Labnet C1001 Mini Plate Spinner Instruction Manual
- June 13, 2024
- Labnet
Table of Contents
Labnet C1001 Mini Plate Spinner
Product Information
Capacity | Not specified |
---|---|
Dimensions | Not specified |
Weight | Not specified |
Power supply | Not specified |
Maximum speed | Not specified |
Maximum RCF | Not specified |
Temperature range | Not specified |
Electrical Protection class | Not specified |
Product Usage Instructions
Installation
- Place the Labnet MPS-1001 Mini Plate Spinner on a clean, flat surface with sufficient space around it for ventilation.
- Plug the power cord into the back of the unit, and connect the other end to the power socket.
- To turn on the unit, set the power switch to the On (I) position.
The Labnet MPS-1001 Mini Plate Spinner should be placed on a clean, flat surface with sufficient space around it for ventilation. Plug the power cord into the back of the unit, and connect the other end to the power socket. To turn on the unit, set the power switch to the On (I) position.
Operation
To avoid spillage of samples, only plates that have been heat or tape sealed should be placed in the centrifuge. Plates are loaded into the rotor in a vertical position.
-
Lift the clear centrifuge lid to gain access to the rotor. If the lid is locked, press the Open Door button to release the lid.
-
Carefully slide the first plate into position so that the sealed side faces the center of the rotor and the bottom of the plate faces out. Be sure that the plate is pushed all the way down into the slot.
-
Load the second plate in the same manner as the first, making sure that it is all the way down in the rotor slot.
-
Lift the clear centrifuge lid to gain access to the rotor. (If the lid is locked, press the Open Door button to release the lid).
-
Carefully, slide the first plate into position so that the sealed side faces the center of the rotor and the bottom of the plate faces out. Be sure that the plate is pushed all the way down into the slot.
-
Load the second plate in the same manner as the first, making sure that it is all the way down in the rotor slot. If spinning only one plate, it is extremely important to load a dummy plate of the same brand, style, and sample volumes (water can be used) as a balance.
-
The time setting range is from 0 to 99 minutes. Press the time increase/decrease button to adjust the time parameter. Select 0 to cancel the timing function, and press the START/STOP button to save the data
-
To begin a run, close the lid of the centrifuge. Press the START/STOP button to run. Allow the unit to run for 15-20 seconds. This is sufficient time to spin the samples down to the bottoms of the wells.
-
To stop, press the START/STOP button and the rotor will decelerate. Once the rotor has stopped, press the Open Door button to open the centrifuge lid and take out the sample plate(s).
Notes:
- A slight vibration may be felt when the centrifuge reaches full speed. Excessive vibration indicates that the rotor is not balanced. Do not run the centrifuge unbalanced. Always use a balance plate of the same brand, style, and sample volumes.
- The SHORT button is the electric mode button. Holding down the SHORT button will run the unit at the highest speed. After it has reached maximum speed, the unit will start counting up in seconds. To stop operation, simply release the SHORT button.
Cleaning
Be sure to disconnect the power cord before cleaning. After the rotor is stopped, clean the surface of the instrument with a dry cloth and 75% alcohol. Make sure that all parts of the instrument are completely dry before operating. Do not immerse the unit in liquid.
Limited Warranty
In the event this product fails within the specified period of time because of
a defect in material or workmanship, contact Corning Customer Service at:
USA/Canada 1.800.492.1110, outside the U.S. +1.978.442.2200, visit
www.corning.com/lifesciences, or
contact your local support office.
Corning Incorporated (Corning) warrants that this product will be free from
defects in material and workmanship for a period of one (1) year from date of
purchase. CORNING DISCLAIMS ALL OTHER WARRANTIES WHETHER EXPRESSED OR IMPLIED,
INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR OF FITNESS FOR A
PARTICULAR PURPOSE. Corning’s sole obligation shall be to repair or replace,
at its option, any product or part thereof that proves defective in material
or workmanship within the warranty period, provided the purchaser notifies
Corning of any such defect. Corning is not liable for any incidental or
consequential damages, commercial loss or any other damages from the use of
this product. This warranty is valid only if the product is used for its
intended purpose and within the guidelines specified in the supplied
instruction manual. This warranty does not cover damage caused by accident,
neglect, misuse, improper service, natural forces or other causes not arising
from defects in original material or workmanship. This warranty does not cover
motor brushes, fuses, light bulbs, batteries or damage to paint or finish.
Claims for transit damage should be filed with the transportation carrier. In
the event this product fails within the specified period of time because of a
defect in material or workmanship, contact Corning Customer Service at:
USA/Canada 1.800.492.1110, outside the U.S. +1.978.442.2200, visit
www.corning.com/lifesciences, or
contact your local support office. Corning’s Customer Service team will help
arrange local service where available or coordinate a return authorization
number and shipping instructions. Products received without proper
authorization will be returned. All items returned for service should be sent
postage prepaid in the original packaging or other suitable carton, padded to
avoid damage. Corning will not be responsible for damage incurred by improper
packaging. Corning may elect for onsite service for larger equipment. Some
states do not allow limitation on the length of implied warranties or the
exclusion or limitation of incidental or consequential damages. This warranty
gives you specific legal rights. You may have other rights which vary from
state to state. No individual may accept for, or on behalf of Corning, any
other obligation of liability, or extend the period of this warranty. For your
reference, make a note of the serial and model number, date of purchase, and
supplier here.
Introduction
The Labnet MPS-1001 Mini Plate Spinner is designed for quick spins of samples
in PCR plates. Primarily designed for use with 96- or 384-well plates and
small volumes, it will accommodate skirted, non-skirted and semi-skirted
styles. With the use of the adapter you can spin a half plate, 8 or 12 x 0.2
µL strips, and individual PCR tubes. No decontamination is required for the
equipment. Before using any cleaning or decontamination methods except those
recommended by the manufacturer, check with the manufacturer that the proposed
method will not damage the equipment.
Symbols and Conventions
- The electrical warning indicates the presence of a potential hazard which could result in electrical shock.
- CAUTION: This symbol refers you to important operating and maintenance (servicing) instructions within the product Instruction Manual. Failure to heed this information may present a risk of damage or injury to persons or equipment.
Safety Information
- Never use the centrifuge in any manner not specified in these instructions.
- Never put hands in the rotor area unless the rotor is completely stopped.
- Never move the centrifuge while the rotor is spinning.
- Never lean on or leave the centrifuge during operation.
- Never use solvents or flammables near this or other electrical equipment.
- Never centrifuge flammable, explosive or corrosive materials.
- Never centrifuge hazardous materials outside of a hood or proper containment facility.
- Never centrifuge in a strong ultraviolet environment.
- Never continue to operate a centrifuge that is vibrating excessively.
- Always place the centrifuge on a level, sturdy surface.
- Always load the rotor symmetrically. Each plate should be counterbalanced by another plate of the same brand, style, and volume. Always use plates that have been sealed in some manner to avoid spillage.
- Always load plates so that the sealed side faces the center of the rotor and the bottom of the plate faces out.
- Always be sure the plates are pushed all the way to the bottom of the rotor slot before operating the centrifuge.
- Always locate the centrifuge within easy access to an electrical outlet.
WARNING: This unit is intended for short run usage only, requiring 15-20 seconds to spin down samples.
Specifications
Capacity | 2 x 96-/384-well PCR plates |
---|---|
Dimensions | 7.5 x 8.6 x 7.3 in. (190 x 220 x 186 mm) |
Weight | 5.7 lbs. (2.6 kg) |
Power supply | 100 to 240V~, 50/60 Hz |
Maximum speed | 2,500 rpm ±10% |
Maximum RCF | 500 x g |
Temperature range | 5°C to 40°C (up to 80% RH at 31°C, decreasing |
linearly to 50% RH at 40°C)
Electrical| 100V, 50/60Hz, 0.6A|
120V, 50/60Hz, 0.6A
230V, 50/60Hz, 0.3A
Protection class| IP21
Changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
The Labnet MPS 1001 Mini Plate Spinner is designed to be safe when operated
under the following conditions:
- Indoor use
- Altitude up to 2,000 meters
- Pollution Degree 2
NOTE: This equipment has been tested and found to comply with the limits
for a Class A digital device, pursuant to Part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial environment. The
equipment generates, uses, and can radiate radio frequency energy and, if not
installed and used in accordance with the user manual, may cause harmful
interference to radio communications. Operation of equipment in a residential
area is likely to cause harmful interference in which case the user will be
required to correct the interference.
CAUTION: This equipment is not intended for use in residential
environments and may not provide adequate protection to radio reception in
such environments.
NOTE: The equipment should be used in a basic electromagnetic environment
only.
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme à la norme NMB-001 du Canada.
Equipment Disposal
According to Directive 2012/19/EU of the European Parliament and of the
Council of 4 July 2012 on waste electrical and electronic equipment (WEEE),
this product is marked with the crossed-out wheeled bin and must not be
disposed of with domestic waste. Consequently, the buyer shall follow the
instructions for reuse and recycling of waste electronic and electrical
equipment (WEEE) provided with the products and available
atwww.corning.com/weee.
To request certificates, please contact us at
www.labnetlink.com.
Warranty/Disclaimer: Unless otherwise specified, all products are for
research use or general laboratory use only. Not intended for use in
diagnostic or therapeutic procedures. Not for use in humans. These products
are not intended to mitigate the presence of microorganisms on surfaces or in
the environment, where such organisms can be deleterious to humans or the
environment. Corning Life Sciences makes no claims regarding the performance
of these products for clinical or diagnostic applications. For a listing of
US medical devices, regulatory classifications or specific information on
claims, visit www.corning.com/resources.
Corning Incorporated
Life Sciences
www.corning.com/lifesciences
NORTH AMERICA
t 800.492.1110
t 978.442.2200
ASIA/PACIFIC
Australia/New Zealand t 61 427286832
Chinese Mainland
t 86 21 3338 4338
India
91 124 4604000
Japan
81 3-3586 1996
Korea
82 2-796-9500
Singapore
65 6572-9740
Taiwan
t 886 2-2716-0338
EUROPE
CSEurope@corning.com
LATIN AMERICA
grupoLA@corning.com
Brazil
55 (11) 3089-7400
Mexico
t (52-81) 8158-8400
For a listing of trademarks, visit
www.corning.com/clstrademarks. All
other trademarks are the property of their respective owners. © 2023 Corning
Incorporated. All rights reserved. 8/23 CLSLN-AN-1040DOC REV1
www.labnetlink.com