MACHEREY-NAGEL Formaldehyd 10 Tube Test
Overview
The test is suitable for the photometric determination of Formaldehyde.
The test is suitable for surface water, groundwater, drinking water, and
engineered wood.
-
Measuring range:
0.20–10.0 mg/L HCHO (method 0461) 0.20–10.0 ppm HCHO (method 0462 / 0463) 50-mm semi micro cuvette: 0.02–1.00 mg/L HCHO (method 1461) -
Number of tests: 20
-
Wavelength for photometric determination: 412 / 436 nm
-
Shelf life: 24 months
-
Reaction time: 10 minutes
-
Storage temperature: 15–25 °C
-
Storage conditions: upright
Method
Formaldehyde reacts in aqueous solution with ammonium and acetylacetone to form a yellow dye. Analogous to DIN EN ISO 12460-5 for the determination of formaldehyde in chipboard.
Interferences
The foreign materials shown here do not interfere with the test up to the
indicated concentrations (in mg/L). The cumulative effect of different
interfering ions has not been tested.
Data in mg/L:
- Pb2+, Fe2+, Fe3+, NO3 -, S2-: 10
- Zn2+: 500
- Ni2+: 200
- Acetaldehyde: 100
- Cu2+: 20
- Cr(VI): 5
- Cr3+: 2
- Fe3+: 1
The method can be applied for analyzing seawater.
Turbidities cause higher measurement values.
Reagents and accessories
Contents of reagents set:
- 20 test tubes R0
- 2 reagent R1
Required devices:
- MACHEREY-NAGEL photometer
- MACHEREY-NAGEL heating block
- Digital piston pipette 1–5 mL (REF 916909) with pipette tips (REF 916916)
- Digital piston pipette 200–1000 µL (REF 91671) with pipette tips (REF 91667)
- Additionally needed: Special filter 412 nm
Sampling and preparation
See DIN EN ISO 5667-3-A 21.
Adjust to pH 3–10 prior to analysis.
Quality control
The measurement of a blank value and a standard is recommended before every measuring series as quality control measure.
Quality data:
The following data were determined during production according to ISO 8466-1
and DIN 38402-A51:
- Number of LOTs: 8
- Standard deviation of the method: ± 0.04 mg/L HCHO
- Coefficient of variation of the process: ± 0.98 %
- Confidence interval: ± 0.1 mg/L HCHO
- Specified data for procedure:
- Sensitivity (absorbance of 0.010 A corresponds to): ± 0.03 mg/L HCHO
- Accuracy of a measurement value: ± 0.3 mg/L HCHO
- LOT-specific certificates are available at www.mn-net.com
Procedure
- Open test tube
- Pipette 2 mL of sample into test tube
- Add 1 mL R2
- Seal test tube and shake vigorously
- Heat for 10 min at 60°C
- Take the tube from the heating block
- Cool to room temperature
- Wait 60 min
- Swirl
- Clean outside of test tube
- Measure
Implementation in a 50-mm semi-micro cuvette
Measurement against zero value (distilled water instead of sample) necessary
- Open test tube
- Wait 5 min
- Add 1 mL R2
- Seal test tube and shake vigorously
- Heat for 10 min at 60°C
- Take the tube from the heating block
- Cool to room temperature
- Wait 60 min
- Swirl
- Transfer the contents of the test tube into a 50-mm semi-micro cuvette
- Clean outside of test tube
- Measure [method 1461]
Notes
When using other photometers, make sure measurements are possible in test
tubes (16 mm OD) and calibrate the method.
Use the correction value when measuring cloudy or colored samples (see
photometer handbook).
For the determination of formaldehyde in chipboard, please request special
instructions.
To increase the accuracy, it is recommended to perform the measurement of a
reagent blank value.
Smaller concentrations can be determined by using semi-micro cuvettes 50 mm
(REF 91950). 10/2020
Customer Support
www.mn‑net.com
MACHEREY-NAGEL GmbH & Co. KG
Neumann-Neander-Str. 6–8
US Tel.: +1 484 821 0984sales-us@mn-net.com
Management
System
EN ISO 13485:2016
ISO 9001 :2015
www.tuv.com
ID 9000008338