BRAUN IRT6515 ThermoScan 6 Ear Thermometer Owner’s Manual
- June 13, 2024
- Braun
Table of Contents
ThermoScan® 6
Ear thermometer
Owner’s Manual
IRT6515
A004165R2
05JUL23
IRT6515 ThermoScan 6 Ear Thermometer
Product description
1. Hygienic lens filter (Box of 20)
2. Probe tip
3. Probe
4. ExacTemp® Stability Indicator
5. Hygienic lens filter ejector| 6. Display
7. Power button
8. Start button
9. Battery door
10. Protective case
---|---
The Braun ThermoScan Ear thermometer has been carefully developed for
accurate, safe, and fast temperature measurements in the ear .
The shape of the thermometer probe prevents it from being inserted too far
into the ear canal, which can hurt the eardrum .
However, as with any thermometer, proper technique is critical to obtaining
accurate temperatures .
Therefore, read the instructions carefully and thoroughly .
Indications for Use / Intended Use
Braun Thermoscan® IRT6000 Series Infrared ear thermometers are indicated for
the intermittent measurement of human body temperature by consumers of all
ages in a home use environment .
The hygienic lens filter is used as a sanitary barrier between the thermometer
and the ear canal .
Warnings and Precautions
The operating ambient temperature range for this thermometer is 50 – 104 °F
(10 – 40 °C) .
DO NOT expose the thermometer to temperature extremes (below –13 °F / –25 °C
or over 131 °F / 55 °C) or excessive humidity (15- 95 % RH) .
This thermometer must only be used with genuine Braun ThermoScan Hygienic lens
filters (LF 40) .
To avoid inaccurate measurements ALWAYS use this thermometer with a new, clean
hygienic lens filter attached . If the thermometer is accidentally used
without a hygienic lens filter attached, clean the lens (see «Care and
cleaning» section) . Keep hygienic lens filters out of reach of children .
This thermometer is intended for household use only .
Use of this thermometer is not intended as a substitute for consultation with
your physician .
This thermometer is not intended for pre-term babies or small-for-gestational
age newborn babies (birth to 4 weeks of age) . Pre-term is defined as babies
born before 37 weeks of gestation . Small-for-gestational-age is defined as a
newborn baby, born at 37 weeks or later, with a weight below the 10th
percentile for newborn babies of the same gestational age .
Nursing mothers should not take a temperature measurement on themselves or
have their temperature measured by another individual while or immediately
after nursing a baby, as nursing can affect body temperature .
This thermometer is not intended to interpret hypothermic temperatures . If
the device displays a temperature of 97 .5°F (36 .4°C) or less, and the
individual is exhibiting atypical symptoms or behaviors, contact your doctor
or health care professional .
DO NOT allow children under 12 years to take their temperature unattended
.
DO NOT modify this equipment without authorization of the manufacturer .
People who are on antibiotics, analgesics, or antipyretics should not be
assessed solely on temperature readings to determine the severity of their
illness .
Temperature elevation may signal a serious illness, especially in neonates and
infants, or in adults who are elderly, frail, or have a weakened immune system
. Please seek professional advice immediately when there is a temperature
elevation and if you are taking temperature on:
- Neonates and infants under 3 months (Consult your physician immediately if the temperature exceeds 99 .4 °F or 37 .4 °C) .
- Individuals having diabetes mellitus or a weakened immune system (e .g ., HIV positive, cancer chemotherapy, chronic immunosuppressant treatment, splenectomy) .
- Individuals who are bedridden (e .g ., paraplegia, quadriplegia, stroke, chronic illness, surgical recovery, nursing home patient, stroke, chronic illness, recovering from surgery) .
- A transplant recipient (e .g ., liver, heart, lung, kidney) .
This thermometer contains small parts that can be swallowed or produce a choking hazard to children Keep out of reach of children under 12 years .
Body temperature
Normal body temperature is a range . It varies by site of measurement, and it
tends to decrease with age . It also varies from person to person and
fluctuates throughout the day .
Therefore, it is important to determine normal temperature ranges . This is
easily done using Braun ThermoScan . Practice taking temperatures on yourself
and healthy family members to determine the normal temperature range .
Note: When consulting your physician, communicate that the ThermoScan
temperature is a temperature measured in the ear and if possible, note the
individual’s normal ThermoScan temperature range as additional reference .
Why measure in the ear?
The goal of thermometry is to measure core body temperature 1 , which is the
temperature of the vital organs . Ear temperatures accurately reflect core
body temperature, 2 since the eardrum shares blood supply with the temperature
control center in the brain, 3 the hypothalamus . Therefore, changes in body
temperature are reflected sooner in the ear than at other sites . Axillary
temperatures measure skin temperature and may not be a reliable indicator of
core body temperature . Oral temperatures are influenced by drinking, eating
and mouth breathing . Rectal temperatures often lag behind changes in core
body temperature and there is a risk of cross-contamination .
Color-coded fever guidance
Temperature guidance helps you to better understand the meaning of your
child’s temperature with the color indicated on the display . The screen
displays a green, yellow, or red backlight depending on the temperature
reading .
It is important to note that a patient’s temperature may be impacted by age,
gender, or medical condition, or by factors such as ambient temperature, time
of day, medications taken or the temperature measurement site4 .
Note: Backlight is white with no fever indication when temperatures of 93 .2</ T</96 .3 are registered
Color range | Reading | Meaning |
---|---|---|
Green | >96 .3 °F – 99 .4 °F | |
(>35 .7 °C – 37 .4 °C) | Normal | |
Yellow | >99 .4 °F – 101 .3 °F | |
(>37 .4 °C – 38 .5 °C) | Fever | |
Red | >101 .3 °F | |
(>38 .5 °C ) | High fever |
1 Guyton A, Textbook of Medical Physiology, W .B . Saunders, Philadelphia,
1996, p 919 .
2 Guyton A, Textbook of Medical Physiology, W .B . Saunders, Philadelphia,
1996, pp 754-5 .
3 Netter F, Atlas of Human Anatomy, Novartis Medical Education, East Hanover,
NJ, 1997, pp 63, 95 .
4 Herzog L, Phillips S . Addressing Concerns About Fever . Clinical Pediatrics
. 2011; 50(#5): 383-390 .
How does Braun ThermoScan work?
The Braun ThermoScan measures the infrared heat generated by the eardrum and
surrounding tissues . When placed in the ear, the Braun ThermoScan
continuously monitors the infrared energy and the ExactTemp® Stability
Indicator displays a reading only if the thermometer detects that an accurate
and stable temperature measurement has been taken .
How to use your Braun ThermoScan
-
Remove the thermometer from the protective case.
-
Push the Power button .
During an internal self-check, the display shows all segments .
Then the last temperature taken will be displayed for 5 seconds . -
The hygienic lens filter indicator will blink to signal one is needed .
To achieve accurate measurements, make sure a new, clean hygienic lens filter is in place before each measurement .
Attach a new hygienic lens filter by pushing the thermometer probe straight into the hygienic lens filter inside the box and then pulling out.
Note: The Braun ThermoScan will not work unless a hygienic lens filter is attached.
The thermometer is ready to take a temperature when the display looks like the following image: -
Fully insert the IRT probe into the ear canal and point it towards the opposite temple . Make sure the IRT is pointed toward the ear drum and not the canal walls .
Once positioned correctly, press and release the start button . -
ExacTemp Stability Indicator light will pulse while temperature measurement is in progress . The light will remain solid for 3 seconds to indicate that a successful temperature reading has been achieved .
NOTE: If the probe has been properly inserted into the ear canal during the measurement, a long beep will sound to signal the completed measurement .
If the probe has NOT been constantly placed in a stable position in the ear canal, a sequence of short beeps will sound, the ExacTemp Stability Indicator light will go out and the display will show an error message (POS = position error) .
See Errors and troubleshooting section for more information . -
The confirmation beep indicates that an accurate temperature measurement has been taken . The result is shown on the display . Temperature reading displayed is an equivalent of oral temperature .
To repeat Go to step 3 . -
For the next measurement, press Eject button to remove, and discard used hygienic lens filter, and put on a new, clean hygienic lens filter .
The Braun ThermoScan Ear thermometer turns off automatically after 60 seconds of inactivity . The thermometer can also be turned off by pressing the Power button .
The display will briefly flash OFF and it will go blank .
Temperature-taking hints
ALWAYS replace disposable hygienic lens filters to maintain accuracy and
hygiene . A measurement taken in the right ear may differ from a measurement
taken in the left ear . Therefore, ALWAYS take the temperature in the same ear
. The ear must be free from obstructions or excess earwax build-up to take an
accurate reading .
External factors may influence ear temperatures, including when an individual
has:
Factor | Affects |
---|---|
Poor probe placement | |
Used hygienic lens filter | |
Dirty lens | |
In the cases below, wait 20 minutes prior to taking a temperature . | |
Extreme hot and cold room temperature | |
Hearing aid |
Use the untreated ear if ear drops or other ear medications have been placed
in the ear canal.
If an individual has been laying down on their side with an ear face down on a
pillow, use the ear that has been facing upwards .
Last reading recall
- The last temperature taken is stored in the memory and will be automatically displayed for 5 seconds when it is turned on again.
How to illuminate the display
The thermometer includes a convenient night light to illuminate the display in
a dim environment .
The light will turn on when you press any button . It will stay on until the
thermometer is inactive for up to 15 seconds even after temperature is taken.
How to change temperature between °C and °F
Your Braun ThermoScan is shipped with the Fahrenheit (°F) temperature scale
activated . If you wish to switch to Celsius (°C) and/or back from Celsius to
Fahrenheit, proceed as follows:
- Make sure the thermometer is turned off .
- Press and hold down the Power button After about 3 seconds the display will show this sequence: °F / SET / °C / SET .
- Release the Power button when the desired temperature scale is shown.
There will be a short beep to confirm the new setting, then the thermometer is turned off automatically .
Care and cleaning
The probe tip is the most delicate part of the thermometer . It must be clean
and intact to ensure accurate readings . If the thermometer is ever
accidentally used without a hygienic lens filter, clean the probe tip as
follows:
Very gently wipe the surface with a cotton swab or soft cloth moistened with
alcohol . After the alcohol has completely dried out, you can put a new
hygienic lens filter on and take a temperature measurement .
If the probe tip is damaged, please contact Consumer Relations .
Use a soft, dry cloth to clean the thermometer display and exterior . DO NOT
use abrasive cleaners . NEVER submerge this thermometer in water or any other
liquid . Store thermometer and hygienic lens filters in a dry location free
from dust and contamination and away from direct sunlight .
Additional hygienic lens filters (LF 40) are available at most stores carrying
Braun ThermoScan .
Replacing the batteries
The thermometer is supplied with two 1 .5 V type AA (LR 06) batteries .
- Insert new batteries when the battery symbol appears on the display .
- Open the battery compartment . Remove the batteries and replace with new batteries, making sure the poles are in the right direction.
- Snap battery door into place .
Only discard empty batteries . To protect the environment, dispose of empty
batteries at your retail store or at appropriate collection sites according to
national or local regulations .
Do not mix alkaline, standard, or rechargeable batteries .
Calibration
The thermometer is initially calibrated at the time of manufacture . If this
thermometer is used according to the use instructions, periodic re-adjustment
is not required . If at any time you question the accuracy, please contact
Consumer Relations .
Errors and troubleshooting
Error message | Situation | Solution |
---|---|---|
No hygienic lens filter is attached . | Attach new, clean hygienic lens |
filter .
| The thermometer probe was not positioned securely in the ear .
An accurate measurement was not possible .| Check if the positioning of the
probe is correct and remains stable .
Replace hygienic lens filter and reposition . Press Start button to begin a
new measurement .
| Ambient temperature is not within the allowed operating range (50 – 104 °F
or 10 – 40 °C) .| Allow the thermometer to remain for 30 minutes in a room
where
the temperature is between 50 and 104 °F or 10 and 40 °C .
| Temperature taken is not within typical human temperature range (93 .2 –
108 °F or 34 – 42 .2 °C) .
HI = too high
LO = too low| Make sure the probe tip and lens are clean and a new, clean
hygienic lens filter is attached . Make sure the thermometer is
properly inserted .
Then, take a new temperature .
| System error – self-check display flashes continuously and will not be
followed by the ready beep and the ready symbol or screen is blank .
If error persists, If error persists, If error still persists,| Wait 1 minute
until the thermometer turns off automatically, then turn on again .
. . . reset the thermometer by removing the batteries and putting them back in
.
. . . replace the batteries with new batteries .
. . . please contact the Consumer Relations .
| Battery is low, but thermometer will still operate correctly .| Insert new
batteries .
| Battery is too low to take correct temperature measurement .| Insert new
batteries .
Product specifications
Displayed temperature range: | 93 .2 – 108 °F | 93 .2 – 108 °F |
---|---|---|
Operating ambient temperature range: | 50 – 104 °F | 50 – 104 °F |
Storage temperature range: | -13 – 131 °F | -13 – 131 °F |
Storage relative humidity: | Storage 15-95% RH (non-condensing) | |
Operating relative humidity: | Operating 10-95% | |
Display resolution: | 0 .1 °F or °C | |
Accuracy for displayed temperature range | Maximum Laboratory Error | |
35 – 42 °C (95 – 107 .6 °F): | ±0 .4 °F | (±0 .2 °C) |
Outside this range: | ±0 .5 °F | (±0 .3 °C) |
Clinical repeatability: | ±0 .26 °F | (±0 .14 °C) |
Reference Body Site: | Oral Equivalent | |
Battery life: | 2 years / 350 measurements | |
Service life: | 5 years |
Clinical accuracy characteristics and calibration procedures are available
upon request .
This thermometer is specified to operate at a pressure of 1 atmosphere or at
altitudes with an atmospheric pressure up to 1 atmosphere . (700 – 1060 hPa) .
Subject to change without notice .
This appliance conforms to the following standards:
This infrared thermometer meets requirements established in ASTM Standard E
1965-98 as it pertains to infrared thermometers . Full responsibility for the
conformance of this product to the standard is assumed by Kaz USA, Inc . ASTM
laboratory accuracy requirements in the display range of 37 to 39 °C (98 to
102 °F) for IR thermometers is +/- 0 .2 °C (+/- 0 .4 °F), whereas for mercury-
in-glass and electronic thermometers, the requirement per ASTM Standards
E667-86 and E1112-86 is +/- 0 .1°C (+/- 0 .2 °F) .
EN ISO 80601-2-56: Medical electrical equipment – Part 2-56: Particular
requirements for basic safety and essential performance of clinical
thermometers for body temperature measurement .
EN IEC 60601-1: Medical electrical equipment – Part 1: General requirements
for basic safety and essential performance .
EN ISO 14971: Medical devices – Application of risk management to medical
devices .
EN ISO 10993-1: Biological evaluation of medical devices – Part 1: Evaluation
and testing within a risk management process .
EN IEC 60601-1-2: Medical electrical equipment – Part 1-2: General
requirements for basic safety and essential performance – Collateral standard:
Electromagnetic compatibility – Requirements and tests .
EN ISO 15223-1: Medical devices . Symbols to be used with medical device
labels, labeling and information to be supplied . General requirements .
EN IEC 60601-1-11: Medical electrical equipment – Part 1-11: General
requirements for basic safety and essential performance – Collateral standard:
Requirements for medical electrical equipment and medical electrical systems
used in the home healthcare environment .
This product conforms to the provisions of the EC directive 93/42/EEC .
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC . For
detailed description of EMC requirements please contact your authorized local
Service Center (See guarantee card) .
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT .
Please do not dispose of the product in the household waste at the end of its
useful life .
To protect the environment, dispose of empty batteries at appropriate
collection sites according to national or local regulations .
Limited three year warranty
You should first read all Instructions before attempting to use this product.
A . This 3 year limited warranty applies to repair or replacement of product
found to be defective in material or workmanship . This warranty does not
apply to damage resulting from commercial, abusive, unreasonable use or
supplemental damage . Defects that are the result of normal wear and tear will
not be considered manufacturing defects under this warranty . KAZ IS NOT
LIABLE FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY NATURE, ANY IMPLIED
WARRANTY OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE ON THIS PRODUCT IS LIMITED IN DURATION TO THE
DURATION OF THIS WARRANTY. Some jurisdictions do not allow the exclusion or
limitation of incidental or consequential damages or limitations on how long
an implied warranty lasts, so the above limitations or exclusions may not
apply to you . This warranty gives you specific legal rights, and you also may
have other rights which vary from jurisdiction to jurisdiction . This warranty
applies only to the original purchaser of this product from the original date
of purchase .
B . At its option, Kaz will repair or replace this product if it is found to
be defective in material or workmanship .
C . This warranty does not cover damage resulting from any unauthorized
attempts to repair or from any use not in accordance with the instruction
manual .
Call us toll-free at 1-800-477-0457 or
visit Braun Healthcare .com/contact-us Please be sure to specify a model
number .
NOTE: IF YOU EXPERIENCE A PROBLEM, PLEASE CONTACT CONSUMER RELATIONS FIRST OR SEE YOUR WARRANTY. DO NOT RETURN THE PRODUCT TO THE ORIGINAL PLACE OF PURCHASE. DO NOT ATTEMPT TO OPEN THE THERMOMETER HOUSING YOURSELF, DOING SO MAY VOID YOUR WARRANTY AND CAUSE DAMAGE TO THE PRODUCT OR PERSONAL INJURY.
Guidance and manufacturer’s declaration — electromagnetic immunity
The IRT6515 is intended for use in the electromagnetic environment specified
below.
The customer or the user of the IRT6515 should ensure that it is used in such
an environment.
Phenomenon| Basic EMC
standard or test
method| IMMUNITY TEST LEVELS
HOME HEALTHCARE
ENVIRONMENT
ELECTROSTATIC DISCHARGE| IEC 61000-4-2| i 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Radiated RF EM fields a)| IEC 61000-4-3| 10 V/m f)
80 MHz — 2,7 GHz b) 80 % AM at 1 kHz c)
Proximity fields from RF wireless
communications equipment| IEC 61000-4-3| See RF wireless communication
equipment immunity table below
RATED power frequency
d
magnetic fields d) e)| IEC 61000-4-8| 30 Alm g)
50 Hz or 60 Hz
a) The interface between the PATIENT physiological signal simulation, if used,
and the ME EQUIPMENT or ME SYSTEM Shall be located within 0,1 m of the
vertical plane of the uniform field area in one orientation of the ME
EQUIPMENT or ME SYSTEM.
b) ME EQUIPMENT and ME SYSTEMS that intentionally receive RF electromagnetic
energy for the purpose of their operation shall be tested at the frequency of
reception. Testing may be performed at other modulation frequencies identified
by the RISK MANAGEMENT PROCESS. This test assesses the BASIC SAFETY and
ESSENTIAL PERFORMANCE of an intentional receiver when an ambient signal is in
the passband. It is understood that the receiver might not achieve normal
reception during the test.
c) Testing may be performed at other modulation frequencies identified by the
RISK MANAGEMENT PROCESS.
d) Applies only to ME EQUIPMENT and ME SYSTEMS with magnetically sensitive
components or circuitry.
e) During the test, the ME EQUIPMENT or ME SYSTEM may be powered at any
NOMINAL input voltage, but with the same frequency as the test signal (see
Table 1).
f) Before modulation is applied.
g) This test level assumes a minimum distance between the ME EQUIPMENT or ME
SYSTEM and sources of power frequency magnetic field of at least 15 cm. If the
RISK ANALYSIS shows that the ME EQUIPMENT or ME SYSTEM will be used closer
than 15 cm to sources of power frequency magnetic field, the IMMUNITY TEST
LEVEL shall be adjusted as appropriate for the minimum expected distance.
Guidance and manufacturer’s declaration — electromagnetic emissions
The IRT6515 is intended for use in the electromagnetic environment specified
below The.customer or the
user of the IRT6515 should ensure that it is used in such an environment.
Emissions Test| Compliance| Electromagnetic environment —guidance
RF Emissions
CISPR 11| Group 1| The ME equipment uses RF energy only for its internal
furction.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions
CISPR 11| Class B| Complies
Harmonic emissions
IEC 61000-3-2| Not
Applicable| The ME equipment is solely battery powered.
Voltage fluctuations/
flicker emissions| Not
Applicable
Guidance and manufacturer’s declaration — RF wireless communication equipment
immunity
Test
frequency
(MHz)| Band a)
(MHz)| Service a)| Modulation b)| Maximum
power
(W)| Distance
(m)| IMMUNITY
TEST LEVEL
(V/m)
385| 380 —390| TETRA 400| Pulse
modulation b)
18 Hz| 1,8| 0,3| 27
450| 430 — 470| GMRS 460,
FRS 460| FM c)
± 5 kHz deviation
1 kHz sine| 2| 0,3| 28
710| 704 — 787| LTE Band 13,
17| Pulse
modulation b)
217 Hz| 0,2| 0,3| 9
745
780
810| 800 — 960| GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5| Pulse
modulation b)
18 Hz| 2| 0,3| 28
870
930
1720| 1 700 —
1 990| GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS| Pulse
modulation b)
217 Hz| 2| 0,3| 28
1845
1970
2450| 2 400 —
2 570| Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7| Pulse
modulation b)
217 Hz| 2| 0,3| 28
5240| 5 100 —
5 800| WLAN 802.11
a/n| Pulse
modulation b)
217 Hz| 0,2| 0,3| 9
5500
5785
a)For some services, only the uplink frequencies are included.
b)The carrier shall be modulated using a 50 % duty cycle square wave signal.
c)As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be
used because while it does not represent actual modulation, it would be worst
case.
Contact Us
Questions or Comments?
Call: 1-800-327-7226
Visit: BraunHealthcare.com/contact-us
www.braunhealthcare.com
Certain trademarks used under license from The Procter & Gamble Company or its
affiliates .
ThermoScan® and ExacTemp® are trademarks owned by Helen of Troy Limited and/or
its affiliates .
© 2023 All rights reserved .
Kaz USA, Inc ., a Helen of Troy Company 400 Donald Lynch Blvd ., Suite 300
Marlborough, MA 01752
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>