HT INSTRUMENTS HT4011 AC Clamp Meter User Manual
- June 4, 2024
- HT INSTRUMENTS
Table of Contents
HT INSTRUMENTS HT4011 AC Clamp Meter User Manual
PRECAUTIONS AND SAFETY MEASURES
The instrument has been designed in compliance with directive IEC/EN61010-1
relevant to electronic measuring instruments. For your safety and in order to
prevent damaging the instrument, please carefully follow the procedures
described in this manual and read all notes preceded by the symbol with the
utmost attention.
Before and after carrying out the measurements, carefully observe the
following instructions.
- Do not carry out any voltage or current measurement in humid environments.
- Do not carry out any measurements in case gas, explosive materials or flammables are present, or in dusty environments.
- Avoid any contact with the circuit being measured if no measurements are being carried out.
- Avoid any contact with exposed metal parts, with unused measuring probes, circuits, etc.
- Do not carry out any measurement in case you find anomalies in the instrument such as deformation, breaks, substance leaks, absence of display on the screen, etc.
- Pay special attention when measuring voltages higher than 20V, since a risk of electrical shock exists.
In this manual, and on the instrument, the following symbols are used:
Warning: observe the instructions given in this manual; improper use
could damage the instrument or its components.
High voltage danger: electrical shock hazard.
Double-insulated meter.
AC voltage or current
DC voltage
Connection to earth
PRELIMINARY INSTRUCTIONS
- This instrument has been designed for use in environments of pollution degree 2.
- It can be used for CURRENT and VOLTAGE measurements on installations with measurement category CAT III 600V. For a definition of measurement categories, see
- We recommend following the normal safety rules devised to protect the user against dangerous currents and the instrument against incorrect use.
- Only the leads supplied with the instrument guarantee compliance with the safety standards. They must be in good conditions and replaced with identical models, when necessary.
- Do not test circuits exceeding the specified current and voltage limits.
- Check that the battery is correctly inserted
- Before connecting the test leads to the circuit to be tested, make sure that the switch is correctly set.
- Make sure that the LCD display and the switch indicate the same function.
DURING USE
Please carefully read the following recommendations and instructions:
CAUTION
Failure to comply with the Caution notes may damage the instrument and/or its
components or be a source of danger for the operator.
- Before activating the switch, remove the conductor from the clamp jaw or disconnect the test leads from the circuit under test.
- When the instrument is connected to the circuit under test, do not touch any unused terminal.
- Avoid measuring resistance if external voltages are present. Even if the instrument is protected, excessive voltage could cause a malfunction of the clamp.
- When measuring current with the clamp jaws, first remove the test leads from the instruments input jacks.
- During current measurement, any other current near the clamp may affect measurement precision.
- When measuring current, always put the conductor as near as possible to the middle of the clamp jaw, to obtain the most accurate reading.
- While measuring, if the value or the sign of the quantity being measured remain unchanged, check if the HOLD function is enabled.
AFTER USE
- When measurement is complete, switch OFF the instrument.
- If you expect not to use the instrument for a long period, remove the battery.
EFINITION OF MEASUREMENT (OVERVOLTAGE) CATEGORY
Standard “IEC/EN61010-1: Safety requirements for electrical equipment for
measurement, control and laboratory use, Part 1: General requirements” defines
what measurement category is. § 6.7.4: Measured circuits, reads: (OMISSIS)
Circuits are divided into the following measurement categories:
-
Measurement category IV is for measurements performed at the source of the lowvoltage installation.
Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units. -
Measurement category III is for measurements performed on installations inside buildings.
Examples are measurements on distribution boards, circuit breakers, wiring, including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and equipment for industrial use and some other equipment, for example, stationary motors with permanent connection to fixed installation. -
Measurement category II is for measurements performed on circuits directly connected to the low-voltage installation.
Examples are measurements on household appliances and similar equipment. -
Measurement category I is for measurements performed on circuits not directly connected to MAINS.
Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS-derived circuits. In the latter case, transient stresses are variable; for that reason, the standard requires that the transient withstand capability of the equipment is made known to the user.
GENERAL DESCRIPTION
The instrument carries out the following measurements:
- DC and AC voltage up to 600V
- AC current up to 400A
- Resistance and continuity test with buzzer
- Capacitance
- Frequency with leads
- Duty Cycle
- Diode test
- Temperature with K probe
- Detection of presence of AC voltage with and without contact with in-built sensor.
Each of these functions may be selected through a 8-position rotary switch, including the OFF position and a key for enabling the HOLD function. The instrument also has a “MODE”, “Hz%” and “REL” key. For their use, please refer to § 4.2. The selected quantity appears on the LCD display with the indication of the measuring unit and of the enabled functions.
MEASURING AVERAGE VALUES AND TRMS VALUES
Measuring instruments of alternating quantities are divided into two big
families:
- AVERAGE-VALUE meters: instruments measuring the value of the sole wave at fundamental frequency (50 or 60 Hz).
- TRMS (True Root Mean Square) VALUE meters: instruments measuring the TRMSvalue of the quantity being tested.
With a perfectly sinusoidal wave, the two families of instruments provide
identical results.
With distorted waves, instead, the readings shall differ. Average-value meters
provide the RMS value of the sole fundamental wave; TRSM meters, instead,
provide the RMS value of the whole wave, including harmonics (within the
instruments bandwidth). Therefore, by measuring the same quantity with
instruments from both families, the values obtained are identical only if the
wave is perfectly sinusoidal. In case it is distorted, TRMS meters shall
provide higher values than the values read by average-value meters.
DEFINITION OF TRUE ROOT MEAN SQUARE VALUE AND CREST FACTOR
The root mean square value of current is defined as follows: “In a time equal
to a period, an alternating current with a root mean square value of 1A
intensity, circulating on a resistor, dissipates the same energy that, during
the same time, would be dissipated by a direct current with an intensity of
1A”. This definition results in the numeric expression:
The root mean square value is indicated with the acronym RMS.
RMS value: This value changes with the signal waveform, for a purely sinusoidal wave it is 2 =1.41. In case of distortion, the Crest Factor takes higher values as wave distortion increases.
PREPARATION FOR USE
INITIAL CHECKS
Before shipping, the instrument has been checked from an electric as well as
mechanical point of view.
All possible precautions have been taken so that the instrument is delivered
undamaged.
However, we recommend generally checking the instrument in order to detect possible damage suffered during transport. In case anomalies are found, immediately contact the forwarding agent.
We also recommend checking that the packaging contains all components
indicated in 6.3. In case of discrepancy, please contact the Dealer.
In case the instrument should be returned, please follow the instructions
given in § 7.
INSTRUMENT POWER SUPPLY
The instrument is supplied with two 1.5V AAA LR03 batteries, included in the
package.
The “ “ symbol appears when the battery is nearly flat. Replace the battery by
following the instructions given in § 5.2.
The instrument is also equipped with an Auto Power OFF function (which cannot be disabled) which automatically switches off the instrument approx. 30 minutes after the last operation was carried out.
CALIBRATION
The instrument has the technical specifications described in this manual. The
instruments performance is guaranteed for 12 months.
STORAGE
In order to guarantee precise measurement, after a long storage time under
extreme environmental conditions, wait for the instrument to come back to
normal condition.
OPERATING INSTRUCTIONS
INSTRUMENT DESCRIPTION
Description of the controls
Fig. 1: Instrument description
CAPTION :
- Inductive clamp jaw
- AC voltage detector
- Rotary selector switch
- Jaw trigger
- HOLD key
- MODE key
- REL key
- Hz% key
- LCD display
- Input terminal COM
- Input terminal
V CAPHz%Temp
DESCRIPTION OF FUNCTION KEYS
HOLD key
Short pressing the “HOLD” key activates the function Data HOLD, i.e. the value
of the measured quantity is frozen. The message “HOLD” appears on the display.
This operating mode is disabled when the “HOLD” key is pressed again or the
switch is operated.
REL key
With the instrument’s switch set to positions , , and AC current, this key
allows zeroing the displayed value and carrying out a relative measurement of
the quantity being tested. When pressing the REL key for the first time, the
value of the quantity being tested is saved as offset for the following
measurements. The message “REL” appears on the display. The instrument shows
the relative value obtained as current value – offset. This function is not
active in resistance measurements, continuity, temperature and capacitance
tests, duty cycle tests and diode test. Press the REL key again or turn the
rotary switch to exit the function.
Hz% key
With the instrument’s rotary switch set to positions , Hz, by pressing
the Hz% key you can switch to frequency (Hz) or duty cycle measurement (%).
MODE key
The MODE key is used to select resistance measurement, continuity test with
buzzer, and diode test with the instrument’s rotary switch set to position and
to select AC and DC voltage measurement with the switch set to position , ,
Hz.
DESCRIPTION OF ROTARY SWITCH FUNCTIONS
DC voltage measurement
CAUTION
The maximum input DC voltage is 600Vrms. Do not measure voltages exceeding the
limits given in this manual. Exceeding these limits could result in electrical
shocks to the user and damage to the instrument.
Fig. 2: Use of the clamp for DC voltage measurement
- Select the position.
- Insert the red cable into input terminal V CAPHz%Temp and the black cable into input terminal COM (Fig. 2).
- Position the test leads in the desired points of the circuit to be measured. The display shows the value of voltage.
- If the symbol “O.L” is displayed, this indicates overload status.
- To use the HOLD and REL functions, please refer to § 4.2.
CAUTION
- Due to the high input impedance, it may take some time for the instrument to zero the display.
- The changing value shown on the display with open input terminals is not to be considered a problem of the instrument and these values are not added by the instrument while carrying out a real measurement.
AC voltage measurement
CAUTION
The maximum input AC voltage is 600V. Do not measure voltages exceeding the
limits given in this manual. Exceeding these limits could result in electrical
shocks to the user and damage to the instrument.
Fig. 3: Use of the clamp for AC voltage measurement
- Switch on the instrument in any position of selector, near it an AC source and look for the red LED at the clamp’s base (see Fig. 1 – Part 2) to turn on. This indicates that the instrument has detected the AC source’s presence
- Select the position Hz.
- Insert the red cable into input terminal V CAPHz%Temp and the black cable into input terminal COM (Fig. 3).
- Position the test leads in the desired points of the circuit to be measured. The display shows the value of voltage.
- If the symbol “O.L” is displayed, this indicates overload status.
- To use the HOLD and REL functions, please refer to § 4.2.
CAUTION
- Due to the high input impedance, it may take some time for the instrument to zero the display.
- The changing value shown on the display with open input terminals is not to be considered a problem of the instrument and these values are not added by the instrument while carrying out a real measurement.
Frequency and Duty Cycle measurement
CAUTION
- When measuring frequency with leads, the maximum input AC voltage is 600Vrms. Do not measure voltages exceeding the limits given in this manual. Exceeding these limits could result in electrical shocks to the user and damage to the instrument.
- When measuring frequency with the clamp, make sure that all the instrument’s input terminals are disconnected.
Fig. 4: Use of the clamp for frequency measurement and duty cycle
- Select position Hz for measuring frequency with leads.
- Press the Hz% key cyclically until the symbol “Hz” is displayed for measuring frequency or the symbol “%” for duty cycle measurement.
- Insert the red cable into input terminal V CAPHz%Temp and the black cable into input terminal COM (Fig. 4 – left side) for measuring frequency with leads. The value of frequency (Hz) or of duty cycle (%) is shown on the display.
- If the symbol “O.L” is displayed, this indicates overload status.
- To use the HOLD function, please refer to § 4.2.
Resistance measurement
CAUTION
Before attempting any resistance measurement, remove power from the circuit
under test and discharge all capacitors, if present.
Fig. 5: Use of the clamp for Resistance measurement
- Select the position
- Insert the red cable into input terminal V CAPHz%Temp and the black cable into input terminal COM.
- Position the test leads in the desired points of the circuit to be measured (Fig. 5). The display shows the value of resistance.
- If the symbol “O.L” is displayed, this indicates overload status.
- To use the HOLD function, please refer to § 4.2.
Capacitance measurement
CAUTION
Before carrying out capacitance measurements on circuits or capacitors, cut
off power supply from the circuit being tested and let all capacitance in it
be discharged.
Fig. 6: Use of the clamp for capacitance measurement
- Select the CAP position.
- Press the MODE key cyclically until the symbol “nF” is displayed.
- Insert the red cable into input terminal V CAPHz%Temp and the black cable into input terminal COM.
- Position the test leads in the desired points of the circuit to be measured (Fig. 6). The display shows the value of capacitance. When measuring capacitance, the analogue analogue graphic bar is deactivated.
- If the symbol “O.L” is displayed, this indicates overload status.
- To use the HOLD and REL functions, please refer to § 4.2.
Continuity test and diode test
CAUTION
Before attempting any resistance measurement, remove power from the circuit
under test and discharge all capacitors, if present.
Fig. 7: Use of the clamp for continuity test and diode test
- Select the position
- Press the MODE key cyclically until the symbol “ “ is displayed to activate continuity test.
- Insert the red cable into input terminal V CAPHz%Temp and the black cable into input terminal COM and carry out the continuity test of the object to be measured (see Fig. 7– left side). An buzzer sounds when the measured value of resistance is lower than 30
- Press the MODE key to select diode test. The symbol “ “ appears on the display.
- Connect the red lead to the anode of the diode and the black lead to the cathode in case direct polarization measurement is carried out (see Fig. 7 – right side). Invert the position of the leads in case reverse polarization measurement is carried out.
- Values on the display between 0.4V and 0.7V (direct) and “OL” (reverse) indicate correct connection. A value “0mV” indicates that the device is short-circuited, while “OL” in both directions indicated an interrupted device.
Temperature measurement in °C and °F
CAUTION
Do not put the temperature probe into contact with live surfaces. Voltages
higher than 30Vrms or 60VDC imply electrical shock hazard.
Fig. 8: Use of the clamp for temperature measurement
- Select the Temp°C or Temp°F position.
- Insert the K wire probe provided into input terminal V CAPHz%Temp and COM by means of the relevant adapter, respecting the polarity indicated in Fig. 8. The display shows the value of temperature.
- To use the HOLD function, please refer to § 4.2.
AC current measurement
CAUTION
Before attempting any measurement disconnect all the test leads from the
circuit under test and from the meter’s input terminals.
Fig. 9: Use of the clamp for AC current measurement
-
Select position 40A or 400A
-
Insert the cable in the middle of the clamp jaws, in order to obtain accurate measures.
The display shows the value of AC current. -
If the symbol “O.L” is displayed, this indicates overload status. In this case, position the rotary switch to a higher measuring range.
-
To use the HOLD and PEAK functions, please refer to § 4.2.
CAUTION
A possible value displayed with the instrument not in measuring mode is not to
be considered a problem of the instrument and these values are not added by
the instrument while carrying out a real measurement.
MAINTENANCE
GENERAL INFORMATION
- The instrument you purchased is a precision instrument. While using and storing the instrument, carefully observe the recommendations listed in this manual in order to prevent possible damage or danger during use.
- Do not use the instrument in environments with high humidity levels or high temperatures. Do not expose to direct sunlight.
- Always switch off the instrument after use. In case the instrument is not to be used for a long time, remove the battery to avoid liquid leaks that could damage the instruments internal circuits.
REPLACING THE BATTERY
When the LCD display shows the symbol “ “, it is necessary to replace the
battery
CAUTION
Only expert technicians should perform this operation. Before carrying out
this operation, make sure you have removed all cables from the input terminals
or the cable being tested from inside the clamp jaw.
- Turn the rotary switch to the OFF position.
- Disconnect the cables from the input terminals and the cable being tested from the clamp jaw.
- Loosen the battery cover fastening screw and remove the cover.
- Disconnect the battery from the connector.
- Connect the new batteries to the connector (see § 6.1.2) and pay attention to correct polarity.
- Position the battery cover back over the compartment and fasten it with the relevant screw.
- Do not scatter old batteries into the environment. Use the relevant containers for battery disposal.
CLEANING THE INSTRUMENT
Use a soft and dry cloth to clean the instrument. Never use wet cloths,
solvents, water, etc.
END OF LIFE
CAUTION: this symbol found on the instrument indicates that the appliance, its accessories and the battery must be collected separately and correctly disposed of.
TECHNICAL SPECIFICATIONS
TECHNICAL CHARACTERISTICS
Accuracy is calculated as ± [% reading + (number of digits) x resolution]. It
is referred to a temperature 18°C 28°C with relative humidity <75% RH.
AC Voltage (Auto range)
Range| Resolution| Accuracy| Input impedance|
Bandwidth| Over voltage protection
---|---|---|---|---|---
4,000V| 0,001V| ±(1.8%rdg + 8digits)| 10MW| 50-400Hz| 600V
DC/ACrms
40.00V| 0.01V
400.0V| 0.1V
600V| 1V| ±(2.5%rdg + 8digits)
Integrated sensor for AC voltage detection: LED on for phase-earth voltage > 100V, 50/60Hz.
DC Voltage (Auto range)
Range| Resolution| Accuracy| Input impedance| Over
voltage protection
---|---|---|---|---
400.0mV| 0.1mV| ±(0.8%rdg+2digits)| 10MW| 600AACrms
4.000V| 0,001V| ±(1.5%rdg + 2digits)
40.00V| 0.01V
400.0V| 0.1V
600V| 1V| ±(2%rdg + 2digits)
AC Current
Range| Resolution| Accuracy| Bandwidth| Over voltage
protection
---|---|---|---|---
40.00A| 0.01A| ±(2.5%rdg + 8digits)| 50-60Hz| 400AACrms
400.0A| 0.1A| ±(2.8%rdg + 8digits)
Resistance and continuity test (Autorange)
Range| Resolution| Accuracy| Buzzer| Over voltage
protection
---|---|---|---|---
400.0W| 0.1W| ±(1.0%rdg + 4digits)| <30W| 600VDC/ACrms
4.000kW| 0.001kW| ****
±(1.5%rdg+2digits)
40.00kW| 0.01kW
400.0kW| 0.1kW
4.000MW| 0.001MW| ±(2.5%rdg+3digits)
40.00MW| 0.01MW| ±(3.5%rdg+5digits)
Continuity test current: < 0.5mA
Capacitance (Auto range)
Range | Resolution | Accuracy | Over voltage protection |
---|---|---|---|
40.00nF | 0.01nF | ±(4.0%rdg+20digits) | 600VDC/ACrms |
400.0nF | 0.1nF | ±(3%rdg+5digits) | |
4.000mF | 0.001mF | ||
4000mF | 0.01mF | ±(4.0%rdg+10digits) | |
100.0mF | 01mF |
Diode test
Range | Test current | Open-circuit voltage |
---|
| 0.3mA typical| 1.5VDC
Frequency with leads (Autorange)
Range| Resolution| Accuracy| Sensitivity| Over
voltage protection
---|---|---|---|---
10.00Hz ¸ 49.99Hz| 0.01Hz| ±(1.5%rdg+2digits)| ³15Vrms| 600VDC/ACrms
50.0Hz ¸ 499.9Hz| 0.1Hz
0.500kHz ¸ 4.999kHz| 0,001kHz
5.00kHz ¸ 10.0kHz| 0.01kHz
Duty cycle (Autorange)
Range | Resolution | Accuracy |
---|---|---|
0.5% ¸ 99.0% | 0.1% | ±(1.2%rdg + 2digits) |
100s Pulse width 100ms ; Pulse frequency: 100Hz 150kHz; Sensitivity >10Vrms
Temperature with K probe (Autorange)
Range| Resolution| *Accuracy ()| Over voltage
protection**
---|---|---|---
-20.0 ¸ 399°C| 0.1°C| ±(3%rdg+5°C)| 250VDC/ACrms
400 ¸ 760°C| 1°C
-4 ¸ 400°F| 0.1°F| ±(3%rdg+9°F)
400 ¸ 1400°F| 1°F
(*) Accuracy of K-probe not considered
Reference standards
- Safety: IEC/EN61010-1
- Insulation: double insulation
- Pollution level: 2
- Max operating altitude: 2000m (6562 ft)
- Over voltage category: CAT III 600V to ground
General characteristics
Mechanical characteristics
- Size (L x W x H): 200 x 66 x 37mm ; (8 x 3 x 1in)
- Weight (battery included): 205g (7 ounces)
- Max. cable diameter: 30mm (1in)
Power supply
Battery type: 2 batteries x 1.5V AAA
Low battery indication: The display shows the symbol “ “ when the voltage
supplied by the battery is too low.
Auto Power Off: after 30 minutes (may not be disabled)
Display
Characteristics: 4 dgt LCD max 4000 point, sign and decimal point
Sampling rate: 2 measurements per second
Conversion type: Mean value
ENVIRONMENT
Environmental conditions for use
- Reference temperature: 18°C 28°C ; (64°F 82°F)
- Operating temperature: 5 ÷ 40 °C ; (41°F 104°F)
- Allowable relative humidity: <80%RH
- Storage temperature: -20 ÷ 60 °C ; (-4°F 140°F)
- Storage humidity: <80%RH
This instrument satisfies the requirements of Low Voltage Directive 2006/95/EC (LVD) and of EMC Directive 2004/108/EC
ACCESSORIES PROVIDED
- Pair of 2mm test leads
- Adapter + K-type wire probe
- Carrying bag
- Batteries (not fitted)
- User manual
OPTIONAL ACCESSORIES
Model| **Description| Temperature range| Accuracy (at
100°C)| Probe length (mm)| Probe diameter (mm)
---|---|---|---|---|---
TK107| Air and gas temperature| -40 ¸ 800 °C| ± 2.2rdg| 200| 1.5
TK108| Internal temperature of semisolid substances| -40 ¸ 800 °C| ±
2.2rdg| 200| 3
TK109| Internal temperature of liquids| -40 ¸ 800 °C| ± 2.2rdg| 200| 4
TK110| Surface temperature| -40 ¸ 400 °C| ± 2.2rdg| 200| 5
TK111**| Surface temperature with fixed tip at 90°C| -40 ¸ 400 °C| ±
2.2rdg| 260| 5
SERVICE
WARRANTY CONDITIONS
This instrument is warranted against any material or manufacturing defect, in compliance with the general sales conditions. During the warranty period, defective parts may be replaced. However, the manufacturer reserves the right to repair or replace the product.
Should the instrument be returned to the After-sales Service or to a Dealer,
transport will be at the Customers charge. However, shipment will be agreed in
advance.
A report will always be enclosed to a shipment, stating the reasons for the
products return.
Only use original packaging for shipment; any damage due to the use of non-
original packaging material will be charged to the Customer.
The manufacturer declines any responsibility for injury to people or damage to
property.
The warranty shall not apply in the following cases:
- Repair and/or replacement of accessories and batteries (not covered by warranty).
- Repairs that may become necessary as a consequence of an incorrect use of the instrument or due to its use together with non-compatible appliances.
- Repairs that may become necessary as a consequence of improper packaging.
- Repairs which may become necessary as a consequence of interventions performed by unauthorized personnel.
- Modifications to the instrument performed without the manufacturers explicit authorization.
- Use not provided for in the instruments specifications or in the instruction manual.
The content of this manual cannot be reproduced in any form without the manufacturers authorization.
Our products are patented and our trademarks are registered. The manufacturer reserves the right to make changes in the specifications and prices if this is due to improvements in technology.
SERVICE
If the instrument does not operate properly, before contacting the After-sales Service, please check the conditions of batteries and cables and replace them, if necessary. Should the instrument still operate improperly, check that the product is operated according to the instructions given in this manual.
Should the instrument be returned to the After-sales Service or to a Dealer,
transport will be at the Customers charge. However, shipment will be agreed in
advance.
A report will always be enclosed to a shipment, stating the reasons for the
products return.
Only use original packaging for shipment; any damage due to the use of non-
original packaging material will be charged to the Customer.