Drager Alcotest 7510 Professional Digital Breathalyser User Manual

June 15, 2024
Dr ger

Drager Alcotest 7510 Professional Digital Breathalyser

Product Information

Specifications

  • Product Name: Breath Alcohol Monitor
  • Model: Alcotest 7510
  • Intended Use: Fast determination of alcoholisation of people through the measurement of the breath alcohol concentration
  • Options: Can be factory-equipped with standard or classic mouthpieces, wrist strap, power supply unit, and infrared interface

Product Usage Instructions

Preparing the device for operation
To prepare the device for operation:

  1. Install the wrist strap:
  2. Guide the mounting strap (thin section) of the wrist strap through the housing drills on the underside of the measuring device and form a loop.
  3. Pull the thicker section of the wrist strap through the formed eye.

Display, Symbols, and Keys
The device’s display shows various symbols according to its status:

  • PC: Indicates a fault or warning
  • Database is 90% full: Indicates that the device’s database is almost full
  • Database is 100% full: Indicates that the device’s database is full and no more data can be stored. New data will overwrite old data.
  • GPS system active: Only applicable to devices with GPS system
  • Battery or charge state: Indicates the battery or charge state of the rechargeable battery, ranging from full to empty

FAQ (Frequently Asked Questions)

Q: What is the intended use of the Breath Alcohol Monitor?
A: The Breath Alcohol Monitor is designed for fast determination of alcoholisation of people through the measurement of the breath alcohol concentration.

Q: What options can the Alcotest 7510 be equipped with?
A: The Alcotest 7510 can be factory-equipped with standard or classic mouthpieces, wrist strap, power supply unit, and infrared interface.

Q: How do I prepare the device for operation?
A: To prepare the device for operation, follow these steps:

  1. Install the wrist strap by guiding the mounting strap through the housing drills on the underside of the measuring device and forming a loop. Then, pull the thicker section of the wrist strap through the formed eye.

For your Safety

Strictly follow the instructions for use Any use of the device requires full understanding and strict observation of these instructions. The device is to be used only for the purpose specified here. Maintenance The device must be inspected and serviced regularly by trained service personnel. Repairs to the device may only be carried out by trained service personnel. We recommend that a service contract be obtained with Dräger Safety and that all repairs also be carried out by Dräger Safety. Only authentic Dräger parts may be used for maintenance. Observe chapter “Maintenance intervals” on page 57. Accessories Use only accessories shown in the Order List, see page 67 and page 68. Safe coupling with electrical devices Electrical connections to devices which are not listed in these Instructions for Use should only be made following consultation with the respective manufacturers or an expert. Not for use in explosion hazard areas The device is not approved for use in explosion hazard areas.
Safety symbols used in these Instructions for Use
These Instructions for Use use a number of warnings for risks and hazards which might occur when using the device. These warnings contain “signal words” which will alert you to the degree of hazard you may encounter. These signal words and corresponding hazards are as follows:
CAUTION Bodily injuries or damage to property may occur in a situation of potential danger unless appropriate precautions have been taken. Can also be used to warn against any wanton actions.
NOTICE Additional information for the use of the device.

Intended use

Dräger Alcotest® 7510 ­ Breath alcohol measuring device for the
fast determination of alcoholisation of people through the measurement of the breath alcohol concentration.
­ For portable, battery-powered operation.
The Alcotest 7510 can be factoryequipped with the following options:
­ Operation with lithium-ion batteries Features: high energy density, which enables a large number of measurements over the entire established temperature range.
­ with GPS (Global Positioning System) for the satellite-based determination of the device location.
­ with absolute pressure sensor for the compensation of atmospheric pressure during dry gas calibration.
­ Mouthpiece receptacle at right or left on the “Standard” mouthpieces, with or without non return valve or mouthpiece receptacle at top on the”Classic” mouthpieces, with or without non return valve.

Intended use
11:53:47 04.12.07
Alcotest 7510
M
OK
D
Alcotest 7510

00223996_01.eps

————————® Alcotest is a registered trademark of Dräger.
39

What is What?
What is What?
Scope of Delivery

1a

1b
2a 2b

1a Measuring device: Dräger Alcotest 7510 Mouthpiece receptacle at right or left
2a “Standard” mouthpieces with non return valve, or without non return valve, 3 units
or
1b Measuring device: Dräger Alcotest 7510 Mouthpiece receptacle at top
2b “Classic” mouthpieces with non return valve, or without non return valve, 3 units

40

M
OK
D
Alcotest 7510

M
OK
D
Alcotest 7510

3 4

3 Wrist strap
4 Optional: Power Supply Unit 600 mA for device with NiMH or lithium-ion rechargeable batteries
5 Inserted in the case cover: ­ Instructions for Use ­ Quick Guide ­ System CD

5

00323996_01.eps

D D

DDrrääggeer ArlcAotlecsot®t7e5s10t ® 7510

Atemalkohol-Messgerät de

Kurzanleitung Der Betrieb des Dräger Alcotest 7510 setzt die genaue Kenntnis und Beachtung der

de

Gebrauchsanweisung Dräger Alcotest 7510 voraus. Diese Kurzanleitung ersetzt nicht die GGeebrbaruachuscahnswaeinsuwnegidseusnDgrä, gSeer AitlecoXtesbti7s5X10!

Breath Alcohol Monitor en en

Quick User Guide Any use of the Dräger Alcotest 7510 requires full understanding and strict observation of the tIiInonsnstsrtufrocurticoUtnsisoenfoosfrtUhfoseerDDUräräsggeeer,rApAlcalcogoteteesstXt775t51o100X! . This Quick User Guide does not replace the Instruc-

00123996_01.eps 00233040_01.eps

What is What?

Measuring device

1

1 Mouthpiece receptacle

available at right or left 2 Display 3 Mouthpiece ejector

11

11:53:47

04.12.07

2

at right or left dependent on the mouthpiece receptacle

Alcotest 7510

3

4 – key (UP)

10

5 OK – key (ON/OFF, Confirm)

6 Opening to the audible alarm

9

M

4

7 Charging contacts and charging con-

OK

5

nection (jack plug)

8

8 USB connection

9 – key (DOWN/MENU) M
10 Infrared interface

D
Alcotest 7510

11 LED red, yellow, and green

6

Optional: Measuring device with 12 Mouthpiece receptacle
at top

11:53:47 04.12.07

7 12

Alcotest 7510
7
6
M
OK

00423996_01.eps

02423996_01.eps

41

Preparing the device for operation
Preparing the device for operation
Dräger Alcotest 7510 is delivered ready for operation and factory-calibrated. Has the calibration date expired. Display » Service date expired «: Calibrate the device or have it calibrated
by Dräger Safety or another authorised organization. The date of the last calibration can be viewed in the menu (see page 63).
Installing the wrist strap
Guide the mounting strap (thin section) of the wrist strap through the housing drills on the underside of the measuring device and form a loop.
Pull the thicker section of the wrist strap through the formed eye.
42

00523996_01.eps

Preparing the device for operation

Display, Symbols and Keys

The breath alcohol measurement value is specified in units of mg/L in the basic setting of the device. The device can also be converted to all established measurement units, see Technical Handbook Dräger Alcotest 7510.

02523996_01.eps

Symbols According to the device status, the following symbols are displayed in the top row, next to the time and date information (one below the other on the right side):

PC

11:53:47 04.12.2007

There is a fault There is a warning

PC

The device is communicating with a PC

Database is 90 % full Database is 100 % full, no more data can be stored Database is 100 % full and will be overwritten with new data

GPS system active (only on devices with GPS system)

Battery or charge state of rechargeable battery: full Battery or charge state of rechargeable battery: 2/3 full Battery or charge state of rechargeable battery: 1/3 full Battery or charge state of rechargeable battery: empty

Keys
OK
M

“OK” key ­ Switch on/off key for the device. ­ Confirms the entry of the selected menu function(s). ­ Triggers passive or manual sampling.
“DOWN / MENU” ­ Sets values and selects menu functions. ­ Access to the menus.
“UP” ­ Sets values and selects menu functions.

43

00723996_01.eps

Preparing the device for operation

Switch on and off
Switch on Press and hold the OK key for approx. 1 second until the start
screen appears. ­ A self-test is completed. ­ Device information is displayed.
If the next service date has expired: Have the device calibrated (see “Fault, Cause, Remedy” on
page 54).

11:53:47 04.12.2007
Alcotest 7510

Switch off
Hold the OK key pressed. ­ The device shuts down automatically after approx. 2 seconds.

11:53:47 04.12.07

Alcotest 7510 shut-down in 2 sec.

Automatic shut down
In the default setting, the device shuts down automatically 4 minutes after the last operation.
The device also shuts down automatically when the batteries or rechargeable batteries are fully discharged.

11:53:47 04.12.07
Alcotest 7510 auto shut-down

00823996_01_en.eps

00923996_01_en.eps

44

Attaching the mouthpiece
­ According to the version of the device, the mouthpiece can be attached with right or left orientation (Standard) or with top orientation (Classic).
Remove the mouthpiece from the packaging. For hygienic reasons, leave the mouth contact area protected by the packaging, until the mouthpiece sits firmly on the device.
Depending on the version of the device, see page 41: With the chequered mark (= pressure sur-
face) facing upwards, put the “Standard” mouthpiece into the mouthpiece receptacle at an angle and push it into the mouthpiece receptacle. The mouthpiece audibly engages. ­ The inserted mouthpiece can be illuminated by the switched on device.
or From above, push the “Classic” mouth-
piece firmly into the mouthpiece receptacle.
Then completely remove the packaging.

Preparing the device for operation

1

2

00623996_01.eps

02623996_01.eps

45

Performing the measurement
Performing the measurement
Conditions
­ Insert a new mouthpiece for each test subject. ­ Do not allow any high alcohol concentrations on the sensor, e.g., by rinsing the mouth with
a high proof alcohol shortly before the measurement. Doing this reduces the service life of the sensor. ­ Maintain an adequate distance from mobile phone antennas and transmitting stations.
Requirements of the test subject
­ Observe a waiting period of at least 15 minutes after the last oral alcohol intake! Residual alcohol in the mouth can distort the measurement. Distortions can also occur with aromatic beverages (e.g. fruit juice), alcoholic mouth sprays, medicines and drops, and after burping and vomiting. Also observe a waiting period of at least 15 minutes in these cases. A mouth rinse with water or non-alcoholic beverages does not replace the waiting period!
­ The device can detect residual mouth alcohol during the delivery of the breath sample, if this option is activated. If residual mouth alcohol is detected, the measurement is aborted and a relevant message is displayed.
­ The test subject should breathe normally and steadily prior to the sampling.Rapid inhalation and exhalation through the mouth should be avoided.
­ The test subject must be able to produce the required minimum breath volume (default setting 1.2 L). During testing, the breath flow must fall within a preset minimum blowing period (default setting 2 seconds).
46

Performing the measurement

Data input before or after the measurement
Depending on the device configuration, it may be necessary to enter the data before or after the measurement. The following examples illustrate the various data input possibilities. The data input can be skipped with the provided selection item “Skip” and its selection by
pressing the OK key.

Data input using the selection list Example: Input of the sex of the person to be tested.

Select the appropriate setting with the Press OK key.

and key. M

­ The selected setting is indicated with . The cursor jumps to the

bottom row “Confirm”.

Press OK key, the selected setting is applied,

or

move the cursor back to the setting with the key in order to

modify it.

Gender Skip
Male Female

11:53:47 04.12.07

Confirm

Data input of free text with the and – keys M
Example: Input of the age of the person to be tested.
Use the and to enter the first digit. M
Press OK key, the cursor jumps to the second position of the digit to be entered.
Enter the second digit with the and key. M
Press OK key, cursor jumps to the bottom row “Confirm”. Press OK key, the setting is applied,
or move the cursor back to the setting with the key in order to
modify it.

Age Skip
2 _

11:53:47 04.12.07

Confirm

02823996_01_en.eps

02923996_01_en.eps

47

03023996_01_en.eps

Performing the measurement

Menu-driven data input In the menu-driven data input, numbers and letters are combined in the first level, into horizontally arranged groups.

Example: Input of the officer ID.
­ The cursor stands at left of the empty input field. With the and key, select the group that contains the de-
M
sired number or letter. Press OK key. ­ The selected group fans out in the vertical level.

With the group.

and key, select the number or letter from the M

Press OK key.

­ The first input position is set and displayed at the left of the input

field next to the cursor.

Enter additional positions in the same way.

11:53:47 04.12.07
Officer ID I Move cursor left Move cursor right Delete Characters 0 abc ABC 1 def DEF 2 ghi GHI 3 jkl JKL
11:53:47 04.12.2007
Officer ID I Move cursor left Move cursor right Delete Characters 0 a b c

When the input is complete: With the key point the cursor on the bottom row “Confirm”.
M
Press OK key, the setting is applied.
The position of the cursor can be changed during input: With the and key position the rows “Move cursor left” or
M
“Move cursor right”. Press OK key, the cursor position is moved one position to the left
or right.
Incorrect input can be deleted: With the and key point the cursor to the row “Delete
M
characters”. Press OK key, the character at left next to the cursor is deleted.

11:53:47 04.12.07
Officer ID 08351| 4 mno MNO 5 pqrs PQRS 6 tuv TUV 7 wxyz WXYZ 8 #+9 ,.
Confirm

03123996_01_en.eps

03223996_01_en.eps

48

01023996_01_en.eps

Performing the measurement

Automatic Measurement
The sampling is automatically triggered after attaining the minimum breath volume and the minimum blowing period.

Preparation
Switch on the device (see page 44).
Insert a new mouthpiece in the mouthpiece receptacle (see page 45).
After 2 seconds the following appear in the display: ­ customer specific info (if activated),
or ­ Input of data, see “Data input before or after the measurement” on
page 47 or ­ » WAIT «. The customer specific info can also be accessed by pressing the
key. End with the OK key.
After a temperature-dependent heating period, the following appears in the display: » READY « and a short audible signal beeps. The current test number is displayed at the same time. The right LED flashes green.
The device is ready for measurement.

11:53:47 04.12.2007
READY
Test No. 26

Procedure Prompt the test person to blow evenly and continuously into the
mouthpiece. A sufficient breath flow is confirmed by a continuous audible signal and the green LED.
During the delivery of the breath sample the following message appears in the display: » BLOW «.
If the breath sample is sufficient (minimum breath volume and the minimum blowing period were reached), the sampling is triggered. Then the green LED goes out, the audible signal stops.
The following message appears in the display: » WAIT Analyzing «

11:53:47 04.12.2007
BLOW

01123996_01_en.eps

49

Performing the measurement

01223996_01_en.eps

Test Result
After 3 to 20 seconds (dependent on the detected concentration) the measurement result appears in the display. ­ The measurement result is displayed with the corresponding unit. ­ In the default setting, the unit is set to mg/L (milligram ethyl alcohol
per liter breathing air) and the solution is set to two positions after the comma. The default setting can be changed on the main menu (see Technical Handbook Dräger Alcotest 7510).
In addition to the measurement result, an audible signal which is dependent on the measurement result is output, for example: measurement result below the limit value (PASS) = 1 x, measurement result between the limit values (ALERT) = 2 x, measurement result above the limit value (FAIL) = 3 x
If the measurement results are positive, it is recommended to perform a second measurement after 10 to 15 minutes in order to confirm the measurement result.

11:53:47 04.12.07
0.00 mg/L
Confirm

Documentation of the Results: ­ A result protocol can be generated with the Dräger Mobile Printer,
see page 53.

Additional Measurement Pressing the OK key prepares a new measurement.
During the recovery time of the sensor (see “Technical Data” on page 65) the following appears in the display: » WAIT «
NOTICE
Periodically switching off the device does not reduce this waiting period!

Error during the delivery of the breath sample
If the volume of the breath sample is insufficient, the following appears in the display: » Insufficient volume «. A short audible signal beeps and the red LED flashes. To repeat, press the OK key. Waiting period for a new testing standby is 4 to 20 seconds.
In the case of highly irregular exhalation, e.g. sudden termination or suction at the end of the breath sample, the following appears in the display: » Blow interruption «. A short audible signal beeps and the red LED flashes. To repeat, press the OK key. Waiting period for a new testing standby is 4 to 20 seconds.

11:53:47 04.12.07
Insufficient volume
Confirm

50

01323996_01_en.eps

01523996_01.eps

Remove the mouthpiece Depending on the version of the device, see page 41: Push the eject button upwards. The
mouthpiece is pushed out of the mouthpiece receptacle. or
Remove the mouthpiece upwards out of the mouthpiece receptacle.

Performing the measurement

CAUTION
For hygienic reasons, do not directly touch the mouth contact area of the used mouthpiece. Risk of infection!
Dispose of used mouthpieces in accordance with the local regulations.

02723996_01.eps

51

Additional Sampling Possibilities

01623996_01_en.eps

Additional Sampling Possibilities
In the menu, the functions can also be switched on or off in combination (see Technical Handbook Dräger Alcotest 7510).

Passive Sampling
With passive sampling, the sampling is manually triggered by the user of the device.Possible applications include the testing of the ambient air or the testing of the exhaled air of non-cooperating (e.g. unconscious) persons for the presence of alcohol.

Hold the ready to measure device without mouthpiece in the air to be analysed.
Briefly press OK key.
Display » NO ALCOHOL «: if no alcohol is present in the air to be analyzed.
or:
Display » ALCOHOL «: if alcohol is present in the air to be analyzed.

11:53:47 04.12.2007
NO ALCOHOL

Documentation of the Results: A result protocol can be generated with the Dräger Mobile Printer, see page 53.

Passive Confirm

Manual Sampling

The sampling can also be manually triggered if the test subject cannot produce the required breath volume.
Preparation and procedure as per automatic measurement (see page 49).
If » BLOW « appears in the display after the start of the exhalation and the test subject is no longer able to continue with the breath sample, briefly press the OK key. The sampling is now triggered.
During the result display, the message » Manual « also appears in the display.
Documentation of the Results: A result protocol can be generated with the Dräger Mobile Printer, see page 53.

11:53:47 04.12.07
0.00 mg/L
Manual Confirm

52

01723996_01_en.eps

Print the result protocol (basic setting)

Print the result protocol (basic setting)
With the Dräger Mobile Printer, see page 68. Preparation: Switch on the Dräger Mobile Printer.
If the Dräger Alcotest 7510 displays the test result: Place the device next to the Dräger Mo-
bile Printer in such a way that the infrared interfaces (IR) directly face each other. ­ The printing of the protocol automatically begins.
If the printing is done: Press OK key to start a new measure-
ment.
Printing stored test results: see Technical Handbook, function
» Show last tests «.

IR

OK
D
Alcotest 7510

03823996_01.eps

53

01823996_01_en.eps

Fault, Cause, Remedy

Fault, Cause, Remedy

An error message appears if a device fault occurs (see table).
The number that appears below the message is used for service functions. If the fault remains after repeatedly switching the device on and off, contact Dräger Safety.

11:53:47 04.12.07
Internal system error
88888888

Fault Device automatically switches off
Battery symbol empty
Date / time incorrect Menu is automatically exited
Not possible to switch on the device Insufficient volume / invalid sample No continuous tone during blowing

Cause

Remedy

Discharge the rechargea- Replace batteries, page 57

ble battery / batteries

or charge the rechargeable

batteries, page 58.

Device was ready to meas- Switch on device again. ure longer than 4 minutes

Discharge the rechargeable battery / batteries

At the next opportunity, replace batteries, page 57 or charge the rechargeable batteries, page 58.

Discharge backup battery Set date / time, page 63. Work with charged batteries.

The device automatically returns to the measurement function after 120 seconds

Select the menu item again.

Discharge the rechargea- Replace batteries, page 57

ble battery / batteries

or charge the rechargeable

batteries, page 58.

Test subject blows too weakly or non-continuously

Blow more strongly and continuously.

Device is not yet ready to Wait until the device is ready

measure

to measure.

54

Fault, Cause, Remedy

Error message in the display » Mouth alcohol «
» No printout possible «
» Temperature out of range « » Service date expired « The device alerts and blocks the operation, if set » Calibration error « » Accuracy check failed « » Sensor failure «
» Error datalogger «
» RFI error « » GPS ­ position not found «

Cause

Remedy

Residual alcohol in oral Repeat the measurement

cavity not completely reab- again approx. 15 minutes after

sorbed.

the last alcohol intake.

Measured value exceeds Use lower concentrations. display area.

Dräger Mobile Printer not Maintain a visual connection recognised or protocol and close proximity between not completely processed. the IR interfaces of the
devices during the entire printing period.

Device temperature is outside of the allowed temperature range.

Heat or cool the device and start the measurement again.

The service interval is expired.

Calibrate device; by trained service personnel, e.g. from Dräger Safety.

Calibration not successful.

Calibrate device again; by trained service personnel, e.g. from Dräger Safety.

Calibration test not successful.

Calibrate device; by trained service personnel, e.g. from Dräger Safety.

Sensor not within the specification.

Switch the device off and on again. If the error message appears again: have the sensor replaced by trained service personnel, e.g. from Dräger Safety.

Data storage not possible. Read the data memory and re-initialise; by trained service personnel, e.g. from Dräger Safety.

Electromagnetic fault is Correct fault source and

too strong.

repeat measurement.

No free visibility to the satellites.

Move to a location with free visibility and hold the device with the display directed upwards.

55

Fault, Cause, Remedy

Error message in the display » Internal System error « with fault number and text
Examples: . . . heating system . . . EC-sensor (offset) . . . sampling system
. . . backup battery . . . GPS communication » No Charging Temperature out of range «
» Wrong batteries «

Cause

Remedy

Device fault.

Remove batteries for a short period of time (2 seconds) and reinsert (only possible with alkaline batteries or NiMH rechargeable batteries). If the error continues to occur:make a claim with Dräger Safety Service.

Device is not ready to measure

Call Dräger Safety Service.

Device is not ready to measure

Call Dräger Safety Service.

Frequency determination

not successful.

Call Dräger Safety Service.

Pressure values not OK.

RTC Error

Call Dräger Safety Service.

Module Error

Call Dräger Safety Service.

Charger is connected, but the device temperature is outside of the allowed charging temperature range.

Heat or cool the device. Start the charging process again.

Only with alkaline batte- Check the orientation (polar-

ries or NiMH rechargea- ity) of the alkaline batteries or

ble batteries: cells

NiMH rechargeable batteries

incorrectly inserted (polar- and insert correctly if neces-

ity swapped).

sary.

Only with alkaline batteries or NiMH rechargeable batteries: individual cells defective.

Check the alkaline batteries or NiMH rechargeable batteries and replace if necessary.

56

Maintenance
Maintenance
Maintenance intervals
As required: Replace batteries, see page 57 or charge rechargeable batteries, see page 58.
Every six months: Check calibration. Only by authorised persons, see Technical Handbook Dräger
Alcotest 7510. — An extension of the calibration interval, e.g. to 1 year, is possible.
Prerequisite: Analysis of the sensor data and other quality assurance measures by the Service department of Dräger Safety.
Storage
The Dräger Alcotest 7510 also uses some power when it is in a deactivated state. Therefore, remove the batteries or rechargeable batteries from the device when not in use for a longer period of time.
CAUTION During long storage periods: remove the batteries or rechargeable NiMH batteries from the device when the device is not used for longer than 6 months. This avoids damage to the device caused by leaking battery acid.
Replacing the batteries
For devices with alkaline batteries or NiMH rechargeable batteries. Lithium- ion rechargeable batteries cannot be replaced.
If the battery smybol in the display stands at » « (empty) insert new batteries or charged NiMH rechargeable batteries, or place the device in the charger and charge:
Rotate the catch on the battery cover anticlockwise 90o and remove the battery cover.
­ On devices with NiMH rechargeable batteries, the catch is secured on the battery cover with a screw.
If the NiMH rechargeable batteries need to be replaced: In order to open the battery cover, first
loosen the screw. Then rotate the catch 90o and remove the battery cover.
57

01923996_01.eps

02023996_01.eps

Maintenance

Remove used alkaline batteries or NiMH rechargeable batteries and insert new alkaline batteries or charged NiMH re-chargeable batteries ­ observe correct polarity!
Replace the battery cover and lock the catch with a clockwise rotation. On devices with NiMH rechargeable batteries, secure the catch on the battery cover with the screw.

­

­ +­

+­+
Required batteries:: 4 units 1.5 V alkaline type (Mignon, LR6, AA)
or
4 units 1.2 V NiMH rechargeable batteries
­ In order to immediately reinsert the device after discharging the alkaline batteries or the NiMH rechargeable batteries, it is recommended to have replacement batteries or charged rechargeable batteries on hand.

CAUTION
Do not throw used alkaline batteries or NiMH rechargeable batteries into fire, charge them again, or open them by force. Danger of explosion!
Dispose of used alkaline batteries or NiMH rechargeable batteries only as hazardous waste, in accordance with local waste disposal regulations. Further information can be obtained from the relevant local authority and from appropriate waste disposal companies.

Charging the installed rechargeable batteries
For devices with NiMH or lithium-ion rechargeable batteries. If the battery symbol in the display stands at » « (empty) or if the device automatically switches off before reaching operational readiness: Charge the rechargeable batteries in the device or replace the NiMH rechargeable batter-
ies with charged rechargeable batteries, see page 57.
Only charge the device at temperatures between 5 and 40 oC (avoid exposure to direct sunlight). At extremely high or low temperatures, the device automatically stops the charging process.
58

Charging on the mains

Maintenance

OK
D
Alcotest 7510

OK
D
Alcotest 7510

03623996_01.eps

For charging on the mains, the power supply unit 600 mA (Part No. 83 16 991) and, optionally, the desktop charging cradle, is required. Connect the power supply unit 600 mA to the mains (100 to 240 V AC). Insert the jack plug into the charging socket of the device.
or Insert the jack plug into the charging socket on the rear side of the charging cradle and
place the device in the charging cradle with the front side facing forwards. Allow the carrying strap to dangle out of the front of the device, so that it does not block the contacts.
Car Charging

OK
D
Alcotest 7510

OK
D
Alcotest 7510

03723996_01.eps

For charging in the car, the car charger 600 mA (Part No. 83 20 252) and, optionally, the desktop charging module, is required. Insert the connector of thecar charger into the cigarette lighter receptacle. Insert the jack plug into the charging socket of the device.
or Insert the jack plug into the charging socket on the rear side of the charging cradle and
place the device in the charging cradle with the front side facing forwards. Allow the carrying strap to dangle out of the front of the device, so that it does not block the contacts.
59

03523996_01_en.eps

Maintenance

During the entire charging process on the mains or in the car: ­ The following message appears in the display: » Charging «
The battery symbol also cyclically switches the charge state from empty to full.
­ A discharged rechargeable battery is completely recharged in approx. 6 to 8 hours.
­ The message » Charging complete « appears in the display.

11:53:47 04.12.2007
Charging

CAUTION The charging accessorie listed in the oder list, page 67, is to be used only.
NOTICE Obtain the charge state of the rechargeable batteries. For NiMH rechargeable batteries: To ensure the operational readiness of the device even when it is not in use, completely charge the rechargeable battery at least every 3 months.
For lithium-ion rechargeable batteries: When not in use for a longer period of time, charge to at least 50 % every 6 months in order to prevent a reduction in the service life (see function » 50% charging « in the quick menu on page 63.

60

Maintenance
Maintenance
The device does not need any special maintenance. If dirt and deposits are present, the charging contacts must be carefully cleaned.
Cleaning
Do not submerge in liquids, no liquids can touch the connectors! For cleaning: wipe with a disposable towel. do not use any detergents that contain alcohol or solvents.
CAUTION The use of detergents with a high alcohol content can greatly reduce the service life of the internal sensor or generate an erroneous positive trest result in the short-term.
Disposal of Electric and Electronic Equipment
EU-wide regulations for the disposal of electric and electronic appliances, which have been defined in the EU Directive 2002/96/EC and in national laws, are effective from August 2005 and apply to this device. Common household appliances can be disposed of using special collecting and r ecycling facilities. However, as this device has not been registered for household usage, it must not be disposed of through these means. The device can be returned to your national Dräger Safety Sales Organization for disposal. Please do not hesitate to contact the above if you have any further questions on this issue.

Menu Functions
Activation
Switch on the device. After the message » WAIT « or » READY « appears, the menu functions can be selected. For the ” Quick Menu “: briefly press the key.
M
­ The ” Main Menu ” is only accessible to authorised persons with PIN entry, see Technical Handbook Dräger Alcotest 7510.
Operation
­ Select functions with the key and the key: M
­ The selected function is displayed inverted. ­ Activation of the selected function with the OK key. ­ Selecting ” Back to measurement” brings the device back into the measurement
mode by pressing the OK key. If no key is pressed within 120 seconds, the device automatically returns to the measurement mode.
Symbols in the display
Graphical symbols (icons) simplify the navigation through the various menus: Return to next menu level Menu item with submenu Open menu item Selected function Other entries below visible entries Other entries above visible entries
Function Activated function or setting. Current display (1) of the total display number (6)
62

02323996_01_en.eps

Menu Functions

The Quick Menu
The following functions are active in the default setting on device delivery.
» Show last tests « Select the last tests with the , and keys.
M
Go back with the OK key. The data records of the last tests can be transferred to a PC with the PC program Dräger Diagnostics.
» Show cal. dates « Show last and next calibration date. Go back with the OK key.
» Set contrast/backlight « Setting of contrast and backlight of the display background light. Modify with the , and keys.
M
Go back with the OK key.
» Show last error « The last occurred error is displayed. Go back with the OK key.

11:53:47 04.12.07
Back to measurement Show last tests Show cal. dates Set contrast/backlight Show last error

The size of the quick menu can be expanded with the PC software Dräger Diagnostics, e.g.:
» Set date « Set with the , and keys.
M
Go back with the OK key.
» Set time « Set with the , and keys.
M
Go back with the OK key.
» Set language « Select with the , and keys..
M
Go back with the OK key.
» Accuracy check « After selecting the accuracy check with the OK , the test gas that must be used appears in the display. Continue the procedure according to the messages in display. After supplying gas to the instrument the result will be displayed, if applicable the device functions will be locked/unlocked dependent on deviation to target value.
» 50% charging « Complete 50% charging of lithium ion batteries for periods of longer storage.

11:53:47 04.12.07
Back to measurement Show last tests Show cal. dates Set contrast/backlight Show last error Set date Set time Set language Accuracy check 50% charging

03323996_01_en.eps

63

Technical Data

Technical Data
Measuring Principle Measuring range (Default setting)
Breath alcohol concentration
Blood alcohol concentration

Electrochemical sensor
0.00 to 3.0 mg/L (Ethyl alcohol mass per breath volume at 34 oC and 1013 hPa.)
0,00 bis 6,00 (Ethyl alcohol mass per blood volume or blood mass at 20 oC and 1013 hPa. By giving this data, a conversion factor of 2100 is used between the breath alcohol concentration and blood alcohol concentration.)

Display Area
The display is identical with the measuring range. If the measurement result is outside of the measuring range, three arrows are shown in the display » « = Measuring range exceeded.

Ambient conditions during operation
during storage

­10 to 50 oC 10 to 100 % r.H. (non-condensing) 600 to 1100 hPa
­40 to 70 oC

Sampling minimum blowing time
minimum volume

adjustable preset at 2 seconds
adjustable preset at 1.2 liter

64

Technical Data

Analysis Time

Time from the end of sample delivery to the displaying of the results. The analysis time is

dependent on the ambient temperature.

at room temperature:

0.00 mg/L

<4 seconds

0.50 mg/L at 0 oC:

<10 seconds

0.00 mg/L

<4 seconds

0.50 mg/L

<20 seconds

Measuring accuracy Standard deviation of the repeatability with ethyl alcohol standard in the temperature range ­5 bis 40 oC)

to 0.50 mg/L 0,50 mg/L

0.008 mg/L or 1.7 % of measured value whichever value is greater

to 1.00 1.00

0.017 or 1.7 % of measured value whichever value is greater

Sensitivity drift

typically 0.6 % of the measurement value/month

Waiting period between measurements

The sensor is heated in relation to the outside temperature. This means that the waiting periods are significantly reduced in comparisson with unheated systems.

Waiting period after a previous measurement with: (specification in breath alcohol concentration)

0.00 mg/L

0.00 mg/L, independent of the concentration

<10 seconds <20 seconds

Values for other units of measurements can be calculated and displayed accordingly.

65

Technical Data

Dimensions (H x W x D)

with alkaline batteries or NiMH

approx. 185 mm x 90 mm x 44 mm

rechargeable batteries

with lithium-ion rechargeable batteries approx. 185 mm x 90 mm x 49 mm

Weight

approx. 433 g (with batteries / rechargeable batteries)

Power supply Alkaline batteries NiMH rechargeable batteries lithium-ion rechargeable battery

4 units 1.5 V alkaline type (Mignon, LR6, AA) approx. 1500 measurements, or
4 units 1.2 V NiMH (Mignon, LR6, AA) approx. 1500 measurements, or
integrated in the device approx. 50 % more measurements possible than with NiMH rechargeable batteries

CE marking

Electromagnetic compatibility (Directive 89/336/EEC)

66

Order List
Name and Description
Dräger Alcotest 7510 Standard
Dräger Alcotest 7510 Customer specific configuration

Order List
Part No. 83 19 760 83 19 700

Accessories, Spare Parts and Consumables
“Standard” mouthpiece without non return valve (1000 units) “Standard” mouthpiece with non return valve (1000 units)
“Classic” mouthpiece without non return valve (25 units) “Classic” mouthpiece with non return valve (25 units)
Alkaline battery (1 unit) NiMH rechargeable battery (1 unit) Power Supply Unit 600 mA for charging devices with NiMH rechargeable batteries or with a lithium-ion rechargeable battery Car charger 600 mA Desktop charging cradle, compl. Alcotest 7510 to charge the Dräger Alcotest 7510 with power supply 83 16 991. Consisting of: charging cradle base unit, charging cradle adapter, wall bracket
Holster Set, compl. Alcotest 7510 consisting of: holster, belt clip, protective cover plus div. velcro fastening (e.g. to attach inside vehicles)

68 10 830 68 11 065 68 05 700 68 05 703
13 35 804 18 90 092 83 16 991
83 20 252 83 19 762
83 19 761

67

Order List

Name and Description

IR-Modul Alcotest 7510 Infrared interface adapter for Holster Alcotest 7510 for data transfer via IR interface. The adapter can be connected via an USB cable to a PC.
IR-Module + Holster Alcotest 7510 consisting of: Holster and IR-Modul Alcotest 7510 (83 19 763). Use in cojunction with 83 19 765 for mobile use. Use in conjunction with 83 19 762 for stationary use.
Holster Set Alcotest 7510 consisting of… belt clip, protective cover plus div. velcro fastening as add-on of 83 19 764 resp. also spare part of 83 19 761
Device case for the Dräger Alcotest 7510 and Dräger Mobile Printer. Space for mouthpieces, batteries, paper rolls, charging cable, communication cable, keyboard, etc.
Dräger Mobile Printer
Dräger Diagnostics PC program for the data communication with Dräger Alcotest 7510 USB cable 1)
Wrist strap
Packaging unit for Dräger Alcotest 7510
Quick Guide Dräger Alcotest 7510
Technical Handbook Dräger Alcotest 7510

Part No. 83 19 763
83 19 764
83 19 765
83 20 251
83 19 310 83 19 342 83 18 657 83 17 922 83 19 747 90 33 040 90 23 997

————————1) When using a different USB cable, observe a maximum length of 3 m for CE reasons.
68

Dräger Safety AG & Co. KGaA Revalstraße 1 D-23560 Lübeck Germany Tel. +49 451 8 82- 0 Fax +49 451 8 82- 20 80 www.draeger.com