IDEXX SNAP Pro Analyzer User Guide
- June 13, 2024
- IDEXX
Table of Contents
**IDEXX SNAP Pro Analyzer User Guide
**
Setting up the analyzer
-
Open the battery compartment on the back of the analyzer and then place two batteries (provided), negative end first, into the compartment.
-
Connect the appropriate adapter for your 2 region to the power cord using these steps: a. Remove the protective insert while sliding the locking key toward the cord
b. Insert the adapter top-side first and then snap into place.
North America Japan
Australia
United Kingdom
Europe South America
-
Connect the power cord to the analyzer and an appropriate electrical outlet. The status light should appear steady yellow. Note: It will take up to 8 hours to fully charge the analyzer (the status light will turn green when fully charged). The analyzer may be used while charging.
-
If you are planning to connect the analyzer wirelessly to an IDEXX VetLab Station, proceed to step 5.
Note: IDEXX VetLab wireless router required. OR
If you are connecting the analyzer to an IDEXX VetLab Station using an IDEXX VetLab* wired router, connect the analyzer to a numbered port on the router using a CAT5 cable (provided). -
Press the gray Power button at the top of the back of the SNAP Pro Analyzer to power it on. After the startup screen, follow the on-screen instructions.
-
When prompted, do the following:
a. Select your location and then tap the Next icon.
b. If you are connecting the analyzer to an IDEXX VetLab Station, tap Yes, select the IDEXX network from the list, and then tap the
Next icon. If the IDEXX network is not listed, tap the Refresh icon.
OR
If you are not connecting the analyzer to an IDEXX VetLab Station, tap No and then tap the Next icon. When prompted, confirm that you are not connecting the analyzer to an IDEXX VetLab Station.
c. Select the desired date format, enter today’s date using the numeric keypad, and then tap the Next icon.
d. Select the desired time format, enter the current time using the numeric keypad, and then tap the Next icon.
e. If you want the analyzer to beep when there is an error, warning, or other information displayed, select the Sound icon, select the check boxes next to the desired options, and then select if you would like the sound to be Soft or Loud. Then tap the Check Mark icon.
OR
If you do not want the analyzer to beep, select the Mute icon and then tap the Check Mark icon.
Note: Critical errors will always beep.
f. When prompted, tap OK to restart the analyzer.
Need help? Visit idexx.com/prohelpfor helpful installation videos and support documentation.
Running a SNAP test
*With an IDEXX VetLab Station**
-
Tap Analyze Sample on the IDEXX VetLab Station Home screen to enter the patient information and start a run.
OR
If the IDEXX VetLab Station is integrated with a practice management system, proceed to step 2 or select the patient from the Pending or Census list on the IDEXX VetLab Station Home screen.
Note: All patients that appear in the Pending list on the IDEXX VetLab Station also appear on the SNAP Pro* Home screen.
OR
Tap New Patient on the SNAP Pro Home screen to enter the patient’s name and client ID. When you have finished entering the information, tap the Check Mark icon and proceed to step 3. -
Once the patient’s name appears in the list on the SNAP Pro Home screen, tap the patient’s name and then tap the Check Mark icon.
-
Prepare the SNAP* test, but do not activate the test.
-
Gently slide the SNAP test into the port at the bottom of the analyzer (well end first) until it clicks into position. When analysis is complete, results are shown on the results screen and are sent to the IDEXX VetLab Station automatically.
Tip: The color of the status light indicates the results of the run: Green—all results are negative Blue—at least one result is positive Yellow—visual confirmation needed (use the instructions on the next page)
Without an IDEXX VetLab Station
- Tap New Patient on the SNAP Pro Home screen to enter the patient’s name and client ID. When you have finished entering the information, tap the Check Mark icon.
- Prepare the SNAP test, but do not activate the test.
- Gently slide the SNAP test into the port at the bottom of the analyzer (well end first) until it clicks into position (see illustration on left).
When you need to confirm results
If there are very light spots or unexpected color development (such as debris or a blue tint) in the results window, you may be prompted to confirm some or all of a test’s results. Results that need to be confirmed are highlighted yellow.
-
If you believe the highlighted result is positive, tap the plus icon.
OR
If you believe the highlighted result to be negative, tap the minus icon. -
When you have finished confirming the results, tap the Check Mark icon. The results will then be saved on the analyzer and sent to the IDEXX VetLab Station (if applicable).
Note: To cancel the test and return to the Home screen, tap the close icon .
Cleaning the analyzer
To clean the case and screen (perform as needed)
Clean the case and screen using a lint-free cloth dampened with 70% isopropyl
alcohol, mild liquid soap and water, or a 10% bleach solution. DO NOT wipe the
vents underneath the power button or the USB/power/Ethernet ports.
To clean the internal components (perform weekly or as needed)
-
Power off the analyzer and disconnect the power and/or Ethernet cord
-
Use a flat-tip screwdriver or coin to turn the 2 silver fasteners counterclockwise a quarter turn.
-
Remove the base from the analyzer.
-
Clean the entire base using a lint-free wipe or cotton swab dampened with 70% isopropyl alcohol, mild liquid soap and water, or a 10% bleach solution.
-
Clean the 3 white reflectors, the axle, and the activator using a new lint-free wipe or cotton swab dampened with 70% isopropyl alcohol, mild liquid soap and water, or a 10% bleach solution. DO NOT clean the lens.
-
After the base has dried, reattach it to the analyzer and ensure it is locked into place.
-
Turn the silver fasteners clockwise a quarter turn to lock the base into place.
-
Ensure the base and silver fasteners are flush with the analyzer and locked into place. Not locking the base and fasteners properly can impact analyzer performance.
Interpreting SNAP Pro icons
*SNAP Pro Icons**
Icon | Description |
---|---|
Wireless signal status (5 gray bars is the best signal; 5 red bars |
indicates no signal)
| Connected and communicating with the IDEXX VetLab Station via a wired
connection (a red icon indicates the analyzer is offline)
| Battery at 100% power
| Battery at 75% power
| Battery at 50% power
| Battery at 25% power
| Battery is critically low
| Battery is empty
| Battery is charging
| A software upgrade is available (SmartService Solutions customers only)
| Warning notification (a gray icon indicates the warning has been viewed and
is unresolved)
| Error notification (a gray icon indicates the error has been viewed and is
unresolved)
| Lost communication with IDEXX VetLab Station
Technical specifications and safety information
External USB power supply adapter
- Input: 100–240 V AC, 47–63 Hz, 0.3 A
- Output: 5 V DC, 1 A
Consumption
- Operating: Maximum 5 W
Batteries
Main batteries
- Part number: 79-91922-01
- Output: 3.7 V DC, 2,600 mAh
Watch battery
- Part number: BR1225
- Output: 3 V DC, 48 mAh
Operating conditions
- Operating temperature: 15°C–30°C (59°F–86°F)
Safety precautions
- The SNAP Pro* Analyzer contains no user-serviceable components. Do not disassemble. Call IDEXX Technical Support for any service issues.
- Do not insert anything in the port except SNAP* tests approved for use with the SNAP Pro Analyzer.
- Protect the analyzer from sources of heat or flames.
- Use only the supplied power adapter.
- Use only IDEXX-supplied batteries.
- Dispose of used batteries only in accordance with your local ordinances. These batteries contain lithium. Recycling is required in the European Union and is strongly encouraged in all other countries.
- If the equipment is used in a manner other than specified, the protection provided by the analyzer may be impaired.
- The SNAP Pro Analyzer should be placed on a flat surface during operation.
Notice: To satisfy Industry Canada (IC) radio frequency (RF) exposure requirements for mobile and base station transmission analyzers, a separation distance of 20 cm or more should be maintained between the antenna of this analyzer and persons during operation. To ensure compliance, operation at closer than this distance is not recommended. The antenna(s) used for this transmitter must not be colocated or operating in conjunction with any other antenna or transmitter.
Authorized representative in the European Community
Manufacturer’s declaration that product meets applicable European Community directives
Product complies with applicable Electromagnetic Compatibility (EMC) standard for Australia
This symbol is placed in accordance with the European Union Directive 2002/96 on the Waste Electrical and Electronic Equipment (the WEEE Directive). If disposed of within the European Union, this product should be treated and recycled in accordance with the laws of your jurisdiction implementing the WEEE Directive. For additional information, visit idexx.com/weee.
Temperature limitation
Consult instructions for use
Recyclable
References
- SNAP Pro Analyzer Support and Resources - IDEXX US
- WEEE Compliance in Europe Recycling Information for Customers and Recyclers - IDEXX UK