IDEXX SNAP Pro Analyser User Manual
- June 13, 2024
- IDEXX
Table of Contents
*Cleaning the SNAP Pro Analyser
User Manual**
SNAP Pro Analyser
To clean the internal components (perform weekly or as needed)
- Switch off the analyser and disconnect the power and/or Ethernet cord.
- On the bottom of the switched-off analyser, use a flat-tip screw driver or coin to turn the 2 silver fasteners counterclockwise a quarter turn.
- Remove the base from the analyser.
- Clean the entire base using a lint-free wipe or cotton swab dampened with 70% isopropyl alcohol, mild liquid soap and water, or a 10% bleach solution.
- Clean the 3 white reflectors, the axle, and the activator using a new lint-free wipe or cotton swab dampened with 70% isopropyl alcohol, mild liquid soap and water, or a 10% bleach solution. DO NOT clean the lens.
- After the base has dried, reattach it to the analyser and ensure it is locked into place.
- Turn the silver fasteners clockwise a quarter turn to lock the base into place.
- Ensure the base and silver fasteners are flush with the analyser and locked into place. Not locking the base and fasteners properly can impact analyser performance.
To clean the case and screen (perform as needed)
Clean the case and screen using a lint-free cloth dampened with 70% isopropyl
alcohol, mild liquid soap and water, or a 10% bleach solution. DO NOT wipe the vents underneath the power button
or the USB/power/Ethernet ports.When cleaning, do not expose the analyser to:
Organic solvents
Ammonia-based cleaners
Ink markers
Sprays containing volatile liquids
Insecticides
Disinfectant
Polish
Air freshener