POWER PROBE PPDMM CATIV Digital Multimeter Instruction Manual

June 13, 2024
POWER PROBE

**POWER PROBE PPDMM CATIV Digital Multimeter Instruction Manual

**

SERVICE

Thank you for purchasing our product!
As our customers are our top priority and we strive for 100% satisfaction, we would appreciate your feedback on the product.

Please contact us if you have any questions regarding the product. Qur professional support team is always ready to answer your enquiries and provide assistance.

GENERALSPECIFICATIONS

PPDMM is a stable, safe, reliable compact digital handheld 6000 count auto-ranging multimeter. This meter can measure AC/DC voltage, AC/DC current, resistance, capacitance, frequency, temperature, diodes and continuity. This meter is ideal for many situations, whether you’re a professional or causal user.

  • Operating Altitude: 2000m
  • Relative Humidity: 75% max operating
  • Operating Temperature: 0°C~40°C/32°F~104°F (<80% RH)
  • Storage Temperature: -10°C~60°C/14°F~140°F (<70% RH)
  • Accuracy Temperature: -18°C~28°C/64°F~82°F (<80% RH)
  • Temperature Coefficient: 0.1x(specified accuracy)/°C (<18°C or >28°C)
  • Sampling Frequency: approx. 3 times/sec.
  • Fuse Protection: /A/mA input: FE00mA/600V 10A input: F10A/600V
  • DC/AC Voltage: 600V
  • DC/AC Current: 10A
  • Resistance: 60MQ
  • Capacitance: 60mF
  • Frequency: 10kHz
  • Diodes: 2.7V « Continuity: <50Q
  • Temperature: -20°C~1000°C/-4°F~1832°F
  • LCD Display: 3 % digit display (6000 counts)
  • Product Supply: 3×1.5V AAA batteries
  • Product Size: 160mmx74mmx49mm / 6.3’%2.9″x1.9″
  • Product Weight: 4829/ 1.06lb
  • Safety Rating: CAT IV 600V; pollution degree: Il
  • Safety Standards: IEC61010-1
  • Pollution Degree: 2
  • Accuracy: +(of reading + # of least significant digits)

WARNINGS

To avoid electric shock and injury or damage to the meter, observe the following safety methods:

  •  Check the meter before use to make sure there was no damage during transit.
  • Check that the insulation on the test leads is not damaged and/or wires are not exposed.
  • If any faults or abnormalities are observed, the meter should not be used and should be checked out prior to use.
  • Never exceed the protection limit values indicated in specifications for each range of measurement.
  • Always be careful when working voltages above 60V DC or 30V AC rms, keep fingers behind the probe barrier while measuring.
  • Make sure the rotary switch is in the correct position before measurement.
  • Never use the meter in an environment with explosive gas, vapor or dust.
  • Aiways keep fingers behind probe barriers when making measurements.
  • When connecting test leads to a circuit, connect the black test lead first, then the red lead. Disconnect in the opposite order.
  • Turn off power and discharge all capacitors first before measuring resistance, continuity or diodes.
  • Failure to follow safety guidelines could compromise the safety features of this meter.
  • Do not use the meter without the battery cover in place.
  • Replace the batteries as soon as the low battery symbol “ ” to avoid false reading that could lead to electric shock and injury.
  • Remove test leads from all circuits before opening the battery cover.

SAFETY SYMBOLS

Important safety information
AC(alternating current)
DC (direct current)
DC/AC Voltage or Current
Gmund
Double insulation protection
Fuse
Compliance with EU regulations|

FEATURE DETAILS

  1. LCD Display

  2. Hold/Backlight Button
    Press. % to hold the current reading on the display. Press the button again to release the hold Hold “~” to turn on the backlight. Hold the button again to manually turn off the backlight.

  3. Max/Min Button
    In all modes (except continuity, diode, capacitance), press “Max/Min” and the display will show the maximum value recorded since the button was pressed. Press the button again and the display will show the minimum value recorded. Pressing the button a third time will show the difference between the max and min value. Hold “Max/Min” to return the display to normal readings.

  4. Function Switch Button
    Press “FUNC” to switch between functions or between AC/DC current.

  5. Frequency/Duty Cycle Button (Hz/%)
    In AC voltage/current modes, press “Hz%” and the display will show the frequency measurement. Press the button again to switch to duty cycle. Press the button a third time to return to normal display.

  6. Range Button
    In voltage, current and resistance modes, the default range is auto.
    To enter manual range, press “RAN”. Each press of the button increases the range, and returns to the lowest range when pressed in the highest range. Hold “RAN” to return to auto range. (Only auto range is available in capacitance mode)

  7. Relative Measurement Button
    In all modes (except resistance, continuity, diode), press “REL” and the display will show the relative value, i.e. the difference between the stored value when the button was pressed and the currently measured value. (REL = stored value — currently measured value) Press the button again to return the display to normal. (In REL mode, auto range is disabled)

  8. Rotary Switch

  9. Input Jack (all measurements; current below 600mA) 10.A Jack (current measurements between 600mA-10A only)

  10. Common Jack (all measurements)

DISPLAY FUNCTION INSTRUCTIONS

  1. Direct Current
  2. Alternating Current
  3. Negative DC Value
  4. Auto Power Off
  5. Data Hold
  6. Auto Range Active
  7. Maximum Display
  8. Minimum Display
  9. Relative Display
  10. Diode Test
  11. Continuity Test
  12.  Low Battery
  13. Temperature in Celsius
  14. Temperature in Fahrenheit
  15. Resistance
  16. Frequency

ROTARY SWITCH FUNCTION INSTRUCTIONS

  1. DC Voltage:<600V

  2. AC Voltage:<600V

  3. 1 DC Current (large):<10A

32 AC Current (Jarge): <10A

  1. 1 DC Current (Middle):<600mA

  2. 2 AC Current (Middle): <600 mA

2 AC Current (Small): <600uA

  1. Resistance: <60MQ

2 Continuity:<500

  1. Capacitance: <60mF

  2. Temperature: -20°C~1000°C/-4°F-1832°F F

ELECTRICAL SPECIFICATIONS

DC Voltage Measurement|
---|---
Range| Resolution| Accuracy
600mV| 0.1mV| ±(0.5% of reading +2 digits)
6V| 0.001V
60V| 0.01V
600V| 0.1V

  • Input impedance: 10M
  • Max. input voltage: 600V rms

AC Voltage Measurement|
---|---
Range| Resolution| Accuracy
600mV| 0.1mV| ±(1.0% of reading +5 digits)
6V| 0.001V
60V| 0.01V
600V| 0.1V

  • Input impedance: 10M
  • Max. input voltage: 600V rms
  • Frequency response: 40-400Hz, calibrated to rms of sine wave (average response)

DC Voltage Measurement|
---|---
Range| Resolution| Accuracy
600jiA| 0.10| ±(1.0% of reading +5 digits)
60000| 1pA
60mA| 0.01pA
600mA| 0.1pA
10A| 10mA| ±(2.0% of reading +10 digits)

  • Overload protection: μA/mA input: Fuse (F600mA/600V) 10A input: Fuse (F
  • Max. input current: μA/mA input: 600mA rms 10A input: 10A rms

Resistance Measurement| |
---|---|---
Range| Resolution| Accuracy
6000| 0.10| ±(0.8% of reading +5 digits)
6k0| 0.001k0
60k0| 0.01k0
600k0| 0.1k0
6M0| 0.001 MO
60M0| 0.01 MO| ±(2.0% of reading +5 digits)

  • Overload protection: μA/mA input: Fuse (F600mA/600V)
  • Frequency response: 40-400HZ, calibrated to rms of sine wave (average response) 10A input: Fuse (F10A/600V)
  • Max. input current: μA/mA input: 600mA rms 10A input: 10A rms

Continuity Test

Overload Protection| Open Circuit Voltage
600V RMS| Appx. 3.0V

  • Max. input voltage: 600V rms

Diode Test

Overload Protection Test Current Open Circuit Voltage
600V RMS Appx. lmA Appx. 3.0V DC
  • Max. input voltage: 600V rms
Capacitance Measurement Accuracy
Range Resolution
1 nF 0.001 nF ±(4.0% of reading +10 digits)
1 OnF 1 0.01 nF ±(3.0% of reading +10 digits)

100nF I 0.1 nF
1pF 1 nF
10pF| 10nF
100pF| 100nF
1mF ; 1pF
±(4.0% of reading +10 digits)
10mF 10pF
60mF 1 10pF

  • Max. input voltage: 600V rms
  • Max. input voltage: 600V rms
  • Signal input range: 20.2V AC rms (voltage input will increase as frequency increases)

Frequency (cont.) (AC voltage)|
---|---
Range| | Resolution| Accuracy
99.99Hz| | 0.01 Hz| ±(1.5% of reading +5 digits)
999 9Hz| 0.1Hz
9.999kHz|  0.001kHz

10kHz| 1 0.01kHz| Reference only

  • Input impedance: 10M
  • Max. input voltage: 600V rms
  • Signal input range: μA: ≥60µA rms mA: ≥6mA rms A: 20.6A rms (current input will increase as frequency increases)
  • Max. input current: 10A rms

|
---|---
|
Range| Resolution| Accuracy
-20°C~1000°C| 1°C| ±(2.O% of reading +3 digits)
-4°F~1832°F| 1°F

  • Overload protection: Fuse (F600mA/600V)