygrometry, atmospheric pressure, CO2
KP TrackLog| 1| Differential pressure| Differential pressure
KTT TrackLog| –| 4| Inputs for thermocouple probes| Temperature

  • Input which allows to plug different compatible SMART PLUG probes: see optional probes page 12.

Device descriptionsauermann-Tracklog-Data-Loggers-01
\(9\)

Display sauermann-Tracklog-Data-Loggers-01 \(11\) sauermann-Tracklog-Data-Loggers-01 \(12\)

  • The selected values to display during the configuration with the application will scroll on the screen every 3 seconds.
  • The display can be activated or deactivated via the KILOG software.
  • At extreme temperatures, the display can become hardly readable and its display speed can slow down at temperatures below 0°C. This has no incidence on the measurement accuracy.

Keys

  • OK key: allows to start or stop the dataset or change of scrolling group. This key also allows to join LoRa® network. See page 14.
  • Selection key: allows to scroll values in the groups scroll. See page 14.

LEDs

Alarm LED
If the red “Alarm” LED has been activated, it has the following states:

  • Always OFF: no setpoint alarms has been exceeded.
  • Flashing quickly (5 seconds): a threshold is currently exceeded on at least one channel.

Operating LED
If the green “ON” LED has been activated, it flashes every 10 seconds during the recording period.

Connections
The communication between the device and the computer is carried out via an USB cable and with the female micro-USB connector. sauermann-Tracklog-Data-
Loggers-01 \(15\)

Dimensions (mm)

Devicessauermann-Tracklog-Data-Loggers-01 \(16\) Wall mount

Devices features

| KT TrackLog| KTT TrackLog
---|---|---
Units displayed (1)| °C, °F, °Ctd, °Ftd, % RH| °C, °F
Resolution| 0.01°C, 0.01°F, 0.01% RH| 0.1°C, 0.1°F
External input| Micro-USB female connector
Input for probe| 2 inputs for interchangeable probes(2)| 4 inputs for thermocouple probes (K, J, T, N, S)
Internal sensor| Temperature| –
Type of sensor| NTC| Thermocouple
| | K: from -200 to +1300°C

Measuring range

| Measuring range of the internal sensor(3): From -40 to +70°C|

  • J: from -100 to +750°C
  • T: from -200 to +400°C
  • N: from -200 to +1300°C

| | S: from 0 to 1760°C
| | K: ±0.4°C from 0 to 1300°C ±(0.3% of the reading +0.4°C) below 0°C
| | J: ±0.4°C from 0 to 750°C ±(0.3% of the reading +0.4°C) below 0°C
Accuracies (4)| ±0.4°C from 0 to 50°C

±0.8°C below 0°C or above 50°C

| T: ±0.4°C from 0 to 400°C ±(0.3% of the reading +0.4°C) below 0°C
| | N: ±0.4°C from 0 to 1300°C ±(0.3% of the reading +0.4°C) below 0°C
| | S: ±0.6°C
Setpoints alarm| 2 setpoint alarms on each channel
Frequency of measurement| From 1 minute to 24 hours
Operating temperature (5)| From -40 to +70°C| From -20 to +70°C
Storage temperature| From -20 to +50°C
Battery life| 3 years(6)
European directives| 2011/65EU RoHS II; 2012/19/EU WEEE; 2014/30/EU EMC; 2014/35/EU

  1. Some units are available only with some optional probes.

  2. Input which allows to plug different compatible probes see optional probes on page 12.

  3.  Other measuring ranges are available according to the connected probe: see optional probes on page 12.

  4. All accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurement carried out in the same conditions, or carried out with
    calibration compensation. The accuracies of the selected probe must be added to to the device accuracies.

  5.  At temperatures below 0°C, the display can become hardly readable and its display speed can slow down. This has no incidence on the measurement accuracy.

  6. Non-contractual value. Based on 1 measurement each 15 minutes at 25 °C. A correct operation of the device and the storage conditions must be respected.

TECHNICAL FEATURES

| KCC TrackLog
---|---
Units displayed| °C, °F, %RH, hPa, ppm
Resolution| 0.01°C, 0.01°F, 1 ppm, 0.01% RH, 1 hPa
External input| Micro-USB female connector
Internal sensor| Hygrometry, temperature, atmospheric pressure, CO2
Type of sensor| Temperature: CTN Hygrometry: capacitive
Atmospheric pressure: piezoresistive
CO2: NDIR
Type of sensor| Temperature: from -40 to 60°C Hygrometry: from 0 to 100% RH
Atmospheric pressure: from 800 to 1100 hPa
CO2: from 0 to 5,000 ppm
Accuracies (1)| Temperature: ±0.2°C
Hygrometry(2): ±1.5% RH (from 10 to 80% RH and from 10 to 50°C(3)) Hysteresis: 0.8% RH at 25°C
Atmospheric pressure: ±3 hPa
CO2: ±50 ppm ±3% of the reading (at 25°C). Temperature dependence: ±1 ppm CO2/°C
Setpoints alarm| 2 setpoint alarms on each channel
Frequency of measurement| From 1 minute to 24 hours
Operating temperature (4)| From -40 to +60°C(5)
Storage temperature| From -20 to +50°C
Battery life| 1 year(6)
European directives| 2011/65EU RoHS II; 2012/19/EU WEEE; 2014/30/EU EMC; 2014/35/EU

  1. All accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurement carried out in the same conditions, or carried out with
    calibration compensation.

  2. Accuracy in HR is linked to the temperature: typical ±2% RH below 10°C and above 50°C. Time drift: < 0.5% RH per year in normal conditions of use (from 5 to 60°C and from 20 to 80% RH, apart from indoor pollutants).

  3. The sensor has a better performance when it is used in normal recommended temperature and hygrometry ranges, i.e. respectively 5°C to 60°C and from 20% RH to 80% RH. A
    prolonged exposure in conditions beyond the normal ranges, especially in conditions of high humidity, can temporally cause an RH measurement drift (offset) (i.e. +3% RH after 60 hours continuous at >80% RH). After returning to normal temperature and humidity ranges, the sensor will by itself progressively come back to its original calibration state. A prolonged
    exposure to extreme conditions can accelerate its ageing.

  4. At temperatures below 0°C, the display can become hardly readable and its display speed can slow down. This has no incidence on the measurement accuracy.

  5. Except for the CO2 parameter: 0 to 60°C.

  6. Non-contractual value. Based on 1 measurement each 15 minutes at 25 °C. A correct operation of the device and the storage conditions must be respected.

| KP TrackLog
---|---
Units displayed| Pa
Measuring range| ±1,000 Pa
Resolution| 1 Pa
Accuracies (1)| ±0.5% of the reading ±3 Pa(2)
Tolerated overpressure| 21,000 Pa
External input| Micro-USB female connector
Inputs for probe| 2 pressure connections
Internal sensor| Differential pressure
Setpoints alarm| 2 setpoint alarms on each channel
Frequency of measurement| From 1 minute to 24 hours
Operating temperature| From 5 to +50°C
Storage temperature| From -20 to 50°C
Battery life| 3 years(3)
European directives| 2011/65EU RoHS II; 2012/19/EU WEEE; 2014/30/EU EMC; 2014/35/EU

  1. Provided that an autozero at 0 Pa has been carried out (the device must be disconnected) at the temperature at which the measurement will be performed (±5°C).

  2. All accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurement carried out in the same conditions, or carried out with
    calibration compensation.

  3. Non-contractual value. Based on 1 measurement each 15 minutes at 25 °C. A correct operation of the device and the storage conditions must be respected.

Housing features

LoRa ® **frequency** 868 Mhz
Dimensions 110.2 x 79 x 35.4 mm
weight KT TrackLog, KCC TrackLog, KP TrackLog: 206 g KTT TrackLog: 200

g.
Display| 2 lines LCD screen.
Screen size: 49.5 x 45 mm
2 indication LEDS (red and green)
Control| 1 OK key 1 Selection key
Material| Compatible with food industry environment ABS housing
__

Protection

|

  • IP 65: KT TrackLog IP 65: KP TrackLog(1)
  • IP 54: KTT TrackLog(2) IP 40: KCC TrackLog

PC communication| Micro-USB female connector USB cable
Battery power supply| 2 lithium AA 3.6 V batteries – 2600 mAh(3)
Environmental conditions of use| Air and neutral gases
Hygrometry: in non-condensing conditions (<80% RH) Maximum altitude: 2000 m

  1. With the pressure connectors plugged.
  2. With all the thermocouple probes plugged.
  3. When replaced, please use the same high performance batteries type SAFT LS 14500 (Li-SOCl2) 3.6 V – 2600 mAh.

Features of optional probes
All probes for KT TrackLog have an automatic recognition of of probes and the adjustment parameters storage make them 100% interchangeable.

Part number Description Measuring range

External or ambient thermo-hygrometric probes
KITHA-2| Interchangeable hygrometry and ambient temperature probe|

Hygrometry: from 0 to 100% RH Temperature: from -20 to +70°C

KITHP-130-2| Remote interchangeable hygrometry and temperature probe
KITHI-150| Remote interchangeable hygrometry and temperature probe| Hygrometry: from 0 to 100% RH Temperature: from -40 to +180°C
General use or insertion Pt 100 temperature probes
KIRGA-50 / KIRGA-150| IP65 immersion probe (50 or 150 mm)|

From -40 to +120°C

KIRAM-150| Ambient probe 150 mm
KIRPA-150| Penetration probe IP65| From -50 to +250°C
KIPI3-150-E| IP68 penetration probe with handle
KITI3-100-E| IP68 penetration probe with T-handle
KITBI3-100-E| IP68 penetration probe with corkscrew handle
KIRV-320| Velcro probe| From -20 to +90°C
KICA-320| Smart adapter for Pt100 probe| From -200 to +600°C according to the probe
Thermocouple probes
All the thermocouple temperature probes for the KTT TrackLog have a class 1 sensitive element as per IEC 584-1, 2 and 3 standards.
For more details about the available thermocouple probes, please see the “Thermocouple probes” datasheet.

For more details, please see the “Measuring probes for KT TrackLog” and “Thermocouple probes” datasheets.

To connect a probe:

  • Open the mini-DIN connection cap on the bottom of the data logger.
  • Connect the probe in such a way the mark on the probe is in front of the user.

Connection to LoRa® network

  • TrackLog device is synchronised ex-factory for an automatic connection to LoRa® network via the Gateway. Connect the Gateway to the Ethernet network (see the procedure described below), and the device connects automatically to the Gateway.
  • However, if you encounter any difficulties to connect your TrackLog device to LoRa® network, please refer to the procedure described below.

Gateway

  1. Connect the gateway to the mains and connect the Ethernet jack.

  2. The LED flashes when it is connected to the electricity grid.

  3. The fixed LED is then connected to the LoRa® network.
    This operation can take a few minutes. Please, wait…

  4. Log in to https://tracklog.inair.cloud to configure your TrackLog and launch a dataset.

Device

  • When starting a measurement dataset, or during the batteries replacement:
    • The “Cloud” icon flashes on the TrackLog display and “Sync” is displayed on screen. Then “Sync” disappears, the cloud becomes permanent and OK is displayed when TrackLog is successfully connected to LoRa® network.

OK disappears after 5 seconds.

  • If the “Cloud” icon keeps flashing and “Sync Err” is displayed, it is necessary to attempt to connect again to LoRa® network.
  • Press the TrackLog OK key.
  • If the connection to LoRa® network does not work, please check that the 1700 port is open in UDP on the Ethernet network on which the Gateway is plugged.

Function of keys

OK key: allows to start, stop the dataset or change of scrolling group like described in the following tables. It also allows to reach LoRa® network.
Selection key: allows the scroll values in the scrolling group like described in the following tables. sauermann-Tracklog-Data-Loggers-01
\(20\) sauermann-Tracklog-Data-Loggers-01
\(21\)

sauermann-Tracklog-Data-Loggers-01 \(22\) * Please see the summary table of the groups organisation on the following page.

Groups organisation
The table below summarises the groups organisation and measured values available during a measurement dataset:

| Group 1| Group 2| Group 3
---|---|---|---
Measured temperature| Max. value in temperature Min. value in temperature| High alarm threshold in temperature Low alarm threshold in temperature
Measured hygrometry
| Max. value in hygrometry Min. value in hygrometry| High alarm threshold in hygrometry Low alarm threshold in hygrometry
Measured CO2| Max. value in CO2 Min. value in CO2| High alarm threshold in CO2 Low alarm threshold in CO2
Measured differential pressure
| Max. value in differential pressure Min. value in differential pressure| High alarm threshold in differential pressure Low alarm threshold in differential pressure
Measured atmospheric pressure| Max. value in atmospheric pressure Min. value in atmospheric pressure| High alarm threshold in atmospheric pressure Low alarm threshold in atmospheric pressure
Parameter 1 of probe 1
| Max. value in Parameter 1 of probe 1 Min. value in Parameter 1 of probe 1| High alarm threshold in Parameter 1 of probe 1 Low alarm threshold in Parameter 1 of probe 1

Parameter 2 of probe 1*

| Max. value in Parameter 2 of probe 1 Min. value in Parameter 2 of probe 1| High alarm threshold in Parameter 2 of probe 1 Low alarm threshold in Parameter 2 of probe 1
Parameter 1 of probe 2| Max. value in Parameter 1 of probe 2 Min. value in Parameter 1 of probe 2| High alarm threshold in Parameter 1 of probe 2 Low alarm threshold in Parameter 1 of probe 2
Parameter 2 of probe 2
| Max. value in Parameter 2 of probe 2 Min. value in Parameter 2 of probe 2| High alarm threshold in Parameter 2 of probe 2 Low alarm threshold in Parameter 2 of probe 2

Measurements scroll
According to the selected parameters during the configuration and according to the type of device, the measurement scroll is carried out like following :sauermann-Tracklog-Data-Loggers-01 \(24\) * Parameters available according to the device and probe type.

Examples:

  • KT TrackLog with a thermo-hygrometric probe (channel 1) and a temperature probe (channel 2):sauermann-Tracklog-Data-Loggers-01 \(25\)
  • KCC TrackLog:sauermann-Tracklog-Data-Loggers-01 \(26\)

The measurements scroll can be carried out by pressing the “Select” button of the data logger or wait about 3 seconds and the display scrolls automatically.

PC communication
Connecting the device to the PC allows to carry out a complete configuration of the Data Logger:

  1. Plug the male USB connector of the cable to an USB connection on your computer*.
  2. Open the USB cap on the right side of the Data Logger.
  3. Connect the male micro-USB connector of the cable to the female micro-USB connector of the device.sauermann-Tracklog-Data-Loggers-01 \(27\)

Data Logger configuration, data download and processing with the web and mobile applications
Please refer to the web & mobile applications user manual: sauermanngroup.com

  • The date and time updates automatically when a new configuration is loaded with the mobile Application.
    • The computer must be in compliance with the IEC60950 standard.

WIRELESS CONNECTION FUNCTION

  • TrackLog devices have the wireless connection function allowing to communicate with a smartphone or a tablet (Androïd or iOS) via the application.
  • The device is named “TrackLog” in the list of available devices of the tablet or smartphone.
  • The wireless connection function, always active, allows to carry out a complete configuration of the Data Logger and download the data recorded in the device.

DEVICE UPDATE

To carry out the device update:

  • Plug the device in USB on the computer (see page 17).
  • The device is displayed on the computer as a removable storage.
  • Copy the supplied update file.
  • Paste the update file on the device, on the storage root.
  • Please wait 10 to 15 seconds without unplugging the device until the percentage is displayed.
  • The device can be unplugged once the update starts.
  • Please wait until the end of the update (100% is displayed) before using again the device.

MAINTENANCE

Replace the batteries
With a 1 to 3 years battery life*, the TrackLog device guarantees long-term measurement.

  • When replaced, please use the same high performance batteries SAFT LS 14500 (Li-SOCl2) 3.6 V – 2600 mA type.
  • Warning: in case of polarity non-compliance, an overheat can occur on the device, which will not damage the device, but can cause a risk of burning.
  • Do not throw the batteries with domestic waste. Dispose of used batteries at the recycling facility set up for this purpose or seek an approved body.

Replace the batteries on KT, KP and KTT TrackLog
To replace batteries:

  • Unscrew the unlosable screw on the battery hatch on the backside of the TrackLog with a cross-head screwdriver.
  • The battery hatch opens. Remove the old batteries.
  • Insert the new batteries** respecting the polarity.
  • Replace the battery hatch.
  • Screw it.sauermann-Tracklog-Data-Loggers-01 \(28\)

Replace the batteries on KCC TrackLog

  • Unscrew the unlosable screw on the battery hatch on the backside of the TrackLog with a cross-head screwdriver.

  • The battery hatch opens. Remove the old batteries.

  • Press simultaneously OK and Select keys of the device until the 2 LEDs light on.

  • The LEDs light off after a few seconds.

  • Wait 1 minute before replacing the batteries.

  • Insert the new batteries** respecting the polarity.

  • “Test Bat” is displayed.

  • When “Test bat” disappears, the device is operational again and will continue from the point of the disruption.

  • Replace the battery hatch.

  • Screw it.

  • Non-contractual value. Based on 1 measurement each 15 minutes at 25 °C. A correct operation of the device and the storage conditions must be respected. ** The device battery life is calculated, therefore we recommend to use new batteries.

Device cleaning

  • Please avoid any aggressive solvent.
  • Please protect the device and probes from any cleaning produce containing formalin, that may be used for cleaning rooms and ducts.
  • For hygrometry probes:
    • It is strongly discouraged to remove the protection cap from hygrometry probes, since the sensor inside is very fragile. Any contact can damage it. However, if the cap protection removal is required, please take the maximum precautions and do not touch the sensor.
    • To remove the protection cap, unscrew it.
    • During a prolonged use of the hygrometry probe, it is recommend to protect the remote probe with the CPH protection cap, available in option.

SAFETY LOCK WALL MOUNT WITH PADLOCK

Mount the safety lock support on the required place.

  1. Present the Data Logger on the support starting with the inferior part.
  2. Clip the Data Logger on the support by falling back the superior part.
  3. Insert the padlock to ensure the safety lock function.sauermann-Tracklog-Data-Loggers-01 \(29\)
  • The padlock can be replaced by a fail-safe sealed
  • The data logger can be placed on the screw-mount without the safety lock function
  • To remove the data logger from the support, proceed on reverse order.

CALIBRATION

  • A calibration certificate is available as option under paper format.
  • We recommend to carry out a yearly checking.

KP TRACKLOG: PERFORM AN AUTO-ZERO

  • The device must be placed in a convenient location to perform the auto-zero. It is possible to reset the device during a recording dataset:
    • Unplug the pressure tubes of the device.
    • Press the “Selection” key during 5 seconds to carry out the auto-zero.
  • The instrument resets. The screen displays “…”
  • Plug the pressure tubes.
  • The device continues the measurements and the dataset recording.
  • It is possible to reset the device when values are measured but not recorded:
    • Unplug the pressure tubes of the device.
    •  Press the “Selection” key to display the measurement.
    • “Selection” key during 5 seconds to carry out the auto-zero. The instrument resets. The screen displays “…”
  • Plug the pressure tubes.
  • The device continues the measurements.

ACCESSORIES

Accessories Part numbers Illustrations
1 AA lithium 3.6 V battery KBL-AA
Safety lock wall mount with padlock KAV-320

Gateway| TrackLog Gateway|
Cloud subscription offer| See subscription sheet| –
Calibration certificate| Option| –
Thermocouple temperature **probes*| See| –
Temperature and hygrometry ambient probes**| See specific data sheet| –
Protection caps for hygrometry probes| See specific data sheet| –
Wired extensions for probes
In polyurethane, 5 m length with male and female mini-DIN connectors
Note: several extensions can be wired in order to obtain up to 25 m cable length| KRB-320|
USB micro-USB cable which allows to plug your device to your PC| CK-50|

Only the accessories supplied with the device must be used.

  • for KTT TrackLog only. ** for KT TrackLog only.

TROUBLESHOOTING

Problem Probable cause and possible solution
No value is displayed, only the icons are present. The display is configured

on “OFF”. Configure it on “ON” with the web or mobile application.
The display is completely off* and there is no communication with the computer.| The battery has to be replaced. See page 19.
The display indicates —— instead of the measured value.| The probe is disconnected. Plug it again to the data logger.
There is no LoRa® communication with the Data Logger.| Check that the gateway is connected on the Ethernet network and the 1700 port in UDP is open, then press OK key to attempt to connect again to the LoRa® network.

  • On KT and KTT TrackLog only.

BE CAREFUL! Material damages can happen, so please apply the precautionary measures indicated.
sauermanngroup.com

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

sauermann User Manuals

Related Manuals