IDEXX Coag Dx Prothrombin Time PT
Product Information
The IDEXX Coag Dx Prothrombin Time (PT) is a unitized coagulation test designed for in vitro use. It is used to perform a quantitative, one-stage prothrombin time. The test is performed using fresh whole blood from dogs, cats, or horses. The PT test is compatible with the Coag Dx Analyzer and the SCA2000* Coagulation Analyzer. This test is intended for veterinary diagnostic use only.
- Summary and Explanation
- The PT test measures the extrinsic and common coagulation pathways.
- It provides quick results, typically in less than 2 minutes, as no sample processing is required.
- Patient-side testing is particularly valuable during procedures and therapeutic interventions, hemostasis assessment before or after blood transfusions, and diagnosis of anticoagulant rodenticide toxicity.
- Principle of Operation
- The clot detection mechanism of the PT test consists of a series of LED optical detectors aligned with the test channel of the cartridge.
- The movement of the blood sample between the detectors is measured. As clot formation begins, blood flow is impeded, causing the movement to slow down.
- The analyzer recognizes the clot endpoint when the movement decreases below a predetermined rate. The PT whole blood clotting time is reported in whole seconds.
- Reagents
- Each box of PT test cartridges contains a self-contained disposable test chamber preloaded with a dried preparation of thromboplastin, stabilizers, and buffers.
- Each cartridge is individually packaged in a pouch with a lot-specific expiration date. Used test cartridges should be handled with care and disposed of properly due to potential infectiousness.
- Storage and Stability
- The PT test cartridges should be stored at room temperature. The maximum storage duration is 4 weeks, but it must not exceed the marked expiration date. If stored at room temperature, the outer box should be marked with the new expiration date.
Product Usage Instructions
Sample Collection
To ensure accurate coagulation testing and prevent contamination, blood
samples should be collected using the following guidelines:
- Do not collect samples until the analyzer indicates “Add Sample and Press Start”.
- If using a syringe sample from an indwelling line, the amount of blood required to adequately flush the line until it is free of contaminants depends on the volume of solution in the line. Longer lines may require greater volumes to clear them.
- Poorly collected blood samples with visible clotting or debris accumulation should be discarded, and a fresh sample should be collected.
Intended Use
- The PT is a unitized coagulation test intended for in vitro use in performing a quantitative, one-stage prothrombin time.
- The PT test is performed using fresh canine, feline, or equine whole blood.
- This test is to be used with the Coag Dx Analyzer and is also compatible with the SCA2000 Coagulation Analyzer.
- This test is for veterinary diagnostic use only.
Summary and explanation
- The PT, a test of the extrinsic and common coagulation pathways, can be performed usingfresh whole blood at the patient’s side.
- Since no sample processing is required, PT results are obtained in less than 2 minutes.
- Patient-side testing is especially valuable during procedures and therapeutic interventions, for hemostasis assessment before or after blood transfusions, and for diagnosis of anticoagulant rodenticide toxicity.
- Traditionally, the events leading to the formation of a fibrin clot have been simplified in coagulation theory into two coagulation pathways: the intrinsic and extrinsic, both leading to the common pathway and the formation of a stable fibrin clot.
- The PT is a measure of the extrinsic and common coagulation pathways.
Principle of operation
- The analyzer utilizes a mechanical endpoint clotting mechanism in which clot formation occurs within the disposable PT cartridge.
- Following whole blood sample introduction, the analyzer precisely measures 15 μL of blood and automatically moves it into the test channel within the PT cartridge.
- The remainder of the blood sample, not needed for testing, is automatically drawn into the waste channel of the cartridge.
- Sample/reagent mixing and test initiation are also performed automatically, requiring no operator interaction. After mixing with the reagent, the sample is moved back and forth within the test channel and observed for clot formation.
- The clot detection mechanism consists of a series of LED optical detectors aligned with the test channel of the cartridge. The speed at which the blood sample moves between the detectors is measured.
- As clot formation begins, blood flow is impeded and the movement slows. The analyzer recognizes that the clot endpoint has been achieved when the movement decreases below a predetermined rate.
- The PT whole blood clotting time is reported in whole seconds.
Reagents
Each box of PT test cartridges contains the following:
- 10 pouches, each containing one PT test cartridge and one desiccant
- The PT test cartridge is a self-contained disposable test chamber preloaded with a dried preparation of thromboplastin, stabilizers, and buffers. Each cartridge is individually packaged in a pouch. Cartridge pouches are stamped with a lot-specific expiration date.
- Caution: All used test cartridges should be considered as potentially infectious, handled with care, and disposed of properly.
Storage and stability
When refrigerated (36˚F–46˚F/2°C–8°C), the foil-pouched PT cartridges are
stable until the marked expiration date. Room temperaturestorage
(59˚F–86˚F/15°C–30°C) is optional for unopened, pouched cartridges. PT
cartridges should not be exposed to temperatures in excess of 98.6˚F/37°C.
Note: Room temperature redating is to a maximum of 4 weeks but must
never exceed themarked expiration date. Redating is necessary if stored at
room temperature. Mark the outer box with the new expiration date when
cartridges are stored at room temperature.
Sample collection
Blood samples to be used for coagulation testing must be collected in the
following manner to prevent contamination with tissue thromboplastin or
indwelling intravenous (IV) solutions that interfere with the coagulation
assays. Poorly collected blood samples with visible clotting or debris
accumulation must be discarded and a fresh sample collected. Patient
excitement should be minimized as this can increase platelet count, platelet
aggregation, and the levels of von Willebrand factor (vWF), fibrinogen, and
factors V and VIII. Prolonged venous stasis and excessive probing for the
vessel should be avoided. Use of the cephalic or saphenous veins are advised
as bleeding is easier to control from these sites.1 If a syringe is used, it
should have a 23-gauge needle or larger. Use of excessive force when expelling
the blood specimen through the needle may cause hemolysis.
Note: Samples must not be collected until the analyzer indicates “Add
Sample” and “Press Start.”
Syringe sample, from indwelling line
Note: The amount of blood required to adequately flush the line until it
is free of contaminants is dependent on the amount of solution contained
within the line. Greater volumes will be required to clear longer lines. Using
a tuberculin (1 mL) or 3 mL syringe, collect a minimum of 0.2 mL of blood from
a previously flushed access port. Do not allow bubbles to form in the syringe.
Syringe sample, from a venipuncture
- Prepare the venipuncture site by cleansing with alcohol and allowing to air-dry completely.
- Obtain a minimum of 0.2 mL of blood.
Operating instructions
Before performing any assay, refer to the IDEXX Coag Dx* Analyzer Operator’s Guide for detailed operating instructions.
Material provided
- PT test cartridges
Material required (not provided)
- Coagulation analyzer
- Plastic syringes
- 23-gauge needle or larger (for syringe sampling)
Note: PT test cartridges must be at room temperature prior to use. Once removed from the refrigerator, this may take up to 60 minutes. For best results, the pouch should be opened immediately prior to testing.
Test procedure
Refer to the IDEXX Coag Dx* Analyzer Operator’s Guide if any fault message should appear during this procedure.
- Insert a test cartridge into the cartridge opening of the analyzer. The cartridge must be inserted with the blood reservoir facing up. The analyzer will automatically identify the test cartridge and display the test type.
- During the warming stage, observe the display for fault messages. The analyzer emits an audible tone when it is ready and alternately displays the “Add Sample” and “Press Start” messages. The analyzer remains in the “Ready” mode for 5 minutes before a “START…TIMED OUT” message displays. If this occurs, a new test cartridge must be placed in the analyzer.
- Obtain the blood sample. (See the Sample Collection section for more information.)
- Immediately dispense one drop of blood into the center sample well of the test cartridge; fill from the bottom of the well up. This may be done either with or without a needle. A sufficient quantity of blood must be addeddirectly to the center sample well to fill it flush to the top. Should a large drop of blood extend above the center sample well, push it into the outer sample well.
- Press the Start key. A single beep signals the start of the test. The analyzer automatically mixes the sample with the reagent and detects clot.
- The analyzer emits a single beep when the test is complete. The test result, in seconds for whole blood, remains on the screen until the test cartridge isremoved from the analyzer and for 120 seconds following its removal.
Operating precautions
Do NOT use cartridges that are past their marked expiration date or that have been improperly stored. Do NOT force a cartridge into the analyzer. If resistance to insertion is encountered, gently remove the cartridge and examine the cartridge slot. Remove any obstruction before attempting further use of the analyzer. Sample collection and handling for all coagulation testing requires careful adherence to guidelines. As with all diagnostic tests, test results should be scrutinized in light of a specific patient’s condition and anticoagulant therapy. Any results exhibiting inconsistency with the patient’s clinical status should be repeated or supplemented with additional test data. The PT is affected by poor technique, including blood collection and the transfer of blood to thesample well. The accuracy of the test is largely dependent upon the quality of the blood sample, which may be affected by the following:
- Foaming of the sample
- Hemolysis of the sample
- Clotted or partially clotted blood
Performance characteristics
Reference intervals
Whole blood samples were obtained from normal healthy animals and tested with
the PT test. Reference intervals were developed as follows:
Species | Canine | Feline | Equine |
---|---|---|---|
Reference interval (seconds) | 11–14 | 13–22 | 15–18 |
Limitations
Whole blood PT test results under 7 seconds or over 100 seconds are not reported. Instead, either an “Out of Range-Lo” or “Out of Range-Hi” message is displayed.
Reference
- Green RA, Thomas JS. Hemostatic disorders: coagulopathies and thrombosis. In: Ettinger SJ, Feldman EC, eds. Textbook of Veterinary Internal Medicine. 4th ed. Philadelphia, PA: WB Saunders; 1995:1946–1963.
Suggested reading
- Evatt BL, Brogan D, Triplett DA, Koepke J, Waters
- G. Effect of thromboplastin and instrumentation on the prothrombin time test.
- Clin Lab Haematol. 1981;3(4):331–342.
- Gralnick HR, Kessler CM, Palmer R. The prothrombin time: variables affecting results.
- In: Triplett DA, ed. Standardization of Coagulation
- Assays: An Overview. Skokie, IL: College of
- American Pathologists; 1982.
- Tseng LW, Hughes D, Giger U. Evaluation of point-of-care coagulation analyzer for measurement of prothrombin time, activated partial thromboplastin time, and activated clotting time in dogs.
- Am J Vet Res. 2001;62(9): 1455–1460.
© 2018 IDEXX Laboratories, Inc. All rights reserved. Coag Dx is a trademark or registered trademark of IDEXX Laboratories, Inc. or its affiliates in the United States and/or other countries. SCA2000 is a trademark of Symbiotics, Inc.
- One IDEXX Drive
- Westbrook, Maine 04092
- idexx.com.
- 06-12895-03